You are on page 1of 174

FOUNDRY-MASTER PRO

O X F O R D I N S T R U M E N T S A N A L Y T I C AL G M B H
AN OXFORD INSTRUMENTS COMPANY
Y O U R P AR T N E R F O R Q U AL I T Y
Contents

0.1 - Introduction Page 4


0.2 - Using this manual Page 5
0.3 - Safety instructions Page 8

Section 1

START-UP AND INSTALLATION


1.1 - Installation - Overview Page 10
1.2 - Utilities menu Page 12
1.3 - Burn analysis (quick start) Page 31

Section 2

ANALYSIS
2.1 - Analysis: General information Page 41
2.2 - Analysing Page 60
2.3 - Recalibration Page 70
2.4.1 - Grade mode Page 79
2.4.2 - Grade search Page 93
2.5 - Parameter settings Page 109
2.6 - Generating pseudoelements Page 117
2.7 - Producing a report file Page 128

Appendix A

A.1 - Spare parts and accessories Page 148


A.2 - Technical data Page 151

Page 2 Contents
Appendix B

B.1 - Maintenance Page 152


Problems/Frequently asked questions
(FAQs)
B.2 - Cleaning the lens Page 154
B.3 - Cleaning the window Page 159
B.4 - Changing the electrode Page 162
B.5 - Washing bottles Page 166
B.6 - Notes on nitrogen analysis Page 168
B.7 - Support Page 172

Appendix C

C.1 - Quality deviations Page 173


C.2 - Acceptance Certificate Page 175

Page 3
SECTION 0.1

Introduction

Dear Customer,

Congratulations on your purchase of the FOUNDRY-MASTER PRO.

The FOUNDRY-MASTER PRO is a precision measuring instrument that is


remarkably easy to use. This manual is designed to tell you everything you need to
know to get the most out of the spectrometer and keep it in good working order.
Please take the time to read it – you will reap the benefit later on when using the
FOUNDRY-MASTER PRO.

The FOUNDRY-MASTER PRO is a reliable, compact yet high-performance


laboratory spectrometer.

In analytical mode the FOUNDRY-MASTER PRO generates quantitative analyses


of materials. The analytical result is displayed on the screen and can be stored in
the internal database. By comparing the result with stored material data, you can tell
immediately whether the examined material meets the defined specification or how it
deviates from the specification.
The analytical data can be exported to another PC. Certificates can easily be
prepared and printed.
Important:
The reliability of your analysis is directly dependent on the quality of your
sample preparation.

The FOUNDRY-MASTER PRO requires only a small amount of maintenance


for efficient and reliable operation. The maintenance operations which you as user
can carry out are described in Appendix B of this manual. Information about spare
parts and accessories can be found in Appendix A.1 "Spare parts and accessories".

Oxford Instruments Analytical GmbH wishes you every success with your
FOUNDRY-MASTER PRO.

Introduction

Page 4
SECTION 0.2

Using this manual

Important:.

Because each individual spectrometer is configured according to the customer's


specific needs, the matrices and/or materials displayed on your screen may differ
from the illustrations in this manual.

Unless otherwise indicated, we assume the user has already started Windows™.
Basic knowledge of how to use software running under Windows™ is assumed.

As settings can be customised in Windows™ according to user preferences, the


appearance of menus and dialog boxes may differ from the examples shown here
(which are in "Windows™ Classic View"). We therefore recommend you set your PC
accordingly. However, we cannot guarantee that all illustrations in this manual will
necessarily match those on the screen of your PC.

1 - Minimise or close all applications running on the PC.


2 - Right-click in a vacant area of the Windows™ Desktop.
3 - On the context menu, click Properties to open the "Display Properties"
window.
4 - Click the "Themes" tab, from the listbox select "Windows™ Classic".
5 - Press "OK" to confirm the changes.

Important: Pay special attention to any bold or underlined text, or any passages
marked with Important or Note.

The names of executable commands are displayed in bold.

For example: "Print"

All menu and dialog box names are printed in bold and without exclamation mark.
For example: "Create / Change alloy".

Using this manual

Page 5
All figures are numbered by sections. If a particular figure or section is referred to in
the text, it is printed in bold.
For example: Section 2.3, Figure 10.

If an area is bordered in red, the subsequent text or arrow relates to the whole of the
area inside this marking, see Figure 1.

Figure 1 is not an illustration from the manual.


Select
option

Select to
activate
Figure 1

Tolerances of the
individual
elements.
Double-click to
These texts and
edit, see Figure
arrows…
11

…relate to the area


marked in red

Element lines with


intensity values:
Uneditable
information

Page 6 Using this manual


To simplify data input, our WASlab software features a useful screen editor (screen
keyboard). You open the screen editor by double-clicking the relevant field and
operate it with the mouse. You also have the option of using an external keyboard.
To make it easier to find your way around the software, the editor (keyboard) is
object-oriented. The current reference is shown to the left of the input field, in this
case File name (Figure 2).

Input field

Figure 2

Displayable columns on the screen

The number of visible columns depends on the screen resolution set on your
system.

Using this manual


Page 7
SECTION 0.3

Safety instructions

The FOUNDRY-MASTER PRO has been developed especially for use in


industrial production environments and is highly resistant to dust,
temperature changes and vibrations.
Nevertheless, the following points should be observed:

Never place liquids on top of the FOUNDRY-MASTER PRO.

Never place unduly heavy objects (>10 kg) on the cover of the FOUNDRY-MASTER
PRO.

Never place unduly heavy objects (>10 kg) on the spark stand of the FOUNDRY-
MASTER PRO.

In burn mode the FOUNDRY-MASTER PRO emits electromagnetic waves. This


radiation can cause radio interference. To minimise emitted interference, it is
necessary to set up and operate the unit in accordance with the instructions in this
manual. In the event of electromagnetic incompatibility, the user may need to carry
out additional measures. The following method may help to minimise interference:
- Place the unit further away from "receivers" (network/signal/
telecommunications cables, radio and television receivers)
- Specific matching with other systems
- Additional filtering measures for the mains connection
- Variation of the selected excitation parameters
- In certain circumstances, low-inductance earthing of the sample being analysed
may reduce interference.
Screening of the whole unit may be considered in certain cases.

The German Accident Prevention Regulations according to VDE guidelines must be


observed when using the unit.

Maintenance or repair must only be carried out by qualified experts. Never open the
housing. There are no user-serviceable parts inside the housing.

Regular inspection of the unit by a qualified expert will increase reliability and
prolong service life.

Only original parts should be used for repair.

Safety instructions

Page 8
Internal or external safety devices of the unit must on no account be disabled. This
applies especially to the overpressure valve on the cover of the vacuum tank.

The unit must only be used in a dry environment (protection class IP20).

The unit must be set up on a stable surface. Make sure it is far enough away from
the wall. Align the unit so that it is straight.

When fully shutting down the unit, be sure to close the ball valve between the
vacuum pump and vacuum tank and switch off the pump. Otherwise there is a
danger of the pump overheating as the housing fans are no longer working.

If you touch the unit at the same time as touching electrically conductive parts, such
as machines, instruments, etc., possible potential differences may give rise to
voltages which, though perceptible, are not dangerous.

Temperatures below 0°C or above 50°C may impair the operation of the TFT
screen.

Please observe the warning signs and notices on the unit.

In mains operation the unit must only be operated on systems with PE protective
earth.

Never carry out burn processes on wet or damp surfaces, in view of the danger of
short-circuit.

Be careful not to touch the burn electrode when the excitation source is switched on.

It is essential to turn off the excitation source when working on the sample stand,
e.g. when changing the electrode.

To avoid condensation in the unit, it should be acclimatised to the ambient


temperature before starting up.

Never look directly at the electric arc. Wear suitable safety goggles to protect your
eyes when working.

We recommend regular data back-up.

This unit (PC) has a battery containing potentially harmful


substances. It must not be thrown away with domestic waste.
At the end of its service life, it should only be disposed of
through the customer service department of WAS/Oxford
Instruments or a suitable collection facility.

Page 9
SECTION 1.1

Installation - Overview

Read through the manual carefully before using the unit.

The commissioning process involves a complete function test of the unit and user
training by a Oxford Instruments Analytical GmbH customer service representative.
If you wish to install the system yourself, please following these instructions.

Before installation

Check the delivery to ensure it is complete.


If any parts are missing or damaged, contact your local Oxford Instruments
Analytical GmbH partner immediately.

Examine the unit for visible damage.


Remove the transport securing of the optical system (only trained personnel should
do this).
Connect the computer to the instrument.
Connect the keyboard, mouse and printer to the computer.
Connect the mains cable to the computer.
Connect the mains cable to the spectrometer (230 V).
Connect the argon cylinder to the instrument with the supplied hose.
(Use only argon suitable for spectrometry, 99.998% - 99.999% Ar).
Traces of other gases (e.g. nitrogen) in the argon may impair precision.
A supply pressure of at least 3.0 bar and not more than 4 bar is required.
Connect the vacuum pump using all the hose connectors provided for that purpose.
Close the valve to the tank, switch on the vacuum pump. First evacuate the hose and
then open the valve to the vessel. Always close the valve to the tank before
switching off the vacuum pump.
First boot up the computer; only then should you switch on the FOUNDRY-MASTER
PRO.

Page 10
Initial check

First switch on the computer. The system starts Microsoft Windows™ XP.
Wait until Windows™ has fully started up.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

Switch on the FOUNDRY-MASTER PRO.

Select "Analyse" mode.


Select the desired matrix.
Click "OK" to confirm.

Select "Recal" (recalibrate instrument).


Select the "Full" option.
The name of the recalibration sample (e.g. RE12) appears in the centre of the
screen. Burn the properly prepared (freshly ground, grease-free) sample. Avoid
carrying out a second burn in the same spot. The RSD (relative standard deviation)
value in the right-hand column will be displayed after the third burn and should be
less than or equal to 10.

If this is the case, carry out further burns. The burn results with the largest
deviations from the average value are deleted (average value displayed in the right-
hand column).
To delete individual burns, select the column (double-click) and press "Delete".
Warning: Deletion is irrevocable and cannot be undone.

Click "Next" at the top left of the screen and burn the next recalibration sample.
Once recalibration has been completed (message "Recalibration successful") the
recalibration factors are checked: Choose "Mode" from the toolbar and select
Recalibration factors from the listbox. All factors should be between 0.5 and 2.0,
ideally close to 1.
The Windows folder Start / Documents contains the file offset.dat: At an ambient
temperature of approx. 20°C the values in the last columns should be between –10
and +10.

Important: If the factors and/or the offset values are outside these ranges,
please contact your Oxford Instruments Analytical Service Partner.

After installation, back up the "Master" folder so that all relevant data can be
restored if necessary. It is the customer's responsibility to back up data on a
regular basis. Oxford Instruments Analytical GmbH cannot be held
responsible if data is lost. Data loss can be caused by hard drive failure,
computer viruses etc.

Installation - Overview
Page 11
SECTION 1.2

Utilities menu
This section gives a brief description of the options on the Utilities menu; for
more detailed information, refer to the appropriate section.

This section explains the Utilities menu at user level 4.


If you see only a limited menu, you are at a lower user level.
Enter the necessary password, see Figure 4. Use of a password prevents
unintentional changes to the software settings.
If you do not know the correct password, consult the responsible person.

Important: User levels 1 to 3 are restricted. Therefore some of the options


described in this section are not available. If you want to change the settings,
consult your superior.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

Utilities menu

Page 12
FOUNDRY-MASTER PRO Start menu

Select
"Utilities"

Figure 1

Utilities menu

Utilities menu; password level 4:

Figure 2

Return to main menu - Close this menu, go back to the main menu.

Enter password - Opens a window to enter a password.

Options - Opens the Options window.


User level 4 only

Passwords - Opens a menu for assigning passwords.


User level 4 only

Utilities menu

Page 13
Delete files - Allows you to delete files.

Create / Change alloy - Opens the Create / Change alloy window

If a reduced Utilities menu appears as shown in Figure 3, check the user level.

Restricted Utilities menu

Figure 3

Enter password

If password protection is active, at program start-up the Password dialog window


appears, see Figure 4. Enter the password and confirm with "OK" to start the
WASLab software.

Further information on password management can be found in the "Passwords"


section later on in this section.

Figure 4 Enter
password

Confirm with
"OK"

Utilities menu

Page 14
Options

The Options window is only available at user level 4.

Figure 5

Number of burns
before the cleaning
reminder appears.
Reduction of argon
flow in burn mode.
Value is entered in
minutes.

Start in analysis mode, When enabled, on start-up the WASLab software


in program automatically opens the analytical program which
has been selected from the dropdown list on the
right. If this option is not selected, the standard start
window opens instead.
Further information about using an analysis -
program can be found in Section 3.2.

