POWER PLANT LABORATORY
Power Plant
Laboratory
A Manual
1
Specifications and Start Up
2 Sampling of Lube Oils
3
Calibration and Maintenance
Introduction
Calibration strategies
Confidence tests
Cleaning the PPL and Tests Cells
Storage and transportation
4 Bunker Samplers
B User Instructions
5
Console
6
Heated Viscometer
7
Water in Oil Test (Reagent B / EasySHIP)
8
Insolubles Test
9
TBN Test
10
TAN Test
14 Pour Point
A Manual
MANUAL
Console
Console
Display:
8 digit LED type
Keyboard:
Membrane type with tactile buttons
Interfaces:
Measuring Cell socket with inductive power circuit and Infra
Red data link. Infra-Red data link on side of unit for viscometer.
RS232 data port on the rear of the console for data download to PC.
Memory:
Capacity to store 256 readings in non volatile storage
Clock:
Realtime clock with internal backup battery
Power:
110 to 240AC 50/60Hz 20VA
Fuse Rating:
SPECIFICATIONS
Heated Viscometer
Heated Viscometer
Operational fluid density:
870kgm-3 ≤ ρ ≤ 1000 kgm-3
Range:
20 - 810 cSt at 50°C (ISO Fuel Grades RMA10 to RML55)
20 - 810 cSt at 40°C (lube oils SAE 5 through SAE 50)
Test time:
Heating from 25°C: 10minutes
Viscosity at 40°C: 3 minutes (unheated)
Repeat test: maximum 30 seconds
Calculations:
Viscosity at 50°C or 40°C (heated)
Viscosity at 40°C (unheated, corrected to 40°C)
Viscosity at 100°C (calculated)
Calculated Carbon Aromaticity Index (CCAI)
Density correction from 50°C to 15°C in vacuo
Variable Viscosity Index (for unheated mode)
SPECIFICATIONS
Accuracy:
Typically with +/ - 3%(20 - 450 cSt) or +/- 2 cSt
Power:
110 to 240 VAC 50/60Hz 200VA
Fuse Rating:
2.5A 20mm 250VAC HRC A/S (T) Ceramic
Density MeterTest
Compatibility Compatibility
Salt/Fresh Water
Range: Range:
800 to 1010kg/m3 at 15°C (ISO As per ASTM D4740
8217 Fuel Grades DMA to RML55) Test time:
Test time: 20 minutes (unattended)
Heating from 15°C: 10 minutes. Accuracy:
Repeat test: maximum 30 seconds Variation of 1 rating in 20 repeat
Cleaning: 1 minute tests
Test temperature: Power:
Selectable 50°C or 70°C 110 to 250VAC Autoselected
Calculations: Fuse rating:
Density at 15°C in vacuo, centiPoise 2.5A 20mm 250VAC HBS A/S (T)
to centiStokes Calculated Carbon Ceramic
Aromaticity Index (CCAI)
Accuracy:
Typically with ±0.1%
(800 - 1010kg/m3)
Power:
110 to 250VAC Autoselected
50/60 Hz 200VA
SPECIFICATIONS
Fuse rating:
2.5A 20mm 250VAC HRC A/S (T)
Ceramic
SPECIFICATIONS
Sampling of
Lube Oils
Taking samples
l Sample directly from the
top of the main oil supply
reservoir. Always take
samples from the same point.
l Ensure that the total
quantity of oil in circulation
is approximately the same
during each sampling.
Additional sampling
Hydraulic systems: Middle of main reservoir or system return line.
Beware of breaking into high pressure lines.
SAMPLING OF LUBE OILS
Calibration and
Maintenance
Calibration Strategies
Check the viscometer oil temperature display, is this correct?
Is there any possibility of fuel dilution in the oil sample?
Could the reference oil sample be wrongly labelled?
Is the correct 23mm metal ball being used?
finished.
l A very high or low reading or a reading of zero indicates the
pressure transducer is damaged. (Check with another test cell that
the console can display a reading).
l The pressure transducer is similar to a human eye. It will last a
lifetime and can be gently rubbed. However, if it is pressed with
a sharp object or pushed hard with a finger when cleaning the
cell, it will break.
l Rolling some paper into a tube and gently pressing this on the
pressure transducer will cause the console reading to increase.
