Document No DFL-QDOC-014

Diamond Fabrics Limited


Date of Issue March,2021
Date of Rev. -
26 Km Faisalabad road Feroze Wattoan Dist. Sheikhupura Rev No. N/A

QUALITY ASSURANCE LABORATORY- DENIM DIVISION

Standard Operating Procedure/Policy for ETP Water Testing.


Page 1 of 1
Sample collection point
 ETP Inlet Point
 ETP Discharge point Valve
 Minimum sample 500 ml
Frequency of testing
 3 Time per week (Monday, Wednesday, Friday)
Parameters & standard methods
Limits Optional
Parameter
Standard Standar
s Foundational Progressive Aspirational PEQS d
▲15 / ▲10 / ▲5 / ▲3 /
Temperature By thermometer(@time of sampling)
Max. 35°C Max. 30°C Max. 25°C Max. 37°C
pH APHA-4500-H+B 6.0~9.0 6.0~9.0 6.0~9.0 6.0~9.0 ISO 10523
TDS APHA 2540 C Not specify Not specify Not specify 3500 mg/l
COD APHA 5220D 150 mg/l 80 mg/l 40 mg/l 150 mg/l ISO 6060
ISO 5815-
BOD APHA 5210B 30 mg/l 15 mg/l 5 mg/l 80 mg/l 1,-2
TSS APHA 2540D 50 mg/l 15 mg/l 5 mg/l 200 mg/l ISO 11923

List of testing Equipment with make & model


Equipment Name Make Model
pH/mV meter Mettler Toledo Switzerland S220
Conductivity meter Mettler Toledo Switzerland FE30
Seven2Go DO Meter Mettler Toledo Switzerland S4-Skit
Multipurpose COD/BOD/TSS(70-750mg/l) analyzer Macherey-Nagel- Germany PF12
COD Reactor (12 holes for test tubes with 16mm OD) Macherey-Nagel- Germany Vario Compact
Mili-pore Suction assembly for TSS/MLSS(Capacity 4L/hr) UK n/a
Weighing Balance Mettler Toledo Switzerland ME1002-E

Results Compilation:
After testing results data will be re-view by QA Lab head & informed to ETP manager on same
day for their further action.

Prepared by: DM Quality Reviewed by: Sr. Manager Approved by: GM