Argon saver The argon flow in the spark stand is reduced when
a user-defined period of time has elapsed since the
last analysis. The value is entered in minutes.

Stand clean warning A reminder is displayed when the number of burns


entered here is reached (Figure 6).

Utilities menu

Page 15
Clean stand reminder

You are recommended to clean the stand as soon as this message appears.

You can skip cleaning by clicking "Remind me later", but this is not advisable.

Figure 6

Page 16
Further options

Figure 7

Display analysis horizontally - To display the analytical results in rows, showing


percentage of chemical elements by mass.

Automatic average - Active: The average value of the analytical


(only available if Display results is automatically displayed (in percentage
analysis horizontally has been analysis mode only).
selected) Inactive: Average value display can be turned on
and off by pressing the corresponding button in
the toolbar.

Print in color - may be selected (colour printer must be


connected).

Minimum recalibration burns - Here you can specify the minimum number of
for spark burns during recalibration, see 2.3.

External programs
After every burn start - Not part of the standard package.
When pressing Store start - Activation possible according to customer
requirements.

Utilities menu

Page 17
Error settings for bad samples, allocation of colour display

The colour of the error message window is user-definable. The error message
window appears as soon as a burnt sample breaches the defined
specifications/limits.
The warning window is displayed if the analysis of the burnt sample lies just within
the defined limits.
The colour of the warning message can be modified to suit the environment during
operation, e.g. an intense colour for a bright environment.

The word Error can be displayed in any colour. Click the top left button -
"Foreground" to open the colour properties window. Select the desired colour and
confirm with "OK".

Repeat the process for the "Background" and the warning window.

Colour properties are


modifiable:
Windows™ menu

Figure 8

Select option.
The colour properties
window is opened

Once all settings have been carried out: Confirm


the changes.

Utilities menu

Page 18
Passwords
Passwords can only be assigned at user level 4.
Therefore access to user level 4 must be password-protected.

It is advisable to protect all settings against accidental changes.


Define access rights for the relevant user level by setting RED or GREEN bars and
assigning a password for the user concerned, see Figure 9. Simply click a bar to
change its colour.
If passwords have been assigned, the user must enter his password on starting the
WASLab software, see Figure 4 and Figure 12. The program only starts once the
correct password has been entered.

Oxford Instruments Analytical GmbH cannot be held responsible for any damage
caused by maloperation. It is absolutely essential to restrict access to the WASLab
software.

Important: Passwords must be kept safely.


Without a passw ord you cann ot access the softw are.
If you have forgotten your WASLab password, speak to your superior or
contact your Oxford Instruments Analytical GmbH Service Partner.

Assigning passwords and access rights Enter passwords for the relevant
user level in the fields**

Figure 9

GREEN: Mode available.

RED: Mode not available.

Utilities menu

Page 19
To explain the assignment of password and access rights, in the following example
we change the permission to Create/Change alloy for user level 2.

Set the bars in the Edit password window (Figure 10) for user level 2 by clicking on
them. In this way you enable or disable the assigned options for that user.
Enter password; passwords may consist of letter, numbers and symbols.

In user level 2: Click the bar for the Create /


Change alloy option to change it to red.

Assign a password for user level 2:


We use "TEST2" merely as an example.

Figure 10

Confirm with "OK" to return to the Utilities menu at user level 4, see Figure 11.
To activate the modified settings, restart the WASLab software by clicking in the
Utilities menu on "Return to main menu" and then in the Start menu "End".
Restart the WASLab software with a double-click, see Figure 12.

Utilities menu

Page 20
FOUNDRY-MASTER PRO Utilities menu

Exit the Utilities menu

Click “END”
to return to
Windows™

Figure 11

Restart the WASLab software: Now the software starts with password protection.

FOUNDRY-MASTER PRO Start window with password protection

Enter the password assigned for


user level 2 and select “OK”

Figure 12

FOUNDRY-MASTER PRO Utilities menu

In this example the WASLab software now has a limited Utilities menu.

Page 21 Utilities menu


The Create / Change alloy mode is not displayed and therefore cannot be
changed.

Check the
user level

Figure 13

Delete files
The WASLab software has a central Delete files window. This allows you to delete
all types of files created with WASLab.

Note: The programs listed in the Delete files window, see Figure 14 ff., are
displayed for information only and are not available to you.
Here you can find the programs you have generated while working with the
instrument.

As an example we will delete the file "Report file1". This file is located in the Report
files group.
In the Utilities menu select the "Delete files" option: The Delete files window is
opened.

Utilities menu
Page 22
Delete files menu

Open list

Figure 14

Delete files: All available file groups are displayed in the dropdown list on the left
(see arrow "Open list"). Select a group in order to view the relevant files in the list on
the right.

Display of all available


file groups
Figure 15

From the dropdown list select the file type you wish to delete.
In our example we wish to delete a file from the Report files group.
Selecting a file group causes the relevant files to appear in the right-hand window.

Utilities menu

Page 23
Select file group

Figure 16 List of files in this


group

From the right-hand window select the file you wish to delete. In our example this is
the file (WASDemo).

Only click the "Delete" button once you have selected a file.
Select the file (WASDemo). Confirm removal of the file with "Delete".

Select file

Figure 17

Confirm with
"Delete"

To finally delete the selected file, close the Confirm window by clicking "Yes". The
file is now deleted - or click "No" to go back to the previous window.

Only delete files you no longer need. Deleted files are permanently removed, they
are not sent to the Windows™ Recycle Bin. They are not recoverable.

Utilities menu

Page 24
Figure 18

Confirm with "Yes"

Exit the Delete files window and click Close to return to the Utilities window.

Create / Change alloy

The Select analytical program menu allows you to create a new analytical
program or open an existing program. For more information about setting up an
analytical program, see Sections 2.1 -2.7.

Figure 19

Utilities menu

Page 25
Vacuum gauge
The vacuum gauge is only available in conjunction with the optional vacuum sensor.
This is located in the pump compartment of the unit and is used to monitor the
vacuum tank: The unit is only in a measurable condition if and as long as the
pressure in the vacuum tank is less than 0.1 mbar.

If a vacuum sensor is present, a coloured "Vacuum" button appears in the start


screen of the WASLab software, see Figure 20. Clicking on the vacuum button
opens the vacuum gauge (pointer with scale).

Vacuum
gauge

Figure 20

Vacuum
button

The colour of the vacuum button gives an indication of the current pressure in the
vacuum tank: If the pressure goes below a threshold of 0.1 mbar, the colour of the
button changes from red to green (and vice versa). The scale of the vacuum gauge
is accordingly divided into red and green areas. Stable and correct analysis of
certain elements requires high optical transparency, including in the wavelength
below 180 nm (the so-called vacuum-UV range), which is only guaranteed at an (air)
pressure below 0.1 mbar.

However, the tank pressure rises by several tenths of a mbar with every burn, as
some argon is led into the tank in the process. As a result the pressure may fall
slightly below the 0.1 mbar threshold. This is not a problem, though, because argon
(unlike air) doe not impair optical transparency in any way. Provided the system is
intact, the pressure in the tank falls back below the threshold within a few minutes; if
the vacuum button stays green within this time, the system remains in a measurable
condition.
Therefore the gauge is not so much for short-term monitoring as medium and long-
term monitoring of the tank pressure:
− During the (re-)commissioning of the unit or after reactivating the pump / after
opening the ball valve in the pump compartment, you can see whether the
required vacuum in the tank has been reached. If the tank has been
Utilities menu

Page 26
ventilated, e.g. for maintenance purposes, this can take several hours. On
pumping out the air, if the required pressure of 0.1 mbar is not reached, the
system does not enter the measurable state, i.e. measurements are blocked.
− If the pump was in a measurable state, it will cease to be so as soon as the
tank pressure rises above 0.1 mbar and stays there (for longer than 10
minutes after the last burn).

In such cases, make sure that the vacuum pump is switched on, that the ball valve in
the pump compartment is fully open and that there are no leaks in the vacuum
system, e.g. due to incorrectly connected vacuum components or incorrectly fitted
tank cover. The vacuum pump must be serviced if necessary; follow the
corresponding instructions in the operating manual for the vacuum pump.

If the problem cannot be rectified, the vacuum pump may be faulty or the vacuum
system may have a leak. In this case please contact the Oxford Instruments
Analytical Service Centre.

Utilities menu
Page 27
Excitation parameters

By assigning parameters for excitation it is possible to control several things during


burning, including the times for the individual sections of a measurement, the
amount of energy used in the measurement and the connected argon flow.

Proceed as follows:

1. Open WASLab and press the "Analysis" button

Figure 21

Analysis

2. Next a window appears in which you can choose the base to be measured.
Confirm with OK.

Figure 22

For example:
FE

Utilities menu

Page 28
3. Then you will see the measurement window, with the following buttons along the
top of the screen:

Figure 23

Press the CHANGE


button

4. This opens a window in which you can access all the parameter settings. To get
to the excitation parameters, click the Edit button opposite Parameter file.

Figure 24

Utilities menu

Page 29
5. The following window opens:

Figure 25

It is very important to retain the preset parameters in this window, because these
parameters are matched to one another and your unit has been calibrated with
them.
Changes to these settings may mean that your samples are no longer excited
correctly, thereby invalidating the measurement results.

Utilities menu

Page 30
SECTION 1.3

Spark analysis (quick start)


ANALYSIS IN SPARK MODE (ARGON)

The flowcharts on the following pages should help you to produce reliable analyses.
It is not enough simply to burn the sample; certain conditions must also be met.

If the unit has not been used for a long time, the argon system must be adequately
flushed with argon to avoid contamination with oxygen. The argon used must
contain less than 5 ppm O2 and H2O (argon for spectral analysis 99.998% -
99.999%). The analysis of certain elements such as nitrogen or carbon is greatly
impaired by impurities in the argon. With argon quality of 4.6 (99.996%), analysis
may no longer be possible. Further information about nitrogen analysis can be found
in Appendix B.6 "Notes on nitrogen analysis".
The most critical point in spark analysis is the sample preparation. The
importance of sample preparation is often underestimated, even though the quality
and reliability of the analysis directly depend on it. Upon discharge, an area of the
sample surface is melted and evaporated by the spark. The large amount of energy
in the spark creates a plasma. This plasma emits light which is spectrally analysed
and measured with the aid of CCDs (Charge Coupled Devices, light-sensitive
semiconductor chips). The computer works out the analysis on the basis of this
data. Since the spark discharge only evaporates the surface of the sample and does
not penetrate deep into the material, you will also measure any impurities on the
surface. Dirt, oil, oxides and even fingerprints can disturb and even prevent the
formation of the plasma. This gives a bad burn spot with very low light efficiency.
The result is a faulty analysis, i.e. the analysed composition of the sample does not
conform to its actual composition. Freshly ground, clean sample surfaces are
therefore a fundamental requirement for accurate analysis.

The abrasive paper should be changed as soon as you change the matrix (e.g. from
Cu to Fe) or as soon as satisfactory grinding is no longer possible with the paper.
Generally speaking you can prepare one set of recalibration samples with one
grinding disc, always beginning with the pure sample; this is the sample with the
lowest number in the marked identification (e.g. "RA10" for Al matrix). Do not use
carborundum (SiC) paper as the presence of silicon and carbon will invalidate the
analysis of these elements. We recommend aluminium oxide (corundum) paper or
zirconium paper.

To burn a sample, place it with the prepared, flat side on the spark stand in such a
way that the hole in the spark stand plate located above the electrode is as fully
covered as possible, see Figure 1.

S pa r k an al ys i s ( qui c k st a r t )

Page 31
Figure 1

In addition, the sample must be pressed down with the swivelling, height-adjustable
hold-down device mounted on the left side of the spark stand, see Figure 2. Choose
the pressure point where the hold-down device presses on the sample in such a
way that the sample seals the spark stand hole as tightly as possible; this is usually
the case if the pressure point is located centrally over the electrode. If the sample is
not located correctly, there will be a gap between the sample and the spark stand
plate, allowing air to penetrate from outside into the argon-flushed analytical space
(space between electrode and sample), thereby invalidating the analysis. At the
same time, the sound generated during the burn process (measurement) will reach
the outside more easily; a quiet burn noise is therefore a good indication of an
optimally located sample. Light from the spark often reaches the outside as well; do
not look at the spark without eye protection as you may be dazzled by the
light and your eyes may be damaged by the high-energy UV radiation.

On no account bring an electrically conductive object near the spark stand


hole or remove it from there during the burn. Otherwise there is a danger of
electric shock.

S pa r k an al ys i s ( qui c k st a r t )

Page 32
Figure 2

To avoid accidental starting of the burn process without having a sample in place,
measurement is prevented unless the hold-down device is in the position shown in
Figure 2. This is monitored by a proximity switch in the hold-down mechanism. If a
sample cannot be pressed with the hold-down device because of its size, it is
possible to lift, turn and lower the hold-down device to place it in the position shown
in Figure 3, in which burning is possible without direct contact between the hold-
down device and the sample.

