You are on page 1of 114

GE Healthcare

BRIVO CT315/325
THEORY OF OPERATION
Operating Documentation

5306516-1EN
Rev 4
© 2009-2012, General Electric Company,
All Rights Reserved.
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的
伤害。

• THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


WARNING
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE
(EN) OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO
PROVIDE TRANSLATION SERVICES.

Attention
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS
SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE
SERVICE PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC
SHOCK, MECHANICAL OR OTHER HAZARDS.

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL
copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event,
within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
To file a report:
Fill out a report on http://egems.med.ge.com/edq/home.jsp
Contact the local service coordinator.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be
performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall
be performed by qualified GE Medical personnel. The products involved (and the accompanying electrical installations) are
highly sophisticated, and special engineering competence is required. In performing all electrical work on these products,
GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

IMPORTANT! . . . X-RAY PROTECTION


X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should be thoroughly
read and understood by everyone who will use the equipment before you attempt to place this equipment in operation. The
General Electric Company, Medical Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no practical
design of equipment can provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully acquainted with the
recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports
available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the
International Commission on Radiation Protection, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems Group, its agents, and
representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

LITHIUM BATTERY CAUTIONARY STATEMENTS

CAUTION
Danger of explosion if battery is incorrectly replaced. Replace only with the same or equivalent
type recommended by the manufacturer. Discard used batteries according to the manufacturer’s
instructions.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEMS CQA Process to report all omissions, errors, and defects in this publication.

ATTENTION
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

REVISION HISTORY
REV Date Primary Reason for Change
1 08/06/2009 Initial release.

2 11/16/2009 System Updated system options. (Sec. 1)


Added Power Control Timing Chart. (Sec. 2)
Updated Safety Loop that can be controlled by the TGP and DTB boards. (Sec. 2)

OC Updated OC Block Diagram. (Sec. 1)

Revision History
Updated SATA Port number of Mainboard. (Sec. 2)
Updated SCB Board Block Diagram. (Sec. 3)

Table/Gantry Updated Illustration 2-8 Schematics Diagram of Relay Control in System. (Sec. 2)
Updated system thickness. (Sec. 2)

3 03/09/2012 System Updated figures for safety loop and E-Stop. (Sec.2)

4 06/08/2012 Appendix Updated storage temperature from -30~50 to 0~30.


Updated storage humidity to 0~70%.

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

REVISION HISTORY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SYSTEM

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - GENERAL DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 SYSTEM OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-3 SYSTEM SPECIFICATIONS AND DATA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-4 SYSTEM OPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

Table of Content
SECTION 2 - POWER DISTRIBUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-1 POWER DISTRIBUTION UNIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 SAFETY LOOP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-3 E-STOP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

SECTION 3 - SCAN OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 WARM-UP SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 AXIAL SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-2-1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-2-2 Scan Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
3-2-3 Asymmetric Scan (Only for Dual Slice System). . . . . . . . . . . . . . . . . . . . . . . 3-6
3-2-4 Twin Cine Scan (only for Brivo CT325) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-2-5 Auto mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
3-3 SCOUT SCANS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9

TABLE OF CONTENTS i
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

ii TABLE OF CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 1 - GENERAL DESCRIPTION

1-1 INTRODUCTION
The features of this CT (Computed Tomography) system, one of the CT scanners of this series, include the
following:
Workstation type information processing system,
Solid-state x-ray intensity detector,
Continuous rotation type gantry with slip rings and high frequency coupling.

SECTION 1 - GENERAL DESCRIPTION


This CT system is comprised of the following main components (called subsystems):
• Operator Console (Host)
• Scanning Gantry
This subsystem further includes the following subsystems:
- DAS/Detector
- X-Ray Generator (XG)
- Power Distribution Unit
• Patient Table
The system may be connected with some of the following customer option equipments:
• Advantage Windows Image Workstation
• Image Camera

1-2 SYSTEM OVERVIEW


The Operator Console (OC) controls the entire system, according to the operator’s operations. The OC sends
instructions to the processor of the TGP board, which then controls the gantry and table subsystems according to
the instructions.
The TGP board processor also passes the OC instructions to the processor of the OGP board which is equipped on
the gantry rotative frame. The OGP board controls the DAS subsystem, the collimator aperture, or positioning lights
according to the passed instructions.
The OGP board processor also passes the instructions from the TGP board to the processor of the x-ray generator
subsystem. The XG processor controls the x-ray generator according to the instructions (originally from the OC).
Reversely, the OC receives status information from the TGP board or other processors (via the TGP board).

SYSTEM 1-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

Illustration 1-1 System Block Diagram

Tube

JEDI
SR
AC115V
Auxiliary Unit Power Unit XG Power
AC380V
PDU BOX

Temperature
Gantry
Control
Rotational
OC
AC220V

Detector Axial Drive


AC220V

DAS DCP Transmitter Receiver DTB Board

Z-Axial Photo Gantry Cntr/ Axial


Step Motor Servo Motor
Sensor Disp Board Drive

Collimator

TGP Board
Aperture
Step Motor
Encoder

Gantry
OGP Board Stationary

SCB
Monitor
Keyboard

Step Motor Display


Step Motor PS2/USB USB
Driver Card
Cradle COM
Host Processor
Speaker Encoder NIC1

NIC2
Table Connector SATA
Latch Switch

HDD DVD
Table OC

1-2 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-3 SYSTEM SPECIFICATIONS AND DATA


The following tables describe system specifications and data.

Table 1-1 Scan Time

Scan Time [sec]


-
0.7 (Half scan)
1.0
1.5

SECTION 1 - GENERAL DESCRIPTION


2.0
3.0
5.0

Table 1-2 Scan/Recon/Cal FOV

Scan/Recon FOV Cal FOV


P-Head (18) Small (25)
Head (25) Small (25)
Body (43) Large (43)

Table 1-3 Image Spatial Resolution Related

No. of Actual Scan Views No. of Detector Active


Recon Matrix
Full Scan Half Scan Channels
685 X 1 (Brivo CT315) 512 X 512
972 635
685 X 2 (Brivo CT325) 256 X 256

SYSTEM 1-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-3 SYSTEM SPECIFICATIONS AND DATA (CONTINUED)

Table 1-4 KV-MA Stations

KV MA
10, 20, 30, 40, ..., 180, 190, 200
120
(in 5 mA increments)
10, 20, 30, 40, ..., 140, 150, 160
140
(in 5 mA increments)

Table 1-5 Slice Thickness

Single Slice Thickness [mm] Dual Slice Thickness [mm]


1i mode (1 image/rotation) 2i mode (2 images/rotation)
1, 2, 3, 5, 7, 10 1.2, 2, 4, 6, 10 0.6, 1, 2, 3, 5, 7, 10

1-4 SYSTEM OPTIONS

Table 1-6 Main System Options

Category Name Description


60 sec Helical -
Helical scan 90 sec Helical -
Scan/Recon
120 sec Helical -
Collimator 0.6 mm Slice thickness -
Denta Scan -
SmartPrep -
Including Volume Analysis, Volume
Volume Viewer
Rendering, Navigator
Software CTC -
AVA -
Perfusion II -
1s Scan -

1-4 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 2 - POWER DISTRIBUTION

2-1 POWER DISTRIBUTION UNIT


The power distribution unit (PDU) provides powers to all the subsystems.
• Provide AC power for X-Ray Generation, three-phase, 380V.
• Provide AC power for Gantry Axial Drive Motor, single-phase, 220V.
• Provide AC power for Gantry and Table control PS (TGP, DTB and Cradle Motor Driver), 220V.

SECTION 2 - POWER DISTRIBUTION


• Provide AC power for Gantry Power Supply of DAS, OGP, Collimator Control and RF Transmitter, Single
phase, 115.
• Provide AC power for OC, single phase with grounding, 220V.
• Provide a means for emergency stop of all X-Ray and drive power circuits, including JEDI, Axial Drive
Motor and Table Cradle Drive Motor (24V).

Illustration 2-1 Power Distribution to Subsystems

13 14
5 6
3 4 R3
1 2 R2
K11 R1
A1 A2
RMT J8

TS1 F1 TS2
CB1 NZMB1 A80 13 14
L1 1 2 5 6 L1
F2
L2 3 4 3 4 L2
L3 5 6 1 2 L3
F3
N K1
RMT J8 A1 A2

GND

CB6 D10 CB5 D4


1 2 220V 1 2
RMT J5
115V 1 2
SR 115V
CB4 D16

T1
CB7 D10
1 2
OC 220V
CB3 D16
1 2
Sv 220V
CB2 D6
1 2
Gantry P.S 220V

NEUTRAL
K2 N10
1 2
P.S 24V Table 24V

RMT J1 RMT J8

A1
K2
N A2

SYSTEM 2-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-1 POWER DISTRIBUTION UNIT(CONTINUED)


Circuit breakers CB1, CB2, ... , CB7 can turn OFF powers, as written below:
• The ‘SR115/120’ output is not controlled by any contacts; the power is always present in the CT system
unless circuit breakers CB4 or CB6 are turned off.
• The relay K1 is turned on some delay time later after K11 is turned on. The resisters connected to the
relay K11 suppress rush currents.
• Centralized Control Power Switch: CB1 serves as whole system power control switch for service and
operator.
• Gantry/Table Power Switches:
Gantry PS Switch (CB2): Power of RMT (24V), table driver and gantry fans will be cut off when CB2 is
off. And as a result of RMT 24V was cut off, the contactors K1 (K11) and K3 will be in state of break and
cut off, the power of x-ray generator and axial driver will be cut off, too. In case of CB2 is off, only slip
ring (CB4) and OC (CB7) could be power on.
Axial Drive (CB3): When CB2 is ON, CB3 can control power of Axial Drive & Motor.
Slip Ring (CB4): CB4 can control the power of 115V on slip ring, which provides power for DAS Power
Supply, Tube pump and fan (through JEDI) on rotation part of gantry. Detector heating is also controlled
by CB4. CB4 is valid when CB6 is switch ON (power to transformer is ON).
RMT (CB5): CB5 can control 220VAC used by RMT, which provides AC working current for contactors
in PDU box. In case of CB2 is ON, control voltage is valid, contactor can work normally and be controlled
by programmable control units. Except affected by CB2, CB5 is affected by CB6, CB5 is valid only when
CB6 is switch ON (power to transformer is ON).
Operator Console (CB7): CB7 can control the power of OC, and other breaks do not affect CB7 except
centralized control power switch (CB1).
The XG Power Switch is a node of safety loop. The contactor K1 (K11) is off when XG power is off, as a result, no
power on JEDI and axial drive.

2-2 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-1 POWER DISTRIBUTION UNIT(CONTINUED)

Power Control Timing Chart

Illustration 2-2 Power Control Timing Chart

INPUT Control

PON SW

E-OFF

Safety-Loop

SECTION 2 - POWER DISTRIBUTION


RESET Emer

OUTPUT Control

XG400(K1)
Resistor(k11)

Note
‘Safety Loop’ is always closed in normal conditions after the system is switched ON or reset; this
means that ‘XG Power’ is always supplied to the gantry in normal conditions.
In the illustration:
‘E-OFF’ turns ‘close’ or ‘open’ when either ‘E-OFF-O’ or ‘E-OFF-G’ turns ‘close’ or ‘open’. That ‘Safety Loop’ turns
‘close’ or ‘open’ means the system safety loop is closed or opened, including both ‘SAFE-O’ and ‘SAFE-G’ turns
‘close’ or ‘open’. (See the ‘Safety Loop’ (3-2) description)
‘Reset Emergency’ turns ‘close’ or ‘open’ when either ‘RST-EM’ or ‘RST-EMG’ turns ‘close’ or ‘open’.

SYSTEM 2-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 SAFETY LOOP


Illustration 2-3 shows a diagram of the system safety loop.
As shown, components on the gantry rotational block are not involved in the safety loop.

Illustration 2-3 Safety Loop

Gantry
DTB

J4.5 J4.6

CN17.5 CN17.6
TGPL

CN2.1 CN5.9

CN2.4 CN5.21

PDU On
J2 - 1 J2 - 4
Gantry
RMT
Board
J1.24 J1.23

SCB
Computer
K19 XG Power
will be off
OC
K5 K1
XG Power
Relays will be off
K11

Open of Safety Loop


The safety loop can be opened by any of the following:
• Control by the TGP and DTB boards:
The TGP board opens the safety loop in any of the following cases:
- When the TGP board receives from the OGP board a safety loop open demand due to an
overtime of x-ray exposure.
- When the TGP board detects abnormal communication with the OGP board or the operator
console (host processor) during x-ray exposure.
- When DTB detects scan abort command from OC is failed.
• Control by the host processor (i.e., system software):
The host processor opens the safety loop in any of the following cases:
- When the host processor receives a safety loop open demand from the TGP board.
- When the host processor detects an overtime of x-ray exposure.
- When the host processor detects an extra scan (other than scans which the host processor
instructed the TGP board to perform) performed.

