CALIBRATION CERTIFICATE

NUMBER: 23401
Page 1 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900
laboratorio@circutor.com
www.circutor.com

DESCRIPTION Power Analyzer CVM-A1500A-ITF-485-ICT2

MANUFACTURER CIRCUTOR, S.A.


MODEL CVM-A1500A

IDENTIFICATION CVM-A1500A/084222110008
SERIAL NUMBER 084222110008
CUSTOMER CIRCUTOR, S.A.
Vial Sant Jordi s/n
Viladecavalls
08232

CALIBRATION DATE 28th March 2022

Stamp Date of issue: 28th March 2022

Area Manager

Firmado digitalmente por: 39180593R


PERE SALLENT (C:A08513178)
Fecha y hora: 28.03.2022 15:44:25

This certificate is issued in accordance with the conditions of the accreditation granted by ENAC which has evaluated the laboratory’s
calibration and measurement capabilities and its measurement traceability to the SI system of units or other internationally accepted references
(when traceability to SI is not feasible).

ENAC is one of the signatories of the Multilateral Agreement of the European Cooperation for Accreditation (EA), and the international
Laboratories Accreditation Cooperation (ILAC)

Page 1 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022


ENVIRONMENTAL DATA:
Temperature (23 ± 2) °C
Relative humidity < 70%

CALIBRATION DATA
Calibration by: Alberto Matute
Procedure used: PS03-902-P001 Ed.5.0 PS03-902-P003 Ed.5.0
Calibration results: Within specifications
Comments:

DECISION RULE FOR THE STATEMENTS OF CONFORMITY


Non-binary declaration with security zone of w = 1U

Pass: Acceptance limit = Tolerance - Uncertainty = Pass (The measured values were observed in tolerance at
the points tested. The specific false accept risk is up to 2.5%).

Fail: Acceptance limit = Tolerance + Uncertainty = Fail (One or more measured values were observed out of
tolerance at the points tested. The specific false reject risk is up to 2.5%).

Conditional Pass: Acceptance limit = Tolerance + Uncertainty, Tolerance = Conditional Pass (The measured
values were observed in tolerance at the points tested. However, a portion of the expanded measurement
uncertainty intervals about one or more measured values exceeded tolerance. When the measured result is
close to the tolerance, the specific false accept risk is up to 50%.)

Conditional Fail: Acceptance limit = Tolerance, Tolerance + Uncertainty = Conditional Fail (One or more
measured values were observed out of tolerance at the points tested. However, a portion of the expanded
measurement uncertainty intervals about one or more measured values were in tolerance. When the measured
result is close to the tolerance, the specific false reject risk is up to 50%).

nc: No compliance

UNCERTAINTY
The expanded uncertainty measurement corresponds to the standard uncertainty measurement multiplied by
the coverage factor (k=2), so that the probability of coverage corresponds to approximately 95%. The typical
uncertainty of measurement has been determined according to the guide EA-4/02 M.

TRACEABILITY
The traceability of the measures refers to our reference standards calibrated periodically in inter (national)
laboratories accepted by laboratories accredited by ENAC or accredited by entities recognized by ENAC.

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 2 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

STANDARDS

INSTRUMENT SERIAL No. ASSET No. CAL. DATE CERTIFICATE No.


FLUKE 6105A 419673602 LAB-FLK-018 28/07/2021 082262
FLUKE 6106A 419673603 LAB-FLK-019 30/07/2021 082265
FLUKE 6106A 422673626 LAB-FLK-020 01/11/2021 083598
FLUKE 6106A 422673648 LAB-FLK-021 29/07/2021 082266

