GC955
MANUAL
JUNHO 2013
SYNTECH SPECTRAS Analyser GC955
Synspec b.v.
De Deimten 1
9747 AV Groningen
Nederland
info@synspec.nl
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 2
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 3
Contents overview
INDEX ________________________________________________________________
_________________________________________________________________
_________________________________ 131
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 5
Chapter 1 Technical
Technical instrument date ( if applicable)
1.1 Technical-
Technical- and order specification
This manual for the Syntech Spectras introduces you to the installation, the running and simple
maintenance
maintenance of the Syntech Spectras gas chromatographs.
The instrument has been built for semi-continuous measurement of hydrocarbons in air and other gases.
WARNING:
The instrument is to be used only by trained personnel.
Please note that the following risks have to be considered when working with the instrument:
• A relevant selection of the gases nitrogen, helium, hydrogen and air at a pressure from 0.5 to 5 bar will
be connected to the instrument. Personnel must be schooled in working safely with these gases.
• The instrument works with a 220VAC power supply. Inside the instrument there is a high voltage supply of
2 kV in a protective housing.
• The pre concentration unit will be at a temperature of 100 to 260 °C during desorption.
The instrument has been tested for accordance to the EMCdirective 89/336/EMC, test specification EN 50081-1:
1991 and EN 50082-2: 1994.
There is a warranty on the instrument of one year after delivery. The warranty does not include filters, pre
concentration tubes and capillary columns. For a PID lamp there is a two-year warranty.
WARNING: If you do not copy your data in a regular cycle, loss of data is at your own risk! Of
course the hard disk itself is replaced under warranty! We expect the life span of your hard disk
to be about 5 years!
NOTICE:
• Please inspect the instrument for transport damage after receiving it. In case of damage we ask you to
contact your supplier within 8 days after receiving the instrument.
• Be sure to read chapter 2 before installing the instrument, and chapter 3 and 4 before switching the
instrument on.
• If you have questions about running the instrument, please contact your supplier or Synspec. (Tel. +31
50 526 6454, fax +31 50 525 6540, e-mail info@synspec.nl)
• Always disconnect the gas and power supply before opening the instrument
• Do not move the instrument when the internal computer is working to avoid damage to the hard disk.
• Beware of electrostatic shocks to the internal computer: do not drop metal parts in it, take care to use
appropriate equipment.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 9
Standard 220 VAC (210 - 230 VAC), we recommend 600 VA, however for normal start-up the maximum demand
is about 300 VA, at continuous use the demand is about 90 VA. (A 110VAC, 60Hz version is avaliable)
For standard applications, unless otherwise stated, room temperature may vary between 5 and 35 °C . However, it
is not advised to use a room temperature regulation that results in temperature changes of more than 5 oCper
hour. In this case stable retention times are not reached.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 10
Must be 20 - 90%.
2.2.4 Dimensions
The instrument is built for a 19” rack. We advise to give the instrument a ventilation space of 1 standard HU at the
bottom and top. Mounting on a rail is advised. If the GC is placed on a metal or wooden board, please pay
attention to good ventilation at the bottom: either by using a ventilated housing or by using a perforated board.
Do not forget to reserve a place for keyboard and mouse, preferably a board on slides, which can be 10 to 50 cm
below the GC.
All fittings to the instrument must be 1/8” Swagelok-compatible, for the gas supply in brass, for the sample in
stainless steel (this also prevents confusion).
As regulators and the connections to bottles are not standardized, the best idea is to get advice from a good gas
supplier in your country. Explain what you want to do with the equipment before ordering. Never order pressure
regulators in another country, this is simply a waste of money!
For the carrier gas nitrogen, gas chromatographic quality means: use a nickel-plated bronze regulator with steel
membrane, range 0 - 4 or 0 - 10 bar. Use bottles with a maximal pressure of 200 bar.
For calibrating gases a stainless steel regulator with high quality steel membrane is preferred.
Nitrogen of GC-quality (quality 5.0 or higher, i.e. 99.999% pure) must be used, unless otherwise specified. It
should be connected by a 1/8” FEP or Teflon tube, maximum length 2 m, to the carrier gas port. The pressure at
the port should be between 2.5 bar and 4 bar.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 11
Two of our detectors use no other gases besides the carrier gas: the PID and the TCD.
To generate the signal the FID has a hydrogen flame, which uses hydrogen and air. For hydrogen, bottles
or a generator can be used. In a lab a bottle is a good solution, provided it can be stored well. For a
mobile station we think generators are superior: this way the user has no responsibility for transporting
and storing bottles in a field situation.
A hydrogen generator is expensive. But, considering its lifetime, in the end it is not more expensive than
bottles.
The PDECD uses very pure helium to generate a strong UV-light, which is used to ionise xenon. Xenon is
dosed as a 3% xenon in helium mixture.
Both gases must be provided in quality 6.0. The helium must also have an extra catalytic purifying step.
All materials in connection with the gases must be ultrapure and must all be stainless steel.
The internal diameter of the inside tubing is mostly 0.5 to 1.0mm. So the sample has to be filtered. We advise a
Teflon filter membrane of 5 micron, preferably in a 1/8” housing, alternatively in a 6 mm or 1/4” housing with good
connections to the 1/8” tubing. Alternatively disposable nylon filters with glass frit or Teflon filters can be used.
Clean inert tubing at the sample-port is essential for obtaining good results.
Use only PTFE (Teflon), FEP or stainless steel, even for short connecting pieces. Do not use other plastics or
bronze.
Do not use the same tubing for taking ambient air samples and calibration samples. Use two tubing systems, that
are only connected the last 5-15 cm before the sampling port.
Of course the tubing can be connected to central sampling systems with a large volume pump. Contact us if you
have doubt about the GC ‘s position within a larger sampling system.
We advise a daily span check. If these checks are stable, a basic calibration is only needed one to three months.
Without an automatic daily check a manual span check is advised every two weeks at least until the instrument is
stable, where after a monthly check is the minimum.
See chapter 4 for more details about the calibration.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 12
Before the GC is switched on after disuse it should be inspected. After normal transport it may be presumed that
the instrument will function correctly.
The instrument should be inspected for loose parts, as they can do much harm in the oven ventilator and the
computer.
After transport it is useful to check
- the correct position of the computer’s boards in the connections, this is simply done by pressing the boards
down into their connectors.
The gas lines of the instrument have been closed for transport. Remove the metal plugs at the backside, keep
them to use when the instrument has to be moved without gas lines connected to it. The ends of the two sample
gas lines are also protected for some columns: with a plastic plug at the back (sample outlet) and a similar
plastic plug in the computer compartment at the front protecting the end of the detector output.
See the scheme of the rear side: connect carrier gas to carrier gas port. The stainless steel Tee that is delivered
for the sample tubing must be connected to port 4 and 5; then the sample tubing must be connected to it. Place
the sample filter in the last piece of tubing before the sample Tee.
For an FID: Clean, dry compressed air must be connected to the AIR-port, pressure at the port at least 3.0 bar
but less than 4.0 bar. Hydrogen must be connected to the hydrogen port, pressure at the port 2.5 bar.
Test for leaks!!
Turn on:
1. GC and PDECD can be started independently. The best thing is first to start the PDECD and then the
GC.
2. Put the switches of the PDECD in the following position:
Temperature 100 °C
Mode: at PDECD
Range: at 1
Power:-
Amplification (Attenuation) at 1
3. First of all switch the power supply of the PDECD on.
4. Wait until the detector temperature is stable (for 5 to 10 minutes), then put the discharge at ‘on’.
5. As the lamp starts, the background will fall from 200 to circa 1-10 mA.
6. Start the gas chromatograph.
If the helium has been disconnected this can take up to circa 1.5 hour. Otherwise circa 5 minutes.
When the analyser will not be used for a longer time, the choice are :
• To let it run on without doing anything with the results: this will enable you to start up in minutes,
the only problem is that it consumes power and gas.
• Stopping the run and leaving it ready for restart: this is generally better. There is less
consumption of gas and power, but still a very quick restart. Not advised with cold traps, as
condensation may give problems in the trap when it is not heated out frequently.
• Switching it off completely: in this case take care that no dust or humidity comes in.
A special case is the PDECD, which is the most difficult to turn on.
1. In case of a PDECD detector: during short interruptions its the best option is to let the gas supply
run normally and keep the PDECD switched on.
2. If the PDECD must be switched off, first switch off the discharge.
3. Then disconnect the current of the PDECD unit.
4. During lengthy interruptions disconnect the power of the purifiers and close off the gas bottles.
This will prolong the startup time to circa 2 hours.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 14
4 3
8
2
5
1 9
10
6 7
1 Carrier gas
2 Output pump
3 Sample inlet
4 Stripper outlet
5 External connectors
6 Digital input and output connectors
7 Analogue input and output connectors
8 Oven ventilation
9 Instrument ventilation with filter
10 Power, 220 V
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 16
9 4
6
5
1 Oven
2 Lamp /detector
3 10-port valve
4 Computer
5 Gas regulation cabinet
6 Hard disk
7 High Voltage board and computer power supply. Danger:
anger high voltage
8 Power supply
9 Flow sensor
10 Hard disk
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 17
3
2
1
6 5
9 8
11
7
12 10
GC 855 combined with stream selector and complete gas supply system for automatic validation
A1
A2 1 (spa ngas)
2 3 4 5
A3
6 7 8 9 10 (Zer o - air)
Exhaust
Purge
Fuse 1,5 A
Sample out
air
Strippe r out 10 p -valve Samplep ump
Sample in
helium Calibration
gas bottle
SAM PLE PUM P OUT
cylinder
Carrier gas
Stream selector 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 5 6
Compressor
3.0 Introduction
In section 3.1 the operating principles of the gas chromatograph are explained: sampling, separation and
analysis. Section 3.2 gives an overview of the detection and separation possibilities, section 3.3 of the
electronics, software and the data handling
3.0.1 Overview of the options for the Syntech Spectras GC955 series
The gas chromatograph GC955 is produced in 7 different types, adapted to the different requirements of
the compounds that have to be measured, the field conditions and the different uses by the customers.
Generally instruments are sold adapted to the requirements of the customer and calibrated for the
compounds that will be measured.
The different combinations of the options in points (4) to (6) lead to 7 different series.
The instrument can also be supplied without an LCD screen.
For sampling from up to 10 different points a stream selector can be added.
This manual is written for the series 200, 400 , 600 and 800, as they all have pre concentration. Although
column type, temperature regulation and detector which are used may be quite different, the gas flow
schemes are quite similar.
A gas chromatograph for the automatic measurement of air samples and other gases has a special
method to take a sample: not by injection, but by leading the sample through a loop. Either the sample is
pumped, or the sample flows in under a slight overpressure. By switching a valve the sample is
introduced into a gas chromatographic column where it is separated. Then the separated compounds
are measured in the detector. The flow of carrier gas and sample through the system is complicated. By
way of a sophisticated 10-port valve, which is switched between two modes -position 03 and 04 - they
are sent in the right directions. Figure 3.1 and 3.2 give an overview of the gas lines in both modes of the
valve.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 22
3.1.1 Sampling
Each cycle, the gas chromatograph takes a fresh sample of the air or gas that is to be analysed into a
sample loop. This is done by a small membrane pump in the instrument. The capacity of the pump is 1.5
l/min. Depending on the distance of the instrument from the source the pump can be switched on from
10 seconds to several minutes.
The dimensions of the sample loop depend on the concentration of the compounds that have to be
analysed: from 75 microliters for the high ppm range to 3 ml for the low ppb range.
The machine works semi-continuously: the first step is flushing the sample tubing by drawing air (or
sample gas) through it with a pump. Then the pump is switched off and with the help of an indirect piston
system a volume of 35 ml sample gas is preconcentrated on a Tenax® or Carbograph column. This
procedure can be repeated until enough sample material has been drawn. Before each sampling the
pump can be switched on again for a short time. The pre concentration tube can be purged with carrier
gas to remove oxygen and water if desired.
Fig. 3.3: principle of pre concentration: the dark arrows show the airflow for emptying the piston, the white arrows show the
backflush, during which the sample is sucked through the pre concentration tube
For hydrocarbons with boiling points below 20 °C the absorption capacity of the medium can be a
problem, because the compounds can break through during sampling: in this case they are not trapped
completely and in the time between trapping and desorption they may evaporate. Enhancement of the
capacity is possible by cooling the pre concentration tube during absorption. Bij 3.1.2
The pre concentration tube can be cooled by radiation. To do this, the pre concentration tube is
embedded in an aluminium block. This block is cooled on both sides by Peltier elements. To avoid
freezing or condensation in the system, drying of most samples will be needed. The lowest temperature
that can be reached in practice is – 5 oC.
The sample is desorbed in a short time by heating the pre concentration tube quickly, while flushing it
with carrier gas. By way of the 10-port valve in position 03 the sample is brought on the separation
column. (See the diagram 3.1)
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 24
The separation column consists of two parts : a stripper column and an analysis column. These are
usually two identical columns, with different lengths. The goal is to prevent a prolonged analysis, which is
normally caused by the necessity of waiting for elution of the highest boiling compound. Therefore the
time is set for the slowest eluting compound of interest to get through the column. After this time the flow
in the stripper column is reversed by switching the 10-port-valve.
Figure 3.4: Gas chromatographic separation with a sample loop and stripper/analysis column: the air is pumped into the loop, after
which the whole sample is brought onto the stripper column. After the passing of the desired compounds the rest is flushed back.
The earlier compounds are separated on the analysis column.
3.1.4 Separation
The rest of the analysis is performed with the 10-way valve in position 04. ( See the diagram 3.2 ) The
compounds which already passed through the stripper column continue their way through the analysis
column to the detector. (see 3.2) The gases, which already were seperated somewhat in the stripper
column, are seperated as far as possible in the analysis column .
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 25
The analysis can be done at a constant temperature or with a temperature program. With a cyclic
changing temperature compounds with more differing boiling points can be separated and determined
within an acceptable cycle time. During separation the oven is heated, usually 15 - 60 °C above the basic
temperature. After the heating phase cooling down has to take place before the next cycle can start. This
takes a few minutes.
For automatic measurements with the GC the cleaning of the system is essential to prevent deterioration
of the column, which would result in detoriation of the peaks. The backflush column is flushed every cycle
for some minutes separately from the analysis column in the reversed direction. To clean the analysis
column each cycle has to last until no more signals are detected.