2-4 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
- When DAS data were not transferred to OC without error on time.

SECTION 2 - POWER DISTRIBUTION

SYSTEM 2-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-3 E-STOP
Illustration 2-4 shows a diagram of the system E-Stop.

Illustration 2-4 E-Stop

Frame 1 Frame 2 Gantry

Front Cover

CN6.1 CN6.2 TGPL


CN3.2
CPU
CN3.1

CN2.3

CN2.6

CN5.20 CN5.8

PDU On
Gantry
J2 - 3 J2 - 6

RMT J1.28 J1.27


Board

Gantry/Table
Power will be off SCB
K6 Computer
K2

K19
OC
K5 K1
XG Power
Relays
will be off
K11

2-6 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 3 - SCAN OPERATION

3-1 WARM-UP SCANS


The system requires the ‘Warm-up’ scans to warm-up the x-ray tube prior to starting the first scan after power-on,
or when more than two hours have elapsed since the last scan.
The warm-up scans also have to be performed before performing phantom calibration which updates the calibration
files (CAL files), if more than three hours have elapsed since the last scan.
The operator selects either the Tube Warm-up or Daily Calibration warm-up scans, as listed in Table 3-1.

SECTION 3 - SCAN OPERATION


Table 3-1 Warm-up Scans

No. of Scans
Type Purpose
Brivo CT315 Brivo CT325
(Until the Anode temperature (Until the tube case temperature reaches
Tube Warm-up for x-ray tube warming reaches 48%.) 10%.)
(Until the Anode temperature (Until the tube case temperature reaches
for x-ray tube warming reaches 48%.) 30%.)
for calibration and x-ray 11 13 (Max.)
alignment - 10mm scan only - Including 8 scans (calibration) + 5 scans
- Using 10mm scan data, warm- (X–ray alignment)
up error is analyzed. - Calibration scans contain 10mm, 7mm,
5mm, 3mm, 2mm, 1mm, 0.6mm, and
Asymmetric scans. (0.6mm is optional.)
Daily Calibration - Using 7 mm scan data, warm-up error is
analyzed.
for system diagnostics 4 6
(IQ Performance Mea- - 1scan for Offset Mean and SD - 1 scan for Offset Mean and SD
surement Scan) - 3 scans for Ratio - 3 scans for Ratio
- 1 scan for A/B ratio
- 1 scan for Z-axis collimator step
operation

The ‘Tube Warm-up’ type scans or the first series of scans of ‘Daily Calibration’ warm-up the x-ray tube.
This means that Tube Warm-up only warms the x-ray tube.
During the scans for calibration (eight scans) the system collects data to analyze it, and update CAL files:
The system stores the data and checks whether the data falls within the specified range. If the data is out of range,
the system reports a ‘WARM-UP ERROR.’ If the data is within range, the system compares the data with the previ-
ous warm–up data and calculates correction factors for CAL files, and updates them.
During the scans for x-ray alignment (three scans), the system collects data to analyze the x-ray alignment.
IIllustration 3-1 shows the warm-up scan sequence.

SYSTEM 3-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-1 WARM-UP SCANS (CONTINUED)

Illustration 3-1 Warm-up Sequence

Start

Tube warm-up
scans
Error Message

Error & Yes


Error code Any error?

No Raw Data Disk

Scans &
Data collection

Raw data
files
Data check

Error Message

Warm-up No Within
Error range?
CAL
Yes files
Calculates
correction
factors System Disk

CAL files
correction

End

3-2 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-1 WARM-UP SCANS (CONTINUED)

IQ Performance Measurement Scan


During the IQ Performance Measurement scans the system collects data to analyze it, and displays the result in
the ‘Status’ column.
Analyzed data: Offset mean, Offset SD, Stability of air calibration, mA-dependent DAS linearity variation, Scan
time–dependent DAS linearity variation, A/B ratio.
For details, refer to the ‘Home’ descriptions, in Section ‘Service Tool’, System tab, of Diagnostics manual.

Warm-up Error Analysis


The analysis for detecting warm-up error is performed using 7 mm scan data (for Twin) or 10 mm scan data (for

SECTION 3 - SCAN OPERATION


Single):

Analysis Items Descriptions Specifications


Ref Count Reference count after offset file correction 3000 ~ 60000
SD/MEAN SD/Mean value after offset file and reference corrections 0 ~ 0.03
Mean after offset file and reference correction / Mean at last
Ratio 0.95 ~ 1.05
daily calibration
HPF (Ratio) HPF of Ratio – 0.01 ~ 0.01
Ratio between [Channel data at the 1/4 intervals from 1st
Aperture Check 0.5 ~ 1.5
view] and [Mean value of all views]

When you analyze a warm-up calibration file, convert the related vector file into the Mean file, then use the Generic
System Analyzer of the Service tool.

SYSTEM 3-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 AXIAL SCANS

3-2-1 General
Prior to the actual x-ray exposure, the system collects 256 views of offset data generated from the DAS. The offset
data is used to correct actual x-ray data.
The system performs a full 360 deg. scan (clockwise direction only) during 1.0, 1.5, 2.0, 3.0 or 5.0 sec. These are
called scan speeds or scan times, and not all of them are available to all the systems; the available scan speeds or
times vary according to system models or options installed on the systems.
During a scan, the system collects 972 views of data. A view period differs according to the scan speed. The
following table shows this relation:

Scan Speed (sec) View Period (µsec/view)


1.0 1029
1.5 1543
2.0 2058
3.0 3086
5.0 5144

The view period is synchronized with the azimuth encoder pulse.


Axial scans are initiated from either 0 deg. or 180 deg. azimuth angle except for helical scans; during helical scans,
scans can be initiated from any azimuth angle.
The system knows gantry azimuth position by azimuth encoder pulse counts and the Gantry Pulse which indicates
that the gantry azimuth is at this moment at 0 deg., (i.e., home position).

3-2-2 Scan Sequence


Illustration 3-2 shows the axial scan sequence.

3-4 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 AXIAL SCANS (CONTINUED)

Illustration 3-2 Axial Scan Sequence

Parameters (Yes)
change?
Scan parameters setting (KV, MA, Slice Thickness, Scan Time, Slice Interval)
(No)
Cradle positioning (Patient is moved into the gantry scan plane)

Rotor start (The rotor is accelerated to operation speed)

SECTION 3 - SCAN OPERATION


Gantry acceleration (Gantry is accelerated to the constant scan speed)

Offset data collection (256 views of offset DAS data are collected )

High Voltage ON (Start x-ray exposure)

Das enable (DAS is enabled to collect data)

Data collection (972 views of actual DAS data are collected;


view period is synchronized with gantry azimuth position)

High Voltage OFF

Gantry deceleration
(Yes)
Next scan?
(No)

End

SYSTEM 3-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 AXIAL SCANS (CONTINUED)

3-2-3 Asymmetric Scan (Only for Dual Slice System)

Asymmetric Mode
In scanning, you can choose either symmetric or asymmetric slices from the twin-detector.
There are three choices in the Asymmetric Mode.
• [OFF] mode: (Symmetric is OFF) acquires slices with the same thickness (symmetric).
• [10:1] mode: acquires slices with different thickness (asymmetric).
- [10:1]: provides 10mm and 1mm slices.
- [10:2]: provides 10 mm and 2 mm slices.
- [6:1]: provides 6 mm and 1 mm slices.
Illustration 3-3 shows how asymmetric slices are obtained.
In Asymmetric Mode, the slice, whose thickness is shown at the Axial Thickness field, asymmetrically straddles the
center of the twin-detector. This slices is called “fused”, whereas the other thinner slices is derived from one detec-
tor.

3-6 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 AXIAL SCANS (CONTINUED)

Illustration 3-3 Asymmetric Slice

Asymmetric slices of [10:1]


Collimator

Radiation area
Center between
detectors A and B

SECTION 3 - SCAN OPERATION


1mm slice
9mm
Detector A Detector B

10mm slice

Asymmetric slices of [10:2]


Collimator

Radiation area
Center between
detectors A and B

2mm slice
8mm
Detector A Detector B

10mm slice

Asymmetric slices of [6:1]


Collimator

Radiation area
Center between
detectors A and B

1mm slice
5mm
Detector A Detector B

6mm slice

SYSTEM 3-7
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 AXIAL SCANS (CONTINUED)

3-2-4 Twin Cine Scan (only for Brivo CT325)


In Cine Scan, continuous multiple 360-degree data acquisition is achieved without table travel or at the same loca-
tion.
In Twin Cine Scan, images are displayed in the 1i mode during prospective recon; on the other hand, in retrospective
recon, images from Detector A and images from Detector B can be displayed alternately (2i mode).
The 1i mode display is also possible in retrospective recon.

3-2-5 Auto mA
The Auto mA feature enables the system to reduce mA and patient dose with negligible effect on image quality. The
system varies mA according to patient shape and a predetermined algorithm for scan location.
The Auto mA feature requires one Scout scan, either AP or LAT in the illustration below, to obtain necessary infor-
mation on X-ray attenuation. (The last scout scan data is used for the Auto mA scan.)
The system automatically increases mA at the scan locations where X-ray attenuation is relatively high and decreas-
es mA at the scan locations where X-ray attenuation is relatively low. In the example below, the mA is set low at the
start scan location (lung area), then the mA will increase at the location of liver area, and then it will decline at the
location of abdomen area.

Illustration 3-4 Auto mA

0–degree Scout (AP)

Tube

Patient 90–degree Scout (LAT)


Tube

Start End

mA

0 Scan location
Start End

3-8 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-3 SCOUT SCANS


Prior to the actual x-ray exposure, the system collects 64 views of offset data generated from the DAS. The offset
data is used to correct actual x-ray data.
The system advances the cradle and collects data from 685 active channels. The cradle speed is 75 mm/sec, and
data collection timing is synchronized with the cradle encoder pulse.
Illustration 3-5 shows the scout scan sequence.

Illustration 3-5 Scout Scan Sequence


Scan parameters setting (KV, MA, Scan Range)

Cradle positioning (Cradle is moved to the Start + 20 mm position)

SECTION 3 - SCAN OPERATION


Rotor start (The rotor is accelerated to operation speed)

Offset data collection (64 views of offset DAS data are collected )

Cradle acceleration (Cradle is accelerated to the constant scan speed)

High Voltage ON (Start x-ray exposure)

Das enable (DAS is enabled to collect data)

Data collection (Actual DAS data are collected;


view period is synchronized with cradle position)

High Voltage OFF

Cradle Stop Start


position
Cradle
move start Cradle stop

High DAS
Voltage enable
ON
Data collection
Cradle
acceleration

SYSTEM 3-9
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

3-10 SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

OPERATOR CONSOLE

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - GENERAL DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 Main Component Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-3 Operational Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

SECTION 2 - HOST PROCESSOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Table of Content
2-1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 Built-in I/O (Input/output) Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

SECTION 3 - SCAN CONTROL BOX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 FUNCTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2-1 Safety Block. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3-2-2 Scan Control and Communication Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-2-3 Audio Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
3-3 CONTROL KEYS/BUTTONS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

SECTION 4 - OTHER OC COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4-1 KEYBOARD, MOUSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

TABLE OF CONTENTS i
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

ii TABLE OF CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 1 - GENERAL DESCRIPTION

1-1 OVERVIEW
Single stand along workstation is used as Operator’s Console in Brivo CT315/325 system. No special or customized
peripheral equipments. The connections between OC and other subsystem are all standard interface, which
including RS232 (COM), USB, and Ethernet with RJ45 interface.
Standard keyboard and mouse are used as system input device.

SECTION 1 - GENERAL DESCRIPTION


Scan Control Box is used to provide scan control buttons, such as start scan, table movement, stop scan, Dr’s Mic
and Speaker, etc.
The information processing system (operator console) is based on the Intel Core2 Duo CPU architecture.
The host processor of the operator console is a Linux computer, which is on the market.
(You connect it to a monitor, standard keyboard, scan control box and mouse to configure a working system.)
Illustration 1-1 shows a block diagram of the operator console of this CT system.

