Factory Temperature Calibration Certificate EXO Conductivity/Temperature Sensor Part# 599870
Factory Temperature Calibration Certificate EXO Conductivity/Temperature Sensor Part# 599870
Serial #: 21A104772
Primary Range: -5ºC to 35ºC
Primary Uncertainty: ±0.01ºC
Secondary Range: 35ºC a 50ºC
Secondary Uncertainty: ±0.05ºC
Calibration Date: 12 Feb 2021
Calibration Equipment:
Standard Platinum Resistance Thermometer (SPRT)
Instrument #: 162CE-5234
Calibration Date: 4 Jun 2019
Hart Scientific 1594A Temperature Bridge
Instrument #: B51261
Calibration Date: 15 Jan 2021
Hart Scientific 7321 High Precision Bath
Instrument #: 1
Ref Temp (ºC) 21A104772 Result (ºC) Residual (ºC) Uncertainty (ºC)
Traceability:
We certify that we have calibrated this product following manufacturing standards with
calibrations that are traceable to the National Institute of Standards and Technology (NIST) or
are derived from accepted values of natural physical constants. As manufactured, the product
meets or excedes its current specifications.
Uncertainty:
Stated uncertainty calculations follow the guidelines described in the NIST Technical Note 1297,
Guidelines of Evaluating and Expressing the Uncertainty of NIST Measurement Results and are
expressed as expanded uncertainties with k=2, approximately 95% confidence.