The time setting for cleaning is done for normal ambient air. In a very dirty atmosphere a longer time may
have to be set for to clean the stripper.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 26
3.2.2 PID
Standardly the instrument will be provided with a Photo Ionisation Detector, 10.6 eV, with a measuring
cell of 50 µl.
This detector will achieve a high resolution and allow the measurement of very low concentrations: for
aromatic hydrocarbons, alkadienes and sulfur compounds down to 150 ppt, i.e. 0,4 µg/m3.
Photo ionisation is the process by which a photo excited electron absorbs enough radiant energy to be
ejected from the atom or molecule. Photo ionisation is initiated by the absorption of a 10.6 eV photon by
the molecule. If the molecule has an ionisation potential equal to or less than 10.0 eV, the following
process occur.
R + hv -----> R+ + e-
where hv is a photon with approximately 10.6 eV and R is the species of interest. Molecules with
ionisation potentials greater than 10.6 eV will yield a lesser response
The detector source is a low pressure discharge lamp filled with noble gas. It provides a stable
monochromatic source of high energy photons suitable for ionisation of the various components in gas
chromatography.
The UV source is contained in a vacuum-tight envelope with a UV grade magnesium fluoride window
maintaining isolation of the UV source from the gas chromatograph's carrier gas. The lamp’s output is
approximately 100 microwatts of 10.6 eV photons contained in a beam 6 degrees wide.
The PID-signal is amplified, the signals are then stored as a value of 0 - 10 V. On the LCD-screen the
chromatogram can be followed. Peak height and peak area can be quantified and stored for those peaks
that are of interest.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 27
Figure 3.5: An analysis of an airsample loaded with paint residues as a result of painting activities one day before
The PID has a low sensitivity for saturated hydrocarbons. This detector is however specifically sensitive
for aromatic hydrocarbons, unsaturated compounds and heterogeneous hydocarbons like sulphur
compounds. Because of this, a group of interesting substances in the fields of environmental pollution,
process monitoring and industrial hygiene can be assessed.
The photo ionisation detector is ideally suited for continuous monitoring with a gas chromatograph: it is a
stable, sensitive detector, requiring no special gases.
The selective sensitivity makes possible monitoring of a range of (eco) toxics in the ppt to ppm-range.
Nitrogen or helium can be used as a carrier gas.
3.2.3 FID
The flame ionisation detector is the almost universal detector for the gas chromatograph, often used in
the laboratory.
Its sensitivity makes it suitable for the low ppb to ppm range. Needing a hydrogen flame and pure air as
a source of oxygen, the running cost of this detector is higher than for a PID. Hydrogen is shut off when
the flame does not burn, the flame temperature is monitored for this. Nitrogen or helium can be used as a
carrier gas.
The combination of a PID and an FID gives the opportunity to monitor complex mixtures satisfactorily.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 28
Principle of flame ionisation: in a flame of oxygen and a burning compound ions are formed. When other
molecules than hydrogen pass through a hydrogen flame, these will also burn and increase the amount
of ions formed compared to pure hydrogen.
The ions and electrons generated by the ionisation are attracted to the positive and negative side of an
electrical field in the detector cell. The electrical current in the detector cell changes, and this change is
measured.
Fig. 3.6 With a double detector system, combining an FID with a PID or a TCD, the flow scheme for a system 600
will be as follows:
first detector,
TCD or PID,
only carrier gas flows through
3.2.4 TCD
The thermal conductivity detector is the oldest detector in gas chromatography. Its sensitivity is less than
for the PID’s or the FID’s, which means the TCD is suitable to monitor compounds in the low ppm to %
range.
However, the TCD can measure compounds not detected by the other two (see section 5). The carrier
gas is usually helium.
Principle: The temperature of a conductive material wire influences the resistance. A gas flowing through
a heated chamber cools the wire.
Depending on the molecules in the gas the uptake of heat changes, this factor is called the thermal
conductivity. Changes in the conductivity of the wire are measured.
3.2.5 PDECD
The newest detector in this range is the Pulsed Discharge Electron Capture Detector. It is very sensitive
to those compounds that can trap electrons easily and not sensitive to others.
This means in practice that the PDECD is sensitive in the ppb-range to hydrocarbons containing chlorine,
bromine, fluorine . This detector needs a steady supply of purified helium and of a dopant gas.
Principle: helium is ionised by a pulsed discharge. High energy photons are emitted in this ionisation.
The photons in their turn ionise the dopant, which is Xenon. This results in a Xenon ion and an electron.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 29
The electrons give a standing current, which is diminished when they are captured by the sensitive
compounds. The pulsed discharge principle is an electron source without the conventional radioactive
Ni-element. Users do not need training in handling radiation risks.
With the combination PID/PDECD a range of interesting environmental and toxic compounds can be
measured and identified.
Fig. 3.7 A flow scheme for a system with sample dryer looks as follows:
first detector,PID,
only carrier gas flows through
second detector,
Sample dryer Preconcentration Stripper column Analysis column PDECD
tube
Carrier gas
3% Xenon in Helium
Gas supply to a double detector system with PID and ECD Helium
The separation is determined by the column, the basic temperature, the programmed temperature rise
and the flow velocity of the carrier gas. The separation capacity of a column is determined by its length,
the type of coating and the quotient of the coating thickness and the inner diameter.
All commercially obtainable packing materials can be used in stainless steel columns of 2.1 mm ID up to
5 meters long and in fused silica or steel capillary columns of up to 60 m ( at I.D. 0,32 mm) and up to 30
m (at ID 0.53 mm.)
By choosing the column material and the temperature area, many different mixtures can be separated.
3.3 Electronics
Electronics and software
The GC955 uses an industrial Pentium computer. The electronics boards specially made for the
instrument are the power supply and the central function and signal control board. The instrument is
EMC-compatible . The power supply and the central board are situated at the underside.
The instrument has electronic sensors to measure the flow of the carrier gas and the pressure of the
valve switching gas. For sampling there are sensors for the sample pressure and temperature, to enable
correction for outdoor temperature and outdoor air pressure of the sample volume. An optosensor
checks the functioning of the piston.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 30
The control of the cycle program is done by a user-friendly program under Windows 98. In a menu the
present cycle can be followed, or a stored chromatogram can be reloaded and settings of the cycle
program can be changed.
The GC955 is delivered with settings, which are adapted to the application as specified.
Changes in the program can easily be made in a bar diagram, in which is visible directly, what the
consequences of the proposed changes would be for the program.
Of course it is also possible to change the program for the determination of other compound mixtures.
The software has its own help function.
The software has the following menu structure:
Animate data
Data selection
Animate the
Stop animation
The measured values are saved on the instrument’s hard disk. Measured cycles can be restored on the
LCD - screen, but also at the same time on another connected PC-screen (in a network, or on a
connected workstation).
Height and area of measured peaks are calculated for selected areas of interest in every cycle, for
instance for methylmercaptane and ethylmercaptane. The program has a default setting for integration:
start and end of peaks are determined by calculating and evaluating the first and second derivatives of
the detector-signal. If desired, the setting can be adjusted for each individual signal.
These values are saved as value per cycle on the hard disk. Besides this the values are passed on via
the RS232-port in the datastructure as specified by the customer
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 32
The measured raw data can also be read out directly via three analogue ports at the backside of the
GC955.
Calibration data can be entered through a separate menu. With these calibration values a calibration line
is calculated. In this way the measured data can be read out and stored as concentration values.
The raw data and the calculated concentrations may both be stored to allow for later corrections.
Communication is possible via the digital and analogue in- and outputs, via the RS232-connectors,USB
and ethernet. Several functions of the GC are checked continuously; if there are problems an error
message is given. Status information can be sent to and from the instrument .
RS232-port
The same messages can be send over the RS232-ports and a modem connection towards a control
station. By online control further information about the problem can be obtained, errors in the setting of
the program can be fixed, or other optional changes can also be made by modem communication. GC’s
with connected equipment for calibration can also be calibrated in this way.
Remote control of the instrument is best done with a Windows based remote control program. Synspec
recommends the use of Symantec's PC-anywhere (version 8.0 or higher). The remote control can in this
case take place over modem, or over a ethernet-LAN.
With a normal LAN connection, a shared disk can be obtained, to retrieve ASCI-data from the GC.
The USB ports are used to connect external (zip)drives, CD-reader and writers etc.
Overview of data transfer and of communication options: for each option the useful elements are
explained, if an option does not fulfill a task it is labeled “no”.
Data transfer Status information Instrument control
Analogue outputs
Concentration values will be No no
transmitted
Analogue inputs
Can be used to read in data no no
from other instruments
Digital outputs
no Yes, all set parameters in Limited: some actions,
the digital like runstart, or start
communication protocol calibration can be set.
will be transmitted.
RS-232-
Concentration values will be Yes, all set parameters in Limited: some actions,
communication transmitted the communication like runstart, or start
protocol will be calibration can be set.
transmitted.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 33
RS-232 modem
Yes, all data can be copied. Yes, as the GC screen Can be used to change
and PC Anywhere Speed depends on modem can be seen settings and then
and phone line activate them. Only
possible if phone line is
stable.
Ethernet
yes no Not directly
connection Can be used to check on
remote PC
measurements and then
to copy back improved
programs.
USB connection
Can be used to copy data no no
from the hard disk to zip
drive, memory stick, CD-
writeable etc.
The GC955 is adapted for use as early-warning system for toxic compounds. Levels for low and high
alarm can be set. Alarms can be given over modem, or over digital outputs. The GC can also be
provided with a buzzer or lamp for an alarm on site.
4.0 Introduction
The GC955 is a gas chromatograph for automatic measurements of compounds in air and other gas
mixtures. In this chapter the normal functioning of the instrument is explained: the handling of the gas
chromatograph, the calibration and the data handling. At the end of several sections there is a
summarizing table.
There are a lot of possibilities to change the instrument’s program. Data handling and communication
can be set to adapt to all sorts of protocols. All this is covered in chapter 5. For normal use these
parameters are set for the customer and do not have to be changed.
Visual inspection
Before the GC is switched on after disuse, it should be inspected. After a normal transport it may be
presumed that the instrument will function correctly. The instrument should also be inspected for loose
parts, as they can do much harm in the oven ventilator and the computer.
After transport it is useful also to check the flow of the stripper column with a flow meter.
Connect helium via a purifier to the inlet on the top of the detector, helium doped with xenon via the other
purifier to the dopant inlet. Please do not open the inlets until just before connecting them, and only open
the bottles when they are connected.
To lengthen the lifetime of the LCD-screen’s backlight the Windows screensaver is activated. This means
that the screen will go dark if mouse and keyboard are not used for longer than 30 minutes. Under the
Windows 98 Main panel\Settings\Control panel\Desktop\Power saving features this setting can be
changed.
1. Chapter 1 identifies a lot of manufactured parts in the instrument. Many of these are delivered with
instruction leaflets about use and maintenance. As far as they are useful to the customer they are
supplied with the manual.
2. The GC comes with a range of settings which are saved in the GC955id.txt. These are, for each serial
number individually, the calibrations of the following sensors: flow sensor, pressure sensor for pneumatic
air, and pressure sensor for the outdoor air-pressure. There is also a list of 10 names of compounds that
can be selected. The factors for conversion from ppb to µg/m3 are also set here, as they differ from
country to country. When sensors are replaced or other compounds must be measured, this file has to
be corrected..
3. The application as ordered is saved as one or more run-programs. These work with a flow setting that is
noted in the technical instrument data (chapter 1).
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 38
The instrument can be switched on once the gases are connected and the filter is in place. The GC must
be connected to a mouse and a keyboard. The oven cover must be closed. The L-shaped cover must be
put on.
For a TCD
TCD: always make sure that the supply of carrier gas is sufficient.
For an FID: the instrument will not start up completely if the flame does not start; so check the
supply of air and hydrogen.
For a PDECD: see to it that the switches are set in the following way:
Temperature : 100 °C
Mode: on PDECD
Range: on 1
Current: -
Attenuation: 1
First, switch on the power supply of the PDECD. Wait until the detector temperature is stable (5-
10 min.) then switch on the pulsed discharge. When the lamp starts the background current
drops from 200 to ca 1 - 10 mA. If the helium has been disconnected the lamp may take up to 1.5
hour to start. Otherwise it takes 5-10 min. When the lamp has started, start the gas
chromatograph itself.
After starting the GC the Windows 98 program starts and the 955 screen appears. A self-test is
performed on electronics and PID lamp and the oven is heated to the base temperature of your default
application program. Then the PID lamp is started.
For an FID: the instrument will not start up completely if the flame does not start: so check the supply of
air and hydrogen. When starting, the hydrogen valve is opened. With newly connected lines, the
hydrogen flow limits the burning, and the instrument will not start in the first run. For safety the hydrogen
valve shuts after 10 seconds if the flame does not burn. This process is repeated 6 times, then an error
message will appear. Close the program. Start anew until the flame burns. After transport this may take 3
to 4 repeats of the procedure. The burning can be checked by holding a mirror to the FID-exhaust at the
rear of the instrument. If the connections have been left in place, generally the instrument will start
quickly.
If the instrument is set to start automatically after power failure, this will also work. This somewhat long
procedure is designed to guarantee a safe start.
If the instrument was working, a new start can only begin after the flame has been exhausted: until then
you will see the message: “reset state not low”.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 39
If you have taken off the hydrogen or air on a working instrument, the hydrogen safety valve will close. To
restart the FID, click options\actual run\restart FID.
The carrier gas flow through the capillary column should be between 1.0 and 5.0 ml/min., through the
packed column between 12 and 25 ml/min. This will be set by the mass flow controller. Checking for
leaks or obstructions is advised after long transports, or after working without adequate filters. The flow is
set for the application.
A pressure sensor is used to check if the pressure to switch the diaphragm valve is correct.
The program contains seven different functions: actual run, data file, program edit, calibration, graphs,
autoliniarisation and IO-state diaphragm. See also the overview of the program structure in Chapter 3.
In the view mode pop-up-menu you can select the desired mode.
1. In actual run the present measurement is shown on screen with the measured values below.
2. In the data file mode you can recall all measured chromatograms to the screen to compare them.
(See a data-file in figure 4.1)The hard disk (>3 GB) has a large capacity to store chromatograms.
