S KLINOSKOP H

Service Manual

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Main Index

KLINOSKOP H
4W 5245 0420723001

1. Pre - Installation

2. Installation

3. Circuit Diagram

4. Circuit Description

5. Service Fault Finding / Repairs

6. Spare Parts

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

List of Pages

010 - 1 List of Pages

010 - 2 Typical Layout of X - Ray Room

010 - 3 Power Supply Requirement

010 - 4 General Guidelines

010 - 1 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation
Typical Layout Of X Ray room

4
950 mm

12
N
Y
3

CONTROL
6

E
P
O
8

11

A
7
D
5

9
2

C
1

B M
X

010 - 2 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

Recommended Room Dimensions

A = 5030 mm ( 16 ‘ - 6 “ )
B = 4270 mm ( 14 ‘ - 0 “ )
C = 2370 mm ( 7 ‘ - 9 “ )
D = 1220 mm ( 4 ‘ - 0 “ )_
E = 1980 mm ( 6 ‘ - 6 “ )

Minimum Room Height

(Measured below the wooden beam to floor level)

i) Without Sirecon I = 2590 mm (8 ‘- 6 “ )


ii) With Sirecon I = 2740 mm (9 ‘- 0 “ )

Legend w.r.t Layout

1* Main Switch 1524 mm (5 ‘- 0 “ ) above floor.


2* Mains contactor 1524 mm (5 ‘- 0 “ ) above floor.
3* Push button station
4 RED Radiation indicator above door level.
5 Phase Transformer ( Static Balancer )
6 Control Console
7 Diagnostic Examination Table
8 High Tension Transformer
9 Chest Stand
10 Protective Screen - Lead lined.
11 Orientation point of examination table.
12 Wall thickness should be 230 mm (9 “) thick.

NOTE : 1 * As per recommendation depending on the X - Ray Generator.


2 Layout plan not to scale.

010 - 2 Pg. 2 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

Power Supply Requirement

Schematic Layout

E RSTN

ELECTRIC ELECTRICAL
MOTOR WIRING
FOR
X-RAY
MACHINES
EARTH
WIRE
12
S.W.G.BARE
COPPER

MAINS
SWITCH
FUSE
UNIT(1)
EARTH
WIRE
12
S.W.G.BARE NO
COPPER PUSH
BUTTON
STATION(3)
MAINS NC
CONTACTOR(2)
EARTH
WIRE
12
S.W.G.BARE
COPPER

STATIC CONTROL EXAMINATION


BALANCER(5) CONSOLE TABLE

EARTH
WIRE
12
S.W.G.BARE
COPPER

010- 3 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

Power Supply Requirement

Sr. Generator Power Line Power Mains Switch Mains Contactor Push button Mains Wire Size Earth
mA in Use Requirement Capacity Requirement Requirement station SWG or Sq. Mm Wire Size
requirement in Copper
1 300 mA 415 V , 50 Hz 40 KVA Siemens Make Siemens make Siemens make 7/16 or 10 12 SWG
Phase to Power Guard SICOP Power Push Button Bare
Phase Fuse Switch Unit Contactor of 63 Station Type copper
type 3KL5037 - A 3TF47 O2- 3SBO8 O2-8AC
1YA100 with HRC OA..ZAO1 with or Equivalent
Fuse type two “SIGUT” No.
3NA1018 of 63 A 3TX7 460-OE or
or Equivalent Equivalent
2 500 mA Same as above 65 KVA Same as above Same as above Same as above 19/18 or 12 Same as
above
3 800 mA Same as above 100 KVA Siemens Make Siemens make Same as above 19/18 or 25 Same as
Power Guard SICOP Power above
Fuse Switch Unit Contactor of 110
type 3KL5237 - A 3TF50 O2-
1YA100 with HRC OA..ZAO1 with
Fuse type two “SIGUT” No.
3NA1010 of 125 A 3TX7 460-OE or
or Equivalent Equivalent

010 – 3 Pg. 2/2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

General Guidelines

Civil Work for Installation of X - Ray Machines

1 All the wall of the X - Ray examination room will have to be 9 inches (230 mm) thick for radiation
safety.. If the walls are of 4.5 inches ( 114 mm ) thickness , they will have to be plastered with
barium sulphate mixed with cement in proportion of 5 : 1 ( 5 parts of cement : 1 part of barium
sulphate ) . For further details contact BARC, Trombay.

2 The main entrance door of the examination room to be lead lined with 0.5 mm lead sheet.

3 A seasoned wooden beam of size 6 inches X 2.5 inches ( 150 mm X 65 mm ) to be fixed at a


minimum height with clearance of 8 feet 5 inches ( 2565 mm ) from finished floor , at location X -
Y as shown in the model layout plan . Length of the wooden beam will depend on the length of
the examination room. Necessary supports to be provided to prevent sagging of the wooden
beam.

NOTE : For installation with Image Intensifier the minimum clearance required is 9 feet
( 2740 mm )

4 A trench with dimension 3 inches wide X 4 inches deep ( 75 mm X 100 mm ) covered with
seasoned wood , having cross section of 0.5 inch X 6 inches ( 13 mm X 150 mm ) , flushed to the
floor , is to be provided at location O - P as shown in the model layout plan .

5 For Fluoroscopic applications with fluorescent screen , the X - ray room windows , door and any
other openings will have to be made light leakage proof .

6 For Image Intensifier Installation , additional seasoned wooden beam of size 6 inches X 2.5
inches ( 150 mm X 65 mm ) to be fixed at a minimum height with clearance of 9 feet ( 2740 mm
) from finished floor , at location M - N as shown in the model layout plan.

7 Dark room with standard accessories to be provided before electrical installation of the
equipment. Guidelines should be obtained from the manufactures / Suppliers of the Dark room
accessories.

RADIATION SAFETY

• Lead lined Screen to be provided in front of the X - ray control for the safety of the operator.

• Radiation safety measures to be adhered to (e.g. use of lead aprons and other radiation
protection measures) as per Atomic Energy Regulatory Board regulations for the safety of
patients and operator.

010-4 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Pre - Installation

Electrical Work for Installation of X - Ray Machines

A Mains power supply to be provided and terminated in the Main Switch at a Height of 5 feet
(1270 mm) from finished floor, at the location (1), as shown in the model layout plan.

