Professional Documents
Culture Documents
20
MONITORING VENTILATOR
FUNCTION
Richard D. Branson, RRT
From the Division of Trauma and Critical Care, Department of Surgery, University of
Cincinnati, Cincinnati, Ohio
Support Features
Monitoring Features
Clinician-set features
Positive pressure tidal breaths Essential Essential Essential
Mandatory rate Essential Essential Essential
Flow or l:E or inspiratory time Essential Recommended Recommendedt
Expiratory pressure (PEEP) Essential Essential Optional
Foo2tto 1 Essential Essential Optional
Patient-interactive features
Patient spontaneous breath Essential Recommended Optional
(e.g., CPAP, IMV)
Breath-triggering mechanism Essential Recommended Recommendedt
(flow or pressure sensors to
initiate a ventilator breath)
Flow-timing interaction (e.g., Recommended Optional Optional
pressure support)
Feedback control
(e.g., mandatory minute Optional Optional Optional
ventilation)
*Essential = considered to be necessary for safe and effective operation in the majority of patients
in the specified setting; recommended = considered to be necessary for optimal management of
virtually all patients in the specified setting ; optional = considered to be possibly useful in limited
situations but not necessary for a majority of patients in the specified setting.
tF002 = oxygen concentration delivered by device; F002 = F102 when patient demand (inspiratory
flow rate) is met.
:j:Essential feature if patient has intact ventilatory drive and respiratory muscles or possibility of
partial or complete ventilator independence is anticipated.
From AARC-ARCF Consensus Conference: Consensus statement on mechanical ventilators. Respir
Care 37:1000-1008, 1992; with permission.
Pressure
Ppeak Essential Essential Essential
P mean Essential Optional Optional
pplateau Essential Optional Optional
PEEP (set) Essential Essential Optional::f:
Intrinsic PEEP (auto-PEEP) Recommended Optional Optional
Volume
Vr expired machine Essential Recommended Optional
VE machine Essential Optional Optional
Vr expired spontaneous Essential Recommended Optional
VE spontaneous Essential Optional Optional
Vr inspired spontaneous Recommended Optional Optional
Timing
Flow mechanical Recommended Optional Optional
Flow spontaneous Optional Optional Optional
l:E ratio Essential Recommended Optional
Rate mechanical Essential Recommended Optional
Rate spontaneous Essential Recommended Optional
Gas Concentration
Foo2§ Essential Optional::f: Optional::f:
Lung Mechanics
Effective compliance Optional Optional Optional
lnspiratory airways resistance Optional Optional Optional
Expiratory airways resistance Optional Optional Optional
Maximal inspiratory pressure Optional Optional Optional
Circuit Characteristics
Tubing compliance Recommended Optional Optional
*Essential = considered to be necessary for safe and effective operation in the majority of patients
in the specified setting; recommended = considered to be necessary for optimal management of
virtually all patients in the specified setting; optional = considered to be possibly useful in limited
situations but not necessary for a majority of patients in the specified setting.
tMonitors need not be integral part of ventilator.
tEssential if feature is used on a specific patient.
§F 002 = oxygen concentration delivered by device; F002 = F102 when patient demand (inspiratory
flow rate) is met.
VT = tidal volume ; VE = set minute ventilation.
From AARC-ARCF Consensus Conference: Consensus statement on mechanical ventilators. Respir
Care 37:1000-1008, 1992; with permission.
Alarm Features
Level 1
Power failure (including when battery in Electrical control system*
use)
Absence of gas delivery (apnea) Circuit pressures,* circuit flows, timing
monitor, carbon dioxide analysis
Loss of gas source Pneumatic control system*
Excessive gas delivery Circuit pressures,* circuit flows , timing
monitor
Exhalation valve failure Circuit pressures, circuit flows , timing
monitor
Timing failure Circuit pressures, circuit flows , timing
monitor
Level 2
Battery power loss (not in use) Electrical control system*
Circuit leak* Circuit pressures,* circuit flows
Blender failure F102 sensor
Circuit partially occluded Circuit pressures, circuit flows
Heater/humidifier failure Temperature probe in circuit
Loss of/or excessive positive end- Circuit pressures
expiratory pressure
Autocycling Circuit pressures, circuit flows
Other electrical or preventive subsystem Electrical and pneumatic systems monitor
out of limits without immediate overt
gas delivery effects
... Level 3
Change in central nervous system drive Circuit pressures, circuit flows , timing
monitor
Change in impedances Circuit pressures, circuit flows, timing
monitor
Intrinsic positive end-expiratory pressure Circuit pressures, circuit flows
(auto) >5 cm H20
*Alarms currently defined in the International Standards Organization and ASTM 5 standards.