This provides confidence in the cell electronics and console
display.
POWER PLANT LABORATORY
General calibration:
3
l Using 25ml of Water Test Reagent, add 50ul of water with
micro pipette (1% contamination). Perform the Water Test and
expect a result between 0.8% and 1.2%
l Do not pour water into the cell and perform a test. This will
over-pressurise the transducer and damage the cell.
The test chemicals tend to degrease the thread so periodically lubricate the
cap to avoid it jamming after a test.’
TBN Cell
Confidence Test
The TBN Test functions by comparing the cell pressure generated by
the new engine oil (Reference Sample), to the cell pressure generated
by the used engine oil. As the engine hours build the engine oil TBN will
slowly deplete giving a lower cell reading. If the engine oil is topped up
CALIBRATION AND MAINTENANCE
with new oil the TBN will rise and generate a higher cell reading.
The cell reading for the used engine oil is higher than the reference
reading for the new oil
Check handbook for correct oil test quantity and check that the same
amount was used in both the reference and engine oil tests.
Has the RUT oil or reference type changed from that in the engine?
New engine oil could give slightly higher reading (<5%) due to
l Insert a reagent tube with an acid mixture (red*). (Add 1 ml TAN oil
to the green* sample tube). Check a reading of higher than 1000.
Reagent D is not green* (i.e. pink/red)
Absorbing Carbon Dioxide can alter the pH or alkalinity of this reagent over
time. Add one drop at a time of Reagent E or F (Potassium Hydroxide) to the
CALIBRATION AND MAINTENANCE
bottle of Reagent D. Shake after each drop until the colour just changes from
red* to green*.
The Reagent E or F can be added to the Reagent D in the sample tube to just
turn it green*, but this is not so easy to do.
Do not overdose the Reagent D. It only has to just turn green* and should not be
‘dosed’ to obtain dark blue/green*.
l Pour out the oil from the Viscometer into a container, using the sieve
to catch the metal ball. Place the sieve and ball on some tissue
paper.
l Wipe the open end of the Viscometer with tissue paper, then push this
into the tube with the cleaning rod. Be careful not to scratch or push
the Red Sensors at either ends of the tube.
Cleaning the Water Cell:
l Gently wipe out the Cell and cap with soft tissue paper.
Note: Do not press on the pressure transducer. Do not use water to
clean this Cell.
l Tip contents into a slops bucket. Wipe out the Cell and cap with soft
tissue paper.
l Wipe out the tube holder with a soft tissue if fluid is spilled.
Note: For a large spillage leave cell inverted on tissue paper to allow
spillage to drain and then wipe clean. Do not use solvents in the tube
holder or immerse the cell.
Bunker Samplers
BUNKER SAMPLERS
Installation
The ISO fuel specification ISO 8217 refers to the place of custody
transfer, which is normally at the vessel’s manifold. In other words,
at the vessel’s end of the bunker delivery hose.
However, it is often more practical to use the sample taken by the
supplier at the barge end of the hose.
Provided that the sample is witnessed throughout the bunkering,
then properly mixed and split, it doesn’t matter which end of the
delivery hose is used.
Delivery
pipe
to
bunker
manifold
540mm
working
BUNKER SAMPLERS
20mm
Direction
of flow from
delivery
to storage
12.5mm dia
l If the Cubitainer fills during the bunkering period, remove it and add
a numbered tamper evident cap.
Place an empty Cubitainer on the sampler and continue to draw a
sample.
l Some delivery conditions can cause a vacuum in the line e.g. suddenly
stopping the pumps on the barge. Close the sampler valve in these
circumstances or the sample will be drawn back into the delivery line.
If a sample is required for on board testing only, then the sample can be
USER INSTRUCTIONS
Console
CONSOLE
The console auto selects the power supply to suit the mains
voltage. No settings are required.
l Turn on the mains supply and then turn on the power supply using
the switch next to the socket.
