You are on page 1of 148

Service 

Manual 
(Engineering Edition) 

URIT­8021A 
Automatic Chemistry Analyzer 
V1.0 

URIT Medical Electronic Co., Ltd.
URIT‐8021A Service Manual 

Catalogue 
Copyright ....................................................................................................................................................................1 
PREFACE......................................................................................................................................................................2 
CHAPTER 1 INSTRUMENT INSTRUCTION .......................................................................................................................3 
1.1 Brief Intrduction .......................................................................................................................................................... 3 
1.2 Intended Use................................................................................................................................................................ 3 
1.3 Structure ...................................................................................................................................................................... 3 
1.4 Function of Main Part.................................................................................................................................................. 5 
CHAPTER 2 PERFORMACE AND TEST FLOW...................................................................................................................6 
2.1 Technical Parameter..................................................................................................................................................... 6 
2.2 Operational Environment ............................................................................................................................................ 6 
2.3 Test Process ................................................................................................................................................................. 7 
2.3.1 Typical Test FlowChart...................................................................................................................................... 7 
2.3.2 Test Process Instruction .................................................................................................................................... 8 
CHAPTER 3 INSTALLATION AND DEBUGGING............................................................................................................... 12 
3.2 Installation ................................................................................................................................................................. 14 
3.2.1 Instrument Inspection .................................................................................................................................... 14 
3.2.2 Installation ...................................................................................................................................................... 15 
3.2.3 Environmental Requirements ......................................................................................................................... 15 
3.2.4 Location Requirements................................................................................................................................... 16 
3.2.5 Power Requirements ...................................................................................................................................... 16 
3.2.6 Instrument Connection................................................................................................................................... 17 
3.3 Installation of Aspirating probe and Stirrer ............................................................................................................... 18 
3.3.1 Aspirating probe Installation........................................................................................................................... 18 
3.3.2 Stirrer Installation ........................................................................................................................................... 21 
3.4 Saftware Installation.................................................................................................................................................. 22 
3.5 Instrument Adjustment .............................................................................................................................................. 26 
3.5.1 Parameter Setup............................................................................................................................................. 26 
3.5.2 Setup of Motor Speed..................................................................................................................................... 30 
3.5.3 Parameter Setup of Liquid Path...................................................................................................................... 31 
3.5.4 Setup of Initial Position................................................................................................................................... 33 
3.5.5 Adjusting Position Parameter to Washing Pool .............................................................................................. 38 
3.5.6 Adjusting Position of Reagent /Sample Tray ................................................................................................... 40 
3.5.7 Debugging of Optical Path .............................................................................................................................. 41 
CHAPTER 4 MECHANICAL UNIT MODULE .................................................................................................................... 44 
4.1 Shell and Structure of the Machine ........................................................................................................................... 44 
4.1.1 Shell ................................................................................................................................................................ 44 
4.1.2 Disassembly and Instructure .......................................................................................................................... 45 
4.2 Aspirating Unit .......................................................................................................................................................... 50 
4.2.1 Function.......................................................................................................................................................... 50 
4.2.2 Structure ......................................................................................................................................................... 50 
4.2.3 Aspirating probe Installation........................................................................................................................... 52 
4.3 Stirrer Unit................................................................................................................................................................. 52 
4.3.1 Function.......................................................................................................................................................... 52 
4.3.2 Structure ......................................................................................................................................................... 53 
4.3.3 Stirrer Installation ........................................................................................................................................... 54

URIT‐8021A Service Manual 

4.4 Reagent /Sample Tray Unit........................................................................................................................................ 54 
4.4.1 Function.......................................................................................................................................................... 54 
4.4.2 Instructure ...................................................................................................................................................... 55 
4.4.3 Install and Disassemble Reagent/Sample Tray................................................................................................ 55 
4.4.4 Install and Disassemble Refrigeration module ............................................................................................... 59 
4.5 Reaction Unit............................................................................................................................................................. 60 
4.5.1 Function.......................................................................................................................................................... 60 
4.5.2 Structure ......................................................................................................................................................... 60 
4.5.3 Install and Disassemble Reaction Tray ............................................................................................................ 61 
4.6 Washing Hand Unit.................................................................................................................................................... 62 
4.6.1 Function.......................................................................................................................................................... 62 
4.6.2 Structure ......................................................................................................................................................... 62 
4.6.3 Installation and Disassembly .......................................................................................................................... 62 
4.7 Photoelectric Detection Unit ..................................................................................................................................... 63 
4.7.1 Function.......................................................................................................................................................... 63 
4.7.2 Structure ......................................................................................................................................................... 63 
4.8 Install and Disassemble Injection Pump.................................................................................................................... 66 
CHAPTER 5 LIQUID PATH ............................................................................................................................................ 67 
5.1 Function..................................................................................................................................................................... 67 
5.2 Liquid Path Diamgram .............................................................................................................................................. 69 
5.3 List of Valve and Pump....................................................................................................................................... 69 
CHAPTER 6 ELECTRONIC CIRCUIT OF HARDWARE ........................................................................................................71 
6.1 Summary ................................................................................................................................................................... 71 
6.2 List of Circuit Board.................................................................................................................................................. 71 
6.3 Position of Circuit Board........................................................................................................................................... 71 
6.4 Principle of Circuit Board.......................................................................................................................................... 73 
6.5 Function of Circuit Board.......................................................................................................................................... 73 
6.5.1 Main board ..................................................................................................................................................... 73 
6.5.2 Main Board ..................................................................................................................................................... 75 
6.5.3 Terminal Board................................................................................................................................................ 78 
6.5.4 Power Board ................................................................................................................................................... 79 
6.5.5 Transfer Board of Pump and Valve.................................................................................................................. 81 
6.5.6 Motor Transfer Board ..................................................................................................................................... 83 
6.5.7 Signal Amplify Board....................................................................................................................................... 85 
6.5.8 Liquid Level Sensing Board ............................................................................................................................. 87 
6.5.8 Light Transfer Board........................................................................................................................................ 89 
CHAPTER 7 SOFTEWARE PARAMETER .........................................................................................................................90 
7.1 General Parameter in Registry Table ......................................................................................................................... 90 
7.1.1 Way to Entrance.............................................................................................................................................. 90 
7.1.2 Sign of Simulation Mode................................................................................................................................. 90 
7.1.3 Sign of Having Water in Cuvette ..................................................................................................................... 90 
7.1.4 Path of Software Registration ......................................................................................................................... 90 
7.2 Common Files of Software........................................................................................................................................ 91 
7.2.1 Main Procedure .............................................................................................................................................. 91 
7.2.2 Test Program ................................................................................................................................................... 91 
7.2.3 Data Base ........................................................................................................................................................ 91 
7.2.4 Test Data Folder .............................................................................................................................................. 91

II 
URIT‐8021A Service Manual 

7.2.5 Language Package Folder................................................................................................................................ 91 
7.3 Method to Open Common Function.......................................................................................................................... 92 
CHAPTER 8 MAINTENANCE ........................................................................................................................................ 93 
8.1 Preparation................................................................................................................................................................. 93 
8.2 Daily Maintenance..................................................................................................................................................... 94 
8.2.1 Check Distilled Water...................................................................................................................................... 94 
8.2.2 Check Detergent Barrel................................................................................................................................... 95 
8.2.3 Clean Waste Solution Barrel ........................................................................................................................... 97 
8.2.4 Check Cleaning Position of Reagent and Sample Tray, Check Dilution Position ............................................. 99 
8.2.5 Check/Clean Aspirating Probes, Stirrer and Cuvettes..................................................................................... 99 
8.2.6 Check Printer and Printing Paper.................................................................................................................. 100 
8.3 Weekly Mintenance......................................................................................................................................... 100 
8.3.1 Aspirating Probes and Stirrer........................................................................................................................ 101 
8.3.2 Clean Reagent Tray and Sample Tray ............................................................................................................ 102 
8.3.3 Clean Washing Pool ...................................................................................................................................... 103 
8.3.4 Clean Reaction Tray....................................................................................................................................... 105 
8.3.5 Clean Washing Hand..................................................................................................................................... 106 
8.3.6 Clean Panel ................................................................................................................................................... 106 
8.3.7 Deep Clean Cuvette ...................................................................................................................................... 107 
8.3.8 Check AD Value and Save Cuvette Blank ...................................................................................................... 107 
8.4 Monthly Maintenance.............................................................................................................................................. 109 
8.4.1 Clean Incubation Bath................................................................................................................................... 109 
8.4.2 Clean Distlled Water Barrel........................................................................................................................... 111 
8.4.3 Clean Waste Solution Barrel ......................................................................................................................... 113 
8.4.4 Clean Detergent Barrel ................................................................................................................................. 116 
8.4.5 Clean Drive Shaft of Aspirating Probe........................................................................................................... 118 
8.4.6 Clean Drive Shaft of Stirrer ........................................................................................................................... 118 
8.4.7 Check Washing Hand .................................................................................................................................... 118 
8.5 Maintenance for Every Three Months............................................................................................................. 118 
8.5.1 Change Cuvette and Cuvette Bracket ........................................................................................................... 118 
8.6 Non­Periodical Maintenance ................................................................................................................................... 122 
8.6.1 Clear/Replace Aspirating Probe .................................................................................................................... 122 
8.6.2 Replacing Stirrer ........................................................................................................................................... 124 
8.6.3 Replacing Lamp............................................................................................................................................. 125 
8.6.4 Solenoid Valve Rinse..................................................................................................................................... 127 
8.6.5 Check/Replacing BNC Line............................................................................................................................ 127 
8.6.6 Replacing Waste Solution Tube..................................................................................................................... 128 
8.7 Long­time Disuse Maintenance............................................................................................................................... 128 
9 TROUBLESHOOTING .............................................................................................................................................. 130 
9.1 Troubleshooting Guide ............................................................................................................................................ 130 
9.2 Obtaining Technical Help ........................................................................................................................................ 130 
9.3 Troubleshooting Method.......................................................................................................................................... 131 
CHAPTER 10 Measurement Method ......................................................................................................................... 135 
10.1 Method of Measurement........................................................................................................................................ 135 
10.2 Endpoint ................................................................................................................................................................ 135 
10.3 Two­Point End Method (Double Reagent Endpoint)............................................................................................. 136 
10.3 Two­Point Rate Method (Fixed­Time) .................................................................................................................. 137

III 
URIT‐8021A Service Manual 

10.4 Rate Method (Kinetic method) .............................................................................................................................. 138 


10.5 Substrate Exhaust .................................................................................................................................................. 139 
10.5.1 Substrate Exhaust Method 1(Substrate exhaust Limit) .............................................................................. 140 
10.5.2 Substrate Exhaust Method 2(Slope Ratio).................................................................................................. 141 
Appendix A      URIT‐8021A Circuit Schematic Diagram .............................................................................................. 143

IV 
URIT‐8021A Service Manual 

Copyright 
Copyright: URIT Medical Electronic Co., Ltd. 

URIT Medical Electronic Co., Ltd. owns the intellectural property rights to this service manual. No part of this 
manual may be reproduced, stored or transmitted in any form, or by any means without the express written 
permission of URIT. 

Customer Service 
URIT Medical Electronic Co., Ltd. 
ADD: No.3 Fuhe Alley, Zhonghua Road, Guilin, Guangxi 541001, PR China 
Customer Service hotline:  86 773 2288566  86 773 2825742 
400­ Service hotline: 400  727  2288


URIT‐8021A Service Manual 

PREFACE 
This document is the service manual for URIT­8021A discrete random access chemistry analyzer. It 
describes the structure, operation, maintenance and troubleshooting concerning the instrument in details. 
Users should read carefully the manual and get special training before operating to guarantee instrument 
precision, normal operation and personal safety. 

Sign Illustration 
Meaning of the signs used in the URIT­8021A is as following. 
Caution. Refer to the 
Caution. Electric shock 
accompanying document 

Caution. Hot surface  Biohazard 

Ground  Power on 

In vitro diagnostic 
Power off 
medical device 

Environmental protection  Keep away from heat and 
lifetime  radioactive source 

Serial number  Manufacturer 

Recovery  May cause personal injury 

Refer to the operating manual


URIT‐8021A Service Manual 

CHAPTER 1 INSTRUMENT INSTRUCTION 
1.1 Brief Intrduction 

URIT­8021A is clinical chemistry instrument with the characteristics of open, full­automatic, discrete, STAT 
priority and computer­controlled. URIT­8021A is intended for use in conjunction with reagents to measure 
quantitatively certain chemical items in serum, urine and cerebrospinal fluid. Please read the operating 
manual carefully before using since it is a high sophisticated instrument. 

1)  Work Unit consists of optical unit, mechanical operation unit, liquid path control unit, hardware circuit 
unit and operating unit. 

2)  Mechanical operation unit consists of aspirating system, stirring system and eight­phase washing 
system.aspirating system includes sample tray, reagent tray, aspirating mechanism, aspirating probe 
(the inner core of syringe is made of ceramic which aspirating accurately and maintenance­free ) and 
sample probe wash station.The sample tray has 45 positions, which contains 35 rountine positions, 5 
standard positions, 3 QC positions and 2 STAT positions. Tube or sample cup could be placed on 
routine position. The reagent tray could place the reagent bottle in different specifications. The optical 
unit is a whole sealing, statics, array and rear spectrophotometry optical system, which includes 120 
cuvettes, high­resolution filter and halogen light. 

3)  The operating unit is a external computer (CPU 2.0 GHz or above, 16XDVD drives, memory: 1GB or 
above) , the application software should be setup under the Windows XP operation system. 

4)  Liquid path control unit consist of vacuum pump, solenoid valve, syringe, rinse system and tube system. 

5)  The instrument is easy to operate. The layout of the screen menu is reasonable, name is simple. Such 
as testing parameter setup, patient’s information input, quality­control, reagent, data query, standard, 
running test and hardware parameter. After setting, put the sample and reagent to the instrument and 
begin to analyze. Print out the result by the external printer at last. 

1.2 Intended Use 

The instrument is only for professional, in vitro use in hospitals, clinics and laboratories. 
CAUTION 
Please  contact  the  reagent  manufacturer  or  distributor  if  some  samples  may  not  applicable  to 
analytical test according to the test parameter and reagent. 

1.3 Structure 

The instrument structure mainly consists of optical system, reaction system, rinse system, sample system,

URIT‐8021A Service Manual 

reagent system, liquid path system and software system. 

The instrument is mainly composed of analyzer, computer and printer. Printer is optional accessory. 

The analyzer mianly consists of reagent tray, sample tray, reaction tray, aspirating system, stirring system, 
optical system, rinse system, liquid path system and hardware circuit. 

User could select cabinet as needed. 

Figure 1.3.1 Overview of Analyzer 

Figure1.3.2 Top View of Analyzer


URIT‐8021A Service Manual 

1.4 Function of Main Part 

Main workflow: 
1)  Mechanism unit startup the initialization progress 
2)  The eight­step wash mechanism washing the cuvettes automatically (No this step, if you do not select 
the clean before test option). 
3)  A washed and clean cuvette rotated to reagent aspirating position. 
4)  The reagent tray convey first reagent to aspirating position. The aspirating probe (aspirating mechanism) 
turn to the aspirating position and move down to absorb first reagent, then deliver it to cuvette. 
5)  First reagent in the cuvette preheated several cycles and then moved to sample aspirating position. 
6)  The sample tray conveys the sample to aspirating position. The aspirating probe (aspirating mechanism) 
turn to the aspirating position and move down to absorb sample, then deliver it to the cuvette. 

7)  Cuvette which with reagent and sample solution moved to stirring position to stir. 

8)  for the item which use double reagent for testing: 

a)  Constant  speed  mode:  After  fixed  cycle,  cuvette rotated  to  the  relative  position,  the  reagent tray 
convey second reagent to aspirating position. The aspirating probe (aspirating mechanism) turn to 
the aspirating position and move down to absorb second reagent, then deliver it to the cuvette. 

b)  Mixed mode: After arrive the incubate time of first reagent, cuvette rotated to the relative position, 
the  reagent  tray  convey  second  reagent  to  aspirating  position.  The  aspirating  probe  (aspirating 
mechanism) turn to the aspirating position and move down to absorb second reagent, then deliver it 
to the cuvette. 

9)  Cuvette moved to stirring position to stir after the addiction of second reagent. 

10)  During each cycle, each of the cuvettes transmitted to the photometric position (the light source position) 


for automatic signal collecting and absorbance calculation. 

11)  Eight­step washing mechanism washes the cuvettes automatically after finishing the test. 

Table 1.4.1: Function of main part 
Unit  Function 
Aspirating probe unit  Absorb or inject the sample and reagent for all the biochemical items. 
Sample tray unit  45 sample positions for all samples, STAT samples, controls and calibrators. 
Reagent tray unit  60 reagent positions for all reagents, detergent and diluents. 
Reaction tray unit  120 cuvettes to provide reaction and detection place. 
Stirring unit  Stirring and mixing the liquid in cuvette filled with sample or second reagent. 
Optical test system  It’s contains ten wavelengths and test the absorbance of liquid in the cuvette. 
Automaic wash unit  Provide eight­step auto wash cuvette function.