OPERATOR CONSOLE 1-1


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
Illustration 1-1 Operator Console Block Diagram
HIS/RIS

10/100/1000 Mbps

Video Card
10/100/1000 M bps
Monitor
PS/2
Host Processor
Mouse Autovoice USB
USB Port
RS232

USB Port
SATA-1 SATA-2 SATA-4
SCB
Audio System Raw Data DVD Driver
Disk Disk

PS/2
STD Keyboard

Service Key

Scan Room
RS422
TGP

Gantry Mic
Table Speaker

DAS
DTB
Heart Gate
Trigger

1-2 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-2 MAIN COMPONENT DESCRIPTION

Host Processor
The host computer performs all of the following operations:
• User interface
• Image processings/display
• Communication
• Database control
• Raw data/image data flow control

SECTION 1 - GENERAL DESCRIPTION


• Image storage
• Scan/recon control
• Recon operation

This system disk mainly stores the followings:


• System and application software
• Images
• Calibration files
• System parameters

System Disk Capacity


• Image Storage: 150000 images

DVD Drive
Host PC shall provide DVD recordable drive (CD-R, DVD-RAM, DVD-R media)

Raw Data Disk


A hard disk drive which is a high-speed one for storing raw data.
• Raw Data Storage: <5000 Slice

Scan Control Box


The Scan Control Box is a separate part placed on worktable, convenient for operator’s operation. It provides not
only the system control functions, such as X-Ray on, table movement, scan termination, safety loop forward, patient/
doctor voice switch, but also voice amplifier, audio volume control and USB interface with host.

OPERATOR CONSOLE 1-3


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-3 OPERATIONAL DESCRIPTION

Scan Data Main Flow


Illustration 1-2 shows:

The raw data is then stored to a hard disk drive which is a high-speed type and exclusively used for raw data storage;
recon processing is performed to generate images.
The reconstructed images are then read by the host processor which converts the image data to video signal, as
well as stores the image data to the system disk, which also stores system software.

As shown, raw data and image data are mainly transferred over the network.

Illustration 1-2 Scan Data Main Flow

Monitor

Video Signal

System/Image Host
Disk Processor

Image Data
Network

Raw Data

Raw Data DTB


Disk
Raw Data from DAS

1-4 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 2 - HOST PROCESSOR

2-1 OVERVIEW
The host processor is a Linux computer.
The host processor is the central unit within the operator console, performing various operations such are described
below:
• Controls data transfer between data storage units and memory devices, etc.
• Controls the entire scan sequence; it sends necessary instructions to relevant devices, and receives

SECTION 2 -HOST PROCESSOR


status information from those devices.
• Recon operation.
• Performs graphics operations.
The host processor requires AC 220V power input.

Table 2-1 Main Specifications of the Host Processor

Component Specification
Central Processor Model Inter Core2 Duo CPU
Operating Frequency 2.83 GHz
Cache Memory 3 MB L2 Cache
Other 1066 MHz System Bus
Hyper Threading Technology
Main Memory Capacity (standard) 4 GB (2 GB X 2)
Type DDR2 667
Hard Disk Drive Capacity 320 GB (7200 rpm) SATA X 2
(System Disk)
DVD Type Super format: DVD+R, DVD+RW, DVD-RAM, CD-R, SATA
Graphics Card Resolution Nvidia FX570
256 MB DDR Memory: PCI-E X 16
Built-in I/O - PC (PS/2) compatible keyboard and mouse
(Input/Output) 1 x Serial Port
Parallel Port
Two 10/100/1000 Ethernet standard (RJ-45)
PCI-E X 16 Slot
PCI-E X 8 (X 4) Slot
4 PCI Slot
Built-in Audio - Stereo input/output (YES)
Operating System - CTT Linux 6.3.9 or later

OPERATOR CONSOLE 2-1


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 BUILT-IN I/O (INPUT/OUTPUT) DESCRIPTION


The host processor has input/output ports for the following:

Display Card (Graphics Card)


DVI signals are output from this port.

Keyboard/Mouse
Standard keyboard and 3-button Optical Mouse are connected these ports, respectively.

10/100/1000 Ethernet
A diagnostic console, an independent console, or an HIS/RIS is connected to this port.
DICOM compatible; the operator console can send (push) images to those devices using DICOM protocols at a 10
Mbits/sec maximum (the transfer speed varies according to receiver models, the size of transfer data, protocols,
etc.).

TCP/IP Protocal
Receive raw data from DTB by using TCP/IP protocal.

SATA Port
SATA Port 1: 320GB SATA WDC WD3201ABYS (System)
SATA Port 2: 320GB SATA WDC WD3201ABYS (Rawdata)
SATA Port 4: PLEXTOR DVD-RW

Built-in Audio
Audio in and Audio out is connected to Scan Control Box.

2-2 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 3 - SCAN CONTROL BOX

3-1 OVERVIEW
Scan Control Box is a stand along box to supplement control functions to motion parts (Table) and X-Ray, which is
related to patient safety, and must be ensured being controlled by operator. Which is placed on worktable, and
connected to OC and TGP. To simplify connections between subsystem, signals between OC and TGP were
forwarding (pass through) by Scan Control Box, and besides, Scan Controlo Box provides voice communication
and control function cooperated with OC sound adapter card and software. The electricity power (24V) is provided
by gantry power supply and through TGP forwarding.

SECTION 3 - SCAN CONTROL BOX


Illustration 3-1 Scan Control Box

3-2 FUNCTION
Scan Control Box provides following functions:
• User interface to interact with system to control table movement and touch off x-ray exposure with oper-
ator’s supervision.
• Voice communication and control between operator and patients, including microphone to collect doc-
tor’s voice and transfer to patient, receive patient voice and present patient voice by speaker, and patient
voice volume manually control.
• Signal forwarding between TGP and OC.

OPERATOR CONSOLE 3-1


Brivo CT315
Brivo CT325
PROPRIETARY TO GENERAL ELECTRIC COMPANY THEORY OF OPERATION
REV 4 5306516-1EN

3-2 FUNCTION (CONTINUED)

Illustration 3-2 SCB Board Block Diagram

SCB
E_stop
PDU Safety
E_reset
Safety Loop Block
Host
TGP Control
TGP Control (RS422)
TGPL X-ray On Scan
USB TO
Control Scan Commun Scan Box
TGP Reset Block ication USB
Block

Power 5V
Audio
and 12V
Mux ctrl

Key button

From Patient Record Autovoice


Audio In
To Patient Audio Play Autovoice Audio Out
Block

To Operator From Operator

The SCB board is connected to the OC and TGP, and as shown in Illustration 3-2.
The SCB board functionally consists of the following three blocks:
• Safety block
• Scan control and communication block
• Audio block

3-2-1 Safety Block


As shown in Illustration 3-2, this block includes the following circuits:
• Emergency Stop Button:
This is a normal connection (NC) node button. When the button is pressed the PDU will stop gantry pow-
er supply until the Emergency reset.
• Emergency Reset:
The Emergency reset is used for reset the PDU from Emergency status.
• Safety Loop:
The safety loop is a relay controlled by host computer. When the data transfer betweem the DTB and
computer works well, the safetyloop is closed. Otherwise, the safetyloop is opened. There is a LED
(safety) at the service window show the safety loop status.

3-2 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2-2 Scan Control and Communication Block


Keys used for scanning, and a microphone used for intercom, are incorporated into the SCB.
These keys are connected to the host processor via USB to uart chip. FPGA is used to control the communication
and implement the protocol between host processor. There is a download connector at the service window. It could
be used to update firmware.

3-2-3 Audio Block


The audio mixer block of the SCB board includes microphone, speaker, power audio amplifier for external speakers.

Illustration 3-3 Audio Block


Voice-detector

SECTION 3 - SCAN CONTROL BOX


Band-pass From patient
Filter volume
G Mic Analog Switch 2:
PreAmp OP. Amp Talkon PWR Amp OC SPK
From
Gantry Mic

OC Mic Analog Switch 1:


PreAmp OP. Amp Talkon* OP. Amp To Host Audiocard
OC Mic (record autovoice)
Gantry
PWR Amp
SPK
To patient
volume
From Host O P. A m p OP . A m p
Switch 2: Talkon
Audiocard (play
autovoice)

As shown, audio signals from several sources are mixed in this block.
• Audio Input Sources:
- Microphone on the Gantry (Intercom: from patient)
- Microphone in the SCB (Intercom: to patient)
- Audio output on the Host Processor (Auto voice play)
• Audio Output Destinations:
- Speaker on the Gantry
- Speaker in the SCB
- Audio input on the Host Processor (Record auto voice messages)

OPERATOR CONSOLE 3-3


Brivo CT315
Brivo CT325
PROPRIETARY TO GENERAL ELECTRIC COMPANY THEORY OF OPERATION
REV 4 5306516-1EN

3-2-3 Audio Block (Continued)

Audio Input Switch


As shown in Illustration 3-3, analog switches are used to switch audio input sources, and are controlled by the 'Talk
On' button on the SCB.
• An intercom function is provided between the OC room and the gantry room. The 'Talk On' button on
the SCB switches communication direction, as described below:
- While 'Talk On' is pressed, communication of "OC room  Gantry room" is enabled; and both
of Auto Voice (in the gantry room) and sound from the gantry to OC room are disabled.
- While 'Talk On' is released, communication of "Gantry room  OC room" is enabled. And when
a big enough sound trigger the detector circuit , the sound will amplifier out in the OC room. A
jumper at the service window is supplied to disable voice detector function. That means the OC
speaker always amplify out the sound in the gantry.
• Auto Voice messages can be recorded by using the microphone equipped in the SCB .

While Auto Voice is being played, communication of "Gantry room  OC room" is not disabled. The autovoice will
feedback from the gantry microphone to SCB speaker.

Volume Control
The following two kinds of volume control are provided.
• The following can be adjusted with the manual volumes on the keyboard:
- Volume of communication of "OC room  Gantry room" (through the SCB microphone)
- Volume of communication of "Gantry room  OC room" (through the gantry microphone)
• The electronic volumes are controlled by software, controlling the following:
- Auto voice volume in the OC room

3-4 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-3 CONTROL KEYS/BUTTONS


The following keys/buttons place on the Scan Control Box.
• Move to Scan: Table control
• Stop Move: Table control
• Start Scan: X-Ray touch off
• Pause: X-Ray interrupt, after 1 scan finished, and could be resumed
• Stop Scan: X-Ray interrupt, immediately, and could not be resumed
• Emergency Stop: Provide emergency power cut off when being pushed
• Emergency Stop Reset: Different button with emergency stop used to recover power on capability after

SECTION 3 - SCAN CONTROL BOX


Emergency Stop

OPERATOR CONSOLE 3-5


Brivo CT315
Brivo CT325
PROPRIETARY TO GENERAL ELECTRIC COMPANY THEORY OF OPERATION

3-6 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 4 - OTHER OC COMPONENTS

4-1 KEYBOARD, MOUSE


The keyboard used for this operator console integrates the following components:
• Standard Keyboard, Mouse:
This standard keyboard and the mouse have a PS/2 interface for connecting themselves to the host
processor.

SECTION 1 - OTHER OC COMPONENT


4-2 DISPLAY
• LCD (19”): The resolution used for LCD Displayis 1280x1024 matrix.

OPERATOR CONSOLE 4-1


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

4-2 OPERATOR CONSOLE


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

TABLE/GANTRY

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - GENERAL DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 COMPONENTS INTERCONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-3 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

SECTION 2 - SUB-ASSEMBLY DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Table of Content
2-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 GANTRY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2-1 Stationary Gantry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2-2 Rotationary Gantry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2-2-3 SLIP RING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-3 TABLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14

TABLE OF CONTENTS i
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

ii TABLE OF CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 1 - GENERAL DESCRIPTION

1-1 INTRODUCTION
The TGP (Table Gantry Processor) board is the main controller of the Table/Gantry subsystem. The TGP board
receives commands from the Operator Console and passes the OC instructions to the OGP (On Gantry Processor)
board, which is installed on the rotational Gantry Frame. The OGP board controls the DAS subsystem, the collima-
tor aperture, or positioning lights according to the passed instructions. The OGP board also passes the instructions
from the TGP board to the X-ray Generator subsystem. The XG processor controls the X-ray Generator according

SECTION 1 - GENERAL DESCRIPTION


to the instructions (comes from the OC). Reversely, the OC receives status information from the TGP board or other
processors (via the TGP board).
The Table/Gantry consists of the following components:
• TGP board
• OGP board
• Servo Amp/Servo Motor
• Gantry Display
• Gantry Switch
• Collimator
• Positioning Light
• Cradle drive system

TABLE/GANTRY 1-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-2 COMPONENTS INTERCONNECTION


The illustration 1-1 shows the components of the Table/Gantry.