3. The program edit menu enables you to change the program, (chapter 5),
4. the calibration program is used for calibrating (4.3).
5. The graph is used for a quick view of the measured compounds.
6. The option autolinearisation can be used for a multi-point calibration (chapter 5)
7. The IO-state diagram shows the instrument’s present state of functioning.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 40
The default mode is actual run. After an automatic check of the settings, the program is standby for run.
The oven must reach the desired starting temperature before the first cycle is started by clicking start run
with the mouse. Now the background zero adjust signal is set, then the program starts. The detectors
need some time to stabilize, so the first reset may take a minute. (In automatic mode it will start either
immediately or as soon as synchronization is reached.)
The measured values can be seen on the chromatogram.
In the status lines below the chromatogram information is given : clock time, cycle time, the actual
measured value for the detectors, ambient temperature, oven temperature.
Further extra sensor signals, from left to right: pressure of valve switching air, flow of carrier gas.
The ambient temperature is measured on the underside of the instrument.
The base line is normally set at about 10 mm above the edge of the chromatogram window. This can be
changed under actual run\options\adjust detector Y-offset.
With two detectors it is most practical to keep one on offset zero and the other one at 30. With unstable
baselines you may want to put it even higher.
Timing
The sample that is analyzed is always the sample that has been preconcentrated in the last run.
Each application has its own cycle time. There are two possibilities: either the next cycle starts
automatically after the end of the last run, or the instrument measures synchronously. See section 5.5 for
details. If you want to stop measuring click with the mouse on stop run.
Choose view mode, then data file in the pop-up menu to recall files for comparison with the present
measurements. Choose the directory for the correct month (for instance d_9505 for May 1995)
The chromatograms are saved in month-directories, with day-subdirectories, with a filename derived from
the measurement time in days, hours, minutes, prefixed by an m: e.g. m_281435.bin.
The chromatogram appears if you double-click a file. The scale in the x-axis and the y-axis is the same as
in the chromatogram that is being measured, to make a comparison simpler. You can move freely
between the modes actual run and data file.
For automatic measurements of these peaks in coming runs, you must set the peak windows.
1. Click options, then edit peak windows and set the windows you want to determine. Edit peak
windows allows you to choose from a list of compounds and to set the window time in seconds.
2. The effect can be seen after clicking select screen elements under options and then checking the
box show peak windows. With each peak window can also be shown: the retention time of the
measured peak, the integrated area and the calculated concentration. See chapter 4.4, automatic
measurements, for the finer points of peak window setting.
3. To use these windows in the following runs, return to actual run and click start run.
Fig. 4.4: A calibration line for benzene from a permeation tube source, diluted to a range from 1.2 to 11 ppb
A calibration is used to determine the conversion factor for automatic calculation of concentrations from
area values. It is also used to check the correct functioning of the instrument.
The frequency of calibrations, the number of points determined and the number of repetitions per point
depend on your quality control scheme. For calibration we provide several facilities, see 4.3.4..
Before starting the calibration you should set the peak windows (see 4.2.2), this must be done on the
peaks of a file in data file.
Fig. 4.5: The window calibration under actual run, used to start or stop the calibration mode
Start sampling the calibration gas during the next actual run. The following run the calibration gas mixture
will be analyzed, to use this for your calibration curve you must click set calibration active under options
in actual run.
There are three choices: start new calibration, add data to current calibration and add quality control point.
If you click start new calibration the present calibration is no longer active. To prevent mistakes this
choice must be confirmed before it is activated. Add data to current calibration can be used if a
calibration had to be discontinued for some reason. A calibration must still be completed within 24 hours.
A validation point can be made with add quality control point, if for quality control the validity of a
calibration has to be checked. (See 4.3.5)
The measured calibration values are used to make a calibration line. The correlation coefficient of the line
will be calculated, to indicate the linearity.
Fig 4.6: Calibration selection dialog, used to select the files which will be used to calculate the calibration line
Fig. 4.7: Settings for calibration: choose the unit in which the data are to be saved
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 46
5. The calculation of the calibration line can be done in three ways. For normal use the linear regression with
fixed base point is the best. For measurements where the very low range and high range of higher boiling
compounds are important (i.e. with toluene below 0.5 ppb and above 50 ppb) the non-linear regression is
better as there is with higher boiling compounds a slight deviation from linearity in the higher range.
6. After this calculation has been done, this calibration is used to calculate concentrations out of raw
measured data until the next time that start new calibration is activated.
We advise a daily span and zero check. If these are stable, a basic calibration is only needed every
month up to three months. Without an automatic daily check, it is advised to do a manual span and zero
check at least every two weeks.
To obtain low calibration concentrations, normally diluting of commercially obtainable calibration gas with
zero-air is needed.
To save calibration gas you can switch off the sample pump during calibration by clicking disable sample
pump under options in actual run
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 47
The quality control system often requires a regular calibration check. For this you should enter a
calibration sample during maximally half an hour. How many runs this is, depends on your program. If
the option add quality control point (under actual run\ options\set calibration mode) is chosen, these
values will be drawn on the calibration line. The calibration itself will not change.
There is also an option to automatically make a quality control point at regular frequencies. For this
please contact Synspec or your distributor about the details of the program we must prepare. When you
use this option the calibration gas has to be switched on automatically and the way this is done will vary
with the customers’ own choice of equipment.
The program is already prepared to set the time for a validation run. See for this chapter 5.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 48
A well adjusted instrument will run for two to four weeks without needing attention. Of course, the gas
supply must be sufficient and calibrations should be made following the quality standard.
It is advised to set the program for automatic restart after power failure and for synchronization of cycle
time with clock time, see 5.5, run settings. However, you can always leave this mode by clicking stop run.
Now the machine stays standby until start run is clicked again.
In the auto-mode the peaks in the windows are determined. The setting of these windows must be done
carefully. this is done under: options\edit peak windows, where you can choose the windows you want to
adjust:
1. They should not be too broad, to prevent interference from other components. However, they
can be narrower than the real peak, as the program calculates beginning and end .
2. A few seconds extra must be taken, because of small variations in retention time depending on
variations in the room temperature.
3. You should not determine these details on the data measured in the first cycle after starting the
instrument because the temperature in the first run differs slightly from later runs.
To check the current setting, a sample is taken (preferably a low concentration calibration gas-mixture,
an ambient air mixture is also possible) and this is checked with the temperature program and the
window setting.
To do this, choose view mode\data file and go to options\select view items, and there are the boxes oven
temperature and oven temperature (set point).
Follow the compliance of the measured temperature with the set point: if this is not correct yet, then wait
for another run, until the temperature is stable.
Follow the signal from the detectors on screen::
The exact times depend on: temperature, column type, and flow of the carrier gas. The window timing is
determined for each individual column, carrier gas flow, and temperature program. After adjustment of
your GC, these times do not change very much, so normally the application program should give the
expected peaks in the expected windows.
The oven temperature should be as programmed.
In case of complicated chromatograms, it may be difficult to set a satisfying window for the start and end
of each peak. Adjustment of the integration method may help in some cases.
In data file, under options\edit peak windows there is a button attributes.
There is a default setting for integration, which is applied to each peak window if you do not click the
attribute button in the listing. Start and end of peaks are determined by calculating and evaluating the first
and second derivatives of the detector signal.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 49
The area under the peak, with a start and end within the given time setting of the peak window is
calculated. Typical area values for small peaks just above the instrument noise are under 500.
An advanced integration option is the use of retention time locking. By locking the chromatogram on an
always identified peak, like Butane for C2C6 or toluene for C6C10, all shifts due to temperature or flow
differences will be corrected. See 5.9 for more information.
For special cases the setting can be adjusted for each individual signal. Typical problems are:
incomplete separation leading to ‘shoulders’, or even to two maximums in one peak, and baseline
changes in a peak window. Settings are:
1. Integration result is 2x first half of peak area: this helps quite well if the peak to be determined is
only disturbed by a later peak.
2. Integrate with [forced] horizontal baseline from start of peak: in this case a strict horizontal
baseline is integrated, independent of possible drift.
3. Include possible shoulders: this is useful in the presence of small peaks under the main peak.
4. Integrate down to detector values = 0 [virtual baseline]: this can be used if there is an peak in
front of the peak of interest and the detector signal had not yet gone down to baseline before the
start of the peak
5. Integrate highest peak in window: this prevents smaller peaks nearer to the center of the window
to be mistaken for the peak of interest. If this is not active, then the peak with the lowest hit rate,
i.e. the peak nearest to the center of the window, is integrated.
6. Integrate the complete window: this causes an integration of the whole area>PID=0 to be
integrated from start to end of the window.
Integrate with [forced] horizontal baseline from end of peak: in this case a strict horizontal baseline is
integrated, independent of possible drift.
1 2 4
The attributes may help to integrate as well as possible. Of course, a better way is to change the
chromatogram by changing flow, temperature or perhaps even the column type. Do not hesitate to ask
Synspec or your distributor in case of separation problems.
Changes in the attributes have to be made before a calibration is started. As the described problems
mostly only occur with the sample air and not with the calibration sample, decision to use one of the
alternative attributes has to be made before a calibration is made.
In a normal series of measurements selected peaks are integrated and after a calibration the
concentrations are calculated.
These values are saved in the file rundata.txt. To select the desired data format click view mode, program
edit, options, run data output text format. In the window that appears choose between the saving of either
or both of raw data and of measured retention times together with the concentration values. Also choose
the desired delimiter between the data.
On the following page you see an overview of the data registration of the measured chromatograms:
there are three kinds of saved files: the normal measurements of sample air, the measurements of
calibrating gas and the measurement of validation points: they are saved as respectively m-, c- and v-
files. They are bin-files, binary in type. (See figure 4.9 for an overview of the data structure)
For each kind, a directory is made, with monthly directories and daily subdirectories.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 51
In the chromatograms all information is saved, and they can be viewed the same way as actual run. Even
reintegration of windows is possible.
D_yymm
D_yymmdd
M_ddhhss
M_ddhhss Datafiles
M_ddhhss
1 day-datafiles
D_yymmdd
M_ddhhss Datafiles
1 day-datafiles
1month-datafiles
CALIBDAT
C_yymmdd
C_ddhhss
Calibration datafiles
(measurements)
C_ddhhss
1 complete calibration
C_yymmdd
C_ddhhss
Calibration datafiles
(measurements)
C_ddhhss
1 complete calibration
VALIDAT
V_ddhhss
Calibration datafiles
(measurements)
V_ddhhss
series of daily validations
rundata.txt
format:
DateTime ; IntVal-1 ;Conc-1 ;IntVal-2 ;Conc-2 ; IntVal-3 ;Conc-3 ; Comment
09-07-9519:53 ;46232 ;7.9 ;30302 ;9.9 ; 512031 ; 3.0 ;This is a testrun
G
C
5
8 (o
)5
rt C
G d
i5
8 C
G
yre
o
tc d
i5
8
c
u
rts yre
o
tcu c
u
rts e
ru
t
D m
y_ D
y_m d
h
M
_m sh e
lysa
-d
i1
D
ft e
lsifa
td
C T
L
ID
B
A D
y_m d
h
M
_m sh e
o
lyn
sa
-d
m
i1
D
ft e
lsifa
td
-h
C
y_m C
d
h
m
_ s a
d
e
o
n
ltic
p
m
r(b
C
1
u
s o
n
e
iltsa
rfb
)
T
L
ID
A
V C
h
d
_ s a
d
e
o
n
ltic
p
m
r(b
C
1
u
s o
n
e
iltsa
rfb
)
At present the hard disk has >6.000 MB space, of which >5000 MB can be used to save data. Regular
data transfer is advised. This can be done by copying the rundata.txt, to obtain the measured values.
WARNING: we are not able to guarantee the life span of your hard disk! If you do not copy your data in a
regular cycle, loss of data is your own risk! Of course the hard disk itself is replaced under warranty!
To transfer chromatograms, copy the bin-files unto diskettes, a ZIP drive or a recordable CD.
These can be connected via the USB connectors.
The easiest way to do this, is using Windows Explorer. Data can also be transferred by remote control
(see 4.7).
After transferring the data you should do probability checks on the integrated values and the
concentration values.
It is useful to look at the chromatograms to check the following:
• unexpected extra peaks
• unstable baseline
• shifting retention times
• changes in the set parameters for temperature, pressure and flow
This can be done at another PC after loading the GC demo-version. It can also be done in the animation
mode. If there have been changes that made the measured data unusable, by reprocessing the original
data files often a better result can be obtained. This is done under reprocess data files (4.5.5).
After checking a data series the chromatograms should be deleted from the GC’s hard disk, again using
Explorer. The rundata.txt can be left as it is, and will just grow longer. If you wish to start a new
rundata.txt, then rename or delete the present one (after saving the data on a diskette).
4.5.3 Animation
A time series of chromatograms, played as an animation is one of the simplest ways to make the above
checks. A series of BIN-files is displayed as a film, making it possible to view a week of data in just a few
minutes.
Copy the BIN files you want to animate to the directory C:\GC955\ANIMATE. Be sure that no other BIN
files are in this directory.
Choose the desired attenuation of the y-axis and the scale of the x-axis before starting. These are then
fixed during the animation.
Click on view mode\data file\ animate data\ data selection. Now, the data have been loaded for the film.
This takes time, as each file is loaded and processed. When ready, the animation can be viewed by
selecting view mode, data file, animate data, Animate them .
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 53
The animation appears and starts when you press the button start. Speed can be adjusted with the
speed bar. To stop the film press stop.
Leave animation by pressing view mode\data file\animate data\stop animation.
An animation can be saved under a separate filename, and therefore it can be copied. The extension
should be .ADF.
4.5.4 Graphs
For simple evaluation of the measured concentrations the graph mode can be used. After clicking view
mode\graph a simple graph of the measured concentrations against time appears on screen. This
should only be used as a quick check, as it does not have the scope of normal data validation programs.
This makes it possible to reinterpret (reprocess) the data, without changing the original data files.
For instance: a peak arises in the chromatogram, which could only be assigned after the measuring
period had finished. Reprocessing makes it possible to give accurate information on this newly identified
peak, even though it is not appearing in the original RUNDATA.TXT.
