Professional Documents
Culture Documents
Viale Industria 4
12051 Alba (CN) Italy
T +39 0173 31 41 11 F +39 0173 34 872
www.bianco-spa.com
FLUO LM2
OPTICAL CENTRING DEVICE
Attachment INTRODUCTION
Table of contents
1 TECHNICAL DATA AND EQUIPMENT IDENTIFICATION......................................... 2
2 DESCRIPTION OF THEEQUIPMENT ......................................................................... 4
2.1 Main parts .................................................................................................................... 4
2.2 Compatibility with the previous system, FLUO LM1 .............................................. 5
3 CONDITIONS OF USE ............................................................................................ 6
4 OPERATOR WORKSTATION ................................................................................... 6
5 INSTALLATION AND START-UP .............................................................................. 7
5.1 Equipment assembly ................................................................................................. 7
5.1.1 Preliminary check ............................................................................................................ 7
5.1.2 Anchoring the sensor tube ............................................................................................. 7
5.1.3 Vibrations .......................................................................................................................... 7
5.1.4 Installing the control board ............................................................................................ 8
5.1.5 Accessibility ...................................................................................................................... 8
5.1.6 Heat sources ..................................................................................................................... 8
5.1.7 Cleaning the sensor......................................................................................................... 8
5.2 Electrical system ......................................................................................................... 9
5.2.1 Line speed reference ...................................................................................................... 9
5.2.2 Exchange of enabling signals and messages to the outside ................................... 9
5.3 Machine commissioning. ........................................................................................ 10
5.3.1 Preliminary checks ......................................................................................................... 10
5.3.2 System start-up and calibration .................................................................................. 10
6 DISPOSAL............................................................................................................. 11
______________________________________________ Page 2 of 11
Sensor for maximum ambient temperature: 40°C when using the polycarbonate
cover, upon request we can provide you with different tube covers that can be
used at high temperatures.
Note: for operation at altitudes over 1000 metres a.s.l. it is necessary to reduce the
maximum permitted ambient temperature by 0.5 °C every 100 metres more.
The equipment is supplied with an identification plate, affixed on the side of the
panel or on the internal side of the door, reporting the following identification data:
Indicative picture:
The sensor has an identification label placed on the side of the sensor tube
indicating its serial number and production date, along with some other data.
______________________________________________ Page 4 of 11
2 Description of theequipment
2.1 Main parts
FLUO LM2 sensor is made of 2 parallel bars between which slides the fabric.
The distance between the bars is fixed, but the mechanical length and the inner
quota that determine the active reading areas are defined based on the final use
of the machine.
One of the two bars is a blinking infra-red light emitter, the other one receives the
light possibly attenuated by the fabric sliding between the tubes.
An automatic system determines the intensity of the light through two appropriate
DARK and LIGHT signals, the self-adapting system is suitable for any type of fabric,
ensuring optimal reading.
The reading tube is divided in 2 parts: RIGHT and LEFT, the electronic system
determines the direction in which the fabric will be moved and triggers an
adjustment signal proportional with the run-out, allowing the actuators connected
to the adjustment outputs to restore the fabric to its appropriate position.
The maximum length of the cables connecting the FLUO LM2 sensor to the B016
control board is of 10 metres, should you require grater connection distances,
please contact Bianco S.p.A.
CAUTION: The cables used for connecting the sensor tube to the control board are
screened cables, please consult the wiring diagram for their connection.
CAUTION: Make sure that the tube configuration is suitable for installation, the
sensor is available in different covers: polycarbonate or coated in silicone for high
temperatures, you can also choose from several types of side plugs, based on the
necessary IP protection level and on the final installation site.
______________________________________________ Page 5 of 11
Control board B016 communicates with FLUO LM2 sensor. with the external
analogue and digital I/O and with the centring adjustment actuators.