Illustration 1-1 Table/Gantry

GANTRY
SWITCH GANTRY DISPLAY
PANEL

CRADLE POSIT ION AXIAL MOTOR


POTENTIOMETER
APERTURE
CRADLE ENCODER TGP BOARD MOTOR
OGP BOARD
SERVO AMP/MOTOR CRADLE
STEPPING MOTOR DRIVER MOTOR
GANTRY PULSE

APERTURE SENSORS
LATCH SWITCHES

EMERGENCY SWITCHES

Z–AXIS LIMIT SENSORS


DAS
Z–AXIS MOTOR

1-2 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

1-3 SPECIFICATIONS
The specifications of the Table/Gantry is as follows:

Gantry Rotation Speed


• 1.0, 1.5, 2.0, 3.0, 5.0 sec/rot.

Cradle Speed

• Slow: 20 mm/sec ± 3%


Fast: 100 mm/sec ± 3%

SECTION 1 - GENERAL DESCRIPTION


Scout: 75 mm/sec ± 3%
Home: 100 mm/sec ± 3%

TABLE/GANTRY 1-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

1-4 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 2 - SUB-ASSEMBLY DESCRIPTION

2-1 OVERVIEW
The Table/Gantry consists of Table and Gantry. The Gantry is divided into Stationary Gantry and Rotationary
Gantry. The communication between stationary and rotationary Gantry is made by slip-ring, a rotation mechanism
that permits electrical power and signals exchange.
The illustration 2-1 shows the Table/Gantry block diagram.

Illustration 2-1 Table/Gantry Block Diagram

SECTION 2 - SUB-ASSEMBLY
GANTRY

DESCRIPTION
STATIONARY GANTRY

PDU

ROTATIONARY GANTRY

POWER
GANTRY
OC DISPLAY
TGP BOARD
GANTRY COLLIMATOR
SWITCH
OGP BOARD

POSITIONING
LIGHT

AXIAL
SLIP
AXIAL DRIVE MOTOR RING
DAS XG

TABLE
Step Motor Driver Board Step Motor

Speaker

Cradle Pot

Cradle Encoder

Latch Switches

TABLE/GANTRY 2-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 GANTRY

2-2-1 Stationary Gantry


The stationary Gantry is mainly controlled by the TGP (Table Gantry Processor) board. The Illustration 2-2 shows
the main components of the stationary Gantry.
The TGP board communicates with Operator Console and Table via cables and with the OGP board located in
rotationary Gantry via cable and slipring.

Illustration 2-2 Stationary Gantry main components

PS
㪜㪄㪪㪫㪦㪧

PDU X-DISP ASSY

㪩㪼㪸㫉㩷㪚㫆㫍㪼㫉

MICRPHONE
㪫㫆㩷㪦㪚

㪫㫆㩷㪫㪘㪙㪣㪜 TGP 㪜㪄㪪㪫㪦㪧

㪫㫆㩷㪪㪣㪠㪧㪩㪠㪥㪞 S-DISP ASSY


㪝㫉㫆㫅㫋㩷㪚㫆㫍㪼㫉
DRIVE
MICRPHONE

㪤㪦㪫㪦㪩
㪜㪄㪪㪫㪦㪧
DTB

PS1
Switch Power Supply provides +24VDC to TGP.

2-2 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-1 Stationary Gantry (Continued)

TGP Board
TGP board is a communication core on the stationary gantry. It receives the commands from OC and passes the
OC instructions to the OGP board, which is installed on the rotational gantry frame. OGP board controls the DAS
subsystem, the collimator aperture, or positioning lights according to the passed instructions. OGP board also
passes the instructions from the TGP board to the processor of the X-ray Generator subsystem. The XG processor
controls the X-ray Generator according to the instructions.
The TGP Board issues commands to and receives status from the components it controls. The TGP oversees the
following functions:
• Axial Drive Control

SECTION 2 - SUB-ASSEMBLY
• Cradle In/out Drive Control
• Gantry Display Control

DESCRIPTION
• Gantry Control Panel
• Error Detection
Illustration 2-3 TGP Block Diagram
TGP BD Emergency
PDU
Rot Safe
PDU Safety
G-Pulse
Loop
Rot Count C
CAN
SRAM M
Driver
Rotate Enc
monitor
Flash
Test 0,1 Memory
Reset SW Slice Count
C
Status
Reset LED
Patarn DAS Slip
Trig Ring

OC OC Comm NIOS
unication 9600bps Slip
SIO Ring
OC OC Direct CT Lock
Line &Alarm Relay
Control
OC

Flash
Add Cont

16bit Cradle Pot


ADC
Cradle
Cradle
Motor M
Control
Driver
Enc

Display Gantry
Control Display

Gantry
SW
Table
SW

Mat SW

Audio Mic
Amp
Spk

TABLE/GANTRY 2-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-1 Stationary Gantry (Continued)


There are four service switches on the TGP Board which are used to control the Gantry Rotation manually.

Illustration 2-4 TGP Board Service Switches

System Jog
Service ON FAST
Mode Rotation

OFF
Manual Continuous SLOW
Service OFF
Mode Rotation

SW1 SW2 SW3 SW4

• SW1
- Service ON: When the switch is in this position, axial driver would be disabled, and Gantry can
be rotated by hand manuallly.
Note
When TGP finds this switch is toggled to “Service ON” state, TGP should give +0VDC to driver J4-
2 (DINP1) to disable driver. Meanwhile, TGP should give +24VDC to SBRK_+ to open the brake,
thus service guys can rotate the Gantry manually.
- Service OFF: When the switch is in this position, axial driver would be enabled and external
force can’t pull the Gantry. Gantry can only rotate by OC instruction or service switches.
Note
When TGP finds this switch is toggled to “Service OFF” state, TGP should give +24VDC to driver
J4-2 (DINP1) to enable driver. Meanwhile, TGP should give +24VDC to SBRK_+ to open the brake.
• SW2
- System Mode: When the switch is in this position, only instruction from OC can make Gantry
rotate. SW3 and SW4 can’t make Gantry rotate under this condition. (SW1 should be in
“Service OFF” position).
- Manual Mode: When the switch is in this position, TGP do not response to OC instruction. SW3
and SW4 can make Gantry rotate. (SW1 should be in “Service OFF” position).
• SW3
- JOG Rotation: This switch is a spring loaded switch. When the switch toggles to this position,
Gantry will rotate continuously under the speed set by SW4. When releasing this switch, Gantry
rotation will stop.
- OFF/Home: When this switch turns to this position from JOG mode, gantry will stop rotation
immediately. And when this switch turns to this position from continue mode, gantry will stop
rotation at Home position.
- Continuous Rotation: When the switch is in this position, Gantry will rotate according to the
setting of SW4.
2-4 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-1 Stationary Gantry (Continued)


• SW4
- FAST: When the switch is in this position, Gantry will rotate with speed 1.0 second/revolution.
- SLOW: When the switch is in this position, Gantry will rotate with speed 5.0 second/revolution.

DTB Board
DTB is used on the Gantry Subsystem for data transfer, see below diagram.

Illustration 2-5 DTB Block Diagram

SECTION 2 - SUB-ASSEMBLY
Transceiver Transceiver

DESCRIPTION
DTB Board performs the following functions:
• Read data from fiber interface.
For single system, the raw data volume is 972 x 740 x 2 byte/s (the fastest speed, 1s scan)
For dual system, the raw data volume is 972 x 740 x 2 x 2 byte/s (the fastest speed, 1s scan)
• Send these raw data to OC by using TCP/IP protocol. Refer to software requirements for more detail
information.
• LED is required to report data transfer status.
• DTB is needed to send “open safety loop” signal to TGP in case of “continuous bad raw error”, “no raw
data after X-ray on error” and “ FIFO overrun error”.
• CPU configures FPGA after power on.
• ISP (In System Programming) function is needed for DTB firmware/FPGA image download.
• Log mechanism and should have a way to transfer the log to OC.

TABLE/GANTRY 2-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-1 Stationary Gantry (Continued)


DTB Hardware Electrical Requirements
• Fiber Connector: A multimode 62.5/125um fiber transceivers (1x9 Package Style) as the raw data re-
ceive. The wavelength is 1300nm.
• RJ45 Connector: A 10/100M Base-TX RJ45 connector for data transmission to OC, and also used for
instructions transfer between DTB & OC.
• Signals between TGP & DTB
Signal Signal Description Active Output/Input
Reset_N Reset DTB, differential 422 signal L I
X-Ray On Inform DTB that X-Ray is on, differential 422 signal H I
Safety Loop Open Inform TGP to open safety loop, differential 422 signal H O

2-6 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-1 Stationary Gantry (Continued)

S-DISP
The Illustration 2-6 shows the block diagram of S-DISP. The control signals and power signals are transferred to
S-DISP from TGP. The CPLD can decode the control signals and generate the drive signals to control S/I/X-ray
indicators and 5 digit 7-segment displays. When a function key is pressed, the corresponding active signal is trans-
ferred to TGP. Function keys signal also connect with CPLD for test function.

Illustration 2-6 S-DISP Block Diagram

S-DISP
S/I/X-ray indicator 5 digit 7-seg displays

SECTION 2 - SUB-ASSEMBLY
TGP Interface CPLD
CN6

DESCRIPTION
(for test)

Emergency Display drive


5 function keys
interface

The S-DISP Board Assy enables user to control the movement of the cradle and shows some indication information.
• X-Ray Indicator
The X-ray indicator lamp illuminates during X-ray exposure.
• S/I Indicator
The “S” LED is illuminated to indicate the current scanned position is outside the internal landmark.
The “I” LED is illuminated to indicate the current scanned position is inside the internal landmark.
• 5 Digit 7-segment Displays Control
The 5 displays show the reading of the distance (mm) between the landmark and the anatomy in the X-
ray path.
• 5 Function Keys
There are 5 function keys on S-DISP board assy to control the movement of the cradle or turn on/off the
position light.
- In key
Use this key to move the cradle towards the Gantry
- Out key
Use this key to move the cradle away from the Gantry
- Fast key
Use this key to increase the speed of cradle in/out
- PLSW key
Use this key to turn on or off the positioning lights
- EXLMR key
Use this key to move the anatomy beneath the positioning light

TABLE/GANTRY 2-7
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 STATIONARY GANTRY (CONTINUED)


• Emergency Stop Interface
The S-DISP interfaces with the emergency button through one 2-pin connector. The signal of emergen-
cy button is transferred to TGP through S-DISP.

X-DISP

The Illustration 2-7 shows the block diagram of X-DISP. The control signals are transferred to X-DISP from TGP to
control the three LEDs on X-DISP board assy.

Illustration 2-7 X-DISP Block Diagram

X-DISP

TGP

CN10

The main function of this board is to get +24V signal and XLED signal and to illuminate the three LEDs on this board.
The three LEDs are illuminated during X-ray exposure.

2-8 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 STATIONARY GANTRY (CONTINUED)

Relay
There are three sets of relay that are used in system, the functions are described below.
• Relay 1: Detect the door state. When the door is closed, the switch on the door is closed also, so the
coil of relay is powered on and the contact of the relay is closed. The TGP can get the state of the door.
• Relay 2: Control the room lamp. The CT system will turn on /off the room lamp by controlling the relay 2.
• Relay 3: Control the ready status lamp. The CT system will turn on/off the ready status lamp by control-
ling the relay 3.