To do a reprocessing, the data files must be transferred to a PC.
WARNING: do not use any name in the path of all main- and subdirectories that has more than 8
characters.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 54
In the PC-version of the GC955 program, the Help file contains an extended explanation of reprocessing.
The names and path of the data files, which are to be reprocessed, are placed in a text file.
The program starts loading the first data file and re-integrates this file with the current settings.
The data files themselves remain unchanged.
The following list is an overview of settings, which are important in reprocessing:
• Peak window settings
• Integration attributes
• Calibration
A simulation of a calibration can be done on the PC.
The results of the reprocessing are placed in an output file. The format of this file is like RUNDATA.TXT’s.
As an option, there is a possibility also to generate a report file for each chromatogram. In this file all
found peaks are described, with areas, labels (if known) retention times etc.
These report files can be used in other statistical software for peak identification and pattern recognition.
As soon as a valid input filename is entered the data processing can be started by pressing the start
button. The status line indicates the total number of files to be processed and the number of the actual
processed file, giving an idea on how long the processing will take.
It is possible to stop the processing by pressing the stop button.
All output files are placed in the directory GRAPHDAT, a subdirectory of the directory where GC955.EXE
is located.
4.6 IO-
IO-state
Warning:
Warning the IO diagram is designed for the standard instruments with one valve and a PID/FID detector
combination. If you have a special instrument, we will not incorporate the details of this in the diagram.
So, you may miss important things, like your TCD, or your second valve. For these instruments, we refer
to the manual for the diagram.
The remote control of the instrument works with the program Symantec (Norton) PC Anywhere. If you
want to use this, the following components are needed:
• The program PC-Anywhere for Windows 98 (version 8.0 or higher)
• A modem (or, if you make a direct connection, a null-modem cable.)
• Follow the instructions in the manual of PC_Anywhere.
Alternatively, the connection can be made via the ether net connector, over a LAN network.
For use in environments where an alarm signal is desired whenever concentrations of toxic compounds
rise too high, an alarm function is added. It is possible to attach a buzzer or a lamp to an output port at
the backside. For low alarm, high alarm, and status alarm, the signals are all different.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 56
The compounds of interest should be calibrated at first. Then a low and a high alarm level can be set
under program edit\settings\alarm setting. (As an alternative area levels can be set, but this is less
exact.) If concentrations exceed the set levels the word alarm appears on the screen. In alarm status
under actual run\options, the cause for the alarm can be found. By clicking on accept alarm, you show
that the alarm has been seen: the buzzer or light is switched off. The word onscreen disappears as soon
as safe values have been measured again.
By clicking reset, the alarm function is reactivated.
If there is malfunctioning of the apparatus, the words “system alarm” will appear on the screen. Most
times the instrument will continue to function.
The alarm will also be noted in the log-file.
4.9 Password
The GC955 can be secured against influence from outsiders by the use of a password. Under actual run
the password can be activated, for this you must click on options\password protection. The password
can be changed in program edit mode. (See chapter 5).
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 57
5.6 Communication______________________________________________________ 68
5.6.1 Analogue outputs ________________________________________________________________
_____________________________________________________________________
_____________________________________ 68
5.6.2 Extended 16 channel RS232 analogue output multiplexer _____________________________________
_____________________________________ 69
5.6.3 Digital outputs ________________________________________________________________
_______________________________________________________________________
_______________________________________ 70
5.6.4 RS232 Communication settings _________________________________________________________
_________________________________________________________ 70
5.6.5 Analogue inputs ________________________________________________________________
_____________________________________________________________________
_____________________________________ 72
5.0 Introduction
There are a lot of options to change the instrument’s program. Data handling and communication can be
set to adapt to all sorts of protocols. All this is covered in chapter 5. For normal use these parameters are
set for the customer and do not have to be changed.
The beginner is advised to read through the parts 5.1 to 5.5. The setting of communication, alarms and
auto-validation is more advanced work. (Section 5.6 to 5.9)
By clicking the program edit menu under view mode you get access to a whole range of ways to change
the program settings.
The easy way to do this, is to load a program file: click file, load program file, select a program by
double-clicking.
The program is loaded in the program edit menu, where you can edit it if you wish. To activate a program
change for the next actual run, click stop run, and click start run after the end of the run.
Before making your own programs you should have a good understanding of how gas chromatography
works. If you are still unsure how to effect certain changes in the measurements, please contact your
dealer or Syntech Spectras and we shall send you the desired program, or give you instructions how to
make minor adjustments.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 59
Changing the program yourself is very easy once you understand something about chromatography. As
long as the programs are saved under different names, nothing can go wrong: returning to the old
program is always possible.
The central part of the program is represented by the seven bars on the program edit screen. (See figure
5.1) Together they build the most important functions of the program: cycle time, injection time, pre
concentration heating, purge time, sample pump, sample strokes, oven temperature.
To change bars, click on them, and then move the pointers with the mouse. The cycle time determines
which elements are used: if they are outside of the cycle time bar, they will not be in use, as the next
cycle starts before they are reached.
Any change as described in the sections 5.2-
5.2-5.5 is not activated, unless the program is stopped and then
restarted.
The time needed to separate the compounds in a sample depends on their boiling points, their
absorption on the column material, the temperature and the carrier gas flow. The injection time is set as
the time, which the slowest compound you want to measure needs to pass through the stripper column.
The cycle time is the time this compound needs to pass through the stripper and the analysis column
plus time needed to bring the system back to the state needed for the start of the next run.
Without temperature programming, the time a compound needs to pass through stripper column and
analysis column is relative to the length of the two columns. So if you make the cycle time longer, you
must also lengthen the injection time by the same factor.
When you use temperature programming, these time settings cannot be calculated, but are made
experimentally, as described below.
To try out the settings for new compounds, use a long cycle time and a relatively long injection time for
instance:
For a 2 M stripper, and a 13 M analysis column: start with the injection time ¼ of the total cycle time.
Make sure all compounds you want to measure are out of the analysis column before the end of the
cycle time with an isothermal program, and before cooling down with a temperature programmed
instrument.
Adjust the cycle time to whole minutes minus 15 seconds; this makes the planning of continuous
measurements easier.
(See for the planning of automatic measurements also 5.5, run settings) Now you can shorten the
injection time by steps of 5 seconds until a peak of interest disappears. Then set the time 5 seconds
longer again.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 60
The gas flow is one of the factors that determine the separation and the cycle time. By setting the flow
higher, the cycle becomes shorter, but the separation is decreased.
Only change the flow when making separations of new mixtures. This is done by going to actual run
\options \change column flow. Here you can type the column flow in ml/min. The flow is regulated by a
mass flow controller, and will change within a few seconds. See 7.3.3 for flow programming.
Long sampling times are useful if a mean concentration has to be calculated over longer time spans, e.g.
1-hour means. For this we deliver many instruments with cycle times from 10 minutes up to 45 minutes
The pre concentration then is charged during many sample strokes. This is not advisable with unstable
compounds: they can react with oxygen, water or other compounds, and the adsorbing medium may
even act like a catalyst.
A solution can be, to use a system with cooled pre concentration: a system of the series 800. (See 5.3.4
and chapter 7). The other option is to take only 1 or 2 sample strokes at the start of a cycle and to inject
into the column immediately after this.
It is advised to purge especially with these unstable compounds: this removes water and oxygen, and so
prevents reaction with these during desorption. However, it does take time, during which desorption and
reaction may occur.
In the program, there must be at least 20 seconds between purging and injecting. (Purging influences the
pressure conditions in the system, which must have time to readjust before injection.)
If the sample amount is small, for instance when Tedlar bags are analyzed, the internal pump is not used
in the program. It is advised to test, how many sample strokes are needed with the system used, to
reach a sufficient sensitivity.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 61
An example of a program:
Figure 5.2: 2 sample strokes to preconcentrate, and then a short purging, before injection. Used for
unstable compounds.
The temperature programmed oven allows you to make quicker analysis with less peak broadening of
compounds with greater boiling point differences.
To change the temperature programming, click oven temperature under program edit.
The temperature program appears as a graph of temperature against time. To change the settings, click
on a corner and drag this in the desired direction. The base temperature, however, must be changed as
described under isothermal ovens. The following limitations must be observed:
• Start temperature and end temperature must be the same.
• Heating velocity can be set from 0 o C to 20 o C per minute. You cannot drag a slope to heat
more than 20 o C per minute.
• Cooling down is done by forced ventilation and depends on the room temperature. After cooling
down the oven is closed again and after 20 sec. the heating up to oven base temperature is
made.
• The last peak of interest must have left the column before cooling down, else this last peak will
be split up.
Before the next cycle is started, the stable temperature must be held for minimally 30 seconds.
With differences of more than 30 o C in the cycle it should be checked that the cooling down part of the
temperature program functions by following at least three cycles. Check the compliance with the real
temperature with the set temperature on the data file under view mode, choose under options the view
items oven temperature set and programmed and check if your program leaves enough time for
temperature stabilisation. If not, use a longer cycle.
Temperature and flow largely determine the components’ retention time. With the new electronic mass
flow controller, you can use a combination of temperature and flow programming in the GC955.
At the moment, the flow programming is not made in the window program edit, but in the file GC955id.txt.
Enter the following parameters:
Startflow = (in ml/min. multiplied by 10)
Endflow = (in ml/min. multiplied by 10)
Flowrampstart = (in seconds multiplied by 5, so 5 minutes is 5x60x5 = 1500)
Flowrampend = (in seconds multiplied by 5, so 10 minutes is 5x60x10 = 3000)
A good start flow is in the range of 1 to 2.5 ml/min. For the standard mass flow controller we use, the
maximum end flow is 10 ml/min.
The flow starts to increase after flowrampstart and reaches its highest point at flowrampend. It stays at
that level until one minute before the end of the cycle.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 63
On the pre concentration tube the hydrocarbons from the sample are absorbed. With quick thermal
desorption, these are evaporated. The pre set value for heating is 30 seconds, with a start 3-6 seconds
before injection. The tube is normally cleaned in 30 seconds. The heating time can be adjusted by
clicking and dragging the pre concentration heating bar. Do not heat longer than the injection time.
Check after changing this time if the pre concentration is cooled down before the sampling starts. This
can be done in data file, choose pre concentration temperature under options\select view items .
The temperature can also be adjusted under program edit\settings\instrument settings. The standard
temperature of 180 o C is too high for compounds that polymerize or decompose easily. Choose a lower
temperature, like 100 o C. You must, however, heat for a longer period to clean the tube completely.
Fig. 5.3: checking the cooling of the pre concentration before a new sample is taken
With low-boiling and unstable compounds the results can improve with a cooled pre concentration. For
normal pre concentration, the temperature is in the range of 25 to 45 °C , depending on the internal
temperature of the instrument. Of course, this also depends on the room temperature.
The pre concentration system can be cooled by radiation. For this the pre concentration tube is
embedded in an aluminum block. This block is cooled on both sides by Peltier elements. After startup of
the GC, the block needs a few minutes to reach the correct temperature. The minimum that can be
reached by the present version is –5 °C. To obtain this in continuous mode, the room temperature around
the instrument should be around 20 °C, else the minimum will be higher. However, under most
conditions, a temperature of +5 °C will be possible.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 64
To avoid freezing or condensation in the system, drying of most samples will be needed. Synspec
advises the use of a Perma Pure dryer, based on Nafion. (Please be aware of the limitations of this and
other types of drying system, when trying to measure water soluble hydrocarbons: they will tend to go
with the water.)
Adjustment of the pre concentration temperature can be done by changing the value precontemp in the
file GC955id.txt. This value can be set from –5 to +30.
The amount of sample strokes determines the amount of air that is pre concentrated and thus the
sensitivity. Each sample stroke has a volume of 35 ml.
The pre concentration material should not be saturated. If you are in doubt, please ask your distributor or
Syntech Spectras about the capacity for your compounds. You can also check the saturation yourself, by
running a series with a constant concentration of a calibrating gas, adding an extra sample stroke each
time. For the materials Tenax and Carbopack, saturation data are available for many compounds,
however, these data are not complete.
In data files, the sampling information can be visualized by selecting the menu item options\select view
items in data mode and clicking sample.
Warning: a sample must be completed, before you start the next one. Else, the correction for temperature
and sample is incorrect. If only one stroke is used, and this one stroke is not completed, a fatal program
error occurs.
The sensitivity of the instrument can be set by changing the ionization voltage.
This is done under settings\instrument settings.
The ionization voltage can be adjusted from -600 (very insensitive) to -5000 mV (very sensitive).
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 67
In the first case the machine will wait with starting the next measurement until the computer clock reaches
the right minute to synchronize the cycle time. If you use a synchronized cycle of 10 minutes, then you
must set the cycle time just under 10 minutes, for instance 15 seconds shorter. This is done, to allow for
data saving and zero-adjustment before the next cycle begins.
In the second case, the machine will always start immediately after you click on the start run button, and
it will start the next run, after a run has ended.
The instrument can be set for automatic restart after power up. If this button is not activated, the program
is loaded up to the point where the button start run has to be clicked.
Under settings\change instrument panel, you can change the look of the area below the chromatograms
on the LCD. The screen is preset for your instrument. Do not activate this item, unless you exchange
detectors, or exchange the trap for a loop or a cooled trap.
By clicking on the symbol, the instrument panel screen opens up:
You now can change the elements shown on the bottom line, their units, the names or abbreviations
shown.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 68
Fig. 5.9 instrument panel configuration for the setting of the indications on the lower part of the LCD
As long as the screen is active, you can also reorganize the placement of the elements.
To save and get out, again click on change instrument panel.
If you have messed it up completely, you can get back to a standard setting by clicking on Default.
These settings are saved in the configuration file (.CFG).
5.6 Communication
At the rear of the instrument is the analogue output connector with four pins. These can be attached to
the Phoenix female connector provided with the instrument.
The outputs are 1 - 10V, maximal current 1 mA.
The first three analogue outputs are dedicated to the first three peak windows. If you want other peaks to
appear at the analogue outputs, change the order in edit peak windows, under data file \settings.
The range can be set.
The last pin is dedicated to the detector output signal.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 69
The 4 analogue outputs can only transfer the data from 4 peaks. For systems analyzing many peaks, like
the ozone precursor application and many others, this is only a fraction of the complete information. For
such applications, we advise you to use the RS232 communication protocols, as they are much more
flexible. Alternatively, collect the data by copying the rundata.txt file.