CAUTION: Board B016 should be supplied using +24VDC and not 24VAC as for the
previous C212 control board of FLUO LM1 system
3 Conditions of Use
Make sure that the ambient temperature does not exceed the operating limits set
for the sensor and control panel, make sure that there are no vapours, as they may
affect the measuring or the system performance. If using the machine in humid
environments or if the temperatures do not match the ones intended, please make
sure to use a suitable configuration of FLUO LM2 tube sensor.
4 Operator workstation
Under normal conditions of use, the only task of the operator is to decide whether
to operate the centring system in AUTOMATIC mode or in MANUAL mode using the
keys MAN+/MAN-, and to define the central position of the fabric by setting it using
the BALANCING potentiometer connected to B016 control board.
______________________________________________ Page 7 of 11
Remove the packaging from the equipment and make sure all parts are intact.
Check if the power supply is suitable.
CAUTION: Board B016 should be supplied using +24VDC and not 24VAC as for the
previous C212 control board of FLUO LM1 system
CAUTION: Make sure that the tube configuration is suitable for installation, the
sensor is available in different covers: polycarbonate or coated in silicone for high
temperatures, you can also choose from several types of side plugs, based on the
necessary IP protection level and on the final installation site.
Fasten FLUO LM2 sensor tube to the machine structure, making sure that the fabric
passes between the tubes without touching them.
If possible, avoid installing the system in places with strong lighting or heating
sources, as they might affect the system performance.
Fasten the pair of sensor tubes using the suitable side brackets and connect the
grounding strips.
5.1.3 Vibrations
Fasten the control board B016 or the electric panel in which it is placed, connect
the sensor tube, the supplies, and the auxiliary accessories such as potentiometers,
buttons, limit switches, actuators, reference or exchange signals.
CAUTION: Carefully follow the wiring diagram supplied with the machine.
5.1.5 Accessibility
You should be able to open completely the access doors to the control panel in
order to carry out the maintenance operations. You should be able to reach the
sensor upper cover in order to remove it if necessary and access the internal
boards.
The bulbs installed in the control panel should be visible.
Avoid placing the electric cabinet near direct heat sources or vapour jets. Avoid
direct exposure to sunlight.
Do not use water jets to clean the sensor, unless the tube configuration allows it.
Clean using a wet cloth, do not use solvents as they may dull the protective
covering of the tube.
______________________________________________ Page 9 of 11
CAUTION: Place the speed sensor near a drive roller or an idle roller in order to
ensure correct belt speed detection. Do not use the rollers placed in sections with
speed boost settings as the detected data will be distorted.
FLUO LM2 system can communicate with the line on which it is installed,
sharing the adjustment signals, measuring signals, contacts for exceeding set
tolerance thresholds, enabling signals or any other type of signals.
______________________________________________ Page 10 of 11
CAUTION: Check all wirings, overall power supply from control panel, proper
connection of sensor tubes and all other connections related to B016 board.
CAUTION: Board B016 should be supplied using +24VDC and not 24VAC as for the
previous C212 control board of FLUO LM1 system
Power board B016 and make sure that it starts up, on the 3-digit display should
appear the following message (scrolling):
Where:
dd-mm-yyyy represents the issue date for the software version installed on the B016
board
Proceed with the calibration of FLUO LM2 sensor tube, only after carefully reading
the next paragraphs in this manual in order to understand the operating mode of
the B016 board and how to set the software functions or check and set the
configuration parameters.
______________________________________________ Page 11 of 11
6 Disposal
DECOMMISSIONING
If the equipment becomes inactive, it is recommended that you remove it from
service or to affix to the same a warning note, clearly visible.
The sensor and the electronic boards should be disposed of in compliance with the
relevant national laws in force.
