Professional Documents
Culture Documents
Date: 19/03/2020
Name of Faculty: Mr. Darshan Salunke
Lecture No : 5 (as per TT)
pH is defined as pH log H
pH pOH pK w 14
pKw is constant for a given temperature
Neutral pH varies with temp. (7.5 at 0C & 6.5 at 60C) If pH increases pOH
decreases and vice-versa
Natural water pH is in the range of 4-9
Natural waters are slightly basic due carbonates and bicarbonates
Relationships exist between pH, acidity and alkalinity
pH measurement: pH meter
pH meter is used
– Involves potentiometric measurement of hydrogen ion activity
– capable of reading both pH and millivolts
– A pH meter with good electrodes measures pH with 0.1 pH units accuracy under normal
conditions
pH meter has
– A potentiometer
– A glass electrode
– A reference electrode
– A temperature compensation device
pH meters usually have two controls
– Intercept control – parallelly shifts the response curve, between emf and pH, for giving 0
emf with pH 7 buffer
– Slope control – rotates response curve about isopotential point
pH measurement: pH meter
Glass electrode
– A sensor electrode
– Electro motive force (emf) produced in the glass electrode
system linearly varies with the pH of the sample
– Using buffers of known pH values emf is measured by glass electrode system and plotted
against pH for calibrating the meter
– With the calibrated meter emf produced by the sample is measured and pH is estimated by
extrapolation and interpolation
Reference electrode
– A half cell providing constant electrode potential
– Calomel electrode or silver: silver chloride electrode is used
– Has a liquid junction
– The electrode is filled by an electrolyte to proper level to ensure proper wetting of the liquid
junction
Combination electrode: both glass electrode and reference electrode are
incorporated into a single probe
Electrode storage
Keep the electrodes wet when the pH meter is not in use
– Follow manufacturer’s instructions
– Use tap water with conductivity >4000 µmhos/cm rather than
distilled water for short-term electrode storage
– pH 4 buffer is best for glass electrode storage
– Saturated KCl solution is good for glass electrodes and combination electrodes
Before use remove electrodes from storage solution, rinse
with distilled water, and blot dry with soft tissue
– Rinsing and blotting dry are also needed for electrode transfer from one solution to the
next
Prior to use, conditioning of the electrode in a small portion
of the sample for a minute is recommended
– in case of poorly buffered samples the conditioning can be in 3 or 4 successive portions of
the sample
– The conditioned electrode is not rinsed, it is only blot dried
Calibration of pH meter
Transfer electrode(s) into a standard buffer of neutral pH and set isopotential
point on the meter (point of ‘0’ emf)
Transfer electrode(s) into 2nd standard buffer of pH within 2 units from the sample
– Ensure same temperature for both sample and 2nd buffer
– Record temp. and adjust temp. on the meter
– Adjust meter pH to that of the buffer
Transfer electrode(s) to 3rd standard buffer of pH <10 and within 3 pH units from
the sample
– Check if the meter pH is within 0.1 units of the actual - If not then the pH meter is faulty
Calibration frequency
– needed prior to each set of pH measurements
– If pH values vary widely within a set, check with a 3rd buffer of pH within 1 or 2 units from the
sample is needed
pH measurement: Buffer solutions
Use of special low sodium error electrodes is needed for measuring pH >10, at
high temperature, accurately
Liquid membrane electrodes are good for pH below 1
pH measurement can not be made accurately in non- aqueous media,
suspensions, colloids or high ionic strength solutions
Temperature can affect pH measurement by
– Changing the properties of the electrodes
– Brining in chemical equilibrium changes
Buffer solutions deteriorate from mold growth or contamination hence replace
them every 4 weeks once
Store buffer solutions and samples in polyethylene bottles
1. Explain the concept of pH
2. Explain the construction of pH meter.