Professional Documents
Culture Documents
HX-DxH Series
Service Manual
Installation Manual
HX-DxH Series
Service Manual
Installation Manual
Read this service & installation manual thoroughly before you use the product.
Keep this instruction manual for future reference.
Manual Notation
The number of this manual is entitled on the basis of the following standard:
X M X - Y Y Y Y - Z
(1) (2) (3) (4) (5)
(1) S : Service
O : Operation
(2) M : Manual
(3) K : Korean
E : English
(5) Z : Version
X-ray equipment, if not properly used, may cause injury. Accordingly, the
instructions herein should be thoroughly read and understood before
attempting to place this equipment in operation. We are glad to assist and
cooperate in placing this equipment in use.
In case of any changes in its user or location, ensure that this service & installation
manual is always kept together with the equipment. Periodically check to be sure
that the manual and the warning labels are not missing or damaged. If they are,
contact your DK Medical Systems service representative for replacement.
Notice
• All copyrights regarding this manual are property of DK Medical Systems.
Neither all nor part of these contents shall be reproduced or duplicated without the
express permission of DK Medical Systems.
• Content of this manual may be changed for improvement without prior notice.
Although every possible effort has been made to avoid errors while creating this
manual, immediate revision may not be possible in the event that errors or missing
information are detected.
• Screen images and illustrations contained in this manual may differ from those in
actual use, and are intended for example purposes only. Illustrations may also use
partial images.
DANGER
Indicates an imminently hazardous situation which, if not avoided, will result in serious
injury or death.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in serious
injury or possibly death.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor to
moderate injury or equipment damage.
NOTE
Emphasizes additional information that is provided to ensure the proper use of this
product.
Preface 8
Operating Precautions
(for Both the Safety and the Prevention of Danger)
on Using Electric Medical Equipment
1. Only personnel with the following experience and knowledge can use the equipment.
1) Medical radiographic training (if particular qualifications are required in the country
concerned, those qualifications must be held).
2) A capacity to read and understand the instruction manual.
2. When installing the equipment, pay attention to the following items:
1) Do not install it near water faucet or similar equipment.
2) Install it away from potential sources of problems such as abnormal pressure, temperature,
humidity, drafts, direct sunlight, dust, chlorine or sulphur gas.
3) During transportation and operation of the equipment, avoid tilting, vibration and sharp
impact against it.
4) Keep the equipment away from area where chemical or gas is stored.
5) Use only correct electrical power source with matching frequency, voltage and current
(or wattage).
6) Check the conditions of the battery power source (power and polarity) before operating the
equipment.
7) Properly ground the equipment.
8) When using this equipment, an operator shall be in an X-ray protection room.
3. Before operating the equipment, pay attention to the following items:
1) Check the conditions of switch contacts, polarity, dial settings, and meters, and make sure
the equipment performs smoothly.
2) Confirm that the ground is connected properly.
3) Check all wirings for proper and correct connections.
4) Pay attention when using more than one unit at a time, because it may lead to an incorrect
diagnosis and cause a hazard.
5) Check the condition of external electric circuit, which will be directly connected to a patient.
6) Check the condition of battery power source.
4. While operating the equipment, pay attention to the following items:
1) Do not exceed the time or the amount of equipment use needed for a diagnosis or a
therapy.
2) Observe the equipment and a patient continuously, for early detection of a problem.
3) When a problem is detected with the equipment or a patient, take proper actions to stop
the equipment without harming the patient.
4) Do not let a patient touch the equipment.
WARNING
· The responsibility of proper use and maintenance of the machine lies with a user.
· Only a diagnostic radiography technician and/or any personnel with a certificate
indicating equal proficiency can use the machine.
· Since an inspection and a repair of the inside of the machine are dangerous, if an error
occurs, make sure to put up a sign of defect of the machine. Then, contact your DK
Medical Systems service technician or a personnel with a certificate indicating equal
proficiency to repair the apparatus.
· In general, modifications of the machine are strictly prohibited by the Regulatory
Requirements of the Law of the country where the device is installed. Nevertheless, if
any modification should be made, consult it first with the manufacturer and/or seller.
Preface 10
Perform Periodic Inspection
Preventive maintenance is required to maintain safety and performance of this system for a
long time. As to the maintenance and inspection that specially trained experts exclusively
can perform, utilize the maintenance agreement offered by our company.
Wa r r a n t y
The warranty of this system is valid for one (1) year as from the date of purchase.
Please note that our company is not responsible for the following failures and damages:
Service after the expiration of warranty is available at reasonable cost and should be performed
by your DK Medical Systems service representative.
In some jurisdictions, some of the foregoing warranty disclaimers or damage limitations may not
be applied. DK Medical Systems will be indemnified for any claim, liability, or damage arising out
of the misuse or non-compliance with this manual by the purchaser or lessee of the system.
WARNING
WARNING
WARNING
Preface 12
WARNING
Repair and maintenance of the inside of the equipment can only be performed by
an engineer assigned by DK Medical Systems.
Maintenance must be assigned to specially trained expert. Contact your DK Medical
Systems service representative for repair and maintenance.
WARNING
To ensure the safety of equipment when using it, thoroughly read the manual
provided with the equipment. Also, fully understand all of precautions written in
this manual regarding the equipment use.
WARNING
If connecting this equipment to another one, make sure that another equipment is
certified by DK Medical Systems, does not cause degrading any system
performance and/or guarantees its safety.
CAUTION
Make sure that operator and patient can communicate while taking X-ray
irradiation.
If the equipment use can put a patient in danger, stop the examination or treatment
immediately.
DANGER
WARNING
Preface 14
WARNING
Do not put this equipment in a place where any liquid may exist, nor put such
liquid into the equipment. Avoid letting any liquid be spilled on the surface of the
equipment, nor enter into the equipment.
Otherwise, an electric shock may occur.
In case of liquid being entered inside of the equipment, turn off the power immediately,
and contact DK Medical Systems Service Representative.
WARNING
In case when any abnormality is found during the system operation, stop operating it
immediately and contact our service agents. Repair and inspection of the system need to
be conducted by them.
WARNING
WARNING
When moving the equipment, always be very careful to avoid contact with a
patient or an operator and ensure that they do not become caught between the
equipment and any neighboring devices.
Otherwise, it may cause injury.
Do NOT use the equipment in a place where metal fragments may enter.
Otherwise, it may result in an electric shock.
WARNING
Prior to conducting a study, check that the equipment is functioning correctly and all
system components and attachments are securely fastened.
WARNING
WARNING
Preface 16
NOTE
Cautions on Storage
CAUTION
WARNING
Be sure to turn the equipment power OFF before cleaning and disinfecting the
equipment.
Otherwise, malfunction may occur in the equipment, or the equipment may operate in
an unintended way. Thoroughly ventilate the room before turning ON the power after
disinfection work is completed.
CAUTION
Be sure to clean and disinfect the equipment periodically.
Cleaning and disinfection are very important to ensure that the equipment is used
hygienically and safely. Strictly follow the methods described in this manual.
CAUTION
Do NOT directly apply rubbing alcohol or water nor spray it onto the equipment.
Wipe off the surface of the equipment with a cloth soaked with rubbing alcohol.
If rubbing alcohol gets inside the equipment, it can cause failure or accidents.
CAUTION
Observe the following precautions when cleaning and disinfecting a touch panel:
• Turn the equipment power OFF before cleaning and disinfecting.
• Clean and disinfect the touch panel periodically.
• Wipe the surface of the equipment with a cloth soaked with disinfectant or cleaner.
If liquid gets inside the equipment, it can cause failure or accident.
• Do NOT rub or hit the panel surface because it can be scarred easily.
• Wipe the panel gently with moistened soft cloth to remove the dust from the surface.
Preface 18
CAUTION
CAUTION
CAUTION
On completing the work, check the following points before switching the power ON
again.
WARNING
When turning the power ON after cleaning, make sure the examination room is
properly ventilated.
