Professional Documents
Culture Documents
1. Connect the 32-channel scanner (Falcon Senior) to Falcon 2000 instrument box
with 100-ft cable.
2. Connect the Instrument box to the Computer having ‘Tank 32E’ software
through custom serial port cable.
3. Connect 110 V AC power supply to the Instrument box and to the Computer.
5. Adjust the gap between the scanner and floor as close as possible.
10. Select Gain, so that the ‘pickup strength’ yellow bar graph reads at least half
height.
11. Select Frequency lower than 12 Hz for plate thickness ¼” and above, and
viceversa.
14. Calibrate the instrument after placing a sheet of FRP of same coating thickness
over the Standard plate having same thickness and metallurgical properties and
artificial defects machined over it
18. Start scanning, not exceeding the speed 6” per second and also maintaining the
overlap.
19. Mark the area(s), which show indications above the threshold level on the plate
for further investigations.
21. Prepare a test report indicating the level of corrosion on each plate.
22. Clean the probe, coil and the instrument and store it in the appropriate location.