From AARC-ARCF Consensus Conference: Consensus statement on mechanical ventilators . Respir
Care 37 :1000-1008, 1992; with permission.
Home or Skilled
Level:j: Critical Care Transport Nursing Facility
*Essential = considered to be necessary for safe and effective operation in the majority of patients
in the specified setting; recommended = considered to be necessary for optimal management of
virtually all patients in the specified setting; optional = considered to be possibly useful in limited
situations but not necessary for a majority of patients in the specified setting.
tAlarms need not be integral components of the ventilator.
:j:Levels are defined in Table 4.
§Redundancy is required only if ventilator is providing total support.
From AARC-ARCF Consensus Conference: Consensus statement on mechanical ventilators. Respir
Care 37:1000-1008, 1992; with permission.
Airway Pressure
With every patient-ventilator system check, airway pressures
should be observed and recorded. This includes peak inspiratory pres-
sure (PIP), PEEP, mean airway pressure (Paw), plateau pressure (Ppiat),
and auto-PEEP. If a combination of modes is used, such as intermittent
mandatory ventilation (IMV) and PSV, the PIP of mandatory and sup-
ported breaths should be measured and recorded.
During volume control ventilation (VCV), PIP provides a rough
estimate of lung compliance. 51 At a given inspiratory flow (V1), tidal
volume (VT), and PEEP, PIP increases as compliance and resistance
increase. Elevated PIP may be the result of a kink in the endotracheal
tube or ventilator circuit, secretions in the airway, bronchospasm, the
patient fighting the ventilator (dys-synchrony), or a true reduction in
lung compliance.
Because elevated PIP often is associated with barotrauma and may
cause further lung injury, mimicking adult respiratory distress syndrome
(ARDS), a change in PIP of greater than 8 to 10 cm H 2 0 or 15% of a
previously reported value should be investigated actively.42 If a mechani-
cal problem (e.g., secretions, kinked tube) cannot be found, the physician
should be notified. Trending of values is important, but any single value
should be assessed in light of a group of values obtained over the
preceding 6 to 8 hours.
Reductions in PIP may be caused by technical problems, including
leaks in the circuit, leaks in the humidifier, or an improper tracheal tube
cuff seal. The RCP should begin evaluation of a low PIP at the patient,
working back toward the ventilator, looking for sources of a leak. Low
gas inlet pressures or disconnection of source gases also can lead to low
PIP. If the delivery system is secure, a reduction in PIP may be the result
of improved compliance or reduced airway resistance.
Ppiat is measured by adding an inspiratory pause of 1 to 2 seconds
such that a static pressure can be measured at zero flow. 51 Most ventila-
tors are capable of making P plat measurements if the RCP sets the appro-
priate pause time. In addition, P plat should be measured on an unassisted
breath using a constant inspiratory flow. Patient effort during a Pplat
measurement can cause the true pressure to be underestimated.
P plat is thought to provide the closest approximation of peak alveolar
pressure. Recent evidence from laboratory studies suggests that a peak
136 BRAN SON
r Exp.
PIP
Expiratory
l
Hold
Figure 1. Volume , flow , and pressure waveforms during an expiratory hold maneuver to
measure auto-PEEP. PEEPE = set PEEP; PEEP, = intrinsic or auto-PEEP ; PEEPT =
total PEEP.
138 BRANSON
Alarms
Proper alarm settings are important for patient safety and to reduce
false alarms. Frequently, the alarm section of the patient-ventilator sys-
tem check is relegated to a cursory position; the only entry concerning
alarms is a check mark or the designation that alarms are "on."
Current microprocessor-run ventilators have a never-ending array
of alarms. When appropriate, these alarms should be set and the settings
recorded. Of particular importance are low- and high-pressure alarms
during VCV and low- and high-volume alarms during pressure control
ventilation.
140 BRANSON
Miscellaneous
Other variables that should be recorded include the mode of ventila-
tory support, breath sounds, quality and quantity of respiratory secre-
tions, endotracheal tube cuff pressure, position of the endotracheal tube,
and a brief narrative of pertinent observations. All entries should be
signed by the responsible RCP.
