Professional Documents
Culture Documents
OPERATING MANUAL
ULTRAOX
MODELS: 3, 8, 12, 30, 50, 70, 90, 100, 120, 180, 260, 370, 440, 560
- Index –
-2-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
1. GENERAL INFORMATION
ULTRA-CONTROLO - Projectos Industriais, Lda., was founded in 1987 and is a Portuguese leader
company in Vacuum systems, Pressure producing and Air treatment technologies in Portugal.
ULTRAOX® is ULTRA-CONTROLO's specifically designed plant for medical use.
The extensive scientific knowledge gained from early in the departments of research and development
of its principals and by the research department of the ULTRA-CONTROL, knowledge of international
law, standards and technical rules led to the creation of certified devices intended for medical use,
namely MEDIAR ®, ULTRAAR ®, ULTRAVAC ®, ULTRASEG ® and now ULTRAOX ®.
COPYRIGHT:
This manual contains text, images, diagrams and other illustrations that are protected by
copyright. No part of this document may be reproduced without the written permission of the
Manufacturer. Any violation will be punished by law. All rights reserved.
ULTRA-CONTROLO, reserves the right to make modifications to this operating manual without
prior notice.
1.1. MANUFACTURER
ULTRA-CONTROLO PROJECTOS INDUSTRIAIS, LDA.
PARQUE INDUSTRIAL QUINTA LAVI, ARMZ.8
ABRUNHEIRA
2710-089 SINTRA PORTUGAL
-3-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
USERS MUST READ THIS MANUAL CAREFULLY BEFORE USING THE DEVICE. IT CONTAINS VITAL
INFORMATION FOR THE CONSERVATION AND SAFETY OF THE DEVICE.
OBSERVE THE WARNINGS SHOWN IN ORDER TO PREVENT MALFUNCTIONS OR RISK OF PERSONAL
INJURY.
ENSURE THAT THE OPERATING INSTRUCTIONS ARE ALWAYS ACCESSIBLE FOR INSPECTION NEAR
THE DEVICE.
ULTRA-CONTROLO IS NOT LIABLE FOR ANY DAMAGE OR INJURY RESULTING FROM THE
NON-OBSERVANCE OF THESE SAFETY INSTRUCTIONS OR NEGLIGENCE OF THE USUAL CARE AND
ATTENTION REQUIRED DURING HANDLING, OPERATION, MAINTENANCE OR REPAIR, EVEN IF THIS
IS NOT EXPLICITLY MENTIONED IN THESE OPERATING INSTRUCTIONS.
IF ANY OF THE REGULATIONS CONTAINED IN THESE INSTRUCTIONS - ESPECIALLY WITH REGARD
TO SAFETY - DOES NOT CORRESPOND TO THE LOCAL LEGAL PROVISIONS, THE STRICTER OF BOTH
SHALL PREVAIL.
THESE SAFETY REGULATIONS ARE GENERAL AND VALID FOR VARIOUS TYPES OF MACHINES AND
EQUIPMENT. IT IS THEREFORE POSSIBLE THAT SOME REFERENCES DO NOT APPLY TO THE UNIT(S)
DESCRIBED IN THESE INSTRUCTIONS.
IN ADDITION TO THE OPERATING INSTRUCTIONS, ALL OTHER GENERALLY APPLICABLE LEGAL AND
OTHER MANDATORY REGULATIONS RELEVANT TO ACCIDENT PREVENTION AND ENVIRONMENTAL
PROTECTION MUST BE ADHERED TO AND PASSED ON TO OTHERS!
THESE COMPULSORY REGULATIONS MAY ALSO DEAL WITH THE HANDLING OF HAZARDOUS
MATERIALS OR THE ISSUING/WEARING OF PERSONAL PROTECTIVE EQUIPMENT, OR TRAFFIC
REGULATIONS.
-4-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
ULTRA-CONTROLO manufactures not only conventional but also facilities to the needs of its customers. This
ability to design, planning and execution, made the ULTRA-CONTROL pioneer in many areas of technology of
compressed air and vacuum as well as treatment and filtration of gases in Portugal, with a long and successful
series of innovative systems that will mark its definitive presence in this millennium.
In ULTRAOX ®, the oxygen is produced by pressure swing adsorption and the quality is monitored by a oxygen
sensor before supplying the medical pipeline system.