PIC
On/off switch
Mains socket
CONSOLE
Cell socket:
This socket features our unique inductive coupled power supply.
The cells get their power when placed upon the socket via an
electromagnetic link. The measurement data is transmitted to the console
via an infra red optical link built into the socket. This enables the cells to be
designed without wires, batteries or connectors so they are very rugged.
RS232 port:
The console has the ability to log results. These can be downloaded to
a PC via a standard RS232 data connection. The Kittiwake download
software is required to perform this operation. This is available from
Kittiwake at www.kittiwake.com
RS232 Port
Cell Socket
Reset
Key
Mode Key
Temperature
Zero Key
Up or down
Arrow keys Return Key
Speed Key
CONSOLE
Cleaning
Disconnect the power lead. Wipe console with tissue or a soft rag.
A small amount of reagent A on a cloth will remove marks.
Maintenance
If the unit fails to power up disconnect the power lead and check
the fuses are OK. We have supplied replacment fuses of the correct
type in the spares pack. Do not use any other type of fuse.
The Real Time Clock has a battery internal to the console. This
should last approximately 5 years and is automatically replaced
if the console is returned to the factory for re-calibration and
refurbishment. It is not recommended that you replace the battery
in normal use, but if the Real Time Clock stops operating and you
are using the data logging facility, the battery can be replaced with
button cell type CR1620.
The operation of each mode is described in the section for that test. Some
of the modes have extra options such as reporting the results in a different
CONSOLE
manner. These are accessed using the arrow keys and are described in the
section on each test.
l Pressing the return key again will confirm that the result set
is to be stored. Any other key will cancel the store
operation.
l Select Upload.
l The ID and hours will now be saved with all result sets until they
are altered.
Each error has been assigned a unique code and this code may be
interpreted using the table below:
Heated Viscometer
HEATED VISCOMETER
Before connecting the power supply to the mains check that the correct
supply voltage is selected. Failure to do this may damage the instrument.
Remove fuse box using a Select correct voltage Replace box, ensuring
screwdriver. supply. white arrows are aligned.
l Connect the yellow power supply lead to the viscometer and lock into
place using by rotating the outer collar.
l Connect the mains lead into the socket on the side of the power
supply, checking the correct mains voltage is selected. Plug the
HEATED VISCOMETER
l Turn on the mains power and then turn on the power supply
using the switch next to the socket, the green light should come on and
the viscometer display should illuminate.
Location
Make sure the power supply is located towards the rear of the
workbench where the cable cannot be caught during operation.
HEATED VISCOMETER
There is an internal circuit that controls the heating of the oil to allow
measurements to be taken on higher viscosity oil. The display will
request a ‘TILT’ for measurement only when the temperature is stable to
ensure accurate results. The processing circuitry compensates for the
temperature of the oil and allows display of Centistokes adjusted to 40,50
or 100 oC. There is a calculator feature, which allows the reading to be
adjusted for Density and Viscosity Index for improved accuracy and to
display the CCAI.
End cap
Reset key
Return key
Mode key
Arrow key: up
Speed key
Connection for power Arrow key: down
supply (opposite side)
To change fuse
HEATED VISCOMETER
Remove fuse box using a Carefully remove and Replace box, ensuring
screwdriver. replace fuse. white arrows are aligned
to correct voltage supply.
l Turn off the power supply and disconnect the yellow lead from the
viscometer. Carefully open the end cap to empty the oil out of the
tube. CAUTION: The oil may be hot (50 oC).
l Using the sieve supplied to catch the metal ball, tip the oil out into
a container.
l Using the rod supplied push a wad of clean tissue down the centre
of the tube ensuring all remaining oil is cleaned out. Replace the
metal ball into the tube and fit end caps for safe keeping.
Make sure the power supply is disconnected from the mains. Wipe down
the instrument with a clean dry soft cloth. Do not immerse in water, if
necessary to remove stubborn marks use a cloth soaked in warm soapy
water.
