Professional Documents
Culture Documents
O X F O R D I N S T R U M E N T S A N A L Y T I C AL G M B H
AN OXFORD INSTRUMENTS COMPANY
Y O U R P AR T N E R F O R Q U AL I T Y
Contents
Section 1
Section 2
ANALYSIS
2.1 - Analysis: General information Page 41
2.2 - Analysing Page 60
2.3 - Recalibration Page 70
2.4.1 - Grade mode Page 79
2.4.2 - Grade search Page 93
2.5 - Parameter settings Page 109
2.6 - Generating pseudoelements Page 117
2.7 - Producing a report file Page 128
Appendix A
Page 2 Contents
Appendix B
Appendix C
Page 3
SECTION 0.1
Introduction
Dear Customer,
Oxford Instruments Analytical GmbH wishes you every success with your
FOUNDRY-MASTER PRO.
Introduction
Page 4
SECTION 0.2
Important:.
Unless otherwise indicated, we assume the user has already started Windows™.
Basic knowledge of how to use software running under Windows™ is assumed.
Important: Pay special attention to any bold or underlined text, or any passages
marked with Important or Note.
All menu and dialog box names are printed in bold and without exclamation mark.
For example: "Create / Change alloy".
Page 5
All figures are numbered by sections. If a particular figure or section is referred to in
the text, it is printed in bold.
For example: Section 2.3, Figure 10.
If an area is bordered in red, the subsequent text or arrow relates to the whole of the
area inside this marking, see Figure 1.
Select to
activate
Figure 1
Tolerances of the
individual
elements.
Double-click to
These texts and
edit, see Figure
arrows…
11
Input field
Figure 2
The number of visible columns depends on the screen resolution set on your
system.
Safety instructions
Never place unduly heavy objects (>10 kg) on the cover of the FOUNDRY-MASTER
PRO.
Never place unduly heavy objects (>10 kg) on the spark stand of the FOUNDRY-
MASTER PRO.
Maintenance or repair must only be carried out by qualified experts. Never open the
housing. There are no user-serviceable parts inside the housing.
Regular inspection of the unit by a qualified expert will increase reliability and
prolong service life.
Safety instructions
Page 8
Internal or external safety devices of the unit must on no account be disabled. This
applies especially to the overpressure valve on the cover of the vacuum tank.
The unit must only be used in a dry environment (protection class IP20).
The unit must be set up on a stable surface. Make sure it is far enough away from
the wall. Align the unit so that it is straight.
When fully shutting down the unit, be sure to close the ball valve between the
vacuum pump and vacuum tank and switch off the pump. Otherwise there is a
danger of the pump overheating as the housing fans are no longer working.
If you touch the unit at the same time as touching electrically conductive parts, such
as machines, instruments, etc., possible potential differences may give rise to
voltages which, though perceptible, are not dangerous.
Temperatures below 0°C or above 50°C may impair the operation of the TFT
screen.
In mains operation the unit must only be operated on systems with PE protective
earth.
Never carry out burn processes on wet or damp surfaces, in view of the danger of
short-circuit.
Be careful not to touch the burn electrode when the excitation source is switched on.
It is essential to turn off the excitation source when working on the sample stand,
e.g. when changing the electrode.
Never look directly at the electric arc. Wear suitable safety goggles to protect your
eyes when working.
Page 9
SECTION 1.1
Installation - Overview
The commissioning process involves a complete function test of the unit and user
training by a Oxford Instruments Analytical GmbH customer service representative.
If you wish to install the system yourself, please following these instructions.
Before installation
Page 10
Initial check
First switch on the computer. The system starts Microsoft Windows™ XP.
Wait until Windows™ has fully started up.
If this is the case, carry out further burns. The burn results with the largest
deviations from the average value are deleted (average value displayed in the right-
hand column).
To delete individual burns, select the column (double-click) and press "Delete".
Warning: Deletion is irrevocable and cannot be undone.
Click "Next" at the top left of the screen and burn the next recalibration sample.
Once recalibration has been completed (message "Recalibration successful") the
recalibration factors are checked: Choose "Mode" from the toolbar and select
Recalibration factors from the listbox. All factors should be between 0.5 and 2.0,
ideally close to 1.
The Windows folder Start / Documents contains the file offset.dat: At an ambient
temperature of approx. 20°C the values in the last columns should be between –10
and +10.
Important: If the factors and/or the offset values are outside these ranges,
please contact your Oxford Instruments Analytical Service Partner.
After installation, back up the "Master" folder so that all relevant data can be
restored if necessary. It is the customer's responsibility to back up data on a
regular basis. Oxford Instruments Analytical GmbH cannot be held
responsible if data is lost. Data loss can be caused by hard drive failure,
computer viruses etc.
Installation - Overview
Page 11
SECTION 1.2
Utilities menu
This section gives a brief description of the options on the Utilities menu; for
more detailed information, refer to the appropriate section.
Utilities menu
Page 12
FOUNDRY-MASTER PRO Start menu
Select
"Utilities"
Figure 1
Utilities menu
Figure 2
Return to main menu - Close this menu, go back to the main menu.
Utilities menu
Page 13
Delete files - Allows you to delete files.
If a reduced Utilities menu appears as shown in Figure 3, check the user level.
Figure 3
Enter password
Figure 4 Enter
password
Confirm with
"OK"
Utilities menu
Page 14
Options
Figure 5
Number of burns
before the cleaning
reminder appears.
Reduction of argon
flow in burn mode.
Value is entered in
minutes.
Argon saver The argon flow in the spark stand is reduced when
a user-defined period of time has elapsed since the
last analysis. The value is entered in minutes.
Utilities menu
Page 15
Clean stand reminder
You are recommended to clean the stand as soon as this message appears.
You can skip cleaning by clicking "Remind me later", but this is not advisable.
Figure 6
Page 16
Further options
Figure 7
Minimum recalibration burns - Here you can specify the minimum number of
for spark burns during recalibration, see 2.3.
External programs
After every burn start - Not part of the standard package.
When pressing Store start - Activation possible according to customer
requirements.
Utilities menu
Page 17
Error settings for bad samples, allocation of colour display
The colour of the error message window is user-definable. The error message
window appears as soon as a burnt sample breaches the defined
specifications/limits.
The warning window is displayed if the analysis of the burnt sample lies just within
the defined limits.
The colour of the warning message can be modified to suit the environment during
operation, e.g. an intense colour for a bright environment.
The word Error can be displayed in any colour. Click the top left button -
"Foreground" to open the colour properties window. Select the desired colour and
confirm with "OK".
Repeat the process for the "Background" and the warning window.
Figure 8
Select option.
The colour properties
window is opened
Utilities menu
Page 18
Passwords
Passwords can only be assigned at user level 4.
Therefore access to user level 4 must be password-protected.
Oxford Instruments Analytical GmbH cannot be held responsible for any damage
caused by maloperation. It is absolutely essential to restrict access to the WASLab
software.
Assigning passwords and access rights Enter passwords for the relevant
user level in the fields**
Figure 9
Utilities menu
Page 19
To explain the assignment of password and access rights, in the following example
we change the permission to Create/Change alloy for user level 2.
Set the bars in the Edit password window (Figure 10) for user level 2 by clicking on
them. In this way you enable or disable the assigned options for that user.
Enter password; passwords may consist of letter, numbers and symbols.
