• True Discrete Random Access Specifications Sysmex CA-500
Principles Clotting : Scattered Light Detection Method • High Performance Automation Percentage Test End Point Detection Method Chromogenics : Colorimetric Method • Safety and Space Saving by Parameters PT, APTT, Fbg, TT, Extrinsic Factors (II, V, VII, X), Intrinsic Factors (VIII, IX, XI, XII), Design AT-III, PLG, α2-AP, PC, others to be confirmed Throughput Max. approx. 50 tests per hour (PT/APTT simultaneous analysis) • Extensive Data Management Sample Volume PT, APTT 50µL Intrinsic Factors 5µL Options Fbg 10µL AT-III, α2-AP, PLG 16µL TT 100µL PC 15µL Extrinsic Factors 5µL Reaction Tubes Capacity : 60 tubes Sample Aspiration The sample probe, with a liquid surface sensor, quantitatively aspirates plasma from centrifuged whole blood samples Quality Control X control, Levy-Jennings control 180 data points x 12 files x 7 parameters in memory Data Storage 300 samples (1,500 tests) results with the latest 600 tests of coagulation cureves Print Out Built-in thermal graphic printer Dimensions Main Unit : Approx. 540 x 487 x 470 W x H x D (mm) Weight (kg) Main Unit : Approx. 45 Power Requirement 117 / 230V AC ± 15% (50 or 60Hz) Power Consumption 400VA or less (50 or 60Hz)