If the unit fails to power up disconnect the power lead and check the fuse
is OK. We have supplied replacement fuses of the correct type in the
spares pack. Do not use any other type of fuse. There are no other user
serviceable parts inside the unit, if the unit still does not operate return to
the supplier for repair.
l Fill the tube with the oil sample to just above the V-plate. Slowly push
in and rotate the plug with the valve open until all air is expelled.
Do not use excess force.
l Loosely fit the bleed valve and place the Viscometer in an upright
position. Leave to stand for 5mins
HEATED VISCOMETER
l Tighten the bleed valve and wipe off excess oil from the plug. Place
Viscometer back on its base and connect the power cable.
The Viscometer is now ready for use but do not tilt yet.
Note: If air is not fuly expelled from the unit, it may affect results
V-plate
HEATED VISCOMETER
Slowly push in
and rotate the end cap
until all air is expelled.
Tighten the bleed valve.
2 00
125.4cS
HEATED VISCOMETER
indicator
cSt at 100°C 12.54cS
l Select Mode 1.
Use Speed and Arrow keys to input the oil Density
(i.e. Density at 15° in vacuo).
l Select Mode 1.
Use Speed and Arrow keys to input the oil Density
(i.e. density at 15° in vacuo).
The Density of many Synthetic Oils is nearer to 1.0 than 0.9.
l Select Mode 5.
Enter the approximate Viscosity Index (VI).
The VI of Multigrade Oils will be higher than the default VI of 100.
HEATED VISCOMETER
Density = 0.9
1c .900
5 xxx
125.4cS
Density = 0.9
1c .900
Density = 0.992
1c .992
5 126
132.5cS
1c .900
1u .970
1c .992
1c .992
1u .970
3 133
HEATED VISCOMETER
3 147
4 864
Note: Different Cells are required for the 2.5% and ppm ranges.
l Place the Cell on a level surface and remove the end cap.
Add Reagent A up to the internal lip (20ml).
Add 5ml of the test oil using the syringe provided.
CAUTION: Take care not to inhale the powder and avoid contact
with the eyes.
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
Note: The test will not work unless shaken very thoroughly,
especially in cold temperatures.
WATER IN OIL TEST
4 C 1
% v/v 4 0 . 3 3
Press the Up Arrow to select Mode 4.1 (the correct Mode for using the
PPM Water Cell).
Note: Use the Arrow keys to toggle between Modes before the test
starts Different Cells are required for the 2.5% and ppm ranges.
l Place the Cell on a level surface and remove the end cap.
Add Reagent A up to the internal lip (20ml).
Add 5ml of the test oil using the syringe provided.
CAUTION: Take care not to inhale the powder and avoid contact
with the eyes!
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
Note: The test will not work unless shaken very thoroughly,
WATER IN OIL TEST
l The display will read directly in parts per million water in the oil
(6000ppm range) Note:The two ranges require different cells and it
is not possible to toggle between the calibrations.Variation on
repeat reading on the 6000ppm cell of +/- 50ppm is quite normal.
POWER PLANT LABORATORY
7
0-6000 ppm
4.1 -————
4.1 C 1
WATER IN OIL TEST
ppm
4.1 480
POWER PLANT LABORATORY
7 Water in Oil Test (Reagent B) 0-3000 ppm
Silver Cell
Only suitable for oils <120 cSt @ 40°C
To improve the accuracy of the 0-3000 ppm test, it is possible to
compensate for the residual water in Reagent A.
Note: Use the Arrow keys to toggle between Modes before the
test starts.
l Place the Cell on a level surface and remove the end cap.
Add 5ml of the Reagent A using the syringe provided.
Add 20ml of the test oil using the syringe provided (4 x 5ml).
Note: Note the quantities are different from the other water
tests
CAUTION: Take care not to inhale the powder and avoid contact
with eyes!
l
Press the Zero button to zero the reading and start the test.
% v/v 4 —————
0-3000 ppm
4.2 —————
(Gloves not shown for clarity)
4.2 C 1
Note: The test will not work unless shaken very thoroughly,
especially in cold temperatures.
The reading will stabilise giving the result for the range
0 - 3000 ppm.