Ancillary Equipment
inspect the circuit for soiling, and change the circuit as required by
hospital policy. At the University of Cincinnati, we have successfully
gone to every-7-day ventilator circuit changes without a significant
change in the incidence of nosocomial pneumonia.
References
2. Accidental breathing circuit disconnections in the critical care setting. HHS publication
no. FDA 90-4233. Rockville, MD, FDA, 1990
3. Al-Saady N, Bennett ED: Decelerating inspiratory flow waveform improves lung
mechanics and gas exchange in patients on intermittent positive-pressure ventilation.
Intensive Care Med 11:68-75, 1985
4. American Association for Respiratory Care: Clinical Practice Guideline: Patient-
ventilator system checks. Respir Care 37:882-886, 1992
5. American Na tional Standards Institute. Standard for humidifiers and nebulizers for
medical use. ANSI Z79:9, 1979
6. Bartel LP, Bazik JR, Pawner DJ: Compression volume during mechanical ventilation:
Comparison of ventilators and tubing circuits. Crit Care Med 13:851-854, 1985
7. Bedsole SC, Kempf J: More faulty Bear check valves. Respir Care 29:1159, 1984
8. Benson MS, Pierson DJ: Auto-PEEP during mechanical ventilation of adults. Respir
Care 33:557-568, 1988
9. Berman LS, Downs JB, Van Eeden J: Inspiration :expiration ratio: Is mean airway
pressure the difference? Crit Care Med 9:775-777, 1981
10. Branson RD: Enhanced capabilities of current ICU ventilators: Do they really benefit
patients? Respir Care 36:362-376, 1991
11. Branson RD: Humidification of inspired gases during mechanical ventilation. RT
3:55-66, 1991
12. Branson RD: Flow triggering systems. Respir Care 39:138-144, 1994
13. Branson RD, Davis K Jr, Campbell RS, et al: Humidification in the intensive care unit; a
prospective study of a new protocol utilizing heated humidification and a hygroscopic