Illustration 2-8 Schematic Diagram of Relay Control in System

SECTION 2 - SUB-ASSEMBLY
+24VDC
㩺㪈㪏 㪮㪟㪫

DESCRIPTION
㩺㪈㪏 㪙㪣㪢
㩺㪈㪏 㪩㪜㪛
㩺㪈㪏 㪰㪜㪣
TGP
㩺㪈㪏 㪙㪩㪥
㩺㪈㪏 㪙㪣㪬

㪈㪋
㪂 㪚㪦㪤
㪩㪜㪣㪘㪰㪈 㪐

㪛㪚㪉㪋㪭
㪠㪥㪧㪬㪫㪈 㩿㪛㪦㪦㪩㩷㪪㪮㪠㪫㪚㪟㪄㪛㫆㫆㫉㩷㪺㫃㫆㫊㪼㪀
㪮㫀㫉㪼㩷㪫㪼㫉㫄㫀㫅㪸㫃 㪌 㪈
㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㩺㪈㪏 㪩㪜㪛
㩺㪈㪏 㪙㪣㪬 㪈㪊 㪥㪅㪦㪅

㪫㪈㪄㪈 㪫㪈㪄㪉 㩺㪈㪏㩷㪞㪩㪥
㪦㪬㪫㪧㪬㪫㪈 㩿㪙㪘㪰㩷㪣㪘㪤㪧㪀 㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㪫㪉㪄㪈
㩺㪈㪏㩷㪞㪩㪥
㪫㪉㪄㪉 㩺㪈㪏㩷㪙㪣㪢 㪚㪦㪤
㪉㪌㪇㪭㪘㪚㩷㪤㪸㫏㪃㩷㪌㪘㩷㪤㪸㫏 㪫㪊㪄㪈 㪫㪊㪄㪉 㪈㪊 㪄

㪩㪜㪣㪘㪰㪉
㪛㪚㪉㪋㪭

㪫㪋㪄㪈 㪫㪋㪄㪉
㪦㪬㪫㪧㪬㪫㪉 㩿㪩㪼㪸㪻㫐㩷㪪㫋㪸㫋㫌㫊㩷㪣㪘㪤㪧㪀 㪘㪮㪞㪉㪉㪄㪘㪮㪞㪈㪉 㪫㪌㪄㪈 㪫㪌㪄㪉
㩺㪈㪏㩷㪮㪟㪫 㪈 㪌

㪫㪍㪄㪈 㪫㪍㪄㪉 㪥㪅㪦㪅㪈㪋 㪂


㪉㪌㪇㪭㪘㪚㩷㪤㪸㫏㪃㩷㪌㪘㩷㪤㪸㫏
㪫㪎㪄㪈 㪫㪎㪄㪉
㩺㪈㪏㩷㪙㪩㪥 㪚㪦㪤
㪫㪏㪄㪈 㪫㪏㪄㪉 㪈㪊 㪄

㪩㪜㪣㪘㪰㪊
㪛㪚㪉㪋㪭

㩺㪈㪏㩷㪰㪜㪣㩷 㪈 㪌

㪥㪅㪦㪅㪈㪋 㪂

TABLE/GANTRY 2-9
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2 STATIONARY GANTRY (CONTINUED)

Axial Drive and Motor


The Axial Drive and the Axial Motor work together with the TGP, TGP sends the rotate command to the drive through
Canopen protocol by the control cable, besides the TGP will also supply the power (+24VDC) of brake to the motor
directly. The motor drive supplies three-phase power to the motor through power cable, sends command and re-
ceives feedback through signal cable.

Illustration 2-9 Gantry Rotation Control Diagram

TGP BD ASSY

Control cable
Gantry
Brake cable

Power cable
Signal cable Axial Motor
Driver

Belt Gear Ratio 20:1


Axial Motor with 10000line/rev Encoder

24V Power Supply


The switch power supply transfers the input power of AC 220V to the output power of 24V DC to provide the power
for the parts on gantry subsystem.

Illustration 2-10 24V Power Supply Distribution Diagram


+ 24V DC
220V AC
HW300-24A

DTB BD Step Motor TGP BD Left Top Fan Right Top Fan
RF RCV Signal Brush RMT Board
ASSY Drive BD ASSY ASSY ASSY

Gantry Microphone
Located at upper portion of the entrance of the Gantry opening, at the front and the rear side.

Emergency Switch
Located at left and right sides of the Gantry Frame.

2-10 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-2 Rotationary Gantry


The Illustration 2-11 shows the main components of the rotationary Gantry.

Illustration 2-11 Rotationary Gantry main components

PHOTO
DAS SENSOR
MOTOR
MOTOR
DRIVER

GPLS

MOTOR
DRIVER MOTOR
TO TGP COLLIMATOR
(SLIP RING)

SECTION 2 - SUB-ASSEMBLY
ENOCODER
OGP BOARD
POSITIONING
LIGHT

DESCRIPTION
PHOTO SENSOR

RELAY

SWITCHING
POWER
SUPPLY
TEMP
CONT
115VAC ASSY

TO DTB DCP DAS/DETECTOR


RF XMT

OGP Board
The OGP (On Gantry Processor) performs the following functions:
• X-ray exposure on/off control
• DAS data transfer timing
• Collimator control
• Beam Tracking Control
• Positioning light control
• Scan control
• The automatic shut-off of the Tube Fan/Pump Power Supply after 30 minutes when the OC has been
powered down.

TABLE/GANTRY 2-11
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-2 Rotationary Gantry (Continued)

Switching Power Supply


Supply power to the OGP board, DCP board, RF transmitter, Temp cont, DAS, and positioning light.
Supply power to the DAS and Temperature Control Box.

Collimator
Located at the exposure window of the X-ray tube, controls the aperture of the X-ray beam by moving the blades.
For Brivo CT315: The aperture can be 1, 2, 3, 5, 7 or 10 mm.
For Brivo CT325: The aperture can be 1.2, 2, 4, 6 or 10 mm in 1i mode.
The aperture can be 0.6, 1, 2, 3, 5, 7 or 10 mm in 2i mode.

Motor Driver
Generates five phase pulses to control the stepping motor of the collimator. Receives pulse command from OGP
board and convert it to other type of pulses.

Encoder
Sends the feedback pulse to the OGP to show the moved amount of the blades.

Photo Sensor
Detects when the collimator blades are in 1mm aperture

DAS
Data Acquisition System. Processes the data received from detector and send it to operator console. The detailed
explanations are in the DAS/Detector sections.

Temp Cont Assy


Device that controls the Detector temperature

DCP
Receives parallel data from DAS and convert them to serial data. The detailed explanations are in the DAS/Detector
sections.

RF XMT
Converts Serial signal from DCP2 to RF signal.

GPLS
Notify TGP and OGP board when the Gantry is at zero degree tube position.

2-12 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-2-3 SLIP RING


The Slip Ring transfers the following signals;
• DAS Data
• DAS Trigger
• TGP and OGP Communication message
• Safety Loop

MECHANISM
This Gantry slip ring system utilizes:
• Three 380VAC power high-power rings (nine high power brushes)

SECTION 2 - SUB-ASSEMBLY
• Two 120V power rings (six low power brushes)
• One safety ground ring (three ground brushes)

DESCRIPTION
• Two signal rings (four signal brushes)
• RF transmitter / receiver modules

Illustration 2-12 Simplified CT Contactless Slipring Assembly

HSDCD Transmitter HSDCD Receiver


Module Module
DC Power DC Power

TX RX

CT DAS Serial Datastream CT DAS Serial Datastream


From DAS To OC
(Network)
(Fiber)

Illustration 2-13 Control Signal Transfer Strategy

Six channels
TGP OGP

Mux Slipring two Mux


Stator channels Rotor
Cable A Cable B

Two channel cables

TABLE/GANTRY 2-13
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-3 TABLE
The Illustration 2-14 shows the main components of the table.

Illustration 2-14 Table

LATCH SWITCHES

TGP Board

SPEAKER

E
STEP ENCODER
MOTOR
From Gantry NF2 (CRADLE)
SW M
P.S. STEP MOTOR
DRIVER ENCODER
POTENTIOMETER

Cradle
Portion of the Table that executes longitudinal movement. It is controlled by cradle driver board.

Latch Switches
Positioned at both right and left sides of the Table, these switches release or latch the cradle. When both switches
are pressed, the cradle is latched. When one of the switches is depressed, the cradle is released. They are for safety
purpose.

Potentiometer
There is a potentiometer which is used for cradle absolute position.

Encoder
There is a encoder which is used for cradle relative position.

Speaker
Located at the table front bracket, it’s used for broadcasting operator’s voice.

2-14 TABLE/GANTRY
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

DAS/DETECTOR

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - GENERAL DESCRIPTION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

SECTION 2 - DETECTOR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1


2-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Table of Content
2-2 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-3 HILIGHT DETECTOR MODULE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-4 CHANNEL DISTRIBUTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-5 DAS/DETECTOR CONNECTOR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-5-1 DAS/Detector Connector for Single Slice System . . . . . . . . . . . . . . . . . . . . . 2-5
2-5-2 DAS/Detector Connector for Dual Slice System . . . . . . . . . . . . . . . . . . . . . . 2-6
2-6 QCAL CHANNELS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-7 AFTER GLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-8 HEATER AND TEMPERATURE CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8

SECTION 3 - DATA ACQUISITION SYSTEM (DAS) - SINGLE . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 DATA FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-3 DCP BOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-3-1 DCP Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-3-2 DCP Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-4 CONVERTER BOARD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9

SECTION 4 - DATA ACQUISITION SYSTEM (DAS)_DUAL. . . . . . . . . . . . . . . . . . . . . . . . . 4-1


4-1 OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 DATA FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3 DCP CARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3-1 DCP Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4-3-2 DCP Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4-4 CONVERTeR BOARD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9

TABLE OF CONTENTS i
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

ii TABLE OF CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 1 - GENERAL DESCRIPTION

1-1 OVERVIEW
DAS/Detector subsystem consists of Solid State Detector and Data Acquisition System (DAS) that are located on
the rotationary Gantry. X-ray data acquired by the detector is converted to light, then to electrical signal in the
detector and then sent to DAS. The DAS digitizes, serializes and performs offset corrections on the signal and sends
to the Operator Console for image reconstruction.

SECTION 1 - GENERAL DESCRIPTION


The X-ray exposure through the scanning object is detected by these modules, converted to light, then converted
to electrical signal and sent to to the DAS. The DAS digitizes and sends this data by slipring and ethernet cable to
the Operator Console, where it is processed for image reconstruction.

Illustration 1-1 X-ray Exposure and Data Flow Block Diagram

OBJECT
X-RAY

DATA DATA
X-RAY DETECTOR ACQUISITION DAS Data PROCESSING
TUBE SYSTEM UNIT

Illustration 1-2 X-ray Exposure and DAS/Detector

X-RAY TUBE

X-RAY BEAM

DETECTOR
DAS

DAS/DETECTOR 1-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

1-2 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 2 - DETECTOR

2-1 OVERVIEW
The detector consists of housing, collimator and detector modules. The detector module is made of scintillator
(lumex crystal), photo-diode and connector. When the X-ray exposure is performed, the X-ray enters through the
collimator and hits the scintillator. The scintillator is made of a material that emits light when hit by X-ray. The
scintillator is in contact with photo-diode so that the light generated by X-ray is converted into electrical current. The
generated current is sent to the DAS (Data Acquisition System) to be processed.

SECTION 2 - DETECTOR
2-2 SPECIFICATIONS
Distance from Focus to ISO: 541 mm.
Distance from Focus to Detector surface: 949.075 mm.
Detector Channel Pitch: 1.15 mm.
Detector Channel angle: 0.06946 degree.
View Number:
• Offset: 256 views
• Active: 972 views
Channel Distribution:

Single Slice Dual Slice


Slice - A Slice B Slice
Active Channel 685 685 685
Reference Channel 3+3 3+3 3+3
QCAL Channel 3 3 3

DAS/DETECTOR 2-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-3 HILIGHT DETECTOR MODULE


The module consists of Scintillator cells, Collimator, Photo Diode, Reflex Films and Connector (illustration 2-1 and
2-2). The X-ray enters through the collimator (tungsten plates) and is absorbed by the Scintillator. The Scintillator
is made of Lumex Crystal, the material that emits light when absorbs X-ray radiation. The Scintillator is optically
coupled with Photo Diode which converts light to electrical current. The Photo Diode is electrically connected to
DAS through the connector.

Illustration 2-1 DETECTOR MODULE (Single Slice)

COLLIMATOR (IN THE


DETECTOR HOUSING)

X-RAY

HILIGHT MODULE
LUMEX CRYSTAL

LIGHT

PHOTO DIODE

CONNECTOR (TO DAS)

2-2 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-3 HILIGHT DETECTOR MODULE (CONTINUED)

Illustration 2-2 DETECTOR MODULE (DUAL SLICE)

COLLIMATOR (IN THE


DETECTOR HOUSING)

X–RAY

SECTION 2 - DETECTOR
HILIGHT MODULE LUMEX CRYSTAL

LIGHT

PHOTO DIODE

Slice B

Slice A

CONNECTOR (TO DAS)

DAS/DETECTOR 2-3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-4 CHANNEL DISTRIBUTION


The illustration 2-3 shows the channel distribution.