WATER IN OIL TEST
Note: Different Cells are required for the 2.5% and ppm ranges.
l Place the Cell on a level surface and remove the end cap.
Add Reagent A up to the internal lip (20ml).
Add 5ml of the test oil using the syringe provided.
l Place the Agitator into the cell. Replace the cap tightly.
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
Note: The test will not work unless shaken very thoroughly,
especially in cold temperatures. EasySHIP Test Reagent must
be above 18°c
WATER IN OIL TEST
% v/v 4 —————
(Gloves not shown for clarity)
4 C 1
% v/v 4 0 . 3 3
l Place the Cell on a level surface and remove the end cap.
Add Reagent A up to the internal lip (20ml).
Add 5ml of the test oil using the syringe provided.
l Place the Agitator into the cell. Replace the cap tightly.
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
l The display will read directly in parts per million water in the oil
(6000ppm range)
0-6000 ppm
4.1 -————
4.1 C 1
ppm
4.1 480
POWER PLANT LABORATORY
75
Water in Oil Test (EasySHIP) 0 – 6000 ppm
Green Cell
l Select Mode 4, Water in Oil 0 - 2.5% (Standard Cell)
The display will read as shown.
Press the Up Arrow to select Mode 4.1 (the correct Mode for using
the 0 - 6000 ppm Water Cell).
Note: Use the arrow keys to toggle between modes before the test
starts. Different cells are required for the 0 - 2.5% and 0 - 6000 ppm
ranges.
l Place the Cell on a level surface and remove the end cap.
Add 30ml of Reagent A using one of the syringes provided.
Add 3ml of the test oil using a different syringe.
Note: The quantities are different from the other water tests
l Place the Agitator into the cell. Replace the cap tightly.
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
NB: The test will not work unless shaken vigorously, especially
in cold temperatures. EasySHIP Test Reagent must be above 18°C
WATER IN OIL TEST
4.1 C 1
l Place the Cell on a level surface and remove the end cap.
Add 35ml of Reagent A using one of the syringes provided.
Add 10ml of the test oil using a different syringe.
Note: Note the quantities are different from the other water tests
l Place the Agitator into the cell. Replace the cap tightly.
% v/v 4 —————
0-3000 ppm 4.2 -————
(Gloves not shown for clarity)
4.2 C 1
l Remove the cell from the console and shake until the display
reaches 120, then replace on the console.
Insolubles Test
INSOLUBLES TEST
Take the Insolubles tube and fill with Reagent J to the fill line
(10ml).
Gently insert the Insolubles tube into the test cell. A short length of
tube will protrude from the test cell.
l Rotate the Zero knob on the side of the cell fully anticlockwise
and then slowly clockwise until the reading reaches zero
(overshooting reads [- - - -] or 9.9).
The display will read as shown.
% w/w 2 — — — — —
% soot
index 2 .1 ———
2 X.X
2.1 X.X
2 0.0
2.1 0.0
INSOLUBLES TEST
% w/w
2 1.8
% soot
index 2.1 0.9
TBN Test
TBN TEST
l Place the cell on a level surface and remove the end cap.
Shake the bottle of Reagent C and attach the pourer spout.
Add Reagent C to the TBN Cell until the level reaches the
internal lip inside the cell bowl. Add test oil:
[New Oil TBN 2 - 20] add 10 ml Test Oil
[New Oil TBN 21 - 40] add 10 ml Test Oil
[New Oil TBN 41 - 100] add 2.5 ml Test Oil
Replace end cap and tighten.
Note: The test will not work correctly unless the test cell is
shaken very thoroughly, especially in cold temperatures.
Determination of the Test Value (used oil)
Routine test procedure for used oils using the Reference value from
your log book (see earlier in this test procedure).
l Repeat the Reference Value Procedure, but with the same quantity
TBN TEST
3 C 1
Reference
Value 3 XXX
TBN TEST
Test
Value 3 XXX
POWER PLANT LABORATORY
9 Calculating the Used Oil TBN
Example:
New Oil TBN = 40
Reference Value = 140
Test Value = 93
l Start by testing the used oil to get the Test Value as described
on the previous page.
l Press Enter.
l The display will ask for the Reference Value, use the Up/Down/
Speed keys to display this value.
l Press Enter
l The display will ask for the New Oil TBN, use the Up/Down/Speed
keys to display this value.