Figure 10
Confirm with "OK" to return to the Utilities menu at user level 4, see Figure 11.
To activate the modified settings, restart the WASLab software by clicking in the
Utilities menu on "Return to main menu" and then in the Start menu "End".
Restart the WASLab software with a double-click, see Figure 12.
Utilities menu
Page 20
FOUNDRY-MASTER PRO Utilities menu
Click “END”
to return to
Windows™
Figure 11
Restart the WASLab software: Now the software starts with password protection.
Figure 12
In this example the WASLab software now has a limited Utilities menu.
Check the
user level
Figure 13
Delete files
The WASLab software has a central Delete files window. This allows you to delete
all types of files created with WASLab.
Note: The programs listed in the Delete files window, see Figure 14 ff., are
displayed for information only and are not available to you.
Here you can find the programs you have generated while working with the
instrument.
As an example we will delete the file "Report file1". This file is located in the Report
files group.
In the Utilities menu select the "Delete files" option: The Delete files window is
opened.
Utilities menu
Page 22
Delete files menu
Open list
Figure 14
Delete files: All available file groups are displayed in the dropdown list on the left
(see arrow "Open list"). Select a group in order to view the relevant files in the list on
the right.
From the dropdown list select the file type you wish to delete.
In our example we wish to delete a file from the Report files group.
Selecting a file group causes the relevant files to appear in the right-hand window.
Utilities menu
Page 23
Select file group
From the right-hand window select the file you wish to delete. In our example this is
the file (WASDemo).
Only click the "Delete" button once you have selected a file.
Select the file (WASDemo). Confirm removal of the file with "Delete".
Select file
Figure 17
Confirm with
"Delete"
To finally delete the selected file, close the Confirm window by clicking "Yes". The
file is now deleted - or click "No" to go back to the previous window.
Only delete files you no longer need. Deleted files are permanently removed, they
are not sent to the Windows™ Recycle Bin. They are not recoverable.
Utilities menu
Page 24
Figure 18
Exit the Delete files window and click Close to return to the Utilities window.
The Select analytical program menu allows you to create a new analytical
program or open an existing program. For more information about setting up an
analytical program, see Sections 2.1 -2.7.
Figure 19
Utilities menu
Page 25
Vacuum gauge
The vacuum gauge is only available in conjunction with the optional vacuum sensor.
This is located in the pump compartment of the unit and is used to monitor the
vacuum tank: The unit is only in a measurable condition if and as long as the
pressure in the vacuum tank is less than 0.1 mbar.
Vacuum
gauge
Figure 20
Vacuum
button
The colour of the vacuum button gives an indication of the current pressure in the
vacuum tank: If the pressure goes below a threshold of 0.1 mbar, the colour of the
button changes from red to green (and vice versa). The scale of the vacuum gauge
is accordingly divided into red and green areas. Stable and correct analysis of
certain elements requires high optical transparency, including in the wavelength
below 180 nm (the so-called vacuum-UV range), which is only guaranteed at an (air)
pressure below 0.1 mbar.
However, the tank pressure rises by several tenths of a mbar with every burn, as
some argon is led into the tank in the process. As a result the pressure may fall
slightly below the 0.1 mbar threshold. This is not a problem, though, because argon
(unlike air) doe not impair optical transparency in any way. Provided the system is
intact, the pressure in the tank falls back below the threshold within a few minutes; if
the vacuum button stays green within this time, the system remains in a measurable
condition.
Therefore the gauge is not so much for short-term monitoring as medium and long-
term monitoring of the tank pressure:
− During the (re-)commissioning of the unit or after reactivating the pump / after
opening the ball valve in the pump compartment, you can see whether the
required vacuum in the tank has been reached. If the tank has been
Utilities menu
Page 26
ventilated, e.g. for maintenance purposes, this can take several hours. On
pumping out the air, if the required pressure of 0.1 mbar is not reached, the
system does not enter the measurable state, i.e. measurements are blocked.
− If the pump was in a measurable state, it will cease to be so as soon as the
tank pressure rises above 0.1 mbar and stays there (for longer than 10
minutes after the last burn).
In such cases, make sure that the vacuum pump is switched on, that the ball valve in
the pump compartment is fully open and that there are no leaks in the vacuum
system, e.g. due to incorrectly connected vacuum components or incorrectly fitted
tank cover. The vacuum pump must be serviced if necessary; follow the
corresponding instructions in the operating manual for the vacuum pump.
If the problem cannot be rectified, the vacuum pump may be faulty or the vacuum
system may have a leak. In this case please contact the Oxford Instruments
Analytical Service Centre.
Utilities menu
Page 27
Excitation parameters
Proceed as follows:
Figure 21
Analysis
2. Next a window appears in which you can choose the base to be measured.
Confirm with OK.
Figure 22
For example:
FE
Utilities menu
Page 28
3. Then you will see the measurement window, with the following buttons along the
top of the screen:
Figure 23
4. This opens a window in which you can access all the parameter settings. To get
to the excitation parameters, click the Edit button opposite Parameter file.
Figure 24
Utilities menu
Page 29
5. The following window opens:
Figure 25
It is very important to retain the preset parameters in this window, because these
parameters are matched to one another and your unit has been calibrated with
them.
Changes to these settings may mean that your samples are no longer excited
correctly, thereby invalidating the measurement results.
Utilities menu
Page 30
SECTION 1.3
The flowcharts on the following pages should help you to produce reliable analyses.
It is not enough simply to burn the sample; certain conditions must also be met.
If the unit has not been used for a long time, the argon system must be adequately
flushed with argon to avoid contamination with oxygen. The argon used must
contain less than 5 ppm O2 and H2O (argon for spectral analysis 99.998% -
99.999%). The analysis of certain elements such as nitrogen or carbon is greatly
impaired by impurities in the argon. With argon quality of 4.6 (99.996%), analysis
may no longer be possible. Further information about nitrogen analysis can be found
in Appendix B.6 "Notes on nitrogen analysis".
The most critical point in spark analysis is the sample preparation. The
importance of sample preparation is often underestimated, even though the quality
and reliability of the analysis directly depend on it. Upon discharge, an area of the
sample surface is melted and evaporated by the spark. The large amount of energy
in the spark creates a plasma. This plasma emits light which is spectrally analysed
and measured with the aid of CCDs (Charge Coupled Devices, light-sensitive
semiconductor chips). The computer works out the analysis on the basis of this
data. Since the spark discharge only evaporates the surface of the sample and does
not penetrate deep into the material, you will also measure any impurities on the
surface. Dirt, oil, oxides and even fingerprints can disturb and even prevent the
formation of the plasma. This gives a bad burn spot with very low light efficiency.
The result is a faulty analysis, i.e. the analysed composition of the sample does not
conform to its actual composition. Freshly ground, clean sample surfaces are
therefore a fundamental requirement for accurate analysis.
The abrasive paper should be changed as soon as you change the matrix (e.g. from
Cu to Fe) or as soon as satisfactory grinding is no longer possible with the paper.
Generally speaking you can prepare one set of recalibration samples with one
grinding disc, always beginning with the pure sample; this is the sample with the
lowest number in the marked identification (e.g. "RA10" for Al matrix). Do not use
carborundum (SiC) paper as the presence of silicon and carbon will invalidate the
analysis of these elements. We recommend aluminium oxide (corundum) paper or
zirconium paper.