B Double earthing to be provided and terminated in copper strip below the main switch.

C Mains Contactor ( Wherever applicable ) to be installed at a height of 5 feet ( 1270 mm )


from finished floor , at the location ( 2) , as shown in the model layout plan .

D Push button station ( Wherever applicable ) to be installed at a height of 5 feet ( 1270 mm )


from finished floor , at the location ( 3 ) as shown in the model layout plan .

E A radiation warning lamp to be installed above the entrance of the X - ray room door at the
location (4), as shown in the model layout plan.

F For three phase mains supply , where static balancer is used , wiring from mains contactor
(2) to the static balancer (5) is to be provided.

G In case of Image Intensifier installation, a servo controlled voltage stabilizer of, 240 V, 50 Hz,
1 KVA Capacity will have to be provided.

CAUTION

• All items from (1) to (9) will have to be interconnected with the earth wire terminating in the
copper strip with proper size of copper wire.

• The power supply to the X - ray room is to be drawn by the shortest possible route from the
main distribution transformer or the main source of power supply located in the Hospital /
Clinic.

• In case the line resistance exceeds the rated value required to give full output of the X - ray
generator, the output of the X - ray generator will have to be de - rated.

010-4 Pg. 2 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

List of Pages

020 - 1 List of Pages

020 - 2 Safety Instructions

020 - 3 Protective Measures

020 - 4 Check List of Packed Material

020 - 5 Preparation of site & Installation

020 - 6 Physical Installation of Equipment

020 - 7 Start Up Instructions

020 - 8 Adjustments

020 -1 Pg 1/1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Safety Instructions
General Safety Instructions
Before proceeding with the installation, make sure that mains voltage and frequency specified
are available. Check availability of the pre - filtration on the X - Ray tube assembly. Remove
packing materials and protective grease residue from all parts before assembling. Apply thin
film of oil on to rails .Re - install all the mounting bolts / screws as soon as possible during
installation to prevent later mix - up (This does not apply to the bolts of transport packing.)

Operating Safety Instructions


X Rays are hazardous to the operating personnel and the patient. This equipment should be
operated only by authorized personnel in the medical institutions having knowledge of radiation
hazards and methods of protections. Before operating the equipment, the operator must
familiarize himself with the operating instructions.

General Recommendations

1 During Radiography, the operator should stand behind a protective screen and as far
away as possible (not less than 2 Meters) from the X ray tube head and the patient.

2 The X ray beam should be properly confined by the use of collimating devices and cones.
This also improves the picture quality. During fluoroscopy, the X ray beam should be
confined to the limits of the fluoroscopic screen protected by lead glass.

3 During Fluoroscopy, the operator should wear protective clothing, e.g. lead apron,
protective gloves. The protective ray guard provided should be mounted on the spot film
device

NOTE

Servicing and repairs of this equipment should be undertaken by


authorized personnel only.

020 -2 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Safety Notes
To ensure the safety of the patient and the operator, the equipment must be checked regularly
to maintain its reliability and serviceability. More frequent checks are required if the equipment
is operated under unusual conditions.

Before using the equipment for examinations, the user must ascertain that all the safety related
devices are functioning properly and that the unit is ready for operation. Special attention must
be given to the displays and pilot lights (Visual Checks). Normally during radiography/
fluoroscopy the radiation / H. T. ‘ON ‘Warning indication lamp lights up. In case of
malfunctioning of the control component, the warning lamp may continue to glow. In such case,
the complete unit must be switched ‘OFF’ and the concerned service department should be
informed for fault correction.

As manufacturers of electro medical products, the safety related performance of the equipment
is strictly maintained. The repairs, modifications and maintenance are carried out only by us.
For this purpose the replacement parts, as and when required are changed with the original
components only.

020 -2 Pg. 2 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Protective Measures / Safety Instructions


Room Installations
The room installation shall be in compliance with the AERB regulations. All necessary works have
to be carried out according to the installation documents.

Make sure that all protective - ground wires provided by the manufacturer are connected properly
before starting the equipment after installation. In case of a fault, they must always be connected
to the protective - ground wire. To ensure this, the connection between these covers and the
protective ground wire must be made with screws and lock washers, or via the protective ground
wires provided. This applies for example to all metallic covers which may develop a dangerous
contact voltage in the protective ground wire between the different components of the equipment
and the main supply must be connected according to the wiring diagram. After completion of the
operation, the protective ground wires must be tested with the Multimeter.

Personal Protection Measures ( Occupational Safety )


The regulations governing labour protection and accident prevention must be observed, primarily
in the interest of the person performing the work. These regulations are based primarily on the
accident prevention regulations of the occupational safety regulations in the relevant country. The
advice below must be regarded a primarily additional information.

Working Under Electrical Voltage :


No work shall be performed on parts above voltage (> 42 V). This prohibition does not apply for
measuring and adjustment procedures. In these procedures special care must be exercised. Only
such tools and measuring instruments may be used which are suitable for the intended procedure.

Body Protectors :
It is mandatory to use the appropriate body protectors.

Radiation Protection :
The safety instructions described concerning radiation protection must be observed during
adjustment and start - up procedures for X - ray systems.

Mandatory Reporting :
The engineering supervisor in charge must be informed at once if there has been an accident or if
there are any hazards which may cause an accident. Make sure that the ground connection is
correct. The protective conductor shall be connected to the ground terminal of the generator as
indicated in the wiring diagram. At the end of the installation, these ground connections shall be
checked again.

