1. General
Variations of load cell materials and manufacturing processes can cause the
parameters of a load cell to vary. The potentiometer in junction box can be adjusted to
maintain stable load cell sensitivity and the balance of a weighing instrument.
2. Model names
Model Meaning Details
C Caisun
J Junction box
4 numbers of load cell can be connected 2-12 pcs load cell
G Connector material G is normal connector, and Y is IP68 connector
C Junction box type Type A, B, C, D, T, and Type XH
E Junction box operation E is excitation voltage adjustment
SP is signal adjustment by normal resistance
SJ is signal adjustment by precision resistance
DL is to connect digital load cell
J1 J2 J3 J4
-E
+IN
-IN
2nd and 3rd points: To adjust potentiometers W2 and W3, and simultaneously monitor
U2, U3 to 24.05mV.