Table of contents
1 FLUO LM2: CONTROL BOARD B016 ...................................................................... 2
1.1 General description ................................................................................................... 2
1.2 Compatibility with the previous system, FLUO LM1 .............................................. 2
1.3 Functional limitations for B016 board...................................................................... 3
1.4 Terminals description ................................................................................................. 3
1.4.1 Terminal M1 – Solenoid valves – .................................................................................... 5
1.4.2 Terminal M2 – Power supplies –...................................................................................... 5
1.4.3 Terminal M3 – Sensor tubes FLUO LM2 TX and RX – .................................................... 5
1.4.4 Terminal M4 – analogue inputs and outputs – ............................................................ 6
1.4.5 Terminal M5 – relay auxiliary functions –....................................................................... 6
1.4.6 Terminal M6 – opto-insulated digital inputs – .............................................................. 7
1.4.7 Terminal M7 – auxiliary serial lines .................................................................................. 7
1.5 Fuses installed on board B016 .................................................................................. 8
1.6 Wire bridges, Switches, Buttons and Jumper for Board B016 ............................. 9
1.7 LED Board B016 ......................................................................................................... 11
1.8 Trimmer and Test Point Board B016 ....................................................................... 13
______________________________________________ Page 2 of 13
Control board B016 communicates with FLUO LM2 sensor. with the external
analogue and digital I/O and with the centring adjustment actuators.
CAUTION: Board B016 should be supplied using +24VDC and not 24VAC as for the
previous C212 control board of FLUO LM1 system
Terminals use:
CAUTION: Please connect / disconnect the terminals only after cutting off the
machine power supply, in order to avoid damaging the board or any other
equipment connected to it.
______________________________________________ Page 4 of 13
CAUTION: Board B016 should be supplied using +24VDC and not 24VAC as for the previous
C212 control board of FLUO LM1 system
NOTE: The wire bridge that connects 0V and PE (GROUND) should be performed in a single
point of the machine, usually inside the main panel, near the +24Vdc power supply of the
board B016.
Avoid such a connection by maintaining the 0V suspended only if noticing that the overall PE
connection is poor and that the disturbances caused to the PE by the motors, inverter or
screened cables, affect system performance, in such conditions, please keep the 2 potentials
PE and 0V separate in order to avoid affecting the measuring accuracy and in extreme cases,
damaging the devices.
F2 F4
F1 F3
1.6 Wire bridges, Switches, Buttons and Jumper for Board B016
B8 B9 SW1 SW2
NOTA1: To view the default configuration of these switches, please follow the
machine wiring diagram.
NOTE2: When impulse adjustment mode is enabled, the valves output is enabled
by impulses, allowing small movements.
NOTE3: If SW2.4 is OFF parameters display and editing mode is not accessible, After
installing and commissioning the machine, set SW2.4 to OFF to make sure that no
parameters can be accidentally changed.
NOTE4 : If the switch is ON the sensitivity value is doubled for parameters or for
analogue input.
NOTE5: If the switch is ON during board start-up or RESET the parameters are
restored to their default values, after performing this operation you cannot return to
the previously set values. After resetting the parameters, always set SW.6 to OFF
otherwise the parameters will be reset each time you start the device.
NOTE6: By setting SW2.7 to ON you can perform the system calibration. please
consult the relative paragraph in the user manual.
NOTE7: By setting SW2.7 to ON you will enable the board's functional testing mode.
This procedure should not be performed by the end user, it should be performed
only at BIANCO S.p.A. headquarters and it requires particular skills and devices.
Proceeding with this procedure at the end user's site can be dangerous because
the board I/O do not follow the machine's operating logic and may damage the
board or the machine.
Always leave SW2.8 set to OFF.
______________________________________________ Page 11 of 13
LD1
LD3
NOTE1: This LED blinks whenever a digital input, a dip-switch or a button changes
state, it can be used to make sure that the accessories work properly, if this LED
does not signal the state change (for example when a button is pressed), there
might be a connection error or the control board might be damaged.
NOTE2: This LED blinks when there is no fabric between the TX and RX sensor tubes.
It also blinks during the sensor tubes calibration procedure.
NOTE3: It turns on if there is a system error such as: board B016 overheating, wrong
level of supply voltage, loss of already set parameter values, other system errors. If
an error is detected the LED blinks and the error description scrolls on board B016
display, the system continues to display the error message even if the cause of the
error was removed, in order to make it disappear you have to RESET the board.
