Professional Documents
Culture Documents
Mech
Mech
2.1 DETECTOR UNIT A thermistor for temperature compensation is located in the detector
unit.
2.1.1 Outline
Transducer Chamber
The Detector Unit consists of the Mixing Chamber for whole blood Mixing Chamber
dilutions, the open type Transducer Chamber and the Hgb Flow Cell.
2.1.2.2 Transducer Chamber (Detector Block No.40 Assy) Hgb Flow Cell
(1) Sheath flow transducer with a ceramic jet nozzle of 0.2 mm inner Figure 2-1: Detector Unit
diameter (front sheath only).
(2) One aperture of 75 um diameter is commonly used for WBC and
RBC detection. 2.1.3 Signal Input/Output
(3) Two nozzles dispense the diluent and lyse from over the transducer
chamber. The volumes of these reagents are determined to 500 uL With PCB No.3057T1 (Analog, Logger, Driver board)
by syringe. - Thermistor analog signals
(4) Lowered detection voltage level (to 58V). - Hgb detection signals
(5) The Hgb Flow Cell (Hgb Unit No.16) is combined with the - Cell detection resistance pulses
transducer.
One sample tube is set in the open-door type holder with a tube Refer to Section 1.7.9 for the collection tubes.
adapter. Refer to Section 1.7.10 for the adaptors for each sample tube.
The piercer is held by the arm located in the top, and the stepping motor Jump to Section 1.7.9 ]
drives it down to pierce the sample cap. Open tube is also applicable. Jump to Section 1.7.10 ]
The control tube or micro tube requires special adaptors. 2.2.2.2 Door Lock
Reference: The CP Unit is shown below. The Cap Piercing Unit of When the piercer is moving down and “Aspirating” is displayed on the
pocH-100i is named as “Aspiration Unit No. 11”. LCD, the sample door is locked by solenoid for security.
2.2.2.3 Sensors
Door Sensor
Adaptor Sensor
Piercer Home
Position
(Upper End)
2.3.2 Features
STM1 STM3
Diluent Syringe Sample Syringe STM2
Piercer Up/down
STM4
Piercer Back/Forth
STM6
Pinch Valve
STM5
W-Pinch Valve