Professional Documents
Culture Documents
KN 120-294 Service Manual
KN 120-294 Service Manual
KN 120 – KN 294
SERVICE MANUAL
1
CONTENTS Page
SECTION 1 INTRODUCTION
2.1.General Overview 4
2.2. Explanations of Functions 4
2.2.1. Power Supply Values 4
2.2.2. Main PCB 4
2.2.3. Control Panel 5
2.2.4. Cooling unit 6
2.2.5. Lightening Unit 6
2.2.6. Battery Charging Unit 6
SECTION 3 SERVICING
2
SECTION 1
INTRODUCTION
1.1. PURPOSE OF THE SERVICE MANUAL
This manual includes servicing and maintenance information for KN 120 and KN 294. It
is prepared to be used by technicians who were formerly trained by Nuve. This manual
informs the technicians about the operating principles, diagnosing and repairing
methods and spare part replacing.
In the case that any problem which is not identified in this manual exists, please contact
Nuve.
Factory
Esenboga Yolu 22. Km.
Akyurt 06287 ANKARA TURKEY
Tel: 00 90 312 399 28 30 (3 lines)
Fax: 00 90 312 399 21 97
e-mail: service@nuve.com.tr
Control units are located at the top of the instrument. The air inside is sucked by the fan
that is located at the top of the chamber then passes through the evaporator that is
located at the top, between the chamber and the body and again enters to the chamber
from the opennings at the back side of the chamber. The cooling function and the air
circulation is provided by this way. The KN series blood bank refrigerators are
microprocessor controlled units.
3
SECTION 2
OPERATING PRINCIPLES
The KN series Blood Bank Refrigerators may be split into 4 main groups;
• Power supplies
• Control unit
• Cooling unit
• Lightening unit
Power
Glass Fuse Power inlet
Consumption
The microprocessor controlled main PCB operates on the “On/Off” controlling system and all
excluding the circular chart recorder all the control elements are connected to the main PCB.
These are;
• Temperature sensor
• Lightening unit terminals
• Keyboard and display PCB
• Cooling group terminals
• Circulation motor terminals
• Remote alarm and NC contact inlets
4
2.2.3. Control Panel
Figure 1
1. Temperature Display
2. Cooling Led
3. Door Open Led
4. Value Increase and Decrease Keys
5. Alarm Led
6. Alarm Mute Button
7. Lightinh Button
8. Set Button
9. Open/close Key Lock
5
2.2.4. Cooling Unit
The cooling function is obtained by circulating the refrigerant gas within a close circuit system.
The elements of this close circuit system are;
• Compressor
• Condenser Cooling Fan
• Condenser
• Dryer
• Capillary Pipe
• Evaporator
Figure 2
Samples can easily be observed by the lightening lamp inside the chamber. The lightening unit
elements are connected to the main PCB. They are;
• Lightening switch
• Ballast
• Starter
• Lightening lamp
In case of cut of electricity chamber temperature and the other displays are kept working by the
aid of a battery which is charged automatically. Battery charging elements are;
6
SECTION 3
SERVICING
Caution: Before servicing the instrument, please take the necessary precautions both
for your health and environmental safety. Please respect the warnings on the unit!!
The components on the main PCB must not be replaced even the failure is caused by one of
the components on the main PCB. In this case, please send the failed PCB to factory service
along with a note on which the failure explanations are written.
Before replacing the PCB or any control element, please make sure that the failure is not
caused by weak wire and terminal connections.
7
3.2. GENERAL FAILURE
1 The on / off switch is on Power supply inlet failure. Check that mains voltage is supplied.
but the power led does not Check that the terminals and power supply
lighten and the display is sockets are well connected.
blank.
The glass fuse has blown. Replace the glass fuse and check the short
circuit that causes the blowing of the fuse.
2 The on / off switch is on The main PCB / display Remove this connection cable and re-place
and the power led lightens PCB connection cable is not it carefully.
but the display is blank or connected properly.
some segments are blank.
The main PCB / display Replace the connection cable.
connection cable is defective.
3 The unit blows fuse Short-circuit exists. Check the electrical terminals and
frequently. cables for short circuit.
4 The circulation motor does The motor cable connection Check these terminals if they are connected
not work. terminals are loose. properly.
8
FAILURE PROBABLE CAUSES SOLUTION
6 The condenser fan does not Motor cable connection terminals Check that all terminals are
work. are loose or broken. connected properly.
The cooling terminal on the main Check the outlet with a multimeter. If
PCB does not outlet to the SSR 12 V DC does not exist, replace the
(solid state relay). main PCB.
7 The compressor does not work. The compressor cable Check the cable connection
connection terminals are loose or terminals and ensure that they are
broken. connected properly.
The cooling terminal on the main Check the outlet with a multimeter. If
PCB does not outlet to the SSR. 12 V DC does not exist, replace the
main PCB.
Compressor windings are
burned. Replace the compressor.
9
3.3. TEMPERATURE REGULATION OPERATING FAILURE
1 Temperature does not There is excessive heat loss. Check that the chamber gasket is tight enough.
reach the set value.
The cooling terminal on the Check the outlet with a multimeter. If 12 V DC
main PCB does not outlet to does not exist, replace the main PCB.
the SSR.
2 Actual temperature inside Temperature sensor reads Replace the temperature sensor.
differs from the displayed incorrect values.
one.
The main PCB calibration Apply the calibration procedure, if the problem
fails. continues replace the main PCB.
3 “Er1” appears The temperature sensor is Check that the endings and clamp connections
continuously on the broken or the PCB are not loose.
temperature display. connection is loose.
10
3.4. COOLING SYSTEM FAILURES
CAUTION: Cooling system failures should only be handled by the trained and
experienced technicians.