020 -3 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

1 * Check List of Klinoskop H Table Component’s

Case I
Dim. - 2100 mm X 1020 mm X 1075 mm
Approx. Net Wt. 315 Kg. Approx. Gross Wt. 450 Kg.
Klinoskop H Table + Bucky with Bucky Grid + Cassette Tray , Single tube telecoupling
Gear Box Assembly ( Motor + Gear Box ) , Foot rest, Cough guard , Ray guard ,
Compression cone , Diaphragm plate, Patient Hand Grips, Ring nuts ,Straight cuff , Cone
12 cms, Motorised Collimator with connecting cable, Foot switch , Immobilizing device, Base
marking template , Guide rod, Interface cable, Packet of Fastener ,Operating Manual

Case II
Dim. - 1150 mm X 650 mm X 300 mm
Approx. Net Wt. 45 Kg. Approx. Gross Wt. 70 Kg.
Explorator H with cable & Screen Grid

Case III
Dim. - 1150 mm X 1000 mm X 650 mm
Approx. Net Wt. 70 Kg. Approx. Gross Wt. 118 Kg.
L - Carriage with compression carriage

Case IV
Dim. - 725 mm X 260 mm X 175 mm
Approx. Net Wt. 80 Kg. Approx. Gross Wt. 100 Kg.
Counter weights

Case V
Dim. - 685 mm X 260 mm X 175 mm
Approx. Net Wt. 70 Kg. Approx. Gross Wt. 85 Kg.
Counter weights

020 - 4 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
2 * Check List of Column Stand Components

Case I
Dim. - 6150 mm X 425 mm X 450 mm

Column Stand
Cross Arm Tube
Column Extension Assembly
Floor Rail
Ceiling Rail
Extension Floor Rail
Extension Ceiling Rail

Case II
Dim. - 1100 mm X 775 mm X 550 mm

Column Base
Cross Carriage
Set of Scales
Column Carriage Cover
Cable Holder
Cable Clamp
Flat Piece
Handle for column Carriage
Handle for cross Arm
Counterweights
Packet of Fasteners

020 - 4 Pg. 2 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
Preparation Of Site

Check the approximate dimensions of room in which Klinoskop H is to be installed. Refer


dimensional drawing provided.

Check the availability of required power supply. Plan lay out by referring pre - installation
instructions. In Accordance with the dimensional drawing and project drawing, establish the
orientation point and central line of table. Do the marking on the floor.

Preparation Of Installation

1 * Take all packed casings in a room nearer to area, where they are going to be installed.

2 * Open all COLUMN Cases and check the received parts are as per check list, given in
020 - 4 Pg. 2 / 2.

3 * Mark the positions of horizontal beam, Ceiling rail & flooring rail. Drill the holes
accordingly and fix them with given fasteners (Refer Dimensional & Project drawing).

4 * Install Column stand (Installation of Column stand is not given in this manual
descriptively. )

5 * Start unpacking all table cases.

Caution : Don’t remove supporting beams on which table is rested .

Only take out cardboard packs. Check all material is as per check list given in document
020 - 4 Pg. 1 / 2

6 * Align the Foundation template and mark holes in the floor. Drill the fixing holes. Grout
the foundation bolts supplied along with the unit

NOTE

The floor must be leveled. Check with spirit level.

020 - 5 Pg 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Physical Installation of Equipment


1 * Installation of Gear Box
1) Bring table which is supported by
wooden beams on wooden base parallel
to the installation area so that transferring
unit Klinoskop H from wooden base to floor
will be easier .
2) Take out bottom cover ( 1 fig. 1 ) fixed on
leg assembly at the base for installation of
Gear box ( 2 fig. 2 ).
3) Remove Gear box assembly (Gear box +
motor) separately packed and mount into the
5 3 1 4 base unit such that sprocket teeth of Gear box
engages in bed segment.
4) Before fixing with M8 x 28 hex. Hd. Bolts -2
Fig. 1
X & M 8 x 25 Soc. Hd. Screw - 2 x with
washers. Check engagement by physically
checking whether sprocket has gone inside
or not .Fix Hex. Hd. Bolt M8 X 80 with Hex. Nut
M8 in such a way that movement of gear box
assembly is restricted and proper engagement is
done. After assuring of engagement insert
6 Dowell pins 5 X 25 - 2 X provided to engage
gear box on leg base assembly ( 5 fig. 1).

2 Carefully remove wooden support beam (4 Fig.


1). Screw out the frame nuts (3 fig. 1) from unit
base.

Installation of Table frame on Floor


6) Transfer the base unit at the location
Fig. 2 prepared with the help of at least 6 - 7 casual
labourers by carefully physically lifting the
unit .Bolt the base unit to the floor with M 10 x
30 bolts ( Foundation Bolts ) - 4 x with
Segment washers and lock washers.
7) Install safety switch assembly on Gear box
Housing such that actuator provided on segment
Switch Connection for horizontal position presses the lever provided
Ferrules on switch ( centrally located in safety switch
S 10 S 10 - 101, 102 assembly ) provided for sensing horizontal
position ( 6 fig. 2 ).
S 11 S 11 - 111, 112

S9 S9- 91, 92

020 - 6 Pg. 1 / 12
Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
2 * Installation of Horizontal Carriage Assembly

Unscrew and remove profile rails, Spacer strip


and take off the table top.

NOTE
8
Do not unscrew the transport protector (7 fig.
10 6) provided to hold compression carriage (8 fig.
3) on transverse carriage (9 fig. 3). Temporarily
unscrew the stopper (10 fig. 3) from L - Carriage
assembly.

Fig. 3

Position the L - Carriage assembly (Transverse


Carriage + Compression Carriage ) ( 11 fig. 4 )
parallel to the unit ( 12 fig. 4 ) and carefully slide
11
it on , with the help of two people ( fig. 5 ).
Screw the stopper back into position (Access is
from bottom)

Fig. 4

Fig. 5 Fig. 6
020-6 Pg. 2 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

3 * Installation of Counter-weights

Counter weight carriage is fixed to bed longitudinal


member by transport protector (13 fig.7) to restrict
movement of horizontal carriage in
transportation. Unscrew and remove transport
protector carefully to relieve horizontal carriage.

13

Fig. 7

Install the weights supplied in the counter weight


carriage ( Install it by number wise ) and bolt
into place . Observe the white strip coding on the
weights (Fig. 8)

Fig. 8

020 - 6 Pg. 3 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
4 * Installation of Spot Film Device

Place the spot film device on to a suitable base.


