Laser Light Localizer


SP

Installation Instructions
Option

Installation Instructions

Valid for:
© Siemens, 2004

SIREMOBIL Compact
ARCADIS Varic

03779001

© Siemens, 2004
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: RXR2-130.812.01.02.02 English


Replaces: RXR2-130.812.01.01.02 Doc. Gen. Date: 01.05
CS SD 24
n.a.

2004
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2004“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

Laser Light Localizer RXR2-130.812.01.02.02 Page 2 of 20 © Siemens, 2004


01.05 CS SD 24
Table of Contents 3
0Table of Contents

1 _______ Safety Information _______________________________________________ 4

Protection against laser beams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


Labels for U.S.A.: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Labels for all other countries: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 _______ Prerequisites and Preparation _____________________________________ 7

Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Required parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Required aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Removing the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

3 _______ Installation, Wiring, and Programming _____________________________ 10

Wiring on the SIREPHOS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10


Installing the localizer unit on the SIREPHOS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1a. Changing the configuration (only SIREMOBIL Compact): . . . . . . . . . . . . . . . . . . . . 11
1b. Changing the configuration (only ARCADIS Varic): . . . . . . . . . . . . . . . . . . . . . . . . . 12

4 _______ Functional Test and Final Steps___________________________________ 13

Functional test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Attaching the cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Attaching the cover panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Installing the I.I. attachment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Affixing the identification labels for the U.S.A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Affixing the identification labels for Great Britian and other countries . . . . . . . . . . . . . . 17

5 _______ Changes to Previous Version _____________________________________ 19

© Siemens, 2004 RXR2-130.812.01.02.02 Page 3 of 20 Laser Light Localizer


01.05 CS SD 24
4 Safety Information
1-
Protection against laser beams
1Safety Information

This product contains two Class 2 lasers (USA: Laser class 2).
Observe the safety information in ARTD-002.731.03.xx.xx.
When working with the laser light localizer, do no look directly into the laser beam.

Labels for U.S.A.: 0

Fig. 1: Laser-Klasse-Piktogram

Fig. 2: Certification label Fig. 3: Identification label

Fig. 4: Label Fig. 5: Laser aperture label

Laser Light Localizer RXR2-130.812.01.02.02 Page 4 of 20 © Siemens, 2004


01.05 CS SD 24
Safety Information 5

Labels for all other countries: 0

Warning label according to IEC 825

Fig. 6: Warning label according to IEC 825


(use local language)

Fig. 7: Warnschilder nach IEC 825

© Siemens, 2004 RXR2-130.812.01.02.02 Page 5 of 20 Laser Light Localizer


01.05 CS SD 24
6 Safety Information

Identification label:

Fig. 8: Identification label

Fig. 9: Label

Laser Light Localizer RXR2-130.812.01.02.02 Page 6 of 20 © Siemens, 2004


01.05 CS SD 24
Prerequisites and Preparation 7
2-
Note
2Prerequisites and Preparation

These installation instructions only apply to retrofitting SIREMOBIL Compact or ARCADIS


Varic systems with a laser light localizer.

© Siemens, 2004 RXR2-130.812.01.02.02 Page 7 of 20 Laser Light Localizer


01.05 CS SD 24
8 Prerequisites and Preparation

Required parts 2.1

Laser light localizer complete for ARCADIS Varic 80 79 720


Laser light localizer complete for SIREMOBIL Compact 37 79 001 G5429

Fig. 10: Overview of required parts


Tab. 1 List of required parts

1 Localizer unit
2 I.I. attachment 23 cm (9") I.I. (only ARCADIS Varic)
2 I.I. attachment 17 cm (7") and 23 cm (9") I.I. (only SIREMOBIL Compact)
4 SIREPHOS cover
5 Sterile covers
7 Adjustment device
8 Warning labels according to IEC 825 (text and symbol)
8 Identification label
8 For U.S.A.: “Laser class 2” warning logotype and DHHS certification label
8 Various small parts

Laser Light Localizer RXR2-130.812.01.02.02 Page 8 of 20 © Siemens, 2004


01.05 CS SD 24
Prerequisites and Preparation 9

Required aids 0

• Service PC (only SIREMOBIL Compact)


• Toolbox (including Allen key)

Removing the cover 0

• Disconnect the system power plug.


• Remove the cover plate from the basic unit.
• Remove the SIREPHOS cover by carefully loosening the bonding surfaces on the long
sides of the cover.

© Siemens, 2004 RXR2-130.812.01.02.02 Page 9 of 20 Laser Light Localizer


01.05 CS SD 24
10 Installation, Wiring, and Programming
3-
Wiring on the SIREPHOS
3Installation, Wiring, and Programming

• Loosen and remove the cover from the tube collimator.


• Route the spare leads (white and gray) up and out through the back.
• Replace the cover of the tube collimator.

Installing the localizer unit on the SIREPHOS 0

Fig. 11: SIREPHOS localizer unit


Pos. 5 Screws
Pos. 6 Threaded holes
Pos. 7 Solder terminal
Pos. 8 Cable ties

• Loosen the screw (5/Fig. 11 / p. 10).