Turning the power ON when any flammable gas remains in the examination room could
lead to fire, smoke, explosion or electrocution.
CAUTION
Preface 20
Cautions on Electromagnetic Compatibility (EMC)
WARNING
WARNING
WARNING
Do NOT use this equipment adjacent to, or stacked with other equipment.
If adjacent or stacked use is necessary, be sure that this equipment works properly in
such environment.
WARNING
Do NOT bring any cellular telephone or related device into the examination room
with their power ON.
Such devices can exceed the EMC standard limitations, and under some conditions it
can impair the proper functioning of the equipment. In the worst case, it can cause
serious injury or clinical error.
Cautions on Radiation
WARNING
Do not allow any personnel to use the equipment unless he/she is a qualified one (such
as a doctor, X-ray technician or X-ray engineer). The equipment can only be used under
a supervision of such qualified personnel.
WARNING
Preface 22
WARNING
During X-ray irradiation, ensure that the X-ray irradiates the necessary region only.
To avoid unnecessary exposure to X-ray radiation, narrow down the area to be shoot by
a collimator, and take protective measures like wearing a lead apron.
WARNING
Always check the X-ray exposure region by using the collimator lamp.
If irradiating outside of the required region of a patient, there may be a risk of exposing
the patient to unnecessary radiation.
WARNING
A user of the equipment is liable for a patient and/or user’s injury or wound caused from
an X-ray exposure. Since the manufacturer provides all necessary information
regarding the equipment and its possible risk if not being properly used, the
manufacturer is not liable for managing the equipment after its installation nor for safety
practices that a user does.
Also, it is emphasized that the manufacturer has no responsibility of maintenance
and/or services of any equipment if it is not maintained properly along with the
instruction, or is modified and/or changed.
In this manual, there are proper information including warning, caution and product
labels attached to the surface of the equipment. And the manufacturer does not have
any liability for a user’s equipment use or violence against the safety rules/
recommendations.
Also, the manufacturer is not liable for any patient or individuals over-exposed to X-ray
resulted from a user’s misuse of the equipment.
This X-ray unit may be dangerous to both patient and operator unless safe exposure
factor, operating instruction and maintenance schedule are observed.
CAUTION
Do not exceed the maximum tube operation limits shown in the X-Ray tube data of its
manufacturer. Intended life and reliability may not be obtained unless a generator is
operated within published specifications.
WARNING
To avoid the risk of electric shock, this equipment must be connected to supply mains
with protective ground.
CAUTION
In order to minimize the absorbed dose, make the distance to the focal spot as far as
possible (Minimum 45 cm).
Preface 24
CAUTION
CAUTION
CAUTION
Keep pressing the radiography preparation button (or the first stage of hand switch) and
the radiography button (or the second stage of hand switch) until exposure time is
completed. Releasing them before completing X-ray irradiation will stop the procedure in
the middle, and thus a suitable radiography image will not be acquired.
WARNING
Do not allow operation of this apparatus by any person other than qualified personnel
(physician, radiotherapy engineer and clinical X-ray engineer), or allow a person to use
this equipment under the supervision of qualified personnel.
WARNING
The equipment can be operated only by qualified personnel, such as radiography
technician or a person with equivalent qualification.
WARNING
No person but only a patient is allowed to stay in the examination room during X-
ray irradiation.
If the equipment is not used correctly, operator, patient, and others may receive a greater
dose of radiation than necessary. If someone has to be in the examination room with a
patient for some reason, then that person must take adequate measures to protect him/
herself against the radiation (e.g., wearing a protective apron, screen, etc.).
* During radiography, the radiography indicator illuminates and the buzzer rings for
audible warning.
WARNING
Perform X-ray irradiation carefully according to doctor's direction when using the
equipment with expectant mother, woman who is suspected as pregnant, lactating
woman, or children.
Particular ways of using the equipment may increase the scatter dose absorbed into the
patient, which may cause a radiation hazard.
Preface 26
WARNING
Always check the X-ray exposure region by using the collimator lamp.
If irradiating outside of the required region of a patient, there may be a risk of exposing
the patient to unnecessary radiation.
WARNING
During X-ray irradiation, ensure that the X-ray irradiates the necessary region only.
To avoid unnecessary exposure to X-ray radiation, narrow down the shooting area of
collimator and take protective measures, such as wearing a lead apron.
WARNING
CAUTION
Be sure to carry out aging (running-in operation of the X-ray tube unit) before
taking an X-ray radiography.
Refer to the ‘aging’ procedure described in the X-ray high voltage generator manual.
CAUTION
CAUTION
In order to minimize the radiation dose on a patient, make the distance between
the focus and the patient's body surface as far as possible (Minimum 45 cm).
The shorter the distance becomes, the greater the amount of scatter dose will be
absorbed into the patient, which may cause a radiation hazard.
CAUTION
Pay extra attention, when irradiating X-ray for a long time or repeatedly.
It may cause a radiation hazard.
Preface 28
Cautions on Operation
CAUTION
NOTE
CAUTION
Aging is effective to confirm that there is no failure in the X-ray tube unit as well as to
prevent electric discharge. In the event of abnormal operation, turn off the power of unit
and distribution board, and then contact your DK Medical Systems representative.
1. Select large focus and perform radiographic operation following the order described
below. Pause for 60 seconds between radiographic operations.
2. If an error occurs during aging, wait for a minute and reduce the tube voltage by 30
kV and perform radiographic operation. If the error keeps existing, contact to DK
Medical Systems.
3. For an X-ray tube unit with two or more focuses, make sure that they can be
changed during radiographic operation.
4. If there are two or more X-ray tube units, perform aging for each unit.
5. After three times of exposure, operator shall take time for cooling X-ray tube for five
minutes. The table below is the chart for X-ray tube Aging.
2nd Exposure three times with 100 kV/40mA with 60sec interval
Preface 30
Service Life
The equipment lifetime is 10 years (based on DK Medical Systems’ criteria) assuming the
specified maintenance checks are performed.
Disposal Precaution
CAUTION
Language(s) Provided
DK Medical Systems provides technical documents in English. Providing language(s) of
each country will be added at DOC and label and packaging chapter.
Software Revision
This manual applies to the software revision 1.0 ∗
Attenuation equivalent should not be greater than 1.6Vmm Al but thinner than 15mm
thickness. All parts of the cassette such as hinges, locks and marking device shall be
within the minimum radii for corners and edges. The flatness shall be 1mm warp or less.
WEEE Mark
With your co-operation, we are aiming to reduce contamination from waste electronic
and electrical equipment and preserve natural resource through re-use and recycling.
Do not hesitate to ask your DK Medical Systems service representative, if you require
further information.
Preface 32
Abbreviations
Abbr. Full Description Abbr. Full Description
Acryl Acryl MOT Motor
ATR Actuator Off-CNT Off-Centered
Base Base PTM Potentiometer
BR Bearing PRS Pressure
BK Bracket Rail Rail
CRG Carriage SFT Shaft
CNT Center SOL Solenoid
CHN Chain SPG Spring
Clutch Clutch SPT Sproket
CSH Cushion STD Stand
DTR Detector STP Stopper
EMG Emergency SW Switch
EXT Extend TB Timing Belt
FXD Fixed TP Top Plate
Lock Lock WIT Weight
Mcr SW Micro Switch Wire Wire
Options
The "Options" described in this manual may be provided as standard, or may be
unavailable, depending on its sales format.
Preface 34
Warning and Caution Labels
The following safety labels, which describe handling precautions, are attached to the
equipment. With adequate understanding of the contents on these labels and the warning
/caution items in this manual, operate the equipment safely. Inspect the safety labels
periodically (once a year).
If any label is peeled or unreadable by stain or scratch, replace it with a new one. For
new labels, contact your DK service representative.
Following Labels are attached as seen on the Figure below.
z 2
Preface 36
This page is intentionally blank.