Illustration 2-3 DETECTOR CHANNEL DISTRIBUTION

Single Slice System Channel Distribution


J01 J05 J06 J07 J08 J44 J45 J46 J47 J51
1 MOD

1 MOD
1 MOD
1 MOD

1 MOD
1 MOD
Detector

1 MOD 17~19 20~32 33~35 36~48 49~51 52~61 62~64 65~640 641~656 657~672 673~688 689~704 1 MOD
CH CH CH CH CH CH CH CH CH CH CH CH

1~3CH 4~12CH NC REF REF REF


QCAL REF
NC NC
3CH 3 3 3 4CH 3CH 13+ 3 13+ 3 10 3 576CH 16 12+ 4 2+ 14 10 6 7 3 3 3

1:1 3:1 1:1 3:1 2:1 2:1 2:1 1:1 1:1 2:1 3:1 3:1 3:1 NC 3:1 1:1 3:1
DAS

CH: 3 1 3 1 8 8 5 3 576 14 2 8 2 1 1 3 1

Dual Slice System Channel Distribution


J01 J05 J06 J07 J08 J44 J45 J46 J47 J51
1 MOD

1 MOD
1 MOD
1 MOD

1 MOD
1 MOD
Detector

1 MOD 17~19 20~32 33~35 36~48 49~51 52~61 62~64 65~640 641~656 657~672 673~688 689~704 1 MOD
CH CH CH CH CH CH CH CH CH CH CH CH

1~3CH 4~12CH NC REF REF REF


QCAL REF
NC NC
3CH 3 3 3 4CH 3CH 13+ 3 13+ 3 10 3 576CH 16 12+ 4 2+ 14 10 6 7 3 3 3

1:1 3:1 1:1 3:1 2:1 2:1 2:1 1:1 1:1 2:1 3:1 3:1 3:1 NC 3:1 1:1 3:1
DAS

A Slice

1:1 3:1 1:1 3:1 2:1 2:1 2:1 1:1 1:1 2:1 3:1 3:1 3:1 NC 3:1 1:1 3:1
B Slice

CH: 3 1 3 1 8 8 5 3 576 14 2 8 2 1 1 3 1

3:1 Three detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the three detector channel current.

2:1 Two detector channels are connected in parallel into one DAS channel.
DAS channel current is the sum of the two detector channel current.
2-4 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-5 DAS/DETECTOR CONNECTOR

2-5-1 DAS/Detector Connector for Single Slice System


The detector module consists of 16 channels. The illustration 2-4 shows the flexible PWB that connects detector
modules to the DAS Converter Board and its pin configuration.

Illustration 2-4 DAS/DETECTOR CONNECTOR


Connector Side View

Connector Side View


d c b a
TO DETECTOR
1
2
24 PIN 3

SECTION 2 - DETECTOR
CONNECTOR 4
5
6

FLEXIBLE
PWB

Back Side View


a b c d
48 PIN
CONNECTOR 1
2
3
4
5
6
TO DAS 7
8
9
10
11
12
Back Side View

DAS/DETECTOR 2-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-5-2 DAS/Detector Connector for Dual Slice System


The detector module consists of 32 channels (A slice and B slice). The illustration 2-5 shows the flexible PWB that
connects detector modules to the DAS Converter Board and its pin configuration.

Illustration 2-5 DAS/DETECTOR CONNECTOR

48 PIN A B C D
CONNECTOR
1 2SG 2SG 2SG 2SG
D C B A
2 2SG 2SG 2SG 2SG
1
2 3 A–1ch B–1ch A–16ch B–16ch
3
4 4 A–2ch B–2ch A–15ch B–15ch
5
6 5 A–3ch B–3ch A–14ch B–14ch
7
6 A–4ch B–4ch A–13ch B–13ch
8
9
7 A–5ch B–5ch A–12ch B–12ch
10
FLEXIBLE 11
PWB 8 A–6ch B–6ch A–11ch B–11ch
12
9 A–7ch B–7ch A–10ch B–10ch

10 A–8ch B–8ch A–9ch B–9ch

11 FG FG FG FG

12 FG FG FG FG

Note
SG: Signal Ground
FG: Frame Ground

2-6 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-6 QCAL CHANNELS


The X-ray exposure can move along the Z-axis. It can be due to mechanical adjustment or to tube temperature (the
angle of the X-ray output can vary according to the temperature). There is three channels in the left most positioned
detector module that is used for Z-axis data correction. The outputs of the three channels are tied together so that
only one data is taken for three channels. The corresponding intensity according to the position of the X-ray
exposure along the Z-axis is detected (as can be seen at right side of the illustration 2-6), so that the system
recognizes the current X-ray position (along the Z-axis). There is a remaining data for Z-axis data correction in the
system. Therefore, the correction is made according to the data acquired by the Q-CAL channels.

Illustration 2-6 Q-CAL CHANNELS

Q–CAL CHANNELS

SECTION 2 - DETECTOR
A Slice X–RAY INTENSITY
A Slice
DETECTED AC-
X–RAY
CORDING TO THE X–RAY EXPOSURE
B Slice BEAM POSITION EXPOSURE
B Slice
ALONG THE Z–AXIS
Z
Q–CAL CHANNELS

2-7 AFTER GLOW


The Lumex crystal scintillator has a property represented in the illustration 2-7. The scintillator continues emitting
light even after the X-ray radiation is stopped. This phenomenon is called AFTER GLOW. This remaining scintillation
provides incorrect data, but the system performs the error correction to eliminate this portion.

Illustration 2-7 AFTER GLOW

X–RAY ON

scintillation

X–RAY OFF

DAS/DETECTOR 2-7
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-8 HEATER AND TEMPERATURE CONTROL


The detector comprises a heater unit. The heater keeps the detector temperature constant, what is needed for
detector accuracy. The heater is controlled by the Temperature Control Box which supplies power to the heater. The
detector also comprises a thermistor that feeds back the temperature information to the Temperature Control box.
Approximately three hours is needed after turning ON the power to the Temperature Control box to stabilize the
detector temperature. The temperature is controlled to 35  0.5 C. The illustration 2-8 shows its block diagram.

Illustration 2-8 TEMPERATURE CONTROL BLOCK DIAGRAM

Thermistor DETECTOR

Heater DC24V 60W

INPUT POW
ER 24V DC
Driver for
Heater

PS
PS (+12V -> +5V)
(+24V -> +12V)

(OP Amp CPU ROM


Coparato)

Switch LED indication

TEMPERATURE CONT BOX

2-8 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 3 - DATA ACQUISITION SYSTEM (DAS) - SINGLE

3-1 OVERVIEW
The DAS consists of the following boxes: RBB (Right Back Board), CBB (Center Back Board), and LBB (Left Back
Board). Each box contains the set of boards as follows:

• RBB - 1 Converter Board


• CBB - 3 Converter Boards
• LBB - 1 Converter Board

SECTION 3 - DATA ACQUISITION


The Illustration 3-1 shows the DAS boxes.

Illustration 3-1 DAS (RBB, CBB, LBB)

SYSTEM
RBB ASSY
LBB ASSY CBB ASSY

Co rd #6
DC

CONNECTER

Bo
nve
PA

a
(TO DETECTOR)
Boa verter
Con

rte
SS

rd #

r
Y

10

Co rd #
Boa verter

Bo
Co

nve 7
a
n
rd #

Board #8
Converter

rte

DAS CONTROL BOARD


r
9

The DAS channels number is as follows:

Active Channels 685


Reference Channels 3+3
QCAL Channels 3
NC Channels 26

DAS/DETECTOR 3-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-1 OVERVIEW (CONTINUED)


The illustration 3-2 shows DAS block diagram.
The data coming from the detector is taken by the Converter boards, sequentially channel by channel, digitized,
converted to parallel, then sent to the Operator Console. The DCP receives control and timing signals, and gener-
ates the time signals for data gathering. The DCP Board also generates the test pattern.

Illustration 3-2 DAS/Detector Structure

#45 #35 #34 #27 #26 #19 #18 #11 #10 #1

#10 #9 #8 #7 #6 Converter
Boards
POWER DC+/-5V Analog
SUPPLY
DC+5V
Digital
Data
Flow
Data Flow
DCP
BOARD
Control Signal

3-2 DAS/DETECTOR
REV 4

RBB
GE HEALTHCARE

JP1 1-1 JP4

JP1
Illustration 3-3

2-2

......
8-8 JP4

CBB

JP3 1-1 JP5


JP3

2-2

......

8-8
JP5

CN80
DAS Power Connection

15A
L 4 6 6
110V AC 5 7 7
N 6 8 8
+5V ANALOG
1 3 3

DAS/DETECTOR
ANALOG GND
2 4 4 LBB
-5V ANALOG
3 5 5
+5V DIGITAL
7 1 1
AC + DIGITAL GND
8 2 2
+/- 5 V COM
AC ANALOG -

AC +
+5V
DIGITAL
AC -
5306516-1EN

3-3
THEORY OF OPERATION
Brivo CT325
Brivo CT315

SECTION 3 - DATA ACQUISITION


SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

Illustration 3-4 DAS Signal Connection

Collimator Driver

Collimator
LBB ASSY
FC45
J51 CN45
FC44
OGP

J47 CN44

DCP
Converter
Board #10

FC35
J38 CN35
JS4

CBB ASSY
JS3
FC34
Converter J37 CN34
Board #9

DETECTOR
Converter
Board #8

Converter FC11
Board #7 J14 CN11
JS2

RBB ASSY
JS1
FC10
J13 CN10

Converter
Board #6

FC2
J5 CN2
FC1
J1 CN1

3-4 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-2 DATA FLOW


The illustration 3-5 shows the data flow. The Detector generates an electrical current proportional to X-ray intensity.
The electrical current is converted to voltage signal in the Converter board. The voltage signal is amplified to an
appropriate level, converted to digital data (serial). The data is then sent to to DCP board where it is converted back
to serial, conditioned and then sent to Operator Console through the RF Slipring.

Illustration 3-5 DAS block diagram

DAS Offset
Trim
DAC
FPGA

SECTION 3 - DATA ACQUISITION


Electrical
Current Fold
DETECTOR DCP Board
Control

SYSTEM
3-3 DCP BOARD

3-3-1 DCP Functions


The DCP board performs the following functions:
• Interfaces with the OGP for Rx reception and scan completion via RS422 bus.
• Sets up gain and offset trim and controls the converter cards, via the converter CAN interface.
• Receives triggers and starts acquisitions with the converter cards.
• Performs serial-to-parallel conversion on data streams from the converter cards, does parity checking
on the data, and runs it through a translation table for view data ordering.
• Adds Forward Error Correction (FEC) to the channel data and sends it across the slip ring to the OC via
the high speed serial data interface.
• It provides the A/D converter chip, it acquires the kV and mA values for each scan.
• Monitors the DAS subsystem for various faults.
• Communication with Detector Heater Control Board to monitor the Detector Temperature.
• Trigger jitter detection.

DAS/DETECTOR 3-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-3-2 DCP Interface


The DCP is the main control board for Data Acquisition System (DAS). It is applicable for DAS converter cards. It
packages all the converter cards data into a single high speed serial data stream to OC, compatible with the receive
function on the DASIFN at 95/110 Mbps slip ring communication. Illustration 3-6 shows the interface of DCP.

Illustration 3-6 DCP Interface

SDCOM/+-
DSCOM/+- CAN_LO
CAN_HI
DASTRIG/+-
OGP GPLS2/+- DATA_ODD/+-
XRON/+- DATA_EVEN/+-
AC/OFF/+- STRIG/+- DAS
RESET/+- DCP SH_CK/+-
DASENABLE/+- CN_FLT CVTR
CN_RST
X-PROG
RAW+5VA W-PRTCT
RAW-5VA IREF.EXT
POWER RAW+5VL AGND
SUPPLY AGND
LGND

Fibre Optical
Cable
Slip Ring
KV Signal
From JEDI
0~10V mA Signal

3-6 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-3-2 DCP Interface (Continued)

Inputs
From Converter Boards:
• Inner Row Serial Data Streams (Differential Pairs)
• Converter CAN Bus communication for status, faults, chassis temperature readings, and serial number
information (CAN)
• Converter Board Fault Line (TTL)
From OGP:
• Input View Triggers (Optical Isolated)

SECTION 3 - DATA ACQUISITION


• Active/Offset (Optical Isolated)
• High Voltage on to X-ray tube (Optical Isolated)
• DAS enable signal (Optical Isolated)

SYSTEM
• Gplus2 signal (Optical Isolated)
• OGP reset to DCP (Optical Isolated)
• TG CMD send/receive serial port (Optical Isolated)
From Generator (JEDI):
• KV & MA analog signals
From Detector Heater Control Board:
• Serial port communication protocol on differential pair signals. (RS422)
From Backplane:
• Power Supply

Outputs
To Converter Boards:
• Shift clock (Differential Pairs)
• Trigger signal (Differential Pairs)
• Converter CAN Bus communication for control information (CAN Differential Pairs)
• Reset signal (TTL)
• External reference current (Analog)
To OGP:
• TG CMD transmits serial port (RS422)

DAS/DETECTOR 3-7
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-3-2 DCP Interface (Continued)


To Detector Heater Control Board:
• Serial port communication protocol differential pair output signals. (RS422)
To Slip-ring (OC):
• High-speed serial data stream containing the view data with embedded FEC (High speed optical fiber
link).