URIT‐8021A Service Manual 

CHAPTER 2 PERFORMACE AND TEST FLOW 
2.1 Technical Parameter 

1)  System: full­automatic, discrete/optional, STAT priority function 


2)  Testing speed: 280 tests/h (pure biochemistry) 
3)  Item storage: more than 1200 
4)  Data storage: Infinite storage according to the storage space of computer. 
5)  Method method: Endpoint, Rate(kinetic method), 2­point endpoint, 2­point rate (2­point kinetic), double 
wavelength, blank(reagent blank, sample blank), immunoturbidimetry etc. 
6)  120 cuvettes, the cuvettes colorimetric directly. 
7)  Calibration method: linear and non­linear calibration, multi­point linear. 
8)  Sample tray: The primitive tube which specification is 5mL, 7mL and 10mL or specified sample cup. 
9)  Reagent tray: 60 reagent positions with refrigeration function, which could place reagent bottle in 
various specification. 
10)  Sample Volume: 2  to 100  , variable in 0.1 
11)  Min. reaction volume: 150 
12)  Optical system: whole sealing, static, array and rear spectrophotometry optical system, 10 wavelengths 
for selecting.The wavelength accuracy is  2nm. 
13)  Light source: halogen lamp, 12V/20W. 
14)  Washing system: unique washing system for aspirating probe and stirrer. Eight­phase washing system. 

2.2 Operational Environment 

1)  Water consumption: 3L/H distilled water 
2)  Temperature control: 37  0.1 degree centigrade 
3)  Power: AC220, 50/60Hz, 500VA 
4)  Fuse: T8AL 250V 
5)  Relative Humidity: not exceed  85 percent 
6)  Environmental temperature: 10 to 35 degree centigrade 
7)  Instrument Dimension: 102cm  83cm  112cm (L  W  H)


URIT‐8021A Service Manual 

2.3 Test Process 

2.3.1 Typical Test FlowChart 
NOTE 
Two kinds of test modes in URIT­8021A are mixed mode and constant speed mode, each 
period time of them are respectively 18s and 24s. Choose the mode according to need. 
Instrument could be chosen to automatic wash cuvette before test or directly test. 

2.3.1.1 Mixed Mode 

Figure 2­1 Mixed Mode Flowchart 

Instruction of the mixed mode flowchart is as following: 
1­Start  6­Stir sample  11­Finish the test 
2­Initialization  7­Double reagents test  12­Washing cuvettes 
3­Washing cuvettes automatically  8­Single reagent test  13­Stop automatically 
4­Add first reagent  9­ Add second reagent 
5­Add sample  10­Stir reagent


URIT‐8021A Service Manual 

2.3.1.2 Constant Speed Mode 

Figure 2­2 Constant Speed Mode Flowchart 

The meaning of the numbers in the figure 2­2 is the same as the figure 2­1, please refer to 2.3.1.1 

2.3.2 Test Process Instruction 

2.3.2.1 Sample Aspirating Procedure 

The procedure of sample aspirating in each period is as follows: 
1)  The aspirating probe uplift from washing pool. 
2)  Turn to the above of sample tray. 
3)  Down to the sample cup (or tube) and absorb sample. 
4)  Uplift and turn to the above of reaction tray. 
5)  Down to the cuvette and inject the sample. 
6)  Uplift from cuvette and turn to the above of washing pool. 
7)  Down to washing pool for bubble type washing. 
(procedure of sample aspirating in next period) 

2.3.2.2 Reagent Aspirating Procedure 

The procedure of reagent aspirating in each period is as follow: 
1)  The aspirating probe uplift from washing pool. 
2)  Turn to the above of reagent tray. 
3)  Down to the reagent bottle and absorb first reagent. 
4)  Uplift and turn to the above of reaction tray.


URIT‐8021A Service Manual 

5)  Down to the cuvette and inject the reagent. 
6)  Uplift from cuvette and turn to the above of washing pool. 
7)  Down to washing pool for bubble type washing. 
(procedure of reagent aspirating in next period) 
The procedure of second reagent aspirating is the same as first reagent. 

2.3.2.3 Stirrer Work Procedure 

The stirrer work procedure in each period is as follow: 
1)  The stirrer uplift from washing pool. 
2)  Turn to the above of reaction tray. 
3)  Down to the cuvette. 
4)  Stirring the mixed solution in cuvette. 
5)  Uplift from cuvette and turn to the above of washing pool. 
6)  Down to washing pool for bubble type washing. 
(procedure of stirring in next period)


URIT‐8021A Service Manual 

2.3.2.4 Reaction Tray Work Procedure 

Figure 2­3 Reaction Layouts 

1­No.35, Square Wiper                                        6­No.81 light spot 
2­Adjusting left/right for optocoupler                  7­No.44, the first nozzle initial position 
3­Adjusting front/rear for optocoupler                8­Optocoupler bracket 
4­No.1, aspirating probe initial position              9­Light source 
5­No.110, stirrer initial position 

2.3.2.4.1 Mixed Mode Procedure 

The sequential procedure of reaction tray in mixed mode: 
Reaction tray provides 120 positions for cuvettes, and rotates twice in a clockwise direction and stops two 
times in a period. After the second stop, the reaction tray has completed the following procedures: 
1)  First reagent: 
a)  First stop: Aspirating probe absorbs sample from sample cup and then injects to cuvette. 
b)  Second stop: Aspirating probe absorbs first reagent from reagent bottle and injects to cuvette. Then

10 
URIT‐8021A Service Manual 

the stirrer turns to cuvette to stir the mixed solution. 
2)  Second reagent: 
a)  First stop: Aspirating probe absorbs second reagent from another reagent bottle and injects to 
cuvette. 
b)  Second stop: Then the stirrer turns to cuvette to stir the mixed solution. 

2.3.2.4.2 Constant Speed Mode Procedure 

The sequential procedure of reaction tray in constant speed mode: 
Reaction tray provides 120 positions for cuvettes, and rotates three times in a clockwise direction and stops 
three times in a period. After the third stop, the reaction tray has completed the following procedures: 
1)  First stop: Aspirating probe absorbs second reagent from reagent bottle and injects to cuvette. 
2)  Second stop: Aspirating probe absorbs sample from sample cup and then injects to cuvette. Then the 
stirrer turns to cuvette to stir the mixed solution. 
3)  Third stop: Aspirating probe absorbs first reagent from another reagent bottle and injects to cuvette, 
then stirrer turns to cuvette to stir the mixed solution again.

11 
URIT‐8021A Service Manual 

CHAPTER 3 INSTALLATION AND DEBUGGING 

3.1 Safty Precautions 

Please comply with the following rules for safety and active use. 

Preventing Breakage and Flammability 
Please observe the following precaution for prevention breakage. 
CAUTION 
1)  Installation must be complied with the installed environment in the manual 
2)  If relocation is necessary, contact your local distributor or URIT firstly. 

Preventing Injury 
Please observe the following precautions for preventing injury. 
CAUTION 
1)  While the instrument is in motion, DO NOT touch the moving parts, such as sample probe, 
reagent probe and stirrer, etc. Otherwise, you may be injured. 
2)  Before lamp replacement, turn off the power switch and wait until the lamp is cooled down. 
Otherwise, you may receive burns. 

Chemical and Biological Safety 
Please comply with the following matters for chemical and biological prevention. 
Biological Hazard 
Inappropriate  handling  of  sample  may  lead  to  biohazardous  infection.  Do  not  touch  the 
sample, mixed solution and waste directly with your hands. Be sure to wear gloves and lab 
coat,  if  necessary,  goggles.  In  case  your  skin  contacts  the  sample,  please  follow  standard 
laboratory safty procedure and consult a doctor. 

CAUTION 
Some reagents are strong acid or alkaline. Please use them carefully avoiding direct contact. If 
the  reagent  sticks  to  the  human  body,  immediately  wash  it  off  with  water  and  soap.  If  the 
reagent splashes into eyes accidentally, wash it off with water and consult an oculist. 

Disposing Wastes 
Please comply with the following matters when treat the wastes in order to avoid personal injury and protect 
environment.

12 
URIT‐8021A Service Manual 

Biological Hazard 

Some substances contained in reagent, control, calibrators and waste are subject to regulations 
of contamination and disposal. Dispose of the waste in accordance with your local or national 
rule for biohazard waste disposal. 

Operational Environment 
CAUTION 
1)  Please install the instrument according to the specified installed instruction in the manual. 
Otherwise, the results may not reliable even may cause system damage. 
2)  Please contact URIT if system state changed is necessary. 

Systematic Usage 
CAUTION 
1)  The  operator  must  receive  training  before  operating  the  instrument.  Please  follow  the 
instruction of the manual to operate. Improper operation may cause personal injury, system 
damage and improper result. 
2)  Please make a calibration and quality­control test when use the system for the first time to 
ensure it can be used normally. 
3)  A  quality­control test  must be  done  when use  the  system.  Otherwise,  the  reliability of  the 
result could not be guaranteed. 
4)  The  communication  interface  of  analytic  part  is  set  to  connect  with  the  communication 
interface of operation part. Please use the cables of URIT for connecting. 
5)  The computer should be for the instrument exclusive use. DO NOT run any other software 
with the computer while it is connected with the instrument. Inappropriate manner may result 
computer virus infection. 

Caution on Electromagnetic Wave Interference 
CAUTION 
Keep the instrument away from strong noise source and electromagnetic wave. Turn off mobile 
phones  and  transmitter­receiver  when  operating  the  instrument  as  the  electromagnetic  wave 
may cause an adverse effect on instrument. 

Other Cautions 
CAUTION 
1)  DO NOT touch the keyboard, indicator and mouse when your hands is wet, also includes the 
chemistry. 
2)  Check samples for contamination (dust, or fibrinogen) and air bubble before analyses. 
3)  To  make periodic  maintenance,  test  and  replacing  according  to  the  manual for  getting  the 
exact result. 
4)  For replacements of major parts, such as light source lamp, aspirating probe, reaction cuvette 
etc, please contact URIT. 
5)  For  settings  of  sample  volume,  reagent  volume,  wavelength,  standard  values  etc,  please 
refer to the instruction in reagent kit as well as this operating manual. Take note of checking 
the  quality  of  distilled  water  and  detergent,  calibration  results,  control  results,  and  sample 
results. Make sure there is no air bubble in the liquid paths.

13 
URIT‐8021A Service Manual 

3.2 Installation 

WARNING 
The URIT­8021A analyzer should be installed only by technicians of or authorized by URIT. 
Only the technicians of or authorized by URIT can perform the installation of URIT­8021A, users shall make 
preparation for satisfying the installation requirements in accordance with this manual before installation. If 
relocation is necessary, please contact your local distributor or URIT. 

3.2.1 Instrument Inspection 

Please check the carton according to the following procedures: 
1)  Loosen 8 buckles in the bottom of wooden box and uplift the box then take out cabinet. Open the door of 
cabinet and take out distilled water barrel, waste solution barrel and wash solution barrel. See figure 3­1 

Figure 3­1

14 
URIT‐8021A Service Manual 

2)  Take out the analyzer and put on the cabinet. See figure 3­2 

Figure 3­2 
3)  Open the package of liquid path and take out. 
4)  If any loss or damage exists, contact the distributor or manufacturer immediately. 

3.2.2 Installation 

The  instrument  is  high  sophisticated  thus  proper  installation  is  very  important  to  its  performance.  User 
should guarantee  the environment  and electrical condition  are  comply  with  the  recommended  conditions. 
Provide a distance of 50cm at least for each side for operating and maintaining. 

3.2.3 Environmental Requirements 

1)  Keep away from direct sunlight. 
2)  Only for indoor installation. 
3)  Dust free. 
4)  Installed on horizontal ground (Gradient less than 1/200). 
5)  Ground load weight: 250kg
15 
URIT‐8021A Service Manual 

6)  Room temperature: 10 to 35 degree centigrade. 
7)  Relative Humility: not exceed 85 percent (No condensation). 
8)  86kPa~106kPa; Atmosphere Pressure: 86 to 106 KPa. 
9)  The  reliability  of  the  data  could  not  be  guaranteed  if  the  fluctuation  of  temperature  and  humility 
exceeded the certain range. 
10)  Good ventilation and do not face air conditioner. 
11)  No obvious vibration. 
12)  Keep away from electromagnetic field and electricity interruption, such as cellphone, radio transceiver, 
Brush­type engine and the electrical equipment frequently turn on and off. 
13)  The instrument should be near to the power. 

3.2.4 Location Requirements 

The instrument installation layout is below. Surrounding distance is the recommended maintenance space. 

Measure: cm 
Instrument Dimension: 89  73  112cm (L  W  H) 
Dimension of Operating Board (Only for reference): 70  50  80cm (L  W  H) 

3.2.5 Power Requirements 

The following power must be prepared; switchboard should be located within 10m. 
1)  Power 
Voltage: AC 230V, 50/60Hz 
2)  Grounding 
Using three­pin power plug 
3)  Plug board 
A 15A output plug board with more than three 5A sockets. Heavy­duty devices should not share the plug 
board with the instrument, such as refrigerator, air conditioner etc. 
4)  3 core power cable cat is using; the type of wire and plug is depended on voltage.
16 
URIT‐8021A Service Manual 

CAUTION 
Make sure the instrument is grounded properly. Poor grounding may cause bad effects on test 
result and even damage to the instrument. 

3.2.6 Instrument Connection 

Figure 3­3 Instrument Connection 
The instrument connection detail steps are as follow: 
1)  Take out the power line from accessory box, one end inserted into the power interface of instrument, the 
other end connected to the power. 
2)  Using the COM port communication line which provided by URIT for connection. One end connected to 
the RS232 Port of instrument, the other end connected to the COM port of computer. And please tighten 
by the screws. 
3)  Tighten communication interfaces by screws to avoid falling off and inaccurate communication. 
4)  Connecting the BNC interfaces on the rear of instrument to the corresponding level sensing interface on 
the barrel. 
Tube connection of instrument and liquid path box: 

To  connect  the  liquid path  tubes  according to  the  marks  on  tubes  and the  silk­printing on  the  rear  of  the 
instrument. 

CAUTION

17 
URIT‐8021A Service Manual 

If you drain the waste solution into sewer directly without waste barrel, please short connecting 
the interface of waste solution BNC to prevent the alarm that waste solution is full. Otherwise, 
the liquid level sensing buzzer will alarm. 

3.3 Installation of Aspirating probe and Stirrer 

In order to prevent damage during transportation, aspirating probe and stirrer are disassembled and packed 
individually, then reinstalled and debugged after arrive the destination. 

3.3.1 Aspirating probe Installation 

1)  Uplift the arm of the aspirating probe to the highest position and then rotate it to the above of reagent 
/sample tray. 

Figure 3­4 Sampling Arm 
2)  Gently pinch the sampling arm cover and remove the cover from aspirating arm.

18 
URIT‐8021A Service Manual 

Figure 3­5 Take down Sampling Arm Cover 
(3)Take out the aspirating probe from the accessory box. See as the figure 3­6, 3­7. 

Figure 3­6 

Figure 3­7 
3)  Remove the retaining screw, insert the aspirating probe downwards into the positioning hole and tighten 
the  retaining  screw,  then  connect  the  liquid  level  sensing  connector  on  the  circuit  board.  See  as  the 
figure 3­8.

19 
URIT‐8021A Service Manual 

Figure 3­8 
NOTE 
1)  Keep the spring carefully, when remove the retaining screw. 
2)  Uplift sampling arm to the highest point before installing aspirating probe. Make sure the 
anti­collision  plate  installed  between  anti­collision  optocouple  and  spring  pressed  on 
location plate. 

Figure 3­9

20 
URIT‐8021A Service Manual 

3.3.2 Stirrer Installation 

1)  Gently pinch the stirrer cover and remove it. 

Figure 3­10 

NOTE 
Keep the flat washers and spring washers carefully to prevent fall into the instrument, when 
remove the restaining screws of stirrer. 

2)  Take out the stirrer from the accessory box. 

Figure 3­11 
3)  Remove the retaining screws; insert it downwards into the positioning hole, then tighen the restaining 
screw.

21 
URIT‐8021A Service Manual 

Figure 3­12 
NOTE 
The  positioning  hole  for  adjusting  the  front  and  rear  position  of  stirrer.  Two  screws  on  the 
adjusting cover do not need to fix tigh before adjusting the initial position. 

3.4 Saftware Installation 

1)  Click  a message box is displayed as figure 3­13.

22 
URIT‐8021A Service Manual 

Figure 3­13 
2)  Click Next, the dialog box to choose the installation location is shown as figure 3­14. 

Figure 3­14 
3)  Click Next, the message box is displayed as figure 3­15.

23 
URIT‐8021A Service Manual 

Figure 3­15 

4)  Click Next, the message box is displayed as figure 3­16 to choose whether create a desktop icon or not. 

Figure 3­16 
5)  Click Next, the message box is shown as figure 3­17. Then click Install botton to perform the installation 
process.

24 
URIT‐8021A Service Manual 

Figure 3­17 
6)  The message box will be shown as the figure 3­18 after complete the installation, then click Finish to 
exit. 

Figure 3­18 

7)  Double­click  to run the software.

25 
URIT‐8021A Service Manual 

8)  Click Help  Software Registration  to enter registration interface shown 


as figure 3­19 

Figure 3­19 
Send the User ID to After­Sale Service to obtain the License Number, and then input the number in blank 

box. Click  , a prompt box will popup, shown as figure 3­20. 

Figure 3­20 
Click OK, then close the software and restart according to prompt, the installation is completed. 

3.5 Instrument Adjustment 

This section mainly introduces the instrument adjustment method. Please make sure the correct connection 
of piping and wiring. 

3.5.1 Parameter Setup 

1)  Double­click  , then select Server for adjusting and input the password 3112772 into the main 


interface. 
2)  Turn on the red power switch, then green testing switch. And check that if abnormal for instrument.