Remove stop bolt (14 Fig. 9) at the top end of the
spot film unit (15 fig. 9) guide rail. Unscrew and
17 remove compression carriage unit cover
(16 fig.10)

NOTE
15
Watch out for switch S35 (17 Fig. 9) while
14 inserting the spot film unit into the compression
carriage (In received condition you may find tape
Fig. 9 on S35 switch). Don’t remove tape from this
switch at this point of time. Slide the spot film unit
as shown in (fig. 11) across the table top
and in the direction of arrow ( Guide the rails
between the ball-bearings ) into the compression
carriage unit until fully engaged . Screw the stop
bolt (14 fig. 9) into the spot film unit guide rail
again. Remove tape from S35 switch. Unscrew the
transport protector from L - Carriage
( Compression carriage is hold to L - Carriage
16 by transport protector ) . Screw the spacer tube
back again. This serves as a stop for the
compression carriage

Fig. 10

Fig. 11 020 - 6 Pg. 4 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

5 * Installation of Explorator Accessories

Insert compression cone in diaphragm plate such


that it gets locked inside (Fig. 12).

Insert the loops of radiation guard (Flap type)


through the angle profile behind the spot film unit.
(Fig. 13).

Fix cough guard with two screws supplied in front of


the spot film unit as shown in (Fig. 14)

Fig. 12

Fig. 13 Fig. 14

020 - 6 Pg. 5 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
6 * Installation of Counter-weights for
compression carriage & spot film unit

Remove the cover plate by unscrewing M3


screws - 12 xs from L - Carriage rear side. Install
the counter - weights as shown in Fig. 15.

Fig. 15

Check balancing of compression carriage along with spot film (Including Cassette) by moving it up or
down. After leaving Explorator (Spot film unit) at particular position it should remain in same position
after removing your hand. Then it is called as perfect balancing.

Before going further in installation, make sure of following points

1* Fixing of Table to Floor.

2* Interlocking of Gear box sprocket to Bed segment.

3* L - Carriage Installation.

4* Counter - weights for L- Carriage & Bucky

5* Spot film unit Installation

6* Spot film unit accessories installation.

7* L - Carriage Balancing.

Fix previously removed table top; profile rail with spacer strips

020 - 6 Pg. 6 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
7 * Installation of Electrical Connections
Routing of Explorator Interface Cable

Route the hose emerging from spot film unit on L -


Carriage side wall (18 Fig. 16). Remove leg
18 assembly side covers (19, 20 Fig. 17). Route the
cable on leg assembly such that hose flange
provided for fixing will come at the position of cut -
out provided in leg component (21 fig. 17).
Connect Euro - connector socket to X2 connector
on Power Supply & Distribution Pcb Card
mounted on leg tilt assembly ( 22 fig. 18 ) fixed
at side wall of leg.

Fig. 16

21

19
22

20

Fig. 18
Fig. 17

020 - 6 Pg. 7 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
Remove Compression Carriage cover as shown in
23 (23 Fig. 19). Remove vertical Carriage cover as
shown in (24 Fig. 20). Intra - carriage cable is
already routed on L - Carriage assembly. Connect
connector X 24 of Explorator interface cable to
intra - carriage cable connector X 25 (25 Fig. 21).
Connect other end of intra - carriage cable to
longitudinal & transverse brake assembly
(26 Fig. 22). Insert Hose in leg cover in such a
way that groove of hose flange will fit in leg cover
Fig. 19 wall as shown in Fig. 23

25
24

Fig. 20 Fig. 21

26

Fig. 22 Fig. 23

020 - 6 Pg. 8 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
Interconnections of External Interface Cables

Foot Switch & Optional remote KV control switch Connection

Install foot switch by connecting foot switch flying leads to a provided socket
( A part of cable routed from Explorator, 2 way socket ) X 15

Connect optional remote KV control switch cable ( For Multiphos 10 ) to


a provided connector X 17

If the case of two tube execution manual collimator can be interfaced to


two way connector provided in Leg.

Collimator , Room Light , Generator Interface Cable connections

Connections of external Interface cables are to be done as follows.

Sr. No. Type of External Cable Where to connect on Power Supply &
Distribution Pcb Assembly

1 Collimator Cable X5
8 Core , Multicore Grey

2 Generator Interface Cable X3


16 Core , Multicore Grey

3 External room light connection X7

020 - 6 Pg. 9 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Keep the foot rest (27 Fig. 24) into the place;
press the two handgrips (28 Fig. 24) upwards so
that the foot engages in the groove of the profile
rail notches. Pressing the handgrips upwards
releases the locking mechanism; the footrest can
be placed to any desired position.

27 28

Fig. 24

020 - 6 Pg. 10 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation
8 * Explorator H Manual Spot film Device

10

1
2
3

4
5
6

7 9

Fig. 25
Operating Controls

1 Room Light Switch 11 Release for sliding the spotfilm device


into the parked position
2 Fluoroscopic kV remote adjustment 12 Scattered Radiation guard
3 Motorised Collimator Control 13 Cough Guard
4 Fluoroscopy Release 14 Hand Grip
5 Exposure Release
6 Switches for electro-magnetic locking Cassette Programme ( inch )
of spot film device.
7 Unit Tilting 8” X 10 “
8 Guide Handle
9 Handle for moving cassette frame with 10 “ X 12 “
stops for choosing film division
10 Cassette frame for cassette sizes 8” x 14 “ X 14 “
10 “ and 10 “ x 12 “

020 - 6 Pg. 11 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

9 * Installation of Under - couch tube Assembly


(It is better to do first switch adjustment mentioned in 13 if required and then
proceed for this installation).

Attaching the X - ray tube Assembly to the


rigid support :

Move the table into vertical position. Mount the


31 rigid tube support assembly (29 Fig. 26 ) by
33 M6 x 25 Allen bolt - 3 x together with washer and
29 lock washer to L - Carriage base ( 30 Fig. 26 ) . Fix
height adjusting bracket (31 Fig. 26) along with
30 32 adjusting bolt with M6 x 20 Allen bolts - 2 x. Bolt the
X - ray tube assembly (32 Fig. 26) to rigid support
with minus side upwards with
M 6 x 16 Allen bolts - 4 x (33 Fig. 26) with washer
and lock washers.

Fig. 26
Installing Motorised Double slot Collimator :

Place the Motorised double slot collimator with the


clamp sides on the flange of the X - ray tube
34
assembly. Screw in the four bolts with uniform
36 35 tightness. The clamp slides moved by the bolts must
clamp over the flange. These center the diaphragm
to the X - Ray tube assembly. Fasten the extension
cable plug to the socket of Motorised double slot
collimator.

Deploying the voltage cable and rotating anode


cable

Refer Documents supplied along with control.


