You are on page 1of 2

Factory Temperature Calibration Certificate

EXO Conductivity/Temperature Sensor


Part# 599870

Serial #: 21A104772
Primary Range: -5ºC to 35ºC
Primary Uncertainty: ±0.01ºC
Secondary Range: 35ºC a 50ºC
Secondary Uncertainty: ±0.05ºC
Calibration Date: 12 Feb 2021

Calibration Equipment:
 Standard Platinum Resistance Thermometer (SPRT)
Instrument #: 162CE-5234
Calibration Date: 4 Jun 2019
 Hart Scientific 1594A Temperature Bridge
Instrument #: B51261
Calibration Date: 15 Jan 2021
 Hart Scientific 7321 High Precision Bath
Instrument #: 1

Selection of Calibration Results

Ref Temp (ºC) 21A104772 Result (ºC) Residual (ºC) Uncertainty (ºC)

35.276 35.276 -0.000117 0.0057

25.283 25.283 0.000014 0.0061

15.286 15.286 -0.000115 0.0057

1.286 1.286 -0.000020 0.0055

-4.717 -4.717 -0.000046 0.0055

Traceability:
We certify that we have calibrated this product following manufacturing standards with
calibrations that are traceable to the National Institute of Standards and Technology (NIST) or
are derived from accepted values of natural physical constants. As manufactured, the product
meets or excedes its current specifications.

Uncertainty:
Stated uncertainty calculations follow the guidelines described in the NIST Technical Note 1297,
Guidelines of Evaluating and Expressing the Uncertainty of NIST Measurement Results and are
expressed as expanded uncertainties with k=2, approximately 95% confidence.

Calibration Technician Date


13 February 2021
____________________________

Item # 599377 DWG # 599377


Rev B April 2015
Item # 599377 DWG # 599377
Rev B April 2015

You might also like