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 3 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications
TOLERANCE APPLIED IN VOLTAGE
SPECIFICACTION IN PHASE VOLTAGE, ACCORDING TO STANDARD IEC 61000-4-30:2015 Class A 0.1% Unominal
VOLTAGE CA 50 Hz
11 V @ 50 Hz
11,00 V @ L1 11,00 V 0,004 V 0,012 V ±0,230 V Pass
11,00 V @ L2 11,00 V 0,0000 V 0,0062 V ±0,2300 V Pass
11,00 V @ L3 10,99 V -0,0100 V 0,0062 V ±0,2300 V Pass
23 V @ 50 Hz
23,00 V @ L1 23,00 V -0,002 V 0,014 V ±0,230 V Pass
23,00 V @ L2 23,00 V 0,0000 V 0,0068 V ±0,2300 V Pass
23,00 V @ L3 22,99 V -0,011 V 0,014 V ±0,230 V Pass
110 V @ 50 Hz
110,00 V @ L1 110,04 V 0,040 V 0,018 V ±0,230 V Pass
110,00 V @ L2 110,04 V 0,044 V 0,020 V ±0,230 V Pass
110,00 V @ L3 110,04 V 0,040 V 0,018 V ±0,230 V Pass
230 V @ 50 Hz
230,00 V @ L1 230,10 V 0,100 V 0,035 V ±0,230 V Pass
230,00 V @ L2 230,08 V 0,080 V 0,035 V ±0,230 V Pass
230,00 V @ L3 230,07 V 0,070 V 0,035 V ±0,230 V Pass
345 V @ 50 Hz
345,00 V @ L1 345,13 V 0,133 V 0,048 V ±0,230 V Pass
345,00 V @ L2 345,10 V 0,100 V 0,048 V ±0,230 V Pass
345,00 V @ L3 345,11 V 0,110 V 0,048 V ±0,230 V Pass

SPECIFICATION APPLIED IN PHASE VOLTAGE : 0.1 % Reading


480 V @ 50 Hz
480,00 V @ L1 480,10 V 0,100 V 0,068 V ±0,480 V Pass
480,00 V @ L2 480,03 V 0,030 V 0,068 V ±0,480 V Pass
480,00 V @ L3 480,08 V 0,080 V 0,068 V ±0,480 V Pass
600 V @ 50 Hz
600,00 V @ L1 599,95 V -0,047 V 0,077 V ±0,600 V Pass
600,00 V @ L2 599,90 V -0,098 V 0,077 V ±0,600 V Pass
600,00 V @ L3 599,98 V -0,024 V 0,077 V ±0,600 V Pass

SPECIFICACTION IN PHASE VOLTAGE, ACCORDING TO STANDARD IEC 61000-4-30:2015 Class A 0.1% Unominal
VOLTAGE CA 60 Hz
11 V @ 60 Hz
11,00 V @ L1 10,98 V -0,0200 V 0,0062 V ±0,2300 V Pass
11,00 V @ L2 10,99 V -0,009 V 0,013 V ±0,230 V Pass
11,00 V @ L3 10,99 V -0,0100 V 0,0062 V ±0,2300 V Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 4 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

23 V @ 60 Hz
23,00 V @ L1 22,98 V -0,0200 V 0,0068 V ±0,2300 V Pass
23,00 V @ L2 22,98 V -0,018 V 0,014 V ±0,230 V Pass
23,00 V @ L3 22,98 V -0,0200 V 0,0068 V ±0,2300 V Pass
110 V @ 60 Hz
110,00 V @ L1 110,01 V 0,010 V 0,018 V ±0,230 V Pass
110,00 V @ L2 109,99 V -0,010 V 0,018 V ±0,230 V Pass
110,00 V @ L3 110,01 V 0,010 V 0,018 V ±0,230 V Pass
230 V @ 60 Hz
230,00 V @ L1 230,04 V 0,044 V 0,036 V ±0,230 V Pass
230,00 V @ L2 229,99 V -0,010 V 0,035 V ±0,230 V Pass
230,00 V @ L3 230,02 V 0,020 V 0,035 V ±0,230 V Pass
345 V @ 60 Hz
345,00 V @ L1 345,02 V 0,016 V 0,049 V ±0,230 V Pass
345,00 V @ L2 344,98 V -0,020 V 0,048 V ±0,230 V Pass
345,00 V @ L3 345,02 V 0,024 V 0,049 V ±0,230 V Pass

