You are on page 1of 6

TEST 

PROCEDURE OF ENERGY METER 
‘ELSTER’ 
A) SOFTWARE: 
1‐ You can’t make any change on relay software that saved on the relay and also 
can’t send it to another relay. 
2‐ To modify any change of the setting of relay, you must make new program & 
also this new program can send to any relay. 
3‐ To open the file you make, choose file and open. But, take care we work at type 
A3ALPHA‐R (Reactive Power). 
4‐ Reading  from  the  relay  is  done  by  choosing  ‘Diagnostic  Read’  &  ‘Short 
Diagnostic’. 
5‐ Note:  there  is  one  default  program  saved  in  the  PC  software  named  Sample 
A3R, you can work with it(BETTER TO WORK WITH IT). 
6‐ From this program, we have settings change according to the station and other 
fixed settings, 
a) Variable setting: this setting is for CT & VT ratio. 
b) Constant setting: 
1‐ Metering  Quantities:  put  KWH‐Del,  KVARH‐Del,  KWH‐Rec  and  KVARH‐
Rec. 
2‐ Display Values in normal mode: 
‐ Date. 
‐ Time. 
‐ Voltage for each phase. 
‐ Current for each phase. 
‐ System KW. 
‐ System KVAR. 
‐ System P.F. 
‐ System FREQUENCY. 
‐ Current Billing Total KWH‐Del, KWH‐Rec, KVARH‐Del& KVARH‐Rec. 
‐ Previous  Billing  Total  KWH‐Del,  KWH‐Rec,  KVARH‐Del&  KVARH‐Rec 
(OPTIONAL). 
3‐ Register Mul plier = 1000. 

Page | 1  
 
4‐ You  must  make  configuration  for  output  pulse  contact  (FROM  RELAY 
OPTIONS). 
NOTE (IMPORTANT): 
You  have  6  contacts  (A,  B,  C,  D,  E&  F),  but  we  use  only  4  contacts  as 
following: 
• (A) Connect to external relay terminal block to (32‐30). 
• (B) Connect to external relay terminal block to (32‐27). 
• (C) & (D) NOT USED (DISABLE). 
• (E) Connect to external relay terminal block to (32‐29). 
• (F) Connect to external relay terminal block to (21‐22). 
7‐ Sending your program to relay is done by choose PROGRAM then GO. 
8‐ You  can  make  clear  the  data  using  software  by  choose  CLEAR  VALUES  & 
STATUES. 
 
B) CALIBERATION TEST: 
1‐ MIN  OPERATING  CURRENT:    This  is  the  min  current  that  the  relay  can  make 
pulse with it. 
2‐ MEASUREMENT TEST: This test made according to the secondary side and the 
relay display values also according to the secondary side. 

 
3‐ ENERGY MEASUREMENT TEST:  
‐ FIRST,, TAKE CARE in this test must make calibration for Volt, Current 
and  P.F  Angle  that  come  out  from  the  ‘FREJA’,  because  ‘FREJA’  has 
error and this error may be effect to the energy measurement error 
of the relay,, this calibration is done using ‘ZERA’.  
‐ NOTE:  TAKE  CARE,  the  four  decimal  after  the  point  may  change  the 
percentage error of the relay, so you must make adjustment till reach 
the least error from FREJA. 
‐ In this test, our calculations are done according to the primary side. 

Page | 2  
 
 

 
‐ PROCEDURE FOR THE TEST: we have 2 method of procedure for this 
test. 
‐ FIRST METHOD: 
• First, make calibration for the required values as shown above. 
• Then,  make  clear  for  the  saved  results  in  the  relay,  but  Take 
care  to  make  this  step  by  injecting  voltage  only  and  the 
clearing is done by: 
ƒ Push the 3 push bo oms (RESET+ALT+TEST) in the front 
of the relay for 2 seconds, but TAKE CARE you must push 
the 3 bo oms together.  
ƒ Using software, by choose ‘’Clear Values & Statuses’’. 
• Inject voltage only from the first page of ‘FREJA’ for small time, 
me of star ng of relay (20 sec for example). 
• Inject the required values from second page of FREJA. 
• From the third page of FREJA put the time of starting and the 
required time of the test and inject. 
• After injection finish, inject voltage only to the relay and take 
the measurement of the required energy from the software of 
the relay. 
‐ SECOND METHOD: 
• Make clear for the saved results in the relay, but Take care to 
make this step by injecting voltage only. 
•  Inject  voltage  only  from  the  first  page  of  ‘FREJA’  for  small 
me,  me of star ng of relay (20 sec for example). 
• Inject  the  required  values  for  currents,  voltages  and  angles 
without making any calibration from the second page of FREJA. 
• From the third page of FREJA put the time of starting and the 
required time of the test and inject. 