40
40
TBN TEST
TAN Test
TAN TEST
l Shake the oil sample thoroughly. Remove 1ml using a 1ml syringe,
and add this to the TAN tube. Replace the cap and shake well
(if the oil has any TAN the colour will change to red). Allow any
bubbles to settle. Return tube to the TAN cell. The reading will be
higher than zero if there is any TAN.
l Press Return and enter the syringe reading using the Up/Down
arrows and the Speed key.
(The greater the volume of Reagent E used, the lower the syringe
reading).
Mode 5 - IP177 (ASTM D 664) used for all other mineral oils.
TAN TEST
Mode 5.1 - SAE ARP 5088 (modified IP139, ASTM D 974) used for
aviation turbine and synthetic oils.
Toggle between the two by pressing the Arrow keys.
5 ——— ——
5 .1 ——
(Gloves not shown for clarity)
5 XXXX
5 0
5 X X X X
5 5 .05
IP177 5 X.X
TAN TEST
IP139 5.1 X.X
l Shake the oil sample thoroughly. Remove 1ml using a 1ml syringe,
and add this to the TAN tube. Replace the cap and shake well
(if the oil has any TAN the colour will change to red). Allow any
bubbles to settle. Return tube to the TAN cell. The reading will be
higher than zero if there is any TAN.
l Press Return and enter the syringe reading using the Up/Down
arrows and the Speed key.
(The greater the volume of Reagent F used, the lower the syringe
reading).
l Press Return and the display will show the TAN (0 to 3 mg KOH/g)).
Mode 5.2 - IP177 (ASTM D 664).
TAN TEST
5 ——— ——
5 .1 ——
(Gloves not shown for clarity)
5 . 2 X X X X
5 . 2 0
5 . 2 X X X X
5.2 5 .05
TAN TEST
IP177 5.2 X.X
Start point
Add Reagent D
Add Reagent E
or F drop by drop.
Density Meter
DENSITY METER
CAUTION: The Density Meter must not be used for heating any other
liquids such as water, cleaning fluids, or any fuels with a low flash
point temperature (e.g. gasoline).
Note: These instructions apply to density meters with software version 2.xx
The meter auto selects the heating and power supply to match the supplied
mains voltage. This is displayed on the screen at power up.
240ac
DENSITY METER
l Turn on the mains power and then turn on the power supply using
the switch next to the socket, density meter display should
illuminate showing the mains voltage detected.
On/off switch
Fuse
Lead to mains
supply
Power Socket
Location
Pouring spout
Display
Mode key
Reset key Return key
Arrow key: up
Arrow key: down
Speed key
To change fuse
DENSITY METER
Open lid using screwdriver. Carefully remove fuses. Replace and secure lid.
l Carefully pour the oil into a container using the pouring spout.
CAUTION: The oil will be hot (up to 70 oC ). Pour out the contents
carefully. Using the rod supplied use a wad of clean tissue down the
centre of the tube ensuring all remaining oil is cleaned out.
General Cleaning
Make sure the power supply is disconnected from the mains. Wipe down
the instrument with a clean dry soft cloth. Do not immerse in water, if
necessary to remove stubborn marks use a cloth soaked in warm soapy
water.
Maintenance
If the unit fails to power up disconnect the power lead and check the fuse is
OK. We have supplied replacement fuses of the correct type in the spares
pack. Do not use any other type of fuse.
There are no other user serviceable parts inside the unit, if the unit still
does not operate return to the supplier for repair.
l Centigrade digit “ ° “ flashes as the Density Meter heats the oil.
Stir the oil occasionally as it heats up with the stirring rod provided.
When the temperature stabilises at 50°C the meter is ready for
operation, (this takes about 10 minutes).
Testing
l Read the hydrometer at the Hydrometer shaft
meniscus.
