Document No: DD+DIS188.07E


This no. is needed for production line only:
A350271.1 DRYSTAR 5302 / DRYSTAR AXYS
Type 5366 / 5367

Drum Drive Module (MOD06)


Replacement Instructions

► Purpose of this document


This document describes all steps of the replacement procedure for the spare part
Drum Drive Module (MOD06), CM+349270.X.
The main steps are:
• Exchange Drum Drive Module
• Print and evaluate SMPTE test image

• Dispose the Enclosure document after finishing your task


(observe the confidentiality note below).

• For an overview of valid enclosures please refer to MedNet GSO library


General Info => Agfa HealthCare => Publications => Enclosures =>
General - Enclosures
CONFIDENTIALITY NOTE:
Use, dissemination, distribution or reproduction of this document by unauthorized personnel is not permitted and may be unlawful.
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

09-2007 printed in Germany Document Node ID: 17039243


Enclosure_e_template_v02
Agfa Company Confidential Copyright © 2007 Agfa HealthCare N.V.
DD+DIS188.07E Drum Drive Module (MOD07) Replacement

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
(1) Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
(2) Strictly observe all safety directions within the "Generic Safety Directions" and
on the product.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Release 0 DRYSTAR 5302 / DRYSTAR AXYS Page 2 of 5


09-2007 5366 / 5367 Agfa Company Confidential
DD+DIS188.07E Drum Drive Module (MOD07) Replacement

Required Spare Part, Tools & Time 1 Switch off and disconnect all cables

3
80 min.

5366QRM24 5366QRM16

2 Open top cover and remove left cover 3 Remove belt and pulley

1 3

2
2

1
3

5366QRM25 5366QRM26

4 Disconnect 2 connectors and open 5 Remove 3 screws and remove drum


cable holder drive module.

3
2
2
1

1 1
5366QRM27 5366QRM28

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Release 0 DRYSTAR 5302 / DRYSTAR AXYS Page 3 of 5


09-2007 5366 / 5367 Agfa Company Confidential
DD+DIS188.07E Drum Drive Module (MOD07) Replacement

6 Put new drum drive module in place and 7 Re-connect 2 connectors, close cable
mount 3 screws holder.

3
1
1
2

2 2
5366QRM29 5366QRM30

8 Mount new pulley and new belt 9 Mount the side cover and close top
cover

1
3
2
4
2
5
1
3

5366QRM31 5366QRM32

10 Connect cables and switch on 11 Print a SMPTE testimage

1. Press the Key-operator key


3 4 WARMING UP
Please Wait 2. Scroll down to “Print Image”
3. Press the Confirm key
4. Press the Confirm key to select “testimage”
5 Ready
5. Scroll down to “smpte_lin.zip”
6. Press the Confirm key
2 7. Press the Confirm key to select one copy
SELECT CA
1 Print Testimage
3 smpte_1300.zip
4 smpte_lin.zip

5366QRM23 5366QRM07

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Release 0 DRYSTAR 5302 / DRYSTAR AXYS Page 4 of 5


09-2007 5366 / 5367 Agfa Company Confidential
DD+DIS188.07E Drum Drive Module (MOD07) Replacement

12 Check image quality 13 Your printer is ready to use

5366QRM08 5366QRM09

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Release 0 DRYSTAR 5302 / DRYSTAR AXYS Page 5 of 5