Preface
Operating Precautions (for Both Safety and the Prevention of 9
Danger) in the Use of Electronic Medical Equipment
Warranty
Precautions in Use 12
Be Sure To Read the Following To Prevent Explosion, Electric 14
Shock, or Injury
Cautions on Storage 17
Cautions on Cleaning and Disinfection 18
Cautions on Electromagnetic Compatibility (EMC) 21
Cautions Relating to Cellular Telephones 22
Cautions on Radiation
Cautions on Radiography 26
Cautions on Operation 29
Cautions on X-ray Tube Aging 30
Service Life 31
Disposal Precaution
Language(s) Provided
Software Revision
Abbreviations 33
Options
Contents 38
01 Overview
1.1 Application 45
1.2 Contraindications 46
05 Technical Data
183
5.1 Product Specification – HX-D5H
185
5.2 Product Specification – HX-D6H
187
5.3 Exposure Conditions – HX-D5H
5.3.1 Small Focal Spot
188
5.3.2 Large Focal Spot
5.4 Exposure Conditions – HX-D6H 189
5.4.1 Small Focal Spot
5.4.2 Large Focal Spot 190
5.4.3 Time & mA Table 191
5.4.4 Information of Radiation Quantities 192
06 Daily Checklist
6.1 Daily Checklist 195
6.1.1 Starting Inspection Checklist
6.1.2 Ending Inspection Checklist 196
Contents 40
07 Appendix
7.1 Manufacturer Information 199
7.1.1 Regulatory Information
7.1.2 Company’s Quality System
7.1.3 International Standards
7.2 List of Components 200
7.2.1 Fuse List
7.2.2 Accessories 201
7.2.3 List of Cables
7.2.4 Capacity of Starting Capacitor 204
7.3 Attenuation Equivalent 205
7.3.1 Attenuation Equivalent Table
7.4 Connection Diagram 206
Contents 42
This page is intentionally blank.
1.1 Application
1.2 Contraindications
1.1 Application
1. Intended Use
The generator provides manual mode (kV, mA, ms), semi-automatic mode (kV, mAs),
automatic mode (AEC). It can store various APR (Anatomical Programming Radiograph)
by using highly integrated memory, and it can acquire an image with proper kV, mA, ms
values by an automatic control method without worrying about the subject by giving AEC
function. AEC irradiation method is an automatic control, which is convenient for a user,
and has more accurate value and time settings, so to obtain excellent image quality.
In order to use the product safely and reliably for a long period of time, you should
thoroughly familiarize yourself with the functions and operation of the product and read
the maintenance instructions of this manual before using this equipment. Special
attention should be paid to all warnings and cautions of this manual.
The manufacturer is not responsible for any damage to the equipment or a user’s injury
caused by an improper use or a failure in maintaining the equipment.
The responsibility of the device’s use and maintenance lies with a user.
1.2 Contraindications
The HX-D5(6)H should not be used if any of the following contra-indications exists.
CAUTION
1. Introduction 46
Chapter
The operation panel provides the function of setting the tube voltage and tube current
that are applied to the X-ray tube unit, as well as the irradiation time for performing
radiography.
2. Generator
The generator includes a power circuit. It controls the opening and closing of the main
circuit and manages the generation of high voltage to be applied to the X-ray tube
according to the tube voltage signal of the X-ray tube unit. The tube current and
irradiation time can be manipulated on the user panel.
The operating and the interface circuits are necessary ones for the external equipment
and the safety interlock, and these circuits are provided in the operation generator.
3. Essential Performance
Item Specification
Sex No limitation
Nationality No limitation
1. Introduction 48
Chapter
1. Introduction 50
Chapter
Use the equipment under the following ambient conditions in an examination room or
operating room.
· Temperature : 5 to 35°C
· Humidity : 45 to 85 % (Non-condensing)
· Air Pressure : 800 to 1060 hPa
· Environment Luminosity : 150 to 500 lx
· Ambient Noise Level : Under 70 dB
CAUTION
Avoid sudden changes in temperature and humidity, even if such changes are made
within the above range. Otherwise, it can cause condensation that may damage the
equipment. Condensation inside the equipment can cause rust and corrosion. Frost
resulted from low temperatures can damage internal circuits.
CAUTION
DANGER
Do not use a power supply other than the one specified below.
Using a power supply other than the one specified below may cause serious accident
such as a fire, smoke emission or an explosion.
• Frequency: 50/60 Hz
• Standard voltage and permitted voltage range and power-supply impedance under
non-load conditions.
Standard Permitted
Power Supply Impedance
Voltage Voltage Range
1Ø, 220 VAC ± 10 % 0.16 Ω max
• Relationship between cable length and cross-sectional area to match the transformer
capacity.
• Transformer capacity: 40kVA (single phase)
• Frequency: 50/60 Hz
• Standard voltage and permitted voltage range and power-supply impedance under
non-load conditions.
Standard Permitted
Power Supply Impedance
Voltage Voltage Range
3Ø, 380V ± 10 % 0.16 Ω max
• Relationship between cable length and cross-sectional area to match the transformer
capacity.
• Transformer capacity: 50 kVA (3-phase)
1. Introduction 52
• The relationship between cable length and cross-sectional area corresponds to the
transformer capacity.
HX-D5H HX-D6H
Length
40kW 50kW
Below 10m 16㎟ 5.5 ㎟
20m 38 ㎟ 5.5 ㎟
30m 38 ㎟ 5.5 ㎟
40m 60 ㎟ 5.5 ㎟
50m 60 ㎟ 8㎟
60m 60 ㎟ 8㎟
70m 80 ㎟ 8㎟
80m 100 ㎟ 14 ㎟
90m 100 ㎟ 14 ㎟
100m 100 ㎟ 14 ㎟
• Safety Devices
CAUTION
If an earth leakage breaker is used for either power voltage, use an inverter type of earth
leakage breaker to prevent malfunction in high-frequency circuits.
Transport and store the equipment under the following ambient conditions:
· Temperature : -10 to 60 °C
· Humidity : 10 to 95 % (Non-condensing)
· Air Pressure : 800 to 1060 hPa
CAUTION
1. Introduction 54
This page is intentionally blank.
2.1 Rating
2.1 Rating
1. HX-D5H
HX-D5H HX-D6H
40kW 50kW
E7239X; 125kV, 320mA
/80kV, 500mA
(E7239X) 125kV, 320mA E7884,X;150kV, 320mA
Rate (E7239X)
80kV, 500mA /100kV, 500mA
RAD-14; 150kV, 320mA
/100kV, 500mA
Tube Voltage
40kV ~ 125kV (1 step) 40 ~ 150kV (1 step)
Adjustment
Tube Current
10mA ~ 500mA 10 ~ 630mA
Adjustment
NOTE
Nominal max. electric power is the max. tube current allowable to flow for 0.1 sec at the
tube voltage of 100kV multiplied by the tube voltage.
NOTE
The above-specified rating is for X-ray high-voltage generators and is limited by the X-
ray tube ratings when an X-ray tube is used with lower ratings .
Starting voltage for each X-ray tube used for HX-DxHis as follows.
(Refer to the following chart on Starting Capacitor’s Capacity.)
Frequency 50/60Hz
NOTE
When the above specification is not fulfilled, the apparatus function and rating may not
be guaranteed.
Safety Device
Safety Device
Rating of Fuse or
Model Voltage
Type Capacity Circuit Breaker
NOTE
If an earth leakage breaker is to be installed, be sure to use it for inverter which prevents
malfunction of high frequency circuit.
4. Power Consumption
Power Consumption
NOTE
If a length of cable exceeds the above-specified value, the apparatus function and
ratings may not be guaranteed.
