You are on page 1of 2

Calibration certificate 2920625

Device: FT10-ES
Serial number: P451447
Date of calibration: 14.10.2020
Technician ID: 46
Calibration method: NKM-FT10CS-PK-V1.0

Calibration method
The device to be calibrated is connected to the reference device with a dedicated cable harness. A computer
software with a script is used to control both devices to adjust their readings to match at Pt100 temperatures -30
and 300 °C. The device to be calibrated stores the resulting values in its EEPROM memory. Then the device and
the reference readings are compared at several points. This calibration does not include the sensor nor its
inaccuracy.

Calibration standards used


Description Device
Pt100 calibrator Nokeval Ac2002 A157256

Calibrated ranges
Rref

Pt100-4w input

Adjust point Device reading Reference reading Offset


-30 °C -30.000 -30.000 0.000
300 °C 300.000 300.000 0.000

1/2
Calibration certificate 2920625

Pt100-4w input

Test point Device reading Reference reading Offset


-100 °C -100.004 -100.000 0.004
-50 °C -50.005 -50.000 0.005
0 °C 0.017 0.000 -0.017
50 °C 49.989 50.000 0.011
100 °C 100.011 100.000 -0.011
150 °C 150.037 150.000 -0.037
200 °C 199.994 200.000 0.006
250 °C 249.969 250.000 0.031
300 °C 300.017 300.000 -0.017
350 °C 349.972 350.000 0.028
400 °C 399.944 400.000 0.056
450 °C 449.959 450.000 0.041
500 °C 499.949 500.000 0.051
550 °C 549.953 550.000 0.047
600 °C 599.939 600.000 0.061

Pt1000-4w input

Test point Device reading Reference reading Offset


0 °C -0.18277 0.00000 0.18277