SPECIFICATION APPLIED IN PHASE VOLTAGE : 0.1 % Reading


480 V @ 60 Hz
480,00 V @ L1 479,94 V -0,059 V 0,068 V ±0,480 V Pass
480,00 V @ L2 479,89 V -0,110 V 0,068 V ±0,480 V Pass
480,00 V @ L3 479,94 V -0,058 V 0,069 V ±0,480 V Pass
600 V @ 60 Hz
600,00 V @ L1 599,78 V -0,220 V 0,076 V ±0,600 V Pass
600,00 V @ L2 599,73 V -0,266 V 0,077 V ±0,600 V Pass
600,00 V @ L3 599,81 V -0,186 V 0,077 V ±0,600 V Pass

VOLTAGE CA V NEUTRAL
55,00 V @ 50 Hz LN 54,90 V -0,11 V 0,16 V ±0,28 V Pass
110,00 V @ 50 Hz LN 109,89 V -0,11 V 0,20 V ±0,56 V Pass
230,00 V @ 50 Hz LN 229,65 V -0,35 V 0,15 V ±1,16 V Pass
300,00 V @ 50 Hz LN 299,94 V -0,06 V 0,16 V ±1,51 V Pass
400,00 V @ 50 Hz LN 399,73 V -0,27 V 0,19 V ±2,01 V Pass
500,00 V @ 50 Hz LN 499,74 V -0,26 V 0,36 V ±2,51 V Pass
110,00 V @ 60 Hz LN 109,85 V -0,15 V 0,11 V ±0,56 V Pass
230,00 V @ 60 Hz LN 229,72 V -0,28 V 0,14 V ±1,16 V Pass
500,00 V @ 60 Hz LN 499,29 V -0,71 V 0,22 V ±2,51 V Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 5 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

FREQUENCY
45,00 Hz @ L1 45,00 Hz 0,0010 Hz 0,0062 Hz ±0,0090 Hz Pass
50,00 Hz @ L1 50,00 Hz 0,0000 Hz 0,0063 Hz ±0,0100 Hz Pass
55,00 Hz @ L1 55,00 Hz 0,0010 Hz 0,0064 Hz ±0,0110 Hz Pass
60,00 Hz @ L1 60,00 Hz 0,0000 Hz 0,0065 Hz ±0,0120 Hz Pass
65,00 Hz @ L1 65,00 Hz 0,0010 Hz 0,0066 Hz ±0,0130 Hz Pass

CURRENT CA

0.25 A @ 50Hz
0,250 A @ L1 0,250 A 0,00000 A 0,00058 A ±0,00225 A Pass
0,250 A @ L2 0,250 A 0,00000 A 0,00058 A ±0,00225 A Pass
0,250 A @ L3 0,249 A -0,00060 A 0,00070 A ±0,00225 A Pass
0,250 A @ LN 0,249 A -0,00070 A 0,00069 A ±0,00350 A Pass

0.5 A @ 50Hz
0,500 A @ L1 0,500 A 0,00000 A 0,00058 A ±0,00250 A Pass
0,500 A @ L2 0,500 A 0,00000 A 0,00058 A ±0,00250 A Pass
0,500 A @ L3 0,500 A 0,00000 A 0,00058 A ±0,00250 A Pass
0,500 A @ LN 0,500 A -0,00020 A 0,00066 A ±0,00600 A Pass

1 A @ 50Hz
1,000 A @ L1 1,000 A 0,00040 A 0,00072 A ±0,00300 A Pass
1,000 A @ L2 1,000 A 0,00000 A 0,00060 A ±0,00300 A Pass
1,000 A @ L3 1,000 A 0,00000 A 0,00060 A ±0,00300 A Pass
1,000 A @ LN 0,998 A -0,00160 A 0,00072 A ±0,01100 A Pass