<Top view>
unit: mm
<Front view> <Side view>
<Top view>
<Top view>
<Top view>
unit: mm
<Front view> <Side view>
<Top view>
<Top view>
Power Cabinet
400 kCAL/h (170 kCAL/h) 600 kcal/h (170 kcal/h)
High-Voltage Generator
NOTE
These calorific values are for general operating condition. They can differ depending on
actual operating conditions. Values specified in parentheses above are calorific values
of the X-ray tube unit which are included those of the power cabinet and the high-
voltage generator. Values for the console do not include X-ray TV’s calorific value.
Also, note that 1kW equals to 860kcal/h.
3. Environmental Conditions
When the environmental conditions specified below are not met in an installation
room, It is recommended to install a dedicated air-conditioning unit in the room.
Environmental conditions required for the room are as follows:
Atmospheric
Rooms Temperature Humidity Precaution
Pressure
Examination Room 5 to 35°C 45 to 85% 800 to 1060hPa
No dew condensation is
allowed.
Operating Room 5 to 35°C 45 to 85% 800 to 1060hPa
Humidity 10 to 95%
DANGER
Prior to connecting cables, be sure to switch the power OFF and set a knife switch or a
circuit breaker on the power equipment in order to open it.
1) First, connect a grounding cable from the distribution panel to the cabinet. Use
washers to prevent loosening.
2) Be sure to firmly connect the shielding wire of the power cable to the grounding
terminal strip on the cabinet side.
3) When connecting the input power cable, proceed it after removing the fuse from
the socket. (Refer to the picture below.)
A Fuse
A A
1) Connect the earth cable to the 2) Remove the Fuse holder cover
(HX-D5H).
Grounding terminal.
220Vac 0Vac
240Vac 110Vac
2) When it is 240Vac
220Vac 0Vac
240Vac 110Vac
When the input power supply of generator is 240 Vac, change the transformer input
tab with 240 Vac and connect it.
1) This equipment uses a down voltage transformer, and the terminal to be connected
varies depending on the input voltage. This down voltage transformer uses a
different model, depending on the generator model.
2) Connect to 0V and 220Vac terminals when input power is 220Vac. The factory setting
is 220 Vac.
3) When input power is 240Vac input, connect the cable connected to 240Vac terminal
at the factory to 220Vac.
4) When input power is 110Vac input, connect the cable connected to 110Vac terminal
at the factory to 220Vac.
220Vac 0Vac
240Vac 110Vac
Fig. 2.8.
1) This equipment uses a down voltage transformer, and the terminal to be connected
varies depending on the input voltage. This down voltage transformer uses a
different model, depending on the generator model.
2) Connect to 0V and 380VAC terminals when input power is 380Vac. The factory
setting is 380 Vac.
3) When input power is 3phase 400VA input, connect the cable connected to 400Vac
terminal at the factory to 380Vac.
4) When input power is 3phase 220VA input, connect the cable connected to 220Vac
terminal at the factory to 380Vac.
380Vac 0Vac
400Vac 220Vac
Fig. 2.9.
IGBT connection
Resistor connection
Input Filter
1) Connect the connector of OP console to the CN14 on the Main Control Board.
2) OP Console Cable Length: 10 m
3) OP Console Connector Pin: 15 Pin
Fig. 2.13.
Fig. 2.14.
2
1
Remove eight screws from the rear Remove the hand switch connector
of OP console. of J16.
Fig. 2.16.
CAUTION
1) Make sure that the receptacle socket assembly is clean and correctly mounted on
an X-ray tube and a generator according to IEC 60526 or NEMA XR 7.
2) Ensure that the termination plug and mounting accessories are clean.
3) Loosen the Grub Screw and slip the Ring Nut over the sleeve.
4) Place the Divided Flange onto the Ring Flange.
5) Wet the Silicone Washer (both sides) and the Rubber Gasket with the Silicone Oil.
6) Fit the Rubber Gasket onto the Plug against the Ring Flange.
7) Push the Silicone Washer over the contact pins.
8) Insert the Termination Plug into the Receptacle Socket (watch the Nipple on the
Plug to ensure correct positioning of the Contact pins.)
9) Hand tighten the Ring Nut.
10) Tighten the Grub Screw.
* IMPORTANT: Always replace the Silicone Washer by a new one in case the Ring
Nut is loose.
Fig. 2.17.
1 2
Fig. 2.18.
CAUTION
* Change the packing annually. However, be sure to replace it whenever the cable head
is detached, for repair or inspection.
* Be careful with an electric shock when disconnecting the high-voltage cable.
1. PCB Information
It shows the Components -Rotor IF Board and Main Board.
PC
AEC #1
OP
AEC #2
X-ray
Debug
#1
Rotor
IF
Cap
FB
Lock +
Lamp N Lock -
Lamp P
0Vac
110Vac
THR -
Main
THR +
M3 S
M2 M
M1 C
Cap
Bank
Magnet
AUX
AUX
Door
SUB
SW Magnet
Collimator(AC0V)
Collimator(AC24V)
0Vac (FAN)
110Vac (FAN)
Thermal (TH1)
Thermal (TH1)
◎SHIFT (M3)
◎ MAIN (M2)
◎ COM (M1)
GND
Case GND
Cap
FB
Lock +
Lamp N
Lock -
Lamp P
0Vac
110Vac
CAUTION
Bucky1 Grid
4 RET+ RET- 5
Bucky PWR Buck1 DET Bucky2 Grid
Buck2
1 110Vac 0Vac 3 RET+ RET- RET+ RET-
DET
RDY +
RDY -
1 RDY+ RDY- 2 RDY+ RDY-
110Vac RET +
Bucky1 Grid Bucky2 Grid
0Vac RET -
220Vac RDY+ RDY- Bucky +
25Vac Buck1 DET Bucky -
0Vac
Cap
FB
Lock +
Lamp N
Lock -
Lamp P
0Vac
110Vac
CAUTION
Bucky2 Grid
4 RET+ RET- 5
Bucky PWR Bucky1 Grid Buck1 DET
Buck2
110Vac 0Vac RET+ RET- RET+ RET-
DET
RDY +
RDY -
RDY+ RDY- RDY+ RDY-
110Vac RET +
Bucky1 Grid Buck1 DET
0Vac RET -
220Vac 1 RDY+ RDY- 2 Bucky + 3
25Vac Bucky2 Grid
Bucky - 1
0Vac
Cap
FB
Lock +
Lamp N
Lock -
Lamp P
0Vac
110Vac
CAUTION
Cap
FB
Lock +
Lamp N
Lock -
Lamp P
0Vac
110Vac
CAUTION
Fig. 2.26.
Fig. 2.27.
ⓑ Stand
Buck1 DET Bucky2 Grid
RET+ RET- RET+ RET-
Buck2
DET
1 RDY +
Bucky2 Grid
3 RET +
4 RET -
ⓑ Stand
Bucky2 Grid
Fig. 2.31. Rotor IF PCB Ass’y
AEC 1 (Table)
AEC 2
DAP
3 RS 485-B RxD/TxD-N
4 RS 485-A RxD/TxD-P
CAUTION
Connect the power supply cables to the terminals on the output side of the fuse holder.
Fasten the cable JS1 earth lead to the fixing screw at the top-left side of the TABLE
Main board.
TP19 : mA FB
TP7 : kV FB
GND
Fig. 2.34.
Fig. 2.33.
AC220V
INPUT 2.5kW
CAPACITOR BANK
CB-100 &
CB-200
DC 340V
OUTPUT
AC 220V
OUTPUT
CB-100
DC400V/ 10000uF*14EA
CB-200
DC400V/ 10000uF*36EA
Cap Bank
HX-D6H
Fig. 2.38.
D. Connection Diagram
The Cap Bank is connected to the generator in the below green box.
Fig.2.40.
F. How to connect to the Generator
Cap Bank
Generator
60A CB(+)
60A CB(-)
Fig.2.41.
Fig.2.42.
⑤ Connect the 3,500mm AC power input line to the 30A breaker and SUB FUSE as shown
below.
⑥ F.G(Frame Ground) is connected to the frame fixing the Main Board in CB-200, and to FG
frame terminal block with SMPS in AccurayD5.