To burn a sample, place it with the prepared, flat side on the spark stand in such a
way that the hole in the spark stand plate located above the electrode is as fully
covered as possible, see Figure 1.
S pa r k an al ys i s ( qui c k st a r t )
Page 31
Figure 1
In addition, the sample must be pressed down with the swivelling, height-adjustable
hold-down device mounted on the left side of the spark stand, see Figure 2. Choose
the pressure point where the hold-down device presses on the sample in such a
way that the sample seals the spark stand hole as tightly as possible; this is usually
the case if the pressure point is located centrally over the electrode. If the sample is
not located correctly, there will be a gap between the sample and the spark stand
plate, allowing air to penetrate from outside into the argon-flushed analytical space
(space between electrode and sample), thereby invalidating the analysis. At the
same time, the sound generated during the burn process (measurement) will reach
the outside more easily; a quiet burn noise is therefore a good indication of an
optimally located sample. Light from the spark often reaches the outside as well; do
not look at the spark without eye protection as you may be dazzled by the
light and your eyes may be damaged by the high-energy UV radiation.
S pa r k an al ys i s ( qui c k st a r t )
Page 32
Figure 2
To avoid accidental starting of the burn process without having a sample in place,
measurement is prevented unless the hold-down device is in the position shown in
Figure 2. This is monitored by a proximity switch in the hold-down mechanism. If a
sample cannot be pressed with the hold-down device because of its size, it is
possible to lift, turn and lower the hold-down device to place it in the position shown
in Figure 3, in which burning is possible without direct contact between the hold-
down device and the sample.
Caution: With the hold-down device in this position, before and during the
entire measurement process (burn) you must take care to ensure that the
spark stand hole is covered by a sample. During burning with an open spark
stand hole, never place any part of your body near the hole in view of the
danger of electric shock.
The spark stand must be cleaned after every 300 burns. Every time the sample is
changed, the electrode must be cleaned with a steel brush (with brass handle). If
you change the matrix (e.g. from iron to aluminium), the electrode and brush should
be replaced, see Appendix A.1. Otherwise the analysis may be invalidated by
contamination on the electrode and brush.
For every matrix there should be a separate electrode and brush.
S pa r k an al ys i s ( qui c k st a r t )
Page 33
Figure 3
Page 34
Start and process of analysis
Argon 3 bar
Switch on computer
™
Wait until Windows has fully started
Switch on spectrometer
Double-click the
WASLab icon,
spectrometer software
Figure 4 is started
Choose "Analysis",
select analytical program,
e.g. AL_000 or CU_000
Check Not
result OK
Recalibration:
OK
Select the
recalibration method.
"TYPE" or "FULL"
(SEE FLOWCHART
P. 39 OF THIS SECTION)
" AN AL Y S E "
Page 35 S pa r k an al ys i s ( qui c k st a r t )
Check whether the system needs to be recalibrated
Select "Next"
AN AL Y S E
Page 36 S pa r k an al ys i s ( qui c k st a r t )
** Note:
You can find out how "exactly" a spectrometer measures by performing a
comparison with certified (i.e their exact composition is known) CRM (Certified
Reference Material) samples. A clear distinction is made between two statistical
terms:
The analytical certificate supplied with each certified sample shows the element
concentrations including the scatter (standard deviation) between analyses of
different laboratories.
For certification, the samples are analysed by as many as 15 reputable certified
laboratories in this country and abroad. Each laboratory uses the best facilities at
is disposal for the analysis (e.g. wet chemistry, combustion, XRF, AES-ICP and
spark atomic emission spectrometry). The standard deviation resulting from the
comparison of these analyses (1 sigma) does not correspond to the standard
deviations of the individually used analytical measuring instruments, but rather it
is the standard deviation of the averages from the individual laboratories. The
average of all measurements of the individual laboratories is regarded as the best
approximation to the "true" content of an element in the sample and is used as
the target value for the calibration. If such CRM samples are used for calibration,
despite taking the greatest possible care, there is always a certain deviation
between the target values and actual values calculated from the calibration
curves. In other words, there is a systematic deviation between the spectrometer
analysis and the certified value. However, with modern analysis technology these
deviations are small and allow a simplified method of determining whether a
spectrometer measures accurately:
To estimate the expected deviation in the analytical value obtained with an optical
emission spectrometer based on the standard deviation of an element stated in
the certificate, according to experience the value in the certificate can be
multiplied by a factor of 3 (in statistical terms: approx. 99.7% of values lie within a
3-sigma interval).
For example:
⇒ The concentration shown in the average value of an optical emission spectrometer should lie in
the range between 1.082% - 0.024% = 1.058% and 1.082% + 0.024 % = 1.106 %.
S pa r k an al ys i s ( qui c k st a r t )
Page 37
TYPE RECALIBRATION
Select "Recal"
Select "Type"
OK
If one burn deviates too
much from the others:
select it by
Confirm double-clicking in the
column and remove it
with "Delete".
AN AL Y S E
Page 38 S pa r k an al ys i s ( qui c k st a r t )
FULL RECALIBRATION
Select "Recal"
Select "Full".
RSD
should be If one burn deviates too
less than much from the others:
Figure 7 3%. select it by
double-clicking in the
column and remove it
Not OK with "Delete".
Check samples
OK
Select
A N A L Y S I S
Page 39 S pa r k an al ys i s ( qui c k st a r t )
Carry out analyses
Select "NEXT"
Choose "Analysis"
Figure 8
Select "Save"
if you wish to save the
analysis
Page 40 S pa r k an al ys i s ( qui c k st a r t )
SECTION 2.1
Note: Depending on the selected user level / password level, certain functions
of the software are not available. Therefore there are different menu
structures for the different password levels.
This part of the manual explains how to use analysis mode at password level
4, from where you have access to all individual main menus, and also
includes the instructions for password levels 1 - 3.
Choose "Analysis"
Figure 1
Important: A tungsten electrode is always used for analysis in burn mode. Set a
distance of 3.2 mm between the electrode tip and the sample surface.
There is a setting gauge in the accessories case.
Figure 2
To simplify working an analytical mode, the following function keys are pre-
programmed:
Page 42
F10 F2 F4 Del F5 F6 F7 F8 F9 Alt+R Alt+Q
Average
Figure 4 value is
Information bar displayed
in this
column
1 2 3 4 5
For description see below
Information bar
The information bar shows important additional information about the currently
displayed analytical mode.
From left to right:
1 Display mode
Displayed Mode
PA Percentage analysis (% by mass)
CC Channel concentration
IC Corrected intensity ratios
IS Recalibrated intensity ratios
IR Intensity ratios
I Intensities
F Recalibration factors
Page 43
3 Mode e.g. Sample analysis
or Recalibration
5 Number of burns Shows the total number of burns since the last counter
reset. Resetting can only be carried out by an authorised
Oxford Instruments Analytical GmbH Service Partner.
Page 44
Anal ysis: General information
Brief description of function keys
Figure 5
If the "Display analysis horizontally" option has been chosen (see Section 1.2
Utilities menu, Figure 7), and Automatic average is not selected, this button will
be added to the menu, see Figure 6.
Figure 7
-
+
Figure 8
RESULT
Repeat for further burns.