However, sometimes the need exists for more than 4 peaks over analogue output. For this, we provide
an option to multiply the peaks over an RS232 connection.
Warning: for some systems, all RS232 ports are already needed for other functions, like calibration
valves, second GCs etc. In this case the multiplexer cannot be used.
Installing the box is very easy:
1. Connect the multiplexer with a normal (not crossed) rs232cable to the GC to serial port no. n,
where n=1, 2, or 3.
2. Connect the power cable and switch it on.
3. On the GC, open the file GC955_ID.TXT.
4. Add a line <RS232ANAOUT = TRUE>.
5. Add a Line <AOCOMPORT = n> (where n is the no from point 1).
6. Close and save the file.
7. Copy the new software onto the GC’s hard disk (back up the previous GC955_1.EXE). It might be
necessary to delete the file ACT_LIN.DAT.
8. restart the GC program.
9. The first 14 windows of the peak window editor will be visible on the multiplexer. The output scale
is the same as in the GC, under Program\options\ analogue output.
Only calibrated component windows will be seen, non-calibrated component windows will be set at zero.
The original 4 analogue outputs on the GC are still active and the window as described in 5.6.1 will not
change.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 70
The connection for digital signals is also at the rear of the GC. It is a 16-pin connector with 7 digital TTL-
outputs. They are activated over the digital output dialog under settings in program edit
The buttons c1 up to c4 can be set as combinations of two sensors or functions, to make even more
data available over the digital outputs.
Fig. 5.12: Defining combinations of readings of sensors and settings to combinated digital outputs
Over the RS232-connection, through a modem, communication with another PC is possible, by using
PC-Anywhere.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 71
Using specially designed communication programs, communication with central data logging systems is
also possible. As these systems are not at all standardised, adaptations will often be necessary.
However, the basic programming is already included. Please ask Synspec or your distributor for
adaptations that will work on your system.
A choice between several optional settings can be made under communications in settings under
program edit:
Optionally, the German Hessen-Bayern and the Gesytec protocol can be activated. The ASCI-terminal is
more common . The RS232-settings must be deactivated before PC-Anywhere can work.
If changes are made, the program should be saved, then it should be activated by starting a new run,
and then the computer must be restarted, as the settings are activated as soon as the computer starts.
Over the RS-232 status protocols similar to those of the digital outputs can be set:
Fig. 5.14: Example of an error status protocol, for the Hessen-Bayern protocol system.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 72
The GC’s PC can, of course, also be used as a data logger for other instruments. As this is not used
regularly, it is disabled by default. If you are interested in using a running GC to function as a data logger
for analogue signals from other equipment, please contact Synspec or your local distributor.
The alarm function must be programmed under program edit\settings\alarm settings. Here, low and high
threshold values can be set for up to 10 compounds. For each channel the values must be set
individually. Before this is effective, a calibration must be done. Optionally, area values can be entered,
but this is not advised. The system alarm has its own settings.
Once data have been entered in this window, and the “active” boxes are checked, the system will give
alarm warnings on screen, optionally over a buzzer or warning lamp, when the value is higher than the
threshold. The alarm is also automatically noted in the GC955-logfile. If the low and/or high component
alarms are activated, the alarm is also set in the digital output window.
The default password is GC955 (use capitals). To change this password, go to program
edit\settings\change password. Now, the present password must be entered, to get access to the
change option. The new password must be entered and confirmed, after that, it is active.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 73
Warning:
Warning in section 5.8, several options for automatic validation are explained. These all include the
connection of a calibration gas source to the GC. Several different switching systems are introduced.
The following two remarks concerning the gas source must be observed:
1. In all cases the calibration gas must be at ambient pressure: from 900 to 1050 millibar. If the
pressure is outside this range, a proper correction of pressure cannot be calculated. If the pressure
is above 1.5 bar, the GC may be damaged.
2. The flow of the calibration gas from an undiluted bottle must be restricted to a flow of cir. 25 ml/min.
Synspec can provide you with a simple critical orifice. Synspec takes no responsibility for bottles that
are emptied due to a lack of restrictors.
A periodical automatic validation of the calibration can be set with the menu option automatic validation.
When this is switched on, the system will not save the new values as measured samples, but save them
in the validation directory. In the rundata.txt they are prefixed with v.
The cycle can be set from 23 up to 168 hours (7 days).
This feature is very flexible in design. The system returns to the set frequency per week if you enter a
preferred day. The system indicates the time of the next validation. Up to 5 runs can be set, for zero air or
span gas. However, the system does not switch on the validation gas.
There is an option to attach a switching valve to the GC, in this case the providing of the gas can also be
directed from the GC.
Under edit concentrations you can enter in the concentration of the span gas. (see figure 5.13)
5.13)
Under edit conditions the following is possible:
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 74
1. Toggle auto-calibration if the deviation is greater than set range: this means that your old calibration
is deactivated, so do this only with span gas of exactly known concentration!
2. If possible, skip the first values: this gives a greater accuracy, but needs a longer calibration cycle.
3. Use relative factors to calibrate missing compounds: it is possible to calibrate other compounds with
known relative sensitivity for one compound, for instance nonane and decane relative to octane.
4. Alarm function: for the span gas a deviation in concentration is given, for zero gas an area value
greater than a set value can be chosen.
The Syntech Spectras GC can be provided with a 6-port Vici electrical valve for switching between the
sample and a calibration gas. See chapter 6, for the connections that have to be made to the gas system
and to the green digital connector. In this section, the programming is treated. In the auto-validation
menu set the following:
1. use only span gas, no zero gas.
2. set the timing and the amount of cycles according to the instructions in 5.8.0.
3. in edit conditions activate disable sample pump if span gas active.
4. in edit concentrations set the concentrations and activate use.
The valve will now be switched by the GC, provided the digital setting is ok.
5.8.3 Setting the digital outputs for the automatic validation with a 6-port Vici-valve.
To set the digital outputs correctly select the following settings in the menu digital outputs:
under program edit\settings, select digital outputs.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 76
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 5 6
Auto-validation is normally used to check the validity of an existing calibration. However, it can also be
used to calibrate. In the item edit concentrations the actual concentrations of all components that must
be calibrated must be added and in edit conditions the conditions for the switching of calibration valves
from the GC etc.
Then add a line in the file GC955id.txt:
AUTOVALMENU = TRUE.
After restart of the program GC955 an extra item is added in actual run\options: activate auto-validation.
If you click on this a calibration will be performed automatically. An alarm will appear if the values deviate
more than the set condition. After a power failure an automatic validation will start up also.
For some applications a combination of two 3/2 magnetic valves can be used as an alternative to the
high quality Vici 6-port valve.
We provide this as an extra option. While the inertness of the Vici system is guaranteed, this is not the
case for the 3/2 magnetic valves. The customer should test for himself whether his calibration gas does
not deteriorate when lead over this system.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 77
The system can also be mounted afterwards by the supplier or customer, with a special installation
leaflet.
The system will be connected and activated by the digital input pins. The valve’s settings will already be
prepared by Synspec.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 78
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 5 6
5.8.6 Autolinearisation
concentration
The instrument has an option to test the linearity without using more than one concentration of calibration
gas. This is only possible for instruments with a pre concentration tube.
The mode: autolinearisation is loaded from view mode \autolinearisation. Click options and then start
autolinearisation. The instrument now pops up the request to connect the calibration gas. Choose a
stable concentration, and leave it on the instrument for 7 hours.
The instrument will now run 12 cycles, with 9 to 1 complete strokes and 3 partial strokes of the sampling
piston. Set the instrument to actual run and wait until the end of the autolinearisation before doing other
things with the computer: the calculation uses most of the processor capacity.
The second order regression is calculated and shown afterwards in the mode autolinearisation. The
parameters calculated can be used to improve the linearity of the instrument.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 79
If the instrument is still non-linear, two actions should be undertaken: first of all, tubing and rotor must be
cleaned. Also, the detector sensitivity should be increased. This can be done by cleaning the lamp and
the detector. If this does not improve the detector signal, a new lamp may be needed.
If you do not want to use the calculated parameter in your data processing, it is easy to return to the
original setting: close the program GC955, delete the file autolinS3.dat in the directory GC955 and restart
the program. Now the original program is in use again.
5.8.7 Calibrate for all components with a single component calibration mixture
This procedure describes the use of one gas to calibrate several others. This option can well be used for
C8-aromates. The error will be <2% as the sensitivity is identical. Also it is possible to do it for the C6-C8
aromates with only benzene. Here, the error will be in the same range. In any case, these errors are
systematic and will not change with time.
Please note: this method can only be used for calibrating hydrocarbons that have physically similar
properties in their behaviour in the detector and have boiling points that differ less than 100 °C. You
should be certain of the peak identification.
Procedure:
Have the analyser running for at least one hour, and then run 2-3 times with the calibration mixture. The
chromatogram will look like this:
Fig. 5.26: Chromatogram with calibration gas and correct setting of windows
You must know the retention times for all components you want to calibrate, so all is integrated in the
ambient air.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 80
Give all C8 peaks the concentration of para-xylene and save the file.
Do this for all peaks you want to calibrate.
If you do this when the system is still running, the calibration for benzene, toluene and p-xylene will be ok,
but for o-xylene and ethylbenzene the result will be wrong.
Fig. 5.28 Original peak window setting for the xylene peaks
Now the tricky part: the peak window of first ethylbenzene and then o-xylene are shifted to the place of
the p-xylene window,and the system then is calibrated in this position.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 81
You see that now, the window for ethylbenzene is on the place of the p-xylene window.
Go to the window calibration\options\data selection calibration files.
Now when you select ethylbenzene, you will see some points made during the running calibration,
without ethylbenzene in it. This means that the area was low. These points must first be removed by
clicking on remove file until all points have disappeared.
Now you must reload them in the setting with ethylbenzene over the para-xylene. Load them anew by
clicking on the files in add file.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 82
After this, set the windows in the original position and start a new run.
5.9.1 Introduction
Retention time locking is a method to keep peak windows centered over the peak, even if the retention
times are shifting.
This is done by selecting a peak, which can always be identified (this limits the applicability). This peak
will be the reference peak.
When a chromatogram is recorded, the retention time of the reference peak is compared with the
retention time of the reference peak in the original chromatogram where the peak windows were set (the
reference chromatogram).
All peak widonws (including the reference peak) are now recalculated.
With these new calculated peak window settings, the chromatogram is integrated.
To show the user that retention time lock is active, there will be a message in the screen.
Also, the file data will be flagged accordingly.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 83
First of all, load (or record) a file containing all relevant peaks in the view menu data file. Center the peak
windows exactly to the middle of the peaks. Make sure that the window over the peak you want to make
the reference peak is broad (say 30 seconds).
The windows over the other peaks can be smaller than you would normally set them.
In data file select the dialog retention time lock.
Press set to make the actually loaded chromatogram the reference chromatogram.
Select the reference peak in the list.
Press lock now, to lock on the reference peak.
With retention time lock active you can switch between active and not active.
If you want to recheck the retention time lock file, click load, then it will be loaded in the editor.
Every instrument has its own identification file, named GC955.txt. In this file specific settings for the
applications ordered by the customer are activated. They influence the data communication protocol
used, the alarm settings, the temperature regulations, the integration method, etc.
In the table in the appendix you will find an overview of the options.
Notice:
Normally these settings are not to be changed by the customer without consulting with the supplier or with
Synspec, even if you have two Syntech Spectras GC955s with different settings.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 84
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 85
6.0 Introduction
The service and maintenance chapter is intended for trained service personnel. Do not start working
before reading the whole text and contact Synspec or your distributor if you are unsure about the exact
way to proceed.
If you
you must open the instrument, always shut off the gas supply and the electrical power before starting.
Take care of closing the gas supplies, especially the hydrogen supply. After repair, always check the
hydrogen connection for leaks!!
If you have to work in the oven compartment, be sure to discharge the high voltage on the lamp. (See
section 6.1.2) If you have to test pneumatics, ventilation, or the PID, only pull out oven’s own power
supply. (See illustration 6.5)
Troubleshooting and maintenance is a task, which should be shared between distributor and client depending on
their experience, budget, and the required speed of repairs.
Where only the distributor should work, this is marked as ‘D’. Where distributor and client can share the work or
cooperate, this is indicated as ‘C’. In some cases the client can identify the problem, but not make a repair, there
it is indicated as ‘C/D’.
The following points will be covered in the troubleshooting list:
1. Instrument will not start up
2. Leak in gas supply
3. Pressure alarm
4. Flow alarm
5. Retention times are unstable
6. Peaks have bad shape
7. Integration is impossible
8. No peaks in data file
9. Calibration is not linear
10. Big change in sensitivity
11. Program stops at end of cycle
12. Data cannot be copied on floppy disk
13. Bad reproducibility
14. Measuring is not continuous
15. Baseline unstable
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 87
instrument will C Loose connectors Check if the connections in the computer compartment are in place:
not start up they can get loose during transport. Beware of electrostatic shocks to
the system!
C/D blown fuse replace fuse at entrance by spare one, check at next start-up what
happens: there may be an electrical problem in the instrument.
C/D failure of computer check if the connectors are in place, check if the processor ventilator is
C Wrong comport Check in the file GC955id.txt the setting for the MX port. For an internal
connector it must be Mccomport 4. For an external comport it can be 1,
2 or 3. Check the comport, and fill in the number of the port that is used
in the file GC955id.txt. Restart the GC program
C/D hard disk failure check if the pc can make contact to the hard disk at start up: if not
contact your supplier about a new hard disk
C/D bad contact in see below
detector
C/D fault detected at check the error messages, correct for them, shut down the GC955.exe
C Instrument is very Two problems: water condensing on electronics and leading to general
cold errors, and water condensing on lamp and leading to PID lamp fault
message. Both problems will disappear if the instrument is heated to the
temperature of the environment before start.
