Maintenance DD+DIS149.05E

5.2 SW patch installation......................................................................... 16


5.3 Film calibration................................................................................... 17
5.4 Density Reproduction Check ............................................................ 17
5.5 Test image evaluation........................................................................ 19
5.6 Host system image quality check..................................................... 20

6 Completion of the maintenance........................................ 20


6.1 Backup Data on CF Card ................................................................... 20
6.2 Reset of the relative counters in the Infocounter............................ 21
6.3 Customer Conversation .................................................................... 21

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / II DRYSTAR 5302 Edition 1, Revision 0
Agfa Company Confidential (Type 5366/100) 2005-09-09
Supprimer filigrane Wondershare
PDFelement

DD+DIS149.05E Maintenance

1 Safety

⇒ The DRYSTAR 5302 is suitable for continuous operation.


⇒ The DRYSTAR 5302 enables or helps in the diagnosis of a patient.

When operating or maintaining the DRYSTAR 5302 always observe the


following safety guidelines:
• Have electrical or mechanical defects repaired by skilled personnel only!
• Do not override or disconnect the integrated safety features.
• Ventilation openings may not be covered.
• Always switch off the DRYSTAR 5302 and disconnect the power cord
from the outlet before carrying out any maintenance work.
Always take into account the markings provided on the inside and outside
of the printer. A brief overview of these markings and their meaning is
given below:

Safety warning, indicating that the DRYSTAR 5302 User and


Reference Manuals should be consulted before making any
connections to other equipment. The use of accessory equipment
not complying with the equivalent safety requirements of this
printer may lead to a reduced level of safety of the resulting
system. Consideration relating to the choice of accessory
equipment shall include:
Use of the accessory equipment in the patient vicinity.
Evidence that the safety certification of the accessory equipment
has been performed in accordance with the appropriate IEC
60601-1 and IEC 60601-1-2 harmonized national standard.
In addition, all configurations must comply with the medical
electrical systems standard IEC 60601-1-1. The party that makes
the connections acts as system configurator and is responsible
for complying with the systems' standard. If required contact your
Agfa Service Organization.
In order to reduce the risk of electric shock, do not remove any
covers.

Caution hot:
Keep hands clear from the Thermal Print Head.

Type B equipment:
Indicates that the DRYSTAR 5302 complies within the limits for
type B equipment.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0 DRYSTAR 5302 Chapter 9 / 1
2005-09-09 (Type 5366/100) Agfa Company Confidential
Supprimer filigrane Wondershare
PDFelement

Maintenance DD+DIS149.05E

Supplementary protective earth connector:


Provides a connection between the DRYSTAR 5302 and the
potential equalization busbar of the electrical system as found
in medical environments. This plug should never be
unplugged before the power is turned off and the power plug
has been removed.
Intergrounding connector:
Provides a connection between the printer and other
equipment which might exhibit minor ground potential
differences. These differences may degrade the quality of
communication between different equipment. Never remove
connections to this terminal.
Protective earth (ground)
Provides a connection between the printer and the protective
earth of the mains. Do not remove this connection, because
this will have a negative influence on the leakage current.
Power/Reset Button
Note that the power cord has to be disconnected from the wall
outlet in order to disconnect the unit entirely from the mains.
Precautions for use in USA only:
Make sure that the circuit is single-phase center-tapped, if the
printer is connected to a 240 V/60 Hz source instead of a
120 V/60 Hz source.
Transport after installation
Before moving the printer, always switch off the machine. The DRYSTAR
5302 should always be transported by 2 persons. When doing this, the
stability and the structure of the table top have to be taken into account.
The printer should not be placed on a soft surface, since this might
prevent appropriate ventilation and cause overheating. Nothing else
should be installed on the same table, and measures should be taken to
prevent table shock, while this may cause image artifacts. The appliance
shall only be transported with all covers closed. The appliance shall not be
transported continuously from one location to the other. Do not lift the
printer by the output tray.

Waste disposal and environmental regulations


In most countries DRYSTAR film is considered industrial waste and is not
allowed to be disposed of as household waste. Please consult your local
waste disposal regulations. Agfa recommends to have DRYSTAR film
waste hauled away by a licensed company.
After its life span, do not dispose of the DRYSTAR 5302 without
consideration of local waste disposal regulations. Please consult your
local service organization.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 2 DRYSTAR 5302 Edition 1, Revision 0
Agfa Company Confidential (Type 5366/100) 2005-09-09
Supprimer filigrane Wondershare
PDFelement

DD+DIS149.05E Maintenance

2 Introduction
The maintenance steps described in this document are based on the
technically required measures to guarantee the proper operation of the
printer.
The maintenance points have been arranged in a chronologically suitable
order to make the work routines as efficient as possible. The sequence of the
maintenance points in the checklist (see appendix) is identical with the
maintenance instructions.
Additional maintenance steps can be required according to national
regulations.

2.1 Maintenance frequency

The maintenance has to be carried out:


• every 25.000 printed films, or
• every 12 months, whichever comes first.

2.2 Required time

• Time required for the maintenance: ≤ 2 hours,

2.3 Required tools


Description Order Number / Comment
Calibrated densitometer recommended types are
• MacBeth TR-924
• X-Rite 301
• X-Rite 310
• X-Rite 331
• X-Rite 341
• Gretag D200-2
Flash light commercially available
Loop 8x CM+9.9579.9904.0
Service PC with Drystar 5302 Density For Service PC specifications refer to
Reproduction Check files chapter 3.2, Tools.
Density reproduction check files are
available via MEDNET, GSO Library
Waterproof pen commercially available
Wrist strap CM+9.9999.0830.0

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0 DRYSTAR 5302 Chapter 9 / 3
2005-09-09 (Type 5366/100) Agfa Company Confidential
Supprimer filigrane Wondershare
PDFelement

Maintenance DD+DIS149.05E

2.4 Required cleaning material

Cleaning material Order Number


Agfa Cleaning Tissue (as alternative to CM+309931.0
Isopropyl alcohol or Ethanol and Lint
free cloth). Box of 50 pieces.
Isopropyl alcohol or Ethanol Commercially available
Lint free cloth Commercially available
Ink eraser 'Pelikan' 10+9.9999.1176.0
Vacuum cleaner
220/240V, 50Hz CM+9.9999.0895.0
110/115V, 60Hz CM+9.9999.0896.0

2.5 Required spare parts

The following spare parts are part of the Maintenance-assortment


CM+0 5366 0999 733

Description Order Number Case of Exchange/ Usage

Air Filter CM+343350.X Exchange each maintenance

Always have the latest Drystar 5302 software patches available on the
Service PC.
Depending of the urgency and classification of a software patch it might be
necessary to update the printer software.

Printed copies of this document are not controlled.


The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 4 DRYSTAR 5302 Edition 1, Revision 0
Agfa Company Confidential (Type 5366/100) 2005-09-09