Inspection Report

NO:HWKF202204279001

Detector type 3543

Detector Series N 30203543A306 18M108564

Inspection information AGA18M108564


SPN

Angell detector team


Inspector

2022-4-27
Inspection Date
Inspection Repair Content

Feedback
Fault Phenomenon Can not get image
Date:2022/4/21
Analysis staff:
Angell detector
In the testUI software(detector calibrationn software),we can team
not get into AED2 mode of detector,AED2 means detector can
get image when it receive X ray. We try to get a dark image of
the detector, found the dark image abnormal,there is a problem
in the area above the image, suspected of damage or liquid
Fault Cause
infiltration.Ask customer take photo of the detector and found
that the shell at the upper left corner of the detector has been Date:2022/04/27
sunken. It is suspected that it has been impacted or heavily
pressed, which will cause internal damage to the detector,can
see the photo.

Analysis staff:
It needs to be returned to the Angell Technology for Angell detector
Handling Method maintenance. team
Date:2022/04/27
Inspection Repair Content
NO:20220610003
Detector type 3543
Detector Series N 30203543A306 18M108564
Inspection information SPN AGA18M108564
Inspector 邱国定
Inspection Date 2022-6-9
Inspection Repair Content
Analysis staff:
Fault Phenomenon there is not able to come to AED2, raw dark field is strange 施逸君
Date:2022-6-7
Step: Analysis staff:邱国定
1. The dark field TFT image in the repair record is incorrect.
2. First connect the collected image to determine the image
abnormality, and part of the dark field 8000.
3. Open the front face (TFT) of the flat panel detector, and
the wiring between the binding area of the six gates and the
TFT is corroded. This phenomenon cannot be directly
photographed with a mobile phone. Take pictures with the help
of a microscope. Figure A shows the corrosion at the wiring,
and figure B shows the normal picture.
4. The normal use of TFT requires that TFT+ROIC+GATE are
Fault Cause
normal. The damage usually occurs in these three categories. Date:2022-6-9
The damage is caused by the connection between the gate
binding area and the TFT. There is a line in this area. It is
determined that the line has been corroded. Once there is a
problem with the line, the whole TFT+ROIC+GATE can only be
replaced.

TFT is the most important part of the detector which for


imaging,it is weak and will damage when:
1.Corrosion due to liquid infiltration
2.Crush or impact
TFT+ROIC+GATE needs to be replaced. Separate TFT cannot Repair staff:邱国定
Handling Method work. Gate and IC must be bound to work normally
Date:2022-6-9
Repair staff:邱国定
Handling Result
Date:2022-6-9

Not Good A Good B


red circle: corrosion As a comparison
yellow circle: watermark