Each ULTRAOX ®, is tested in an intensive test under the most adverse conditions. These conditions are much
more severe than those required in most medical applications. This test also includes a full simulation of faults,
alarms, a leak test and verification of the conditions of installation on site.
Along with this operating instructions, please read the instructions for use specific to the vacuum pumps.
When operating with ULTRAOX, we recommend the use of PPE, namely ear protector
and face shield.
-5-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
1.5. LABELLING
On the side of the control panel of the plant is affixed the CE marking of the equipment, which is
composed of the information presented below.
1.6. WARRANTY
ULTRA-CONTROLO covers warranty for the period of one year, depending on the models.
Signal lamps, fuses and sensors are not covered.
This warranty period of one (1) year counts from the date of issue of the invoice to the device, under normal
conditions of use and operation, in respect of defects in materials and manpower.
The warranty is not extendable to any component if the defect or malfunction results from improper use or
misuse of the device, such as ambient temperature out of limits, nonconforming installation site, compressed air
in poor conditions and preventative maintenance at fault.
The warranty will be considered void if the plant has been repaired or modified by another entity other than
ULTRA-CONTROLO or its authorized agent.
For other warranty conditions, please see warranty certificate.
Medical Oxygen System is warranty
-6-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
2. ULTRAOX FEATURES
The medical conditions will be those require respiratory treatment with higher concentration of Oxygen
Should not be used for any other purposes since it is classified as a medical device class IIb.
Pressure
Capacity O2
Inlet Outlet
Model (per unit)
(AC) (O2)
l/min m3/h bar bar
3 5,3 0,318 7,5 6
8 13 0,777 7,5 6
12 17,9 1,072 7,5 6
30 51,7 3,1 7,5 6
50 76,7 4,6 7,5 6
70 108,3 6,5 7,5 6
90 133,3 8 7,5 6
100 158,3 9,5 7,5 6
120 206,7 12,4 7,5 6
180 303,3 18,2 7,5 6
260 436,7 26,2 7,5 6
370 608,3 36,5 7,5 6
440 725 43,5 7,5 6
560 928,3 55,7 7,5 6
-7-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-8-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
2.3.1. Compressor
Each ULTRAOX ® is equipped with one, or several high technology compressors with the following main
characteristics:
Inlet air filter
Ensures continuous operation
-9-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
Note: The electric control box can vary according to the type of plant and
the needs of each configuration according to the type of compressor
control and air treatment units.
Optical signals given by the control box:
Information available for remote monitoring through the terminal free voltage:
Presence of voltage Compressor maintenance
Compressor failure Oxygen concentrator failure
-10-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
The plant is not ready to work in potentially explosive environments. Please contact the
manufacturer should such a need arise.
Must be careful when transporting the components of the plant. Always use a forklift or
pallet trucks, pallet components must be belted (figure above). Apply the forks in the
appropriate places.
-11-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-12-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
3. INSTALLATION INSTRUCTIONS
Once the plant is used for medical applications, installer must follow the following notes:
Arrange the right installation of the complete plant according to the manufacturer instructions, assuring
the continuous supply of the oxygen
Verify that the site of installation is suitable for the kind of application, compatible to the produced gas
(ex. by using for the network, only copper degreased pipes, compatible with oxygen, etc). In particular
the quality of the produced gas, the user safety and the environmental conditions shall be considered;
Assure that the oxygen network is appropriate for the use of oxygen, according to local law and
Assure that the final test / commissioning of the complete System of Supply is in accord with the
manufacturer instruction.
Plant must be placed in a plain, clean, dry, well ventilated and protected area, in absence of sparks or
naked flames, inflammable substances (gas, vapour, dusts) that can be fire or explosions cause.
Rooms or areas for supply systems should be dedicated and not used for any other purposes.
ONLY NOMINATED PERSONS SHOULD BE AUTHORIZED TO OPERATE AND ATTEND THE
SUPPLY EQUIPMENT.
The air compressor and the air dryer can be placed in the same room together with the
concentrator.
Environmental temperature must be between 5° and 40° C.
In case the ambient temperature falls below + 10°C, the installation should be equipped with an
auxiliary heating device to guarantee a frost free working conditions.
One group of filled cylinders sufficient for one side of a manifold may be stored in the same room or
area. Empty cylinders disconnected from the supply equipment may be stored pending their removal.
Full and empty cylinders should be segregated and their respective storages areas should be labeled.