2 A @ 50Hz
2,000 A @ L1 2,001 A 0,00120 A 0,00073 A ±0,00400 A Pass
2,000 A @ L2 2,001 A 0,00060 A 0,00076 A ±0,00400 A Pass
2,000 A @ L3 2,001 A 0,00070 A 0,00075 A ±0,00400 A Pass
2,000 A @ LN 1,996 A -0,0042 A 0,0011 A ±0,0210 A Pass

5 A @ 50Hz
5,000 A @ L1 5,002 A 0,0015 A 0,0012 A ±0,0070 A Pass
5,000 A @ L2 5,002 A 0,0019 A 0,0011 A ±0,0070 A Pass
5,000 A @ L3 5,002 A 0,0017 A 0,0011 A ±0,0070 A Pass
5,000 A @ LN 5,001 A 0,0014 A 0,0020 A ±0,0510 A Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 6 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

0.25 A @ 60Hz
0,250 A @ L1 0,250 A 0,00000 A 0,00058 A ±0,00225 A Pass
0,250 A @ L2 0,250 A 0,00000 A 0,00058 A ±0,00225 A Pass
0,250 A @ L3 0,249 A -0,00100 A 0,00058 A ±0,00225 A Pass
0,250 A @ LN 0,249 A -0,00060 A 0,00070 A ±0,00350 A Pass

5 A @ 60Hz
5,000 A @ L1 5,002 A 0,0017 A 0,0010 A ±0,0070 A Pass
5,000 A @ L2 5,002 A 0,0016 A 0,0010 A ±0,0070 A Pass
5,000 A @ L3 5,002 A 0,00190 A 0,00099 A ±0,00700 A Pass
5,000 A @ LN 5,003 A 0,0031 A 0,0017 A ±0,0510 A Pass

ACTIVE POWER
Measuring ratio: 5 / 5 A

110 V, 5 A, cos φ 1 @ 50 Hz
550 W @ L1 550 W 0,00 W 0,59 W ±3,10 W Pass
550 W @ L2 550 W 0,00 W 0,59 W ±3,10 W Pass
550 W @ L3 550 W 0,00 W 0,59 W ±3,10 W Pass

220 V, 5 A, cos φ 1 @ 50 Hz
1100 W @ L1 1100 W 0,00 W 0,62 W ±4,20 W Pass
1100 W @ L2 1100 W 0,00 W 0,62 W ±4,20 W Pass
1100 W @ L3 1100 W 0,00 W 0,62 W ±4,20 W Pass

220 V, 5 A, cos φ 1 @ 60 Hz
1100 W @ L1 1100 W 0,00 W 0,62 W ±4,20 W Pass
1100 W @ L2 1100 W 0,00 W 0,62 W ±4,20 W Pass
1100 W @ L3 1100 W 0,00 W 0,62 W ±4,20 W Pass

110 V, 5 A, cos φ 0.7i @ 50 Hz


385 W @ L1 385 W 0,00 W 0,58 W ±2,77 W Pass
385 W @ L2 385 W 0,00 W 0,58 W ±2,77 W Pass
385 W @ L3 385 W 0,00 W 0,58 W ±2,77 W Pass

220 V, 5 A, cos φ 0.7i @ 50 Hz


770 W @ L1 770 W 0,00 W 0,61 W ±3,54 W Pass
770 W @ L2 770 W 0,00 W 0,61 W ±3,54 W Pass
770 W @ L3 770 W 0,00 W 0,61 W ±3,54 W Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 7 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

220 V, 5 A, cos φ 0.7i @ 60 Hz


770 W @ L1 769 W -1,00 W 0,61 W ±3,54 W Pass
770 W @ L2 769 W -1,00 W 0,61 W ±3,54 W Pass
770 W @ L3 769 W -1,00 W 0,61 W ±3,54 W Pass

110 V, 5 A, cos φ 0.5i @ 50 Hz


275 W @ L1 275 W 0,00 W 0,58 W ±2,55 W Pass
275 W @ L2 275 W 0,00 W 0,58 W ±2,55 W Pass
275 W @ L3 275 W 0,00 W 0,58 W ±2,55 W Pass

