You are on page 1of 18

r

u

h
ü
M
s
n
m
ie
S

Laser Light Localizer


SP

Startup
Option

Startup Instructions

Valid for:
© Siemens, 2004

SIREMOBIL Compact
ARCADIS Varic

03779001
© Siemens, 2004
All documents may only be used for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.

Print No.: RXR2-130.815.01.02.02 English


Replaces: RXR2-130.815.01.01.02 Doc. Gen. Date: 01.05
CS SD 24
n.a.

2004
2 Copyright / Version / Disclaimer
1Copyright / Version / Disclaimer

Copyright
“© Siemens, 2004“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Mindit Magnetic Resonance Ltd. - China, Siemens Shang-
hai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA and/or
Siemens Healthcare Diagnostics Inc. - USA.

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The installation and service of equipment described herein requires superior understand-
ing of our equipment and may only be performed by qualified personnel who are specially
trained for such installation and/or service.

Laser Light Localizer RXR2-130.815.01.02.02 Page 2 of 18 © Siemens, 2004


01.05 CS SD 24
Table of Contents 3
0Table of Contents

1 _______ Laser Light Localizer_____________________________________________ 4

Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Additional notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Protective conductor resistance test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measuring devices and aids required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Adjusting the laser light fan beams in relation to the I.I. attachment . . . . . . . . . . . . . . . . 9
Attaching the adjustment device. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Adjusting the laser diodes in relation to the central beam . . . . . . . . . . . . . . . . . . . . . . . 11
Final steps and functional test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Checking the labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Labels for U.S.A.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Affixing the labels for Great Britain and other countries . . . . . . . . . . . . . . . . . . . . . . 14
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Mechanical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Checking the operating values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

2 _______ Changes to previous version _____________________________________ 17

© Siemens, 2004 RXR2-130.815.01.02.02 Page 3 of 18 Laser Light Localizer


01.05 CS SD 24
4 Laser Light Localizer
1-
Safety information
1Laser Light Localizer

NOTICE This product contains two Class 2 lasers. (U.S.A.: Laser class
2).
¹ Observe the safety information in ARTD-002.731.03...
¹ When working with the laser light localizer, do no look
directly into the laser beam.

NOTE Even though there is no immediate risk of eye injury, you


should not look into the laser beam.

Laser Light Localizer RXR2-130.815.01.02.02 Page 4 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 5

Additional notes 1.1

These setting instructions apply to installation of a laser light localizer on a SIREMOBIL


Compact or ARCADIS Varic system.
This document can also be used for adjustment procedures during service and mainte-
nance work. Store it in a safe location together with the adjustment device.

NOTICE Removing and inserting boards.


¹ Switch off the system before removing or inserting
boards!

© Siemens, 2004 RXR2-130.815.01.02.02 Page 5 of 18 Laser Light Localizer


01.05 CS SD 24
6 Laser Light Localizer

Protective conductor resistance test 1.2

After all work has been completed and all covers have been installed, perform the protec-
tive conductor test according to ARTD-002.731.17. The protective conductor resistance
must not exceed 0.2 ohms.

Laser Light Localizer RXR2-130.815.01.02.02 Page 6 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 7

Measuring devices and aids required 1.3

(per ARTD)
• Toolbox
• Adjustment device for laser light localizer (included with laser light localizer delivery 37
79 001 G5429)
• Protective conductor meter, item no. 44 14 899 RV090

© Siemens, 2004 RXR2-130.815.01.02.02 Page 7 of 18 Laser Light Localizer


01.05 CS SD 24
8 Laser Light Localizer

Start-up 1.4

• Press to switch on the laser light localizer while the C-arm is switched on. The laser
light localizer automatically shuts off after 5 minutes.
• Position the C-arm in the preferred position for operation, i.e. horizontally with the
C-arm lowered.
• Rotate the I.I. attachment so that the crosshairs are aligned horizontally/vertically.

Laser Light Localizer RXR2-130.815.01.02.02 Page 8 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 9

Adjusting the laser light fan beams in relation to the I.I. attachment1.5

Fig. 1: I.I. attachment Fig. 2: Laser light fan beam

• Set the excentrics (1/Fig. 1 / p. 9) and (2/Fig. 1 / p. 9) to the ”zero position”.


• Loosen the screws (3/Fig. 1 / p. 9) until the laser diodes can be turned in their holders.

NOTICE Lens out of focus


¹ Do not turn the lens of the laser diodes.

• Turn the diodes in the holder until the laser light fan beams are reflected on the I.I. as
shown in (Fig. 2 / p. 9). The laser light fan beams start at (X/Fig. 2 / p. 9). This setting
ensures that the laser light fan beams cross near the laser light localizer.
• Retighten the screws (3/Fig. 1 / p. 9).

