Professional Documents
Culture Documents
Bio-Electronics & Measurements II Sheet (1) (Theoretical)
Bio-Electronics & Measurements II Sheet (1) (Theoretical)
Faculty of Engineering
Biomedical Engineering Department
a) The palpation method measures only the ……………………………………. pressure value of ABP.
c) A pressure of 30 mmHg; the equivalent H2O manometer indicate pressure of: ……………………….
…………………………………………………………………….
…………………………………………………………………….
…………………………………………………………………….
…………………………………………………………………….
e) When constant force causing the pressure in the system the pressure is said to be
………………………………………; whereas if the force is varying the pressure is said to be
………………………………………
f) The palpation method measures only the…………………………………………pressure value of ABP ;
whereas the flush method measures the……………………………………………..pressure value of ABP.
g) The zero reference pressure used for measuring a gage pressure
indicates……………………………………….absolute (atm) pressure.
h) Indirect method for blood pressure measurement use an instrument called…………………………….
i) A patient’s ABP is 140/80:
The 140 value indicates……………………………………………pressure.
The 80 value indicates……………………………………………..pressure.
|Page1
Q2: Answer the following:
a) Draw a diagram shows the relationship between cuff and blood pressure.
b) Regardless of the electrical or physical principles involved, direct measurement of BP is usually
obtained by one of three methods; state those methods?
c) Measurement of BP with a catheter using electronic transducer can be achieved in two ways; state
those ways?
d) List the four basic types of pressure amplifiers?
Q3:
The following is cross-sectional view of Arterial Blood pressure transducer; identify its marked
components:
1)………………………
2)………………………………………………………. 7)……………………………………………………….
3)………………………………………………………. 8)……………………………………………………….
4)………………………………………………………. 9)……………………………………………………….
5)………………………………………………………. 10)…………………………………………………….
|Page2