This document is the property of Dubai Municipality and its issue is controlled. No part of this document
should be reproduced or copied without the prior written permission of Dubai Central Laboratory Department.
Dubai Municipality Standard المواصفـات القيـاسية لبلـدية دبـي
DMS 2008: 2010
This copy is issued for “MODERN SURVEYING CALIBRATION & TESTING LABS” use only. No other entity entitled to use this copy.
Table of Contents
Page
Foreword ii
1 Purpose 1
2 Scope 1
3 Warnings and safety precautions 1
4 Definitions 2
5 Principle 4
6 Acceptability criteria 5
7 Reference standard and calibration equipment 5
8 Unit under calibration 5
9 Measurement traceability 5
10 Quality control checks 5
11 Calibration method 6
12 Computation of results 10
13 Calibration frequency 11
14 Results and uncertainty of measurement 11
15 Reporting process including expression of results 11
16 Method validation 12
17 Records 12
Publications referred to 13
Foreword
With a view to have a comprehensive set of Dubai Municipality Standards which would be
consistent and fulfill local needs and yet be at par with International requirements, Dubai
Central Laboratory is developing standards taking guidance as much possible and required
from International and Regional Norms.
This Dubai Municipality Standard (DMS) describes a Standard Operating Procedure useful
for Calibrating Drying Oven / Testers.
The standard has been prepared under the scheme of standardization of calibration/test
procedures for use of DCL laboratories and by other interested persons/organizations. The
standard is issued after validation.
This procedure combines essential features and specification requirements of International
practices. It has been developed in such a way as to provide as much information about the
operating characteristics of the measurand. Attempts have been made to make the contents
user friendly.
1 Purpose
To standardize the method to be used for carrying out calibration of different types of
electrically heated ovens.
2 Scope
This test method is applicable for the calibration of general-purpose electrically heated
and temperature controlled laboratory ovens operating under normal air environment
inside the work place.
• Range: The method is applicable for gravity convection or forced ventilation
ovens designed to operate over full or part of range from 40°C to 200°C.
• Types: Following types of air ovens shall be identified IA, IB, IIA, IIB,
satisfying ASTM E 145 – 94 standard requirements where as IIC and IID are
identified to meet local needs. The technical criteria are defined in clause 4.15
to 4.20.
• Uncertainty > ± 0.2ºC (k =2).
4 Definitions
4.1 Calibration
The set of operations, which establish, under specified conditions, the relationship
between values indicated by the measuring instruments or values represented by
material measure and the corresponding known values of the measurement used.
4.2 Reference standard
A standard generated, generally of the highest metrological quality available at a
given location, from which measurements at that location can be derived.
4.3 Traceability
The property of a measurement whereby it can be related to appropriate standards,
generally National or International standards through an unbroken chain of
comparisons.
4.4 Critical equipment
Equipment, which if not calibrated will affect the quality of the intended result.
4.5 Gravity convection oven
An oven, that is ventilated through gravity convection whether or not they are fitted
with internal fans to assist the circulation of air.
4.6 Forced ventilation oven
An oven which is ventilated with the use of a ventilating fan which forces draught air
into the oven and controls the circulation in a definite pattern.
4.7 Working space
That portion of the internal space, which is not less that 50 mm from any wall
including the roof?
4.8 Oven temperature
Rev. (01) 2 Issue Date 06/12/2010
Dubai Municipality Standard المواصفـات القيـاسية لبلـدية دبـي
DMS 2008: 2010
This copy is issued for “MODERN SURVEYING CALIBRATION & TESTING LABS” use only. No other entity entitled to use this copy.
The temperature measured at the approximate centre of the working chamber of the
oven.
4.9 Ambient temperature
The temperature of the air inside the room where the oven is being tested as
measured by a thermometer placed at least 1 meter from the oven.
4.10 Oven indication
The reading of the temperature indicating devices, which is an integral part of the
oven.
4.11 Temperature variation
The difference at any moment between the temperature at the centre of the working
space and at any other point in the working space.
4.12 Temperature fluctuation
The short term changes in temperature at any point in the working space.
4.13 Time of recovery
The time required for the oven to regain its set temperature after opening for 1
minute and closing again.
4.14 Temperature overshoots
Any excess of actual over desired oven temperature during a heating-up period.
4.15 Type IA oven
An oven ventilated by gravity convection whose maximum temperature variation
value shall be within ±2% of the difference between oven and ambient temperature
and maximum temperature fluctuation within ±2°C.
4.16 Type IB oven
An oven ventilated by gravity convention whose maximum temperature variation
value shall be within ±5% of the difference between oven and ambient temperature
and maximum temperature fluctuation of within ±2°C.
5 Principle
Error shall be determined by comparison of oven temperature condition in various
locations using sensitive, compatible and calibrated PRT sensors with respect to the
temperature indication of the UUC. The error shall be obtained from average value of
several readings.
6 Acceptability criteria
6.1 The oven shall be tested for compliance to the following requirements:
6.1.1 Error of indication: within ± 2% of set point.
6.1.2 Time of recovery: 1°C within 20 minutes at 100°C or nearest.
6.1.3 Temperature overshoot: maximum 1°C at set point 100°C.101°C shall not last
longer than 10 minutes. The peak of overshoot shall not reach 103°C.
6.1.4 In specific cases, the criteria shall be based on manufacturer’s recommendation or on
the requirement of the customer.
9 Measurement traceability
Calibrated multi sensor digital thermometer (PRT probes) and the data logger shall be
used. They shall be traceable to International standards. The calibration shall be
performed in-house or by an accredited laboratory.
Stopwatch shall be calibrated in house to the National reference of time (Etisalat time
clock).