Caution: With the hold-down device in this position, before and during the
entire measurement process (burn) you must take care to ensure that the
spark stand hole is covered by a sample. During burning with an open spark
stand hole, never place any part of your body near the hole in view of the
danger of electric shock.

The spark stand must be cleaned after every 300 burns. Every time the sample is
changed, the electrode must be cleaned with a steel brush (with brass handle). If
you change the matrix (e.g. from iron to aluminium), the electrode and brush should
be replaced, see Appendix A.1. Otherwise the analysis may be invalidated by
contamination on the electrode and brush.
For every matrix there should be a separate electrode and brush.

S pa r k an al ys i s ( qui c k st a r t )

Page 33
Figure 3

Page 34
Start and process of analysis

Argon 3 bar

Connect power supply

Switch on computer


Wait until Windows has fully started

Switch on spectrometer

Double-click the
WASLab icon,
spectrometer software
Figure 4 is started

Choose "Analysis",
select analytical program,
e.g. AL_000 or CU_000

Select menu item “Mode”;


Click the "Argon flush" button; flush
for 1 minute, click button again to stop
flushing.

Check Not
result OK

Recalibration:
OK
Select the
recalibration method.
"TYPE" or "FULL"
(SEE FLOWCHART
P. 39 OF THIS SECTION)

" AN AL Y S E "

Page 35 S pa r k an al ys i s ( qui c k st a r t )
Check whether the system needs to be recalibrated

To determine whether your system needs to be recalibrated, you need a known


(analysed or certified) sample with a similar chemical composition to the sample you
want to analyse.

Select "Next"

Enter the name/number of


the sample, click "OK"

Analyse the sample,


at least 3 - 4 burns

Examine the results

If one burn deviates too


much from the others:
select it by
double-clicking in the
Figure 5 column and remove it
with "Delete".

If 3 values are uniform and


lie within the specification of
the certificate **

If 3 values are uniform and lie


outside the specification of the
certificate **

THE INSTRUMENT MUST


BE
RECALIBRATED
(STANDARDISED)

AN AL Y S E

Page 36 S pa r k an al ys i s ( qui c k st a r t )
** Note:
You can find out how "exactly" a spectrometer measures by performing a
comparison with certified (i.e their exact composition is known) CRM (Certified
Reference Material) samples. A clear distinction is made between two statistical
terms:

1.) The "precision" of measurements. In this case the standard deviation is


determined from a series of successive measurements on the same sample. If
it is related to the average value, we obtain the relative standard deviation.
2.) The "accuracy" of measurements. Here the question is whether the average
value is correct in terms of the stated content.

Generally speaking a spark spectrometer in good working order measures very


precisely, but that does not necessarily mean that the analysis is also "accurate".
Spectrometry is a comparative process, in other words, the composition of an
unknown sample is determined by comparing it with data of known samples. If
errors occur in this comparison ("systematic errors"), the instrument will precisely
apply these deviations to the unknown sample.

The analytical certificate supplied with each certified sample shows the element
concentrations including the scatter (standard deviation) between analyses of
different laboratories.
For certification, the samples are analysed by as many as 15 reputable certified
laboratories in this country and abroad. Each laboratory uses the best facilities at
is disposal for the analysis (e.g. wet chemistry, combustion, XRF, AES-ICP and
spark atomic emission spectrometry). The standard deviation resulting from the
comparison of these analyses (1 sigma) does not correspond to the standard
deviations of the individually used analytical measuring instruments, but rather it
is the standard deviation of the averages from the individual laboratories. The
average of all measurements of the individual laboratories is regarded as the best
approximation to the "true" content of an element in the sample and is used as
the target value for the calibration. If such CRM samples are used for calibration,
despite taking the greatest possible care, there is always a certain deviation
between the target values and actual values calculated from the calibration
curves. In other words, there is a systematic deviation between the spectrometer
analysis and the certified value. However, with modern analysis technology these
deviations are small and allow a simplified method of determining whether a
spectrometer measures accurately:

To estimate the expected deviation in the analytical value obtained with an optical
emission spectrometer based on the standard deviation of an element stated in
the certificate, according to experience the value in the certificate can be
multiplied by a factor of 3 (in statistical terms: approx. 99.7% of values lie within a
3-sigma interval).

For example:

Certificate states: Average 1.082 %, standard deviation 0.008 %.

⇒ The concentration shown in the average value of an optical emission spectrometer should lie in
the range between 1.082% - 0.024% = 1.058% and 1.082% + 0.024 % = 1.106 %.

S pa r k an al ys i s ( qui c k st a r t )

Page 37
TYPE RECALIBRATION

Select "Recal"

Select "Type"

Burn the desired sample (e.g.


BS 81 G-4) at least 3-4 times
(if possible do not burn in the same
burn spot)
Figure 6

OK
If one burn deviates too
much from the others:
select it by
Confirm double-clicking in the
column and remove it
with "Delete".

AN AL Y S E

Page 38 S pa r k an al ys i s ( qui c k st a r t )
FULL RECALIBRATION

Select "Recal"

Select "Full".

Burn the desired samples


at least 3-4 times
(if possible do not spark in the same burn spot)

RSD
should be If one burn deviates too
less than much from the others:
Figure 7 3%. select it by
double-clicking in the
column and remove it
Not OK with "Delete".

Check samples

OK

Continue until all recalibration samples for this matrix


have been burnt.
Click "Next" to continue.

Select
A N A L Y S I S

Page 39 S pa r k an al ys i s ( qui c k st a r t )
Carry out analyses

Select "NEXT"

Enter the sample name/number

Confirm with "OK"

Choose "Analysis"

Burn the sample at least 3-4 times changing


the position

Figure 8

Verify the If one burn deviates too


analysis much from the others:
select it by
double-clicking in the
column and remove it
with "Delete".

If 3 values are consistent

Select "Save"
if you wish to save the
analysis

Go on to the next sample

Page 40 S pa r k an al ys i s ( qui c k st a r t )
SECTION 2.1

Analysis: General information

Note: Depending on the selected user level / password level, certain functions
of the software are not available. Therefore there are different menu
structures for the different password levels.

This part of the manual explains how to use analysis mode at password level
4, from where you have access to all individual main menus, and also
includes the instructions for password levels 1 - 3.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO start window

Choose "Analysis"

Figure 1

Important: A tungsten electrode is always used for analysis in burn mode. Set a
distance of 3.2 mm between the electrode tip and the sample surface.
There is a setting gauge in the accessories case.

Anal ysis: General information


Page 41
Selecting an analytical program
Depending on the configuration of the FOUNDRY-MASTER PRO, the number and
choice of matrices may vary and therefore the number of buttons may differ from the
illustration.
The example menu in Figure 2 shows four matrices. However, the procedure is the
same for all matrices.

Select the required matrix, in this case "FE"

Figure 2

1) Select the required matrix, in this example FE for iron matrix.


2) Select the required analytical program, in this example FE_000.

Select the required program

Figure 3 Confirm with "OK"

To simplify working an analytical mode, the following function keys are pre-
programmed:

Anal ysis: General information

Page 42
F10 F2 F4 Del F5 F6 F7 F8 F9 Alt+R Alt+Q

Number of burns which can be displayed on the


screen simultaneously
(see Section 0.2: Using this manual)

Average
Figure 4 value is
Information bar displayed
in this
column

1 2 3 4 5
For description see below

Information bar
The information bar shows important additional information about the currently
displayed analytical mode.
From left to right:
1 Display mode
Displayed Mode
PA Percentage analysis (% by mass)
CC Channel concentration
IC Corrected intensity ratios
IS Recalibrated intensity ratios
IR Intensity ratios
I Intensities
F Recalibration factors

2 Current analytical program (abbreviation)


CU _000, CU_100, AL_100…..etc.

Anal ysis: General information

Page 43
3 Mode e.g. Sample analysis
or Recalibration

4 Material The material which matches the analysis is shown


here; e.g. if grade search is active

5 Number of burns Shows the total number of burns since the last counter
reset. Resetting can only be carried out by an authorised
Oxford Instruments Analytical GmbH Service Partner.

Function key presets

Start - Activate an analytical process.


New - Enter a new sample name. The results of previous analyses
are deleted from the display.
Average - Display the average value of the current burns.
(not shown in This option is only available if the corresponding option has
Figure 4) been selected from the Options menu, see Figure 6 and
Section 1.2, Figure 7.
Print - Print the current analytical results, indicating:
Name of the alloy, sample name, date, etc.
Delete - Selected, marked data (burn) are deleted.
Save - Option for saving data (analytical results etc.).
Recal - Program for carrying out a recalibration in the active analytical
program.
Mode - Select display mode for analytical results.
Load - Open the Select analytical program window.
Change - Change the parameters of the active analytical program.

RSD - Display the Relative Standard Deviation of at least 3 burns.


Exit - Close the Analysis window, return to Start window.

Page 44
Anal ysis: General information
Brief description of function keys

Figure 5

If the "Display analysis horizontally" option has been chosen (see Section 1.2
Utilities menu, Figure 7), and Automatic average is not selected, this button will
be added to the menu, see Figure 6.

Analysis window in horizontal mode

(Average value display enabled)

The average is displayed above the


analytical results.
Figure 6

Page 45 Anal yse: allgemeine I nformation


Starting an analysis

Figure 7

"Start": Analysis is initiated, the process runs automatically.


Alternatively: Press the green pushbutton on the front of the FOUNDRY-MASTER
PRO.
The analytical process is displayed on the screen, see Figure 8.

-
+

Figure 8

RESULT
Repeat for further burns.

Anal ysis: General information

Page 46
Analyse a new sample

Figure 9

Select "New" to analyse a new sample.


1) If no report file has been set up in the
active program, the dialog window is
Figure 10 displayed (Figure 10); confirm with "OK"
to delete the content of the screen. If you
want to save your analytical results first,
choose "Cancel".

2) If a report file has been assigned to the active program, see Section 2.7, the
dialog window (Figure 11) is shown first.

The number of entries in this dialog depends on the number of questions in the
report file. These entries are copied to the print log. Select "Cancel" to discard the
entries and close the dialog window.

Set up report file

Figure 11

Confirm entries
with "OK"

The user generates this list


when setting up the report file (Section 2.7).

Anal ysis: General information

Page 47
A print log is optional. Not all fields need to be filled in. On printing, any rows not
filled in are disregarded.

Double-click in a text box to open the virtual keyboard.


*
Virtual keyboard
Question Enter required information (optional)

Figure 12

Confirm
with "OK"

Note: The virtual keyboard is displayed anew for each question.

Print results

Figure 13

Print out the current analysis, indicating:

Name of the alloy,


Sample name,
Date etc.

Anal ysis: General information

Page 48
The results are printed on the installed Windows™ default printer. If no printer is
connected, the print job is placed on the queue and can be printed later.

Further information about setting up a print log can be found in Section 2.7.

Delete burns

Figure 14

To delete a no burn that is longer needed, select the relevant column on the screen
with a double-click and then press the "Delete" button. This button only becomes
active when a column is selected.

Save data

Figure 15

This button may be inactive depending on the selected options.


Once a burn has been finished, the analytical result can be saved manually by
selecting this button.
If auto-save is active, the results are automatically save to the defined destination
file, see Figure 28 in this Section.

Anal ysis: General information

Page 49
Confirming saving of analyses

Figure 16
Analyse stored

Recalibration

Figure 17

Choices for recalibrating the FOUNDRY-MASTER PRO.

The WASLab software offers three recalibration methods:


Type recalibration - The current material type (an individual sample) is
recalibrated.
Alloy - An alloy group (e.g. CrNi steel) is recalibrated.
Full - A complete matrix (e.g. FE) is recalibrated.

For further information about recalibration, see Section 2.3.

Presentation of analytical results ("Mode")

Figure 18

Anal ysis: General information


Page 50
Figure 19

Percent analysis - The analytical results are automatically shown as a percent


analysis (default).
Channel - The analytical results are shown in terms of the
concentrations concentration of each individual element channel (analytical
line).
Corrected - The analytical results are shown as the corrected
int ratio intensity ratios.
Recalibrated - The analytical results are shown as the recalibrated
int ratio intensity ratios.
Intensity ratio - The analytical results are shown in terms of the intensity
ratios relative to the reference line.
Intensities - The analytical results are shown as intensities (quantity of
light).
Restandardization - Summary of the last recalibration.
factors

Burn mode

Anal ysis: General information

Page 51
Figure 20

Recall stored - Re-open saved analyses.


analysis
Repeat burns - A definable number of measurements is automatically carried
out on the same sample.
Stand cleaned - Sets the counter for the automatic Stand clean warning to
zero.
Argon flush - After the system has been idle for a long time, the higher
argon flow helps to flush the stand with argon quickly.
Disable alignment - Turns off auto-alignment (only use if necessary).