Fig.2.43.
Fig.2.44.
15
8 16
17
9 18
13
1 10 11 12 3
14 2
Fig. 2.45.
14 Save APR Address Stores APR exposure condition to the APR address.
Enables adjustment of an exposure condition for a
15 Patient Size Selection
patient’s body size.
16 AEC Field Selection Enables selecting AEC field type and using photo timer.
5
4 6
1 2
Fig. 2.46.
ⓓ Factory Reset;
⇒ 4-1 : System Reset
⇒ 4-2 : CAL Data Reset
⇒ 4-3: AEC Data Reset
⇒ 4-4: APR Reset
⇒ 4-5 :1,2,3,4 All Reset
≫ Factory Reset; If Del button is pressed for 2~3 seconds, it will be reset with
a beep sound.
Fig. 2.47.
4) After entering 4.1 using ( ) button, press the TAB for settings.
5) Save DATA after clicking Export / Import. For your information, Export /Import can be
executed only in the 4.1 mode.
Fig. 2.48.
1. Press the ( ) right button of power switch to turn the power on.
Fig. 2.49.
2 3
Fig. 2.41.
3. When the stand-by LED is blinking upon turning on the generator, the Firmware will
be displayed on the window. (Showing Pr. 106 ; as of Mar 2018)
4. Before Calibration, HX-D5Hdata needs to be Imported. (Refer to 2.8.3)
1 2 3 4 5 6 7
No. Model Tube kV Limit SmA/LmA Limit Fix mA DAP Size Cont
0 Small Large All mA None -
6 8D 200 mA 250 mA 6
7 7
8 8
9 9
5. In 3.5, it is possible to set whether you use the whole mA or only a fixed one, in case
when you press the mA up/down buttons during the use of equipment for hours.
8 9 A B C D E F
Non Grid Delay Cap Ban
No. Bucky 1 Bucky 2 Aux AEC1 AEC 2
Bucky Time k
0 None None None None 10 msec None None None
Cap Bank
1 Film Film Sol Film Sol Film Sol 20 msec NonBucky NonBucky
En
2 DR Ch1 Film Mot Film Mot Film Mot 30 msec Bucky1 Bucky1
3 DR Ch2 DR Sol Ch1 DR Sol Ch1 DR Sol Ch1 40 msec Bucky2 Bucky2
7 250ms
8
9 500ms
Enter 1.1 by pressing the ( ) button and move to 1.1, 1.2, 1.3, and 1.4 by pressing
up/down button ( ).
Fig. 2.51.
>>TUBE Calibration<<
- kV : 40, 60, 80, 100, 120kV
- Small mA (6 steps): 40, 50, 63, 80, 100, 125
Large mA (6 steps): 160, 200, 250, 320, 400, 500
Fig. 2.52.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on
window of OP console, with the measured one. By using the value of 1.1 and 1.2
conditions, adjust the value displayed on the OP console to the same as
measured one.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on
window of OP console, with the measured one. By using the value of 1.1 and 1.3
conditions, adjust the value displayed on the OP console to the same as
measured one.
• Set the condition 80kV~120kV, 40mA, 100ms by pressing the kV key. Using the
1.1, 1.3 condition values only, adjust the one displayed on OP console to the same
as measured one.
Fig. 2.53.
• After conducting Ready -> Exposure of X-ray, use only 1.4 and 1.2 conditions and
compare the value displayed on the OP console, with the measured one. Adjust the
one on displayed OP console to the same as measured value. (Do not adjust the
1.1 condition value, since the mean value (address 1.1) in the range of kV small
focus was already set in 40mA step. Since the kV value does not fluctuate
significantly in other mA steps, the CAL can be carried out as the 1.4 value.)
• After conducting Ready -> Exposure of X-ray, compare the value displayed on
window of OP console with the measured one. Use the value 1.4, 1.3 conditions
only when adjusting the value displayed on the OP console to the same as
measured one.
• Set the condition 80kV~120kV, 50mA, 100ms by pressing the kV key. Use the
value 1.4, 1.3 conditions only when adjusting the value displayed on the OP
console to the same as measured one.
Fig. 2.54.
• Press the mA UP button and set the kV and mA, after activating mA. Conduct
the calibration by increasing mA 40, 50, 63, 80, 100, 125 in order – total six
conditions.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on
the OP console window with the measured one. Use only 1.1, 1.7 condition
values to adjust the one displayed on OP console to the same as measured one.
• Press the kV DOWN button to set the condition as 40kV 160mA 100ms.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on
the OP console window with the measured one. Use only the 1.1, 1.3 condition
values to adjust the one displayed on OP console to the measured value with
same value.
• After setting 80kV~120kV, 160mA, 100ms condition, Using only the 1.1, 1.3
condition values, adjust the one displayed on OP console to the same as the
measured one.
Fig. 2.55.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on the
OP console window with the measured one. By using only 1.4, 1.2 condition
values, adjust the one displayed on OP console to the same with the measured
one.
• However, do not adjust the value under 1.1 condition, because the mean value
(address 1.1) of kV Large focus in 160mA step was set. Since the kV value does
not fluctuate significantly in other mA steps, the CAL can be carried out by using
the 1.4 value.
• After conducting Ready -> Exposure of X-ray, compare the value displayed on the
OP console window with the measured one. By using only 1.4, 1.3 condition
values, adjust the one displayed on OP console to the same with the measured
one.
• After setting 80kV~120kV, 200mA, and 100ms condition, Use the value 1.4, 1.3
conditions only when adjusting the value displayed on the OP console to the same
as measured one.
Fig. 2.56.
• Calibrate by increasing mA as 160, 200, 250, 320, 400, and 500 in order – total
six conditions.
• There are lots of X-ray dose in Large Focus steps, so calibration should be
proceeded with enough time.
• Calibration values are divided by two sections which are Small FOCUS and
Large FOCUS. Therefore, it should be calibrated by kV mean value (1.1)
separately.
※ Since 120kV 320mA / 100kV 400mA / 80kV 500mA are the maximum output
conditions of Equipment, calibration should be proceeded with setting ms as
50ms.
1. Connect the CH1 of Oscilloscope to TP7 (kV FB) of Main control board.
2. Connect the CH2 of Oscilloscope to TP19 (mA FB) of Main control board.
3. Oscilloscope setting
a. kV: CH1 VDC Level is 1V per 50kV
b. mA: CH2 VDC Level is 1V per 10mA (In case of Small filament)
CH2 VDC Level is 1V per 100mA (In case of Large filament)
Fig. 2.59.
* To save the value, keep press the ‘Save’ button for 3 seconds.
No. Model Tube kV Limit SmA/LmA Limit Fix mA DAP Size Cont
* To save the value, keep press the ‘Save’ button for 3 seconds
1. Small Filament
a. Connect the CH1 of Oscilloscope to TP7 (kV FB) of Main control board.
b. Connect the CH2 of Oscilloscope to TP19 (mA FB) of Main control board.
c. Set the X-ray condition as 80kV, 100mA and 100ms on Console.
d. Exposure X-ray and measure the mA value.
e. If it need to adjust the mA value, user have to change the value of
1.6 then re-exposure x-ray again to match to 100mA.
2. Large Filament
a. Connect the CH1 of Oscilloscope to TP7 (kV FB) of Main control board.
b. Connect the CH2 of Oscilloscope to TP19 (mA FB) of Main control board.
c. Set the X-ray condition as 80kV, 200mA and 100ms on Console.
d. Exposure X-ray and measure the mA value.
e. If it need to adjust the mA value, Engineer have to change the value
of 1.7 then re-exposure x-ray again to match to 200mA.
* To save the value, keep press the ‘Save’ button for 3 seconds.
Fig. 2.60.