Feed the high voltage cable (34 Fig. 27) and
rotating anode cable (35 Fig. 27) to the x - Ray
tube assembly and connect in accordance with
the wiring diagram of the generator used. Screw
Fig. 27 the two high voltage cables and the rotating
anode cable along with the cable clamp supplied
(36 Fig. 27).

020 - 6 Pg. 12 / 12

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Start Up Instructions

1* Before connecting table to available power supply, check carefully for any loose
connections. Refer wiring and circuit diagrams provided in 4 W 5245.
2* Check the position of MCB on leg tilt assembly. It should be in OFF position.
(Lever should be Downwards )
3* Check positions of Explorator switches. Switches which are mounted horizontally
operating lever should be on left side. Room light switch operating lever should be
on upper side.
4* Check whether collimator and Bucky are interfaced or not. If not then connect
them.
5* Check the incoming power supply by Multimeter. It should be in the limit of
( 415 Vac + 10 % - 15 % )
6* Connect Input supply cable, where phase sequence is clearly indicated. To be on
safer side, check resistance between Phases to Neutral & Phase to Ground. It
should be in Mega - ohms.
7* Switch on the power supply. Check availability of supply at MCB Input.

8* Carefully switch on the MCB.

9* If connected phase is of proper sequence and no mal - function in electrical


assemblies, you observe that Phase Sequence Pcb assembly green LED should
be ON.
10 * If phase sequence Red LED or both Red & Green are On, it shows that phase
sequence of connected cable is wrong. Check and rectify the problem.
11 * Operate tilt switch (provided on control arm of spot film unit) momentarily in one
direction such that table starts tilting. To check whether tilt interfacing is proper or
not.
12 * Operate transverse & longitudinal brakes and check whether movement of L -
Carriage is restricted.
13 * Operate compression brakes and check whether movement of compression
Carriage is restricted.
14 * Move joy - stick lever (Provided on control arm). Observe the collimator horizontal
& Vertical movements of shutters .Press switch for collimator bulb provided on
collimator. Check it should be ON.

020 - 7 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Adjustment Instructions
Adjustment of Limit Switches

If the adjustment is done properly at the time of installing Gear box assembly then following
adjustments are not needed.
Counter - Weight balancing must be accurately carried out with all the accessories like lead
guard, Cough guard & Compression cone.

May be fine adjustments are required because safety switch position depends upon the Gear
Box Installation position.

Go to particular position either Horizontal or Vertical and check whether tilting stops at that
position If not then loose respective actuators provided on segment such that they will press the
plunger of safety switch in turn stops the movement.

S 9 - For Vertical Position

S 11 - For Horizontal Position

S 10 - For Trendelenburg Position

020 - 8 Pg. 1 / 3

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Central ray adjustments

Caution

When working with X - rays, protective clothing must be worn. No parts of the body must
enter the radiation zone. Radiation must be switched off when making mechanical
adjustment.

Centering x - Ray tube assembly and


collimator unit with rigid support.

To do this adjustment table should be in vertical


position. If not bring in vertical position by
operating tilt switch of Explorator control arm.
Mark diagonally center of lead glass fitted on screen
assembly.

Fig. 28 Keep generator in fluoroscopy mode.

Switch ON the collimator bulb, set average field


size and mark the light field on lead glass. Do the
fluoroscopy for small focus (Small time Duration).
Mark the field on lead glass. Do the fluoroscopy for
large focus (For 5 Seconds time exposure). Mark
again the field on the lead glass. Ascertain the shift
in focus. The radiation fields of both focuses must
be centric within the light field marking.

Move the collimator on the flange by means of the 4


holding screws until the above requirements are
met.

Fig . 29

020 - 8 Pg. 2 / 3

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Installation

Vertical Direction

Switch on radiation. Move the spot film unit in the


compression direction and vertically delimit about 1
cm with the collimator. Mark the center of the
collimation on the fluorescent screen, with the spot
film unit decompressed. Now actuate compression
to determine displacement and mark once more.

Rotate the x - Ray tube assembly to reduce this


displacement to a minimum. Ascertain the shift of
the central ray by compressing and de-compressing
a number of times. Fig. 30

Fig. 30

Horizontal Direction

Switch on radiation, compress the spot film unit and collimate horizontally about 1 cm with the
collimator. Mark the center of collimation on the fluorescent screen with the spot film unit de-
compressed. Now operate compression & then ascertain displacement and mark once more.
Adjust the displacement to a minimum by tilting the X - Ray tube assembly. Ascertain the shift
of the central ray be compressing and de - compressing several times. See fig. 30.

Carry out centering of the central ray to the center of the spot film unit using the support.

Check the central ray adjustment in the 0 ° , 90 ° and 12 ° positions of the table top and center
where necessary.

020 - 8 Pg. 3 / 3

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

List of Pages

030 - 1 List of Pages

030 - 2 Power Supply

030 - 3 Brakes

030 - 4 Table Tilting

030 - 5 Collimator

030 - 6 Explorator Control Arm

030 - 7 Generator Table Interface

030 - 8 Phase Sequence Indicator

030 -1 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -2 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -3 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -4 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -5 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -6 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -7 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Diagram

030 -8 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description
List of Pages

040 - 1 List of Pages

040 - 2 Power Supply

040 - 3 Unit Brakes

040 - 4 Table Tilting Circuit

040 - 5 Explorator Control Arm

040 - 6 Control - Explorator Interface

040 - 7 Phase Monitoring Circuit

040 -1 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Power Supply ( W 5245 - 2 - 4 )

Input Transformer provides primary and secondary voltages to the unit.

Component Review:

Transformer, Voltage Supply for basic unit and spot


T1 film device.

- Primary Voltage 0 - 220 V Bet. X 1-3 & X 1-4


- Secondary Voltages as given below

0 V - X 1. 8
17 V - X 1. 7
21 V - X 1. 6
28 V - X 1. 5

F5 28 V Ac for collimator motors and contactors


Adjustment of the light beam collimator lamp voltage
F6
26 V Dc for unit brakes and relay circuits on the PCB.
F7

F 12 13 V Ac for Manual Collimator

040 - 2 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Unit Brakes ( W 5245 - 3 - 4 )


S 14 Switch operates the brakes br1, br2 of the compression carriage and S 15 switch
operates the brakes br3, br4, br5 on the under carriage for transverse and Longitudinal
direction.
When the Cassette is moved in the exposure area during preparation S 17 micro switch is
operated, which operates relay hr 1. Relay hr 1 contacts energizes all the brakes on
Compression carriage and under carriage.
Initially a nominal voltage of 6 - 8 Volts is applied to one coil of all brakes through R 1 and
when the corresponding switches for the magnetic brakes are operated, 26 V is applied to the
other coil of the brakes and thus this brings brakes into operation.