26 
URIT‐8021A Service Manual 

3)  Click  Maintenance  Instrument  Parameter  ,  then  input  the  password 


3112749 to enter instrument parameter interface , see figure 3­21. 

Figure 3­21 
4)  The  instrument  has  been  adjusted  and  the  parameters  restoraged  in  the  main  board  before  leaving 
factory.  Click 

to get parameters from main board when adjust. 
NOTE 
1)  Please check the parameters according to 3.5.2­­­3.5.7 in this manual after parameter get 
for the position of component changed by transportting. 
2)  The Parameter Table in accessories box in each instrument is used in adjusting.

27 
URIT‐8021A Service Manual 

CAUTION 

1)  DO NOT click  before all instrument parameters being set properly. 


2)  On  adjusting  interface,  the  corresponding  mechanism  begins  to  action  only  after  being 

selected and clicking  . 

3)  Please  click  if  you  are  not  sure  the  mechanism’s  direction  of 
movement. 

Meaning of each parameter 
NOTE 
1)  The boxes with red words or grey color are not available. 
2)  Since  the  sample  and  reagent  share  a  aspirating  probe,  please  adjust  the  postions  of 
aspirating probe to cuvette, washing pool, sample, reagent to complete the adjustment of 
sample probe and reagent probe. 

l  Sample tray and reagent tray initial position: 
1)  31­R: Pos. 31 of the reagent tray rotates to the reagent aspirating position. 
2)  1­R: Pos.1 of the reagent tray rotates to the reagent aspirating position. 
3)  1­S: Pos.1 of the sample tray rotates to the sample aspirating position. 

l  Setup of Aspirating probe Movement: 
1)  To cuv.: The aspirating probe will move over to the cuvette 
2)  Vertical: The aspirating probe will move down into the cuvette. 
3)  To cleaning: The aspirating probe will move over to the above position of washing pool. 
4)  Vertical: The aspirating probe will move down into the washing pool. 
5)  R.position to S.position: The aspirating probe will move over sample aspirating position. 
6)  To S.position vertical: The aspirating probe will move down to the cuvette. 
7)  To  31­R:  The  aspirating  probe  will  move  over  to  the  reagent  aspirating  position  of  the  outer  ring  on 
reagent tray. 
8)  To  1­R:  The  aspirating  probe  will  move  over  to  the  reagent  aspirating  position  of  the  inner  ring  on 
reagent tray. 
9)  To R.bottle vertical: The aspirating probe will move down into the reagent bottle. 

l  Cuvette  : 
Cuv.1:  Cuv.1 will move over to the initial position of reaction tray. 

l  Reset: 
Reset:    It does affect any operation if not perform resetting after turn on the power.
28 
URIT‐8021A Service Manual 

l  Setup of Stirring Arm Movement 
1)  R.stirrer to cuv.: The stirrer will move to the cuvette. 
2)  Vertical:The stirrer will move down into the cuvette. 
3)  To cleaning: The stirrer will move to the above position of washing pool 
4)  Vertical: The stirrer will move down to the washing pool. 

l  Fine Adjustment: 
Perform fine adjustment for parameters of selected item. 
There are four commands up, down, right and left to adjust the parameter of selected item. 

l  Cuvette rinsing arm 
Arm to Cuv. Vertical: The cuvette rinsing arm moves down into the cuvettes. 

l  Test: 
It is not available if the button of Test display grey color, click the Adjust button to operate it. 

CAUTION 
The buttons will come into action immediately after clicking. Make sure that all the mechanism is 
in Arm raised state to prevent probe collision. 
1)  Reaction cuvette: The corresponding cuvette will rotate to the reagent aspirating position. 
2)  S. position: The corresponding sample cup will rotate to the sample aspirating position. 
3)  R. position: The corresponding reagent bottle will rotate to the reagent aspirating position. 
4)  Reset: The corresponding sample or reagent injection pump will rotate to the initial position. 
5)  Sam. In: The sample injection pump will aspirate quantitative volume of sample, measured by ul. 
6)  Reg. In: The reagent injection pump will aspirate quantitative volume of reagent , measured by ul. 
7)  Out: The sample or reagent injection pump will dispense the volume aspirated in the previous step. 
8)  S.probe fill water: the liquid path is open for washing the inner wall of sample probe, take times as unit. 
9)  Wash arm fill water: the liquid path is open for washing the inner wall of reagent probe, take times as 
unit. 
10)  Running gear: Running in the moving component of reaction tray. 
11)  Stop: Stop running gear. 

l  The function of other buttons: 
1)  Speed Set: Enter speed setup interface 
2)  Fluid Set: Enter liquid path setup interface 
3)  Switch On: Turn on the corresponding valves and pumps 
4)  Switch Off: Turn off the corresponding valves and pumps 
5)  Stop All Motor: Stop all motions of electric motor 
6)  Sample­detection: Test the communication between instrument and computer. 

l  Functional Buttons:
29 
URIT‐8021A Service Manual 

1)  Adjust: To open the calibrating and testing interface 
2)  Arm raised: All probes and sticks are raised to the highest point. 
3)  Reset: All probes and sticks move down to corresponding washing pool; reaction tray, reagent tray and 
sample tray move to the initial position and liquid path of washing pool begin to draw water. 
4)  Test: To perform the selected action 
5)  Save: To save all existing parameters. 
6)  Return: To exit parameter setup interface. 

3.5.2 Setup of Motor Speed 

Enter instrument parameter interface and click  , the dialog box shown as figure 2­23 will pop 

up 

Figure 3­22 Speed Set Dialogue 
It’s very important to check the parameters according to Parameter Table in accessories box because the 
speed parameters determine the running speed of motors. Just maintain the default value if some 
parameters do not listed in the table.

30 
URIT‐8021A Service Manual 

NOTE 
1)  The actual value of each parameter is consist of two value in the case, such as 250 and 
247, 250 represent big unit and 247 represent small unit(fine adjustment).The bigger the 
value is , the faster the speed of motor is. 
2)  To setup the parameters in this interface according to Parameter Table. 
3)  The maximum speed is 255. 

3.5.3 Parameter Setup of Liquid Path 

1)  Click  to enter liquid path setup interface shown as figure3­23: 

Figure 3­23 
2)  Check if the value of each parameter is consistent with the value in Parameter Table. 
3)  Each serial number in the interface is corresponded with the pump, valve or motor. The corresponding 
value is the opening time of pump, valve or motor. 
4)  The right part of the interface is the function section; the unit of value is given in the bracket. 
5)  It is available for testing whether these components are connected properly. Input the corresponding 
serial number and click Switch On, the corresponding pump, valve and motor will start to work. See 
figure 3­24. 

Figure 3­24

31 
URIT‐8021A Service Manual 

NOTE 
1)  Take ms as unit for parameter which is not indicated unit 
2)  The  serial  number  in  front  of  each  parameter  represents  the  corresponding  number  of 
motor. 
3)  Please consult the technician before performing fine adjustment of parameter 
4)  Click SAVE to save the data after adjusting. 

CAUTION 

Click  to prevent probe collision when adjust the motion position. 

Meaning of liquid path parameter 
1)  Spin cycle: The required times of instrument to complete a cycle procedure, measured by second. 
2)  Cuv.Path Length: The linear distance between two sides of cuvette which the light throuth, measured 
by um. 
3)  Air Before Reg: The quantity of inhalation of air before aspirating reagent, measured by  . 
4)  Air After Reg: The quantity of inhalation of air after aspirating reagent, measured by  . 
5)  Reagent remainder: The quantity of aspirting reagent for rinsing the aspirating probe, measured by  . 
6)  Extra R.volume: The quantity of aspirting reagent added to cuvette, and for calibrating the difference 
value between actual volume and theory volume. Measured by  . 
7)  Air Before Sam: The quantity of inhalation of air before aspirating sample, measured by  . 
8)  Air After Sam: The quantity of inhalation of air after aspirating sample, measured by  . 
9)  Sample remainder: The quantity of aspirting sample for rinsing the aspirating probe, measured by  . 
10)  Extra S.volume: The quantity of aspirting sample added to cuvette, and for calibrating the difference 
value between actual volume and theory volume. Measured by  .

32 
URIT‐8021A Service Manual 

3.5.4 Setup of Initial Position 

Figure 3­25 Initial Position of Motion Component 
1­No.35, Square Wiper  6­No.81 light spot 
2­Adjusting left/right for optocoupler  7­No.44, the first nozzle initial position 
3­ Adjusting front/rear for optocoupler  8­ optocoupler mounting bracket 
4­ No.1, aspirating probe initial position  9­Light source 
5­ No.110, stirrer initial position 

1)  Initial Position of Reaction Tray:

l  Click  in instrument parameter interface. Observe the 
position  of  light  spot  when  reaction  tray  stops,  if  it  is  not  at  the  middle  of  cuv.81;  adjust  the  left/right 
optocoupler mounting bracket as figure 3 ­25, 3­26, 3­27.

33 
URIT‐8021A Service Manual 

Figure 3­26 Adjusting front/rear position for optocoupler 

Figure 3­27 Adjusting left/right position for optocoupler 
NOTE 
If the light spot located on the middle of cuvette, the position is right. 

2)  Initial Position of aspirating probe: 

l  On the R.probe set section, click  . When 
the mechanism unit returns to initial position of optocoupler, adjust the front/back and left/right position 
of aspirating probe manually to make sure the aspirating probe is above the centre position of Cuv.1, 
see as figure 3­28, 3­29. It is available to perform fine adjustment after packing the cover of sampling 
arm.

34 
URIT‐8021A Service Manual 

Figure 3­28 

Figure 3­29 

l  Select  and  input  the  set  parameter  50,  then 

click  . Perform fine adjustment (the range of fine adjustment is 1 to 5) and observe 
the position of aspirating probe in cuv.1. If it is proper, down the aspirating probe till the tip touches the 
bottom of cuvette through gently pressing the arm of aspirating probe. 
l  When the tip is near the bottom of cuvette, please operate the align function step by step. The region of 
fine adjustment shown as figure 3­30:

35 
URIT‐8021A Service Manual 

Figure 3­30 Region of Fine Adjustment 
NOTE 
There are 2­3mm buffer course of anti­collision mechanism of aspirating probe. Observe whether 
the optocoupler barrier is moving or not for adjusting depth. 

3)  Initial Position of Stirrer 

l  On the Stirrer set section, click  . Adjust the 
left/right and front/rear adjustment assembly to make sure that the stirrer is above the centre of cuv.110. 
See figure 3­31 

Figure 3­31 

l  Select  and  input  the  set  parameter  50,  then 

click  ,  perform  fine  adjustment  (the  range  of  fine  adjustment  is  1  to  5).  Upward 
moving 2 to 3 step when the stirrer reaches the bottom of cuvette. 

4)  Initial Position of Washing Hand 
NOTE: It is not necessary to adjust the initial position of washing hand since it has been done before 
delivery. 
l  Keep  the  position  of  reaction  tray,  turn  washing  hand  to  make  sure  the  arrange  radian  of  washing 
probes are  consistent  with  reaction  tray. And  then  to adjust  whole  position of  washing  hand  to  make 
sure the scrape block is above the centre of cuv.35. 

l  Click  and  input  the  set  parameter  50,  then

36 
URIT‐8021A Service Manual 

click  , perform fine adjustment (the range of fine adjustment is 1 to 5), descending 
step by step. The requirement of adjusting washing hand as following: 
a)  The first to sixth porbe’s height is the same, the seventh and eighth are the same and deeper 1to 2 
mm than the former. 
b)  scrap block is not permitted to touch the wall of cuvette(perform fine adjustment of radian or turn 
rotate probes), see figure 3­32: 

Figure 3­32 
c)  The spring anti­collision assembly of seventh and eighth probes raised 1 to 2mm (2 to 3 steps), and 
others do not raised. See figure 3­33: 

Figure 3­33 
Scrap  block  is  not  permitted  to  touch  the  wall  of  cuvette  during  the  adjusting  course.  Adjusting  the 
position of washing hand must be done if the scrape block touch the wall and diverge the cuvette. (the 
position of washing hand has been adjusted before delivery, it is not necessary to adjust on the scene). 
The adjusting methods are as follows: keep the position of reaction tray and loose four locking screws. 
Then adjusting whole position of washing hand to make sure scrape block is above the centre of cuvette 
35.

37 
URIT‐8021A Service Manual 

Figure 3­34 

3.5.5 Adjusting Position Parameter to Washing Pool 

1)  Reagent probe to washing pool: 

l  Select  and  set  turning  parameter  as  45, 

click  and make aspirating probe above the washing pool. Perform fine adjustment 
to adjust left/right of aspirating probe to make sure it is above the centre of washing pool. 

l  Select  and set the vertical as 50, then click  , 
adjust  aspirating  probe  down  to  a  position  in  washing  pool  which  is  easy  to  observe,  then 

click  to adjust aspirating probe above the centre of washing pool. 

l  Select  , then perform fine adjustment (rang of fine adjustment is 1 
to 5) to make the inclined part of sampling part and up edge of inner washing cup at same horizontal line. 
See figure 3­35:

38 
URIT‐8021A Service Manual 

Figure 3­35 
2)  Stirrer to washing pool: 

l  Select  and set turning parameter as 80, click 
to make stirrer above washing pool. Perform fine adjustment to adjust left/right of stirrer to make sure 
stirrer is above the centre of washing pool. 

l  Select  and set parameter as 50 firstly, click 
and adjust stirrer down to a position in washing pool which is easy to observe, then 

click  to adjust stirrer above the centre of washing pool. 

l  Select  perform fine adjustment (rang of fine adjustment is 1 to 5). 
When the up edge of stirring paddle is higher (about 3 to 5mm) than the up edge of inner washing cup. 
See figure 3­36. 

Figure 3­36

39 
URIT‐8021A Service Manual 

3.5.6 Adjusting Position of Reagent /Sample Tray 

1)  Click  , the reagent tray will move to No. 1 reagent postion. 

l  Select  and set parameter as 110, then click  to make aspirating 


probe at the above of reagent bottle which located in No.1 reagent position. Adjusting aspirating probe 
at the above of the centre of No.1 reagent position. 

l  Select  and set parameter as 50, click  and 


adjust reagent probe down to a position in reagent bottle which is easy to observe, then 

click  to adjust aspirating probe above the centre of reagent bottle. 

l  The position of aspirating probe to reagent tray could be adjusted after performing the above processes. 
If  a  large  deviation  still  exist,  the  initial  position  of  reagent  tray  should  be  adjusted.  The  adjusting 
methods are as follows: 
Loose three locking screws on the reagent tray bracket, and then move the reagent tray to make sure 
the aspirating probe at the above of the centre of No.1 reagent position. See figure 3­37: 

NOTE:  It  is  not  necessary  to adjust  the  initial position of  reagent  tray  since  it has  been  done before 
delivery. 

Figure 3­37 

l  Select  and set parameter as 50, click  . Then 


perform  fine  adjustment  (rang  of  fine  adjustment  is  1  to  5),  upward  moving  1  to  2  steps  when  the 
aspirating probe touch the bottom of reagent bottle.
40 
URIT‐8021A Service Manual 

2)  Click  , the reagent tray will move to No. 31 reagent postion. 

l  Select  and set parameter as 90, click  to make aspirating probe at 


the above of reagent bottle which located in No.31 reagent position. Adjusting aspirating probe at the 
above of the centre of No.31 reagent position. 

l  Select  and set parameter as 50, click  and 


adjust aspirating probe down to a position in reagent bottle which is easy to observe, then 

click  to adjust reagent probe above the centre of reagent bottle. The aspirating probe 
could move down to the bottom of reagent bottle and keep the probe do not collide reagent bottle since 
vertical of inner and outer reagent bottle are the same. 

3)  Select  to make No.1 sample cup to the initial position. 

l  Select  , make aspirating probe to the above of 
No.1 sample cup and adjusting left/right to make sure the aspirating probe at the above of the centre of 
No.1 sample cup. 

l  Select  and set parameter as 50, then 

click  . Adjust aspirating probe down to a position in the sample cup which is easy to 

observe, select  and adjust left/right to make aspirating probe at 
the above of the centre of sample cup. 

3.5.7 Debugging of Optical Path 

1)  Wait for at least 30 minutes to make sure the light source is stable after power on. Then click 

to  enter  cuvette  signal  interface,  click  ,  120  cuvettes  will  be  added  water  by 
washing hand. 

2)  Click  in the instrument parameter interface to make the reaction tray move to the

41 
URIT‐8021A Service Manual 

initial position. Then click  , the interface is displayed as figure 3­38: 

Figure 3­38 
3)  Adjusting the potentiometers on the signal board to make the A/D value of corresponding wavelength is 
in the range of 56000  3000. See figure 3­39

42 
URIT‐8021A Service Manual 

Figure 3­39 

4)  Click  to enter the interface shown as figur3­40. Then click  , then the 


interface will show the A/D vaule of 120 cuvettes after reaction tray stop. 

Figure 3­40 
5)  Observe the A/D value and absorbance value. If the A/D value or absorbance is out of range, please 
replace the corresponding cuvette to meet the requirements. Then click SAVE, and the debugging of 
optical path is complete.

43 
URIT‐8021A Service Manual 

CHAPTER 4 MECHANICAL UNIT MODULE 
4.1 Shell and Structure of the Machine 

This section introduces the internal structure of the machine, shell’s composition, remove and installation. 