• Insert the bolts provided into the threaded holes.
• Place the spacers provided over the threaded holes (6/Fig. 11 / p. 10).
• Attach the localizer unit (Fig. 11 / p. 10) and secure to the SIREPHOS using 4 screws.
(5/Fig. 11 / p. 10)(6/Fig. 11 / p. 10)
• Solder spare lead "white" to solder terminal X55.1.
• Solder spare lead "gray" to solder terminal X55.4.
• Secure the two spare leads using cable ties.

Laser Light Localizer RXR2-130.812.01.02.02 Page 10 of 20 © Siemens, 2004


01.05 CS SD 24
Installation, Wiring, and Programming 11

1a. Changing the configuration (only SIREMOBIL Compact): 3.1

• Connect the service PC to the serial interface of the SIREMOBIL.


• The service program must already be installed on the PC.
• Remove the service program floppy disk from the log book and insert into drive A.
• Switch on the system.
• Switch on the service PC.
• Start the service program.
• Click "Get from Unit".
• Select "Laser" in the Adjustment, Parameters menu.
• Select yes for "Laser existing".
• Click "Set Value".
• Click "Put to unit".
• Wait for the system to boot.
• Back up the parameters (Data menu, Backup, Parameter).

© Siemens, 2004 RXR2-130.812.01.02.02 Page 11 of 20 Laser Light Localizer


01.05 CS SD 24
12 Installation, Wiring, and Programming

1b. Changing the configuration (only ARCADIS Varic): 3.2

• Log on to the service menu.


• Click <Main System>-<Configuration>-<Options>.
• Set "Laser Existing" to <yes> and click <Save>.
• Back up the parameters.

Laser Light Localizer RXR2-130.812.01.02.02 Page 12 of 20 © Siemens, 2004


01.05 CS SD 24
Functional Test and Final Steps 13
4-
Functional test
4Functional Test and Final Steps

• Switch on the system.


• Die Funktion des Tasters am Bedienpult des Grundgerätes testen: Bei einge-
schaltetem Laserlichtvisier müssen die beiden Laserdioden und die LED im Taster
leuchten.
• Test the automatic shut-off of the laser light localizer: Switch on the laser light localizer.
The two laser diodes must shut off automatically after 5 minutes.

NOTE Before performing the final steps, the laser light localizer has
to be adjusted according to the adjustment instructions pro-
vided.

© Siemens, 2004 RXR2-130.812.01.02.02 Page 13 of 20 Laser Light Localizer


01.05 CS SD 24
14 Functional Test and Final Steps

Final steps 4.1

Attaching the cover 0

NOTE The laser light localizer has to be adjusted before attaching


the cover.

• Place the cover provided on the SIREPHOS.


• Tighten the cover screws.

Note for U.S.A.: When using the laser light localizer, remove the spacer!

Attaching the cover panels 0

• Replace the cover panels on the SIREMOBIL.

Installing the I.I. attachment 0

Fig. 12: I.I. attachment Fig. 13: I.I. attachment

• Install the I.I. attachment on the I.I. retaining ring using the three clamps (Fig. 12 / p. 14)
(Fig. 13 / p. 14). A cassette holder can no longer be installed. The fluoroscopy grid
remains on the I.I. input.

Laser Light Localizer RXR2-130.812.01.02.02 Page 14 of 20 © Siemens, 2004


01.05 CS SD 24
Functional Test and Final Steps 15

Affixing the identification labels for the U.S.A. 4.2

Fig. 14: Checking the identification labels

© Siemens, 2004 RXR2-130.812.01.02.02 Page 15 of 20 Laser Light Localizer


01.05 CS SD 24
16 Functional Test and Final Steps

• Affix the identification labels provided (Fig. 14 / p. 15) (Fig. 15 / p. 16).


Affix the identification label, "Laser class 2" warning logotype, DHHS certification label,
and the two laser aperture labels.

Fig. 15: Label for C-arm side

Laser Light Localizer RXR2-130.812.01.02.02 Page 16 of 20 © Siemens, 2004


01.05 CS SD 24
Functional Test and Final Steps 17

Affixing the identification labels for Great Britian and other coun-
tries 4.3

Fig. 16: Affixing the warning labels


• Affix the identification labels provided as shown in (Fig. 16 / p. 17) (other countries).

Great Britain: Affix warning label according to IEC 825 (text) in local language,
warning label according to IEC 825 (symbol) and identification label.
Also affix laser aperture labels (Fig. 15 / p. 16).

© Siemens, 2004 RXR2-130.812.01.02.02 Page 17 of 20 Laser Light Localizer


01.05 CS SD 24
18 Functional Test and Final Steps

Other countries: Affix warning label according to IEC 825 (text) in local language,
warning label according to IEC 825 (symbol) and identification label.

Laser Light Localizer RXR2-130.812.01.02.02 Page 18 of 20 © Siemens, 2004


01.05 CS SD 24
Changes to Previous Version 19
5Changes to Previous Version 5-
Cover sheet changed

© Siemens, 2004 RXR2-130.812.01.02.02 Page 19 of 20 Laser Light Localizer


01.05 CS SD 24
20 Changes to Previous Version

Laser Light Localizer RXR2-130.812.01.02.02 Page 20 of 20 © Siemens, 2004