Illustration 3-7 DCP board block diagram

GDAS DCP CLK


32MHz Fiber
CARD ASSY Reset

ARM
ARM RAM
JTAG RAM
AT91
TG com
Flash-ROM
PLL CLK
30MHz
FEC
TAXI
Pulse/
status FPGA CY9689A
Cyclone

CAN
CFG JTAG

ADC LVDS LVDS Iref


AD977 Mx9129 Ds032 Ref200
+5V_Logic
+5V_Analog
-5V_Analog
KV/mA

DHCBCollimator OGP JEDI CVTR Power Supply


Left Backplane Board

3-8 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-4 CONVERTER BOARD


The Converter board performs the following functions:
• Conversion of electrical current signals from the detector to voltage signals.
• Signal level conditioning
• Analog to Digital conversion
• Generation of the timing signals for data gathering
The signal from the detector with the level proportional to the X-ray intensity is fed to the Converter board. The
signal go through the electrical current to voltage signal conversion and conditioning circuits then is fed to the
Analog/Digital(A/D) converter. The A/D converts analog signal to 16 bits digital serial data.

Illustration 3-8 Preamplifier Architecture

SECTION 3 - DATA ACQUISITION


SYSTEM

DAS/DETECTOR 3-9
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

3-4 CONVERTER BOARD (CONTINUED)

Illustration 3-9 Converter Board Block Diagram

2
IREF_INT VOLTAGE
REFERENCE

IREF_EXT
GDAS ASIVS (2)
..
Chan_1
Odd # Chan
Chan_127
. Anti_Alias
Filter CHARGE TO
DIGITAL
CONVERTERS
..
Chan_2
Even # Chan
Chan_128
. (2 x 64 CH.) D0:D9
10

ASIC/FPGA INTERFACE

10
2.5 VOLT BANK

LVDS D_OUT_ODD +/-


Differential D_OUT_EVN +/-
D_IN_ODD +/- LVDS Drivers
D_IN_EVN +/- 4
Differential 3.3 GAB 3.3
S_TRIG +/-
SH_CK +/-
Receivers V V
O (FPGA or ASIC) O
L L CV_FLT*
T T

7 1.5V Core
CVB_ADDR B 2.5, 3.3 VIO B
A A 2
N N I C I/F 2
I C
K K TEMPERATURE
CAN_HI CAN 2 SENSOR
CAN_LO I/F
3.3 VOLT BANK

DSP INTERFACE

Flash Boot
SDRAM SHARC DSP
Prom
CPLD
INTERFACE
CV_RST CPLD 14 Pin JTAG
HEADER

3-10 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 4 - DATA ACQUISITION SYSTEM (DAS)_DUAL

4-1 OVERVIEW
The DAS consists of the following boxes: RBB (Right Back Board), CBB (Center Back Board), and LBB (Left Back
Board). Each box contains the set of boards as follows:
• RBB: 2 Converter Boards
• CBB: 6 Converter Boards
• LBB: 2 Converter Boards

SECTION 4 - DATA ACQUISITION


The Illustration 4-1 shows the DAS boxes.

Illustration 4-1 DAS (RBB, CBB, LBB)

SYSTEM
RBB ASSY
LBB ASSY CBB ASSY

Co rd #6
rter
CONNECTER
DC

nve
PA

(TO DETECTOR)

a
Boa verter

Co r d # 1
rter

B o
5
SS

rd #

nve
Boa verter

a
10
Con

o
Y

B
rd #
Con

Co rd #7
rter
Boa verter
4

nve
Co rd #2
rd #

rter

a
Boa verter

Bo
9

DAS CONTROL BOARD


Con

nve
rd #

Converter

Converter

a
Board #3

Board #8

o
Con

The DAS channels number is as follows:

Slice A Slice B Slice


Active Channel 685 685
Reference Channel 3+3 3+3
QCAL Channel 3 3
GND Channel 26 26

DAS/DETECTOR 4-1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-1 OVERVIEW (CONTINUED)


The illustration 4-2 shows DAS block diagram.
The data coming from the detector is taken by the Converter boards, sequentially channel by channel, digitized,
converted to parallel, go through offset correction, then sent to the Operator Console. The DCP Card is an interface
with OGP (On Gantry Processor) receiving control and timing signals, and generates the timing signals for data
gathering. The DCP Card performs offset correction and generates the test pattern.

Illustration 4-2 DAS/Detector Structure

#45 #35 #34 #27 #26 #19 #18 #11 #10 #1

#5 #10 #4 #9 #3 #8 #2 #7 #1 #6 Converter
POWER
Boards
DC+/-5V Analog
SUPPLY
DC+5V
Digital

Data Flow

Data Flow
DCP Data Flow
Card

Control Signal

4-2 DAS/DETECTOR
4-1
REV 4

RBB

JP1 1-1 JP4

JP1
GE HEALTHCARE

2-2
Illustration 4-3

......
8-8

JP4
CBB
OVERVIEW (CONTINUED)

JP3 1-1 JP5

JP3
2-2

......
8-8
JP5

CN80
15A
L 6
DAS Power Connection

4 6
110V AC 5 7 7
N 6 8 8
+5V ANALOG
1 3 3
ANALOG GND
2 4 4 LBB
-5V ANALOG

DAS/DETECTOR
3 5 5
+5V DIGITAL
7 1 1
AC + DIGITAL GND
8 2 2
+/- 5 V COM
AC ANALOG -

AC +
+5V
DIGITAL
AC -
5306516-1EN

4-3
THEORY OF OPERATION
Brivo CT325
Brivo CT315

SECTION 4 - DATA ACQUISITION


SYSTEM
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-1 OVERVIEW (CONTINUED)

Illustration 4-4 DAS Signal Connection


Collimator Driver

Collimator

LBB ASSY
FC45
J51 CN45
FC44
OGP

J47 CN44

DCP
Converter
Board #5

Converter
FC35
Board #10 J38 CN35
JS4

CBB ASSY Converter


JS3
Board #4 FC34
J37 CN34
4onverter
Board #9

DETECTOR
Converter
Board #3
Converter
Board #8
Converter
Board #2
FC11
CN11
Converter J14
Board #7 JS2

RBB ASSY
JS1
FC10
J13 CN10

Converter
Board #1

Converter
FC2
Board #6 J5 CN2
FC1
J1 CN1

4-4 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-2 DATA FLOW


The illustration 4-5 shows the data flow. The Detector generates an electrical current proportional to X-ray intensity.
The electrical current is converted to voltage signal in the Converter board. The voltage signal is amplified to an
appropriate level, converted to digital data (serial). The data is then sent to DCP Card where it is converted back to
serial, conditioned and then sent to Operator Console through the RF Slipring.

Illustration 4-5 DAS block diagram

DAS Offset
Trim
DAC

SECTION 4 - DATA ACQUISITION


FPGA

Electrical
Current Fold
DETECTOR DCP Board
Control

SYSTEM
4-3 DCP CARD

4-3-1 DCP Functions


The DCP Card performs the following functions:
• Interfaces with the OGP for Rx reception and scan completion via RS422 bus.
• Sets up gain and offset trim and controls the converter cards, via the converter CAN interface.
• Receives triggers and starts acquisitions with the converter cards.
• Performs serial-to-parallel conversion on data streams from the converter cards, does parity checking
on the data, and runs it through a translation table for view data ordering.
• Adds Forward Error Correction (FEC) to the channel data and sends it across the slip ring to the OC via
the high speed serial data interface.
• It provides the A/D converter chip, it acquires the kV and mA values for each scan.
• Monitors the DAS subsystem for various faults.
• Communication with Detector Heater Control Board to monitor the Detector Temperature.
• Trigger jitter detection.

DAS/DETECTOR 4-5
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-3-2 DCP Interface


The DCP is the main control board for Data Acquisition System (DAS). It is applicable for GDAS converter cards.
It packages all the converter cards data into a single high speed serial data stream to OC, compatible with the re-
ceive function on the DASIFN at 95/110 Mbps slip ring communication. Illustration shows the interface of DCP.

Illustration 4-6 DCP Interface

SDCOM/+- CAN_LO
DSCOM/+- CAN_HI
DATA 2_ODD/+-
DASTRIG/+- DATA 2_EVEN/+-
OGP GPLS2/+- DATA 1_ODD/+-
XRON/+- DATA 1_EVEN/+-
AC/OFF/+- STRIG/+- DAS
RESET/+- DCP SH_CK/+-
DASENABLE/+- CN_FLT CVTR
CN_RST
X-PROG
RAW+5VA W-PRTCT
RAW-5VA IREF.EXT
POWER RAW+5VL AGND
SUPPLY AGND
LGND

Fibre Optical
Cable
Slip Ring
KV Signal
From JEDI
0~10V mA Signal
BW LMT

FW LMT
FW+

FW-

BW+

BW-

Collimator & Driver

4-6 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-3-2 DCP Interface (Continued)

Inputs
From Converter Boards:
• 2 Inner Row Serial Data Streams (Differential Pairs)
• Converter CAN Bus communication for status, faults, chassis temperature readings, and serial number
information (CAN)
• Converter Board Fault Line (TTL)
From OGP:
• Input View Triggers (Optical Isolated)

SECTION 4 - DATA ACQUISITION


• Active/Offset (Optical Isolated)
• High Voltage on to X-ray tube (Optical Isolated)
• DAS enable signal (Optical Isolated)

SYSTEM
• Gplus2 signal (Optical Isolated)
• OGP reset to DCP (Optical Isolated)
• TG CMD send/receive serial port (Optical Isolated)
From Collimator:
• Collimator forward and backward position limit feedback (TTL)
From Generator (JEDI):
• KV & MA analog signals
From Backplane:
• Power Supply

Outputs
To Converter Boards:
• Shift clock (Differential Pairs)
• Trigger signal (Differential Pairs)
• Converter CAN Bus communication for control information (CAN Differential Pairs)
• Reset signal (TTL)
• External reference current (Analog)
To OGP:
• TG CMD transmits serial port (RS422)

DAS/DETECTOR 4-7
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-3-2 DCP Interface (Continued)


To Collimator:
• Beam tracking control pulse output (Differential pairs)
To Slip-ring (OC):
• High-speed serial data stream containing the view data with embedded FEC (High speed optical fiber
link).

Illustration 4-7 DCP Card block diagram

GDAS DCP CLK


32MHz Fiber
CARD ASSY Reset

ARM
ARM RAM
JTAG RAM
AT91
TG com
Flash-ROM
PLL CLK
30MHz
FEC
TAXI
Pulse/
status FPGA CY9689A
Cyclone

CAN
CFG JTAG

ADC LVDS LVDS Iref


AD977 Mx9129 Ds032 Ref200
+5V_Logic
+5V_Analog
-5V_Analog
KV/mA

DHCBCollimator OGP JEDI CVTR Power Supply


Left Backplane Board

4-8 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-4 CONVERTER BOARD


The Converter board performs the following functions:
• Conversion of electrical current signals from the detector to voltage signals.
• Signal level conditioning
• Analog to Digital conversion
• Generation of the timing signals for data gathering
The signal from the detector with the level proportional to the X-ray intensity is fed to the Converter board. The
signal go through the electrical current to voltage signal conversion and conditioning circuits then is fed to the
Analog/Digital(A/D) converter. The A/D converts analog signal to 16 bits digital serial data.