© Siemens, 2004 RXR2-130.815.01.02.02 Page 9 of 18 Laser Light Localizer


01.05 CS SD 24
10 Laser Light Localizer

Attaching the adjustment device 1.6

Fig. 3: I.I. attachment Fig. 4: Pointer

Fig. 5: I.I.

• Attach the adjustment device above the laser light localizer as shown in (Fig. 3 / p. 10).

Laser Light Localizer RXR2-130.815.01.02.02 Page 10 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 11

Adjusting the laser diodes in relation to the central beam 1.7

X • Briefly switch on fluoroscopy.


• The pointer (1/Fig. 3 / p. 10) of the adjustment device and the crosshairs of the I.I.
attachment are displayed on the monitor.
X • Readjust the pointer and then briefly release fluoroscopy until it is displayed on the
monitor as shown in (Fig. 4 / p. 10).
• Use the adjusters (6/Fig. 5 / p. 10) to align the laser light fan beams with the lines of the
I.I. attachment. The laser light fan beams must now cross exactly at the hole of the
pointer.

In case of deviation:
• Loosen the screws for laser diode V1 (7/Fig. 5 / p. 10) or the screw for laser diode V2
(8/Fig. 5 / p. 10).
• Shift the laser diode according to the deviation.
• Retighten the screw (7/Fig. 5 / p. 10) or (8/Fig. 5 / p. 10).
• Use the adjusters (6/Fig. 5 / p. 10) to align the laser light fan beams with the lines of the
I.I. attachment. The laser light fan beams must now cross exactly at the hole of the
pointer.
• Repeat the adjustment procedure if necessary.
• Remove the adjustment device from the laser light localizer and store in a safe location.
The adjustment device is required for subsequent service and maintenance work.
• Secure the screws (7/Fig. 5 / p. 10) and (8/Fig. 5 / p. 10) with Loctite.

© Siemens, 2004 RXR2-130.815.01.02.02 Page 11 of 18 Laser Light Localizer


01.05 CS SD 24
12 Laser Light Localizer

Final steps and functional test 1.8

Fig. 6: Cover
• Replace the cover and secure using the cover screws (Fig. 6 / p. 12).
• These setting instructions can be used for adjustment procedures during subsequent
service and maintenance work.
• Switch on the laser light localizer. The two lasers must shut off automatically after 5
minutes.
• Perform the protective conductor test for the complete system.

Laser Light Localizer RXR2-130.815.01.02.02 Page 12 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 13

Checking the labels 1.9

Fig. 7: Checking the identification labels

© Siemens, 2004 RXR2-130.815.01.02.02 Page 13 of 18 Laser Light Localizer


01.05 CS SD 24
14 Laser Light Localizer

Labels for U.S.A. 1.10

• Check the identification label, "Laser class 2" warning logotype, DHHS certification
label, and the two laser aperture labels according to (Fig. 7 / p. 13) and (Fig. 7 / p. 13).

Fig. 8: Label for C-arm side

Affixing the labels for Great Britain and other countries 0

Great Britain Warning label according to IEC 825 (text) in local language. Check
warning label according to IEC 825 (symbol) and identification
label. Also check laser aperture labels (Fig. 8 / p. 14).

Laser Light Localizer RXR2-130.815.01.02.02 Page 14 of 18 © Siemens, 2004


01.05 CS SD 24
Laser Light Localizer 15

Other countries Warning label according to IEC 825 (text) in local language. Check
warning label according to IEC 825 (symbol) and identification
label.

Warning labels according to IEC 825


Use text according to local language:

Fig. 9: Affixing the warning labels

© Siemens, 2004 RXR2-130.815.01.02.02 Page 15 of 18 Laser Light Localizer


01.05 CS SD 24
16 Laser Light Localizer

Maintenance 1.11

Mechanical safety 0

• Ensure that the laser light localizer is labeled as described in this document. Replace
any labels that are damaged.
• Secure the I.I. attachment to the I.I. ring. Ensure that it is seated properly and locked
into place.
• Check the I.I. ring for mechanical damage.
• Check the laser aperture on the tube cover for damage.
• Ensure that the laser light fan beams are set as described in this document.

Checking the operating values 0

• Check the function of the "Laser light localizer on/off" button on the operating console of
the basic unit.
• Switch the laser light localizer back on.
¹ The laser diodes must shut off automatically after 5 minutes.

Laser Light Localizer RXR2-130.815.01.02.02 Page 16 of 18 © Siemens, 2004


01.05 CS SD 24
Changes to previous version 17
2Changes to previous version 2-
Cover sheet changed

© Siemens, 2004 RXR2-130.815.01.02.02 Page 17 of 18 Laser Light Localizer


01.05 CS SD 24
18 Changes to previous version

Laser Light Localizer RXR2-130.815.01.02.02 Page 18 of 18 © Siemens, 2004


01.05 CS SD 24

You might also like