Figure 4­1 Analyzer 

4.1.1 Shell 

The shell of URIT­8021A is mainly composed of the protective cover, panel, left/right cover plate, front shell, 
rear baffle, rear cover plate and cabinet. See figure 4­2:

44 
URIT‐8021A Service Manual 

Figure 4­2(a) 

Figure 4­2(b) 

4.1.2 Disassembly and Instructure 

It’s recommended that the analyzer should be repaired by professional personnel if the malfunction cannot 
be solved according to the adjustment of software and routine maintenance. In case replacing a component 
is needed, please following the steps: 
1)  Turn off the testing power switch. 
2)  Turn off the mian power switch.
45 
URIT‐8021A Service Manual 

3)  Remove the end of gas springs, then take down two hinges to remove the upper cover.(In general the 
upper cover don’t need to remove) 
4)  Loose the  six hexagonal  socket head  cap  screws  on  the  left and  right  sides  of  the front  shell,  and  5 
screws on the front of panel, then pull out the front shell. Drive board, optical box, injection pump could 
be seen. Seen figure 4­3: 

Figure 4­3 
5)  The installation plate fixed in the frame by two hinges and a socket head cap screw. Loose the retaining 
screw and open the installation board. Main board, power board and terminal board are installed on the 
back of installation board. See figure 4­4: 

Figure 4­4 

6)  The rear baffle is used for protecting liquid path. See figure 4­5:

46 
URIT‐8021A Service Manual 

Figure 4­5 
The liquid path could be seen when removing the rear baffle. See figure 4­6: 

Figure 4­6 
1­Pump for supplying distilled water                4­Distilled water tank 
2­Pump forsupplying detergent                        5­Drain waste solution negative pressure tank 
3­Pump for draining waste solution 

7)  After removing rear baffle, loose the screws on the rear cover plate, then down open the rear cover plate,
47 
URIT‐8021A Service Manual 

the inner instructure of instrument could be seen as figure 4­7: 

Figure 4­7 
8)  Loose four M4  10 screws to remove the left cover plate. Then reagent tray refrigeration assembly(left), 
reagent/sample tray, reagent temperature sensor could be seen, see figure 4­7: 

Figure 4­8 
9)  Remove the right cover plate, reaction, power supply box, optical box and a part of washing hand, see 
figure 4­9:

48 
URIT‐8021A Service Manual 

Figure 4­9 
10)  The panel consists of four panels and two tray covers, shown as the figure 4­10. The disassembly 
sequence of panel is: reaction tray cover  reagent /sample tray cover  arc middle panel B  arc 
middle panel A  arc left panel  arc right panel. 
NOTE 
Do not touch the aspirating probe and stirrer when disassembly panels. 

Figure 4­10 

After removing all plates and panels:
49 
URIT‐8021A Service Manual 

Figure 4­11 

4.2 Aspirating Unit 

4.2.1 Function 

Sample probe and reagent probe share a aspirating probe in aspirating unit. The mainly function is aspirate 
the sample from sample cup (tube) to inject into cuvette and the operation of reagent is the same as sample. 
And  it’s  also  provide  liquid  level  sensing,  aspirating  probe  anti­collision,  with  the  amount  of  tracking, 
mechanical limit, power off protection etc. 

4.2.2 Structure 

Aspirating unit consists of drive mechanism and sampling arm. See figure 4­12: 

Drive mechanism used for supporting sampling arm and making aspirating probe to vertical and horizontal 
rotation. The drive mechanism is composed of vertical and rotary movement mechanism which are includes 
stepping motor, synchronization belt and synchromesh gear, to perform the movement of probe by spline. 

Sampling arm is composed of aspirating probe, liquid level sensing board, sampling arm cover and retaining 
screw of aspirating probe, which supported and connected by sampling arm bracket. See figure 4­13:

50 
URIT‐8021A Service Manual 

Figure 4­12 

Figure 4­13

51 
URIT‐8021A Service Manual 

4.2.3 Aspirating probe Installation 

Figure 4­14 
1)  Remove  the  sampling  arm  cover,  and  then  loose  the  retaining  screw  and  spring  to  take  down  the 
aspirating probe since it’s fixed on the sampling arm bracket. 
2)  Loose two screws that fixed on the spline to take out the sampling arm bracket. 
3)  Sampling mechanism installed on baseboard by four inner hexagonal screws, loose the screws to take it 
out. 
4)  The installation steps are contrary to disassembly. 
CAUTION 
1)  Checks  the  movement  of  aspirating  probe  whether  is  smooth  or  not  after  strengthen  the 
retaining screw, if not, loose the retaining screw to adjust. 
2)  Keep the surface of aspirating probe clean when disassemble the mechanism. 
3)  Disconnect the relative electric circuit and liquid path before disassembling the mechanism. 

4.3 Stirrer Unit 

4.3.1 Function 

The stirrer has the functions of mixing the sample and reagent in cuvette, mechanical limit, Power off 
protection.

52 
URIT‐8021A Service Manual 

4.3.2 Structure 

Stirrer unit is composed of drive mechanism and stirring arm. The main structure is shown as figure 4­15. 

Drive mechanism is used for supporting stirring arm and making stirrer to vertical and horizontal rotation.The 
drive mechanism is composed of vertical and rotary movement mechanism which are includes stepping 
motor, synchronization belt and synchromesh gear, to perform the movement of probe by spline. 

Stirring arm is composed of stirring assembly and stirring arm cover, which supported and connected by 
stirring arm bracket. See figure 4­15: 

Figure 4­15

53 
URIT‐8021A Service Manual 

4.3.3 Stirrer Installation 

Figure 4­16 
1)  The  stirrer  assembly  consists  of  stirring  motor,  stirrer  and  micro  adjustment  plate,  and  fixed  on  the 
stirring bracket by two screws. Remove the stirring arm cover, and then loose two retaining screws on 
the micro motor adjustment plate to take down the stirrer. 
2)  Loose two screws that fixed on the spline to take out the stirring arm bracket. 
3)  Stirrer mechanism installed on baseboard by four inner hexagonal screws, loose the screws to take it 
out. 
4)  The installation steps are contrary to disassembly. 
CAUTION 
1)  When installing stirrer assembly, please downwards into the position hole. 
2)  Keep the surface of stirrer clean when disassembling the stirrer assembly. 
3)  Disconnect the relative electric circuit before disassembling the mechanism. 

4.4 Reagent /Sample Tray Unit 

4.4.1 Function 

Reagent/sample  tray  is  a  container  to  load  sample  and  reagent,  and  transfer  accurately  them  to 
corresponding position in sequence. The mianly function as follows: 
1)  Load reagent/sample: The sample cup (tube) with sample placed on the sample tray, and reagent 
bottle with reagent placed in the reagent tray. 

2)  Rotate in sequence: Reagent/sample tray rotates in the sequence that setted by software, transfers

54 
URIT‐8021A Service Manual 

the sample and reagent to aspirating position. 

3)  Reagent refrigeration: Provide and keep refrigeration temperature. 

4.4.2 Instructure 

Reagent/sample tray unit is divided into sample tray bracket, reagent tray bracket, reagent/sample tray drive 
assembly, refrigeration module and reagent/sample tray thermostat assembly. The description of structure 
as following: 
1)  Reagent/sample tray bracket fixed on reagent/sample drive assembly. 

2)  Reagent/sample drive assembly fixed on reagent/sample tray thermostat assembly. 

3)  Reagent/sample tray thermostat assembly fixed on refrigeration module. 

4)  Refrigeration module fixed on baseboard. 

5)  Sample tray bracket: load sample cup (tube), single ring, 45 sample positions. Controlled to rotate in 
sequence by sample tray assembly. 

6)  Reagent tray bracket: load reagent bottle, includes outer and inner rings, each ring contains 30 positions, 


and 60 reagent positions in total are provided by two rings. Controlled to rotate in sequence by reagent 
tray assembly. 

7)  Reagent/sample tray drive assembly: controlled by software, transfer accurately sample/reagent to 
aspirating position. Composed of drive bearing, coded disc, optocoupler, stepping motor, 
synchronization belt and synchromesh gear. 

8)  Refrigeration module: for reagent refrigeration to make reagent meet the requirement of refrigeration 
temperature, and keep stability of reagent. It’s consists of two fancooled refrigeration assemblies and 
reagent/sample tray drive assembly. 

9)  Reagent/sample tray thermostat assembly: Protect the reagent/sample tray and keep the temperature 
of reagent tray. 

4.4.3 Install and Disassemble Reagent/Sample Tray 

There are two types of reagent/sample trays in URIT­8021A: 
1)  Split reagent/sample tray: reagent and sample trays are separated, and each of trays is driven by one 
motor assembly. 
2)  Integrative  reagent/sample  tray:  sample  and  reagent  tray  brackets  are  combined  and  driven  by  one 
motor assembly. See figure 4­17, 4­18, 4­19:

55 
URIT‐8021A Service Manual 

Figure 4­17 Sample tray bracket  Figure 4­18 Reagent tray bracket 

Figure 4­19 Integrative Reagent/Sample Tray 

4.4.3.1 Install and Disassemble Reagent/Sample Tray Bracket 

The disassembly steps of split reagent/sample tray are as following: 
1)  Loose the retaining screw to take out reagent tray bracket. 
2)  Loose six screws in the position hole to take out the sample tray bracket. 
3)  The installation steps are contrary to disassembly. 

The disassembly steps of integrative reagent/sample tray 
1)  Loose the retaining screws to take out the reagent/sample tray bracket. 
2)  The installation steps are contrary to disassembly.

56 
URIT‐8021A Service Manual 

4.4.3.2 Install and Disassemble Reagent/Sample Drive Assembly 

l  Split reagent/sample tray drive assembly. See as figure 4­20, 4­21: 
1)  Refer to 4.4.3.1 to disassemble reagent/sample tray bracket. 
2)  Loose the screws on bracket adjustment plate to take out the plate. 
3)  Remove  four  screws  on  the  left/right  of  reagent/sample  tray  thermostat  assembly  to  take  out 
reagent/sample thermostat and drive assembly. 
4)  The installation steps are contrary to disassembly. 

Figure 4­20

57 
URIT‐8021A Service Manual 

Figure 4­21 Split Reagent/Sample Tray 
l  Integrative reagent/sample tray drive assembly: 
Disassembly steps are the same as split reagent/sample tray. See as figure 4­22, 4­23 

Figure 4­22

58 
URIT‐8021A Service Manual 

Figure 4­23 

4.4.4 Install and Disassemble Refrigeration module 

1)  Refer to 4.4.3 to disassemble thermostat and drive assembly. 
2)  Loose three screws on the peltier clamp, and then take out peltier clamp and heat­insulating block. 
3)  Pull peltier wire to take out peltier. 
4)  Loose four screws that fixed on the radiator from the rear baseboard, then take radiator out. 
5)  The installation steps are contrary to disassembly.

59 
URIT‐8021A Service Manual 

Figure 4­24 Fancooled Refrigeration Assembly 
CAUTION 
1)  The upper and lower surface of peltier shall not be installed in reverse, otherwise, peltier 
does not refrigerate. 
2)  Tightening  alternatively  the  screws  which  fix  peltier  clamp  when  install  peltier  to  avoid 
uneven stress on refrigerating effect. 
3)  Disconnect the relative electric circuit before disassembling the assembly. 

4.5 Reaction Unit 

4.5.1 Function 

Reaction  tray  is  holds  cuvettes  and  provides  a  steady  environment  for  biochemical  test,  which  transmits 
cuvettes to corresponding position in the sequence setted by software. It’s also cooperates with aspirating 
probe, stirrer, washing hand and optical system to work. 

4.5.2 Structure 

Reaction  tray  is  composed of  cuvette  holder assembly,  constant­temperature  incubation  slot and reaction 


tray drive assembly. 

Cuvette holder assembly: it holds cuvette and consists of six groups of cuvette holders, 120 cuvettes and a 
reaction tray cover. Reaction tray cover is installed on reaction tray drive assembly by nut cap. 

Constant­temperature incubation slot: it’s installed on fitting seat by four inner hexagonal screws. Heating 
strip is sealed in slot by cement gel, which connects to socket by a temperature switch.

60 
URIT‐8021A Service Manual 

Reaction tray drive assembly: accurately control the movement of reaction tray. It’s composed of stepping 
motor, optocoupler, coded disc, helical gear, bearing and reaction tray baseboard. 

4.5.3 Install and Disassemble Reaction Tray 

1)  Loose the fitting nuts on a group of cuvette holder, then take out the holder and cuvettes to vacate a 
place. Manually rotate the reaction tray to make washing hand is above this place. 
2)  Loose nut cap to take out reaction tray cover and cuvette holders. 
3)  Remove  the  screws  on  constant­temperature  incubation  slot;  disconnect  the  circuit  of  temperature 
switch, then take out constant­temperature incubation slot. 
4)  Loose the screws on baseboard to take out reaction tray drive assembly. 
5)  Installation steps are contrary to disassembly. 

Figure 4­25 Structure of Reaction Tray

61 
URIT‐8021A Service Manual 

4.6 Washing Hand Unit 

4.6.1 Function 

The function of washing hand is vertical moving to clean the cuvettes automatically. 

4.6.2 Structure 

Washing hand unit is composed of washing hand and drive assembly. Washing hand is fixed on the drive 
assembly by metal thumbscrew and drive assembly fixed on baseboard. 

4.6.3 Installation and Disassembly 

1)  Loose four screws on thumbscrew to take out thumbscrew. 
2)  Upward remove the washing hand. 
3)  Loose the clamping block to remove washing arm. 
4)  Loose three screws between drive assembly and reaction baseboard to remove drive assembly. 

Figure 4­26
62 
URIT‐8021A Service Manual 

4.7 Photoelectric Detection Unit 

4.7.1 Function 

Photoelectric detection unit is one of the centre units which directly determines the accuracy and precision of 
analyze. It’s to produce light, split light, receive signal and convert light signal into electrical signal. 

Optional wavelength: 340nm, 405nm, 450nm, 492nm, 510nm, 546nm, 578nm, 630nm, 700nm, 800nm 
Accuracy of wavelength:  2nm 
Linearity range: 0 to 3.0Abs 
Resolution: 0.0001Abs 
Light source: halogen lamp, 12V/20W. 

4.7.2 Structure 

The photoelectric detection unit is a whole sealing, statics, array and rear spectrophotometry optical system. 
It’s mainly composed of light source holderassembly and optical box assembly. See as figure 4­27: 

Figure 4­27 Optical Box 
The halogen light generates steady continuous spectrum, suitably focus the light by convex lens and 
through the cuvette, irradiate on quasi value mirror; the light will focus again by quasi value mirror and then 
irradiate on monochromator.

63 
URIT‐8021A Service Manual 

There are ten potentiometers in the right open slot of optical box for adjusting bright value. Ten 
potentiometers are corresponded to 10 wavelengths respectively from up to down, they are 340nm, 405nm, 
450nm, 492nm, 510nm, 546nm, 578nm, 630nm, 700nm and 800nm. 

The principle of monochromator is as follows: 
1)  The continuous spectrum through the cuvette and then across quasi value mirror (entrance mirror) into 
rtesting system. 
2)  The spectrum with specific wavelength (other spectrums are assigned to the next beam splitter) through 
beam splitter and splitting by optical filter, the spectrum with single wavelength will be filtered, then 
transformed and enlarged by transducer, at last output electric signal for testing. 
3)  The absorption signal of all wavelengths of liquid in the cuvette will be read and processed by the above 
steps. Please see figure 4­28 for learning inner optical structure. 

Figure 4­28 Optical Structure

64 
URIT‐8021A Service Manual 

4.7.2.1 Light Source Holder Assembly 

Figure 4­29 Structure of Rice Bulb Component 

Figure 4­30 Light Source Holder Assembly 

Figure 4­31 Rice Bulb Holder  Figure 4­32 Halogen Light

65 
URIT‐8021A Service Manual 

4.7.2.2 Optical Box Assembly 

Figure 4­32 

4.8 Install and Disassemble Injection Pump 

The injection pump (500  ) located in the front of analyzer, with the left solenoid valve and aspirating probe 
composes sampling assembly that fixed on the instrument by a mounting plate. The disassembly steps are 
as following: 
1)  Screw out the black connectors on the injection pump. 
2)  Loose socket head cap screws to take out the injection pump. 
3)  The installation steps are contrary to disassembly. 
CAUTION 
1)  Two  black  connectors  should  be  connected  to  corresponding  interfaces.The  connector 
which  connects  with  solenoid  valve  is  connected  to  the  interface  in  the  middle  part  of 
injection pump, and the other one connected to the interface on the top. 
2)  There is a green collar inside the black connector to connect the Teflon pipe. Ensure the 
mouth of teflon pipe parallel to the face of collar and should not extend or retract, otherwise 
the leakage of aspirating probe may be occured. 