Services or containers of combustible gases or liquids should not be permitted within, adjacent to, the
supply system location.
All electrical fittings in supply rooms should be located in fixed positions or protected to minimize the
risk of physical damage.
Clean the area before a definitive placement of the equipment’s that compose the plant.
There should be a minimum distance of 0,8 meters from walls or other obstacles for future
maintenance and parts replacement.
-13-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
Main access must have a distance of 1,0 meters in order to remove components for factory
Technical Assistance Centre major overhauls.
Access to manifold rooms and storage vessels should be level and kept clear.
All doors should be capable of being opened from the inside, at any time, without a key, and should
open outwards.
The doors or gate rooms or enclosures containing the source of supply should be capable of being locked.
At least one emergency exit, which should be free from obstructions at all times and should lead into the
open air or another safe location, should be provided.
A heating system may be used to heat supply system enclosure or storage areas, provided that no part of
the heating system in contact with the air within the room can exceed a temperature of 225˚C and that
cylinders are prevented from coming into contact with the heating system.
Fire-fighting equipment should be provided.
The room or enclosure should be well lit.
o Enclosures (interior or exterior) for supply systems should conform to the following:
o when an enclosure is located near a source of heat such as a furnace, incinerator, or boiler room,
its construction should prevent the cylinder temperature from exceeding 40 ˚ C
o An enclosure should not be located within 3 m of open electrical conductors or transformers
o an enclosure should not be adjacent to oil storage tanks
o an enclosure should comply with local building codes
o an enclosure should have concrete floors
o fences and walls should be of a height no less than 1,75 m
-14-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
Before beginning work on the unit or facility, the following measures should be taken:
Switch off the power.
Take the necessary precautions to prevent it being switched on again.
Make sure that there is voltage.
Ground and short circuit.
Cover or protect neighboring components under electric voltage.
If possible connect the Electric Control Box directly to your General Framework with the supplied cable.
Alternatively, take advantage of the flexible cable that is supplied and place a junction box between the
General Framework and the Electric Control Box with terminal blocks with appropriate section, or install a cable
directly between the two frames.
Make electrical connections in the electrical panel with suitable power cable for the location where ULTRAOX ®
will be installed, should contain three phase conductors, neutral conductor and protective conductor (H07VV-F
5G2,5).
The power cable to the Electric Control Box must be fixed with clamps and glands of good quality, meeting the
rules of sound construction.
The General Board shall have a four-pole circuit breaker curve D 6 or 10 A.
Other operating and automation solutions are possible to apply upon request.
DO NOT MAKE ANY CHANGE IN PARAMETERS AND CONNECTIONS MADE AT THE FACTORY. IF IN
DOUBT CONTACT YOUR SUPPLIER.
-15-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-16-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
FOR A CORRECT USE OF THE PLANT, PLEASE FOLLOW THE FOLLOWING INSTRUCTIONS:
ATTENTION:
THE INTER CONNECTIONS OF THE MAIN COMPONENTS HAVE TO BE MADE ACCORDING TO
THE LABELS INDICATED IN THE FLEXIBLES / CABLES EVENTUALLY SUPPLIED.
CHECK THE FOLLOWING CONDITIONS BEFORE STARTING THE PLANT OR PUT INTO OPERATION:
ATTENTION:
NEVER PLACE ANY BODY PART, INCLUDING HANDS, AT THE ENTRANCE OF ANY VALVES,
YOU CAN HURT YOURSELF IRREMEDIABLY.
Note that at the first starting up, the oxygen pressure vessel is empty and contains air with an oxygen
concentration of 21%, so that the filling up time of the pressure vessel will be longer and some purging time will
be required.
