Multi-parameters Module
MMP12
Multi-parameter
Module
V3.0
Overview
The MMP12 integrated 12 Leads ECG Multi-Parameters module can provide
measurement for ECG, respiration, temperature, blood pressure, blood oxygen,
pulse, etc. The measurement method of ECG: the heart first generates electrical thrill
before mechanical contraction, which generates biological current and conducted to
the body surface. ECG signal is the dynamic curve recording surface potential difference
that caused by different potentials changes in different body site. Impedance respiration
is measured across the thorax. When the patient is breathing or ventilated the volume of
air changes in the lungs, resulting in impedance changes between the electrodes. Respiration
rate (RR) is calculated from these impedance changes. Thermistors are used to monitor skin
temperature. The method of blood oxygen measurement is based on the absorption of selected
wavelengths of light. The light generated in the probe passes through the tissue and is converted
into electrical signals by the photodetector in the probe. The SpO2 module processes the electrical
signal and displays a waveform and digital values for SpO2 and pulse rate. The measurement of non-
invasive blood pressure is based on the principle of oscillation method. Wave of pressure pulse is
detected as the cuff being quickly inflated and deflated; the cuff is tied on limbs. After processing the
oscillation signals and calculation, parameters concerning mean pressure, systolic and diastolic pressures,
pulse rate are obtained.
Application
• SpO2, NIBP, 12 Leads ECG, Temperature, Respiration Rate, PR functions of patient monitor and emergency
transport monitors.
Functions
ECG measurement
• The result of ECG measurement maximum sup- • Calibration for ECG: input 1mv standard voltage
port 12leads ECG waveform includes heart rate, to record waveform amplitude.
body temperature, respiration, and ST-segment
deviation value of channel I, II and V1 as well as • Channel of Heart rate calculation/Arrhythmia
the arrhythmia. analysis: Heart Rate I, II and V1 as well as self-se-
lect are for selection.
• Four measurement modes are provided: diagno-
sis, monitoring, HARDEST and operation. • Notch Mode: 50 Hz, 60 Hz, and 50/60 Hz and off
are for selection.
• Three patient models are supported: adult, pedi-
atric and neonate. • Interval measurement of ECG waveform charac-
teristic are supported: involving measurement of
The range of filtering in mode PR interval, QRS interval, QT interval, QTC, P axis,
Diagnosis
of diagnosis is 0.05hz-130hz
QRS electrical axis and T axis as well as amplitude
The range of filtering in mode measurement of RV5, SV1.
Monitoring
of monitoring is 0.5hz-40hz
• Printable diagnostic analysis are provided includ-
The range of filtering in mode
HARDEST ing arrhythmia, conduction block, myocardial in-
of HARDEST is 5hz-20hz
farction, STT variations, ventricular hypertrophy,
The range of filtering in mode of atrial enlargement, electrical axis deviation and
Operation
operation is 1hz-25hz
other cardiac diseases.
SpO2 measurement
• Real time measurement of SpO2, heart rate and • The ability of anti-motion interference and weak
perfusion index (PI). signal measurement.
Performances
ECG
Range...........................................................................................................................0 – 100 %
Range of measurement Base resistance value........................................................................................ Undefined
Variable resistance value........................................................................................ 1%
Diagnosis........................................................................................................................ ≥90 dB
Common-mode rejection Monitoring.................................................................................................................. ≥100 dB
Operation............................................................................................................... ≥100 dB
Remove the maximum and minimum values of recent 12 RR intervals, and calculate the
Mean algorithm of HR mean value
Tall T-wave rejection The maximum T-wave amplitude (aT) meet the requirement of 1.2mv specified in IEC
capability 60601-2-27
Range................................................................................................................................. 0-300
PVCS (statistics of
Accuracy................................................................................................................................ No
premature beat)
Resolution........................................................................................................................ ±1
Respiration SpO2
Range of RR......................................................... 0-120 rpm
Range of measurement..................................... 0-100%
Accuracy............................... 15-120 rpm: ±2 rpm or ±2 %;
Accuracy........................................................70-100%, ±2%;
Below 15rpm: not Defined
70 % or below, no definition
Base impedance...........................................500-2000 Ω
Resolution..................................................................... 1%
Variable impedance....................................... 0.2-3.0 Ω
Overpressure Protection
Protection pressure Protection pressure
Patient category
of software of hardware
Non-invasive sphygmomanometers-Part 1:
5 EN1060-1 1995
General requirements
Non-invasive sphygmomanometers-Part 3:
6 EN1060-3 Supplementary requirements for electro- 1997
mechanical blood pressure measuring systems
Environment Characteristics
Dimensions
4 LL LL (left leg)
5 LA LA (left arm)
6 RA RA (right arm)
1 RL RL (right leg)
11 V3 Midway between V2 and V4
4 LL LL (left leg)
5 LA LA (left arm)
6 RA RA (right arm)
Ordering
Our integrated 12 Leads ECG Multi-Parameters module part number is MMP12.
For ordering our module, please contact directly our website www.zugmed.com for further information.