Figure 4­34
66 
URIT‐8021A Service Manual 

CHAPTER 5 LIQUID PATH 

5.1 Function 

1)  Liquid path system is divided into two parts: sample aspirating and washing. 
2)  The part of aspirating sample consists of a aspirating probe, injection pump (500  ), solenoid valve and 
stirrer,  which  provided  by  distilled  water  tank.  It  flow  through  solenoid  valve  and  open  to  the  500 
reagent injection pump then open to aspirating probe to wash the inner wall of probe. 
3)  The part of washing includes distilled water supply and waste solution drainage. 
l  Distilled water is draw to distilled water tank, then generate positive pressure. If the pressure is out of 
the range that pressure switch provides, the pump and valve will be turned on to supply water. 
l  Drain  waste  solution  pipeline  is  divided  into  two  parts:  motive  power  drainage  and  gravity  drainage. 
Aspirating  probe  washing  pool  and  stirrer  washing  pool  waste  solution  drainage,  and  reagent 
condensated  water  drainage  are  gravity  drainage  that  draw  waste  by  it’s  gravity.  All  waste  solution 
drainages  are  motive  power  drainage  except  washing  pool  in  the  instrument.  Waste  solution  pump 
generates  motive  power  to  make  waste  solution  vacuum  tank  into  vacuum  negative  pressure.  All 
pipelines are controlled by solenoid valve. When the pipline to draw waste solution into waste solution 
negative pressure tank, the corresponding solenoid valve is turned on. Finally, waste pump drain the 
waste solution from instrument. 
4)  Washing pool for washing the inner and outer walls of aspirating probe, and stirrer with distilled water, 
then the waste will be draw into waste solution barrel. 
5)  Washing hand is to clean cuvettes automatically which composed of eight groups of probe, the front six 
groups are double­probe, and the other two groups are signal­probe. 
l  The  first  group  is  detergent  probe,  the  short  probe  is  for  injecting  detergent  controlled  by  individual 
motive pump and valve, the long probe is for drawing waste solution to waste negative pressure tank by 
vacuum pump. 
l  The  groups  of  second  to  sixth  are  the  distilled  water  probe.  The  long  probes  is  for  drawing  waste 
solution to waste pipe by vacuum pump; Five short probes is injecting distilled water, the motive power 
is provided by water adding vacuum pump, a liquid separate box is used for separating distilled water 
into  five  ways.  Furthermore,  a  pump  back  system  is  setted  for  preventing  liquid  dropping  from  short 
probes.  A  pump  back  solenoid  valve  control  the  system,  both  ends  of  solenoid  valve  connect  with 
distilled  water  injecting  pipeline  and  waste  solution  negative  pressure  tank  respectively.  When  the 
solenoid  valve  is  open,  both  ends  is  connected,  the  excess  distilled  water  in  probe  will  be  draw  by 
negative pressure. 
l  The long probes of first to seventh groups drain distilled water and collect to a pipeline by a solenoid 

valve. 

l  The eighth group is the drier probe which controlled by a individual solenoid valve.

67 
URIT‐8021A Service Manual 

Figure 5­1 
6)  Washing pool liquid path is divided into two paths and each of them controlled by a solenoid valve to 
provide distilled water for washing pool to rinse aspirating probe and stirrer respectively. And all waste 
will be collected to a pipeline and then flow to waste solution barrel by it’s own gravity. 
7)  There is an exhaust valve to keep the gas and liquid in distilled water tank balance, and drain excess 
gas. 
8)  The waste solution from washing pool and condensated water from reagent tray are drained by it’s own 
gravity.

68 
URIT‐8021A Service Manual 

5.2 Liquid Path Diamgram 

Figure 5­2 

5.3 List of Valve and Pump 
List of valve and Pump 
No.  Function 
V1  Contol liquid path to washing the inner wall of aspirating probe 
V3  Solenoid valve for supplying detergent to washing hand(Hand1)
69 
URIT‐8021A Service Manual 

V4  Solenoid valve for supplying distilled water to washing hand(Hand2­6) 
V5  Solenoid valve is control washing hand to inject and draw water (Hand2­6) 
V6  Solenoid valve is control washing hand to draw waste solution (Hand1­7) 
V7  Solenoid valve for supplying distilled water to washing pool of aspirating probe 
V9  Solenoid valve for supplying distilled water to washing pool of stirrer 
V10  Solenoid valve is control washing hand to draw waste solution 
V11  Solenoid valve is control distilled water tank to drain gas 
V12  Solenoid valve for supplying distilled water to distilled water tank 
P1  Pump for supplying detergent 
P2  Vacuum pump for supplying distilled water 
P3  Negative pressure tank vacuum pump 
SP1  Aspirating probe injection pump 
T1  Distilled water tank (Positive pressure) 
T2  Wast solution negative pressure 
Rc1  Washing hand liquid separate box 
Rc2  Wate solution collect box 
M  Washing pool of stirrer 
R  Washing pool of aspirating probe 
WH  Washing hand 
PS  Pressure switch(ON:0.10MPa    OFF:0.12MPa) 
DT  Plug 
K1  Adaptor 
Y70_1­ Y70_3  Tee 
J1­J7  Adaptor(Plastics)

70 
URIT‐8021A Service Manual 

CHAPTER 6 ELECTRONIC CIRCUIT OF HARDWARE 

6.1 Summary 
This chapter mainly introduces the connecting methods and principles of circuit and function of each circuit 
board, etc. 

6.2 List of Circuit Board 
Name and function of circuit board of URIT­8021A are as follows: 
List 6­1 Circuit board of hardware 
No.  Name  Function 
Main  board  is  the  controlling  core  of  analyzer.  It  communicates  with 
1  Main board  computer  with  RS232  and  controls  the  actions  of  lower­position 
machine, receiving and collecting data and signal. 
The function of Drive board is to drive and control each motor, on­off 
2  Drive board 
control of pump and valve. 
3  Terminal board  mainly centralized control power supply 

4  Power board  Supply power to each circuit board of lower position machine. 

Transfer  board  of  Receiving signal of Drive board and control on­off of each pump and 



pump and valve  valve. 
The  function  of  signal  amplify  board  is  to  collect  light  signals,  adjust 
6  Signal amplify board  and do A/D signal transition, then supply the transferred signal to main 
board. 
liquid  level  sensing  Providing  function  of  liquid  level  sensing  and  anti­collision  for 
7  board  of  asipirating  aspirating probe 
probes 
Mainly  receiving  driven  signal  of  controlling  motor  and  transfer 
optocoupler  signal  to  Drive  board.  (Includes  transfer  board  of 
Transfer  board  of 
8  aspirating mechanism, transfer board of reagent/sample tray, transfer 
optocoupler of motor 
board of injection pump, transfer board of stirring mechanism, transfer 
board of washing hand and reaction tray.) 
Transfer board of  Supply power to transfer light. 

light 

6.3 Position of Circuit Board 
The position of each circuit board as follows: 6­1, 6­2, 6­3, 6­4.

71 
URIT‐8021A Service Manual 

Figure 6­1 Drive board, signal amplify board 

Figure 6­2 Main board, power board, terminal board 

Figure 6­3 Transfer board of pump and valuve, optocoupler transfer board of motor

72 
URIT‐8021A Service Manual 

Figure 6­4(A)Liquid level sensing board  Figure 6­4(B)Light transfer board 

6.4 Principle of Circuit Board 

Refer to appendix A for principle and connection of circuit board of URIT­8021A. 

6.5 Function of Circuit Board 

6.5.1 Main board 
The main board is the control center of lower position machine. PC send controlling information by RS232, 
and the information is sent to U16 (RS232 chip) by JP4 interface, then sent to CPU for processing. CPU 
controls  the  action  of  reaction  tray  directly.  Control  signal  is  sent  to  PRT  module  to  control  the  action  of 
motor  by  U6  (signal  is  strengthen  and  reverse)  and  P88  plug.  Other  motors,  pumps  and  valves  are 
controlled by controlling information which send from P88 to the modules of main board. 
Signal of AD value of signal board is sent to main board pass through PAMP signal plug, pass through U5 
and U2 then sent to CPU, CPU sent signal back to PC. Please refer to figure 6­3 and 6­5 for voltage test 
points and line connection of main board. 
Figure 6­2 Table of comparision of integrated chip of main board 
Position of main board  Chip No.  Function of chip 
U19  SP×1117M3­L­3­3  1.0A low power and low dropout regulator 
U1  REF3140  A/D conversion chip provides 4.096V reference voltage 
U5  OPA365  High­precision operational amplifier, as follower in circuit 
U10  74HC14  Hex Schmitt trigger 
U6  CD74HC24OM  Bumper/Line Driver 
U9  MA×813L  Reset, Watch dog and Power monitor chip 
U20  MA×809SEUR  Reset Circuit 
U16  SP3223EEY  MA×232 
U3  MA×3080CSD  RS485 transceiver,receive threshold ­200mv,­50mv 
U2  MA×113  200k/s,16 digit AD analog­to­digital conversion 
UP1  SRV05­4  Data line and port protection

73 
URIT‐8021A Service Manual 

U8  C805IF120  Single Chip(Silicon Labs) 

Figure 6­3: Voltage of test point of main board 
Test point  Voltage range(V) 
+5V  5±0.1 
+3.3V  3.3±0.1 
U5(Pin 4)  ­0.33±0.05 
C15(under capacity )  4.096±0.01 

Note: pin 13 and 
14 are not used 

Power  of 
5V power  main board 
【DY5V】  (PO1) 

4.096V 
test point 

DB.ZG 
The  connector 
for  connecting 
signal  amplifer 
board.【DB.ZG】 

Waste 
solution 
3.3V  test  BNC 
(JP9) 
point  【FBNC】 

5V  test  Detergent 


point  BNC 
( JP8) 
【QBNC】 

Distilled 
waterBNC 
(JP7) 
DB.ZQ  【ZBNC】 

The  connector  for  connecting 


drive board【DB.ZQ】  【RS232】
RS232 (JP4) 

Figure 6­5 Main board connecting and test point 

74 
URIT‐8021A Service Manual 

NOTE 
The content in bracket is the code of plug. 
Power on after connecting power board; test the voltage of each point by using multimeter. Check circuit and 
components if the voltage is not in the range. Please see figure 6­3 for voltage range of test point of main 
board. 

U5 chip needs positive and negative voltage, positive +5V, negative ­0.3V. 4.096V is the reference voltage 
of transfer chip. 

Figure 6­6 Diagram of main board 

6.5.2 Main Board 

Drive board mainly achieve driven of each motor, it is divided into eight modules, one T module, six driven 
modules (A to F) and one module for controlling pump and valve. 

The  main  board  sends  instruction  to  drive  board  by  P1;  the  driven  part  of  reaction  tray  of  T  module  is 
controlled by CPU directly. The controlling signal of washing hand and other six modules is gained by 485 
Bus. The T module only control the motor of washing hand which driven the up and down action, other six
75 
URIT‐8021A Service Manual 

modules could driven two motors. 

The work processing of drive board is as following: CPU send action instruction of one motor to drive board 
by  485  Bus.  Through  DIP  address  selection,  the  corresponding  module  processor  89S52  of  drive  board 
receiving  instruction,  the  output  signal  strengthen  reversely  through  74240,  then  be  sent  to  driven  chip 
THB6064, so that to driven motor. 
Refer to figure 6­4 for control details and DIP jumper setup. 
Figure 6­4 Driven module division and DIP address jumper 
corresponding 
Driven  transferring  Setting No.  DIP address 
Motor  Remark 
module  connection  line  of motor  jumper 
of motor 
Motor of 
Turn  0  Non­dial switch 
reaction tray 

Motor of 
Lift  1  00001 
washing hand 

A  not used 

B  not used 

Motor of sample 
Lift  2  00010 
tray 

Motor of reagent 
Turn  7  00111 
tray 

Motor control up 
and down action 
Lift  8  01000 
of aspirating 
probe 

Motor control 
left and right 
Turn  9  01001 
action of 
aspirating probe 

Motor control up 
and down action  Lift  10  01010 

of stirrer 
Motor control  Turn  11  01011

76 
URIT‐8021A Service Manual 

right and left 
action of 
reagent stirrer 

Sample 
Lift  12  01100 
injection pump 

Reagent 
Turn  13  01101 
injection pump 

Using  multimeter  to  test  voltage  of  each  test  point  of  drive  board.  See  figure  6­7.  Check  circuit  and 
components if the voltage is not in the range. Refer to figure 6­5 for voltage range. 
Figure 6­5 Voltage of test point of drive board 
Test point  Voltage range(V) 
J—Power  24±0.7 
Dr.Bd1  12±0.7 
PJ11(voltage of pin 1)  12±0.7 
P5  5.1±0.2 
PJ4(voltage of pin 3)  12±0.3 

Figure 6­5 Diagram of drive board

77 
URIT‐8021A Service Manual 

6.5.3 Terminal Board 

The function of terminal board is to supply power for analyzer. 
Using multimeter to test output voltage of terminal board after power on. See figure 6­8 and 6­9. The voltage 
range of terminal board is shown as figure 6­6. 
Figure 6­6 output voltage of terminal board 
Test point  Voltage range(V) 
AC1  220  10 
SMPS1  220  10 
ACB  220  10 
ACD  220  10 

Figure 6­8 Diagram of line connection of terminal board 
NOTE 
The content in bracket is the code of plug.

78 
URIT‐8021A Service Manual 

Figure 6­9 Diagram of line connecting of terminal board 

6.5.4 Power Board 
Power board distributes the power to each part of lower position machine. There are four channels to output; 
output of A channel is ±5V, B channel is ±5V, C channel is ±12V, the last is +5V. Test point is the interface of 
pin 1 to pin 14. See figure 6­7. P1 is the switching power input, voltage is ±15V. 

Test  output  of  P1  by  multimeter  after  connecting  transformer.  If  the  voltage  is  not  in  range,  adjust 
potentionmeter until it’s in the range. See figure 6­7 for output voltage of power board. See figure 6­10 and 
6­11 for line connecting and voltage testing. 
Figure 6­7: Output voltage of power board 
Test point  Voltage range(V) 
JP4  12±0.05 
P1 5V1  5.2±0.1 
P1 5V2  5.1±0.1 
P1 +12V  12±0.2 
P1 ­12V  ­12±0.2 
P1 +5VA  5±0.1 
P1 ­5VA  ­5±0.1 
P1 +5VB  5±0.1 
P1 ­5VB  ­5±0.1

79 
URIT‐8021A Service Manual 

Figure 6­10 Test point of line connecting and voltage 

NOTE 
The content in bracket is the code of plug.

80 
URIT‐8021A Service Manual 

Figure 6­11 Diagram of line connecting of power board 

6.5.5 Transfer Board of Pump and Valve 

The transfer board receives instruction which from pump and valve controlling module by JP25 to control 
pumps  and  valves.  Voltage  of  solenoid  and  vacuum  pump  is  24V.  See  figure  6­12  and  6­13  for  line 
connecting. 

Voltage test of transfer board (motor/pump and valve): 

Test the voltage of test point by using multimeter after connecting the lines of drive board. Check circuit and 
components if the voltage is not in the range. See figure 6­8for voltage range of transfer board. 
Figure 6­8: voltage of test point of transfer board 
code  Voltage range 
24V  24±0.2V

81 
URIT‐8021A Service Manual 

Figure 6­12 Diagram of line connecting of transfer board 

Figure 6­13 Diagram of line connecting of transfer board of pump and valve

82 
URIT‐8021A Service Manual 

6.5.6 Motor Transfer Board 

The function of motor transfer board is to receive drive signal from drive board, then separate into two parts, 
includes  LIFT  and  TURN.  Each  part  could  supply  separate  interface  for  connecting  line  of  motor  and 
optocoupler. 

There  are  six  motor  transfer  boards.  Only  aspirating  probe  with  JP2  and  JP3  plug  which  function  is 
anti­collision and liquid level sensing, the others have not. 

See figure 6­9, 6­14, 6­15, 6­16 for detail line connecting method and voltage test point 
NOTE 
The content in bracket is the code of plug. 

Figure 6­9: Voltage test point of moter transfer board 
Code  Voltage range 
+5V  5±0.15V 

Figure 6­14 Transfer board of aspirating probe

83 
URIT‐8021A Service Manual 

Figure 6­15 Transfer board of washing station and reaction tray 

Figure 6­16 Transfer board of stirring mechanism

84 
URIT‐8021A Service Manual 

6.5.7 Signal Amplify Board 

The function of signal board is to convert optical signal into electrical signal.  Then amplify and transfer to 
main board. 
There are ten silicon photocells on signal board, the function of these photocells are to receive optical signal 
and  convert  into  weak  voltage  signal,  then  amplify  the  signal  by  IC  LTC6244  (U1_S3  to  U1_S12,  each 
photocell corresponds one piece of LTC6244), then amplify signal again by OPA365 (U2), select by DG506 
(U3), at last transfer to main board by PAMP1. See figure 6­17 for line connection. 

UT1 is power supply chip which voltage is ­5V; UT2 is power supply chip which voltage is 5V. 

Test the voltage of test point by multimeter after power on. Check circuit and components if the voltage is not 


in the range. See figure 6­10 for voltage range of signal board. 
Figure 6­10 Voltage of test point of signal board 
Code  Voltage range 
+5V  5±0.05V 
­5V  ­5±0.05V 
+12V  +12±0.2V 
­12V  ­12±0.2V 
­0.3V  ­0.3±0.05V

85 
URIT‐8021A Service Manual 

Figure 6­17 Diagram of line connecting of signal board 

NOTE 
The content in bracket is the code of plug.

86 
URIT‐8021A Service Manual 

Figure 6­18 Voltage test point on the back of signal board 

6.5.8 Liquid Level Sensing Board 

The liquid level sensing board consists of liquid level sensing part and anti­collision optocoupler. 

Liquid level sensing adopts capacitive level induction principle. Probes link to U2 (MC3394) by JP2, and U2 
converts  capititance  changes  into  voltage  changes,  then  feedback  to  U1  (MC9S08SH8)  for  processing. 
Alarming of reagent allowance could be achieved by U1. 