Component Review:

br1 , br2 Compression brakes on spot film device.


br3 , br4 Longitudinal brakes on spot film device.
br5 Transverse brakes on spot film device.
S 14 Switch for Compression brakes.
S 15 Switch for Longitudinal & Transverse brakes
hr 1 Relay operates during generator preparation

Current Flow:
(I) Stand - by Condition
26 V ⇒ X6.1 ⇒ R1 ⇒ X25.3 ⇒ X24.3 ⇒ X14.3 ⇒ br3,br4,br5 ⇒ X14.5 ⇒ X24.5 ⇒ X25.5
⇑ ⇓
⇒ X6.6 ⇒ X2.Z14 O V ⇐ X2 . b16
⇓ ⇑
⇒ X13.3 ⇒ X25.2 ⇒ X24.2 ⇒ br1 , br2 ⇒ X13.5 ⇒ X24.4 ⇒ X25.4
(II) When S 14 (NO) is Operated
26 V ⇒ X2.z2 ⇒ X21.4 ⇒ X12.4 ⇒ X21.2 ⇒ X12.2 ⇒ X2.b12 ⇒ X6.13 ⇒ X6.9 ⇒ X2.z12
⇒ X25.1 ⇒ X24.1 ⇒ X13.1 ⇒ br1 ⇒ X13.5 ⇒ X24.4 ⇒ X25.4 ⇒ X2 . b16 ⇒ O V

(III) When S 15 (NO) is Operated


26 V ⇒ X2.z2 ⇒ X12.3 ⇒ X21.3 ⇒ X21.1 ⇒ X12.1 ⇒ X2.b14 ⇒ X25.6 ⇒ X24.6 ⇒ X14.1
⇒ br3 , br4 , br5 ⇒ X14.5 ⇒ X24.5 ⇒ X25.5 ⇒ X2 . b16 ⇒ O V

(IV) When S 17 (NO) is Operated


26 V ⇒ hr1 ( NO) ⇒ V17 ⇒ X6.13 ⇒ X6.9 ⇒ X2.z12 ⇒ X25.1 ⇒ X24.1 ⇒ X13.1 ⇒ br1 , br2
⇒ X13.5 ⇒ X24.4 ⇒ X25.4 ⇒ X2.b16 ⇒ O V

⇒ V19 ⇒ X2.b14 ⇒ X25.6 ⇒ X24.6 ⇒ X14.1⇒ br3,br4,br5 ⇒ X14.5 ⇒ X24.5
⇒ X25.5 ⇒ X2.b16 ⇒ OV

040 - 3 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Table Tilting Circuit ( W 5245 - 4 - 6 )

S7 Foot Safety Switch


S9 Limit Switch for upright position
S10 Limit Switch for Trendelenburg Position
S 11 Limit Switch for Horizontal Position
S 18u Switch (Control Arm) for tilting unit
S 18a Switch (Control Arm) for up righting unit.
hr 4 Bridging over horizontal position.

Operation of switch S 18 a or S 18 u (Control Arm), Contactors S u or S a is actuated for the table


movement. The limit switches S9 (+90) and S10 (- 12) interrupts the contactor control circuit in the
end positions of the unit.
While tilting the table, limit switch S11 interrupts the control circuit on reaching the horizontal
position (O).If the unit is desired to be tilted in trendelenburg position or upright, control switch to
be released and corresponding control arm switch S 18a or S18u to be operated again.
Pots r5 and r 12 to be set at 22 K and 300 ohms respectively. Pot r12 to be adjusted so that the
relay hr5 picks up on operating S 18 u and is chatter free. Pot r5 is to be adjusted so that, the table
crosses over the horizontal limit switch S 11 Smoothly.

For Vertical Position

When the switch is moved right i.e. to operate S 18 a (NO) is actuated and contactor S a is
switched ON which causes the motor to run in one direction and the table steadily rises to the
vertical position.

28 V ac ⇒ X20.9 ⇒ X11.9 ⇒ X11.7 ⇒ X11.8 ⇒ S18a ⇒ X20.8 ⇒ X2.z4 ⇒ X6.5 ⇒ Su.21


⇒ Su.12 ⇒ Sa1.A1 ⇒Sa1 . A2 ⇒ S9 ⇒ X8.1 ⇒ X4.13 ⇒ S7 ⇒ X4.14 ⇒ OV∼

When the table reaches vertical position, the cam actuates switch S9, thus interrupts the supply to
contactor Sa and the table stops in the vertical position.
To acquire any intermediate position the switch on the control arm should be released when
desired position of the table is attained.

For Horizontal Position

When the switch is moved left, i.e. to actuate S18u, the switch S 18u (NO) operates and contactor
Su is switched ON which causes the motor to rotate in the opposite direction i.e. horizontal and the
table steadily lowers down to the horizontal position and stops in this position when the switch S 11
is actuated by the cam and the supply to Su is interrupted.

040 - 4 Pg. 1 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description
28 V ac ⇒ X20.9 ⇒ X11.9 ⇒ X11.7 ⇒ X20.7 ⇒ X2.b4 ⇒ X6.4 ⇒ X2.z4 ⇒Sa.21

⇒ Sa.22 ⇒ Su.A1 ⇒Su.A2 ⇒ S10⇒ X8.3 ⇒ X8.2 ⇒ S11 ⇒ X4.14 ⇒ OV∼

When the horizontal position is reached the micro switch S11 opens and the table now stops in this
position.

To acquire any intermediate position, the switch on the control arm should be released when the
desired position of the table is attained.

For Trendelenburg Position

This position is meant to further lower the table by 12 ° w.r.t. the horizontal position. This is
achieved by operating the switch on the control arm.

The micro switch S11 opens on reaching the horizontal position .When S 18 is operated again
rectifier V 21 energizes the relay hr5.