Use Arrow and Speed keys
in Mode 1 to enter the oil
uncorrected [1u] Hydrometer Read at this point The meniscus
reading at 50°C. where the level
cuts the scale
usually drops down
0.01 graduations
l Press Return.
DENSITY METER
38.6°
3 9 . 0
1u .980
1u .978
DENSITY METER
1c .999
2 280
2 273
3 864
‘P’ indicates
centiPoise option 2P 273
DENSITY METER
Viscosity corrected to
centiStokes 2 279
Compatibility Test
COMPATIBILITY TEST
Intended Use
The Compatibility Oven is use to pre-heat a sample and then dry sample
test paper prior to analysis. It contains chamber heated to 100 oC. It must
not be used for heating any other materials.
The meter auto selects the heating and power supply to match the supplied
mains voltage.
COMPATIBILITY TEST
l
Connect the mains lead into the socket on the side of the power
supply.
l Turn on the mains power and then turn on the power supply using
the switch next to the socket, the timer display should illuminate.
Fuse
On/off switch
Location
COMPATIBILITY TEST
Oven lid
Fuse Reset button
On/off switch
Sample tube
To change fuse
COMPATIBILITY TEST
CAUTION: The oil samples and oven will be hot (up to 100 oC).
Handle with care.
l When cool remove the sample tubes and dispose of oil and
tube.
General Cleaning
Make sure the power supply is disconnected from the mains. Wipe down
the instrument with a clean dry soft cloth. Do not immerse in water, if
necessary to remove stubborn marks use a cloth soaked in warm soapy
water.
Maintenance
If the unit fails to power up disconnect the power lead and check the fuse
is OK. We have supplied replacement fuses of the correct type in the
spares pack. Do not use any other type of fuse. There are no other user
serviceable parts inside the unit, if the unit still does not operate return to
the supplier for repair.
l With the unit fully heated to 1000C, place the tube in one of the
preheat holes. Press Reset and allow the timer to count to 20
(takes 20 minutes).
l Remove the paper and compare the spot against the Spot Reference
Chart (see following pages).
Unit
fully heated 0 0
2 0
Timer reset to
0 minutes 0 0
2 0
COMPATIBILITY TEST
l With the unit fully heated, place the tubes into two of the preheat
holes. Press Reset and allow the timer to count to 5.
l Remove the paper(s) using the tweezers and compare the spot
against the Spot Reference Chart on the next page.
Unit fully
heated 0 0
0 5
Mixture
0 0
1 5
Fuel A Fuel B
0 0
2 0
COMPATIBILITY TEST
Compatible / stable.
Incompatible / unstable.
5
COMPATIBILITY TEST
Incompatible / unstable.
Timer
The equipment is provided with a small display
that will indicate as follows:
Fresh Water.
SALT/FRESH WATER TEST
The Water and TBN Cells have been safety tested to more than 20 times
working pressure. However it is essential that the operator does not
exceed the recommended volumes of reagents or test oils in each cell.
Both Water and TBN Cells are designed to vent excess pressure and
prevent spray before the cap can be fully removed. If the cell is operated
beyond its design pressure then simply slacken the cap to release the
pressure. Hold the cell upright when slackening the cap to prevent any
liquid from being sprayed out as the pressure escapes.
The end plug and bleed valve assembly in the Viscometer are designed
to slide. If cold oil is left in the Viscometer and allowed to expand, there is
danger of an hydraulic lock but damage is prevented by allowing the end
plug to be pushed out.
Chemicals
CHEMICALS
Fire and explosion Hazard data: This describes the flash point, auto ignition
temperature and flammability in air.
The procedure for fire fighting is covered
as are the particular fire and explosion
hazards.
Parker Kittiwake
Hydraulic Filter Division Europe
3 - 6 Thorgate Road Littlehampton
West Sussex BN17 7LU United Kingdom
Tel: +44 1903 731470
Fax: +44 1903 731480
Email: kittiwakeinfo@parker.com
MA-K11152-KW Issue 13 Web: www.kittiwake.com www.parker.com