______________________________________________ Page 13 of 13
TP1
Trimmer
Pos. Description Notes
P1 Trimmer DARK sensor channel calibration Channels “DARK”, “LIGHT”, “LEFT”,
“RIGHT” should be calibrated in
order to adapt or compensate the
P2 Trimmer LIGHT sensor channel calibration reading sensitivity features of each
P3 Trimmer LEFT sensor channel calibration sensor tube.
P4 Trimmer RIGHT sensor channel calibration
Test Point
Item Description Notes
TP1 Test point 0VPOT
TP2 Test point for “DARK” signal These test points are used
when the calibration procedure is
TP3 Test point for “LIGHT” signal performed using an external
TP4 Test point for “LEFT” signal multimeter, referred to as 0V-ADJ
TP5 Test point for “RIGHT” signal (please consult the calibration
TP6 Test point 0V-ADJ procedure).
Table of contents
1 FLUO LM2: SOFTWARE BOARD B016 .................................................................... 2
1.1 General description ................................................................................................... 2
1.2 Software Functions not implemented yet ............................................................. 3
1.3 Main display ................................................................................................................ 4
1.4 Secondary parameters display ............................................................................... 5
1.5 Parameters menu display and editing ................................................................... 6
1.6 Description of the machine parameters ............................................................... 7
1.7 Sensor tube calibration procedure ......................................................................... 9
1.7.1 Calibration using an external multimeter .................................................................. 10
1.7.2 Calibration using the function “Sensor tube calibration parameters display” .... 12
______________________________________________ Page 2 of 13
You can perform any control or setting function using the 3 buttons “-“, “+”, “P”
and the 3-digit red display.
1) MAIN DISPLAY
2) SECONDARY PARAMETERS DISPLAY
3) PARAMETERS MENU DISPLAY AND EDITING
4) SENSOR TUBE CALIBRATION PARAMETERS DISPLAY
This chapter contains detailed descriptions of all software functions that refer to:
Any other Hardware or software versions than the ones indicated in the table might
affect the machine operation mode, please contact Bianco S.p.A. technical
support service if in possession of such versions.
______________________________________________ Page 3 of 13
RESET Control
Control of the meter counter and Speed SENSOR
MODBUS serial protocol
CANBUS serial protocol
Advanced control of the process regulator
Advanced control of the system errors
Zero limit switch control for parking in neutral position
dd-mm-yyyy FLUOrx
Vz.z
Where:
String Description
represents the issue date for the software version installed on
dd-mm-yyyy
the B016 board
dd=day, mm=month, yyyy=year
represents the hardware revision of the B016 board
x
represents the software version currently installed on B016
z.z
board
This message appears only if no system errors are present, otherwise, it is replaced
by the error description
ErruCC 18.9
For example, string “ErrUCC” indicates that the supply voltage exceeds the limits
24V+/-10%, being of 18.9Vdc.
If an error is detected, besides the error message, LED LD18 (red coloured) will also
switch on
The system continues to display the error message even if the cause of the error
was removed, in order to make it disappear you have to RESET the board.
The system errors that are taken into consideration are indicated below:
The errors are only displayed, they do not block the machine operation.
______________________________________________ Page 5 of 13
Starting from the main display, you can view certain operation parameters.
By pressing keys“+” or “-“ you can view and scroll through the following
parameters:
rFb 0.83
P
rFa -9.6
+ - -9.6
UPS 24.2
+ -
S°C 45.9
t°C 35.7
To exit secondary parameters display press at the same time keys “+” and “-“.
Data Description
It represents the difference between the 2 measuring sections
rFb
of the LEFT and RIGHT sensors, when there is no fabric in the
device, a value closer to zero indicates a more accurate
calibration.
It represents the output voltage on RIF+ measuring analogue
rFA
signals, this value takes intro account the set SENSITIVITY level
and the balancing level
It represents the supply voltage of the board B016, expressed
UPS
in volts.
Represents the temperature inside the sensor tube on receiver
S°C
side expressed in °C. The temperature sensor is connected
only in FLUO modules suitable for use at high temperatures; in
standard FLUO the input of the B016 is unused and the data
has no meaning.