• Connect the maximum pressure hose of the gas manifold to the vacuum pump, the
minimum pressure hose to the compressor service line and the gas charge hose to gas
charge cylinder.
• Open the minimum and maximum pressure valves.
• Operate the vacuum pump and vacuum the system for at least 60 minutes.
• Stop the vacuum pump after closing the maximum pressure valve.
11
Figure 3
12
SECTION 4
CALIBRATION PROCEDURE
4.1 HOW TO MEASURE THE ACTUAL TEMPERATURE
• Insert the calibrated external sensor in the chamber from the calibration probe enterance
at the left side of the unit.
• Start the program.
• Please wait approximately 4 hours (after the set temperature) before measuring the actual
temperature by the external temperature sensor.
• Be sure that calibrated external sensor does not touch any metal part or any of the blood
bags, it measures only the air temperature.
• If there is difference between the equipment’s display and the external sensor apply the
calibration procedure as below:
4.2 CALIBRATION
• Press “value increase & decrease buttons” together continuosly until you see “tEC” on the
display.
• Press “SET” and you see “tCP” on the display.
• Press “value increase button” to enter, “000” starts to flash.
• Make this value “211” by using “value increase button”.
• Press “SET” and you see “PrG / Con” flashing alternately.
• Press “SET” and you see “P00” on the display.
• Press “SET” with small intervals until you see “P03” on the display.
• Press “value increase button” to enter and you see a value on the display.
• Change the value that appears on the display according to the difference between the
equipment’s display and the external sensor.
For example, if the set temperature is 4°C and the external sensor measures 7°C, increase
the value, which appears on the display, by adding 3; if the set temperature is 4°C and the
external sensor measures 0°C, decrease the value, which appears on the display, by
subtracting 4 by using “value increase or decrease keys”.
• Press “SET” to validate and you see “P04” on the display.
• Press “SET” with small intervals until you see “PrG / Con” flashing alternately.
• Then press “alarm mute button” to return to the stand-by mode.
13
SECTION 5
SPARE PART REPLACING
CAUTION : Unplug the refrigerator before replacing any part!!
• Disconnect all cable terminals which are connected to the main PCB.
• Unplug the display PCB / main PCB connection cable ending from the main PCB.
• Remove the 4 connection clamps and then the PCB.
• Place a new PCB on the connection clamps and make the terminal connections according
to the electric diagram provided in Section 6.
• Plug the display PCB / main PCB connection cable.
• Disconnect the display PCB / main PCB connection cable from the display PCB.
• Remove the 4 connection screws and remove the display PCB.
• Assemble a new PCB.
• Check that the display keys and the led lamps meet their places on the panel.
14
5.7. REPLACING MAGNETIC GASKET
15
SECTION 6
16
6.1. ELECTRICAL CIRCUIT DIAGRAM
KN 120
17
KN 294
18
SECTION 7
19
20
21
22
7.1 Spare Parts List
MATERIAL
NO NAME
KN 120 QTY KN 294 QTY
1 Outer body upper sheet Z 11. G 04 217/d3 1 Z 11. G 04 260/d4 1
2 Battery Z 12. A 10 001 1 Z 12. A 10 001 1
3 Battery Charging PCB Z 15. E 05 061 1 Z 15. E 05 061 1
4 Ballast Z 12. F 05 004 1 Z 12. F 05 017 1
5 Starter Z 12. F 05 002 1 Z 12. F 05 002 1
6 Transformer Z 15. T 18 006 1 Z 15. T 18 006 1
7 Main PCB Z 15. E 05 060 1 Z 15. E 05 060 1
8 Plastic Panel Z 15. P 15 185 1 Z 15. P 15 185 1
9 SSR (solid state relay) Z 12. R 07 013 2 Z 12. R 07 013 2
10 Shelf Carrier Z 12. K 23 027 10 Z 12. K 23 027 12
11 Door Handle Z 12. K 15 012 1 Z 12. K 15 012 1
12 Shelf Z 11. R 01 081 5 Z 11. R 01 097 6
13 Door Glass Z 11. C 01 029 1 Z 11. C 01 039 1
14 Chart Recorder Z 12. I 01 055 1 Z 12. I 01 055 1
15 Circulation Fan Z 15. F 04 016 1 Z 15. F 04 016 2
16 Evaporator Z 15. S 11 039 1 Z 15. S 11 040 1
17 Glass Fuses Z 14. S 02 025 2 Z 14. S 02 025 2
18 NC Contact Inlet Z 12. F 07 021 1 Z 12. F 07 021 1
19 Remote Alarm Inlet Z 12. F 07 020 1 Z 12. F 07 020 1
20 Compressor Z 12. E 03 021 1 Z 12. E 03 023 1
21 Condenser Z 12. K 20 009 1 Z 12. K 20 013 1
22 Condenser Cooling Fan Motor Z 12. M 06 042 1 Z 12. M 06 042 2
Magnetic Gasket Z 15. C 03 139 1 Z 15. C 03 152 1
Door Lock Z 12. K 27 041 1 Z 12. K 27 041 1
Door Hinge (upper) Z 11. K 14 046 1 Z 11. K 14 046 1
Door Hinge (lower) Z 11. K 14 047 1 Z 11. K 14 047 1
Temperature Sensor Z 15. I 01 062 1 Z 15. I 01 062 1
Seperator Z 15. P 17 009 25 Z 15. P 17 010 30
Fluorescent Lamp Z 12. F 05 025 1 Z 12. F 05 035 1
Display PCB Z 15. G 05 023 1 Z 15. G 05 023 1
Electronic Relay Z 12. R 07 020 1 Z 12. R 07 020 1
Defrost Heater Z 15. I 02 133 1 Z 15. I 02 138 1
23