Date of Issue March,2021 Date of Rev. - 26 Km Faisalabad road Feroze Wattoan Dist. Sheikhupura Rev No. N/A
QUALITY ASSURANCE LABORATORY- DENIM DIVISION
Standard Operating Procedure/Policy for ETP Water Testing.
Page 1 of 1 Sample collection point ETP Inlet Point ETP Discharge point Valve Minimum sample 500 ml Frequency of testing 3 Time per week (Monday, Wednesday, Friday) Parameters & standard methods Limits Optional Parameter Standard Standar s Foundational Progressive Aspirational PEQS d ▲15 / ▲10 / ▲5 / ▲3 / Temperature By thermometer(@time of sampling) Max. 35°C Max. 30°C Max. 25°C Max. 37°C pH APHA-4500-H+B 6.0~9.0 6.0~9.0 6.0~9.0 6.0~9.0 ISO 10523 TDS APHA 2540 C Not specify Not specify Not specify 3500 mg/l COD APHA 5220D 150 mg/l 80 mg/l 40 mg/l 150 mg/l ISO 6060 ISO 5815- BOD APHA 5210B 30 mg/l 15 mg/l 5 mg/l 80 mg/l 1,-2 TSS APHA 2540D 50 mg/l 15 mg/l 5 mg/l 200 mg/l ISO 11923
List of testing Equipment with make & model
Equipment Name Make Model pH/mV meter Mettler Toledo Switzerland S220 Conductivity meter Mettler Toledo Switzerland FE30 Seven2Go DO Meter Mettler Toledo Switzerland S4-Skit Multipurpose COD/BOD/TSS(70-750mg/l) analyzer Macherey-Nagel- Germany PF12 COD Reactor (12 holes for test tubes with 16mm OD) Macherey-Nagel- Germany Vario Compact Mili-pore Suction assembly for TSS/MLSS(Capacity 4L/hr) UK n/a Weighing Balance Mettler Toledo Switzerland ME1002-E
Results Compilation: After testing results data will be re-view by QA Lab head & informed to ETP manager on same day for their further action.