The  anti­collision  optocoupler  has  no  output  voltage  under  normal  condition.  The  optocoupler  will  get 
through if probe collision. Pin 4 will feedback collision signal to drive board, then control signal will be sent 
from drive board. Probe will uplift to the top automatically to protect probe. 

Test voltage of test point by multimeter after connecting liquid level sensing line. See figure 6­19 and 6­20. 
Check circuit and component if the voltage is not in the range. See figure 6­11 for testing voltage range of 
liquid level sensing board.

87 
URIT‐8021A Service Manual 

Figure 6­11 Voltage of test point of liquid level sensing board 
Test point  Voltage range(V) 
﹢12V  12±0.7 
VCC1(IC MC33941 Pin 12_  5±0.1 

Figure 6­19 Line connecting of liquid level sensing board and voltage test point (old version) 

Figure 6­20 Line connecting of liquid level sensing board and voltage test point (old version) 
NOTE 
There is no obvious voltage test point on new version liquid level sensing board; its voltage is 
changed by chip into 5V. If it is need to be test, please contact URIT. 

The number of 1 to 4 is the label of jumper cap joint in figure 6­20; the jumper cap is used for programming 
the procedure of liquid level sensing board and selecting sensitivity of liquid level sensing. See table 6­12 for 
the function of jumper cap. 
Table 6­12 Table of function of jumper cap group 

Function of jumper cap 
Selection of sensitivity of jumper cap
No.  Function 

88 
URIT‐8021A Service Manual 

Joint  for  programming  liquid  Not  plug  jumper  cap­­­the  lowest 



level sensing procedure  sensitivity 
2  Medium sensitivity joint  plug No.3 jumper cap, low sensitivity 

3  Low sensitivity joint  No.2 jumper cap, medium sensivity 

Joint  for  programming  liquid  Plug  No.2+No.3  jumper,  the  most 



level sensing procedure  sensitivity 

6.5.8 Light Transfer Board 

The light transfer board is a “transfer stop” of light power line, which separate light and power and convinent 
for replacing light. 
See figure 6­21 for line connecting of light transfer board 

Figure 6­21 line connecting of light transfer board

89 
URIT‐8021A Service Manual 

CHAPTER 7 SOFTEWARE PARAMETER 

7.1 General Parameter in Registry Table 

7.1.1 Way to Entrance 

Click  Begin  ­­­Run,  input  regedit,  and  then  click  confirm.  Click  my  computer­­ 
HKEY_CURRENT_USER­URIT­DIAGNOSE; it is the location of saving of parameter. 

7.1.2 Sign of Simulation Mode 

In MotorSet parameter table, double click CRAZY_SIMΜLATION_MODE to entry to modify. 
0 means normal motion mode, 1 means simulation mode. 
NOTE 
1)  Under  the  working  condition  of  simulation  mode,  the  test  speed  will  be  accelerated  and 
virtual data will be created. 
2)  Do  not  use  simulation  mode  when  computer  connect  to  analyzer,  otherwise,  will  appear 
probe collision. 

7.1.3 Sign of Having Water in Cuvette 

In System parameter table, double click HaveWater to entry. 
0 means there is no water in cuvette; 1 means there is water in cuvette. 
NOTE 

System will prompt to drain water before test if there is water in cuvette. 

7.1.4 Path of Software Registration 

In System parameter table, double click Directory to entry. That is the registry path of software which must 
be accordance with actual path, otherwise will make mistake.

90 
URIT‐8021A Service Manual 

7.2 Common Files of Software 

7.2.1 Main Procedure 

Icon: 
Function: Main control program 

7.2.2 Test Program 

Icon: 
Function: Test control program 

7.2.3 Data Base 

Name: URIT­DIAGNOSE.mdb 
Function: it is used for storaging all parameters, datas, include reagent parameter, test result, hospital setup, 
QC and calibration datas, except mechanism parameter, instrument status and test curve. 
Application: Copy this document before install new software, and replace the new document in new software, 
so that the old data could be reserved. 

7.2.4 Test Data Folder 

Name: CheckData 
Function: it is used for storaging test curve, a file is generated for one day,its name is the date when the file 
is generated. 
Application: The test curve data could be checked in different computer on URIT software after copying. 

7.2.5 Language Package Folder 

Name: Language 
Function: it is used for storaging all text message, prompt information as Not found will appear if any loss or 
damage. 
Application:  Modifying  the  text  description  according  to  requirement  and  the  text  will  be  displayed  on 
corresponding position.

91 
URIT‐8021A Service Manual 

7.3 Method to Open Common Function 

Different functions are configured for satisfying different requirements. A part of function is shielding since 
different model with different hardware. The opening method and relative function are as follows: 
Click D:\URIT Biochemical Analyzer, then open the Crazy.ini file by notepad or wordpad program. 

1)  Temperature displayed function 
Discription: the temperature of reaction tray and reagent tray will be displayed by open this function 
Character field: TemperatureDisplay 
Opening method: after =, number 0 means closed, 1 means open. 

2)  Version reading and displaying function of lower position machine 
Discription: this function is used for query program version of main board and low position machine. 
Character field: VersionRead 
Opening method: after =, number 0 means closed, 1 means open. 

3)  Cleaning function before test 
Description: cleaning cuvette before test automatically 
Character field: Clean BeforTestModule 
Opening method: after =, number 0 means closed, 1 means open. 

4)  Automatic printing function 
Description: the test data will be printed automatically after completing a test. 
Character field: AutoPrintModule 
Opening method: after =, number 0 means closed, 1 means open. 

5)  Alarming function when sample shortage 
Description: alarming when sample shortage. 
Character field: UseSamRegWarn 
Opening method: after =, number 0 means closed, 1 means open. 

6)  Sample alarming information printing function 
Discription: To print out sample shortage information with test result. 
Character field: PrintSamRegWarn 
Opening method: after =, number 0 means closed, 1 means open. 
NOTE 
1)  Number 1 means open, 0 means closed, carefully execute this operation or contact URIT to 
avoid instrument damage. 
2)  All  the function  must be  matched  with  hardware.  If  not,  instrument  may be  damaged  after 
power on.

92 
URIT‐8021A Service Manual 

CHAPTER 8 MAINTENANCE 
Please strictly according to the operating manual to operate and maintain instrument to guarantee the safety, 
reliability and good working condition of instrument. 
CAUTION 
1)  Please strictly according to the operating manual to operating instrument, otherwise damage 
or personal injury will be occurred. 
2)  People who not trained or authorized by URIT, is not permit to repair the instrument. 
3)  Do not spill water, reagent and detergent on mechanical or electrical components. 

Biological Hazard 
Be sure to put on protective gloves, clothes, or even goggle if necessary. 

8.1 Preparation 
The following tools and detergent may be used in maintenance. 

Tool 
A set of hexagon wrench 
Slotted point screwdriver (medium size) 
Nozzle cleaner 
Clean beaker 
Tweezer 
Clean gauze 
Clean cotton swab 
Brush (used for clean barrel) 

Detergent 
1)  Alkaline detergent, concentrate detergent 
2)  Acid detergent 
NOTE 

1)  The  alkaline  detergent  is  concentrate, dilute  it  with distilled  water,  the  proportion  is 
1:9. 
2)  Please use the detergent supplied by URIT. If not, the test result is not reliable. 

Others 
Absolute alcohol 
Clean beaker

93 
URIT‐8021A Service Manual 

8.2 Daily Maintenance 

Do the following steps everyday to keep instrument under good working condition. 

8.2.1 Check Distilled Water 

1)  To make sure the testing switch is off. 
2)  To check allowance of distilled water in barrel 
If the allowance is enough, do step 5. 
If the allowance is insufficient, do next step. 
3)  Screw off the cover of distilled water barrel in an anticlockwise direction and injecting distilled water. 
4)  Screw on the cover of distilled water barrel in a clockwise direction. 
5)  Check the connection of pipeline of distilled water. 
If there is leakage, press the circle and pull out distilled water pipe. 

Figure 8­1 Pull Out Distilled Water Pipe 
Check the pipe, if there is any damage, cut off the damaged part. If not, insert the pipe and check again. 

Figure 8­2 Insert Pipe 
6)  Check the connection of BNC signal line: check the joint connect well or not.

94 
URIT‐8021A Service Manual 

Figure 8­3 BNC Connection 
If the connection is well, do the next step; if not, connect well as figure 8­3. 

8.2.2 Check Detergent Barrel 

1)  To make sure the testing switch is off. 
2)  Check the allowance of detergent barrel. 
If the allowance is enough, do step 8; if not enough, do the next step. 
3)  Screw off signal line in a clockwise direction. 

Figure 8­4 Before Screwing                                                  Figure 8­5 After Screwing 


4)  Screw off the cover of detergent barrel in an anticlockwise direction; take down it with liquid level sensor.

95 
URIT‐8021A Service Manual 

Figure 8­6 Screw Cover of Detergent Barrel                        Figure 8­7 Take Down Cover of Detergent barrel 


NOTE 
Put the cover with liquid level sensor on the clean desk. 

5)  Dilute  concentrate  alkaline  detergent  with  distilled  water  according  to  the  proportion,  then  pour  into 
detergent barrel. 
6)  Put on cover with liquid level sensor and tighten. 
7)  Connecting BNC signal line. 
Insert BNC signal to the joint and tighten. 

Figure 6­8 Before Screwing                                                          Figure 6­9 After Screwing 


8)  Check the connection of distilled water pipeline: 
If there is leakage, press the circle and pull out distilled water pipeline.

96 
URIT‐8021A Service Manual 

Figure 8­10 Pull out Detergent Pipe 
Check the pipe, if there is any damage, cut off the damaged part. If not, insert the pipe and check again. 

Figure 8­11 Insert and Tighten Distlled Water Pipe 

8.2.3 Clean Waste Solution Barrel 

Biological Hazard 
1)  Be sure to put on protective gloves, clothes, or even goggles if necessary. 
2)  Dispose the waste solution according to local discharge standard, and consult reagent 
supplier or distributor. 

1)  To make sure the testing switch is off. 
2)  Check the allowance of waste solution, if it is too much, do the next step. 
3)  Pull out waste solution pipe: There are two waste solution pipes, the thick one flows low­concentration 
waste solution, and the other (the thin one) flows high­concentration waste solution.

97 
URIT‐8021A Service Manual 

Figure 8­12 Waste Solution Barrel 
4)  Pull out BNC signal line. 

Figure 8­13 Hold the BNC Line  Figure 8­14 Unscrew BNC Line 

Figure 8­15 Pull Out BNC Line 
5)  Screw off cover of waste solution cover. 
6)  Discharge the waste solution into the specified pool.
98 
URIT‐8021A Service Manual 

7)  Insert waste solution pipe. 
8)  Connecting BNC line. 
Insert BNC line into joint and tighten. 

Figure 8­16 Hold the BNC Line                              Figure 8­17 Insert BNC Line 

Figure 8­18 
9)  Screw on the cover of waste solution barrel in a clockwise direction. 

8.2.4 Check Cleaning Position of Reagent and Sample Tray, Check Dilution Position 

Check whether the detergent of No.60 position of reagent tray is enough or not, if not, sufficient it. 
Check whether the detergent of No.59 position of reagent tray is enough or not, if not, sufficient it. 

8.2.5 Check/Clean Aspirating Probes, Stirrer and Cuvettes 

Check aspirating probes, stirrer and cuvettes. If there are any obvious stains, clean it according to weekly 
maintenance. 
Click Execute to perform probes and stirrer cleaning and cuvette rinsing after power on. See figure 8­19.

99 
URIT‐8021A Service Manual 

Figure 8­19 
Performing probe and stirrer cleaning, cuvette rinsing and add water before exit system. See figure 8­20. 

Figure 8­20 

8.2.6 Check Printer and Printing Paper 

Check the work condition of printer everyday and make sure the printing paper is enough. 

8.3 Weekly Mintenance 

Do the following maintenance every week to keep analyzer in best work state.

100 
URIT‐8021A Service Manual 

8.3.1 Aspirating Probes and Stirrer 

CAUTION 

Be careful to operate to avoid being scractched by the needle tip. 

Biological Hazard 
1)  Be sure to put on protective gloves, clothes, or even goggles if necessary. 
2)  Do not discard gauze which used for cleaning aspirating probes. Dispose it according to 
relative regulation. 
1)  To make sure the test switch is off. 
2)  Gently uplift the aspirating probe rocker to the highest point and turn it to a position where convenient for 
operating. 
CAUTION 
Do not touch the surface of probes and stirrer with tweezer when clean to avoid scratch. And 
avoid apply too much pressure to make probe and stirrer bending. 

3)  Pick up a piece of gauze which with absolute alcohol to wipe surface of aspirating probe and stirrer from 
up to down. Especial to wipe the needle tip until the surface is bright and clean. 

Figure 8­21 Clean Aspirating Probe

101 
URIT‐8021A Service Manual 

Figure 8­21 Clean Stirrer 

4)  Switching on testing switch 
5)  Open system software and click Maintenance  Probe and Stirrer Cleaning 

Figure 8­22 
6)  Click RESET, aspirating probes and stirrer will turn to washing pool and washed by distilled water. 

Figure 8­23 

8.3.2 Clean Reagent Tray and Sample Tray 

Biological Hazard 

1)  Be sure to put on protective gloves, clothes, or even goggles if necessary. 
2)  Do not discard gauze which used for cleaning aspirating probes. Dispose it according to 
relative regulation. 
1)  To make sure the test switch is off.

102 
URIT‐8021A Service Manual 

2)  Loosen the two fixed screws on reagent and sample tray. 

Figure 8­24 
3)  Uplift the reagent tray bracket and pull out. 

Figure 8­25 
4)  Using  a  piece  of  gauze  with  absolute  alcohol  to  clean  the  surface  of  reagent  and  sample  tray  and 
screws of holder, until there are no stains on the surface of reagent tray. 
5)  Using a piece of gauze with distilled water to clean the surface of reagent and sample tray and screws of 
holder, until there are no alcohol trace on the surface. 
6)  Install the reagent and sample tray, tighten fixed screw then put on cover. 

8.3.3 Clean Washing Pool

103 
URIT‐8021A Service Manual 

Caution 
Be careful to operate to avoid being scractched by needle tip. 

Biological Hazard 
1)  Be sure to put on protective gloves, clothes, or even goggles if necessary. 
2)  Do not discard the cotton swab which used for clean washing pool. Discord it according 
to the relative regulation. 

1)  To make sure the test switch is off. 
2)  Gently uplift the aspirating probes and stirrer rocker to the highest point and turn to a position where 
convenient for operating. 
3)  Using a piece of cotton swab with alcohol to clean surface and around of washing pool until there are no 
stains. 

Figure 8­26 clean washing pool  Figure 8­26 clean washing pool 
4)  Switching on test switch. 
5)  Open system software and click Maintenance  Probe and Stirrer Cleaning 

Figure 8­27 
Click Reset, aspirating probe and stirrer will turn to washing pool and cleaned by distilled water.

104 
URIT‐8021A Service Manual 

Figure 8­29 

8.3.4 Clean Reaction Tray 

Biological Hazard 

1)  Be sure to put on protective gloves, clothes, or even goggles if necessary. 
2)  Do not discard gauze randomly, please dispose according to relative regulation. 
1)  To make sure the test switch is off. 
2)  Wipe the surface and around of reaction tray by gauze with distilled water until there are no stains. 

Figure 8­30 
CAUTION 

DO  NOT  wipe  reaction tray by  gauze  which  with  absolute  alcohol  or detergent since  it  will 
cause corrosion of reaction tray. All losses thus incurred should borne by users. 
CAUTION 
The gauze should not with too much distilled water to prevent excessive water flow to cuvette.

105 
URIT‐8021A Service Manual 

8.3.5 Clean Washing Hand 

CAUTION 
Be careful to operate to avoid being scractched by needle tip. 

Biological Hazard 
1)  Be sure to put on protective gloves, clothes, or even glasses if necessary. 
2)  Do not discard gauze randomly, please dispose according to relative regulation. 

1)  Be sure the test switch is off. 
2)  Using a piece of gauze with distilled water to clean the surface of washing hand from up to down until 
there are no stains. 

Figure 8­31 Clean Washing Hand 

CAUTION 
Do not touches the surface of probe by tweezer to avoid scratch washing hand; and avoid 
applies too much pressure to bend washing hand probe. 

8.3.6 Clean Panel 

Biological Hazard 
Be sure to put on protective gloves, clothes, or even glasses if necessary. 

Do not discard gauze randomly, please dispose according to relative regulation. 

1)  To make sure the test switch is off. 
2)  Using a piece of gauze with distilled water to clean panel, cover of reaction tray and reagent tray until 
there are no stains. 
CAUTION 
The distilled water contain in gauze should not too much to avoid flow into crack of panel.

106 
URIT‐8021A Service Manual 

8.3.7 Deep Clean Cuvette 

Deep clean cuvette weekly to extend life and clean the deposited dirt on the inwall of cuvette. 
1)  Fill detergent barrel with detergent. 
2)  Click Maintenance  Cuvette rinsing  DeepClean. 

Figure 8­32 

8.3.8 Check AD Value and Save Cuvette Blank 

1)  Switching on test switch 
2)  Click Maintenance  Cuvette signal. 

Figure 8­33 
3)  Click Add Water 

Figure 8­34 
4)  Click Cuv. Blank Reading
107 
URIT‐8021A Service Manual 

Figure 8­35 
5)  Check the AD value in the list 

Figure 8­36 Interface of AD value 
Check the AD value of all cuvettes by clicking the arrows of left and right. 