condenser humidifier. Chest 104:1800-1805, 1993
14. Branson RD, Hurst JM: Laboratory evaluation of moisture output of seven airway heat
and moisture exchangers. Respir Care 32:741-747, 1987
15. Brown DG, Pierson DJ: Auto-PEEP is common in mechanically ventilated patients: A
study of incidence, severity, and detection. Respir Care 31:1069-1074, 1986
16. Burton JDK: Effects of dry anaesthetic gases on the respiratory mucous membrane.
Lancet 1:235, 1962
17. Campbell RS: Managing the patient- ventilator system: System checks and circuit
changes. Respir Care 39:227- 236, 1994
18. Chalan J, Markham JP, Ali MM, et al: The Pall Ultipor breathing circuit filter: An
efficient heat and moisture exchanger. Anaesth Analg 63:566-570, 1984
19. Chamney AR: Humidification requirements and techniques. Anaesthesia 24:602, 1969
20. Chatburn RL: Physiologic and methodologic issues regarding humidity therapy. J
Pediatr 114:416-420, 1989
21. Chatburn RL, Primiano FP Jr: A rational basis for humidity therapy. Respir Care
32:249-253, 1987
22. Ciszek TA, Modanlou HD, Owings D: Mean airway pressure-significance during
mechanical ventilation in neonates. J Pediatr 99:121-126, 1981
23. Cropp AJ, Woods LA, Raney D, et al: Name that tone. The proliferation of alarms in
the intensive care unit. Chest 105:1217-1220, 1994
24. Dahlby RW, Hogg JC: Effect of breathing dry air on structure and function of airways.
J Appl Physiol 61 :312-317, 1980
25. Department of Health and Social Security: Evaluation of heated humidifiers. Health
Equipment Information 177, 1987
26. Dreyfuss D, Djedaini K, Weber P: Prospective study of nosocomial pneumonia and of
patient and circuit colonization during mechanical ventilation with circuit changes
every 48 hours versus no change. Arn Rev Respir Dis 143:738-743, 1991
27. Emergency Care Research Institute: Heated humidifiers. Health Devices 16:741-747,
1987
28. Flynn JT, Bancalari E, Snyder ES: A cohort study of transcutaneous oxygen tension
and the incidence and severity of retinopathy of prematurity. N Engl J Med 326:1050-
1054, 1992
29. Forbat AF, Her C: Correction for gas compression in mechanical ventilators. Anesth
Analg 59:488-493, 1980
30. Gallagher TJ, Banner MJ: Mean airway pressure as a determinant of oxygenat.on
(abstract). Crit Care Med 8:A244, 1980
MONITORING VENTILATOR FUNCTION 143
31. Gedeon A, Mebius C: The hygroscopic condenser humidifier: A new device for general
use in anaesthesia and intensive care. Anaesthesia 34:1043-1047, 1979
32. Hess D, McCurdy S, Simmons M: Compression volume in adult ventilator circuits:
A comparison of five disposable circuits and a nondisposable circuit. Respir Care
36:1113-1118, 1991
33. Jenner W, George BG: Oxygen-air shunt syndrome strikes again. Respir Care 27:604,
1982
34. Leigh JM, White MG: A new condenser humidifier (letter). Anaesthesia 39:492-493,
1984
35. Marini JJ: Monitoring during mechanical ventilation. Clin Chest Med 9:73-100, 1988
36. Marini JJ, Ravenscraft SA: Mean airway pressure: Physiologic determinants and clinical
importance. Part 2: Clinical implications. Crit Care Med 20:1604-1616, 1992
37. Mebius CL: A comparative evaluation of disposable humidifiers. Acta Anaesthesiol
Scand 27:403-409, 1983
38. Moore FA, Haenel JB, Moore EE: Auto-PEEP in the multisystem injured patient: An
elusive complication. J Trauma 30:1316- 1323, 1990
39. Moritz AR, Weisiger JR: Effects of cold air on the air passages and lungs. Arch Intern
Med 75:233-240, 1945
40. Ogino M, Kopotic R, Mannino FL: Moisture-conserving efficiency of condenser humidi-
fiers. Anaesthesia 40:990-995, 1985
41. Operators Manual, Puritan-Bennett 7200a. Carlsbad, CA, Puritan-Bennett Corpora-
tion, 1992
42. Parker JC, Hernandez LA, Peevy KJ: Mechanisms of ventilator-induced lung injury.
Crit Care Med 21:131-143, 1993
43. Pepe PE, Marini JJ: Occult positive end-expiratory pressure in mechanically ventilated
patients with airflow obstruction: The auto-PEEP effect. Am Rev Respir Dis 126:166--
170, 1982
44. Revenas B, Lindholm CE: The foam nose-new disposable heat and moisture ex-
changer: A comparison with other similar devices. Acta Anaesthesiol Scand 23:34-39,
1979
45. Salyer JW, Chatburn RL: Patterns of practice in neonatal and pediatric respiratory
care. Respir Care 35:879-888, 1990
46. Scott LR, Benson MS, Bishop MJ: Relationship of endotracheal tube size to auto-PEEP
at high minute ventilation. Respir Care 31:1080-1082, 1986
47. Shelly MP, Bethune OW, Latimer RD: A comparison of fr.re heat and moisture ex-
changers. Anaesthesia 41:527-532, 1986
48. Slee TA, Paulin EG: Failure of low pressure alarm associated with the use of a
humidifier. Anesthesiology 69:791-793, 1988
49. Smith TC, Marini JJ: Impact of PEEP on lung mechanics and work of breathing in
severe airflow obstruction. J Appl Physiol 65:1488-1499, 1988
50. Society of Critical Care Medicine, Task Force on Guidelines: Guidelines for standards
of care for patients with acute respiratory failure on mechanical ventilatory support.
Crit Care Med 19:275--278, 1991
51. Tobin MJ: Monitoring of pressure, flow, and volume during mechanical ventilation.
Respir Care 37:1081-1096, 1992
52. Yang KL, Tobin MJ: A prospective study of indexes predicting the outcomes of trials
of weaning from mechanical ventilation. N Engl J Med 324:1445-1450, 1991
53. Ziecheck HD: Faulty ventilator check valves cause pipeline gas contamination. Respir
Care 26:1009-1010, 1981
20. Craven DE, Steger KA. Pathogenesis ciency virus, hepatitis B virus, and
and prevention of nosocomial pneumo- other blood-borne pathogens in health
nia in the mechanically ventilated pa- care settings. MMWR 1988;37:377-388.
tient. Respir Care 1989;34:85-97.
21. American Society for Testing and Mate-
rials. Standard specification for ventila-
tors intended for medical use in critical Additional Bibliography
care. ASTM F-29.03.01. Philadelphia:
Arn Soc Testing & Materials.
22. Centers for Disease Control. Update: Salyer JW, Chatburn RL. Patterns of prac-
Universal Precautions for prevention of tice in neonatal and pediatric respira-
transmission of human irnrnunodefi- tory care. Respir Care 1990;35:879-888.