INS DT4 EN Ichroma™ T4 Rev.02
INS DT4 EN Ichroma™ T4 Rev.02
02)
Revision date : March 20, 2020
T4
The detector tube contains contains anti human T4-
fluorescence conjugate, biotin-BSA-fluorescence
conjugate, sodium azide and NaOH in phosphate
buffered saline, bovine serum albumin (BSA) as a
stabilizer and sodium azide as a preservative in
INTENDED USE phosphate buffered saline.
Each detector tube contains 1 granule. 25 detector
ichroma™ T4 is a fluorescence Immunoassay (FIA) for the
tubes are packed in a pouch and packed in a box with
quantitative determination of thyroxine (T4) in human
5.5 mL of detector diluent.
serum/plasma. It is useful as an aid in management and
The Detector diluent is dispensed in a vial.
monitoring of thyroid disorder.
For in vitro diagnostic use only. WARNINGS AND PRECAUTIONS
- Cross-reactivity Accuracy
There was no significant cross-reactivity from these The accuracy was confirmed by 3 different lots testing
materials with the ichroma™ T4 test measurements. ten times each different concentrations.
T4 Recovery
Lot 1 Lot 2 Lot 3 AVG
[nmol/L] (%)
150 149.35 146.51 148.30 148.05 98.702 2. Larsen P.R., Dockalova J., Sipula D., Wu F. M.
75 74.32 74.40 71.90 73.54 98.1 Immunoassay of Thyroxine in unextracted Human
50 47.36 48.77 47.66 47.93 95.9 Serum. J. Clin. Endocrinol. Metabl., 1973, 37(2):177-
25 25.29 24.38 24.23 24.63 98.5 182
12.5 12.33 11.84 12.11 12.09 96.7 3. Wagner M. S., Wajner S. M., Maia A. L. The Role of
Thyroid Hormone in testicular Development and
Function. J. Endocrinol., 2008, 199(3) : 351-365
Comparability: T4 concentrations of 100 serum 4. Wahlin A., Wahlin T. B., Small B. J., Backman L.
samples were quantified independently with ichroma™ Influences of thyroid stimulating hormone on cognitive
T4 and mini VIDAS (BioMerieux Inc. France) as per functioning in very old age. J. Gerontol B. Psychol Sci.
prescribed test procedures. Test results were compared Soc. Sci., 1998, 5 : 234-239
and their comparability was investigated with linear
regression and coefficient of correlation (R). Linear Note: Please refer to the table below to identify various symbols
regression and coefficient of correlation between the
two tests were as follows.
VIDAS T4 vs. i-CHROMA READER
400
R = 0.9935
200
100
0
0.00 100.00 200.00 300.00 400.00
VIDAS T4 [nmol/L]
VIDAS T4 vs ichroma II
400
y = 0.9478x + 1.0741
ichroma™ T4 [nmol/L]
300
R = 0.9935
200
For technical assistance; please contact:
100
Boditech Med Inc.’s Technical Services
0 Tel: +82 33 243-1400
0.00 100.00 200.00 300.00 400.00 E-mail: sales@boditech.co.kr
VIDAS T4 [nmol/L]
y = 0.9922x + 2.1812
300 Tel: +(82) -33-243-1400
R = 0.9843
Fax: +(82) -33-243-9373
200
www.boditech.co.kr
100
Obelis s.a
0
Bd. Général Wahis 53,
0 100 200 300 400
1030 Brussels, BELGIUM
ichroma™ II [nmol/L]
Tel: +(32) -2-732-59-54
Fax: +(32) -2-732-60-03
REFERENCES E-Mail: mail@obelis.net