Page 46
Analyse a new sample
Figure 9
2) If a report file has been assigned to the active program, see Section 2.7, the
dialog window (Figure 11) is shown first.
The number of entries in this dialog depends on the number of questions in the
report file. These entries are copied to the print log. Select "Cancel" to discard the
entries and close the dialog window.
Figure 11
Confirm entries
with "OK"
Page 47
A print log is optional. Not all fields need to be filled in. On printing, any rows not
filled in are disregarded.
Figure 12
Confirm
with "OK"
Print results
Figure 13
Page 48
The results are printed on the installed Windows™ default printer. If no printer is
connected, the print job is placed on the queue and can be printed later.
Further information about setting up a print log can be found in Section 2.7.
Delete burns
Figure 14
To delete a no burn that is longer needed, select the relevant column on the screen
with a double-click and then press the "Delete" button. This button only becomes
active when a column is selected.
Save data
Figure 15
Page 49
Confirming saving of analyses
Figure 16
Analyse stored
Recalibration
Figure 17
Figure 18
Burn mode
Page 51
Figure 20
Page 52
Repeat burns
When the matrix is changed, the first analyses of the new matrix can be affected by
residual condensation (contamination) in the stand from the last burns of the old
matrix. This utility program is used to "burn off" contamination with repeated burn
operations.
Print/store options
Select files
Figure 22
Page 53
Figure 23
Auto search - Compares the result with the result of the last
on last burn measurement and not with the average.
Large grade display Inactive: The result of the automatic grade search is
displayed underneath the analysis window.
Active: The result of the grade search is displayed
in full-screen mode, see Figure 24.
Auto compare and - The two fields are dependent on one another. The
After burns automatic grade recognition depends on the value
entered in the "After burns" field. The number of burns
should be not less than 2.
Strict grade limits - Tolerances are disregarded in the grade search.
Stop at first grade - Only the first matching grade found is displayed. The
found database is not searched further.
Automatic new sample - If active, a new window for entering a further sample
opens automatically.
Grade compare, show - See left
all limits
Apply grade settings to - If selected and confirmed with "OK", all options of this
sub-programs menu apply to all sub-programs of this matrix.
Page 54
Figure 24
Figure 24 shows an example in which the burnt quality does not correspond
to the target grade (red background).
Further information: Grade search, see Section 2.4.2.
Load
Figure 25
Clicking the "Load" button opens the Select analytical program window, see
Figure 26. The window shows all available analytical programs. Select the required
program and confirm.
Page 55
2)….then select the required
Figure 26 program.
Change a program
Figure 27
"Change" opens the Change alloy window for the active analytical program
(Figure 28).
For further information on changing the alloy see Section 2.4.
Page 56
Figure 28 "Browse":
The destination folder
for saving analytical
results is defined here.
Note: Sub-programs of the matrices can have different parameter settings. All
changes to the active programs apply only to these islands. Other sub-programs
remain unchanged.
When you click "OK", confirmed settings are saved and will be available again
when a sub-program is restarted.
All changes to the instrument can have an adverse effect on analytical performance.
Changes should only be carried out by professionals with experience of
spectrometers.
Oxford Instruments Analytical GmbH cannot be held responsible for changes to the
instrument or the instrument settings which have been carried out by unauthorised
operators or for any resulting malfunctions.
Figure 29
Figure 30
Click "Exit" to close the analysis window. The program switches to the Start
window, see Figure 31.
Confirm
Figure 31
Figure 32
Page 58
Section 2.2
Analysing
Important: Depending on the selected user level / password level, certain functions
are disabled. Therefore there are different menu structures for the different user
levels.
This part of the manual explains how to use analysis mode at user level 4, from
where one has access to all individual main menus. Level 4 also includes a guide to
the main menu of password levels 1 to 3.
Figure 1
Choose
"Analysis"
Choose "Analysis"; the Select analytical program window opens, see Figure 2.
Anal ysing
Page 59
Select an analytical program
Select the required matrix by clicking the corresponding symbol. In the following
example we choose Fe (iron).
Highlight the
selected
program
Figure 3
Confirm
with "OK"
Anal ysing
Page 60
Analysis window
To simplify working an analytical mode, the following function keys are pre-
programmed:
Figure 4 Average is
calculated
automatically
Information bar
Page 61
Anal ysing
Recal - Program for carrying out a recalibration in the active analytical
program.
Mode - Select display method for analytical results.
Load - Opens the Select analytical program window.
Change - Change the parameters of the active analytical program.
RSD - Display the Relative Standard Deviation of at least 3 burns.
Exit - Close Analysis window, return to Start window.
Confirm
entries
with "OK"
Figure 5
You have the option of working with report files. If a report file has been assigned to
the analytical program, see Section 2.7, first the dialog box appears (Figure 5).
This dialog box contains precisely those questions which were defined when the
report file was set up. To fill it in, click on the text box (bordered in red in Figure 5).
Not all fields have to be filled in. However, the later printout can only contain the
data entered here.
Finish your entries by clicking "OK".
Anal ysing
Page 62
Enter sample ID here
Figure 6
-
+
Figure 7
Result
Anal ysing
Page 63
Analysis window after 4 burns
Figure 8
Saving analyses:
Anal ysing
Page 64
Deleting burns
2) Click "Delete"
Figure 10
Carry out further burns as necessary. Permanently delete any analytical results that
are no longer needed once and for all as described above.
By double-clicking on the head of a column, you can hide the burn in question
temporarily. The display changes to inactive (it is greyed out), see Figure 11. This
signals that this measurement will not be included in the average calculation. This
means you can quickly assess the influence of each individual measurement on the
overall result.
Anal ysing
Page 65
Deactivating a burn temporarily
Select
"RSD"
Activate display of Relative Standard Deviation (RSD)
Figure 12
Anal ysing
Page 66
Data in the RSD window:
Anal ysing
Page 67
Analysis window with displayed RSD values
Figure 13
Anal ysing
Page 68
Section 2.3
Analysis: Recalibration
Choose "Analysis"
Figure 1
Recalibration
Page 69
Recalibrate analytical program
Select the symbol of the desired matrix. The following example shows a
recalibration with (FE) iron.
Figure 2
Selected sub-
program is
highlighted
Figure 3
Confirm
with "OK"
Recalibration
Page 70
Analysis window
Select "Recal"
Figure 4
Recalibration methods
You have the choice of recalibrating
- a particular alloy group ("Alloy", e.g. low alloy steels within the Fe matrix) or
Recalibration
Page 71
Figure 5
Select "Full"
During the recalibration process it is advisable to carry out at least 3 burns per
sample. When testing a pure sample (e.g. high-purity iron, sample ID beginning with
"RE12"), the sample should be burnt about 10 times, as the first burn results are
often unusable due to the so-called "memory effect". The pure sample is always the
first sample of the set which is requested during the recalibration of a matrix.
The minimum number of burns during recalibration is defined in the Utilities menu,
see Section 1.2, Figure 7.
Recalibration
Page 72
ID of recalibration
sample
Recalibration sample
is requested by the
software
Figure 6
Click on "Start" or press the green pushbutton on the front of the FOUNDRY-
MASTER PRO to burn the recalibration sample now. The process runs automatically
and you can follow it on the screen.
Once burning has finished, the result appears in the Burn 1 column.