-17-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
STARTING UP PROCEDURES:
Connect the electric control box to the ULTRAOX ® components according with diagram;
Power on the refrigerating air drier and wait until dew point of 2 - 3 ˚C is reached;
Open if shut off valve between the compressor and the air drier;
Open shut off valve between the air drier and the compressed and dried air pressure vessel;
Close shut off valve outlet from compressed and dried air pressure vessel;
Confirm by reading the pressure gauge on compressed and dried air pressure vessel until the compressor
stops;
Open shut off valve from compressed and dried pressure vessel to oxygen concentrator and wait until the
green light turns solid and purity level reaches 95%;
Open shut off valve from oxygen concentrator to oxygen pressure vessel and wait until the pressure
gauge on oxygen pressure vessel reaches 6 bar, yellow light blinking;
When yellow light becomes solid, open shut off valve from oxygen pressure vessel to oxygen
concentrator;
Open the shut off valve from oxygen concentrator to the medical gas pipeline system;
As soon as the oxygen purity level reaches 92%, oxygen can be used in the healthcare facility;
From this point on, the plant stays in autonomous oxygen production regime;
ATTENTION:
IF YOU NOTICE ANY NOISE OR VIBRATION, OR ANY DOUBT PERSISTS, PLEASE
IMMEDIATELY CONTACT OUR TECHNICAL SERVICES (INDICATED ON POINT 2.9 OF THIS
MANUAL)
-18-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
5. MAINTENANCE
PLEASE READ THIS MANUAL CAREFULLY BEFORE USING ULTRAOX. IT CONTAINS VITAL
INFORMATION FOR THE CONSERVATION AND SAFETY OF THE DEVICE.
THE SCHEDULED TASK TIME GAPS CAN EVENTUALLY BE CHANGED DEPENDING ON THE
CONDITIONS OF SERVICE IF THEY SO REQUIRE.
THE MAINTENANCE AND SERVICE TIME INTERVALS SHOWN ARE FOR NORMAL OPERATION.
THE INSPECTION PERIOD MAY DIFFER SLIGHTLY FOR DIFFERENT OPERATING CONDITIONS,
SUCH AS HIGH TEMPERATURE AND/OR HUMIDITY. IF THE OPERATING CONDITIONS ARE SEVERE
AND/OR UNSATISFACTORY, REDUCE THE INSPECTION TIME INTERVALS.
BEFORE PERFORMING ANY MAINTENANCE OPERATION, SHUT DOWN THE MAIN POWER TO
COMPRESSOR AND CLOSE THE COMPRESSOR SHUT OFF VALVE. DAMAGES CAUSED BY USE OF
NON-AUTHORIZED PARTS ARE NOT COVERED BY WARRANTY.
-19-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
DAILY
Verify that the Dew Point displayed on the electronic instrument is correct.
Check the proper operation of the condensate drain systems.
Verify the condenser for cleanliness.
-21-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
MAINTENANCE AND CONSERVATION TASKS MUST BE CARRIED OUT BY AUTHORIZED AND QUALIFIED
PERSONNEL.
THE PLANT IS SUBJECT TO WEAR AND DEGRADATION FROM THE START OF OPERATION SO THAT IT
STATUS MUST BE EVALUATED EVERY TWO YEARS BY ULTRA-CONTROLO TECHNICIANS, AND ITS
MAINTENANCE PLAN MUST BE SCRUPULOUSLY RESPECTED
IN ORDER TO AVOID THE RISK OF SOMEONE INADVERTENTLY STARTING THE PUMP WHILE IT IS
BEING REVIEWED OR INSPECTED, WHICH COULD CAUSE SERIOUS DAMAGE TO THE MAINTENANCE
STAFF, THE POWER SUPPLY TO THE ENGINE SHOULD BE COMPLETELY DISCONNECT. UNLESS THE
PUMP IS CLOSED AND FULLY ASSEMBLED, IT CANNOT BE OPERATED.
NEVER HANDLE A COMPRESSOR THAT IS STILL HOT OR NEAR ITS WORKING TEMPERATURE. YOU
CAN BURN YOURSELF WITH VERY HOT PARTS.
FOLLOW THE RECOMMENDATIONS FOR THE PROPER LEGAL DESTINATION OF THE PARTS, SUPPLIES
AND CLEANING PRODUCTS USED FOR PREVENTIVE MAINTENANCE INTERVENTIONS.
-22-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
Concentrator
Ai treatment
O2 treatment
Compressor
Air dryer
Date Parts List Task Description Operator
Signature
-23-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
ON SITE OPERATIONS:
IN ORDER TO AVOID THE RISK OF SOMEONE INADVERTENTLY STARTING THE PUMP WHILE
IT IS BEING REVIEWED OR INSPECTED, WHICH COULD CAUSE SERIOUS DAMAGE TO THE
MAINTENANCE STAFF, THE POWER SUPPLY TO THE ENGINE SHOULD BE COMPLETELY
DISCONNECT.
It is advisable that all technical staff working with ULTRAOX should get in touch with the manufacturer, the
representative or authorized dealers.
