BC-5300 Preventive Kit Manual

1. Part List

FRU
Name Part Number Part Name Qty.
Number
3102-20-69178 probe wipe (6 month) 2EA
BC-5300 3001-20-07263 washer for front bath (6 month) 4EA
801-3101- Preventive 3001-20-18695 washer for rear bath (6 month) 4EA
00062-00 Maintenanc
3001-10-07054 Air Filter (6 month) 10EA
e Kit
082-000027-00 filter assembly (12 month) 1EA
115-002439-00 isolation chamber (contain filter) (12 month) 3EA

2. Part Replacement

2.1 Probe Wipe

2.1.1 Purpose And Interval

In case of contamination or leaking ,replace the probe wipe every 6 months.

2.1.2 Preparation

1. Power off the analyzer if it is running;

2. Turn off the main power switch on the right side of the analyzer and unplug the

power cord;

3. Remove the right side door;

4. Move the sampling module to the back of the analyzer;


2.1.3 Removal

1、Remove the probe wipe from the sampling module (Figure 1)

a) Disconnect the two tubes from the probe wipe;

b) Pull the wipe downward to remove the probe wipe, as indicated by the arrow

in the figure below, and remove the wipe clip;

c) Separate the wipe from sampling module in the direction of arrow;

d) Get out the inlet and outlet pipe from the wipe。

Figure 1 Removing probe wipe

1 ―――pipe clamp 2 ―――presser under the guiding rod


3 ―――probe wipe 4 ―――wipe clip

2、Installation
Repeat the above “removal” step in the reverse order。

2.2 Washers For Front And Rear Baths

2.2.1 Purpose And Interval

In case of leaking caused by ageing, replace the washers every 6 months

2.2.2 Preparation

1、Power on the analyzer to empty the fluidic system. Power off the analyzer and

unplug the power cord from the wall outlet;

2、Remove the left and right doors ,and the top cover;

3、Remove the two screws on the top cover of data board shielding box, and then

remove the top cover.(Figure 2)

4、Disconnect the bath signal lines of the WBC reaction bath or RBC reaction bath

that needs to be removed, and then loosen all buckles fastening the lines in the direction
of wiring ,until the wire is drawn to the bath.

5、Remove the screw on bath shield and take it out, then remove the shield.(Figure

3)

6、Remove the two screws on bath presser and take it out, then remove the bath

(Figure 3)
presser;

7、Pull the bath out with a distance of about 50-80 mm. Disconnect all tubes

connected to the bath. Get the signal wire from wire-protective ring and then remove the
bath assembly.
Figure 2 Removing signal lines of bath

1 ―――M3X6 cross-headed panhead 2 ―――top cover of data board shielding


washer screw box
3 ―――data board

Figure 3 Removing bath assembly

1 ―――shielding plate 2 ―――WBC bath


3 ―――bath presser 4 ―――M3X8 small panhead screw
5 ―――RBC bath 6 ―――shield cover
7 ―――isolation chamber clamp 8 ―――isolation chamber

2.2.3 Replacement

Separate the front bath from the rear and replace the washers . The concave side of

the aperture must face to the front bath。(Figure 4)

Figure 4 Replacement of the washers

2.3 Air Filter

2.3.1 Purpose And Interval

In case that volumetric tube could not drain ,replace air filter every 6 months.

2.3.2 Preparation

1、Remove the left and right side doors;


2、Position revolves pressure pump、V25、V26(Figure 5)

Figure 5 Positions Of Air Filters

2.3.3 Replacement

Separate the wipes from the filters ,and replace them。

2.4 Filter Assembly

2.4.1 Purpose And Interval

In case of clog of probe wipe and P2 waste pump ,replace filter assembly every 12
months.

2.4.2 Preparation

1、Power off the analyzer;

2、Remove the right side door.

2.4.3 Replacement

1、Separate the wipe from filter assembly in the direction of arrow(Figure 6);

2、Remove the screw;

3、Replace the filter assembly.


Figure 6 Removal Of Filter Assembly

2.5 Isolation Chamber

2.5.1 Purpose And Interval

In case that the bath could not drain or clog of P2 waste pump ,replace the isolation
chamber every 12 months.

2.5.2 Preparation

1、Power on the analyzer to empty the fluidic system. Power off the analyzer;

2、Remove the right side door;

(Figure 7)
3、Position revolves DIFF bath、WBC bath and RBC bath。
Figure 7 Positions Of the Isolation Chambers

2.5.3 Replacement

1、Separate the isolation chamber assembly from the clamp;


2、Disconnect all pipes connected to the isolation chamber(Figure 8)

3、Replace the chamber.

Figure 8 Removal Of Isolation Chamber