The numerical values (light intensities) should approximately correspond to the
values listed in the left-hand column after the element channel IDs. Repeat the
process until the required burns have been carried out for all recalibration samples
of the active analytical program and matrix, see Section 1.2 Utilities menu,
Figure 7.
After burning each recalibration sample at least three times and checking the
displayed RSD values, click "Next". The ID of the next recalibration sample appears
on the screen. Repeat these steps until all recalibration samples of the active
analytical program and matrix have been burnt and the message "Recalibration
successful" appears, see Figure 10.
Recalibration
Page 73
Figure 7
If you click "Next" before the recalibration sample has been burnt enough times
(see Options menu, Section 1.2 Utilities menu, Figure 7, Minimum recalibration
burns), you will be asked to carry out the missing burns, see Figure 8. Select "OK"
to proceed with the remaining burns.
Figure 8
Click "OK" to continue
Warning window
If the recalibration process is aborted, the following window appears:
Recalibration
Page 74
Figure 9
The software recognises that the recalibration process has been aborted.
The user now has three options:
Skip this - This means that the active recalibration sample will not be used
standard to calculate the recalibration. The software uses the data of the
last recalibration of this recalibration sample. It continues with
the next recalibration sample.
Burn this - Select this option to continue burning the active recalibration
standard sample.
Abort - The recalibration process is completely aborted. All data of the
recalibration current process is discarded. The software uses the data of the
last saved recalibration.
Figure 10
Once all recalibration samples have been burnt, the results of the recalibration
(restandardisation) are automatically saved.
Recalibration
Page 75
Checking the recalibration data
In the Analysis window, select "Mode"from the top toolbar. This opens the "Burn
mode" window.
In the "Mode" dropdown list, see Figure 11, select "Restandardization factors"
and confirm with "OK".
Information about other options in this list can be found in Section 2.3, Figure
17.
Select "Mode"
Choose
"Restandardization
factors"
Figure 11
Confirm
with "OK"
Recalibration
Page 76
Analysis window
Recalibration data
Figure 12
TIP: Make regular printouts of the recalibration data and keep them in a safe
place together with the equipment documentation.
Recalibration
Page 77
SECTION 2.4.1
Grade mode
Grade mode
This Section explains how to set up individual grade databases.
Which data are already stored in the unit depends on the customer-specific -
equipment configuration.
Select
"Utilities"
Figure 1
Grade mode
Page 78
FOUNDRY-MASTER PRO Utilities menu
Figure 2
Please note:
The number and IDs of the programs displayed in the menu depend on the
equipment configuration.
Select an analytical program (Figure 3).
Figure 3
Analytical program,
Confirm selection
Select e.g. FE_000 Orientation
Grade mode
Page 79
Select
"Grade Sets"
Change alloy
Figure 4
Select the grade for your analysis from the "Select grade set" window.
Grade mode
Page 80
This window has two options:
Option 1: Select an existing Grade set from the list and confirm with "OK". The
Select grade set window is opened, see Figure 5.
Figure 5
Option 2: Assign a new name. (Creating a whole new grade database is possible,
but time-consuming. Therefore we recommend you modify an existing grade
database and save it under a new name.)
Figure 6
Grade mode
Page 81
Virtual keyboard
Enter the name of the grade with the virtual keyboard. Click "OK" to confirm.
Figure 7
Figure 8
Reconfirm the name with "OK". The "Grade selection" window is opened, see
Figure 9.
Grade mode
Page 82
Grade selection
When you create a new grade set, the Grade selection menu is empty. All
parameters of the grade are entered in this window.
Figure 9
Add grade The grade name can contain up to 40 characters. The name
can be alphanumeric, numeric or a combination of both.
Examples: alphanumeric Stainless steel
numeric 316 L
Combination Stainless steel X2CRNIMO17-12-2(316L)
Copy grade Copies the selected grade with all parameters into the same
grade set. The new grade must be renamed before you
change the parameters.
Alphanumeric The list is searched for matching grades alphanumerically.
sort
Intelligent sort The list is searched for matching grades based on
concentration.
Excel in The grade list is imported as a CSV (Excel) file.
Excel out The grade list is exported as a CSV (Excel) file.
Grade mode
Page 83
Virtual keyboard
Entering a new grade name
Enter name
Figure 10
Grade mode
Page 84
The name of the selected active grade is
displayed here
Figure 12
Grade mode
Page 85
Grade selection
See explanation
below
Figure 13
Add grade The grade name can contain up to 40 characters. The name
can be alphanumeric, numeric or a combination of both.
Examples: Alphanumeric Stainless steel
Numeric 316 L
Combination Stainless steel X2CRNIMO17-12-2(316L)
Grade mode
Page 86
Choose
"Add grade"
Figure 14
Figure 15
Grade mode
Page 87
1) Select element , e.g.
Cr,
and...
Edit limits
Double-click to edit
the data*
Figure 18
Grade mode
Page 88
Virtual keypad
Figure 19
For grade type 316 L the concentration range for Cr lies between:
a) Low limit 16.5 %
b) High limit 18.5 %
You must enter these figures.
Depending on the sorting task, you can enter an additional value for the tolerance.
Note: It is not obligatory to enter a tolerance.
Grade mode
Page 89
new low limit 16.0 % (16.5- 0.5)
new high limit 19.0 % (18.5+0.5)
If the analysis lies within this extended interval, the grade is shown with a yellow
background.
Figure 20
Repeat this process to edit all elements important for the characterisation of a grade
(Figures 14 - 20).
Grade selection
Figure 21
Grade mode
Page 90
Once you have completed all entries relating to a grade, close this window. Confirm
the changes with "OK". This takes you back to the "Change alloy" menu (Figure
22).
Werkstoffmodus
Figure 22
Back to the
main menu
Figure 23
Grade mode
Page 91
SECTION 2.4.2
This Section explains the grade search and grade identification functions by
comparing a sample of unknown composition with grade qualities saved in the
database.
Select
"Analysis"
Figure 1
Grade search
Page 92
The following example shows how you can use the Grade search function to
compare an unknown sample, of which you only know that it is an aluminium alloy,
with the qualities saved in the database.
Select AL matrix
Figure 2
Confirm with "OK"
Choose "AL-Global"
Grade search
Page 93
Figure 4
Click on "Start" to carry out a burn; alternatively you can press the green
pushbutton on the front of the FOUNDRY-MASTER PRO. The burn process runs
automatically and you can follow the individual stages (argon flushing, preburn,
measuring) on the screen, see Figure 5.
-
+
Figure 5
Result
The result of the burn is displayed in the first empty column on the left in the
Analysis window, see Figure 6.
Grade search
Page 94
Select "Mode"
Figure 6
1) From the list, select the appropriate grade group (in this example AL_GRADE).
For further information, see Section 2.1, Figure 18 and Figure 19.
Open the Grade options menu (Figure 11) to enter parameters.
Confirm with "OK" and exit the menu.
Click "OK" again to return to the Analysis screen.
Grade search
Page 95
Select
"Search grade"
Figure 8
After every analysis, a search is made within the selected grade group for matches
with stored grade data. Depending on the parameters you have set, the result is
displayed in the line underneath the numerical values (Figure 9) or in full-screen
mode (Figure 13).
Grade search
Page 96
Grade information bar, shows details of the
Figure 9 identified grade
Grade The found or target grade is shown here. The displayed grade
name name corresponds to what the user has entered in the grade
database. Therefore company-specific abbreviations are also
possible. The meaning of the colours is explained above.