1. Small Filament
a. Connect the CH1 of Oscilloscope to TP7 (kV FB) of Main control board.
b. Connect the CH2 of Oscilloscope to TP19 (mA FB) of Main control board.
c. Set the X-ray condition as 40kV, 40mA and 100ms.
d. Exposure X-ray at 40kV with 40, 50, 63, 80, 100, 125 mA.
e. If kV and mA value is whether higher or lower, adjust the value of
“1.3 kV_mA Ref.” to match the mA value of each mA. And adjust the
value of “1.4 mA_kV Ref.” to match the kV value of 40kV.
f. By same procedure, check and adjust the kV and mA as below chart.
* To save the Value, keep press the ‘Save’ button for 3 seconds.
mA
40mA 50mA 63mA 80mA 100mA 125mA
kVp
1.3 1.3 1.3 1.3 1.3 1.3
40kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
60kV
1.1 1.4 1.4 1.4 1.4 1.4
1.6 1.2 1.2 1.2 1.2 1.2
80kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
100kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
120kV
1.1 1.4 1.4 1.4 1.4 1.4
* Remarks; The above table shows the number of addresses to be used for each section.
( ) Number: Calibration Order
Yellow box: When setting Small mA mean value, proceed it first.
mA
160mA 200mA 250mA 320mA 400mA 500mA
kVp
1.3 1.3 1.3 1.3 1.3 1.3
40kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
60kV
1.1 1.4 1.4 1.4 1.4 1.4
1.7 1.2 1.2 1.2 1.2 1.2
80kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
100kV
1.1 1.4 1.4 1.4 1.4 1.4
1.3 1.3 1.3 1.3 1.3 1.3
120kV
1.1 1.4 1.4 1.4 1.4 1.4
* Remarks; The above table shows the number of addresses to be used for each section.
( ) Number: Calibration Order
Yellow box: When setting Large mA mean value, proceed it first.
Fig. 2.61.
2. The value of L,M,H are saved as 88 (1 ~ 99). (For Vacutec chamber saved as 88.)
3. Density Value : -90 ~ +99
4. To change AEC ref Value (1 ~ 99), 2mm Copper (21mm Al) needs to be used.
SID 100
80kV 320mA
It changes value of L = 0 to permit 12msec.
M = 0 is 8msec.
H = 0 is 6msec.
Fig. 2.62.
SID 100
80kV 320mA
It changes value of L = 0 to permit 12msec.
M = 0 is 8msec.
H = 0 is 6msec.
Pull out the wire and connect it to GX-Mini D by using the bottom part of the console.
ⓐ Generator Settings
ⓑ Calibration CAL
ⓒ Chamber (AEC) Settings
ⓓ Use PC
For USB
Communication
1) Adjust the Mean value of kVp per each Small Focus, Large Focus.
2) Adjust the Mean value of mA per each Small Focus, Large Focus.
3) If the value is not matched even though with adjusting mA Mean value, apply to
only the relevant condition value with using in case of Fine adjustment.
4) If the value is not matched even though with adjusting kVp Mean value, apply to
only the relevant condition value with using in case of Fine adjustment.
5) Adjust mA value with adjusting the waveform of Oscilloscope.
6) Using at adjusting Small Focus mA Mean value.
7) Using at adjusting Large Focus mA Mean value.
1 5 7 9
C
6
A B
2 3 4
Fig. 2.64.
Fig. 2.65.
1. Set condition 80kVp 40mA 100ms. (Adjust Mean value at 80kVp)
2. Confirm the value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.6.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.66.
1. Set Condition 40kVp 40mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.68.
1. Set Condition 100kVp 40mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.70.
1. Set condition 80kVp 50mA 100ms. (Adjust Mean value at 80kVp)
2. Confirm the value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.2.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.72.
1. Set condition 60kVp 50mA 100ms. (Adjust Mean value at 80kVp)
2. Confirm the value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.74.
1. Set Condition 120kVp 50mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.76.
Fig. 2.78.
1. Set Condition 100kVp 63mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.80.
Fig. 2.82.
1. Set Condition 60kVp 80mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.84.
Fig. 2.86.
Fig. 2.88.
1. Set Condition 100kVp 100mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.90.
Fig. 2.92.
1. Set Condition 60kVp 125mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.94.
1. Set Condition 120kVp 125mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Button click
→ Change Focus
Fig. 2.95.
1. Set condition 80kVp 160mA 100ms. (Adjust Mean value at 80kVp)
2. Confirm the value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.7.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.96.
1. Set Condition 40kVp 160mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.98.
1. Set Condition 100kVp 160mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.1.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.100.
Fig. 2.102.
Fig. 2.104.
Fig. 2.106.
1. Set Condition 40kVp 250mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.108.
Fig. 2.110.
1. Set Condition 80kVp 320mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.2.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.112.
Fig. 2.114.
1. Set Condition 120kVp 320mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.116.
1. Set Condition 40kVp 400mA 100ms.
2. Confirm the Value with using Exposure.
3. If kVp is not matched, use compensation value at Address 1.4.
4. If mA is not matched, use compensation value at Address 1.3.
5. Match the Value on instrument using the relevant address and Value on Console window.
Fig. 2.118.
Fig. 2.120.
Fig. 2.121.
Fig. 2.122.
Fig. 2.123.
Fig. 2.124.
Fig. 2.125.
Fig. 2.126.
Fig. 2.127.
SMPS
(SNP-Y041)
Filament
PCB Ass’y
Magnetic ROTOR
Contactor LC2 Capacitor
High Voltage
Tank
SMPS
(SNP-Y11T)
Sub-EMI Filter
Discharging
Charging Resistor
Resistor
Magnetic
Contactor LC2
Main Transformer
Magnetic
Contactor LC1
Main Fuse
High Voltage Tank
IGBT Drive
Rotor Capacitor
PCB Ass’y
Rotor Capacitor
Magnetic
Contactor LC1/LC2
Main Transformer
HV Tank
AEC 1 (Table)
AEC 2
Fig. 2.135.
NOTE
The installation screw etc. must not touch an internal substrate. Otherwise, it may cause
a breakdown. When driving it into the wall, the length of part of it that will go through
outside a wall must be shorter than 9mm.
One bucky and one detector can be connected to Rotor & Interface PCB Ass’y.
Refer to the figure below. In our system, Table bucky is set as Bucky 1.
1. Hand Switch
2. Hand Switch Holder
3. Hand Switch Bracket
4. Screw M3 x 8 (2 pcs)
5. Screw M4 x 10 (2 pcs)
Fig. 2.138.
NOTE
First, mount the hand-switch bracket (A) to the control console with the screws specified
as no. 4 in the above.
2
3
5
1
4
No. Name
1 Hand Switch Bracket
2 Hand Switch Holder
3 Screw M3x8 (2 pcs)
4 Screw M4 x 10 (2 pcs)
5 Hand Switch
Hand Switch
(Right Side Type)
Hand Switch
(Left Side Type)
Fig. 2.140.
2 1
Fig. 2.141.
CAUTION
When the fuse is broken, replace it with a new one. It must be the same type and rating.
1. Transformer T1-40
Ref. No Rating
<Fast Acting>
5. SA CONT - 2002 (Option)
Ref. No Rating
F-1 20A 250V
F-2 20A 600V
6. XAT – 2H
Ref. No Rating
F-UN 10A 250V
F-VN 10A 250V
F-WN 10A 250V
3.2 Troubleshooting
E.no 22 H/W +kV Level Error Measured kV value is +20% than the set +kV value.
E.no 23 H/W–kV Level Error Measured kV value is -20% than the set -kV value.
* Remarks;
3. Troubleshooting 168
Chapter
3.2 Troubleshooting
Error No. Code Name Solution
1. Check if the exposure switch is interrupted in the middle of
actual exposure.
2. Check the TP29 (FP_SET) and TP9 (GND) with an oscilloscope.
If the checked value is lower than the value set in the OP,
C.no 01 Exposure Fail
replace the Main_CTRL.
3. Check TP23 and TP25 (GND) with an oscilloscope. If there is no
saw tooth waveform coming from the Ion Chamber, replace the
main board or Ion Chamber.
