1 Imprecision................................................................................................................................................................2
1.1 Within-run Imprecision............................................................................................................................................2
1.2 Day-to-day Imprecision ...........................................................................................................................................2
2 Linearity ....................................................................................................................................................................3
2.1 Linearity ......................................................................................................................................................................3
3 Comparison of Methods .........................................................................................................................................4
3.1 Study 1 ........................................................................................................................................................................4
3.2 Study 2 ........................................................................................................................................................................4
3.3 Study 3 ........................................................................................................................................................................5
4 Comparison of Two Instruments.............................................................................................................................5
5 Stability and Storage ................................................................................................................................................6
5.1 Real Time Stability Reagent ..................................................................................................................................6
5.2 Real Time Stability Calibrators.............................................................................................................................7
5.3 Real Time Stability Controls..................................................................................................................................8
6 Transport Simulation................................................................................................................................................8
CV% 3% 3% 3% 1% 1% 1%
Conclusion:
The above results confirm the low imprecision and the good reproducibility of the system.
2.1 Linearity
To determine the measuring range, a dilution series was produced from 10 routine blood samples mixed 1:1 with
calcium-free phosphate buffer. The hemoglobin concentration was then determined for each individual stage of
the dilution series. Each dilution series comprised 5 dilution stage. Next, the stage with the highest degree of
dilution was mixed again 1:1 with the next highest dilution stage.
18
y = 1,0033x + 0,0544
16
R2 = 0,9968
14
12
A:
10
0
0 2 4 6 8 10 12 14 16 18
B:
y = 1,0107x - 0,1031
1 6,0
R2 = 0,9843
1 4,0
1 2,0
B 1 0,0
8 ,0
6 ,0
4 ,0
2 ,0
0 ,0
0 ,0 2,0 4 ,0 6 ,0 8,0 1 0,0 12 ,0 1 4,0 1 6,0
3.2 Study 2
HumaMeter Hb Plus versus H liquicolor
20
19
18
17
y = 0,9401x +
16
15 R = 0,776
14
A13
12
11
10
9
8
7
6
5
4
3
2
1
0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
B
20
HumaMeter Hb plus comparison with reference method : Hemoglobin Liquicolor
19
18 Hemoglobin Liquicolor Cyan Methemoglobin Method [g/dl] A
respective Method [g/dl] B
17
16 y = 0,8759x + 1,4684
15 R² = 0,9083
14
13
12
y = 0,8138x + 0,4624
B 11 R2 = 0,8968
10
9
hemocue
8
Hbplus rgt 1,5 ml
7
6 Linear (hem ocue)
5
4
3
2
1
0
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
A
Conclusion:
Compared against the reference method cyanmethhemoglobin the Hemocue exhibits a significant
bias in the physiological relevant concentration range. HumaMeter Hb Plus, however, correlates well
in this concentration range.
21
y = 0,9705x + 0,5074
R2 = 0,995
19
Measured concentrations [g/dl] Hbplus 2
17
15
13
11
9
9 11 13 15 17 19 21
Measured concentration [g/dl] Hbplus 1
8
6
% Dev. from target
3 g/dl
4
6 g/dl
2
9 g/dl
0
12 g/dl
-2 3 6 12 18 24 29
15 g/dl
-4
18 g/dl
-6
-8
Storage time, Months
Conclusion:
The storage life tests showed that all of the reagents tested were very stable and suffered no loss of performance
even after a longer period of storage. This fact, combined with the basic chemical characteristics of the substances
tested (Triton-X, NaOH and Hämin) allows the stability of the reagents to be guaranteed for 36 months. The
transport simulation revealed no weaknesses in the reagent vials or boxes. Here, too, no negative influence on the
quality of the products is to be expected.