When the hr5 relay is energized, it switches ON the transistor V11 and thus relay hr3 is ON. When
the hr3 relay is ON, contactor hr4 is also ON and the circuit is completed as below.

28 V ac ⇒ X11.9 ⇒ X11.7 ⇒ X20.7 ⇒ X2.b4 ⇒ X6.4 ⇒Sa.21⇒ Sa.22 ⇒ Su.A1 ⇒Su.A2 ⇒


S10⇒ X8.3 ⇒ hr4 ⇒ OV∼

The table is now further lowered (w.r.t. the horizontal position) until the cam operates the switch S
10 (NC) and the table stops in this position i.e. is the trendelenburg position. The cam is so
adjusted that the table can be lowered only up to 12 ° below the horizontal position.

040 - 4 Pg. 2 / 2

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Control Arm ( W 5245 - 7 - 3 )


The Klinoskop H table is fitted with a Spot film device “Explorator H “. All table, radiation control
and collimator related control functions are installed by actuating the control switches mounted on
the control arm.

The functions of the switches are as given below.

Rocker switch controls the table tilting either in vertical or


S18 u / S18 a horizontal directions.

Toggle switch controls the electromagnetic brakes for the


S 14 movement of Explorator in the compression direction.
Toggle switch controls the electromagnetic brakes for the
S 15 movement of Explorator in both longitudinal and
transverse direction.
S 31 a , b , c , d Joy stick controls the opening and closing of the
diaphragms of double slot Motorised collimator.
Micro switch for sensing the Explorator in the park
S 35 position.
Micro switch for sensing the cassette in exposure area to
S 17 initiate the preparation for an exposure.
Push button switch for initiating the fluoroscopy from the
S 25 control arm of Explorator.
Push button switch to start the exposure from the control
S 22 arm of Explorator.

S 40 Toggle switch actuates the room light.

040 - 5 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Control - Explorator Interface ( W 5245 - 8 - 3 )

Fluoroscopy

When Explorator device is selected at the control 24 V / 220 V is available at connector X 3 / 2.


Micro switch S 35 closes when Explorator is moved out of parking position operating the hr2 relay.
Operating the push - button switch S 25 or foot switch S 25 A will start fluoroscopy.
Before fluoroscopy could be initiated, the X - ray tube shall be coupled to under carriage behind
the table. For single tube tele-coupling, on coupling the tube, micro switch S26 operates and
enables fluoroscopy control circuit through hr2 relay. For 2 - tube tele - coupling S 26 is
permanently shorted through a link between X 24.8 and X 24.7

Fluoroscopy circuit is completed as follows

N1.22 ⇒ X3.2 ⇒ hr2 ( NO ) ⇒ hr1 ( NC ) ⇒ X2.b20 ⇒ S25 / S25 A ⇒ X2.z20


⇒ X3.13 ⇒ N1.46 ⇒ To Fluoroscopy Contactor Ds via Zb2 ( NC ) contact

Exposure Release

When the cassette is moved in the exposure area, a micro switch S 17 is actuated, thus operating
the relay hr 1. Compression and transverse brakes are operated. 24 V / 220 V is sent out to control
through X3 / 12 for generator preparation. 24 V / 220 V generator ready signal is available through
X 3 / 3. Exposure is initiated by pressing the push - button switch S 22.

During preparation mode circuit is completed as follows

26 V ⇒ S 35 ⇒ S 17 ( NO ) ⇒ hr1 ( NO )
N1.22 ⇒ X3.2 ⇒ hr2 ( NO ) ⇒ X3.12 ⇒ N1.47 ⇒ To preparation circuit of control.

The radiographic exposure is completed through

N1.31 ⇒ X3.3 ⇒ X2.b18 ⇒ S22 ⇒ X2.z18 ⇒ X3.11 ⇒ N1.29 ⇒ To the timer ⇒ AS ON

When taking “BUCKY “exposure the circuit is completed as follows

220 V ac ⇒ N1 .32 ⇒ X3.5 ⇒ X4.8 ⇒ V22

V22 ON brings bb relay into circuit which starts Bucky motor and S 1 brings ab relay which
switches ON the timer and hence AS and exposure takes place.

040 - 6 Pg 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Circuit Description

Phase Monitoring Circuit ( W 5245 - 15 - 2 )

This circuit monitors all the three incoming phases on the unit and disables the table movements
(Table tilting) if the phase sequence fails.

Phase ‘U ‘is used to generate 12 V Dc for the circuit. Phase ‘V ‘is isolated and converted into
square wave by J 1. Differentiated by R 5 and C 1. Phase ‘W ‘is also isolated and converted into
square wave by J 2.

The output of J3.15 and J3.16 is ored and then integrated to drive the interlocking relay K1. As
long as phase sequence is correct J3.15 goes periodically high and keeps K1 ON.

If any of the phases sequence fails. J3.15 remains low for all the time, thus K1 drops, which in turn
cut the control supply of motor contactors. The fault condition is indicated by red LED V3. A
normally closed contact of K1 is used to light the red lamp externally visible on the table.

040 - 7 Pg 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

SERVICE (Fault Finding / Repairs)

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Service

Supply Problem
No

Is Mains Voltage Is Mains Voltage


Is MCB Position Is Input
observed at X 1.3 & No observed at X 1.1 & No Yes
O.K. voltage O.K.
X1.4 on D9 PCB X1.2 on D9 PCB

Replace PCB D9 No Yes


Yes

No Change position of
Yes
Is F10 Fuse on D9 Replace MCB
MCB & Check
Blown

Is 28 V ac voltage Is Voltage 28Vac Any visual Find out Supply


present at X 1.6 on D 9 No present at X1.5 on Yes indication on D999 Yes Problem
PCB D999 PCB pcb

No No

Are all phases


Yes Replace D999 PCB Yes
present
Replace Input
No
Transformer

Is 26 V dc Voltage
Is F7 Fuse on D9 Is 21V ac present at
across capacitor C1 on No Yes Yes Replace D9 PCB
PCB O.K. X1.6 on D9 PCB
D1 on D9 PCB

No Replace F7 Fuse

Observation : No Power Supply


050 -2 Pg. 1 / 5

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Service

Yes

Is Voltage 26 V dc
Refer previous page of trouble
present across capacitor C1 on No
shooting
D9 PCB

Yes

Is Circuit path of S14 &


S15 Switch to Brakes & other No Replace respective switches
Electronics is O.K.

