You are on page 1of 8

A B B M E A S U R E M E N T & A N A LY T I C S   |   I N F O R M AT I O N   |   I N F 1 3/ 1 2 4 - E N R E V. B

Navigator 550
Low level dissolved oxygen wet-section

Sensor assembly spares


replacement procedures

Measurement made easy

Navigator 550
low level dissolved 1 Introduction 2 For more information
oxygen wet-section

These procedures must be carried out by a trained Further information is available from:
technician.   www.abb.com/analytical

or by scanning these codes:


Sales Service
2 N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B

3 Safety 4 Disposal

WARNING WARNING
These procedures must be carried out by a trained Dispose of the old components in accordance with the
technician. guidelines contained in the Operating instructions
(OI/ADS550-EN).
ELECTRICAL

• Isolate all high voltage supplies to the transmitter before


performing replacement procedures.

• The wet-section is vulnerable to electrostatic damage.


5 Dissolved oxygen sensor
Wear an anti-static strap or dismantle the wet-section on assembly location
an anti-static workbench.

• Ensure all electrical connections are kept dry at all times.

GENERAL

• Shut off the external sample supply to the wet-section


and drain the flowcell – refer to the wet-section Operating
instructions (OI/ADS550-EN) for flowcell drainage
options.

• When a procedure is complete, restore power to the


transmitter and sample to the wet-section at the correct
flow rate.

• If necessary, calibrate the wet-section – refer to the wet-


section Operating Instructions (OI/ADS550-EN) for
calibration instructions.

• Perform general cleaning of the wet-section using a damp


cloth only – mild detergent can be used as a cleaning aid.
Do not use Acetone or any organic solvents.

• Dispose of the old components in accordance with the


guidelines contained in the Operating instructions
(OI/ADS550-EN). Sensor contained
within flowcell

Figure 1  Low level dissolved oxygen sensor location


N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B 3

6 Replacing the dissolved


oxygen sensor
Part number: low level dissolved oxygen sensor assembly
(AW502 080)

Removing the dissolved oxygen sensor


assembly
Referring to Figure 2:
1 Disconnect (red) sensor connector A.

2 Unscrew clamping screw B and withdraw the sensor


assembly from flowcell body C.

3 Carefully unscrew connector nut G from sensor D and


withdraw connector body E.

4 Remove and discard O-rings F and H.

Figure 2  Removing the dissolved oxygen sensor assembly


4 N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B

…6  Replacing the dissolved oxygen sensor


Fitting a new dissolved oxygen sensor assembly
Note.  Wet-section enclosure,
CAUTION oxygen sensor container and
protective cap not shown for
• Only install the sensor immediately prior to use. The clarity.
sensor has a limited shelf life and, ideally, should NOT be
stored longer than 6 months. Store under cool conditions.

• Take special care to line up the two pins in the sensor with
their respective sockets before making the connection
O-ring 7/8 in ID
and tightening.

• Take care not to damage the delicate membrane on the O-ring 3/4 in ID
face of the sensor.

• Ensure that the mating surfaces (carrying the electrical


connection) of the sensor and connector body are clean
and completely dry.

• Do not overtighten the clamping screw.

1 Remove the top from the new sensor container.

2 Unscrew the protective cap from the rear of the sensor.

Referring to Figure 3:
3 Fit the smaller of the 2 new O-rings A (3/4 in. ID) onto the
connector body B.

4 Locate sensor C onto connector body B, ensuring the pins


and sockets are engaged correctly and tighten connector Optional flowmeter (if fitted)
nut D onto sensor C.

5 Insert the complete assembly into flowcell E, ensuring the


larger of the 2 new O-rings F (7/8 in. ID) is fitted.

6 Screw in clamping screw G using finger-pressure only to


secure the dissolved oxygen sensor assembly in the flowcell
body.

7 Push (red) sensor connector H onto sensor connector block I


firmly and tighten ONE TURN clockwise.
Pt1000 temperature
Sensor shown compensator
fitted connector (blue)

Figure 3  Fitting and connecting the dissolved


oxygen sensor at the flowcell
N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B 5

Notes
6 N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B

…Notes
N AV I G AT O R 5 5 0   |   LO W L E V E L D I S S O LV E D OX YG E N W E T S E C T I O N  |  I N F 1 3/ 1 2 4 - E N R E V. B 7

ABB Limited
Measurement & Analytics
Oldends Lane, Stonehouse
Gloucestershire, GL10 3TA
UK
Tel:  +44 (0)1453 826661
Fax:  +44 (0)1453 829671
Email:  instrumentation@gb.abb.com

ABB Inc.
Measurement & Analytics
125 E. County Line Road
Warminster, PA 18974
USA
Tel:  +1 215 674 6000
Fax:  +1 215 674 7183

abb.com/measurement


INF13/124-EN Rev. B  11.2019

We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.

We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in
whole or in parts – is forbidden without prior written consent of ABB.

© ABB 2019

You might also like