Measured values:
Lot 1
3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
3 0.089 3.11 0.084 2.94 0.085 2.97 0.088 3.08 0.083 2.90 0.085 2.97 0.087 3.04
6 0.172 6.01 0.178 6.22 0.171 5.98 0.168 5.87 0.172 6.01 0.172 6.01 0.174 6.08
9 0.247 8.64 0.266 9.30 0.253 8.84 0.247 8.64 0.252 8.81 0.245 8.57 0.253 8.84
12 0.353 12.34 0.340 11.89 0.347 12.13 0.349 12.20 0.350 12.24 0.346 12.10 0.340 11.89
15 0.433 15.14 0.412 14.40 0.428 14.96 0.428 14.96 0.436 15.24 0.419 14.65 0.421 14.72
18 0.503 17.58 0.515 18.00 0.489 17.10 0.528 18.46 0.531 18.56 0.502 17.55 0.521 18.21
Lot 2
3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
3 0.088 3.08 0.083 2.90 0.084 2.97 0.085 2.97 0.085 2.97 0.087 3.04 0.086 3.01
6 0.169 5.91 0.180 6.29 0.169 5.98 0.179 6.26 0.165 5.77 0.165 5.77 0.168 5.87
9 0.249 8.71 0.261 9.12 0.247 8.84 0.246 8.60 0.262 9.16 0.262 9.16 0.254 8.88
12 0.334 11.68 0.342 11.96 0.337 12.13 0.339 11.85 0.340 11.89 0.349 12.20 0.337 11.78
15 0.442 15.45 0.441 15.42 0.437 14.96 0.428 14.96 0.431 15.07 0.430 15.03 0.434 15.17
18 0.507 17.72 0.497 17.38 0.497 17.10 0.524 18.32 0.519 18.14 0.533 18.63 0.523 18.28
Lot 3
3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
3 0.085 2.97 0.090 3.15 0.089 3.11 0.090 3.15 0.084 2.94 0.089 3.11 0.088 3.08
6 0.171 5.98 0.179 6.26 0.173 6.05 0.175 6.12 0.174 6.08 0.169 5.91 0.175 6.12
9 0.264 9.23 0.264 9.23 0.253 8.84 0.263 9.19 0.250 8.74 0.264 9.23 0.263 9.19
12 0.335 11.71 0.341 11.92 0.352 12.31 0.351 12.27 0.349 12.20 0.350 12.24 0.348 12.17
15 0.427 14.93 0.427 14.93 0.445 15.56 0.420 14.68 0.418 14.61 0.435 15.21 0.425 14.86
18 0.516 18.04 0.494 17.27 0.499 17.45 0.500 17.48 0.498 17.41 0.518 18.11 0.523 18.28
Conclusion:
The real time stability tests of the calibrators showed that is no loss of performance even after storage for 39
months.
Lot 1
fresh 3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
6 0.086 6.01 0.169 5.91 0.180 6.29 0.169 5.91 0.179 6.26 0.165 5.77 0.165 5.77 0.168 5.87
15 0.429 15.00 0.442 15.45 0.441 15.42 0.437 15.28 0.428 14.96 0.431 15.07 0.430 15.03 0.434 15.17
Lot 2
fresh 3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
6 0.086 6.01 0.171 5.98 0.179 6.26 0.173 6.05 0.175 6.12 0.174 6.08 0.169 5.91 0.175 6.12
15 0.429 15.00 0.427 14.93 0.427 14.93 0.445 15.56 0.420 14.68 0.418 14.61 0.435 15.21 0.425 14.86
Lot 3
fresh 3 months 6 months 12 months 18 months 24 months 29 months 39 months
Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb Abs. Hb
g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl g/dl
6 0.086 6.01 0.172 6.01 0.178 6.22 0.171 5.98 0.168 5.87 0.172 6.01 0.172 6.01 0.174 6.08
15 0.429 15.00 0.433 15.14 0.412 14.40 0.428 14.96 0.428 14.96 0.436 15.24 0.419 14.65 0.421 14.72
Conclusion:
The real time stability tests of the calibrators showed that is no loss of performance even after storage for 39
months.
6 Transport Simulation
To determine the suitability and robustness of the reagent vials and boxes, 2 of each of the 3 types of vials (with
the outer box) were placed in an orbital shaker with one vial of each type inserted upside down. A pH indicator
strip was inserted into each box to show any reagent leakage. Additionally, the boxes with vials were subjected to
a 1-meter fall to the laboratory floor once every day. At the conclusion of the 3-week test period all of the reagent
vials and their boxes were undamaged. None of the pH strips showed any evidence of leakage of reagents or
standard.