Yes

Is Mechanical adjustment /
Yes Replace respective Brakes
Fitting of brakes are O. K.

No

Do the adjustment of
brakes and Check
Again

Observation : Activation / De-activation problem of Brakes

050 -2 Pg. 2 / 5

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Service

Yes

Check the availability of


all Supply voltages

Yes

Are Safety switches S9 , S11 ,


No Replace Respective Safety Switches
S 10 functionally O . K.

Yes

Are Coil voltages acoss each contactor


No Replace Respective Contactors
O.K. in respective position of tilt switch

Yes

Is S 18 Tilt Switch O.K. No Replace S18 Switch

Yes

Refer Circuit Diagram

Observation : Table Tilt Not O . K.


050 -2 Pg. 3 / 5

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Service

Yes
Yes

Do Trouble shooting as Is Fuse F5


Is 28 V ac Present No given in previous pages No Replace Fuse F5
O.K.

Yes Yes

Make mech.
Is Fuse F6 O.K. No Replace F6 Fuse Is Joystick lever operating No adjustment to
micro-switches of Joy ensure actuation
stick

Yes
Yes

Is W.W. Resistor
No Replace R5
R 5 O.K. Are Capacitors C1 , Replace Defective
No
C2 of Collimator Capacitors
O.K.

Yes
Yes

Replace defective
Is Collimator Bulb Is Collimator motor No
No Replace the Bulb motor
O.K. O.K.

Observation : Collimator Light Problem Observation : Collimator Diaphragm


movement problem

050 - 2 Pg. 4 / 5

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Service

Is S 35 in closed Is Explorator in Bring Explorator in


No No
condition Radiation field radiation field &
Check again

Yes

Yes
Replace S 35 Switch

Make the
Is Tube coupling switch
No Is connection to S 26 No connection &
S 26 in closed condition
is made Check again

Yes

Check the coupling & if


Yes necessary change
Switch S26

Is preparation
Is cassette inserted in Check contacts should
yes Switch S17 in Yes Yes Replace S17 Switch
explorator frame be open in pressed
closed condition
condition

No

Insert cassette frame in If problem persists check generator Interface cable &
radiation field Connections at generator side .

Observation : Radiography& Fluoroscopy Problem


050 -2 Pg. 5 / 5

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

KLINOSKOP H
MED

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

List of Pages

060 - 1 List of Pages

060 - 2 Spare Parts

060 - 1 Pg. 1 / 1

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

Table Base Assembly

Pos. No. Item Description Part No.


1 Gear Box + Motor 040 3066 001
2 Input Transformer 0420701 001
3 Actuator Bracket 0420711 001
4 Actuator Horizontal 0420712 001
5 Actuator Vertical & Trendelenburg 0420713 001
6 Safety Micro switch 009 5245 014
7 Segment Cover 0407204 001
8 Leg Cover ( Side Cover ) 0416632 001
9 Leg Cover ( Bottom Cover ) 0410118 001
060 - 2 Pg. 1 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

7 6 5 4

Table Tilt Assembly

Pos. No. Item Description Part No.


1 Mcb 3 Ph. , 6 A 0420702 001
2 W.W. Res 150 R 25 W 0404575 005
3 W.W. Res 5 R 15 W 0404575 017
4 Contactor 3TH80 31 -OAC1 0016607 000
5 RC Element 3TX6-406-OD 0016609 000
6 Phase Sequence PCB Assembly 0416576 001
7 Power Supply & Distribution PCB 0410084 001
Assembly

060 - 2 Pg. 2 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

Horizontal Carriage Assembly

Pos. No. Item Description Part No.


1 Compression Brake Assembly 0420753 001

2 Transverse & Longitudinal Brake 0420754 001


Assembly

3 Electromagnet - Complete 0407239 001

4 L - Carriage Cover Front Side 0407154 001

5 L - Carriage Cover Rear Side Sheet 0407156 001

060 - 2 Pg. 3 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

3 1 6 4 5 7

Explorator Control Arm

Pos. No. Item Description Part No.


1 Joy Stick Assembly 0410055 002
2 Knob 0403971 001
3 Toggle Switch SPST 0003698 002
4 Indicator Lamp ( Neon )
5 Push Button Switch 0003704 003
( NO ) Red Knob

6 Push Button Switch 0015251 000


( NO ) Green Knob

7 Rocker Switch 0015472 009


( Center Off )
8 Subminiature Switch E61-50K ( S 35 ) 009 5245 010
9 Subminiature Switch D43CQ1RA ( S 17 ) 009 5245 011

060 - 2 Pg. 4 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts

4
2

4
1

5 6 8

Pos. No. Item Description Part No.


1 Catapult Bucky 0408513 001
2 Cough Guard 0406128 001
3 Hand Grip 0402533 001
4 Foot Rest 0401160 001
5 Ray Guard 0407998 001
6 Ray Guard 0411018 001
7 Foot Switch 0401159 001
8 Collimator 0401238 001

060 - 2 Pg. 5 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.
Back
S KLINOSKOP H

Spare Parts
Miscellaneous

Item Description Part No.


Ball Bearing Complete 0407238 001
Ball Bearing Complete 0407237 001
Ball Bearing Complete 0407236 001
Ball Bearing Complete 0407235 001
Ball Bearing Complete 0409309 001
Ball Bearing Complete 0407358 001
Ball Bearing Complete 0407365 001
Ball Bearing Complete 0407359 001
Ball Bearing Complete 0407367 001
Ball Bearing Complete 0409308 001
Ornamental Screw M 3 X 8 0470582 005
Ornamental Screw M 4 X 8 0470582 002
Ornamental Screw M 4 X 12 0470582 003
Ornamental Screw M 4 X 15 0470582 004
Compression Cone 0410009 001
Handle ( Screen Frame ) 0400380 001
Table Top 0401341 001
Counter Weight Cover 0410121 001

060 - 2 Pg. 6 / 6

Siemens Ltd.
Copyright © SIEMENS LTD. INDIA. All rights reserved.
The information in this CD is not a controlled document and to be used only for reference.
Before ordering any parts HO-Med-CS-SP to be contacted for obtaining the latest information.