Search grade - Enter settings to search the grade database,


see also Section 2.4.2.
Compare grade - Specify a target grade from the grade database, see also
Section 2.4.2.
Grade options - Opens the Grade search options window, see Figure 22.
Print/store options - Opens the window where you can enter settings for Printing
and Saving, see Figure 21.

Confirm changes with "OK".

Anal ysis: General information

Page 52
Repeat burns

When the matrix is changed, the first analyses of the new matrix can be affected by
residual condensation (contamination) in the stand from the last burns of the old
matrix. This utility program is used to "burn off" contamination with repeated burn
operations.

Figure 21 Enter the desired number


of repeats by double-
clicking*

Double-click in the text box to open the virtual numeric keypad.


*

Print/store options

Select files

Figure 22

Grade options dialog

Anal ysis: General information

Page 53
Figure 23

No auto - There is no automatic search of the grade database if


search/compare this button is active
Auto search - If active, the software automatically searches the grade
database.

Auto search - Compares the result with the result of the last
on last burn measurement and not with the average.
Large grade display Inactive: The result of the automatic grade search is
displayed underneath the analysis window.
Active: The result of the grade search is displayed
in full-screen mode, see Figure 24.
Auto compare and - The two fields are dependent on one another. The
After burns automatic grade recognition depends on the value
entered in the "After burns" field. The number of burns
should be not less than 2.
Strict grade limits - Tolerances are disregarded in the grade search.
Stop at first grade - Only the first matching grade found is displayed. The
found database is not searched further.
Automatic new sample - If active, a new window for entering a further sample
opens automatically.
Grade compare, show - See left
all limits
Apply grade settings to - If selected and confirmed with "OK", all options of this
sub-programs menu apply to all sub-programs of this matrix.

Grade search: Large display of found quality

Anal ysis: General information

Page 54
Figure 24

Figure 24 shows an example in which the burnt quality does not correspond
to the target grade (red background).
Further information: Grade search, see Section 2.4.2.

Load

Figure 25

Clicking the "Load" button opens the Select analytical program window, see
Figure 26. The window shows all available analytical programs. Select the required
program and confirm.

Anal ysis: General information

Page 55
2)….then select the required
Figure 26 program.

1) First select the


matrix…

1) Select the required matrix


2) Select the required program.

Note: When changing the matrix:


Clean the whole spark stand.
Change the electrode. Set the electrode-workpiece distance correctly:
3.2 mm.

Caution: Always switch off the (excitation) source when working on


the spark stand and/or the electrode

Change a program

Figure 27

"Change" opens the Change alloy window for the active analytical program
(Figure 28).
For further information on changing the alloy see Section 2.4.

Anal ysis: General information

Page 56
Figure 28 "Browse":
The destination folder
for saving analytical
results is defined here.

Note: Sub-programs of the matrices can have different parameter settings. All
changes to the active programs apply only to these islands. Other sub-programs
remain unchanged.

When you click "OK", confirmed settings are saved and will be available again
when a sub-program is restarted.

All changes to the instrument can have an adverse effect on analytical performance.
Changes should only be carried out by professionals with experience of
spectrometers.

Oxford Instruments Analytical GmbH cannot be held responsible for changes to the
instrument or the instrument settings which have been carried out by unauthorised
operators or for any resulting malfunctions.

RSD (Relative Standard Deviation)

Figure 29

Display of the relative standard deviation (RSD) can be activated.


The result appears in a separate column after at least 3 burns have been carried
out.

Anal ysis: General information


Page 57
Closing the program

Figure 30

Click "Exit" to close the analysis window. The program switches to the Start
window, see Figure 31.

Confirm

Figure 31

Click "OK" to continue.

FOUNDRY-MASTER PRO Start window

Figure 32

Page 58
Section 2.2

Analysing
Important: Depending on the selected user level / password level, certain functions
are disabled. Therefore there are different menu structures for the different user
levels.

This part of the manual explains how to use analysis mode at user level 4, from
where one has access to all individual main menus. Level 4 also includes a guide to
the main menu of password levels 1 to 3.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO Start menu

Figure 1

Choose
"Analysis"

Choose "Analysis"; the Select analytical program window opens, see Figure 2.

Anal ysing

Page 59
Select an analytical program

Depending on the configuration of the FOUNDRY-MASTER PRO, the number of


matrices and therefore the number of symbols in the Select analytical program
window may vary. However, the analytical sequence is the same for all matrices.

Select the required matrix by clicking the corresponding symbol. In the following
example we choose Fe (iron).

Select the required matrix.


Figure 2

Select the alloy type (sub-program)

Highlight the
selected
program

Figure 3

Confirm
with "OK"

Anal ysing

Page 60
Analysis window
To simplify working an analytical mode, the following function keys are pre-
programmed:

F10 F2 F4 Del F5 F6 F7 F8 F9 Alt+R Alt+Q

Number of burns which can be displayed


simultaneously

Figure 4 Average is
calculated
automatically

Information bar

Function key presets

Start - Activates an analytical process.


New - Enter a new sample name. The results of previous analyses
are deleted from the display.
Average - Displays the average value of the current burns.
(not shown in This option is only available if the corresponding option has
Figure 4) been selected from the Options menu, see Figure 6 and
Section 1.2, Figure 7.
Print - Print the current analytical results, indicating:
Name of the alloy, sample name, date, etc.
Delete - Selected, marked data (burn) are deleted.
Save - Option for saving data (analytical results etc.).

Page 61
Anal ysing
Recal - Program for carrying out a recalibration in the active analytical
program.
Mode - Select display method for analytical results.
Load - Opens the Select analytical program window.
Change - Change the parameters of the active analytical program.
RSD - Display the Relative Standard Deviation of at least 3 burns.
Exit - Close Analysis window, return to Start window.

"NEW" opens the dialog for


a new analysis.*

Confirm
entries
with "OK"

Figure 5

You have the option of working with report files. If a report file has been assigned to
the analytical program, see Section 2.7, first the dialog box appears (Figure 5).
This dialog box contains precisely those questions which were defined when the
report file was set up. To fill it in, click on the text box (bordered in red in Figure 5).
Not all fields have to be filled in. However, the later printout can only contain the
data entered here.
Finish your entries by clicking "OK".

Opens the virtual keyboard.


*

Anal ysing

Page 62
Enter sample ID here

Figure 6

Place the prepared sample on the spark stand.


Click Start or press the green pushbutton on the front of the housing to start the
analysis. Repeat the process for further burns.

-
+

Figure 7

Result
Anal ysing

Page 63
Analysis window after 4 burns

Displays the current


sample ID

Figure 8

Saving analyses:

"Store": Saves the analytical results in the


Figure 9 predefined folder.

Anal ysing

Page 64
Deleting burns

Select burns that are no longer needed by double-clicking on the corresponding


column, then click "Delete" on the toolbar to remove the column from the analysis.
1) Select burns to be deleted

2) Click "Delete"

Figure 10

Carry out further burns as necessary. Permanently delete any analytical results that
are no longer needed once and for all as described above.

By double-clicking on the head of a column, you can hide the burn in question
temporarily. The display changes to inactive (it is greyed out), see Figure 11. This
signals that this measurement will not be included in the average calculation. This
means you can quickly assess the influence of each individual measurement on the
overall result.

To reactivate a hidden burn, simply click on the column head again.

Anal ysing

Page 65
Deactivating a burn temporarily

Double-click on the column head to exclude the column


temporarily from the average calculation
Figure 11

Select
"RSD"
Activate display of Relative Standard Deviation (RSD)

Figure 12

Select "RSD". The RSD is displayed (Figure 13).


The RSD display is only available after at least three burns have been carried out
and three burns are active on the screen. Any deactivated measurements as in
Figure 11 must be reactivated.

Anal ysing

Page 66
Data in the RSD window:

SD - Absolute Standard Deviation

RSD - Relative Standard Deviation: standard deviation related


to the average

Minimum - Minimum concentration for x burns

Maximum - Maximum concentration for x burns

Average - Average values of all current burns

Anal ysing

Page 67
Analysis window with displayed RSD values

Figure 13

Click on the RSD button again to return to the previous menu.

Repeat the process for further analyses (burns).

Anal ysing

Page 68
Section 2.3

Analysis: Recalibration

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO Start menu

Choose "Analysis"
Figure 1

Recalibration

Page 69
Recalibrate analytical program

The customer-specific configuration of the FOUNDRY-MASTER PRO defines the


number of matrices. The example in Figure 2 shows 9 matrices. However, the
recalibration process is the same for all matrices on all Oxford Instruments
Analytical GmbH spectrometers.

Select the symbol of the desired matrix. The following example shows a
recalibration with (FE) iron.

Select the matrix, in this example "FE"

Figure 2

Select the alloy (sub-program)


Select the desired sub-
program

Selected sub-
program is
highlighted

Figure 3

Confirm
with "OK"

Recalibration
Page 70
Analysis window
Select "Recal"

Figure 4

Recalibration methods
You have the choice of recalibrating

- only one individual sample ("Type" recalibration),

- a particular alloy group ("Alloy", e.g. low alloy steels within the Fe matrix) or

- a complete matrix ("Full").

Here we choose the option of "Full" recalibration of a matrix.


The procedure for the other recalibration methods is the same; just the number of
samples to be measured may be smaller.

Recalibration

Page 71
Figure 5
Select "Full"

Type recal Type Only the selected sample is


sample recalibrated
Use type Activate Type recalibration data
Undo type Deactivate Type recalibration data
This alloy Only this alloy type (the selected subprogram, e.g.
FE_500) is recalibrated.
Full The complete matrix (e.g. FE) is recalibrated, all sub-
programs are updated.
Cancel Ends recalibration and closes the window.

In the recalibration process, the recalibration samples are automatically requested


by the software in turn.
These samples are supplied with the unit and can be found in the accessories case.

During the recalibration process it is advisable to carry out at least 3 burns per
sample. When testing a pure sample (e.g. high-purity iron, sample ID beginning with
"RE12"), the sample should be burnt about 10 times, as the first burn results are
often unusable due to the so-called "memory effect". The pure sample is always the
first sample of the set which is requested during the recalibration of a matrix.

The minimum number of burns during recalibration is defined in the Utilities menu,
see Section 1.2, Figure 7.

To delete individual burns, highlight the relevant column by double-clicking it and


remove it by clicking " Delete" on the toolbar.

Recalibration

Page 72
ID of recalibration
sample

Recalibration sample
is requested by the
software
Figure 6

RE12 (in Figure 6) is the ID of the recalibration sample to be burnt.

Click on "Start" or press the green pushbutton on the front of the FOUNDRY-
MASTER PRO to burn the recalibration sample now. The process runs automatically
and you can follow it on the screen.

Once burning has finished, the result appears in the Burn 1 column.
The numerical values (light intensities) should approximately correspond to the
values listed in the left-hand column after the element channel IDs. Repeat the
process until the required burns have been carried out for all recalibration samples
of the active analytical program and matrix, see Section 1.2 Utilities menu,
Figure 7.

After burning each recalibration sample at least three times and checking the
displayed RSD values, click "Next". The ID of the next recalibration sample appears
on the screen. Repeat these steps until all recalibration samples of the active
analytical program and matrix have been burnt and the message "Recalibration
successful" appears, see Figure 10.

Recalibration

Page 73
Figure 7

If you click "Next" before the recalibration sample has been burnt enough times
(see Options menu, Section 1.2 Utilities menu, Figure 7, Minimum recalibration
burns), you will be asked to carry out the missing burns, see Figure 8. Select "OK"
to proceed with the remaining burns.

Error message concerning minimum number of burns

Figure 8
Click "OK" to continue

If the recalibration of a particular recalibration sample is not necessary, in the


"Warning" window (Figure 9) select "Skip this standard". The software asks for
the next recalibration sample.

If you want to abort the whole recalibration process, select "Next":


The "Warning" window (Figure 9) appears. Select "Abort recalibration".

Warning window
If the recalibration process is aborted, the following window appears:

Recalibration

Page 74
Figure 9

The software recognises that the recalibration process has been aborted.
The user now has three options:

Skip this - This means that the active recalibration sample will not be used
standard to calculate the recalibration. The software uses the data of the
last recalibration of this recalibration sample. It continues with
the next recalibration sample.
Burn this - Select this option to continue burning the active recalibration
standard sample.
Abort - The recalibration process is completely aborted. All data of the
recalibration current process is discarded. The software uses the data of the
last saved recalibration.

Continue recalibration until all samples have been processed.


After all recalibration samples have been either burnt or skipped, the Information
window "Recalibration successful" appears. Confirm this window with "OK" to
return to the Analysis window.

