Professional Documents
Culture Documents
Compact Pressure Switch: Zse2 / ISE2
Compact Pressure Switch: Zse2 / ISE2
ZSE
ISE
ZSP
PS
ISA
PSE
IS
ISG
ZSM
779
P0756-P0822-E.qxd 08.8.27 6:36 PM Page 780
Series ZSE2/ISE2
How to Order
Positive
pressure ISE2
Vacuum ZSE2
Output specifications
15 NPN open collector
55 PNP open collector
780
P0756-P0822-E.qxd 08.8.27 6:36 PM Page 781
Operating temperature range Operating: 0 to 60°C, Stored: –10 to 60°C (With no condensation and no freezing)
Operating humidity range Operating/Stored: 35 to 85%RH (With no condensation)
Vibration resistance 10 to 500 Hz at whichever is smaller of 1.5 mm amplitude or 98 m/s2 acceleration, in X, Y, Z directions for 2 hrs. each (De-energized)
Impact resistance 980 m/s2 in X, Y, Z directions, 3 times each (De-energized)
Temperature characteristics(Based on 25°C) ±3% F.S. or less
Withstand voltage 1000 VAC for 1 min. (between live parts and case)
Insulation resistance 50 MΩ or more (at 500 VDC by megameter) between live parts and case ZSE
01: R 1/8, M5 x 0.8 T1: NPTF 1/8, M5 x 0.8 ISE
Port size
0X: With suction filter (For mounting on ZX unit) 0R: Base mount type (For mounting on ZR unit)
Mass 35 g (Including 0.6 m-Long lead wire)
ZSP
Grommet type Oil-resistant vinyl cabtire code 3 cores, ø3.4, Cross section: 0.2 mm2, Insulator O.D.: 1.1 mm
Lead wire
Connector type Heat-resistant vinyl electric wire, 3-wire, Cross section: 0.31 mm2, Insulator O.D.: 1.55 mm
PS
Standard Compliant with CE marking
ISA
Output Specifications PSE
781
P0756-P0822-E.qxd 08.8.27 6:37 PM Page 782
Series ZSE2/ISE2
Internal Circuit and Wiring
Calibration Procedure How to Use Connector Example
• Set the ON-pressure by the pressure setting 1. Attaching and detaching connectors -15 NPN open collector
trimmer. Turning clockwise can set the high
pressure/high vacuum pressure. • When assembling the connector to the switch DC ( + )
Brown
housing, push the connector straight onto the
pins until the lever locks into the housing slot. OUT
Main circuit
+
• When removing the connector from the Black
Load
12 to 24 VDC
switch housing, push the lever down to Max. 30 V, 80 mA
LED Red DC (–) –
unlock it from the slot and then withdraw the Blue
Pressure connector straight off of the pin.
setting trimmer Electric circuit diagram
Lead wire 0.2 to 0.33 mm2
Max. coating O.D. -55 PNP open collector
ø1.7 mm DC ( + )
Brown
SET
Main circuit
Socket OUT +
Indicator light
Black 12 to 24 VDC
Hook Load –
LED Red DC (–) Max. 80 mA
Connector Blue
DC polarity indicator
Lever Electric circuit diagram
Slot COM
Housing – +
12 to 24 VDC
Supply pressure
Set point
Steady adsorption • Filter case
Unsteady 2. Crimping of lead wires and sockets Caution
adsorption Strip 3.2 to 3.7 mm at the end of the lead 1. Do not use with thinner, carbon tetrachloride,
Non adsorption wires, insert the ends of core wires evenly into chloroform, acetate, aniline, cyclohexane,
Atmospheric 0 the sockets, and then crimp with a crimping
pressure trichloroethylene, sulfuric acid, lactic acid
tool. When this is done, take care that the and watermiscible cutting fluid (alkaline).
coverings of the lead wires do not enter the 2. Operate it away from direct sunlight.
core wire crimping area.
• Regarding the pressure setting (Crimping tool: model no. DXT170-75-1)
Caution Core wire crimping area Covering retainer
Observe the following precautions for setting Core wire
Socket
the vacuum pressure: Use your fingertips to Lead wire
gently turn the screwdriver. Do not use a
screwdriver with a large grip or with a tip that
Hook Covering
does not fit into the trimmer groove because
this could strip the groove.
Caution
3. Attaching and detaching lead wires
with sockets
Be sure to read before handling.
Hysteresis Refer to front matters 58 and 59 for
• Attaching Safety Instructions and pages 687 to
Insert the sockets into the square holes of
Hysteresis is the pressure difference between
the ON and the OFF pressure of the output
the connector (with +, 0, – indication), and 691 for Pressure Switch Precautions.
continue to push the sockets all the way in
signal. The set pressure is the pressure
until they lock by hooking into the seats in
selected to switch from OFF to ON condition.
the connector. (When they are pushed in
their hooks open and they are locked
pressure
High vacuum
782
P0756-P0822-E.qxd 08.8.27 6:37 PM Page 783
20
H
SET
Vacuum Vacuum pad
L
supply connection port
7 35.6 port
Pressure setting trimmer
Pneumatic circuit
(3000)
600
ZSE
ISE
ZSP
32
58
PS
52
10.2 ISA
6.5
PV
18
PSE
8
10 10 17.6 5.5 IS
Vacuum supply port 2 x ø3.3 Mounting hole
M5 x 0.8 Vacuum pad connection port ISG
M5 x 0.8
ZSM
Connector type: ZSE2-0X-15C
Indicator light (Red)
20
L SET H
- O+
32
59
58
10.2
PV
6.5
18
8
10 10 17.6 5.5
Vacuum supply port 2 x ø3.3 Mounting hole Vacuum pad connection port
M5 x 0.8 M5 x 0.8
783
P0756-P0822-E.qxd 08.8.27 6:37 PM Page 784
Series ZSE2/ISE2
01
Dimensions/Standard Type: ZSE2- T1
SET SET
L H L H
15
15
(3000)
600
(3000)
600
48.5
41
39
39
3.5
3.5
8.5
8.5
2 x ø3.5 13 Piping port 01: R 1/8 2 x ø3.5 13 Piping port 01: R 1/8
Mounting hole M5 x 0.8 depth 5 T1: NPTF 1/8 Mounting hole M5 x 0.8 depth 5 T1: NPTF 1/8
SET SET
L H L H
15
15
9
(3000)
600
600
42
40
49.5
40
3
20 20
18 18
16 16
13
11
13
11
9
Gasket Vacuum supply port ø3.3 Gasket Vacuum supply port ø3.3
784
P0756-P0822-E.qxd 08.8.27 6:37 PM Page 785
SET SET
L H L H
15
15
(3000)
600
(3000)
600
ZSE
ISE
48.5
ZSP
41
39
39
3.5
PS
3.5
ISA
8.5
8.5
2 x ø3.5
2 x ø3.5 13 Piping port 01: R 1/8 13 Piping port 01: R 1/8 PSE
Mounting hole M5 x 0.8 depth 5 T1: NPTF 1/8 Mounting hole M5 x 0.8 depth 5 T1: NPTF 1/8
IS
ISG
ZSM
785