C Error message oven Instrument is very cold and the heating of the oven or the pre
temperature or pre concentration uses more energy than the safety settings of the
instrument allow. Please raise the environmental temperature or set the
concentration
program to a start temp for the oven of 45oC and for the pre
temperature at start
concentration of 180 °C
up
C Error on FID Check hydrogen and air supply and the pressures of these gases: after
a shut down the hydrogen line contains only air, so it will take time to
reach the analyser. Check also if the hydrogen line contains droplets of
water (wet scrubbers), this will blow out the flame when it ignites
C Error on FID A cold FID lights slowly, have patience and try 3 times
C Error on FID To start up the system it is often useful to set the hydrogen pressure a
bit higher.
leak in gas D Bad connection on The connections of the carrier gas system come from the carrier gas
supply carrier gas system connection at the rear, lead through the left compartment to the
pressure and flow controllers with the carrier gas filters and then to the
oven compartment. Small leaks are often found in the oven
compartment, big leaks in the left compartment. Check of the oven
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 88
D leaking tubing The steel tubing is quite strong. The FEP-tubing we use on other parts
can only be damaged by bad handling. Replace tubing by identical
pieces, contact Synspec.
D leak in valve Big leaks can occur in the left compartment or on the 10-port valve
switching parts panel if the tubing has come out of the quick insert fittings. Take out the
tube completely by pressing down the small black or blue ring. Then
push in completely. If this does not help, then the fitting will have to be
replaced.
A leak on the 10-port valve panel needs a trained person for repair, as
the panel holds 10-port valve and detector, and these are essential
parts. First take off the bottom cover. Loosen the valve and the detector,
loosen the panel by unscrewing it and take it out. Check the
connections to the four valves and also to the pressure sensor. If the
pressure sensor is leaking put in a new one.
D leak in carrier gas in leaks on the pressure or flow regulator or on the carrier gas filters are
left compartment also best detected with leak test fluid. The connections on the pressure
regulator are not so easily reached, take you time to inspect and tighten
them. The straight and bended tubing unions in this compartment
should always be tightened by using two wrenches: be sure to hold one
part still while turning the other, in order not to bend the tube instead of
tightening the connection.
switching air too low Check the dimension and length of the supply tubes.
Check for leaks.
or too high
D Pressure sensor or Each sensor is calibrated individually. Do not exchange GC955id.txt.
malfunctioning
flow alarm C pressure of carrier increase pressure to about 3.5 to 4.5 bar.
flow controller system or swing. If it swings, you must increase the carrier gas pressure.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 89
system
retention times C/D problem in mass see problem: flow alarm, cause: problem in mass flow controller.
low pressure on instability: for 0.53 mm columns it must be at least 2 bars, for 0.32
columns at least 2.2 bars, getting higher with longer columns.
carrier gas
C room temperature See to it that the change stays below this, or at least place the
has a variation of instrument away from the airflow of air conditioners, especially the left
side compartment should be protected!
>5 °C per hour
D bad or difficult type Water sensitive and packed columns are more sensitive than the
of column standard silicone column. Either choose another column, or see to it,
that carrier gas always stays on the column, and that the sample is dry.
Contact Synspec for advise on optimising.
D defective Check the temperature if you are in doubt with a different temperature
thermocouple in the sensor and compare with the reading on the GC955-screen.
oven
PID peaks have C/D bad connection in check the blue and yellow connector on the PID for the electrical
bad shape the PID contact. Take off the top of the PID and check inner and outer side of
the electrode for the connection. Tighten all screws.
integration is C long cycle time change minimum peak height, see ch.6, as the system is seeing more
C two peaks are change setting of integration: choose integrate 2 x first half of peak for
C two peaks are choose the setting: start integration at PID-value is 0 for the second
coeluting peak
coeluting
C peak is very broad change setting for minimum slope change ( see ch 6)
C/D peaks are broad heat out the column at 25 °C above the normal temperature
because column is
malfunctioning
C Peak is split up into Check if the peak appears at the moment the oven shutter is opening: in
many small peaks this case the peak is split up in the column.
no peaks in C setting of peak set peak windows on a chromatogram, stop run and start run after end
data file windows has not of cycle. Save the settings also by saving the program under its own
name (like btx5.prg ) in the program edit mode. Do this any time you
been saved
have changed the window settings.
calibration is C use of wrong Use the GC955id.txt and the autolin.dat which belong to the instrument.
not linear settings Linearity of calibration depends on the type of detector cell used: you
should never copy the GC955id.txt from another version.
D New After major changes in the equipment (new lamp, column, valve or
linear range ppm. If you get a wider range and want to measure over all of it, use
non-linear multipoint calibration
C/D absorption effects clean out or change external tubing. If this does not help, clean out
in internal or internal tubing and 10-port valve with diluted ammonia and distilled
water. Dry well and reinstall.
external tubing
D reaction effects in reactive hydrocarbons may react with and on metal and polymer
internal or external surfaces. With which of these they react, depends on the compound.
Contact you supplier or Synspec if you have problems with linearity in
tubing
calibration
big change in C/D lamp failure if a change of more then 20% within 24 hours occurs, it is possible that
PID
C instable calibration check the stability by keeping the bottle at stable temperature. Flush the
pressure reducer,
leading to gradual
decline
C no dust filter on fine dust will coat the lamp window. Clean the lamp by wiping it with
program stops C cycle time too near set for at least 10 seconds between end of cycle and synchronising.
time
C bad contact in see below
detector
C synchronising time check the synchronising time set under program edit\settings\run
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 91
incorrect settings: if you run a 5 min. cycle synchronised to 30 min. etc. the
program will wait after each cycle for the correct time. Stop the program
at the end of the run. Set the timing at the desired frequency, save the
program under its own name and restart.
bad C Leak in sampling Check for this by putting a cap at the sample entrance: the piston
reproducibility system strokes should now not be completed. If the piston moves smoothly
despite closing the sample entrance, you have a leak. Contact the
distributor
D Leak in 10-port Check the connections and the column
valve or column
system
C Dirty sampling lines In this case the concentrations at calibration will rise for a long time.
Change tubing.
computer leading to for appearance of sudden stop. Check hard disk, computer board,
processor
missed
chromatograms
and blue screen
errors
C Error messages in Messages generated by bad communication between electronics of GC
program without and computer board. Contact your distributor for possible update of ICs
on the electronics board.
real reason
C Problem with tasks Windows 98 can be set for multitasking. However, especially network
for Windows 98 connections can interfere a lot. Contact Synspec for improvement of
task settings.
C Problem with writing Check the size of the log file and of data files. Make back-up and
of files remove the original.
Defragment the hard disk after this.
Ethernet Ask your network manager how new files are handled. If each new file is
Below is an overview of the advised frequency for maintenance work on a continuously used Syntech Spectras
GC955 series 600 and 800.
D/ Time in Work Frequency manual needed
C hours at
C 0.5-3 Calibration 1 x every 2 4.3 Calibration gas, automated settings in
weeks to 1 x automatic validation advised to promote
every month quick calibration
C 0.1 Replacement of micro 1x every 2 4.1.3 Teflon filter
dust filter for sample Months
C 0.1 Remove dust from 1x every 2 6.5
ventilator Months
C 1 Clean the lamp 1x every Year 6.1.2 Lamp tissues, lamp cleaning powder
C/ 1 Clean the FID 1 x every year 6 Fine brush
D
C 0.5 Change of carrier gas 1x every Year 6.2.5 2 Filters
filters
D 2 Cleaning of diaphragm 1 x every year By distributor only
C 1 Renew external sample 1x every Year 4.1.3 FEP or Teflon tubing
tubing
C/ 0.5 Replacement of Pre 1 x every Year 6.3.1 Pre concentration tube
D concentration Tenax
tube
D 1 Replacement of cooled 1 x every Year 6.3.1 Cooled trap, resend to distributor for
Pre concentration trap refund
C 1 Optimise hard disk 1 x every year 4
The photo ionisation detector UV lamp, High Voltage Power Supply and amplifier are constructed of the
present state of art devices to assure optimum operation. The overall detector dimensions are 8 cm by 4
cm with 1/16 inch stainless steel inlet and outlet fittings and a reaction chamber volume of 50 microliters.
Electrical connections to the high voltage supply and electrometer amplifier are plug-in.
6.1.1 Installation
The PID detector is mounted in the oven near the 10-way-valve. This reduces condensation within the
detector and provides an overall stable temperature. The maximum operating temperature is about 100°
C. Due to the oven design, a column temperature up to 150° C can be used.
The PID has a cable connection to the High Voltage Supply passing through the oven wall (the red and
black one). On the High voltage supply the inner core (red) must be connected to the positive side of the
supply, which has been marked. Also, the blue and yellow signal cable can be seen, which is plugged
into the amplifier board.
An overview of the connecting cables is seen in figure 6.1 and 6.2, side and upper views of the detector.
Operating
After connecting the power and gas lines the background signal from the PID is automatically checked
and compensated. Due to column bleed and the supplied ionisation voltage a background ionisation will
exist. By reducing the background signal electrically before entering the detector amplifier the output of
this amplifier is almost zero and the whole dynamic range can be used for PID detector signal
amplification.
6.1.2 Maintenance
Once a year with a capillary column or once every month with packed column the lamp window should
be cleaned with a light-source cleaning material.
Once every two years all other components should be cleaned as well.
Switch off the power of both the PID electronic unit and the gas chromatograph. Discharge the HV
output by shortening the + and - connectors of the HV at the lamp - see figure 6.2 and 6.3. Disconnect
gas inlet and outlet lines of the PID detector, remove the high voltage supply cable and signal /ionisation
cable.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 94
red
High Voltage pos
black
High Voltage ground
yellow blue
to amplifier board
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 95
Disassembling
1. For the position of the parts, see figures 6.3 and 6.4: views of the detector when assembled and
when disassembled.
2. First remove the Vici-connectors from the detector. Loosen the input connection at the oven wall
separator, not at the detector.
3. Remove the electrodes by pulling them carefully from their connectors. The connection for the
ionisation voltage and the ground of the High Voltage supply is done in the same way.
4. Loose the three long hex-screws from the upper part of the detector and take off the upper part.
5. Loose the three smaller hex-screws from the clamp-ring and take off the three parts of the clamp
ring.
6. The peek detector housing, inner electrode, in and outlet connectors can be cleaned in a mild
ammonia solution using an ultrasonic bath.
7. Rinse several times with hot water, dry by using clean air and put the parts in a 100°C stove for half
an hour.
8. The lamp is sealed to the detector house with an O ring in the outer grove of the lamp. Carefully, take
the O-ring out with a small sharp screwdriver, be careful to point away from the lamp window to
prevent scratching.
9. Clean the window of the lamp with a light-source cleaning material. Use lens tissue to prevent
scratching the surface. In most cases the use of a cleaning material is not necessary, and the stain
on the window may be removed using lens tissue only.
Assembling
1. Take care not to touch the lamp window and inner electrode with your fingers.
2. Fit the spring and the lamp with the O-ring into the aluminium housing and the three clamp ring
parts. The spring must ascertain the contact of the lamp house to the outer ring of the lamp.
Observe the position for the High Voltage-connector in respect to the other parts.
3. Assemble the lamp and lamp house with the three short fixing hex screws. Pay attention that the
distance between the clamp ring and the detector house must be less than 1 mm when pushed
together by hand, if not the gasket for the lamp might be too thick or not correct in place.
4. Mount the detector with the three long hex-screws. Take care that it is mounted with the connectors
in the correct position.
5. Install the inlet and outlet Vici-connectors.
6. Connect the electrodes according to the scheme in figure 6.1 and 6.2.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 96
LAMP HOUSING
HV positive
out situated at this
position
OUT
in
DETECTOR HOUSING
ionisation
voltage
HV neg
conn like
signal
others
out
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 97
Fig. 6.4: PID (right) and FID (left , below 10-port valve) in the detector compartment.
The main maintenance on the FID is the cleaning, 1 to 2 times per year, of accumulated soot. Take apart
the FID. Then clean the chimney with a brush. Clean carefully the other parts on the FID itself with a small
brush. Take care with the thin wire: this material should not be touched. The FID is mounted deep in the
instrument: take time to remove the connections and to remount them.
Procedure:
1. Switch off the system, remove the gas connections. Be careful with the hydrogen.
2. Take out the T on top of the FID.
3. Take out the tubing to the FID.
4. Carefully take the FID out of its socket.
5. Gently move it up and backwards in the same movement and take the chimney out of the hole.
6. Remove the spring and lay it aside.
7. Gently take off the metal part from the ceramic part. To make this a bit easier, warm it up to hand’s
warmth before starting to turn.
8. Clean the chimney with a brush. Remove any contaminations from the rest of the FID with a brush.
9. The solderings on the FID and the little spiraled wire should be left untouched.
10. Remount all parts. Push the tubing as deep as possible in the Teflon sockets.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 98
6.3 Pneumatics
Safety notice: within the following sections procedures are described, where the power has to be
switched on, while the oven is open. To prevent heating to take place, it is advised to loosen the
connector to the oven heating. To do this, take off the bottom plate by loosening the screws. On the
underside two large electronic prints are visible: the central MX board and the power supply. Take off the
connector for the oven power supply - see figure 6.5. Do not forget to remount them before closing the
underside.
Power supply
10-port valve
front rear
Thermo-couple oven
Detector
and preconcentration
The diaphragm valve can get blocked if dust comes in, or if the column bleeds.
Symptoms: at first, sampling only goes slowly and is not complete. Alarms are given, if no sampling stroke is
complete. Once it is completely blocked no peaks are seen, only a flat line. This changes from one run to the
next.
Prevention: always use a sample filter, and use columns with low bleed. If you are uncertain, contact Synspec or
your distributor for advice on filters and columns.
The valve can easily be opened and replaced. However, caution must be taken, and the whole procedure must
be followed exactly.
Maintenance is done by opening the valve, cleaning the top part and mostly replacing the diaphragm itself. You
need the following: ¼” key, 5/15” key, 9/64” hexagonal key (essential part, without this it is impossible), long
Phillips (cross-headed) screwdriver, ultrasonic bath, neutral soap, ammonia, distilled water, small pincer.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 100
Procedure:
1. Switch off and disconnect the system.
2. Place it on a dust free workspace with enough room to place a lot of small parts that you remove from
the GC.