Confirmation Recalibration successful

Figure 10

Once all recalibration samples have been burnt, the results of the recalibration
(restandardisation) are automatically saved.

Recalibration

Page 75
Checking the recalibration data

In the Analysis window, select "Mode"from the top toolbar. This opens the "Burn
mode" window.
In the "Mode" dropdown list, see Figure 11, select "Restandardization factors"
and confirm with "OK".
Information about other options in this list can be found in Section 2.3, Figure
17.

Select "Mode"

Choose
"Restandardization
factors"

Figure 11

Confirm
with "OK"

Recalibration

Page 76
Analysis window

Recalibration data

Figure 12

TIP: Make regular printouts of the recalibration data and keep them in a safe
place together with the equipment documentation.

Recalibration

Page 77
SECTION 2.4.1

Grade mode
Grade mode
This Section explains how to set up individual grade databases.
Which data are already stored in the unit depends on the customer-specific -
equipment configuration.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO main menu

Select
"Utilities"

Figure 1

Grade mode
Page 78
FOUNDRY-MASTER PRO Utilities menu

Select "Create / Change


alloy"

Figure 2

Steps for creating/editing a grade database

Please note:

The number and IDs of the programs displayed in the menu depend on the
equipment configuration.
Select an analytical program (Figure 3).

Figure 3

Analytical program,
Confirm selection
Select e.g. FE_000 Orientation

Grade mode

Page 79
Select
"Grade Sets"
Change alloy

Figure 4

Choose "Grade Sets"


Which matrices and how many are displayed in this menu depends on the
equipment configuration.

Select the grade for your analysis from the "Select grade set" window.

Grade mode

Page 80
This window has two options:

Option 1: Select an existing Grade set from the list and confirm with "OK". The
Select grade set window is opened, see Figure 5.

Select grade set

Figure 5

Confirm with "OK"

Option 2: Assign a new name. (Creating a whole new grade database is possible,
but time-consuming. Therefore we recommend you modify an existing grade
database and save it under a new name.)

Double-click in the field to


define a new Grade*

Figure 6

Double-click in the text box to open the virtual keyboard


*

Grade mode

Page 81
Virtual keyboard
Enter the name of the grade with the virtual keyboard. Click "OK" to confirm.

Enter desired name

Figure 7

Confirm with "OK"

New name is displayed


here

Figure 8

Confirm with "OK"

Reconfirm the name with "OK". The "Grade selection" window is opened, see
Figure 9.

Grade mode

Page 82
Grade selection
When you create a new grade set, the Grade selection menu is empty. All
parameters of the grade are entered in this window.

Choose "Add grade"

Figure 9

Grade selection options:

Add grade The grade name can contain up to 40 characters. The name
can be alphanumeric, numeric or a combination of both.
Examples: alphanumeric Stainless steel
numeric 316 L
Combination Stainless steel X2CRNIMO17-12-2(316L)

Delete grade Delete the selected grade.

Copy grade Copies the selected grade with all parameters into the same
grade set. The new grade must be renamed before you
change the parameters.
Alphanumeric The list is searched for matching grades alphanumerically.
sort
Intelligent sort The list is searched for matching grades based on
concentration.
Excel in The grade list is imported as a CSV (Excel) file.
Excel out The grade list is exported as a CSV (Excel) file.

Grade mode

Page 83
Virtual keyboard
Entering a new grade name

Enter name

Figure 10

Confirm with "OK"

Grade mode

Page 84
The name of the selected active grade is
displayed here

3) Double-click in these fields


to enter the permitted limits
for the corresponding
Figure 11 element.

2) Select element, in this example the


aluminium (Al) content is important.

1) Highlight the grade; edit the limits and tolerances


for this grade.

Figure 12

Double-click in the input field to


open the numeric keypad*

Enter the limits and tolerances and confirm with "OK".


Repeat the process for all alloy elements of this grade for which you want to set
limits and tolerances.
To save time, when creating a grade database we recommend you take an existing
database and modify it.

Grade mode

Page 85
Grade selection

See explanation
below

Scroll down the list of


grades

Figure 13

Grade selection options:

Add grade The grade name can contain up to 40 characters. The name
can be alphanumeric, numeric or a combination of both.
Examples: Alphanumeric Stainless steel
Numeric 316 L
Combination Stainless steel X2CRNIMO17-12-2(316L)

Delete grade Delete the selected grade.


Copy grade Copies the selected grade with all parameters into the same
grade set. The new grade must be renamed before you
change the parameters.
Alphanumeric The list is searched for matching grades alphanumerically.
sort
Intelligent sort The list is searched for matching grades based on
concentration.
Excel in The grade list is imported as a CSV (Excel) file.
Excel out The grade list is exported as a CSV (Excel) file.

Grade mode

Page 86
Choose
"Add grade"

Double-click to open the virtual


keyboard to edit a grade name.
Then confirm your input.

Figure 14

Enter a new grade


name

Figure 15

Confirm with "OK"

Figure 16 Name has been


accepted

Grade mode

Page 87
1) Select element , e.g.
Cr,
and...

2) ... double-click in the


Figure 17 fields opposite to enter
the limits, see Figure
18

Edit limits

The default value in the Edit limits window is 0.0000 (percent).

Double-click to edit
the data*
Figure 18

Double-click in the field to open the virtual keypad (Figure 19).


*

Grade mode

Page 88
Virtual keypad

Figure 19

For grade type 316 L the concentration range for Cr lies between:
a) Low limit 16.5 %
b) High limit 18.5 %
You must enter these figures.

Depending on the sorting task, you can enter an additional value for the tolerance.
Note: It is not obligatory to enter a tolerance.

To explain the function of the tolerance value, here is an example:


The permitted range for Cr is 2% (16.5% – 18.5%).

If a tolerance of 25% is entered, the new tolerance range is extended by


+
- 0.5 % (= 25% of 2%), therefore:

Grade mode

Page 89
new low limit 16.0 % (16.5- 0.5)
new high limit 19.0 % (18.5+0.5)

If the analysis lies within this extended interval, the grade is shown with a yellow
background.

Figure 20

Confirm with "OK"

Repeat this process to edit all elements important for the characterisation of a grade
(Figures 14 - 20).

Grade selection

Figure 21

Confirm with "OK"

Grade mode

Page 90
Once you have completed all entries relating to a grade, close this window. Confirm
the changes with "OK". This takes you back to the "Change alloy" menu (Figure
22).

Werkstoffmodus

Figure 22

After completing your entries, confirm with "OK"


to return to the Utilities menu.

FOUNDRY-MASTER PRO Utilities menu

Back to the
main menu

Figure 23

Grade mode

Page 91
SECTION 2.4.2

Analysis: Grade search

This Section explains the grade search and grade identification functions by
comparing a sample of unknown composition with grade qualities saved in the
database.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO Start menu

Select
"Analysis"
Figure 1

Grade search

Page 92
The following example shows how you can use the Grade search function to
compare an unknown sample, of which you only know that it is an aluminium alloy,
with the qualities saved in the database.

Select AL matrix
Figure 2
Confirm with "OK"

Choose "AL-Global"

Figure 3 Confirm with


"OK"

Grade search

Page 93
Figure 4

For a description of the other function buttons, see Section


2.1 Analysis: General information

Click on "Start" to carry out a burn; alternatively you can press the green
pushbutton on the front of the FOUNDRY-MASTER PRO. The burn process runs
automatically and you can follow the individual stages (argon flushing, preburn,
measuring) on the screen, see Figure 5.

-
+

Figure 5

Result
The result of the burn is displayed in the first empty column on the left in the
Analysis window, see Figure 6.

Grade search

Page 94
Select "Mode"

Figure 6

Working with grade search

The aim is to identify an unknown grade by comparing it with stored data.

1) From the list, select the appropriate grade group (in this example AL_GRADE).
For further information, see Section 2.1, Figure 18 and Figure 19.
Open the Grade options menu (Figure 11) to enter parameters.
Confirm with "OK" and exit the menu.
Click "OK" again to return to the Analysis screen.

Select grade group


within which the
search will be carried
out
Figure 7
Select search
criteria
(Figure 12)

2) Open the Grade options menu (Figure 12) to enter parameters.


3) Confirm your input and exit the menu.

Grade search

Page 95
Select
"Search grade"
Figure 8

4) Click "OK" to return to the Analysis window.

After every analysis, a search is made within the selected grade group for matches
with stored grade data. Depending on the parameters you have set, the result is
displayed in the line underneath the numerical values (Figure 9) or in full-screen
mode (Figure 13).

Grade search

Page 96
Grade information bar, shows details of the
Figure 9 identified grade

ID of the found grade

Priority and number of the [1/ 2]


found grades Priority Number
Display colours
Green - Matching concentrations for all elements of the identified grade
stored in the database.
Yellow - At least 1 element lies outside the stored high/low limits, but
inside the defined tolerance range.
Red - At least 1 element lies outside the stored high/low limits and
outside the defined tolerance range.
Grade information bar
Grade Choose "Grade" in Grade search mode: A window containing
the found matches appears (Figure 11 left).
Choose "Grade" in Grade comparison mode: The window for
defining the target grade appears (Figure 11 right).
Priority Displays the priority of the identified grade. The best match
produces the highest priority.
Number Shows the number of grades considered for the burnt sample.

Grade The found or target grade is shown here. The displayed grade
name name corresponds to what the user has entered in the grade
database. Therefore company-specific abbreviations are also
possible. The meaning of the colours is explained above.

Grade search
Page 97
Figure 10

Click "Grade" to open Double-click the coloured bar to open the


the list of grades in window showing the limits for the elements
question of the found grades
(Figure 11 left) (Figure 11 right)

Note:
These functions are not available when full-screen mode is active (Figure 13), see
Grade search options.

Double-click on the grade ID in the Displays the elements and their


window to activate the grade. limits
“Click "Cancel" to exit the window (for information only).
without making a selection.

Figure 11

Close

Grade search

Page 98
Optimise grade search

For the most accurate


possible identification
Figure 12 enter a value of not less
than 3.

Confirm with "OK"

Grade search options (based on the grades stored in the database).


No auto - Deactivate grade search or comparison with a target
search/compare grade
Auto search - Automatically search for matches with grades stored in
the database
Auto search for last - Automatic search takes into account the last burn carried
burn out.
Large grade display Inactive: Search result in bottom line of screen
(full-screen mode) (Figure 9 and Figure 10)

Active: Search result displayed in full-screen


mode (Figure 13)
Auto compare and - Both functions: Auto compare and After burns (at least
After burns 2) are linked.
Strict grade limits - Tolerance ranges are disregarded.
Stop at first grade - Only the first match found in the database is displayed.
found
Automatic new sample - A new window is opened for each burn.
Grade compare, show - Displays all limits in the grade comparison
all limits
Apply grade settings to - Chosen settings are applied to all sub-programs of the
sub-programs matrix.

Grade search

Page 99
Search result when full-screen mode is active:
Background colour GREEN, YELLOW or RED depending on the degree of match with the
stored element limits.

Figure 13

If a limit is exceeded (background colour RED) you must acknowledge the result by
clicking "Return to the Analysis menu" in order to continue the analyses.

Direct comparison of an unknown sample with grades from the database

The procedure is the same as when searching the database for a matching grade.
After burning the unknown sample, the result is displayed in the Analysis screen.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

In the following example we compare an unknown sample with the stored grade
2014 AICu4SiMg (3.1255).
Example: Comparison of an unknown sample with a grade from the database

Grade search

Page 100
FOUNDRY-MASTER PRO Start menu

Choose "Analysis"

Figure 14

Select the appropriate matrix. Since 2014 AICu4SiMg (3.1255) is an aluminium


alloy, from the Select analytical program window we select the matrix "AL".

Choose "AL"

Figure 15
Confirm
with "OK"

Grade search

Page 101
Choose AL-Global

Figure 16
Confirm with "OK"

The sample to be identified will now be burnt.


To do this, click "Start" or press the green pushbutton on the front of the
FOUNDRY-MASTER PRO.

-
+

Figure 17

Result

Grade search
Page 102
The result is displayed in the first vacant column on the left-hand side of the window
(Figure 18).

In the main menu select "Mode" to


open the Burn mode window

Figure 18

Burn mode

Note: The equipment configuration is customer-specific. Consequently the


illustrations in this manual may differ from what appears on your screen. The
procedure is the same for all grades in all matrices.

1) Grade comparison. Select the desired target grade from the dropdown list; in
this example 2014 AICu4SiMg (3.1255).
Further information about the Burn mode menu can be found in Section 2.1, Figure
18 and Figure 19.

Grade search

Page 103
Select "AL_GRADE"

Figure 19

Select the desired grade from


the list

Grade search options

2) Open the Grade search options window and select the necessary options.

Figure 20

3) Click "OK" to confirm and exit.

