Professional Documents
Culture Documents
CLINICAL CHEMISTRY 1
PREPARED BY:
LGGRMT
➢ There are many instruments to choose from when doing a chemical
analysis of blood samples.
➢ They differ usually in their principles and applications.
➢ The most common instrument used in clinical chemistry involves the
principle of light absorption.
➢ COLORIMETRY:
The constituent measured is colored. (e.i., it absorbs light within the visible
spectrum (350-700 nm light)
Uses ordinary filters to screen light which will strike the solution.
➢ How to deal?
b a.) Report as “greater than
a (upper limit)”
b.) Dilute the original sample
& re-assay, then compute w/
dilution factor.
concentration
➢ 2. Lower limit of detection
- Corresponds to lower portion of the curve that deviates from the
straight line
- 2 solutions of different concentrations give essentially same
absorbance readings, but at the next highest concentration in the
series follows a linear relationship between absorbance &
concentration
How to deal?
Cu = Au x Cs
As
● Blanks (Correction)
- Sets the disturbance to zero
- Commonly used is distilled H₂O
- Ensures the accuracy of the results by eliminating interfering
substances inherent with the sample or the reagent
- Blanks and Standard = Reference Materials
- While controls are used to test & check the accuracy.
Two types of blanks:
1. Reagent blank
- Contains only 1 reagent (no sample)
- Used to correct for the absorbance readings due to other
materials in the reagents.
A = abc = 2-log%T
A = 2-log%T
A = 2-1 = 1.0
Absorptivity
2
➢ Among possible causes of this deviation from Beer’s Law are:
1. Absorbance of solution is too strong for the instrument to
measure
2. More analyte present than the detecting reagent
➢ Note: it is never valid to simply extend the line all the way to zero.
Always determine the limits of the curve.
➢ SPECTROPHOTOMETRY
Spectrophotofluorometer
*Three types of light scatters occur based on the relative size of the light
wavelength (Gauldie, 1981)
*If the wavelength (𝞴) of light is much larger than the diameter (d) of the
particle, where d > 10𝝺, the light scatters forward owing to the destructive
out-of-phase back-scatter, as described by the Mie theory.
* If the wavelength of light is approximately the same as the particle size,
more light scatters in the forward direction than in other directions, as
defined by the Rayleigh-Debye Theory.
NEPHELOMETER
● Larger particles
produce light
scattering that is
more intense but
irregular, with the
scattering being
greater in the
direction of the
incident light beam
REFRACTOMETER
➢ OSMOMETRY
- Osmometry is the measurement of the osmolality of an aqueous
solution such as serum, plasma, or urine.
- As osmotically active particles are added to a solution causing its
abnormality to increase, four other properties of the solution are also
affected.
- These properties are osmotic pressure, boiling point, freezing point,
and vapor pressure. They are called colligative properties of the
solution because they can be related each other and to the
osmolality.
➢ OSMOMETRY
- As the osmolality of a solution increases, (1) the osmotic pressure
increases, (2) the boiling point elevated, (3) the freezing point is
depressed, and (4) the vapor pressure is depressed.
- Osmometry is based on measuring changes in the colligative
properties of solutions that occur owing to variations in particle
concentration.
- Freezing-point depression osmometry is the most commonly used
method for measuring the changes in colligative properties of
solution.