Page | 3  
 
• Using  ‘ZERA’  choose  ‘W‐REGISTERTEST’  page  and  adjust  the 
option of MM (‘4WA’ for ‘KWH’ &’4WR’ for ‘KVARH’). 
• Start the third page of ‘FREJA’ and ‘ZERA’ in the same time. 
• After  finish  takes  the  reading  from  ‘ZERA’  (this  value  is  the 
expected value in format). 
• Inject  voltage  only  to  the  relay  and  take  the  measurement  of 
the required energy from the software of the relay. 
• Compare between the reading of ‘ZERA’ and the reading from 
the relay (this is the percentage error). 
• NOTE:  THIS  METHOD  IN  MOST  ACCURATE  BUT  MAY  BE  NOT 
CORRECT IN LIGHT LOAD AT 90 AND 270. 
4‐ MAXIMUM DEMAND TEST: 
‐ This test is same as the previous one, but in this one we measure the 
maximum demand (you can take this measure using software also). 
‐ In this test, you must make the se ng for interval in relay = 5 min & 
the se ng of subinterval = 1 min and inject for more than 5 min. 
‐ You  find  the  setting  of  interval  and  subinterval  in  the  software  in 
‘DEMAND’. 
‐ These calculations also are made according to the primary side. 
‐ You don’t need to make clear of the saved results in this test because 
the relay save the new result a er the interval  me in its se ng(i.e, 5 
min). 

 
 
C) ACCURACY OUTPUT TEST(Calibration with output pulses contact): 
1‐ In this test, we get the percentage error in the output pulse contact. 
2‐ We take the contact of the required measured value and connect to ‘ZERA’. 
3‐ ‘ZERA’ get the percentage error by comparing the measured values (from mini‐
clamps & voltage leads) and the energy values from the pulses (from contact). 
4‐ PROCEDURE OF THIS TEST: 
‐ Using  ‘ZERA’  chooses  ‘error  measurement’  and  from  it  you  find  3 
options (constant, pulses& MM). 
a) MM:  this  option  determines  the  value  you  want  to  measure 
(‘4WA’ for MWH OR ‘4WR’ for MVARH). 

Page | 4  
 
b) Pulses: this option determines the required number of pulses you 
want  to  compare  in  one  scan  (choose  ‘SC’&  put  any  number  of 
pulses) BUT TAKE CARE, for more accurate, it’s better to increase 
the number of pulses (for example, 50). 
c) Constant: this is the most important value, if you put this value 
wrong so the error will be wrong. 
o This  value  is  depend  on  the  pulse  speed,  but  here  we  work 
with contact that’s meaning high speed pulse (arrow pulse on 
the screen) 
o  

 
  
d) There is an op on in Se ngs(2) named ‘Ra o for meter constant’ 
option must be off. 

‐ The last point in this test is to choose the configuration of the relay 
contact from the software settings. 
o Relay ‘A’: at relay terminal block meet (32‐30), choose it KWH‐
Del. 
o Relay ‘B’: at relay terminal block meet (32‐27), choose it KVAR‐
Del. 
o Relay ‘E’: at relay terminal block meet (32‐29), choose it KWH‐
Rec. 
o Relay ‘F’: at relay terminal block meet (21‐22), choose it KVAR‐
Rec. 
Page | 5  
 
 

D) GENERAL NOTES: 
1‐ Don’t forget to put the CT ratio and the VT ratio in ‘ZERA’ (you find these ratios 
in Settings). 
2‐ For  more  accurate  in  ‘ZERA’  readings,  you  have  option  for  mini‐clamp  rating 
named    ’I R’ choose this value ‘C1A OR C5A’. 
3‐ Take care from the directions of the mini‐clamps arrow must be right. 
4‐ The percentage error of this relay must be less than ± 0.2 %. 
5‐ To calculate the percentage error of any value 

 
6‐ There is a test named ’creeping test ’ (may be witness ask for it) also named ‘no 
current  test’,  this  test  is  done  by  inject  0  current  and  120  %  of  voltage  for 
certain  me  (for  example,  5  min)  then  check  the  energy  reading  value  from 
relay. 
7‐ The minimum value of voltage that can operate the relay is to inject (96 V) for 
phase‐A. 

By: ENG. Haytham Abdel‐Naby 

S/S 9029 Qassim‐3 Project 

Project Manager: Eng. Emad Radi 

Page | 6  
 

You might also like