Page 104 Grade search


"Percent analysis"
must be set as the
mode

Figure 21 To continue click


"Compare grade"

4) Select "Compare grade", define target grade and return to Analysis window.
Important: 1) If you exit the window by clicking "OK", the comparison will not be
carried out.
2) "Percent analysis" must be set as the mode.

Figure 22

Red background = substantial deviation

Grade search

Page 105
5) Verify the data
After the grade comparison has been completed, all modes can be displayed.

After the "Grade


comparison" has been
carried out, the results
can also be viewed in
other modes.
Figure 23

Analyse further samples in grade mode:

… The virtual keyboard is opened. Enter the ID of a new


sample and confirm.
Select "New"
< … A report file is assigned to the active analytical program.
The input window is displayed. Edit and confirm.
or select "Exit" … Select "OK" and close the Analysis window (Figure 24).
Return to the FOUNDRY-MASTER PRO main menu.

Grade search

Page 106
Click "OK" to
Figure 24
return to the
main menu.

Grade search

Page 107
Section 2.5

Parameter settings

Note: The functions described in this Section allow you to modify important data
and settings. You should only use them if you have appropriate authorisation and
sufficient experience.

Oxford Instruments Analytical GmbH cannot be held liable for malfunctions or


damage caused by changes or interventions in hardware and/or software by
authorised or unauthorised persons.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

Start window

Choose "Analysis"

Figure 1

Parameter settings

Page 108
Setting parameters in Analysis mode

In this Section we will look at how to set parameters in Analysis mode.

For further options refer to Section 2.1 of this manual.

Choose "Change"

Figure 2

Parameter settings

Page 109
Change alloy

Figure 3

Description - Name of the active program.


Matrix element - Select individual element lines for a matrix.
Parameter file - Set software-controlled excitation parameters.
References file - Particular references can be assigned to the element
lines.
Format file - Change screen presentation and printouts.
Curve file - Change the calibration curves of individual lines.
Alloy - Automatically jump from the global program to a
selection file suitable sub-program within a matrix.
Type recal file - Data of individual samples for "Type" recalibration.
Calculation file - Set up and select calculation files.
Report file - Set up and select the report file.
Storage file - Define destination folder for analytical data.
Standards - Select standards for calibration of a matrix.
Grade sets - Enter grades (normally carried out by the operator
according to his own requirements), see also Sections
2.4.1 and 2.4.2.

Parameter settings

Page 110
Excitation parameters
Double-click in the
field to enter a new
value.**

Double-click in the
field to enter a new
value.**

Figure 4

All excitation parameters are preset.


They should only be changed by a Oxford Instruments Analytical GmbH
Service Engineer.
Flush - In this example the flushing duration is 2 seconds before
the start of the burn process. You can change this duration
by double-clicking in the field and entering a new value.

Preburn - Indicates the period of time spent fusing the sample


surface before the actual measurement. The preburn time
should not be more than 30 seconds (risk of overheating
the sample).
AutoSpark - Select AutoSpark to run the analysis automatically.

Double-click in the text box to open the virtual numeric keypad.


**

Page 111
Further excitation parameters
Parameter settings

Figure 5

Alignment and measuring time Optic 1 and 2:


These settings should only be changed by a Oxford Instruments Analytical
Service Engineer.

Note:
Any change to the excitation parameters requires a complete recalibration of the
spectrometer.

Page 112
Parameter settings
Edit references

The data for the element


concerned can be
changed. Double-click in a
white field to open the
virtual numeric keypad.

Figure 6

Define the allocation of reference lines to analytical lines

Reference line = Individual spectral line of the matrix element.


Analytical line = Individual spectral line for an element to be
analysed.

Note: If a different reference is assigned to an analytical line, this can


have serious effects on the analytical result!
Any change should only be carried out by your local Oxford Instruments
Analytical GmbH Partner or following his written instructions.

Parameter settings

Page 113
Add further chemical
elements
Output format

Figure 7 Remove individual


elements from the list.
They can be inserted at
a different point in the
list.
Double-click in a text
field to open the "Edit
value" window*

Double-click in the text box to open the virtual keyboard with the Edit value
* window.

Double-click in the
text box to open the
editing window.

Confirm changes
with "OK"

Parameter settings

Page 114
Edit curve files

Figure 8

Burn standard - Burn standard samples


Edit curves - Opens the window for selecting element lines
whose calibration curves are to be edited.

Delete curves - Delete calibration curves of individual element


lines.
Print all curves - Option for printing all curves or inter-element
corrections.

Note: Any change to the curve data can fundamentally alter the analytical
performance of the spectrometer.

Any change should only be carried out by your local Oxford Instruments
Analytical Partner or by following his written instructions.

Parameter settings

Page 115
Section 2.6

Generating pseudoelements
This Section describes the process for entering individual pseudoelements.

A total of up to ten pseudoelements can be generated, defined by a formula


of your choice.

The following description shows an example of a formula commonly used to


calculate the CE (carbon equivalent) value in steel applications.

Open the WASLab spectrometer software by


double-clicking the WASLab icon

FOUNDRY-MASTER PRO start window

Select "Utilities"

Figure 1

Generating pseudoelements

Page 116
FOUNDRY-MASTER PRO Utilities menu

Click on "Create / Change alloy"

Figure 2

Selecting the required analytical program

Every FOUNDRY-MASTER PRO is configured according to the customer's


requirements. Therefore the number and choice of programs available on your unit
may differ from those illustrated here. For example, the window in Figure 3 shows
11 programs. The procedure for generating a calculation formula is the same for all
programs.

Choose the analytical program from the Select analytical program window in Figure
3, in this example FE_100 – Low Alloy.
Choose the program in
which the calculation
formula is to be defined.

Figure 3

Confirm
with "OK"

Generating pseudoelements

Page 117
Preparing to work with a pseudoelement

Figure 4

Double-click in the
Calculation file field*

Double-click in the text box to open the virtual keyboard; see Figure 5.
*

Enter a name (in this


example "TEST").

Figure 6

Generating pseudoelements

Page 118
The "Edit" button is active by default when using an external keyboard.
*
If you are using the virtual keyboard, the button becomes active when you
click in a white text box.

Figure 7
The name TEST
has been copied for
the active
calculation formula.*

Click "Edit" to
enter a
calculation
formula.

Enter the calculation


formula here.

Double-click here to select the


displayed number of places after
Figure 8 the decimal point.

Pseudoelement erstellen

Page 119
This menu allows the user to enter up to 10 different pseudoelements.

All defined pseudoelements automatically appear in the Analysis screen.

Enter calculation formula for pseudoelements

For example, for CE (carbon equivalent) with the following formula:

CE = C + (Mn+Si)/6 + (Cr+Mo+V)/5 + (Ni+Cu)/15

Important: The notation of the formula must be matched to the software. The
software identifies the individual concentrations of each element on the basis of a
preceding percentage sign. This express must also be placed in brackets.

In the above example, the formula must be entered in the following format:

CE=(%C)+((%MN)+(%SI))/6+((%CR)+(%MO)+(%V))/5+((%NI)+(%CU))/15

Generating pseudoelements

Page 120
Enter and confirm the calculation formula

Use the virtual keypad to enter calculation formula.


Then confirm your entries with "OK"

Figure 9

Confirm entries with


"OK"

Generating pseudoelements

Page 121
Figure 10

Confirm entries with


"OK"

Generating pseudoelements

Return to the
main menu

Figure 11

Generating pseudoelements

Page 122
Choose "Analysis" to
open the program with
the new formula.

Generating pseudoelements

Page 123
Figure 13
Select FE_100

Select the corresponding


matrix to apply the new
calculation formula.

Figure 14
Confirm selection
with "OK"

Generating pseudoelements

Page 124
Analysis screen

Figure 15

The pseudoelement CE
is now automatically
displayed in the Analysis
screen.

The Analysis program (in this example: FE_300) is now ready to start with the new
pseudoelement ("CE").

Generating pseudoelements

Page 125
Automatic pseudoelement calculation

Calculates the pseudoelement automatically from the analytical data,


see Figure 16.

Figure 16

The new pseudoelement "CE"

Note: The displayed CE value is too high in practice. This example is merely
intended to illustrate the entry of complex formulas.

(CE values are mainly calculated for C-steels. Typical values < 0.5).

Generating pseudoelements
Page 126
Section 2.7

Setting up a report file

The WASLab software produces a predefined report file with its own layout.
The layout can be adapted to your company's specific requirements.

The name of the predefined report file is WASDEMO. It is advisable to create a copy
of this file before making any changes to it.

Back up the file “WASDEMO.rep”

1 - Open Windows™ Explorer


2 - Select the MASTER folder
3 - Select the file WASDEMO.rep
4 - Make a Copy of WASDEMO.rep
5 - Rename the copy e.g. "1Test.rep"
6 - Close Windows™ Explorer

Setting up a report file

Open the WASLab spectrometer software by


double-clicking the WASLab icon

Setting up a report file

Page 127
FOUNDRY-MASTER PRO start window

Select
"Utilities"

Figure 1

FOUNDRY-MASTER PRO Utilities menu

Select "Create / Change


alloy"

Figure 2

The Select analytical program window is opened.


Choose the analytical program in which the report is to be printed.

Setting up a report file

Page 128
Select the desired analytical program,
e.g. (FE – Orientation)

Figure 3

Confirm with "OK"

Enter the file name of the report file to be printed.


In the following example we generate a report file called "TEST".

1) Enter file name, e.g. 2) Select "Edit"; continue


"TEST"* to Figure 5**
Figure 4

Confirm with "OK"

Double-click in the text box to open the virtual keyboard.


*
If you are using an external keyboard, the "Edit" button becomes active
** when text is entered. With the virtual keyboard, the "Edit" button is activated
after another text box is selected.

Setting up a report file

Page 129
Report parameters

Figure 5

You are free to choose the print output of the report file in XY coordinates. The X-
position defines the left margin, the Y-position the top margin, in millimetres.

Logo editor
Any logo can be inserted in the report in bmp (bitmap) format. Click on the "Logo"
button and select the required file in Windows™.
The parameters Xpos (140) and Ypos (10) for the logo in Figure 5 define the
position of the logo in an A4 printout, see Figure 6. Experiment with the position
parameters until you are satisfied with the printout. First make a back-up copy of
the original file (WASDEMO.rep), as described at the beginning of this Section.

10 mm from the top edge of the sheet

140 mm
Figure 6 (14 cm) LOGO position

Setting up a report file

Page 130
Editing the report header

The X and Y values define the position of the report header in the printout. Click the
"Font" button to change the typeface.

With this setting the report header is printed at a distance of 2 cm


from the left edge of the page and 1.5 cm from the top of the page.

Enter/edit the report


header by double-clicking in
the text box**

Figure 7

The fonts can be changed using


Windows™ standard fonts.
Confirm entries with "OK"

Double-click in the text box to open the relevant input window, see Figure 8.
**
Double-click in the corresponding white text box to enter or edit data.

Ten editable lines are available to create the report header, see Figure 8.

Setting up a report file

Page 131
Double-click in the corresponding text box
to enter or edit data.*

Figure 8

Once all data have


been edited, click
"OK" to confirm.

Double-click in the text box to open the virtual keyboard.


*

Setting up a report file

Page 132
Edit questions for the output report

Additional information, such as the sample number, heat number, customer name
etc. can be entered in the Questions area.

The user queries the corresponding details before the start of an analysis, provided
a report file has been assigned to the analytical program.

Double-click in the
corresponding text box to enter
or edit data.**

Figure 9

The fonts can be changed using


Windows™ standard fonts.
Confirm entries with "OK"

Double-click in the text box to open the relevant input window, see Figure 10.
**

Setting up a report file

Page 133
Double-click in the corresponding text box to enter
or edit data.*

Figure 10 Normal

Confirm
entries/changes
with "OK"

Double-click in the text box to open the virtual keyboard.


*
20 text lines are available to enter questions, see Figure 10.
Click in one of these white text fields to enter or change data.

Question To enter text, click the relevant line.


Type Double-click repeatedly in the field for further options.
Normal - Entry is printed as entered.
Time - The time of day is added to the entry (Windows™ standard time).
Date - The date is added to the entry (Windows™ standard date).
Grade - The grade data is obtained from the active analytical/sort
program.
Program - The name of the active analytical/sort program is entered
here
POSITION defines the position of the question and answer in the printout.
QX - defines the X position of the question
QY - defines the Y position of the question
AX - defines the X position of the answer
AY - defines the Y position of the answer

Setting up a report file

Page 134
Other layout options

Define the number of


elements in a line of the
output.**

Define the width of the


element display.**

Figure 11

Position of
element name -
(abbreviation)

Number of analytical
results printed per
page.**

In this example the field in which the analytical results are printed is
1.5 cm from the left edge and 12.5 cm from the top of the page.**

Options
Print single - One single analysis per printed page.
analyses
Print average - Averages are printed in the report.
Include grade limits - The entered high and low limits for the element -
concentrations are printed.
Double-click in the text field to open the virtual numeric keypad.
**

Setting up a report file

Page 135
Designing the report footer

The fonts can be changed using


Windows™ standard fonts.
Confirm entries with "OK"

Enter/edit the report


Report footer: Double-
click in the text box.**

Figure 12

With this setting the report footer is printed at a distance of 1.5


cm from the left edge and 22 cm from the top of the page.