Grade search
Page 97
Figure 10
Note:
These functions are not available when full-screen mode is active (Figure 13), see
Grade search options.
Figure 11
Close
Grade search
Page 98
Optimise grade search
Grade search
Page 99
Search result when full-screen mode is active:
Background colour GREEN, YELLOW or RED depending on the degree of match with the
stored element limits.
Figure 13
If a limit is exceeded (background colour RED) you must acknowledge the result by
clicking "Return to the Analysis menu" in order to continue the analyses.
The procedure is the same as when searching the database for a matching grade.
After burning the unknown sample, the result is displayed in the Analysis screen.
In the following example we compare an unknown sample with the stored grade
2014 AICu4SiMg (3.1255).
Example: Comparison of an unknown sample with a grade from the database
Grade search
Page 100
FOUNDRY-MASTER PRO Start menu
Choose "Analysis"
Figure 14
Choose "AL"
Figure 15
Confirm
with "OK"
Grade search
Page 101
Choose AL-Global
Figure 16
Confirm with "OK"
-
+
Figure 17
Result
Grade search
Page 102
The result is displayed in the first vacant column on the left-hand side of the window
(Figure 18).
Figure 18
Burn mode
1) Grade comparison. Select the desired target grade from the dropdown list; in
this example 2014 AICu4SiMg (3.1255).
Further information about the Burn mode menu can be found in Section 2.1, Figure
18 and Figure 19.
Grade search
Page 103
Select "AL_GRADE"
Figure 19
2) Open the Grade search options window and select the necessary options.
Figure 20
4) Select "Compare grade", define target grade and return to Analysis window.
Important: 1) If you exit the window by clicking "OK", the comparison will not be
carried out.
2) "Percent analysis" must be set as the mode.
Figure 22
Grade search
Page 105
5) Verify the data
After the grade comparison has been completed, all modes can be displayed.
Grade search
Page 106
Click "OK" to
Figure 24
return to the
main menu.
Grade search
Page 107
Section 2.5
Parameter settings
Note: The functions described in this Section allow you to modify important data
and settings. You should only use them if you have appropriate authorisation and
sufficient experience.
Start window
Choose "Analysis"
Figure 1
Parameter settings
Page 108
Setting parameters in Analysis mode
Choose "Change"
Figure 2
Parameter settings
Page 109
Change alloy
Figure 3
Parameter settings
Page 110
Excitation parameters
Double-click in the
field to enter a new
value.**
Double-click in the
field to enter a new
value.**
Figure 4
Page 111
Further excitation parameters
Parameter settings
Figure 5
Note:
Any change to the excitation parameters requires a complete recalibration of the
spectrometer.
Page 112
Parameter settings
Edit references
Figure 6
Parameter settings
Page 113
Add further chemical
elements
Output format
Double-click in the text box to open the virtual keyboard with the Edit value
* window.
Double-click in the
text box to open the
editing window.
Confirm changes
with "OK"
Parameter settings
Page 114
Edit curve files
Figure 8
Note: Any change to the curve data can fundamentally alter the analytical
performance of the spectrometer.
Any change should only be carried out by your local Oxford Instruments
Analytical Partner or by following his written instructions.
Parameter settings
Page 115
Section 2.6
Generating pseudoelements
This Section describes the process for entering individual pseudoelements.
Select "Utilities"
Figure 1
Generating pseudoelements
Page 116
FOUNDRY-MASTER PRO Utilities menu
Figure 2
Choose the analytical program from the Select analytical program window in Figure
3, in this example FE_100 – Low Alloy.
Choose the program in
which the calculation
formula is to be defined.
Figure 3
Confirm
with "OK"
Generating pseudoelements
Page 117
Preparing to work with a pseudoelement
Figure 4
Double-click in the
Calculation file field*
Double-click in the text box to open the virtual keyboard; see Figure 5.
*
Figure 6
Generating pseudoelements
Page 118
The "Edit" button is active by default when using an external keyboard.
*
If you are using the virtual keyboard, the button becomes active when you
click in a white text box.
Figure 7
The name TEST
has been copied for
the active
calculation formula.*
Click "Edit" to
enter a
calculation
formula.
Pseudoelement erstellen
Page 119
This menu allows the user to enter up to 10 different pseudoelements.
Important: The notation of the formula must be matched to the software. The
software identifies the individual concentrations of each element on the basis of a
preceding percentage sign. This express must also be placed in brackets.
In the above example, the formula must be entered in the following format:
CE=(%C)+((%MN)+(%SI))/6+((%CR)+(%MO)+(%V))/5+((%NI)+(%CU))/15
Generating pseudoelements
Page 120
Enter and confirm the calculation formula
Figure 9
Generating pseudoelements
Page 121
Figure 10
Generating pseudoelements
Return to the
main menu
Figure 11
Generating pseudoelements
Page 122
Choose "Analysis" to
open the program with
the new formula.
Generating pseudoelements
Page 123
Figure 13
Select FE_100
Figure 14
Confirm selection
with "OK"
Generating pseudoelements
Page 124
Analysis screen
Figure 15
The pseudoelement CE
is now automatically
displayed in the Analysis
screen.
The Analysis program (in this example: FE_300) is now ready to start with the new
pseudoelement ("CE").
Generating pseudoelements
Page 125
Automatic pseudoelement calculation
Figure 16
Note: The displayed CE value is too high in practice. This example is merely
intended to illustrate the entry of complex formulas.
(CE values are mainly calculated for C-steels. Typical values < 0.5).
Generating pseudoelements
Page 126
Section 2.7
The WASLab software produces a predefined report file with its own layout.
The layout can be adapted to your company's specific requirements.
The name of the predefined report file is WASDEMO. It is advisable to create a copy
of this file before making any changes to it.
Page 127
FOUNDRY-MASTER PRO start window
Select
"Utilities"
Figure 1
Figure 2
Page 128
Select the desired analytical program,
e.g. (FE – Orientation)
Figure 3
Page 129
Report parameters
Figure 5
You are free to choose the print output of the report file in XY coordinates. The X-
position defines the left margin, the Y-position the top margin, in millimetres.
Logo editor
Any logo can be inserted in the report in bmp (bitmap) format. Click on the "Logo"
button and select the required file in Windows™.
The parameters Xpos (140) and Ypos (10) for the logo in Figure 5 define the
position of the logo in an A4 printout, see Figure 6. Experiment with the position
parameters until you are satisfied with the printout. First make a back-up copy of
the original file (WASDEMO.rep), as described at the beginning of this Section.
140 mm
Figure 6 (14 cm) LOGO position
Page 130
Editing the report header
The X and Y values define the position of the report header in the printout. Click the
"Font" button to change the typeface.
Figure 7
Double-click in the text box to open the relevant input window, see Figure 8.
**
Double-click in the corresponding white text box to enter or edit data.
Ten editable lines are available to create the report header, see Figure 8.
Page 131
Double-click in the corresponding text box
to enter or edit data.*
Figure 8
Page 132
Edit questions for the output report
Additional information, such as the sample number, heat number, customer name
etc. can be entered in the Questions area.
The user queries the corresponding details before the start of an analysis, provided
a report file has been assigned to the analytical program.
Double-click in the
corresponding text box to enter
or edit data.**
Figure 9
Double-click in the text box to open the relevant input window, see Figure 10.