SECTION 4 - DATA ACQUISITION


Illustration 4-8 Preamplifier Architecture

SYSTEM

DAS/DETECTOR 4-9
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

4-4 CONVERTER BOARD (CONTINUED)

Illustration 4-9 Converter Board Block Diagram

2
IREF_INT VOLTAGE
REFERENCE

IREF_EXT
GDAS ASIVS (2)
..
Chan_1
Odd # Chan
Chan_127
. Anti_Alias
Filter CHARGE TO
DIGITAL
CONVERTERS
..
Chan_2
Even # Chan
Chan_128
. (2 x 64 CH.) D0:D9
10

ASIC/FPGA INTERFACE

10
2.5 VOLT BANK

LVDS D_OUT_ODD +/-


Differential D_OUT_EVN +/-
D_IN_ODD +/- LVDS Drivers
D_IN_EVN +/- 4
Differential 3.3 GAB 3.3
S_TRIG +/-
SH_CK +/-
Receivers V V
O (FPGA or ASIC) O
L L CV_FLT*
T T

7 1.5V Core
CVB_ADDR B 2.5, 3.3 VIO B
A A 2
N N I C I/F 2
I C
K K TEMPERATURE
CAN_HI CAN 2 SENSOR
CAN_LO I/F
3.3 VOLT BANK

DSP INTERFACE

Flash Boot
SDRAM SHARC DSP
Prom
CPLD
INTERFACE
CV_RST CPLD 14 Pin JTAG
HEADER

4-10 DAS/DETECTOR
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

X-RAY GENERATOR

TABLE OF CONTENTS

SECTION PAGE

SECTION 1 - GENERAL DESCRIPTION - I. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1


1-1 GLOSSARY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 JEDI HIGH LEVEL BOLCK DIAGRAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

SECTION 2 - GENERAL DESCRIPTION - II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Table of Content
2-1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 STANDARD FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-3 APPLICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-4 ARCHITECTURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-4-1 Kernel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-4-2 Options Depending on the Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-4-3 A Packaging Architecture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

SECTION 3 - TYPICAL SIGNALS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1


3-1 JEDI TYPICAL SIGNALS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

TABLE OF CONTENTS i
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

ii TABLE OF CONTENTS
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 1 - GENERAL DESCRIPTION - I

1-1 GLOSSARY
Glossary of terms used in this document:

Term Definition
Automatic Brightness Control. Regulation loop which makes the measured
ABC
brightness equal to brightness demand

SECTION 1 - GENERAL DESCRIPTION-I


Automatic Exposure Control. Exposure cut off technique which uses the
AEC
brightness signal to cut the exposure
CAN Controller Area Network. A network used for localized control.
Control Processor Unit. Microprocessor and peripherals which run the
CPU
software/firmware
EPLD Erasable Programmable Logic Device.
Electro Magnetic Compatibility. The EMC function prevents the generator
EMC
from polluting the power source.
Field Programmable Gate Array. It is programmed by the CPU core after
FPGA the reset and handles all the exposure control logic including the system
interface real–time lines.
High voltage variations due to inverter current pulses. Typically a few
HV Ripple
percent.
Software or hardware function which handles the state of a system and
State Machine
authorize to go to the next state upon reception of specific events.
IGBT Insulated gate bipolar transistor. A type of power switch
Ilp HV power inverter parallel resonant current; current in the parallel inductor
Ilr HV power inverter serial resonant current; current in the serial inductor.
MOS Metal Oxide Semiconductor. A type of power switch
OGP On Gantry Processor. Unit which drives the generator in CT systems
PDU Power Distribution Unit
RMS Root Mean Square

X-RAY GENERATOR 1-1


1-2
1-2
REV 4

JEDI GENERATOR / TWIN FUNCTIONAL ARCHITECTURE


GE HEALTHCARE

System Control Bus


Illustration 1-1

System
Interface

Low Voltage
kV Control Heater heater supply Rotation
Power
JEDI HIGH LEVEL BOLCK DIAGRAM

bus
Supply

DC Bus
AC/DC

X-RAY GENERATOR
rotation phases
JEDI Generator / TWIN Functional Architecture

High Voltage High Voltage HV Cables


EMC Inverter Tank
Filter

X-Ray Tube 1
1 phase
3 phase power
tube cooling
input
input
5306516-1EN
THEORY OF OPERATION
Brivo CT325
Brivo CT315
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 2 - GENERAL DESCRIPTION - II

2-1 INTRODUCTION
Jedi is the engineering name for a family of compact high frequency X-Ray generators. This generator family covers
a wide range of applications from mobile equipment up to vascular systems:
JEDI 12-25 kW: Mobile applications

SECTION 2 - GENERAL DESCRIPTION-II


JEDI 24-48 kW: CT applications
JEDI 32-50 kW: RAD applications
JEDI 50-65-80 kW: RF applications
JEDI 100 kW: VASCULAR applications

2-2 STANDARD FEATURES


Jedi is a family of 150 kV generators operating from 12 kW up to 100 kW for all the major radiological, fluoroscopic
and CT applications. The family handles 1 ms to continuous exposures with tube currents ranging from 0mA up to
1000 mA.
The generators feature the very latest technology available:
• Constant potential independent of line voltage variations
• Power generation by a high-frequency converter (High voltage ripple: 40 kHz-140 kHz)
• Distributed micro-processor controlled functions (CAN bus)
Other features include:
• Single phase, three phase or battery power source
• Very low kV and mA ripple, excellent accuracies and dose reproducibility
• Compatible with a wide range of tubes, high speed or low speed, can supply up to 3 different tubes.
Thermal load interactive integrator ensuring optimum use of the heat protection curve of the x-ray tube
• Available in various packaging configurations: gantry, under-table, cabinet
• Serviceability: high reliability, fast installation (no generator calibration), application error codes ensure
fast troubleshooting
• Meets CE marking (and in particular EMC), IEC, UL, CSA, MHW regulations (if required)
• Optional pulsed fluoroscopy

X-RAY GENERATOR 2-1


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-3 APPLICATIONS

Vascular/
RAD Surgery RF CT
Cardiac
3 Points Mode x x x
2 Points Mode x x x
1 Point Mode x x x
0 Point Mode x x
AEC x x x
Tomography x x
AET x x
Continuous/Pulsed Fluoroscopy x x x
Rad Exposures x
Cinema 30 fr/s x
Cinema 90 fr/s x
ABC x x
Variable mA Scans x
Low mA Fluoroscopy x

Legend:
• AEC - Automatic Exposure Control
• AET - Automatic Tomographic Exposure
• ABC - Automatic Brightness Control in fluoroscopy

2-2 X-RAY GENERATOR


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-4 ARCHITECTURE
The Jedi family is composed of 3 elements:

2-4-1 Kernel
• High voltage chain composed of kV control, HV power inverter and HV tank
• Anode rotation function
• Tube filaments heater function
• Control bus for communication between the functions

SECTION 2 - GENERAL DESCRIPTION-II


• DC bus for power distribution to each function
• Input voltage to DC conversion: AC/DC function
• Low voltage power supply
• Application software, running on the kV control board
These functions are the Jedi core. They are present in all versions of the generator.
A function can be unique for all products, or can have several different releases based on product specification.
Examples:
• The anode rotation function is available in 2 releases:
- low speed rotation for applications where the tube has a max rotation of 3000 rpm
- high speed/low speed rotation for applications where at least one of the tubes can use
8000-10000 rpm
• The control bus is unique.

2-4-2 Options Depending on the Application


A System Interface which can be:
• CT interface
• RAD interface (console interface, room interface, AEC management present or not)
• ATLAS interface
EMC function
Grid function (RF, vascular)
Bias function (RF, vascular)
Tube management (2 tubes or 3 tubes option)

X-RAY GENERATOR 2-3


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

2-4-3 A Packaging Architecture


The packaging architecture consists in a set of boxes which can be put together in several ways to make Jedi fit
either in a cabinet, or a console foot, or a table foot. The boxes can also be split in 2 units distant of several meters
(example: CT gantry).
The boxes normally consist of the following:

Auxiliaries Box
Rotation function
Heater function
Low voltage power supply (which can also be in the AC/DC box)
This box is always present.

Power Box
HV tank
HV power inverter
kV control
System interface (for the less complex system interface)
This box is always present.

AC/DC Box
MC filter (optional)
AC/DC function
Low voltage power supply (which can also be in the auxiliaries box)
This box is always present.

2-4 X-RAY GENERATOR


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

Illustration 2-1 Jedi Generator / Packaging Architecture

Inverter Assembly
Dual Snub
AC/DC
Tube Board
Gate Command
HV Board
+ -

KV Measure

SECTION 2 - GENERAL DESCRIPTION-II


HV Tank Rectifier
Block

CT
Inter KV Control Board
face
EMC Filter

POWER BOX AC/DC BOX

LVPS Rotation Board


Heater Board Board

Rotation Rotation
Capacitor Capacitor

AUXILIARIES BOX

X-RAY GENERATOR 2-5


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

2-6 X-RAY GENERATOR


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

SECTION 3 - TYPICAL SIGNALS

3-1 JEDI TYPICAL SIGNALS


The following table describes the main signals of the generator. For each of them, typical values are presented for
the main applications:

Signal/Application PoSpeed AII/EII/FII Values


Anode Rotation:
Phase P current during high speed run 2 A peak / 144 Hz

SECTION 3 - TYPICAL SIGNALS


Phase A current during high speed run 2 A peak / 144 Hz
Filament Drive:
Inverter current in standby 2.5 A 3.3 A peak / 35 kHz
Heater DC input voltage in standby 160V
kV Control:
DC bus measure at 400 VAC 560 V
ILR at full power at 400 VAC 300 A peak / 50 kHz
ILR at min power at 400 VAC 80 A peak / 25 kHz
ILP at full power at 400 VAC 150 A peak / 50 kHz
ILP at min power at 400 VAC 150 A peak / 25 kHz

X-RAY GENERATOR 3-1


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION

3-2 X-RAY GENERATOR


Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

APPENDIX A - SYMBOLS AND CLASSIFICATION

Symbol Publication Description


417-5032 Alternating Current

335-1 Three-phase Alternating Current


3

335-1 Three-phase Alternating Current with neutral

APPENDIX A - SYMBOLS AND


conductor
3N

CLASSIFICATION
Direct Current

417-5019 Protective Earth (Ground)

348 Attention, consult ACCOMPANYING DOCUMENTS

417-5008 OFF (Power: disconnection from the mains)

417-5007 ON (Power: connection to the mains)

Warning, HIGH VOLTAGE

Emergency Stop

APPENDIX 1
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

Symbol Publication Description

Type B

417-5339 X-ray Source Assembly Emitting

417-5009 Standby

Start

Table Set

Abort

Intercom

(on Operator Console)


Power On: light on
Standby: light off

Warning sign :
Radiation of Laser Apparatus

2 APPENDIX
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

Symbol Description

Microphone (Mic)

Contrast

Brightness

APPENDIX A - SYMBOLS AND


CLASSIFICATION
30C System storage prior to installation:
Maintain storage temperature between 0 C and +30 C

0C

System storage prior to installation:


Maintain non-condensing storage
humidity below 70%
DO NOT store system longer than 90 days

Humidity 0-70%
Excluding Condensation

Air Pressure System storage and shipment:


Maintain Air Pressure between 700 and 1060hPa
700-1060hPa

APPENDIX 3
Brivo CT315
Brivo CT325
GE HEALTHCARE THEORY OF OPERATION
REV 4 5306516-1EN

CLASS 1 EQUIPMENT
Any permanently installed equipment containing operator or patient accessible surfaces must provide backup
protection against electric shock,in case the BASIC INSULATION fails. In addition to BASIC INSULATION,Class1
equipment contains a direct connection to a PROTECTIVE(EARTH) CONDUCTOR which prevents shocks when a
person touches a broken piece of equipment or touches two different equipment surfaces simultaneously.

TYPE B EQUIPMENT
CLASS I, II, or III EQUIPMENT or EQUIPMENT with INTERNAL ELECTRICAL POWER SOURCES provide an
adequate degree of protection against electric shock arising from (allowable) LEAKAGE CURRENTS or a
breakdown in the reliability of the protective earth connection.

ORDINARY EQUIPMENT
Enclosed EQUIPMENT without protection against the ingress of water.

OPERATION 0f EQUIPMENT
CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Operation in which EQUIPMENT is connected continuously to the SUPPLY MAINS.The stated permissible loading
time is so short that the long term on-load operating temperature is not attained.The ensuing interval in loading is,
however, not sufficiently long for cooling down to the long term no-load operating temperature.
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR or WITH
OXYGEN or NITROUS OXIDE

CLEANING
The system is NOT WATERPROOF. It is NOT designed to protect internal components against the ingress of
liquids.Clean external system surfaces(Gantry,table,consoles and accessories)with a soft cloth dipped in hot water
and wrung DAMP/DRY. (NOT dripping!) IF NECESSARY, use only mild (dish washing liquid) soap to remove dirt.

Avoid damage to equipment! Some îspray and wipeîcleaners etch and permanently cloud
clear plastic surfaces!! Use only warm water and mild liquid soap to clean surfaces.

4 APPENDIX
www.gehealthcare.com

Imagination at work

You might also like