1. After checking 3-D and 3-E in the CAL Mode, change them
according to the settings.
2. After checking the status of cable connection in the Ion
Chamber, adjust it.
C.no 02 AEC Fail
3. Check TP23 and TP25 (GND) with an oscilloscope. If there is an
input signal in the Ion Chamber, replace the Control Board.
4. Check TP23 and TP25 (GND) with an oscilloscope. If there is no
Ion Chamber input signal, replace the Ion Chamber.
1. Check the status of detector wiring connection.
2. Check whether Rotor board CN20 (RET+,RET-) or CN23
(RET+,RET-) is under short circuit condition in the ADE mode.
Detector It yes, correct it.
C.no 03
Required 3. If there is a contact signal of detector in the ADE mode, replace
the rotor board.
4. If there is no contact signal of detector in the ADE mode, replace
the detector.
1. Check the status of bucky wiring connection.
2. Check whether a rotor board CN19 (RET+,RET-) or CN20
C.no 04 Grid Required
(RET+,RET-) is under short circuit condition. If yes, correct it.
3. If the above are all normal, replace the rotor board.
1. Check the connection status in CN17 (DOOR_SW) on the rotor
C.no 05 Door Open board.
2. If the door contact signal is normal, replace the rotor board.
1. Check the status of wiring connection in CN11 on the rotor
board. If it has a problem, correct it. (Regarding its temperature,
Tube
refer to the Tube Data.)
C.no 06 Temperature
2. Check the temperature sensor in the X-ray tube.
Over
3. If the temperature sensor in the X-ray tube is normal, replace the
rotor board.
1. Check the activation of Cap Bank in the CAL code.
2. Check the status of wiring connection in CN13 on the rotor
Cap-bank
C.no 07 board.
Check
2, Check the status of contact point in Cap Bank.
3. If the above are normal, replace the rotor board.
1. Check whether an operator presses the ready button for longer
X-ray Key than ten seconds. Then, train the operator.
C.no 08
Input Check 2. Check the status of hand switch connection.
3. If the above are normal, replace the OP console.
EP, IP Set 1. When IC U2 and U29 are lost in the main board, replace the
E.no 09
Error board.
3. Troubleshooting 170
Error No. Code Name Solution
2. Check the current between TP15 and TP9 (GND) on main CTRL
board with an oscilloscope. If the measured current has 20% of
difference compared with the established current value, replace
EXPO Fil FB the main CTRL board.
E.no 17
Level Error 3. Check the current between TP13 and TP9 on the main CTRL
board with an oscilloscope; if the set current is lower than 100mA,
then 1V is 10mA. If greater than 100mA, 1V is 100mA.
If abnormal, replace the filament board.
4. If the problem still occurs, contact DK Medical Systems service
representative.
1. Check the voltage in TP7 of the main board with an oscilloscope.
(1V=50kV)
2. If the voltage in TP7 shows more than 20% of difference, check the
EXPO kV FB voltage in TP29 and TP9 whether the set voltage is normal.
E.no 18 (ex: Check whether the input voltage is 1V at 50kV setting. If not,
Level Error
replace the control board.)
3. If the voltage in TP7 has more than 20% of difference under the
normal input power, it is speculated that the HV tank has a
problem.
1. Check the current of TP19 on the main board with an oscilloscope;
if it is lower than the 100mA which is the set current, 1V is10mA. If
greater than 100mA, 1V is100mA. If the measured current has 20%
EXPO mA of difference compared with the established current value, replace
E.no 19 FB Level the main board.
Error 2. Check the current between TP15 and TP9(GND). If the measured
current value is different from the set value, replace the main
board. If it is normal, it is speculated that the HV tank has a
problem.
1. Check the TP7 on the main board with an oscilloscope. if the
H/W kV FB feedback voltage is greater than 3.2V, proceed the calibration.
E.no 20
Max Over 2. After conducting the calibration, if the same abnormality continues,
replace the main board.
1. Check TP19 on the main board with an oscilloscope, if the
feedback voltage at 100mA setting is greater than 8V, proceed the
H/W mA FB calibration.
E.no 21
Max Over
2. After calibration, if the same symptom continues, replace the main
board.
H/W +kV 1. Compare the voltage between TP6, TP8 and TP9 (GND) with
E.no 22
Level Error oscilloscope ch1 and ch2. If two waveforms are not in asymmetry,
H/W -kV it means the HV tank might have a problem.
E.no 23
Level Error 2. If they are in symmetry, replace the control board.
* Remarks;
3. Troubleshooting 172
Chapter
2 APR LED does not turn on. Replace the Membrane sheet.
Mechanism
3 Mini console sheet is broken. Replace the mini console sheet.
Open the Oil Cap bolt on the top of
4 Oil Leakage in HV Tank
HV Tank.
Exposure not available beyond 95 kV
5 Bad IGBT & HV Tank
(at Small /Large)
6 Error E07 when exposing at 320mA Replace the Rotor B/D.
Error E07 at 90kVp 320mA 320ms,
7 Change the Main board (ver. 1.03)
OP console has a malfunction.
Bad contact in Magnet switch.
8 ACCURAY-D5 does not turn ON.
Replace the magnet switch.
Reset the Charging Limit of Main
9 Capacitor Bank is not charged.
B/D in Capacitor Bank.
Calibrated it due to 30% tolerance of
10 CB-200 Cap bank Error happened.
mA.
11 Error E.no 25 happened. Calibrate & set the lower limit.
23 Bad X-Ray exposure with mini console Change the setting value.
No sound or small beep tone during
24 Replace the Mini Console board.
mini console exposure
25 Mini OP Console malfunction Re-install the program.
Replace the firmware v1.0 -> v1.1
26 No light on OP Console kV LED
on ROTOR B/D.
No light on 100 unit LED in OP console Replace the OP console LED
27
LED Ass'y.
Error E.no 19 on specific APR address
28 Replace the F-RAM.
in OP Console
29 X-Ray Key Input Error Matching Code
30 Different X-Ray Dosage X-Ray Calibration
Change the Console position and
31 Different Time in X-Ray
replace IGBT cable.
37 System gets turned down randomly. Repair the Cable from CB-200.
Less conditioned image is acquired
38 Change the generator setting.
randomly.
Replace the board of abnormal
39 Abnormal exposure condition randomly.
condition.
The ‘Ready’ is activated under the
40 Replace the Ready S/W.
Stand-by mode.
Randomly, the exposure process does
41 Replace the Ready exposure S/W.
not work after the READY status.
Unable to hear the Exposure beep tone
42 Replace the Mini Console.
randomly from the Mini console.
3. Troubleshooting 174
Range No. Symptom Solution
1. Tube Rotor activates when the
Replace the Rotor Board & Main
46 power is on.
Board.
2. Less conditioned image is acquired.
Firmware upgrade on Main B/D
Exposure condition does not fit &
47 (V1.05->V1.06). Install a new version
An image is not acquired sometimes.
of XCG.
Different exposure condition is
48 Replace the Main B/D & Calibrate it.
applied, or an image is not acquired.
54 OP Console is not turned ON. Replace the Rotor & Interface board.
Mini Console does not communicate wi Change the Port settings of Device
Commu- 58
th PC. Manager.
nication
Replace the main B/D and connect
The communication between ACC-D5
59 the communication line (Change Ver.
and Elui is cut off randomly.
1.01 -> 1.06)
Upgrade main B/D and connect RS-2
60 Communication is cut off randomly.
32c. Proceed calibration.
62 Mini OP Console is cut off sometimes. Change Main board F/W V1.3>>V1.4.
The daily inspection consists of the starting and ending inspections. Daily inspection is
necessary in order to use the system for a long time. Check whether all inspections
and correct actions have been completed, before using the system.
Refer to the “Starting Inspection Checklist” and “Ending Inspection Checklist” in 6.1
“Daily Checklist”. We recommend that you retain filled out copy of the “Starting
Inspection Checklist” and “Ending Inspection Checklist” for a certain period of time.