Double-click in the text box to open the relevant input window, see Figure 13.
**

Setting up a report file

Page 136
Double-click in the individual fields to enter or
edit data.
Double-click in the text field
10 text lines are available for the footer, see
Figure 13. to edit data.*

Figure 13

Once you have


finished, confirm
your input with
"OK".

Click "OK" to return to the report settings window.

Double-click in the text box to open the virtual keyboard.


*

Setting up a report file

Page 137
Checking the report parameters

Check that you have made all the necessary entries and changes. Select "Preview"
to check the entries and layout, see Figure 15.

Select "Preview";
Figure 14 see Figure 15.

Click "OK" to confirm and return to the main


window.

Setting up a report file

Page 138
Header
LOGO

Questions

Figure 15

Limits
for elements

Analytical data

All the fields in Figure 15 marked with "########" are placeholders for answers.
These are entered by the user (with the unit in analytical mode).
Note: The results shown above (Figure 15) are intended merely as an example.

Click "OK" (see Figure 14) to confirm the changes and return to the "Change
alloy"window, see Figure 16.

Setting up a report file

Page 139
Change alloy

Figure 16

Click "OK" to
return to the
Utilities menu

FOUNDRY-MASTER PRO Utilities menu

Back to the
main menu

Figure 17

Setting up a report file

Page 140
FOUNDRY-MASTER PRO Start menu

Select
"Analysis"
Figure 18

Select the Analysis program for which the new report file was generated.

Figure 19 Select the Analysis program for which the new report
file was generated, in this case: FE – Orientation

Confirm with "OK"

Setting up a report file

Page 141
The Analysis window is opened.

Select "Mode" to active the


report file.
Figure 20

The Burn mode sub-menu is opened.

Select Print/store options.

Figure 21

Select
Print/store options

Setting up a report file

Page 142
Preparing to save and print a report file

In the Print/store options window, select the Report file from the dropdown list.
All settings for saving and printing your results are entered here.
Print single burns

Select the desired report file


and the associated program.

Figure 22
Storage options
(multiple selection)

Click "OK" to confirm


settings and return to the
previous menu
(Figure 24).

Figure 23

Setting up a report file

Page 143
Set up the printer in Windows™; confirm the changes with "OK".

This takes you to the Burn mode window


(Figure 24).

Figure 24

Click "OK" to return


to the previous
menu, see (Figure
25).

After completing your entries you can print out your analytical reports in the desired
format.

Click "New" to begin an analysis.


As a report file has now been assigned to the selected analytical
program, the Dialog window is first opened (Figure 26).

Figure 25

Setting up a report file

Page 144
Enter the corresponding details in each
text field.

Figure 26

The number of questions in the Dialog window depends on your entries in the Edit
questions window, see Figure 9 and Figure 10 in this Section.
Now enter the data and click "OK" to confirm the entries and begin the analysis.
Alternatively: Enter no data and click "Cancel". The analysis which is then carried
out can be printed, but the text fields remain empty.

Select "Start" to begin the analysis and carry out the required number of burns.

Printing the analytical results


So far you have:
- installed and connected your printer,
- set the print options: "Print" menu, see Figure 27,
- carried out the required burns.

Click the "Print" button.

Setting up a report file

Page 145
"Print": print out the
current results

Figure 27

Setting up a report file

Page 146
Appendix A.1

Spare parts and accessories

Spare parts pack small, stock no. 5400000012

Stock number Description Quantity


0000000029 Compression spring 5
0000000180 Hex socket grub screw M3x4 2
0000000287 Electrode cleaning brush round 10
0000000337 Allen key 2.5 1
0000000339 Mains cable cold instruments 2m straight 1
0000000377 Lens biconvex 2
0000000510 Fuse 8A T 20mm 10
0000000547 PS/2 mouse 1
0000000595 O-ring 18x1.5 5
0000000606 Centring ring DN16/KF 2
0000000608 Tensioning ring DN10/16KF Al 2
0000000609 Ball valve Al/Ni 2
0000000714 Plane plate quartz d=5-0,1mm t=1mm 2
0000000869 File brush 2
0000001433 Ferrite for mouse cable 5
0000001468 Tweezers 2
0000001995 Screwdriver 1
0000002039 1 litre pump oil 1
0000002045 Union nut 6mm 5
0000002046 Clamping ring set 5
0000002095 O-ring 3x1 5
0000002103 Extraction tool ceramic insert 2
0000002209 Cleaning brush 2
0000002219 Source DSP 1
3600000034 FMP Button IO 1
0000002407 USB cable 2m for printer 1
0000002481 Tungsten electrode 5
0000003102 Filter mat small 89x89 mm 10
0000003153 PTFE ring 5/3,5x1mm 10
0000003394 OCADs 2
1100000034 Copper tube 6x1 for argon connection 3m 1
1000000005 Knurled screw for electrode holder 2
1100000003 O-ring 6x2.5 5
1100000010 O-ring 30x2.5 5
1100000028 O-ring 25.5x1.5 5
1100000029 O-ring 13x1.5 5
1200000029 Filter mat large 148.5x148.5 mm 10

Ap p e n di x A. 1

Page 147
Spare parts pack small, stock no. 5400000012 (continued)

2000000003 MgF2 lens FMP 1


3000000001 Start button 1
3000000002 Stop button 1
3600000002 OCAD adapter 2
4000000025 FMP spark stand insulation body 1
4000000062 FMP electrode gauge 1
4000000090 FMP lens assembly tool 2
4300000002 Oil filter element EMF 10 1
4300000003 Sealing ring for EMF 1
5010000006 Aluminium oxide with container 1
7000000036 Allen key with T-handle 1.5 1

Spare parts pack large, stock no. 5400000013

Stock number Description Quantity


0000002082 Moxa card 1
0000002221 FMP Ignition Power 1
0000002222 FMP Source 1
3600000012 FMP Power Factor Controller 1
3600000013 FMP Ignition 1
4000000024 FMP spark stand plate 1
4300000001 Oil filter 1
5200000076 FMP Shutter 2
5400000012 FMP spare parts pack small 1

Ap p e n di x A. 1

Page 148
Standard accessories case, stock no. 5010000007

Stock number Description Quantity


0000000029 Compression spring 0.50x5,10x25,00x8.00 2
0000000180 Hex socket grub screw M3x4 vz 1
0000000287 Electrode cleaning brush round 2
0000000337 Allen key 2.5x100 1
0000000339 Mains cable cold instruments 2m straight 2
0000000357 Tool box with organiser 16 inch 1
0000000510 Fuse 8A 20mm T 10
0000000547 MS mouse 2 keys PS-2 1
0000000595 O-ring 18x1.5mm Viton/FKM Shore 80 1
0000000606 Centring ring DN16/KF 1
0000000608 Tensioning ring DN10/16KF Al 1
0000000869 File brush 100x25mm 1
0000001433 Ferrite type TF-50 2
0000001468 Tweezers 1
0000001995 Screwdriver for slotted nuts 1
0000002045 Union nut Swagelok 6 mm 2
0000002046 Clamping ring set (front+rear) 6 mm 3
0000002095 O-ring 3x1 mm Viton Shore 70 3
0000002101 Star grip M6 internal thread 1
0000002103 BG extraction tool ceramic insert 1
0000002114 Union KF25/KF16 1
0000002115 Tensioning ring KF25 1
0000002116 Centring ring with O-ring KF25 1
0000002208 Mouse pad 1
0000002209 Cleaning brush FS 1
0000002407 USB cable A-B approx. 2m 1
0000002481 Tungsten electrode 70 mm 90° 1
0000003102 Filter mat small 89x89 mm 2
0000003153 PTFE ring virginal 5/3.5x1.0 mm 3
1000000005 GN 529 knurled screws M4x10 mm 1
1010000032 Brass dowel MS 6x24 for extraction tool 1
1100000003 O-ring 6x2.5 Viton Shore 80 2
1100000010 O-ring 30x2.5 Viton Shore 70 1
1100000028 O-ring 25.5x1.5 Viton Shore 80 1
1100000029 O-ring 13x1.5 Viton Shore 80 1
1200000029 Filter mat large 148.5x148.5 mm 2
4000000062 Electrode gauge 1
4000000090 Lens insertion tool 1
4000000104 Clamping screw extraction tool ceramic 1
4300000008 Vacuum pump connection kit 1
7000000162 Angled hex socket SW 2.0mm 1
7000000237 Allen key with T-handle 1.5 1
7000000243 Funnel, PE Ø 75 mm, 0.09 l 1

Ap p e n di x A. 1

Page 149
Appendix A.2

Technical data
Dimensions and power consumption FOUNDRY-MASTER PRO
Height approx. 1145 mm
Width approx. 720 mm
Depth approx. 840 mm
Weight (approx, without PC and vacuum pump) 160 kg
Power supply 110 - 230 VAC, 50-60 Hz
Fuse 2 x 16 A slow-blow, 5x20 mm
Power consumption max. (without PC and 1200 VA
vacuum pump)
Power consumption at rest (without PC and 155 VA (source switched on)
vacuum pump) 115 VA (source switched off)

Optical system
Multi-CCD optical system in Paschen-Runge mounting
Focal length 350 mm and 125 mm
Number of grid lines 3000/mm and 1500/mm
Wavelength range 130-780 nm

Excitation
Burn parameters
Frequency 80-500 Hz

Voltage
100-500 V
High Energy Pre-Spark (HEPS)

Options
Wire adapter set
Sample preparation devices
Accessories set
Consumables set

Ap p e n di x A. 2

Page 150
Appendix B.1

Problems/FAQs
Problems/FAQs

Windows™ XP CD-ROM:

The Windows™ XP CD data can be found on the hard disk.


The files are in this folder:

C:\WinXPCD

Printer installation:

Printer driver on Windows™ CD-ROM;


Printer driver on printer CD-ROM;
Download printer driver from the Internet.

All our new units have a USB (Universal Serial Bus) port. If you use a separate PC,
copy the driver to a USB stick and from there to the PC of the FOUNDRY-MASTER
PRO.
Copy the driver to a suitable folder on the C:\ drive (e.g. C:\Windows\TEMP). Then
carry out the installation process.

Appendix B.1

Page 151
Electrodes Clean the electrode using the appropriate brush. We
recommend cleaning after every measurement.
When the matrix is changed, the electrode should also be
changed if necessary.
The electrode should also be cleaned regularly with the
file brush.

Lens The spark stand and vacuum chamber are separated by


a lens. It must be cleaned from time to time, see
Appendix B1-2.

Cleaning the electrode


Cleaning the spark stand
Cleaning the lens

Every 3 months: Check and if necessary change the filter


mats of the fans and outlet filters in the housing

Maintenance to be carried out by the user on a regular basis:

Cleaning the outside of the unit (with a soft cloth)

Do NOT use solvents.


No water must be allowed to get inside the unit.

Appendix B.1

Page 152
Appendix B.2

Cleaning the lens

Open the front and side covers.

Figure 1

Close the ball valve (turn it across the light path).

Figure 2

Appendix B.2

Page 153
Loosen and open the lens cover. (Caution: Be careful not to damage
the optical fibre)

Figure 3

Remove the lens with the tool.

Figure 4

Appendix B.2

Page 154
Loosen the screw coupler with the tool and remove the lens by pressing
on the lens with a clean, soft object.

Figure 5

Clean the lens with a soft cloth and pure acetone. (Observe safety
precautions for working with acetone.)

Figure 6

Appendix B.2

Page 155
Re-insert the lens. Put the flat side on the inner O-ring. The O-rings must
be correctly inserted in the grooves. They must be clean and
undamaged. Then retighten the screw coupler with the tool.

Figure 7

Re-insert the lens holder in the light path. The "collar" must point in the
direction of the vacuum tank.

Figure 8

Appendix B.2

Page 156
Cle
ani
Re-open the ball valve. The spark stand opening should be open when
ng you do this. (If the opening is covered and the lens is not fitted tightly,
there is a danger that water from the washing bottle may be sucked into
the the spark stand.) When opening the ball valve, listen for unusual noises.
When setting down a sample, watch the exhaust hose and the washing
bottles and remove the sample immediately if any water is sucked in.

Activate argon flushing for about 5 minutes.

Appendix B.2

Page 157
Appendix B.3

Cleaning the window


Caution: Always switch off the (excitation) source when working on
the spark stand and/or the electrode.