Figure 8­37 
All the AD value should be in the range of 40000 to 60000. 
Please refer to the relative contents in the chapter of Check/Change Cuvette if value not in the range. 
6)  Click SAVE to save cuvette blank. 

Figure 8­38 
7)  Click Drain water to pump all water out. 

Figure 8­39

108 
URIT‐8021A Service Manual 

8.4 Monthly Maintenance 

Do the followings maintenance every month to keep analyzer in best work state. 

8.4.1 Clean Incubation Bath 

1)  To make sure the test switch is off. 
2)  Screw off the fixed screws on cuvette bracket. 

Figure 8­40 Screw off Fixed Screws on Bracket 
3)  Take out cuvette bracket and put on clean desktop. 

NOTE 
Do not scratch the cuvtte and stain the cuvette when take out the cuvette bracket. 

4)  Take out cuvette bracket in sequence. 

Figure 8­41 Reaction Tray 

NOTE 
The cuvette bracket which under the washing hand is not easy to take out. Rotate the 
reaction tray to a proper position then take out.

109 
URIT‐8021A Service Manual 

5)  Using a clean gauze to clean the incubation slot. 

Figure 8­42 Clean Incubation Slot 

NOTE 
Do not scratch the light hole when clean the incubation slot. 

6)  Put  the  cuvette  bracket  to  the  incubation  slot,  the 
location hole fix to location pin. 

Figure 8­43 Location Hole                            Figure 8­44 Loaction Pin 

Figure 8­45 Two Sides in Level 
7)  Screw on fixed screws and tighten.
110 
URIT‐8021A Service Manual 

8)  To fix six cuvette brackets well in sequence. 
9)  Do the steps as 8.3.8 and save cuvette blank. 

8.4.2 Clean Distlled Water Barrel 

1)  To make sure the test switch is off. 
2)  Pull out distilled water pipe. 
Press the circle and take out distilled water pipe. 

Figure 8­46 Pull Out Joint 

3)  Screw off BNC signal line 

Figure 8­47 Hold BNC Signal Line                    Figure 8­48 Screw off BNC Line

111 
URIT‐8021A Service Manual 

Figure 8­49 Hold BNC Line 
4)  Screw off cover of distlled water barrel. 
5)  Pour out the rest distilled water. 
6)  Clean the inwall of distilled water barrel, using a brush to clean if necessary. 

NOTE 
Do not scratch the liquid lever sensor, catheter and filter when using a brush to clean. 

7)  Using a clean gauze to clean the external and cover of distilled water barrel. 
8)  Insert distilled water pipe. 

Figure 8­50 Insert Pipe 
9)  Connecting BNC line. 
Insert BNC signal line and tighten.

112 
URIT‐8021A Service Manual 

Figure 8­51 Hold BNC Line                                    Figure 8­52 Screw off BNC Line 

Figure 8­53 Hold BNC Line 
10)  Screw on cover of distlled water barrel. 

8.4.3 Clean Waste Solution Barrel 

Biological Hazard 
Be  sure  to  put  on  protective  gloves,  clothes,  or  even  goggles  if  necessary.  Discharge 
waste  solution  according  to  local  discharge  standard  and  consult  relative  reagent 
manufacturer and distributor. 

1)  To make sure the test switch is off. 
2)  Screw off waste solution pipe. 
3)  Pull out waste solution pipe 
Press circle and pull out waste solution pipe.

113 
URIT‐8021A Service Manual 

Figure 8­54 Hold Joint of Pipe 

4)  Screw off BNC signal line 

Figure 8­55 Hold BNC Line                                    Figure 8­56 Screw off BNC Line 

Figure 8­57 Hold BNC Line 
5)  Screw off cover of waste solution barrel. 
6)  Pour waste solution out to the specified pool. 
7)  Clean inwall of waste solution barrel by distilled water, using a brush to clean if necessary.

114 
URIT‐8021A Service Manual 

NOTE 
Do not scratch the liquid level sensor when use a brush to clean. 

8)  Using a clean gauze to clean the external and cover of waste solution barrel. 
9)  Insert waste solution barrel. 

Figure 8­58 Insert Pipe 
10)  Insert BNC signal line. 
Insert BNC signal line to the joint and tighten. 

Figure 8­59 Hold BNC Line                                    Figure 8­60 Insert BNC Line 

Figure 8­61 hold BNC Line 
11)  Screw on cover of waste solution barrel.
115 
URIT‐8021A Service Manual 

8.4.4 Clean Detergent Barrel 

1)  To make sure the test switch is off. 
2)  To pull out detergent pipe. 
Press circle and pull out detergent pipe. 

Figure 8­62 Pull Out The Pipe 
3)  Screw off BNC signal line. 

Figure 8­63 Before Screwing                                                Figure 8­64 After Screwing 
4)  Screw off cover with liquid level sensor of detergent barrel. 

Figure 8­65 Screw off Cover of Detergent Barrel        Fiugre 8­66 Take out Cover with Liquid Level Sensor

116 
URIT‐8021A Service Manual 

NOTE 
Take out cover with liquid level sensor and put on clean desktop. 

5)  Pour the allowance of detergent into a clean container. 
6)  Clean the inwall of distilled water barrel by distilled water, using a brush to clean if necessary. 

NOTE 
Do not scratch the catheter and filter when using a brush to clean. 

7)  Clean liquid level sensor of detergent barrel by distilled water, using gauze to clean if necessary. 
8)  Using clean gauze to clean the external of detergent barrel. 
9)  Screw on cover with liquid level sensor of detergent barrel. 
10)  Connecting BNC signal line. 
Insert the BNC signal line to the joint and tighten 

Figu 
re 8­67 Before Screwing  Figure 8­68 After Screwing 
11)  Insert detergent pipe. 

Figure 8­69 Insert Pipe
117 
URIT‐8021A Service Manual 

8.4.5 Clean Drive Shaft of Aspirating Probe 

1)  To make sure the power switch is off. 
2)  Uplift the arm of aspirating probe to the highest point. 
3)  Using clean gauze to clean drive shaft 
NOTE 

Do  not  use  alcohol  or  other  corrosive  detergent  to  clean  drive  shaft,  otherwise  will  cause 
stuck. Please use specified lubricant oil to maintain drive shaft. 

8.4.6 Clean Drive Shaft of Stirrer 

1)  To make sure power switch is off. 
2)  Uplift the arm of stirrer to the highest point. 
3)  Using clean gauze to clean drive shaft of stirrer. 
NOTE 

Do not use alcohol or other corrosive detergent to clean drive shaft, otherwise will cause 
stuck. Please use specified lubricant oil to maintain drive shaft. 

8.4.7 Check Washing Hand 

1)  Click Maintenance  Cuvette rinsing. 


2)  Check the rinse process to see whether the probe of group 1 to group 6 is parallel and in level or not, to 
check whether probe tip of group 7 is parallel with the lower end surface of block or not. 
3)  To check whether there are stains or crack on probe tip or block. Clean or change probes if necessary. 
NOTE 

Do  not  use  alcohol  to  clean  washing  hand,  using  the  detergent  provided  by  URIT  to 
clean will be better. 

8.5 Maintenance for Every Three Months 

8.5.1 Change Cuvette and Cuvette Bracket 

It will appear scratch, stains on cuvette if used for a period. So it suggests replacing all cuvettes for every 
three months to guarantee the accuracy of test result. 

8.5.1.1 Detect Cuvette 
1)  Click Maintenance  Cuvette Rinsing:

118 
URIT‐8021A Service Manual 

Figure 8­70 

2)  Click Add water. 

Figure 8­71 
Click Cuv. Blank Reading 

Figure 8­72 
3)  Observe the AD value 

Figure 8­73 
Check AD value of all cuvettes by clicking arrows of left and right.

119 
URIT‐8021A Service Manual 

Figue 8­74 
4)  All AD values should be in the range of 40000 to 60000. If not, do the following step: 
a)  If the AD value exceeds 60000, adjust the potentionmeter according to 3.5.7. 
b)  If the AD value under the 40000, take out the cuvette which irradiate by light, observe the AD value, if it 
is in the range, replace the cuvette; if there are no change or reduced, replacing light source. (refer to 
8.6.3) 
c)  If the AD value of one row is too small, do the following steps: 
To confirm the cuvette number which is problematic, the cuvette number corresponds with the number 
on the left of list. 
Screw off the fixed screws on the corresponding cuvette bracket. 

Figure 8­75 Screw off the fixed screws on cuvette bracket 
Take out cuvette bracket and observe if there are any problems on cuvette. 

Figure 8­76 Take out cuvette 
If there still have bubbles in cuvette, do step 1 to 4 again. 
If there still have obvious stain on cuvette, clean cuvette by distlled water, if necessary, soak cuvette for ten 
minutes  by detergent  then  clean  by distilled  water  again.  If the  cuvette  still  has  stains,  refer  to  8.5.1.2 to 
replace cuvette. 
Refer to 8.5.1.2 to replace cuvette if there is fracture or scratch.

120 
URIT‐8021A Service Manual 

8.5.1.2 Replacing Cuvette 
Biological Hazard 
Be sure to put on protective gloves, clothes, or even goggles if necessary. 

1)  Screw off fixed screws on cuvette bracket. 
2)  Take out bracket and cuvette. 
3)  Put new cuvette and bracket to incubation slot, fixed location hole to location pin on the reaction tray. 
The two side of bracket should be in level with the near one. 
NOTE 
Do not touch the wall of cuvette to avoid stain on fingerprint. 

Figure 8­77 Location hole                        Figure 8­78 Location pin 

Figure 8­79 Two side in level with near one 
4)  Screwing on fixed screws and tighten. 
5)  Do step 2 to 4 repeatly. 

8.5.2 Clean Distilled Water Tank and Negative Pressure Tank for Draining Waste Solution 
1)  To make sure power switch is off. 
2)  Open rear baffle.
121 
URIT‐8021A Service Manual 

3)  Empty water in distilled water tank. 
4)  Exhaust negative pressure in the tank which used for draining waste solution. 
5)  Screw off distilled water tank and negative pressure tank by hands or special tool. 
6)  Clean tank 3 to 4 times by water and dry inner by clean cloth. 
7)  Reinstall relative components. 

Figure 8­80 

8.6 Non‐Periodical Maintenance 

8.6.1 Clear/Replace Aspirating Probe 

This chapter mainly introduced the methods when aspirating probe is blocking, staining etc. 

CAUTION 
Be careful to operate to avoid scratch finger. 
Biological Hazard 

Be sure to put on protective gloves, clothes, or even goggles if necessary. 

1)  To make sure the test switch is off. 
2)  Observe the position of aspirating probe over the washing pool. 
3)  Uplift  the  arm  of  aspirating  probe  to  the  highest  point,  then  turn  to  a  position  where  convinent  for 
operation. 
4)  Lightly pinch the block of the probe cover and take out the cover.

122 
URIT‐8021A Service Manual 

Figure 8­81 Take out cover of aspirating probe 

NOTE 
Do not lose the spring and retaining screw when take out the cover of probe. 

5)  One hand gently hold the joint of aspirating probe, the other hand pinches the joint of liquid level pipe 
and take out. 

Figure 8­82 Take out plug of aspirating probe 

NOTE 
Pinch the joint of liquid level pipe with sand paper to increase friction force. 

6)  Using needle cleaner to clean inner of aspirating probe until it is unblocked. 
7)  Injecting water to aspirating probe from joint to inner by syringe until the dirt are all discharged. 
8)  If  the  probe still  block after  cleaing for  several  times,  or  there are  scratch on  inner  and  outer  wall  of 
probe, replacing a new aspirating probe. 
Put the new aspirating probe to location hole from up to down. 
9)  One hand gently holds the joint of aspirating probe, the other hand hold the Teflon tube, insert the probe 
to the tube.
123 
URIT‐8021A Service Manual 

Figure 8­83 The joint of probe completely insert to tube 
10)  Put on the cover of aspirating probe. 
NOTE 
1)  Do not scatch the inner wall of probe when clean the probe by needle cleaner. 
2)  The tube must be plug tightly to avoid leakage. 
3)  If the front of tube is damage, cut a little if necessary. 
4)  Attention the spring can not hook up the aspirating probe, otherwise the anti­collision 
will be caused. And the cover should fasten and sealing. 

11)  Switching on test switch. 
12)  Click Maintenance  Probe and Stirrer Cleaning. 
13)  Observe the position of aspirating probe over washing pool; it should the same with the position which 
observed in step 2. 
If there exsits great deviation, do the above steps again. If the problem still exists, please contact URIT. 

8.6.2 Replacing Stirrer 
1)  To make sure the power switch is off. 
2)  Preparing a new stirrer, using gauze or cotton swab with detergent to clean stirrer, then clean by gauze 
with distilled water. 
3)  Uplift the arm of stirrer to the highest point, and then turn it to a position where convenient to operate. 
4)  Pull out cover of arm of stirrer from down to up. 
5)  Loosen two fixed screws and joint of stirring motor line.

124 
URIT‐8021A Service Manual 

Figure 8­84 

6)  Take out stirrer. 
7)  Loosen two screws on the micromotor adjusting plate, then take out the plate. 
8)  Install the micromotor adjusting plate on the component of new stirrer. 

Figure 8­85 

9)  According to 4.3 to install the stirring component on stirring arm. 
NOTE 
The position of micromotor adjusting plate should the same with origin. 

8.6.3 Replacing Lamp 
The  service  life  of  lamp  is  1500  hours.  Replacing  the  lamp  if  the  lamp  is  damage  or  achieve  rated  life

125 
URIT‐8021A Service Manual 

according to the following steps: 
1)  Turn off light and wait 15 to 20 minutes until it is cool. 
2)  The lamp component is located on the bottom of reaction tray. 
3)  Take out cuvette bracket, then take out cover of reaction tray. See figure 8­86 

Figure 8­86 

4)  Loosen jackscrew and power joint of light, then take out lamp. 
5)  Connecting the lamp well and insert into lamp­socket. 
6)  Turn on power switch and install the cover of reaction tray. 
7)  Run the software, click RESET or Roate to Cuv.1, when the original position of cuvette is confirm, click 
Maintenance  Cuvette signal 
8)  Take out cover of reaction tray again and adjusting the position of lamp, then observe the A/D signal. 
9)  When the A/D signal value is biggest, tighten the jackscrew to fix lamp. 
10)  At last install the cover and other components of reaction tray. 
NOTE 
If the reaction tray does not move to original position, the A/D value could not be read.

126 
URIT‐8021A Service Manual 

Figure 8­87 Lamp components 

8.6.4 Solenoid Valve Rinse 
It will cause water and air leakage if the solenoid valve is block. So it should better to clean valves irregularly. 
1)  Take out solenoid valve and push off circlip. 
2)  Loosen screw cap and take out piston rod. 
3)  Clean cavity and piston rod by distilled water. 
4)  Reinstall and test after power on. 

Figure 8­88 

8.6.5 Check/Replacing BNC Line 
1)  To make sure the test switch is off. 
2)  Check whether there is rupture, wire expose and damage or not, etc. If not, it is not need to replacing 
BNC line, if so, do the following steps: 
3)  Screw off BNC line and replacing a new one.

127 
URIT‐8021A Service Manual 

NOTE 
Attention  one  to  one  correspondence  when  screw  off  the  BNC  line  to  avoid  wrongly 
insert. 

4)  To connect BNC line. 
Insert the line and fixed well. 

8.6.6 Replacing Waste Solution Tube 
The waste solution tube will appear block after long time use. It is suggest to replacing the specified tube. 
There are two specification tubes for URIT­8021A. 

8.7 Long­time Disuse Maintenance 

If the instrument will not use for a long time, do the following steps before power off. 
1)  Click Maintenance  Cuvette Rinsing. 

Figure 8­89 
Click WASH to rinse all cuvettes. 

Figure 8­90 

2)  Repeat step 1 at least two times. 
3)  Pull out distilled water tube. 
Press circle and pull out tube. 
NOTE 
Put the tube on a clean desk after pulling out.

128 
URIT‐8021A Service Manual 

Figure 8­91 Pull out tube 
4)  Click Maintenance  Cuvette Signal; 

Figure 8­92 
Click Water Probe and Water Arm repeatly until there are no waste solution discharge from waste solution 
tube. 

Figure 8­93 

5)  Empty distilled water barrel, detergent barrel and waste solution tube. 

Biological Hazard 
Be sure to put on protective gloves, clothes, or even goggles if necessary. 

According to the local discharge standard to dispose waste solution and consult reagent 
manufacturer or distributor. 

6)  Turn off test switch, main power switch and pull out plug. 
7)  Close the up cover. 
8)  Put on a dust cover.

129 
URIT‐8021A Service Manual 

9 TROUBLESHOOTING 
This  chapter  lists  the  various  malfunctions,  along  with  probable  causes  and  recommended  remedies  to 
correct the problem quickly and easily. If the problem still exists after following the recommended remedy, 
contact URIT for Technical support. 
CAUTION 
Handle malfunction with utmost care and confirm if it is necessary to cut off the power supply at the 
first. 

Biological Hazard 
Put on protective gloves to avoid contacting chemical solution. If it spill to human body, wash it off 
with water immediately. 