It represents the temperature of the environment in which the
t°C
B016 board is installed, expressed in °C
______________________________________________ Page 6 of 13
In order to access the parameters menu make sure that the dip-switch SW2.4 is ON,
otherwise you will not be able to perform any operation.
You can access this menu only from “Main display”, therefore if you are in
“Secondary parameters display” exit this page following the instructions in the
previous paragraphs.
You can access the menu by holding key “P” for 5 seconds.
P P +
InS S.n.S. 27.6 32.0
b.I.L. 27.6
0
-
A.L.L.
C.E.n. 7.5
PId C.t.P. 0
C.A.P. 0.0 P
+ - E.n.C. 0.0
E.C.I. 0
ALL S.F.I. 0
S.E.t. 11
-.-.-. 2.0
-.-.-.
---
1. You can scroll through the desired table using buttons “+” and “-“
2. You can access the desired table by pressing button “P”
3. Each parameter in the table is displayed along with its value, that flashes
alternately with the set value
4. By pressing button “P” once again you can access the parameter editing
menu
5. You can change the parameter value using keys “+” or “-“
6. Press button “P” once again to confirm the parameter set and to return to
table mode view.
7. To exit a table, scroll down until reaching the exit cell, marked with “-.-.-.”,
and press “P” to confirm the action
8. To exit the menu, scroll down through the table list until reaching the exit
cell, marked with “-.-.-.”, and press “P” to confirm the action
If after changing a parameter's value you do not perform the proper table exiting
procedure, as described at points 7, and 8, The new value will not be saved and
will be lost at the next machine restart.
______________________________________________ Page 7 of 13
There are 3 tables with parameters: “Ins”, “PId” and “ALL”, in the current software
version you can manage only some parameters in the table “Ins”
(*) = parameter that cannot be managed in the current software version, the
modification of these parameters does not affect the machine operation.
If setting the dip-switch SW2.5 to ON, the sensitivity set for that parameter or for the
external potentiometer will be doubled.
B.I.L. Balancing
The balancing represents the centring offset, if set to 0 the fabric will be kept in the
centre, if set to a different value, the fabric will be kept on a side, at a distance
proportional to the BALANCING value set, the sign of the parameter indicates the
side on which the fabric will be moved.
The value set is compared to the run-out value indicated by the position of the
“rfB” sensor that can be viewed in “Secondary Parameters” page, this value is not
affected by the values set in parameters SENSITIVITY and BALANCING.
By setting this parameter to 0 the alarm threshold data will be diverted to analogue
input ALARM-AN on terminal M4.6.
The alarm relay will be enabled if the set threshold is exceeded for 1.4 seconds and
disabled if the value falls back within the set range for at least 0.14 seconds.
The default value 11 copies the reading speed set for the previous sensor, model
FLUO LM1 based on board C212.
It may be useful to increase the number of signal averages when the fabric is not
homogenous, but pay attention, because this modification also decreases the
reading speed and in some cases it might cause oscillations during adjustment.
1) The adjustment of the reading contrast is omitted and the power on the
transmitter tubes is forced to maximum.
2) The yellow LED LD17 blinks signalling the absence of the light contrast
adjustment (this LED can blink even before setting switch SW2.7 to ON if no
fabric is being processed)
3) The calibration data display window is enabled.
CAUTION: Before proceeding with the calibration of the sensor tube make sure
that:
A) At least 15 minutes passed from machine start-up in order for the sensors to
reach the operating temperature
B) The tubes are free from any fabric or other objects
C) Both tubes are cleaned
The calibration of the 4 signals (DARK, LIGHT, LEFT, RIGHT) coming from the sensor
tubes can be performed in 2 ways:
We recommend you to perform the calibration using the function “Sensor tube
calibration parameters display” because no external instruments are used (their
accuracy is not known) and the sensor is calibrated taking into consideration the
data detected by the board, compensating for any measuring error.
P1 P2 P3 P4
CAUTION: Before proceeding, carefully read the previous paragraph “Sensor tube
calibration procedure”.