**
Page 133
Double-click in the corresponding text box to enter
or edit data.*
Figure 10 Normal
Confirm
entries/changes
with "OK"
Page 134
Other layout options
Figure 11
Position of
element name -
(abbreviation)
Number of analytical
results printed per
page.**
In this example the field in which the analytical results are printed is
1.5 cm from the left edge and 12.5 cm from the top of the page.**
Options
Print single - One single analysis per printed page.
analyses
Print average - Averages are printed in the report.
Include grade limits - The entered high and low limits for the element -
concentrations are printed.
Double-click in the text field to open the virtual numeric keypad.
**
Page 135
Designing the report footer
Figure 12
Double-click in the text box to open the relevant input window, see Figure 13.
**
Page 136
Double-click in the individual fields to enter or
edit data.
Double-click in the text field
10 text lines are available for the footer, see
Figure 13. to edit data.*
Figure 13
Page 137
Checking the report parameters
Check that you have made all the necessary entries and changes. Select "Preview"
to check the entries and layout, see Figure 15.
Select "Preview";
Figure 14 see Figure 15.
Page 138
Header
LOGO
Questions
Figure 15
Limits
for elements
Analytical data
All the fields in Figure 15 marked with "########" are placeholders for answers.
These are entered by the user (with the unit in analytical mode).
Note: The results shown above (Figure 15) are intended merely as an example.
Click "OK" (see Figure 14) to confirm the changes and return to the "Change
alloy"window, see Figure 16.
Page 139
Change alloy
Figure 16
Click "OK" to
return to the
Utilities menu
Back to the
main menu
Figure 17
Page 140
FOUNDRY-MASTER PRO Start menu
Select
"Analysis"
Figure 18
Select the Analysis program for which the new report file was generated.
Figure 19 Select the Analysis program for which the new report
file was generated, in this case: FE – Orientation
Page 141
The Analysis window is opened.
Figure 21
Select
Print/store options
Page 142
Preparing to save and print a report file
In the Print/store options window, select the Report file from the dropdown list.
All settings for saving and printing your results are entered here.
Print single burns
Figure 22
Storage options
(multiple selection)
Figure 23
Page 143
Set up the printer in Windows™; confirm the changes with "OK".
Figure 24
After completing your entries you can print out your analytical reports in the desired
format.
Figure 25
Page 144
Enter the corresponding details in each
text field.
Figure 26
The number of questions in the Dialog window depends on your entries in the Edit
questions window, see Figure 9 and Figure 10 in this Section.
Now enter the data and click "OK" to confirm the entries and begin the analysis.
Alternatively: Enter no data and click "Cancel". The analysis which is then carried
out can be printed, but the text fields remain empty.
Select "Start" to begin the analysis and carry out the required number of burns.
Page 145
"Print": print out the
current results
Figure 27
Page 146
Appendix A.1
Ap p e n di x A. 1
Page 147
Spare parts pack small, stock no. 5400000012 (continued)
Ap p e n di x A. 1
Page 148
Standard accessories case, stock no. 5010000007
Ap p e n di x A. 1
Page 149
Appendix A.2
Technical data
Dimensions and power consumption FOUNDRY-MASTER PRO
Height approx. 1145 mm
Width approx. 720 mm
Depth approx. 840 mm
Weight (approx, without PC and vacuum pump) 160 kg
Power supply 110 - 230 VAC, 50-60 Hz
Fuse 2 x 16 A slow-blow, 5x20 mm
Power consumption max. (without PC and 1200 VA
vacuum pump)
Power consumption at rest (without PC and 155 VA (source switched on)
vacuum pump) 115 VA (source switched off)
Optical system
Multi-CCD optical system in Paschen-Runge mounting
Focal length 350 mm and 125 mm
Number of grid lines 3000/mm and 1500/mm
Wavelength range 130-780 nm
Excitation
Burn parameters
Frequency 80-500 Hz
Voltage
100-500 V
High Energy Pre-Spark (HEPS)
Options
Wire adapter set
Sample preparation devices
Accessories set
Consumables set
Ap p e n di x A. 2
Page 150
Appendix B.1
Problems/FAQs
Problems/FAQs
Windows™ XP CD-ROM:
C:\WinXPCD
Printer installation:
All our new units have a USB (Universal Serial Bus) port. If you use a separate PC,
copy the driver to a USB stick and from there to the PC of the FOUNDRY-MASTER
PRO.
Copy the driver to a suitable folder on the C:\ drive (e.g. C:\Windows\TEMP). Then
carry out the installation process.
Appendix B.1
Page 151
Electrodes Clean the electrode using the appropriate brush. We
recommend cleaning after every measurement.
When the matrix is changed, the electrode should also be
changed if necessary.
The electrode should also be cleaned regularly with the
file brush.
Appendix B.1
Page 152
Appendix B.2
Figure 1
Figure 2
Appendix B.2
Page 153
Loosen and open the lens cover. (Caution: Be careful not to damage
the optical fibre)
Figure 3
Figure 4
Appendix B.2
Page 154
Loosen the screw coupler with the tool and remove the lens by pressing
on the lens with a clean, soft object.
Figure 5
Clean the lens with a soft cloth and pure acetone. (Observe safety
precautions for working with acetone.)
Figure 6
Appendix B.2
Page 155
Re-insert the lens. Put the flat side on the inner O-ring. The O-rings must
be correctly inserted in the grooves. They must be clean and
undamaged. Then retighten the screw coupler with the tool.
Figure 7
Re-insert the lens holder in the light path. The "collar" must point in the
direction of the vacuum tank.
Figure 8
Appendix B.2
Page 156
Cle
ani
Re-open the ball valve. The spark stand opening should be open when
ng you do this. (If the opening is covered and the lens is not fitted tightly,
there is a danger that water from the washing bottle may be sucked into
the the spark stand.) When opening the ball valve, listen for unusual noises.
When setting down a sample, watch the exhaust hose and the washing
bottles and remove the sample immediately if any water is sucked in.
Appendix B.2
Page 157
Appendix B.3
With the indirect light path there is a small window in front of the optical
fibre which is held in place by a union nut (see photograph).
Because impurities occur when samples are burnt, this window must be
cleaned at certain intervals.
After 300 to 400 burns, remove the window for cleaning. The window
should be replaced after every 3000 burns. If measurements are carried
out on materials with a low melting point, such as Zn grades, the window
cleaning and replacement intervals should be shortened.
Clips
Figure 2
Figure 3
Appendix B.3
Page 159
B.3 • Now, using the supplied tweezers, remove the inner PTFE ring
and the window from the union nut.
Cle
ani
Figure 5
ng
the
win
do PTFE ring Window O-ring Union nut
w
• You can now clean the window with a soft cloth and acetone.
Figure 6 • After cleaning the window, reassemble all the parts, using the
slotted nut screwdriver to screw the union nut back onto the
optical fibre.
Arrangement of O-ring,
window and PTFE ring in
the union nut
Appendix B.3
Page 160
Appendix B.4
When changing the matrix (base) of the samples to be burnt, e.g. from
Fe to Al matrix, it is advisable to clean the spark stand and change the
electrode, so as to avoid contaminations which could falsify the
measurement result.
Figure 1
Then pull the blue protective cover under the spark stands forward a
little and remove it from the spark stand in order to reach the three
knurled screws of the electrode holder.