1. Turn off the power of HV generator, and cut down the electricity.
2. Conduct a visual inspection on the cable connection among component devices.
3. If any rust is found in chemical materials, increase the interval of cleaning component
devices. Use a cloth damped with warm, soapy water for sanitation. Just swipe a cover
or an enclosure of device. Do not use any solvent type of cleaner.
Be sure to turn OFF the system power before cleaning and disinfecting the
system.
Otherwise, a malfunction may occur in the system, or the system may operate in an
unintended way.
WARNING
WARNING
WARNING
Before turning on the power, thoroughly ventilate the room after cleaning a
device, in order to prevent it from accidents.
Flammable gases accumulated in the room may ignite, smoke, explode, or cause an
electric shock when the power is turned on.
WARNING
WARNING
Periodic inspection mainly affects the equipment performance and its internal
mechanism. The inspection requires a deep understanding about the internal
mechanisms which can be dangerous for an untrained personnel. Contact your DK
Medical Systems service representative if you wish to request a periodic inspection.
It is recommended to conduct periodic inspection every six (6) months. A fee is charged
for periodic inspection after a warranty is expired.
Item Description
4.2.1 Fuse
Manufacturer
No. Component Part No. Technical data
/ Type
Generator
1 Main Fuse (F1, F2, F3) 600V, 60A Buss (JT60060)
500V, 10A Cylindrical
2 Magnetic Fuse (F4, F5, F6) Legrand
type (10x38mm)
Generator: Rotor I/F Board
3 Fuse (F1, F4, F5, F6, F7) 250V, 10A (20mm) Glass type
Hand Switch
kV 40 to 125 kV (E7239X)
10 to 500 mA (E7239X)
mA Small : 10, 12.5, 16, 20, 25, 32, 40, 50 ,64, 80, 100, 125mA
Large : 200, 250, 320, 400, 500mA
Short Time Rating Single phase : E7239X: 40kW (125kV, 320mA / 80 kV, 500mA)
Coefficient of
CV (Coefficient Value) ≤ 0.05
Reproducibility
Operation Panel 3 kg
Control Cabinet 82 kg
Nominal Min.
1ms
AEC Exposure Time
Radiography
mAs Limit 500mAs
mA Small : 10, 12.5, 16, 20, 25, 32, 40, 50 ,64, 80, 100, 125,
160mA
Large : 200, 250, 320, 400, 500, 630mA
Setting mAs 0.1 to 500 mAs
Range
0.001 to 10 sec
Operation Panel 3 kg
Control Cabinet 84 kg
Nominal Min.
AEC 1ms
Exposure Time
Radiography
mAs Limit 500mAs
kV mA E7239 kV mA E7239
40 10 40 10
50 8 50 8
63 6.3 63 6.3
40 80 5 90 80 4
100 4 100 2.5
125 3.2 125 1.6
160 2.5 160 0.63
40 10 40 10
50 8 50 8
63 6.3 63 6.3
50 80 5 100 80 2.5
100 4 100 2
125 3.2 125 0.8
160 2.5 160 0.4
40 10 40 10
50 8 50 8
63 6.3 63 5
60 80 5 110 80 2
100 4 100 1.25
125 3.2 125 0.4
160 2.5 160 0.1
40 10 40 10
50 8 50 6.3
63 6.3 63 4
70 80 5 120 80 1.25
100 4 100 0.25
125 3.2 125 0.12
160 1.6 160 0.1
40 10 40 10
50 8 50 6.3
63 6.3 63 4
80 80 5 125 80 1.25
100 3.2 100 0.25
125 3.2 125 0.12
160 1 160 0.1
kV mA E7239 kV mA E7239
125 2.5 125 1.6
160 2.5 160 1.6
200 2 200 1.6
250 1.6 250 0.63
40 100
320 0.5 320 0.2
400 0.5 400 0.1
500 X 500 X
630 X 630 X
125 2.5 125 1.25
160 2.5 160 1.25
200 2 200 1.25
250 1.6 250 0.5
50 110
320 0.5 320 0.16
400 0.5 400 X
500 X 500 X
630 X 630 X
125 2.5 125 0.8
160 2.5 160 0.8
200 2 200 0.8
250 1.6 250 0.32
60 120
320 0.5 320 0.1
400 0.5 400 X
500 X 500 X
630 X 630 X
125 2.5 125 0.8
160 2.5 160 0.8
200 2 200 0.8
250 1.6 250 0.32
70 125
320 0.5 320 0.1
400 0.4 400 X
500 0.16 500 X
630 X 630 X
125 2 125 X
160 2 160 X
200 2 200 X
250 1.6 250 X
80 150
320 0.4 320 X
400 0.25 400 X
500 0.1 500 X
630 X 630 X
125 2
160 2
200 2
250 1
90
320 0.32
400 0.16
500 X
630 X
* Abbreviations
FSD: Focal Source Distance DAP: Dose Area Product
ESD: Entrance Surface Dose EDQ: Effective Dose Quantities
CAUTION
The contents of table above were calculated and prepared by various theories, and data
that is actually calculated might be different slightly, depending on a clinical use. Consider
the above table as data on assessing radiation risks.
Always carry out the inspection described below, prior to using the system.
1. Check the following points visually before switching the power ON. Take proper
actions to correct a problem.
2. Check visually that connections have been made correctly, before switching the
power ON. Correct a nonconformity, if any.
Check Point Action
□ Are all cables connected correctly? Connect them correctly.
3. Turn the power ON. Check the following points visually and aurally. Correct a
nonconformity, if any.
Check Point Action
□ Does the display flicker? Contact our service agent.
□ Is there any unusual smell? Turn the equipment’s breaker OFF,
□ Are all operation-related noises normal? and contact the manufacturer.
Always check the following points, before switching the power OFF. Correct a
problem, if any.
1. Check the following points visually before switching the power OFF. Correct a
nonconformity, if any.
2. Switch the power OFF. Check the following points visually. Correct a nonconformity,
if any.
The company's quality system satisfies the Annex II exclusion section of Annex IV of the
Directive 93/42/EEC as amended by 2007/42EC, which is certified by the Institute for
DNV NEMKO PRESAFE AS.
Manufacturer
No. Component Part No. Technical data EA
/ type
Generator
500V, 10A
2 Magnetic Fuse (F4, F5, F6) Cylindrical type Legrand 3
(10x38mm)
3 Fuse (F1, F4, F5, F6, F7) 250V, 10A (20mm) Glass type 4
5 Fuse (F4, F5, F6, F7, F8, F9) 250V, 2A (30mm) Glass type 6
CAUTION
To avoid causing a fire, it is recommend to use a product that is the same or equivalent
to the above-specified specification, when replacing a product.
7. Appendix 200
7.2.2 Accessories
10 SMPS_Input 1.3m
7. Appendix 202
This page is intentionally blank.
Use a cable specified on the above list; if not, it may occur an error and/or obstacle
against operation. Also it may decrease immunity of the system.
Starting voltage for each X-ray tube used for HX-D5H is as follows. (Refer to the
following chart on Starting Capacitor’s Capacity.)
7. Appendix 204
Chapter
Maximum Value of
No. Component Part Model
Attenuation Equivalent
1 Patient Table TE-90 1.0 mmAl
2 Patient Table TF-90 1.0 mmAl
3 Integration Table ST-90D 1.0 mmAl
4 Detector Stand DS-90 0.5 mmAl
Fig. 7.1.
CAUTION
If an operator does not follow the Attenuation Equivalent of items for the X-ray Beam,
which is in accordance with sub-clause 203.10.101 of IEC 60601-2-54, it may cause
excessive X-ray exposure to the patient. The table 203.104 of Attenuation Equivalent of
items in the X-ray Beam is shown above.
7. Appendix 206
Fig. 7.3. Connection Diagram for HX-D6H