3. Remove the GC column: loosen the nuts and take the column out. Leave the nuts in place.
4. Disconnect the 4 connectors on the detector side of the oven wall
5. Take out the oven wall by removing the two screws fixing it to the oven bottom.
6. When it is out, take out nuts and the special column ferrules.
7. Remove all connectors to the diaphragm valve. This takes experienced people 3 minutes, count on 15
minutes for the first time you do it. The diaphragm has numbers, please put each tube that you take out
on a piece of paper with this number on it. Three tubes are not taken out of the instrument. Take care: the
unions are long, and the tubing coming out of it must not be bended when taking it out, else it will be
difficult to place back. Always start by removing the tubes that you can reach easily, those on the right
side. Carefully remove the two that go to the rear of the GC. Now, on the side to the pre concentration,
the two that go to the pre concentration are the most difficult: carefully loosen them at both ends and
take them out without bending.
8. When all connectors are out, open up the valve by removing the central hex screw with the 9/64” key. If it
is too tight, increase your force by using combination pliers. Remove the screw, and take off the top of
the valve.
9. Inspect the top part and the diaphragm. If the top is dirty between 4 and 5, there is column bleed. If it is
dirty between 1,10,9 and 8, the sampling is the problem. It there are many particles, the sample has to be
filtered better.
10. The top must be cleaned in the ultrasonic bath, until it is clean as a mirror. Afterwards, a last treatment
with distilled water, and then dry it in an oven at 90 ºC.
11. Normally, the diaphragm cannot be cleaned like that, but has to be replaced. Take it off with a pincer.
12. Put on a new one with TOP on top. It fits on the two pins.
13. Carefully put back the dried top part of the diaphragm, with nr. 2 in the direction of the rear wall. Fix it with
the hex screw.
14. Remount all connectors in counter direction. Be careful, not to bend the tubes and to tighten all
connections. Not too tight!
15. Then remount the oven wall and reconnect the tubing on the detector side.
16. Last of all, put the column back in place.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 101
Introduction
Opening a left panel of a GC955 is shown here on an example where a transformer has to be modified from 220
to 110 V settings.
However, for exchanging other elements, like sample piston and mass flow controller, the same procedure is
valid.
Fig. 6.9: Loosen the 4 brass nuts connecting to the tubing on the right side.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 102
Take off the LCD cover by taking out the four little screws.
Fig. 6.11 Then take out the screws indicated on this photo connecting the flow regulation panel to the middle wall.
At the rear panel 3 screws have to be taken out connecting the side to the back panel.
Take out the two attachments on a cooled precon if present (by unscrewing the 2 holders at the left panel
completely)
Then take the 2 screws out of the sample pump so it gets disconnected from the left panel.
Now the left panel can be turned to the side (90 degrees), and the transformer I is freely accessible.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 103
The picture is of a system without a cooler, but will be the same concerning the transformer.
Now resolder the wires on the transformer as indicated on the next photograph.
6.3.3 Pump
The KNF-pump is used to pump the sample air up to the entrance. With inadequate filtering dust may
gather in the unit and the membranes of the pump get blocked.
If the pump does not function, it can be demounted from the left side of the instrument: unscrew the two
bumpers and lift off the pump (See figure 6.8). Open the four screws and clean the square and the two
small round membranes with clean tissue. Mount them again and try to start the pump.
If this does not help, contact your supplier.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 104
The flow of the analysis column and the flow of the stripper column should be checked during
maintenance: they should stay the same between services.
Check them in both configurations of the 10-port valve, take care not to measure at the moments the
oven door is opened and shut, as the flow will then shortly vary.
If the flow decreases sharply, a leak has occurred. This mostly occurs at the fixation points of the
columns.
However, because of vibrations caused by pumps or by transport movements, other unions in the
system may start to leak.
For leaks at the columns see section 6.3.4.
At other places leaks may be sought by using Soppy soap. Retighten the union and check again.
With serious leaks it will also be apparent that the carrier gas consumption has increased. If leak free, the
GC’s Nitrogen pressure should drop only after 30 seconds after shutdown of the Nitrogen supply.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 105
Because of the continuous heating and cooling down of the pre concentration tube loaded with the total
spectrum of ambient air hydrocarbons the pre concentration tube filling will change. Replacement of the
tube after a year is advised.
1. First, shut off gas supply and power, and then remove the bottom cover.
2. New tubes are supplied with their thermocouple attached. Be careful with the connection of the wire
to the tube: it can break off with rough handling, this cannot be repaired.
3. Loosen the two poles of the thermocouple from the MX-print (brown on + , blue on - ). (See figure
6.7)
4. Pull the wire through the valve cabinet, and up to the pre concentration tube. Remove the tie wrap.
To loosen the tube, first unscrew the couplings to the bulkheads.
5. Remove the 1/8” FEP-tubing with the nut and ferrule and carefully push it up to the connectors on
the new pre concentration tube. The end of the FEP-tubing must project about 1 mm, to function as
insulation.
6. Put in the new pre concentration tube and fasten the tubing carefully.
7. Then push the thermocouple wire to the underside, push it through the opening in the valve cabinet
and refasten the poles at the correct configuration.
8. Finally, fasten the thermocouple wire under the one elbow union with a tie wrap.
Maintenance of the cooled pre concentration trap treats the following: how to check if it is working
correctly and how to replace it.
As the trap is expensive due to the elements used for the cooling, we give you the option to exchange the
trap for a new one. We take it apart at Synspec, and reuse all parts except for the inside trap. You cannot
replace the internal trap yourself, as this is a very specialised job.
First of all, look at the bottom line on the screen of a running instrument, see the item precontemp. During
the cooling phase, this temperature should be in the range of the temperature set in the file GC955id.txt.
Then load a data file, look under options\select view items and activate pre concentration temperature.
The red line should look like this at enlargement 25% or 10%: the rise must be very steep, the decline to
cold should be gradual. But, before the sampling starts again, it should be at base temperature again.
Normally you will see the action of the Peltier cooler: it is switched on only every 2 seconds, so you see a
rise and dip.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 106
Fig. 6.15: Heating the cold trap with Peltier cooling afterwards
Errors:
• Trap is not cooling. Usually this means the thermocouple has broken: especially with the –5 to +
300 temperature difference the strain on this soldering is heavy and it sometimes breaks.
However, you must check if the Peltier coolers are warm on the outside. If they are, then they
work. If they’re not warm, they don’t work, and they cause the error.
• Sometimes, the temperature rise is slow and does not go to set-point. This may happen if the
system has been standing idle for a long time, during which it was cooling. Very slowly ice will be
formed inside of the foam isolation. Please contact us, we have a method to try to dry the cooler
out again. Please be aware that this is very much a strain on the thermocouple: it is better to turn
the analyser off instead of having it for a long time on hold.
• Also, compare the sample strokes with an original chromatogram: if the system needs much
longer to complete a stroke, or if the strokes are not finished before the next one starts, the
system is blocked. This may be caused by very fine dust, coming through the external filter, or
because a filter has not been used.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 107
If all functions check out ok and the linearity and reproducibility of the system is ok, there is no reason to
replace the trap.
Order a new one at Synspec giving the serial no. of the instrument and at the same time indicate if you
will be returning the old trap. Fillings for the different applications are not identical, we must produce the
same trap according to our instrument data.
Replacement procedure:
Disconnect all power and gas, place the instrument on a clean surface and open the bottom cover. Place
the instrument on its right side.
The trap is connected at : the left side panel with two mechanical holders, to the sampling system by two
Vici 1/8” connectors on the middle panel, to the transformer by a fixed electrical connection, to the MX
electronics board by a thermocouple, to the sample piston electronics by two connectors.
Easy: take out by hand the connectors on the sample piston electronics board, unscrew the connectors
on the MX board with a no. 0 screwdriver.
The mechanical fixation on the left side panel is not always identical, as the Peltier cooler ventilators
sometimes differ. But, if possible, we use two brass holders. Take out the very small hexagon screws on
the holders and the two screws on the left side panel.
The best way to get room is to take out the two screws fixing the pump on the left side panel. Take care:
on the inside two o-rings are used.
The connection to the transformer is special for this type of trap: two small hexagon screws must be
loosened, then the connector can be taken off the pin. For the top connector this is easily done with a 1.5
mm hexagon wrench. For the one on the bottom, it is best to use an hexagon wrench with a 90º angle.
Now, the last connection is with the Vici nuts. Unscrew these and then carefully take out the unit to the
top. Please return to Synspec for refund of money.
Remounting the system is essentially in the counter direction. Be careful with the tubing on the Vici
connector: it is thin tubing and bends easily, after which it would no longer fit correctly. Also take care
that the electrical isolation in the form of the two pieces of FEP tubing is not displaced, as this would
result in a shortcut of the electrical circuit.
Remounting the transformer again is difficult.
After that, reconnect the thermocouple in the same way as the one for the oven thermocouple.
As last steps remount the pump and put back the special holders of the pre concentration. Set them at
such length, that the trap is vertical, else the lower Vici connector may start to leak.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 108
The 10-port valve has an all-important function for the whole gas system. At the correct actuator pressure
and protected by adequate filters at sample and carrier gas inlet it is guaranteed for 1.000.000 turnings.
When changing tubes or columns, care should be taken to work in a dust free atmosphere, as damage
from dust will most likely occur in the 10-port valve.
The valve is easily blocked by dust or small metal particles. In this case from one run to the next you will
have a straight baseline and no flow during injection.
Opening the 10 port diaphragm valve should only be done by trained technicians.
The procedure is as follows:
1. Remove the column and remove the wall in the oven. For this you have to unscrew the four Vici
connectors to it and two screws at the bottom of the valve compartment.
2. Now you have a good access to the 10 connectors on the 10 port. Unscrew them carefully,
taking care not to bend the tubing.
3. With an hexagon key 9/64” you can easily open the top of the valve. Remove it carefully.
4. Now you see the diaphragm lying on the small pistons, and in your hand you have the polished
surface of the other side. Any dust is normally found on top of the diaphragm or on the polished
surface. Clean with optical tissues.
5. The diaphragm sometimes deteriorates and gets bubbles around the piston ring. In this case it
can be replaced by a new one. Contact Synspec for that.
6. Replace all, now your system should be ok again.
The instrument comes provided with a stripper column and a analysis column. These are precut and
installed. The stripper column is marked with short, the analysis with long sleeves of PTFE-tubing.
The column is made of fused silica and will not break easily.
However, in the bulkheads in the oven wall, breakage may occur for two reasons: heavy shocks to the
instrument, and too strong tightening.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 109
Leaks in these bulkheads are not so easily detected with “Soppy”, because of the low pressure.
However, the decrease in flow can be measured.
In case of leaks caused by a break in the union, the column can easily be reinstalled correctly:
1. Shut off gas supply and power.
2. With a 1/4” wrench open the unions on the bulkheads.
3. First, remove the two unions at the rear; pull out the columns, mostly the fused silica adapters stay in
the union. Remove the nut from the columns, tilt the instrument and catch the fs-adapters with your
other hand. If they do not come out this way, pull them out with a needle.
4. Inspect the ends of the column: if they are broken, cut off 0.5 cm with the ceramic column cutter that
is supplied with the instrument. To cut off a piece just make one medium light stroke with the side of
the cutter and break the column backwards. Inspect under a good lamp, if there is a square cut.
5. Also clean out the fs-adapter, preferably with a very small drill, else with thin wire.
6. Remount these two column ends by following the procedure for fastening a column as described
above. Then proceed with the two at the front.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 110
Normally, there should be enough spare column to cut 0.5 cm from each end. Else it is possible to
rewind the column to allow for more room.
When other separations are desired or when the column deteriorates after several years, Synspec or
your distributor supplies precut columns as a replacement.
1. Remove the unions two by two as described above, then bend the ends of the column into the
bracket, to prevent them getting stuck during removal.
2. Now the column can be pulled out: first lift the column at the front side, then slowly pull it out.
3. When putting in the new column, follow the reversed procedure. Try to avoid scraping of the column
at the oven wall cut-outs at the rearside: scratches on the silica may lead to breaks.
4. The column case is just large enough to give a good grip at the oven walls.
5. Install the column in the bulkheads: the column with the long PTFE-sleeve at the a-, with the short
sleeve at the s-marked bulkhead.
6. Test for the correct flow at both analysis and stripper column.
The instrument comes provided with a stripper column and a analysis column. The lengths are usually
0.4 and 1.6 M, the internal diameter is 2.1 mm. The packings are chosen for their stability, if possible we
use a polymer type without coatings. However, for some analysis the only solution is to use a solid
support, mostly a kind of inorganic material with a coating.
The problems that occur with packed columns are:
• High boiling compounds absorb so well on them, that they are not removed from the stripper
column at the normal temperature. The column becomes saturated, and peaks get broader.
From time to time the column has to be heated for several hours while backflushing. For this you
should set the temperature at 100 °C (see Chapter 5). After cooling down, the peaks should
again be sharper.
• The column slowly looses its coating after several months of use and retention times get shorter.
Here, the solution is either to readjust flows, or to put in a new column.
• The column material breaks, the column gets more and more blocked: this occurs mostly with
carbon packings, which are fragile. The retention time gets longer. Here the solution is to put the
flow somewhat higher.
• Replacing a packed column set is easy: with a ½” key take out the present set and replace it by
a new one. To take it out, first unscrew the connectors on the inside, then those on the side of
the walls.
• The polyimide fittings are flexible and must not be screwed too tightly. Best first screw them on
by hand as far as possible, then tighten them, with only a light pressure of the key. Check
afterwards with leak test soap.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 111
The commercially available Nitrogen of quality 5.0 contains up to 1 ppm hydrocarbons (the next quality,
6.0 may still contain 0.5 ppm hydrocarbons).
To prevent pollution of the system the carrier gas is filtered with activated charcoal (mesh 60/80) in two
filters: one on the analysis and one on the stripper system.
It is advised to replace them after a year, or to replace them when the background ionisation signal
becomes high.
Procedure:
Stop the system and close off the gas supply.
Open the filter holder in the left compartment.
Only filters no. 1 and 2 have to be replaced, no. 3 filters the possible backflush from the sample piston
and has a life time of over 10 years. Loosen the connections of filters 1 and 2 with a ¼” key.
Remove them and replace with cleaned, activated coal filters.
It is quite feasible to reuse the filters: put them in an oven for a night, at 200 ºC, while flushing them
through with Nitrogen. (Heating without flushing does not work).
If then the reactivated filters are capped, they can be reused the next time filters have to be replaced.