Appendix B.4
Page 161
Loosen the three knurled screws (see Figure 2) and lower the
electrode holder out of the stand. Caution: Do not drop the electrode
holder when removing it.
Loosen the three grub screws in the stem of the electrode holder with a
1.5 mm Allen key, see Figure 3. Caution: The spring may force the
electrode to jump out.
Electrode
Small
O-ring
4) Loosen the grub
Figure 3 screw
Large
O-ring
Appendix B.4
Page 162
Take the electrode out of the electrode holder without removing the
spring underneath. Then insert another electrode, which initially sits
loosely on the spring.
With the electrode exactly in that position, tighten the grub screws in the
stem. Caution: Do not overtighten the grub screws.
Figure 4
Appendix B.4
Page 163
Then take the electrode holder out of the gauge and check that the
electrode is firmly secured in the holder.
Make sure that both of the O-rings (see Figure 3) are lying correctly in
their respective grooves.
Now reinsert the electrode holder in the stand from below and fasten it
with the three knurled screws. Tighten the knurled screws firmly to
ensure that the electrode is correctly positioned in the stand and that
the O-rings provide an effective seal.
Finally, reattach the blue cover under the stand and turn the source
back on. If the cover is not properly fitted, to ensure your safety the unit
will not allow you to carry out any measurements and the error
message "Electrode cover off" appears at the bottom of the
measurement window.
Appendix B.4
Page 164
Appendix B.5
Washing bottles
On the left side of the housing of the FOUNDRY MASTER PRO are two washing
bottles, see Figure 1. These filter the medium flowing out of the spark stand and
prevent ambient air from being diffused in the stand via the exhaust hose.
Figure 1
To ensure that the washing bottle set functions correctly, fill the second (in the flow
direction) washing bottle, which has only one hose connected to it, with water until
the tube inside is immersed a few centimetres into the water, as illustrated in Figure
2. Avoid filling the bottle too much; the water level must always be well below the
tube in the first (in the flow direction) washing bottle, which has two hoses connected
to it. The first bottle remains unfilled.
Appendix B.5
Page 165
Check from time to time that the washing bottles are filled correctly and re-fill or
clean them if the filling level can no longer be seen clearly due to dirt.
The caps of the washing bottles can be unscrewed for this purpose. When screwing
them back on, make sure that the bottles are firmly sealed by the caps. To re-fill it is
advisable to use a funnel which can be placed in the hole of the second bottle.
Figure 2
Appendix B.5
Page 166
Appendix B.6
Argon quality
Since nitrogen can also occur as an impurity in the argon being used, the argon
must be very pure. The N2 content should be less than 1 ppm. This is only
guaranteed if the purity is better than 99.9999%, corresponding to a quality1) of 6.0
(sometimes also referred to as 6N0). However, this quality is relatively expensive.
Argon for spectrometry is at least 99.999% pure (Ar 5.0), and may therefore contain
up to 10 ppm of impurities. However, experience shows that there is hardly any
difference between qualities 6.0 and 5.0 in Germany, Europe and the USA (in terms
of N2 content), so that it is usually possible to use the cheaper spectrometer argon.
But in this case high purity is not guaranteed. The reason for the price difference is
the guarantee and the special filling procedure.
Your spectrometer allows you to carry out a simple check on the argon being used,
with the aid of a pure aluminium sample. Provided the Al matrix is part of the
analytical spectrometer programs, the recalibration sample RA10 is used for this
purpose. A pure aluminium sample is supplied with the unit, even if it does not have
an Al matrix. The software is described below. However, it is advisable to have a
small cylinder of Ar 6.0 in case of doubt (see below).
1) The quality tells you the minimum percentage of the gas in abbreviated form, the number before
the decimal point indicating the number of nines in the percentage figure. The number after the
decimal point is equal to the first digit in the percentage figure that is not equal to nine. For example,
Ar 4.6 indicates an Ar purity of at least 99.996%. Ar 5.0 corresponds to at least 99.999% argon.
Page 167
Changing the Ar cylinder
The changing of the Ar cylinder(s) is a critical factor in N analysis. Since the
separation of the valve from the Ar cylinder inevitably results in air getting into the Ar
supply, measures must be taken to expel it from the system.
You can simply flush the Ar system after changing the cylinder. You activate flushing
by selecting "Mode", "Argon flush". This method is very slow, as it can take several
hours before the pipes have been flushed sufficiently to ensure that N-analysis is no
longer impaired.
A better, though more expensive possibility is to use valves which you can buy and
which have a flushing device on the pressure side. This is the safest method.
A third possibility is to follow this procedure:
a) Turn the control valve from 3 bar to 0. This isolates the pressure regulator/
manometer area from the spectrometer.
b) Close the cylinder valve.
c) Unscrew the pressure regulator/manometer from the cylinder.
d) Replace the cylinder.
e) Screw the pressure regulator/manometer onto the new cylinder (do not forget the
seal).
f) Open the cylinder valve and immediately close it again. The primary manometer
should now display the cylinder pressure (usually 200 bar).
g) Loosen the union nut until the pressure between the cylinder and the pressure
regulator/manometer is able to fall to 0.
h) Repeat steps f and g four or five times.
i) Set the control valve to 3 bar.
j) Flush for approximately 2 minutes.
k) Check the Ar quality (N-factor, see below).
Page 168
Checking the argon quality and the equipment condition
An indicator known as the N-factor is used to check the Ar quality. This is equal to
the intensities of the N-line divided by the adjacent spectral background. An Al
sample containing no N is burnt. In the case of units with an Al base, this is RA10.
With others, a corresponding sample is included.
If you switch to a different matrix after testing, remember that the first analyses may
show an increased Al content. This is remedied by "burning clean", i.e. burning the
low sample several (three to five) times.
Figure 1
The sample is burnt and afterwards the N-factor is displayed, see Figure 2.
If it is greater, either there is a leak or the argon being used contains too much
nitrogen. In the second case you can check by replacing the argon being used with a
cylinder of "clean" argon. "Clean" argon can be either Ar 6.0 or a cylinder which has
previously been tested and found good and is kept for such purposes.
In the first case, a leaking connection in the Ar system is the likeliest cause. Use
leak detection spray or alternatively soapy water to check for leaks in the Ar
connections. You should also clean the spark stand and check the O-rings for wear
or damage. If no leak is found and the N-factor is still too high (>1.5), contact our
service department.
Page 169
Figure 2
Page 170
Appendix B.7
Support
If you require support, contact your Oxford Instruments Analytical Service Partner
or call the Oxford Instruments Analytical Customer Support Centre:
Appendix B.6
Page 171
Appendix C.2
AC / QDR
Page 172
Acceptance Certificate &
Quality Deviation Report
Type of instrument:
Installation date:
Company address:
IMPORTANT: It is essential that this document is completed by the end customer directly after the
installation and faxed or emailed to Oxford Instruments in order to both activate and protect your
manufacturers warranty.
Page 173
Quality Deviation Report ( QDR ) No.:_________
Filled by OI CS
Type of instrument:
Instrument serial number:
Installation date / period (from / to):
Representative:
End-Customer:
Deviation Description (please specify any issue(s) to be resolved)
Hardware:
Software:
Analytical Performance:
Incomplete delivery:
Page 174