1) For all measures, please keep the BLACK tip of the multimeter on TP6 “0V-
ADJ” and set the multimeter in Volt DC with 20Vdc full scale
2) By keeping the RED tip on TP2 “DARK” carefully adjust the trimmer P1 until
obtaining 9.00Vdc (+/- 0.01Vdc)
3) By keeping the RED tip on TP3 “LIGHT” carefully adjust the trimmer P2 until
obtaining 9.00Vdc (+/- 0.01Vdc)
4) By keeping the RED tip on TP4 “LEFT” carefully adjust the trimmer P3 until
obtaining 9.00Vdc (+/- 0.01Vdc)
5) By keeping the RED tip on TP5 “RIGHT” carefully adjust the trimmer P4 until
obtaining 9.00Vdc (+/- 0.01Vdc)
After adjusting the 4 trimmers make sure once again that the voltages on the 4
relative test points are 9.00Vdc (+/- 0.01Vdc), otherwise adjust them once again.
2) Insert the fabric between the sensor tubes, the width of the fabric should
exceed by at least 50cm the central part of the tube, not covered by sensors.
3) Make sure that when the fabric is present, the yellow LED LD17 remains off and
does not blink indicating that the light contrast adjustment is properly
performed, suitable for the type of fabric being processed.
4) Start the machine and check the operation of the centring device, if the result
of the adjustment is not appropriate, use the parameters or the external
potentiometers to adapt the SENSITIVITY and BALANCING.
______________________________________________ Page 12 of 13
CAUTION: Before proceeding, carefully read the previous paragraph “Sensor tube
calibration procedure”.
After setting the dip-switch SW2.7 to ON the system switches to “Sensor tube
calibration parameters display” function
By pressing keys“+” or “-“ you can view and scroll through the 4 calibration data
“DARK”, “LIGHT”, “LEFT”, “RIGHT”
d8.9 8.99
P
h9.0 9.00
+ - 9.00
L8.9 8.99
+ -
r9.0 9.00
The name of the channel along with the calibration data (displayed with only one
decimal) and the calibration data at maximum resolution (with two decimals) will
be alternately displayed, blinking.
1) “dx.x” alternated with value “x.xx” represents the calibration value of the
channel “DARK”, press“P” to keep the data display steadily lit (with two
decimals) and use trimmer P1 to adjust 9.00 (+/-0.01)
2) After calibrating the trimmer, press button “+” to start the calibration of the next
channel.
3) “hx.x” alternated with value “x.xx” represents the calibration value of the
channel “LIGHT”, press“P” to keep the data display steadily lit (with two
decimals) and use trimmer P2 to adjust 9.00 (+/-0.01)
4) After calibrating the trimmer, press button “+” to start the calibration of the next
channel.
______________________________________________ Page 13 of 13
5) “Lx.x” alternated with value “x.xx” represents the calibration value of the
channel “LEFT”, press“P” to keep the data display steadily lit (with two
decimals) and use trimmer P3 to adjust 9.00 (+/-0.01)
6) After calibrating the trimmer, press button “+” to start the calibration of the next
channel.
7) “rx.x” alternated with value “x.xx” represents the calibration value of the
channel “RIGHT”, press“P” to keep the data display steadily lit (with two
decimals) and use trimmer P4 to adjust 9.00 (+/-0.01)
8) Use keys “+” and “-“ to make sure that all 4 channels were calibrated,
otherwise proceed with their calibration using a data less accurate than 9.00
(+/-0.01).
5) set switch SW2.7 to OFF and make sure that the display returns to “Main display”
page
6) Insert the fabric between the sensor tubes, the width of the fabric should
exceed by at least 50cm the central part of the tube, not covered by sensors.
7) Make sure that when the fabric is present, the yellow LED LD17 remains off and
does not blink indicating that the light contrast adjustment is properly
performed, suitable for the type of fabric being processed.
8) Start the machine and check the operation of the centring device, if the result
of the adjustment is not appropriate, use the parameters or the external
potentiometers to adapt the SENSITIVITY and BALANCING.