The Syntech Spectra GC955 has a built-in industrial PC. The system is dedicated to the control of one or
two GCs with all their functions.
Also all the data saving and data communication is handled by the PC.
As an option analogue and digital data logging from other instruments is possible.
This PC can of course be used for other things, like word processing, spread sheet evaluation, playing
games, etc. Over the different ports you can load any software. However, in that case we do not
guarantee correct functioning of the system. If Synspec has to repair systems with software problems,
and we find strange software on the system, we charge you for all costs involved.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 112
6.5.2 Data communication with external data loggers and other PCs
Several versions of data logging are available from our side. Also ethernet connection is possible.
A problem most frequently occurs in three cases:
1 With data loggers: only at the end of a cycle a new value will be transmitted through the RS232
port. This is different from many other instruments that may be present at the same site. So, in many
cases, the RS232 port is frequently approached by the customers’ data logger. If this occurs with
instruments that are set to transmit many parameters of measured peaks or of status settings, the
system may become very slow or fail. Please set the data logger frequency to the lowest value you can,
preferably to once every cycle. If you use a modem for data logging, it should have a speed of at least 54
kBps.
2 Some types of data logging programs, like the bus programs, have many subtle variations. The
GC can in principle be set to communicate, but we need detailed information of the data logging
software protocol. We will be pleased to contact the software department of the customer to exchange
information.
3 An ethernet connection enables quick data transfer. We do not advise to log on to an office
network with a central backup. This will mean that each time a new file is saved or a file is changed, a
copy will be made over the network. For a standard GC this means that at the end of each cycle the bin-
file and the rundata.txt must be copied. At the end of a month this may be too much for the system,
which would produce an error message. Because of this message, a line would be added to the file
GC955.log. Therefore GC955.log would also be backed up, which creates another error, and so on, and
the system would get stuck. So, logging on to a central network is only possible if you can avoid
automatic backup.
backup
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 113
Optimal conditions for the PC are: steady power supply, no vibrations, temperature 5-30 °C, humidity
below 80%. But: if necessary, systems can run up to 45 °C, power failures are not so terrible, and good
hard disks can bear a lot of humidity.
If the temperature indicator of ambient temperature is above 45 °C , you should take measures to lower
it: improve ventilation, look for a different place to set the system.
With high humidity and high temperature the hard disk is at risk when the system is stopped. On cooling,
down the internal pressure will be lowered and humid air can be sucked in. This will lead to rust formation
inside the hard disk.
Vibration from pumps is bad for the hard disk.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 114
During transport, sometimes the processor ventilator becomes unstuck. Try to fix it again, but if it is
broken, you can buy standard ventilators for Pentium processors in most countries without a problem.
Customers sometimes change the set-up of Windows on their own. However, it is essential that this
should be as it was originally. Here are the correct settings:
Configuring the outputs is described here more extensively than in Chapter 5. If you want to use the
digital outputs please check if the following text helps you any further, else contact Synspec for more
help on your desired settings:
Under programview\options select Digital outputs..., the following dialog will appear :
On the left side you find the digital output no (1..7) with its corresponding PIN-no at the backside of the
GC.
Digital output no 5 is active at pin 9, but also on the relay-output, pin no 2,3 and 4.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 5 6
Every digital output can be linked to an internal state of the GC, or to a combination of internal states.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 116
Zero-air-data at analogue output The previous run was a run with zero gas
Calibration data at anologue output The previous run was a calibration run
New run start During the first half of the run, this state is high, it toggles to low state
Span gas value out of range Span gas value was higher than the alarm setting in program
edit\settings\auto-validation\conditions
Zero-air value out of range Zero gas value was higher than the alarm setting in program
edit\options\auto-validation\conditions
High component alarm One of the components reached a level higher than defined high level in the
dialog program edit\settings\alarm settings....,
and the active checkbox for this component was checked
Low componentalarm One of the components reached a level higher than defined low level in the
dialog program edit\settings\alarm settings....,
and the active checkbox for this component was checked
High system alarm One of the following situations occurred:
1. The column flow has been below the alarm level defined in GC955-
ID.TXT (default 1 ml/min.) for more than 10 seconds
2. The column flow was more than 10 seconds higher than the alarm
level defined in GC955-ID.TXT (default 4 ml/min.)
3. The air pressure was below 2.0 Bar
4. The air pressure was above 4.0 Bar
5. The hard disk space is below 10 MB
Low system alarm The hard disk space came below 30 MB
Flow too low The column flow was more than 10 seconds below the alarm level defined in
GC955-ID.TXT (default 1 ml/min.)
Air pressure too low The air pressure was below 2.0 Bar
Waiting for remote start Run is waiting to start until digital input 1 is high.
Waiting for time synchronisation Run is waiting for real time synchronization.
Combined alarm
Oven temperature error
Data at output not valid (flagged)
Combination 1
Combination 2
Combination 3
Combination 4
Using combinations :
After selecting one of the buttons C1 to C4 in the dialog above, a new dialog will appear like this :
To install a Vici valve after delivery to the customer, the valve must be mounted on the rear of the
instrument, the stainless steel tubing must be adapted and the digital connector for the GC must be
connected.
Mounting of the valve: with the valve you will receive a mould to bore two holes in the aluminium wall.
Bore the holes at this places, screw on the holder of the valve and mount the valve as tight as possible
on the holder. This is important, else the valve will work itself loose.
The T that connects the 10-port entrance and the pump at the instrument must be removed and the
tubing that is supplied with the valve must be installed. Use for this the ½” and 7/8” wrenches that are
supplied with the GC.
If the calibration gas comes from a gas diluting system, please see to it that the flow is between 1 and 4
l/min. In this case the pump can also function during automatic validation.
If the calibration gas comes undiluted from a bottle, install a restriction between the bottle and the 6-port
valve. Set the pressure in such a way, that the flow is ca. 20 to 40 ml/min.: this is enough, provided the
pump is switched off during calibration.
To check the working of the valve, you can use the manual switch button that is supplied with the valve.
However, normally the valve is meant to be switched automatically from the GC itself. Store the manual
switch after testing, for possible rechecking in case of malfunctioning.
The electrical connection from the valve to the GC must be made as follows:
• Connect the TTL switching signals to the digital outputs of the GC (terminal 5 Pos A, Terminal 6 Pos
B, terminal 16 Ground), using the Green 16 pole Phoenix connector.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 4 5 6
WARNING:
After a power failure, it may happen that the valve is opened half when the power comes on again.
To be sure that everything functions normally, check the result of the auto-validation that always starts
automatically after power failure. Check that after this auto-validation the external sample is measured.
If not, switch the valve a few times manually in actual run\options\switch validation valve and again test
that the external sample is on.
If the valve is connected to an undiluted bottle and the flow of this bottle is not restricted, it may empty
very quickly! Synspec gives no refund on
on emptied bottles if the auto-
auto-validation valve is not checked after
power failure.
To switch the valve properly, the signal for Valve Position A must always be the inverse of Valve Position
B. If one is high, the other one is low and vice versa.
So, in the digital output dialog, press the button C1 to define a combination.
Select for both items "Span gas valve open", check the two boxes NOT, and the radio button "OR". Give it
a name like "No Span gas". The combination gives TRUE, if the span gas valve is not opened.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 120
Now, in the digital output dialog select span gas valve open for output 5, and for output 6 Combination 1
No Span gas.
The edited dialog looks like this :
Of course, if the procedure described above is followed, the valve can also be operated from other
output channels.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 121
7.4 Errors occurring with the stream selector valve __________________________ 129
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 122
7.1 Introduction
7.1.1 Installation
The digital connections must be activated in the program unit: digital outputs. For this, see section 7.3
In this part, a few remarks are made about programming. These are based on knowledge of the chapter
5, which treats the subject deeper. Please read chapter 5.1 to 5.4 before changing the program by
yourself.
In the view mode Program edit two extra bars will appear, one for the external sample pump (located in
the stream selector), and one for setting the channel-step-signal. This is the moment in the run, when the
stream selector switches to the next channel. Of course, this moment must fit with the sampling settings
of the GC program.
Make the adjustments for the step signal step and the external sample pump and restart the program.
For an example see figure 7.1.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 123
Figure 7.1: New bars in Program edit, External sample pump and step signal
Before programming the sequence of the stream selector the channels can be named after the location
they are connected to i.e. "Storage" or "Lab first floor".
This is done in the view mode stream selector under options\edit channel names.
Here you can edit up to 16 different channel names. The names are saved by clicking on ok.
To activate them in the program, click stop run and start run as described in section 5.1.
With the menu-item stream selector\edit stream selector cycle it is possible to program a 20-step duty
cycle for the stream selector.
The GC starts measuring with the first step and continues until an entry “ Restart cycle” is reached or until
step 20 is reached. Then the cycle restarts at step 1.
The stream selector alarm levels can be set for each channel per component. This is important to cover
an area, where the risks are different concerning the acceptable levels and the components that might
cause the problems. However, calibration is only done once.
In the ASCII-file RUNDATA.TXT, for each line the channel number will be inserted.
To save the channel data in separate files, add the following line to GC955_ID.TXT:
PERCHANNEL = TRUE
The filenames will be RUNDATA.Cxx where xx is the channel number.
To store the data in separate files each month, add the following line to GC955_ID.TXT:
RUNDATAMONTH = TRUE
The filenames for each channel will then be RDmmyyyy.Cxx, where mm is the month and yyyy is the year
of the saved data.
7.3.1 Graphs
Stream selector can be used for the calculation and saving of time weighted averages.
Many industrial applications are intended to meet of the workers’ safety regulations, which
usually require the data to be saved as time weighted averages.
A new software feature of the Syntech Spectras, is the calculation per channel of TWA and the setting of
alarms not on last measured value but on the TWA.
This option only works on one component.
To use this option you must have an instrument with a stream selector and the new version of the
software. First the instrument has to be set up and calibrated.
After this, open the window: TWA settings. Channel by channel you can now activate the TWA function.
This means that you are able to set different alarm levels and that the TWA will only be used for those
channels that are in function.
You can choose the time span to be averaged, and name it. Of course, the most well known setting is for
an 8-hour mean.
If you want to use only one setting, you just set it for the first channel and copy it to all other channels.
This colouring enables monitoring personnel to get an impression of the status of the concentration at
some distance of the screen.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 129
As soon as in the GC955_ID.TXT "STREAMSELECTOR = TRUE" is found, the COM2 will be initialised,
and the commands Go1, Go2 etc will be sent to the stream selector.
3) If no error occurs during initialisation, and the stream selector does not react on commands of the GC,
the following has to be checked:
a. the RS232-connection cable between stream selector and GC might not be correctly fitted. In this
case, the pump can still work, as it is directed over the digital connector, but the valve movement cannot
function.
b. If the connection seems to be ok, then a test with the Windows terminal program is used to
check the proper working of the valve. Start up a terminal program (Hyperterminal under Accessories in
Windows 98) select com2, 9600,n,8,1. Type GO2<enter>, the valve should switch to position 2. If it does
not, contact Synspec.
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 131
INDEX
10-port valve, 23, 111 computer, 28 heating power supply, 101
abort run, 40 Computer, 116 Heating velocity, 62
accept alarm, 56 constant temperature, 24 Hessen-Bayern, 71
3
activate a programchange, 58 conversion from ppb to µg/m , 35 heterogeneous hydocarbons, 26
actual run, 37 cooled pre concentration, 60, 63 high voltage supply cable, 96
add data to current calibration, 43 cooling down, 59 identification file, 85
Adjustment of the sampling, 64 Cooling down, 62 Identification file, 35
alarm settings, 72 cycle time, 24, 59 in stainless steel-columns, 28
alarm signal, 55 data registration, 50 incomplete separation, 48
ambient temperature, 38 data transfer, 52 inert microfilter, 35
Analog inputs, 32, 72 datafile, 37, 39 injection time, 59
analog output connector, 68 detector housing, 98 install a Vici valve, 121
Analog outputs, 32 diaphragm valve, 111 instrument panel, 67
analysis column, 23 digital in- and/or output ports, 31 Integrate highest peak, 48
animation, 52 digital output, 118 integration, 31
aromatic hydrocarbons, 26 Digital outputs, 32 integration method, 48
ASCI-terminal, 71 digital TTL-outputs, 70 internal pump, 60
attributes, 48 discharge the high voltage,
voltage 88 IO diagram, 55
Autolinearisation, 80 discrete time-schedule, 67 lamp window, 96
automatic restart, 47, 67 ECD, 9, 27 leaks, 107
autovalidation, 78 edit sample concentrations, 43 Loading of strange software, 115
Autovalidation, 76 electronic boards, 28 machine stays standby, 47
background zero adjust, 38 elution, 23 Maintenance of the cooled pre
base oven temperature, 61 Ethernet connection, 32 concentration, 108
baseline changes, 48 external data loggers, 115 maintenance on the FID, 100
base-temperature, 36 FEP, 35 maintenance work, 95
bulkheads, 108 FID, 9, 26, 36 manometers, 34
Calculating FID-, 7 measured raw data, 31
the conversion factor, 44 flow programming, 62 menu structure, 29
calibration, 37 flowsensor, 37 modem connection, 31
Calibration data, 31 forced cooling, 61 O ring, 98
calibration facilities, 42 fs-adapter, 112 original datafiles, 54
calibration line, 31 fused silica, 28, 111 oven temperature, 38, 59
calibration mixture, 45 gas flow, 60 password, 56, 73
carrier gas consumption, 107 Gesytec, 71 PC-Anywhere, 55
carriergas, 34 graph of temperature PDECD, 27
carriergas filters, 114 against time, 62 periodical automatic validation, 74
channel names, 127 graph of the measured photoionisation detector, 96
chromatograms, 50 concentrations, 53 PID, 25
cleaning of the system, 24 graphs, 37 PID-signal, 25
combinated digital outputs, 70 hard disk, 116 piston, 22
Manual for the Syntech Spectras GC955 series 400 , 600 and 800 single/double 29-1-2013 132
Comercial/Commercial: Assistência/Service:
Note:
SYNSPEC b.v., the owner of this document, reserves the right to make technical changes or modify the contents of
this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. SYNSPEC b.v.
does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.
Copyright© 2011
SYNSPEC b.v. All rights reserved