With the indirect light path there is a small window in front of the optical
fibre which is held in place by a union nut (see photograph).

Union nut with window in


Figure 1 the fitted state

Because impurities occur when samples are burnt, this window must be
cleaned at certain intervals.
After 300 to 400 burns, remove the window for cleaning. The window
should be replaced after every 3000 burns. If measurements are carried
out on materials with a low melting point, such as Zn grades, the window
cleaning and replacement intervals should be shortened.

Page 158 Appendix B.3


To remove the window:
• Undo the clips on the side of the spark stand and remove the
spark stand plate.

Clips
Figure 2

• Then, using the supplied screwdriver for slotted nuts, carefully


unscrew the union nut and window and take them out of the
spark stand.

Figure 3

Screwdriver for slotted nuts

Unscrew the union nut


Figure 4 with the screwdriver

Appendix B.3

Page 159
B.3 • Now, using the supplied tweezers, remove the inner PTFE ring
and the window from the union nut.

Cle
ani
Figure 5
ng
the
win
do PTFE ring Window O-ring Union nut

w
• You can now clean the window with a soft cloth and acetone.

• Warning: When removing the window with the tweezers, be


careful not to scratch the window.

Figure 6 • After cleaning the window, reassemble all the parts, using the
slotted nut screwdriver to screw the union nut back onto the
optical fibre.

Arrangement of O-ring,
window and PTFE ring in
the union nut

Appendix B.3

Page 160
Appendix B.4

Changing the electrode


Caution: Always switch off the (excitation) source when working on
the spark stand and/or the electrode.

When changing the matrix (base) of the samples to be burnt, e.g. from
Fe to Al matrix, it is advisable to clean the spark stand and change the
electrode, so as to avoid contaminations which could falsify the
measurement result.

First of all it is absolutely essential to ensure that the (excitation) source


is turned off. To switch off the source, either turn the source switch (see
Figure 1) anti-clockwise, or turn off the FOUNDRY MASTER PRO with
the main switch on the back of the unit.

1) Turn the source switch to the off position

Figure 1

2) Remove the protective cover

Then pull the blue protective cover under the spark stands forward a
little and remove it from the spark stand in order to reach the three
knurled screws of the electrode holder.

Appendix B.4

Page 161
Loosen the three knurled screws (see Figure 2) and lower the
electrode holder out of the stand. Caution: Do not drop the electrode
holder when removing it.

3) Loosen the knurled


Figure 2 screws

Loosen the three grub screws in the stem of the electrode holder with a
1.5 mm Allen key, see Figure 3. Caution: The spring may force the
electrode to jump out.

Electrode

Small
O-ring
4) Loosen the grub
Figure 3 screw

Large
O-ring

Appendix B.4

Page 162
Take the electrode out of the electrode holder without removing the
spring underneath. Then insert another electrode, which initially sits
loosely on the spring.

Before you retighten the grub screws, it is necessary to press the


electrode against the spring by a precisely defined amount so as to
guarantee that the electrode will be correctly positioned after it is
reinstalled in the spark stand. This is done using the electrode gauge
supplied with the unit (see Figure 4, left or right picture depending on
the version). With the type illustrated in Figure 4 left picture, place the
holder together with the electrode in the gauge so that the electrode
presses against the top inside surface of the gauge. With the type
illustrated in Figure 4 right picture, slip the electrode gauge over the
electrode and press it down as far as it will go.

With the electrode exactly in that position, tighten the grub screws in the
stem. Caution: Do not overtighten the grub screws.

Figure 4

Appendix B.4

Page 163
Then take the electrode holder out of the gauge and check that the
electrode is firmly secured in the holder.

Make sure that both of the O-rings (see Figure 3) are lying correctly in
their respective grooves.

Now reinsert the electrode holder in the stand from below and fasten it
with the three knurled screws. Tighten the knurled screws firmly to
ensure that the electrode is correctly positioned in the stand and that
the O-rings provide an effective seal.

Finally, reattach the blue cover under the stand and turn the source
back on. If the cover is not properly fitted, to ensure your safety the unit
will not allow you to carry out any measurements and the error
message "Electrode cover off" appears at the bottom of the
measurement window.

Appendix B.4

Page 164
Appendix B.5

Washing bottles

On the left side of the housing of the FOUNDRY MASTER PRO are two washing
bottles, see Figure 1. These filter the medium flowing out of the spark stand and
prevent ambient air from being diffused in the stand via the exhaust hose.

Figure 1

To ensure that the washing bottle set functions correctly, fill the second (in the flow
direction) washing bottle, which has only one hose connected to it, with water until
the tube inside is immersed a few centimetres into the water, as illustrated in Figure
2. Avoid filling the bottle too much; the water level must always be well below the
tube in the first (in the flow direction) washing bottle, which has two hoses connected
to it. The first bottle remains unfilled.

This method of filling guarantees that


a) the set performs the above-mentioned functions, and
b) no water can be drawn out of the washing bottle via the exhaust hose into
the spark stand, or even into the optics, if a mistake is made while
cleaning/changing the MgF2 lens (see Appendix B.2, Cleaning the lens).

Appendix B.5

Page 165
Check from time to time that the washing bottles are filled correctly and re-fill or
clean them if the filling level can no longer be seen clearly due to dirt.

The caps of the washing bottles can be unscrewed for this purpose. When screwing
them back on, make sure that the bottles are firmly sealed by the caps. To re-fill it is
advisable to use a funnel which can be placed in the hole of the second bottle.

Fill e.g. via funnel

Figure 2

Appendix B.5
Page 166
Appendix B.6

Notes on nitrogen analysis


In principle the analysis of nitrogen (N), particularly in steel, is no different from that
of other non-metallic elements such as phosphorus and sulphur. The challenge is to
ensure the absence of N2 in the atmosphere. Even a tiny leak in the valves, the
argon supply or the spark stand will immediately cause fluctuating and/or incorrect
analytical values, even though the burn spot – which is generally used as the
assessment criterion for a "good burn" – is satisfactory. The same applies if the N2
content in the argon you are using is too high. To ensure precise and accurate N-
analysis, it is essential to work carefully and to check the condition of the equipment
regularly.

Argon quality
Since nitrogen can also occur as an impurity in the argon being used, the argon
must be very pure. The N2 content should be less than 1 ppm. This is only
guaranteed if the purity is better than 99.9999%, corresponding to a quality1) of 6.0
(sometimes also referred to as 6N0). However, this quality is relatively expensive.
Argon for spectrometry is at least 99.999% pure (Ar 5.0), and may therefore contain
up to 10 ppm of impurities. However, experience shows that there is hardly any
difference between qualities 6.0 and 5.0 in Germany, Europe and the USA (in terms
of N2 content), so that it is usually possible to use the cheaper spectrometer argon.
But in this case high purity is not guaranteed. The reason for the price difference is
the guarantee and the special filling procedure.
Your spectrometer allows you to carry out a simple check on the argon being used,
with the aid of a pure aluminium sample. Provided the Al matrix is part of the
analytical spectrometer programs, the recalibration sample RA10 is used for this
purpose. A pure aluminium sample is supplied with the unit, even if it does not have
an Al matrix. The software is described below. However, it is advisable to have a
small cylinder of Ar 6.0 in case of doubt (see below).

1) The quality tells you the minimum percentage of the gas in abbreviated form, the number before
the decimal point indicating the number of nines in the percentage figure. The number after the
decimal point is equal to the first digit in the percentage figure that is not equal to nine. For example,
Ar 4.6 indicates an Ar purity of at least 99.996%. Ar 5.0 corresponds to at least 99.999% argon.

Page 167
Changing the Ar cylinder
The changing of the Ar cylinder(s) is a critical factor in N analysis. Since the
separation of the valve from the Ar cylinder inevitably results in air getting into the Ar
supply, measures must be taken to expel it from the system.
You can simply flush the Ar system after changing the cylinder. You activate flushing
by selecting "Mode", "Argon flush". This method is very slow, as it can take several
hours before the pipes have been flushed sufficiently to ensure that N-analysis is no
longer impaired.
A better, though more expensive possibility is to use valves which you can buy and
which have a flushing device on the pressure side. This is the safest method.
A third possibility is to follow this procedure:

a) Turn the control valve from 3 bar to 0. This isolates the pressure regulator/
manometer area from the spectrometer.
b) Close the cylinder valve.
c) Unscrew the pressure regulator/manometer from the cylinder.
d) Replace the cylinder.
e) Screw the pressure regulator/manometer onto the new cylinder (do not forget the
seal).
f) Open the cylinder valve and immediately close it again. The primary manometer
should now display the cylinder pressure (usually 200 bar).
g) Loosen the union nut until the pressure between the cylinder and the pressure
regulator/manometer is able to fall to 0.
h) Repeat steps f and g four or five times.
i) Set the control valve to 3 bar.
j) Flush for approximately 2 minutes.
k) Check the Ar quality (N-factor, see below).

Sample preparation and handling


Steel samples are usually ground manually to prepare the surface. However, for N-
analysis we recommend milling or turning, as this ensures that the surface is flat, so
that the hole in the spark stand plate is sealed as well as possible. Despite Jet-
Stream technology, if ground sample surfaces are convex, it is impossible wholly to
avoid nitrogen from the air getting into the spark discharge and being excited there.
This leads to poor precision and false measurement of N-concentration. If turning or
milling is not possible, take care when grinding to ensure that the surface remains
flat. Also, when setting down the sample, make sure there is a sufficient distance
from the edge of the hole in the spark stand to the edge of the sample (the sample
must completely cover the hole), so that no air can penetrate into the spark stand.
About 3 mm is the minimum.

Page 168
Checking the argon quality and the equipment condition
An indicator known as the N-factor is used to check the Ar quality. This is equal to
the intensities of the N-line divided by the adjacent spectral background. An Al
sample containing no N is burnt. In the case of units with an Al base, this is RA10.
With others, a corresponding sample is included.
If you switch to a different matrix after testing, remember that the first analyses may
show an increased Al content. This is remedied by "burning clean", i.e. burning the
low sample several (three to five) times.

To check, proceed as follows:

a) Put down the pure aluminium sample.


b) In the analytical program, click on "Mode".
c) Click "N-factor", see Figure . 1.

Figure 1

The sample is burnt and afterwards the N-factor is displayed, see Figure 2.

The N-factor must be less than 1.5.

If it is greater, either there is a leak or the argon being used contains too much
nitrogen. In the second case you can check by replacing the argon being used with a
cylinder of "clean" argon. "Clean" argon can be either Ar 6.0 or a cylinder which has
previously been tested and found good and is kept for such purposes.
In the first case, a leaking connection in the Ar system is the likeliest cause. Use
leak detection spray or alternatively soapy water to check for leaks in the Ar
connections. You should also clean the spark stand and check the O-rings for wear
or damage. If no leak is found and the N-factor is still too high (>1.5), contact our
service department.

Page 169
Figure 2

Page 170
Appendix B.7

Support

How to get support:

If you require support, contact your Oxford Instruments Analytical Service Partner
or call the Oxford Instruments Analytical Customer Support Centre:

Oxford Instruments Analytical GmbH


Customer Support Centre
Wellesweg 31
47589 Uedem
Germany

Tel.: +49 (0)28 25 93 83 303


Fax: +49 (0)28 25 93 83 100
E-mail: service@oxinst.com

Appendix B.6

Page 171
Appendix C.2

AC / QDR

Page 172
Acceptance Certificate &
Quality Deviation Report
Type of instrument:

Instrument serial number:

Installation date:

Company name (End Customer):

Contact name (End Customer): :

Company address:

Please mark the appropriate line with a cross:


The instrument is now fully operational following the installation. The
entire system is working according to specifications. The user has
received all relevant software and hardware training and is able to
operate the instrument successfully. 

The instrument is now operational following the installation. The entire


system is working according to the specification, but there are issues
to address to achieve total customer satisfaction. The user has

received all relevant software and hardware training and is able to
operate the instrument successfully. Further notes were made and can
be found in the attached Quality Deviation Report (QDR).
The instrument has not been put into operation during the installation.
The Quality Deviation Report contains the details and is attached. 

Site / Office / Location of Instrument:

Representative Name (please print name):

Representative / Installation Engineer Name (please print) & signature:

IMPORTANT: It is essential that this document is completed by the end customer directly after the
installation and faxed or emailed to Oxford Instruments in order to both activate and protect your
manufacturers warranty.

Page 173
Quality Deviation Report ( QDR ) No.:_________
Filled by OI CS

Type of instrument:
Instrument serial number:
Installation date / period (from / to):
Representative:
End-Customer:
Deviation Description (please specify any issue(s) to be resolved)
Hardware:

Software:

Analytical Performance:

Sales / Support issue:

Incomplete delivery:

Page 174

You might also like