10.1 Visual checks on digital thermometer (PRT sensors) for any damage /fault.
10.2 Operational checks on Digital temperature scanner, data logger and PRT/RTD, before
the items are taken outside the DCLD calibration premises.
10.3 Ensure that the probes are connected to the same channels of the data taker as originally
calibrated. For this purpose check the probe number and corresponding channel
number. Both must match.
10.4 The calibration status shall be checked before and after every calibration so as to avoid
use of expired probes.
10.5 The calibration status shall be checked by experiment once a month or in case of doubt.
Such confidence checks shall be carried out at 100 º C (approximately).The difference
shall not exceed 0.2º C.
10.5.1 Annual checks shall be made by resistance measurement.
10.5.2 The spread sheets shall be reviewed, updated once a year or earlier in the event of any
change in calibration status of probes/data taker/bath(s)
10.6 Pre-calibration checks on UUC
10.6.1 Check the oven functions for normal condition, e.g. the setting knob, door is not loose.
10.6.2 Check the ventilation type and condition: Is it opened. Partially opened /closed.
10.6.3 Check for provisions on the oven’s body for sensor probe insertion.
10.6.4 Ensure that sufficient space and fasteners / holders are available for mounting the
probes.
10.6.5 Before starting calibration, ensure that inside compartment of the oven is clean and free
of any foreign object and all the shelves are in position and vents open
10.6.6 Check the electric cables/ connections for any leak/loose/naked un-insulated surface or
connected by long extension cables.
11 Calibration method
11.1 Place the temperature sensors in nine locations inside the compartment of the oven.
Place one sensor at each of the eight corners approximately 50mm from each wall.
Place the ninth sensor in work-space which is within 25mm of the geometric centre
of the oven. The probe locations are shown in following figure (Fig – 1)
In some situations the number of probes and their configuration can vary according
to specific requirement of a particular oven. The basis shall be manufacturer’s
recommendation or customer’s requirement.
Back Side
3 4
1 2
7 8
5 6
of the oven and the reading of the sensor located at the centre of the working space at
each set temperature. However, it will not be required where the knob is not
graduated in terms of temperature units.
Readings shall be recorded at intervals of four minutes for a period of 1 hour
approximately. The error of indication of the oven at the particular temperature
setting is equal to the average of the oven indication readings minus the average of
the actual oven temperature as measured by the sensor located at the centre of the
oven.
While computing error, apply correction to the reference probe reading, which may
be available from the calibration certificate of the probe.
Depending on the intended use of the oven, accuracy of indication may be tested at
other temperature settings as well.
11.6 Temperature variation
With the temperature sensors placed as per Cl. 11.2, temperature variation shall be
measured by monitoring the readings at all points inside the oven. readings from the
nine sensors as well as the ambient temperature shall be taken every 4 minutes for a
period of 1 hour approximately. The maximum difference between the reading at the
centre of the oven and any other points shall be determined from the data and
reported as maximum temperature variation.
11.7 Temperature fluctuation
Temperature fluctuation shall be determined from the data as gathered in Cl. 11.6.
The fluctuation at any point is equal to the difference between the maximum and the
minimum reading at that point during the 1-hour period approximately.
11.8 Time of recovery
They shall be determined at set point 100°C or nearest. Using a stabilized
temperature setting of approximately 100°C, the oven door shall be opened fully for
1 minute and then closed. The time required for the centre temperature to regain its
initial stabilized temperature shall be measured using a stopwatch. The time required
for the oven to reach within 1°C below the original steady set temperature before
opening the door shall be recorded and reported as time of recovery. On opening the
door, the temperature shall start falling. On closing the door it shall start rising. The
time taken from steady 100°C to regain 99°C shall constitute the recovery time. To
determine the recovery time, readings at 10 s interval shall be noted till 99°C is
regained. See Fig -2.
11.9 Overshoot
Overshoot gives an estimate of the time constant of the oven. It shall also be
determined at or nearest to 100°C.
Measurement of overshoot shall follow of the procedure of recovery as described in
Cl. 11.8. After the door of the oven is closed, the chamber temperature shall start
rising. It will continue to rise till a peak is attained and then start falling back to
return to the initial set temperature (100°C). The rise of temperature above the set
value shall be the measure of the overshoot temperature and the time of its recovery
to as the overshoot time.
The peak value shall remain below 103°C. The over shoot temperature shall be 1°C
above the initial set temperature. The overshoot time shall not exceed 10 minutes.
In other words the recovery of 101°C shall be within 10 minutes. Details are shown
in following Figure -2.
Temperature °C
<10min (Overshoot Time)
<103
101
100
99
Temp
Example: The set point is 100°C. The overshoot time shall be the time during which the temperature
returns to 101°C.
12 Computation of results
As specified in above paragraphs Cl .11. 5 – 11 .9
• Temperature fluctuation: All readings collected from various probes and
time intervals shall be corrected by importing the ‘error” from calibration
certificates.
• Temperature variation: All probe readings shall be corrected for error as
in 12.1 above.
• Oven temperature error: Ensure that the corrected value of the reference
probe is used for computing the error of the oven chamber.
• The Recovery time and Overshoot time shall be determined as stated above
under respective head.
IMPORTANT NOTE: While recording above data, care shall be taken to identify
the channel of the data logger and the probe number so as to avoid import of any
wrong error data which is likely to enter due to mis- order of the probes –vis –a-
vis- allotted channel.
13 Calibration frequency
Calibration frequency shall be specified only on customer’s request. It shall not
exceed 12 months in normal circumstances. However, any alteration, modification,
replacement of parts readjustment or any source which may jeopardize calibration
status, shall need re-calibration.
Publications referred to