Address and contact of the nearest Technical Assistance Service can be provided on request.
Waste resulting from installation and maintenance activities must be disposed of in an environmentally friendly
manner and in accordance with the applicable legislation.
The plant, at the end of its life, should be disposed of in an environmentally friendly manner and in accordance
with current regulations.
-24-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
6.1. COMPRESSOR
ULTRACECO SC
1 ADMISSION FILTER
2 SCROLL ROTOR
3 STATIC SCROLL
4 COMPRESSED AIR OUTLE
ULTRASCREW
1 COOLING RADIATOR
2 INTAKE FILTER
3 OIL DRAINING VALVE
4 SCREW UNIT
5 OIL TANK
6 MOTOR
-25-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
MAKE SURE BEFORE COMMISSIONING THAT NOBODY IS IN THE DANGER ZONE OF THE SCROLL
COMPRESSOR.
ONLY OPERATE THE SCROLL COMPRESSOR WITH CLOSED ACCESS PANELS.
Important!
Although this compressor has been subjected to a test run at the factory and has again been thoroughly
inspected before shipment, damage during shipment cannot be excluded. For this reason, every compressor
should be checked once again for damage before being commissioned. In addition, it should be observed
during the first operating hours.
If the compressor is newly connected to a power supply, check the direction of rotation of the drive motor!
Turn the main switch ON, quickly turn On and OFF the compressor in order to verify the motor correct rotation
direction.
-26-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
This procedure should be followed on initial start-up, after periods of extended shutdown or
following maintenance procedures.
Qualified personnel must perform the start-up.
Sequence of operations:
Ensure that the connection to the compressed air system is correct and that the piping is suitably
fixed and supported.
Ensure that the condensate drain pipe is properly fastened and connected to a collection system or
container.
Ensure that the by-pass system (if installed) is open and the dryer is isolated
Ensure that the manual valve of the condensate drain circuit is open.
Ensure that the cooling water flow and temperature is adequate (Water-Cooled).
Remove any packaging and other material which could obstruct the area around the dryer.
Ensure the consumption matches with the values of the data plate.
Ensure the fan work properly - wait for its first interventions (Air-Cooled).
Slowly close the central by-pass valve of the system (if installed).
Ensure the drain is regularly cycling - wait for its first interventions.
-27-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
START UP:
Wait a few minutes; verify that the DewPoint temperature displayed on electronic instrument DMC14
is correct and that the condensate is regularly drained.
SHUT DOWN:
Verify that the DewPoint temperature displayed on electronic controller DMC14 is correct.
After a few minutes, switch off the main switch on the control panel of the dryer (pos. 1).
NOTE:
A Dew Point within 0°C and +10°C displayed on the electronic controller is correct according to the possible
working conditions (flow-rate, temperature of the incoming air, ambient temperature, etc.).
The number of starts must be no more than 6 per hour. The dryer must stop running for at least 5 minutes
before being started up again. The user is responsible for compliance with these rules. Frequent starts may
cause irreparable damage.
-28-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
7. LAYOUT
7.1 SPECIFICATION
The diagram represents the pneumatic flow within Medical Oxygen System ULTRAOX.
The plant consists of one compressor, one refrigerating air drier, one compressed and dried air pressure vessel,
one oxygen concentrator, one oxygen pressure vessel, one cyclonic separator filter, one pre-filter, one micro-
filter, one activated carbon filter, one sterilization filter, pressure gauges, pressure relief valve, non-return
valves and shut off valves
A oxygen sensor measures the percentage in volume of oxygen to be fed to the medical gas pipeline.
C1=Compressor
S1=Fridge Dryer
F#=Filter
R1=Compressed Air Vessel
GO=Oxygen Generator
R2=Oxygen Pressure Vessel
SO#=Shut off valves
-29-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
8. ELECTRIC DIAGRAM
Compressor ON Green H1
Compressor failure Red H2
Compressor maintenance Yellow H3
Oxygen concentrator failure Red H4
Compressed and dried air vessel pressure failure Red H5
Presence of voltage Yellow H6
Sound alarm (optional) shut off Red S1
-30-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-31-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
9. VALIDATION
-32-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-33-
ULTRAOX DTFX07.08 04/07/2019
V A C U U M A N D P R E S S U R E T E C H N O L O G Y
-34-
ULTRAOX DTFX07.08 04/07/2019