9.1 Troubleshooting Guide 

To  eliminate  malfunction  easily  and  correctly,  users  should  read  through  the  Operating  Manual  and  be 
familiar  with  the  routine  operation  and  maintenance  of  URIT­8021A  .In  general,  there  are  three  steps  to 
handle with malfunction: 
Ø  Step 1: Confirming Malfunction 
Users should not only confirm the malfunction, but also clearly know what the normal status should be when 
the malfunction is eliminated. 
Ø  Step 2: Classifying Malfunction 
Malfunction can be categorized into three types. 
l  Malfunction relating to hardware 
l  Malfunction relating to software 
l  Malfunction relating to operation and analyses. 
If  malfunction  relates  to  the  hardware  or  software,  contact  your  local  distributor  or  URIT.  If  malfunction 
relates to the operation and analyses, refer to the troubleshooting table below for solution. 
Ø  Step 3: Eliminating Malfunction 
The maintenance engineer authorized by URIT takes proper measures to correct the problem. 
Users can also eliminate the malfunction under the directions of maintenance engineer. 

9.2 Obtaining Technical Help 

Our Customer Service Office is available to help if the problem is beyond the scope of this manual or if you 
need more technical assistance from URIT. Before calling, please identity the following information: 
1)  The instrument’s model; 
2)  The serial number of instrument; 
3)  The specific symptom and operating condition; 
4)  The data and report relating to the problem.
130 
URIT‐8021A Service Manual 

9.3 Troubleshooting Method 

The  troubleshooting  table  below  presents  the  various  problems  and  malfunctions  that  may  occur  during 
operation. If the problem can not be solved through the recommended methods, contact URIT please. 
NOTE 
For replacement parts of the instrument, refer to Appendix 1. 

Table 9­1 Troubleshooting 
SN  SYPTOM  POSSIBLE CAUSE  REMEDY 
1) Connect power cord correctly. 
1)  Incorrect  connection  with 
2)  Check  if  the  power  receptacle  is  in  good 
power cord. 
condition. 
2)  No  electricity  with  power 
3) Replace fuse(8A) 
Instrument  is  not  receptacle. 
4)Confirm  select  the  correct  connector  at 
active  when  power  3) The safety fuse is fusing 
1  [System  parameter—instrument 
is on.  4)  Improper  COM  interface 
parameter—COM set] 
is selected. 
5) Make sure the RS232 communication cable 
5)  Communication  cable 
is connected to PC correctly. 
error. 
If  the  problem  still  persists,  contact  your  local 
distributor or URIT. 
1) Select [Maintenance—Cuvette rinse], 
Check 
if  reaction  cuvettes  are  dirty  or  damage. 
Cuvette dirty or damage  Replace them if necessary 
2  Cuvette blank error 
Light source aging  2) Replace light source. 
3)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 
1) Check or replace the lamp holder 
1)  Bad  contact  of  lamp 
2) Replace the lamp. 
3  Lamp is dark  holder. 
3)  If  the  problem  persists,  contact  your  local 
2) Lamp is burned out. 
distributor or URIT. 
1)  Check  flow  path  tubes.  Reconnect  or 
1)  Air  leaks  in  the  flow  replace tubes if necessary. 
paths.  2)  Reconnect  meter  regulator.  Eject  air 
Inaccurate 
2) Air bubbles are formed in  bubbles. 
aspirated volume of 
4  meter regulator.  3) Unclog or replace probe. 
reagent or sample. 
3) Probe is clogged.  4)  Check  magnetic  valve  and  replace  it  if 
4) Magnetic valve problem.  necessary. 
5)  If  the  problem  persists,  contact  your  local 
distributor or URIT.

131 
URIT‐8021A Service Manual 

Water or detergent 
1) Flow path tube is leaky. 
does  not  come  out  1) Reconnect the tube or replace it. 
2)  Flow  path  tube  is 
through  the  probe  2) Unclog the tube. 
clogged. 
5  washing  pool  or  3) Replenish water or wash solution. 
3) Water or wash solution is 
stirrer  washing  4)  If  the  problem  persists,  contact  your  local 
used up. 
pool.  distributor or URIT. 

1) Reconnect the tube or replace it. 
1) Flow path tube is leaky.  2) Unclog the tube. 
2)  Flow  path  tube  is  3)  Check  the  vacuum  pump  and  replace  it  if 
Adding  or  draining 
6  clogged.  necessary. 
water is abnormal 
3) Vacuum pump error.  4)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 

A  certain  movable  1)  Check  the  wire of  light  coupling  or  replace 
Light  coupling  is  short 
part  is  out  of  the light coupling. 
7  circuited or broken. 
control.  2)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 
1) Liquid level sensor board  1)  Check  the  liquid  level  sensor  board  and 
Liquid  level  sensor  is defective.  replace it if necessary. 
8  is out of order  2)  Bad  contact  with  liquid  2) Reconnect the liquid level sensor board. 
level sensor board.  3)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 
1) Select Maintenance –­­ Cuvette Rinsing to 
1)  Cuvette  is  dirty  or  check  whether  cuvette  is  dirty.  Clean  or 
breakage.  replace the cuvette. 
2)  Inaccurate  aspirated  2)  Check  sample  injector  and  tube  whether 
volume  of  reagent  or  leakage existed. 
Inaccurate  test 
sample.  3) Replace the lamp. 
result  or  poor 
3) Lamp is deteriorated.  4)  Set  the  parameter  follows  the  operation 
9  repeatability. 
4)  Parameters  of  analytical  manual.  Make  sure  the  instrument  is  well 
item are set improperly.  grounded by means of the ground pole. 
5)  Ground  wire  is  absent  5)  Check  if  the  reagent  is  certified.  Perform 
with power supply.  recalibration. 
6) Reagent problem.  6)  If  the  problem  persists,  contact  your  local 
distributor or URIT.

132 
URIT‐8021A Service Manual 

1) Turn the power switch off, slowly rotate the 
problem parts and observe if there is abnormal 
1) Communication error.  noisy or if it is stuck. 
2)  Mechanical  parts  are  2)  Enter  into  system  parameter  interface, 
loose or stuck.  select Engineer set to adjust parameter. (Only 
Abnormal  motor  3)  Light  coupling  joint  of  accessible  to  professional  person  authorized 
10 
movement  motor is loose.  by URIT). 
4)  Light  coupling  is  3)  Check  the  light  coupling  and  replace  it  if 
defective.  necessary. 
4)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 

1) Replace the stirrer motor. 
Stirrer  does  not  1) Stirrer motor is broken.  2) Reinstall the stirrer motor. 
11  work.  2)  Bad  contact  of  stirrer  3)  If  the  problem  persists,  contact  your  local 
circuit.  distributor or URIT. 

1)  Check  the  flow  path  tube.  Reconnect  or 


replace the tube if necessary. 
1) Flow path tube is leaky.  2)  Check  the  magnetic  valve  and  replace  it  if 
Water  leaks  from 
2) Magnetic valve problem.  necessary. 
12  nozzles  of  cuvette 
3) Vacuum pump problem.  3)  Check  the  vacuum  pump  and  replace  it  if 
cleaner. 
necessary. 
4)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 
1)  Check  the  flow  path  tube.  Reconnect  or 
replace the tube if necessary. 
2)  Check  the  magnetic  valve  and  replace  it  if 
1) Flow path tube is leaky. 
necessary. 
Water drops adhere  2) Magnetic valve problem. 
13  3)  Check  the  exterior  of  probe.  Clean  or 
to the tip of probe.  3)  The  exterior  of  probe  is 
replace the probe if necessary. 
dirty or scathed. 
4)  If  the  problem  persists,  contact  your  local 
distributor or URIT.

133 
URIT‐8021A Service Manual 

1)  Check  if  the  reagent  tray  is  sealed 


completely. 
2)  Check  if  the  heat  dissipation  device  works 
The cooling system is failed  normally. 
Reagent  tray 
or  cooling  temperature  is  3) Check if the refrigerants are used up. 
14  cannot be cooled. 
not low enough.  4)  Check  if  the  circulation  system  of  the 
cooling device works normally. 
5) Replace the peltier. 
6)  If  the  problem  persists,  contact  your  local 
distributor or URIT. 
1)  Enter  into  system  parameter  interface, 
select Engineer set to adjust parameter. (Only 
accessible  to  professional  person  authorized 
1)  Instrument  parameters 
by URIT). 
are set improperly. 
2)  Read  through  the  Operating  Manual  and 
2)  Human  carelessness, 
Probe  and  stirrer  avoid human mistake. 
such as reagent bottle lid is 
conflict  with  each  3)  DO  NOT  put  foreign  objects  on  the 
15  not opened. 
other.  operation panel. 
3) Put foreign objects on the 
4)  Check  if  the  motor  is  installed  and  works 
operation panel. 
properly. 
4) Motor problem. 
5)  Check  the  light  coupling  and  replace  it  if 
5) Light coupling error. 
necessary. 
6)  If  the  problem  persists,  contact  your  local 
distributor or URIT.

134 
URIT‐8021A Service Manual 

CHAPTER 10 Measurement Method 
URIT­8021A  automatic  chemistry  analyzer  is  based  on  the  Lambert’s  Beer  principle  to  analyse  and 
calculate. This chapter will mainly introduce the calculation method. 

10.1 Method of Measurement 
The URIT­8021A performs measurement with the following methods: 
1)  Endpoint method 
2)  Two­point end method 
3)  Two­point rate method 
4)  Rate method(Kinetic method) 

10.2 Endpoint 
In the endpoint measurement, the reaction reaches equilibrium after a period time, it can be considerate that 
all analytes(substrates) have changed into products. It’s best that the absorbance is directly proportion to the 
analytes’ concentration. 
Endpoint method (single reagent) reaction graph see as figure 10­1: 

Figure 10­1 Single Reagent Endpoint Method Graph 

R1: Add first reagent 
S: Add sample 
a: Test start point of absorbance 
b: Test end point of absorbance 
1  5: Test point of reagent­blank absorbance 
The  URIT­8021A  system  records the  absorbance  change  start from  adding  R1 and  end at  the  maximum 
reaction time. 
n  Setup of reagent parameter: 
From S to a: The first incubate time of first reagent 
From a to b: The number of test points 
n  Result calculation:

135 
URIT‐8021A Service Manual 

C =( A 1  ­ A 0 )´ K 
C: The concentration of reactant 
K: Calculation factor 

A1: The average of absorbance from a to b 

A0: The average of absorbance from the first test point to the fifth test point before adding sample. 

(R1 blank) 
NOTE 
When double reagents test use endpoint method, the result calculation is the same as the above 
calculation formula that test point absorbance substracts R1 blank, then  multiplied by K to get the 
concentration. 

10.3 Two‐Point End Method (Double Reagent Endpoint) 


Two­point end method add sample blank and volume correction factor to calculate, maily used in the double 
reagent endpoint. See figure 10­2: 

Figure 10­2 Double Reagent Two­Point End Method Graph 
R1: Add first reagent 
S: Add sample 
R2: Add second reagent 
A: Test start point of absorbance 
B: Test end point of absorbance 
c  d: Special blank test start point, in general to take five test points. System will take 
the blank test points automatically according to the time, if the set time is insufficient 
five test points 
1  5: Test point of reagent­blank absorbance 
The  URIT­8021A  system  records the  absorbance  change  start from  adding  R1 and  end at  the  maximum 
reaction time. 
n  Setup of reagent parameter: 
From S to R2: The incubate time of first reagent

136 
URIT‐8021A Service Manual 

From R2 to a: The incubate time of second reagent 
From a to b: The number of test point 
n  Result calculation: 

C = ( 
[  A 2 ­ A 0 )­ ( A 1  - A 0 ) ´ k1 ] ´ K 
C: The concentration of reactant 
K: Calculation factor 

K1: Volume correction factor 

A1: The average of absorbance from c to d. (Sample blank) 

A2: The average of absorbance from a to b 

A0: The average of absorbance from the first test point to the fifth test point before adding sample. 

(R1 blank) 

VR1 + VS 
k 1  =
VR 1 + VS  + VR 2 
VR1: The volume of first reagent 
VRS: The volume of sample 
VR2: The volume of second reagent 

10.3 Two‐Point Rate Method (Fixed‐Time) 


Two­point  rate  method  (Fixed  ­Time),  namely,  first­order  kinetic  method,  the  reaction  velocity  is  directly 
proportional  to  the  first­order  of  substrate  concentration  within  a  special  period.  As  the  substrate  is 
consumed continuously, the velocity is descreasing gradually, and the absorbance is nonlinear increasing 
descreasing). See the graph as 10­3: 
Two­point  rate  method  allows  the  check  of  substrate  depletion.  When  detecting  substrate  depletion,  the 
system will perform the related operation and flag the test result. 

Figure 10­3 Two­Point Rate Method Graph (Single reagent, positive direction reaction)
137 
URIT‐8021A Service Manual 

R1: Add first reagent 
S: Add sample 
a: Test start point of absorbance 
b: Test end point of absorbance 
The  URIT­8021A  system  records the  absorbance  change  start from  adding  R1 and  end at  the  maximum 
reaction time. 
n  Setup of reagent parameter: 
From S to a: The incubate time of first reagent 
From a to b: The number of test point 
n  Result calculation: 

æ A  - A a  ö
C  = çç b ÷÷ ´ K 
è t b  - t b  ø
C: The concentration of reactant 
K: Calculation factor 

Aa: The absorbance of point a 

Ab: The absorbance of point b 

ta: The time of point a 

tb: The time of point b 
NOTE 
1)  tb­ta  is the time that the number of test points multiplied by the period time. 
2)  The double reagent calculation is the same as single reagent. 

10.4 Rate Method (Kinetic method) 


Rate  method  (kinetic  method)  namely  zero­order  kinetic  method,  the  reaction  velocity  is  not  related  to 
substrate concentration. The analytes absorbance descreasing or inscreasing evenly at a given wavelength, 
and the change rate is directly proportional to the activity of the analytes. The kinetic method is usually used 
to measure enzyme activity. See graph 10­4. 
In fact, it’s impossible for the substrate concentration to be absolutely high, and the reaction will be no longer 
a zero­order reaction when the substrate is consumed to certain degree. Therefore, the substrate depletion 
appears in the rate method. In addition, the reaction can become steady only after a period of time, because 
the reaction is complicated at the beginning and there are miscellaneous reactions due to complex serum 
compositions. Please set the reagent parameter according to the reagent manual that provided by reagent 
manufacturer.

138 
URIT‐8021A Service Manual 

Figure 10­4 Rate Method Graph(Single reagent, negative direction reaction) 
The  URIT­8021A  system  records the  absorbance  change  start from  adding  R1 and  end at  the  maximum 
reaction time. 
n  Setup of reagent parameter: 
From S to a: The incubator time of first reagent 
From a to b: The number of test point 
n  Result calculation: 

C  = DA ab / min ´ K 
C: The concentration of reactant 
K: Calculation factor 

DAab  / min : The absorbance change rate from a to b. 

NOTE 
The double reagent calculation is the same as single reagent. 

10.5 Substrate Exhaust 
In rate method and two­point rate method appears substrate exhaust because the concentration of part of 
sample is high. URIT­8021A provides two methods to check of substrate exhaust, can be choose two or one. 
But two­point rate method only chooses the method of first substrate exhaust. The graph 10­5 is substrate 
exhaust in the single reagent negative direction reaction.

139 
URIT‐8021A Service Manual 

Figure 10­5 

10.5.1 Substrate Exhaust Method 1(Substrate exhaust Limit) 

Make experiment to obtain the absorbance of substrate exhaust and set the limit. The average value of the 
last three points in positive reaction is greater than the absorbance value of substrate exhaust limit. If the 
average value smaller than the absorbance value in negative reaction, it could judge as substrate exhaust. 
The red box in the figure 10­6 is the substrate exhaust setup. 

Figure 10­6 

Figure 10­7 Substrate exhaust (Single reagent, positive direction)

140 
URIT‐8021A Service Manual 

10.5.2 Substrate Exhaust Method 2(Slope Ratio) 

The second method of substrate exhaust is using the slope ratio of reaction curve (absorbance change rate) 
to judge. 

The meaning of each parameter: 
First  slope  (First  read  point  section):  The  slope  in  the  first  read  point  section  that  setted  in  the  reagent 
parameter. 
Second  slope  (Second  read point  section):  The  slope  in  the  second  read point  section that  setted  in  the 
substrate exhaust method 2. 
Begin point: The begin read point of second slope ratio. It should be not greater than the begin read point of 
first  slope  ratio.  The  begin  point  of  single  reagent  is  the  point  that  start  to  add  sample,  and  the  double 
reagents is the point that start to add second reagent 
End point: The end read point of second slope ratio that less than the end read point of first slope. 
Slope ratio: First slope/ second slope 
When the calculated slope ratio is less than the set value, the system will judge for substrate exhaust. This 
method can be used simultaneously with the method 1 or used alone. But two­point rate method only use 
method 1. 
As long as choosing the substrate exhaust method 2, system will automatically take the second slope as the 
calculated  slope  when  the  instrument  judge  for  substrate  exhaust.  The  result  of  the  sample  is:  second 
slope  K. 
Caution 
The begin point of second slope is restricted by begin point of first slope. Begin point should be not 
greater  than  the  begin  point  of  first  slope,  and  end  point  not  greater  than  the  end  point  of  first 
slope.

141 
URIT‐8021A Service Manual 

Figure 10­8 
NOTE 
The judgement method of double reagent is the same as the method in 10.4.2.

142 
URIT‐8021A Service Manual 

Appendix A  URIT‐8021A Circuit Schematic Diagram 

8021A Circuit
Diagram.pdf

143 

You might also like