220 V, 5 A, cos φ 0.5i @ 50 Hz


550 W @ L1 550 W 0,00 W 0,60 W ±3,10 W Pass
550 W @ L2 550 W 0,00 W 0,60 W ±3,10 W Pass
550 W @ L3 550 W 0,00 W 0,60 W ±3,10 W Pass

220 V, 5 A, cos φ 0.5i @ 60 Hz


550 W @ L1 549 W -1,00 W 0,60 W ±3,10 W Pass
550 W @ L2 548 W -2,00 W 0,60 W ±3,10 W Pass
550 W @ L3 549 W -1,50 W 0,72 W ±3,10 W Pass

110 V, 2.5 A, cos φ 1 @ 50 Hz


275 W @ L1 275 W 0,00 W 0,58 W ±2,55 W Pass
275 W @ L2 275 W -0,20 W 0,66 W ±2,55 W Pass
275 W @ L3 275 W 0,00 W 0,58 W ±2,55 W Pass

220 V, 2.5 A, cos φ 1 @ 50 Hz


550 W @ L1 550 W 0,00 W 0,59 W ±3,10 W Pass
550 W @ L2 549 W -1,00 W 0,59 W ±3,10 W Pass
550 W @ L3 550 W 0,00 W 0,59 W ±3,10 W Pass

110 V, 2.5 A, cos φ 0.7i @ 50 Hz


192,5 W @ L1 192,0 W -0,500 W 0,076 W ±2,385 W Pass
192,5 W @ L2 192,0 W -0,500 W 0,076 W ±2,385 W Pass
192,5 W @ L3 192,0 W -0,500 W 0,076 W ±2,385 W Pass

220 V, 2.5 A, cos φ 0.7i @ 50 Hz


385 W @ L1 385 W 0,00 W 0,59 W ±2,77 W Pass
385 W @ L2 385 W 0,00 W 0,59 W ±2,77 W Pass
385 W @ L3 385 W 0,00 W 0,59 W ±2,77 W Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 8 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

110 V, 2.5 A, cos φ 0.5i @ 50 Hz


137,5 W @ L1 137,0 W -0,500 W 0,071 W ±2,275 W Pass
137,5 W @ L2 137,0 W -0,500 W 0,071 W ±2,275 W Pass
137,5 W @ L3 137,0 W -0,500 W 0,071 W ±2,275 W Pass

220 V, 2.5 A, cos φ 0.5i @ 50 Hz


275 W @ L1 275 W 0,00 W 0,58 W ±2,55 W Pass
275 W @ L2 275 W 0,00 W 0,58 W ±2,55 W Pass
275 W @ L3 275 W 0,00 W 0,58 W ±2,55 W Pass

REACTIVE POWER
Measuring ratio: 5 / 5 A

110 V, 5 A, sen φ 0.7 @ 50 Hz


385 var @ L1 384 var -0,70 var 0,69 var ±4,85 var Pass
385 var @ L2 384 var -1,00 var 0,58 var ±4,85 var Pass
385 var @ L3 385 var -0,30 var 0,69 var ±4,85 var Pass

220 V, 5 A, sen φ 0.7 @ 50 Hz


770 var @ L1 770 var -0,30 var 0,71 var ±8,70 var Pass
770 var @ L2 769 var -1,00 var 0,58 var ±8,70 var Pass
770 var @ L3 769 var -0,70 var 0,71 var ±8,70 var Pass

220 V, 5 A, sen φ 0.7 @ 60 Hz


770 var @ L1 770 var 0,00 var 0,61 var ±8,70 var Pass
770 var @ L2 770 var 0,00 var 0,58 var ±8,70 var Pass
770 var @ L3 770 var 0,00 var 0,61 var ±8,70 var Pass

110 V, 5 A, sen φ 0.5 @ 50 Hz


275 var @ L1 274 var -0,80 var 0,66 var ±3,75 var Pass
275 var @ L2 274 var -1,00 var 0,58 var ±3,75 var Pass
275 var @ L3 274 var -0,80 var 0,66 var ±3,75 var Pass

220 V, 5 A, sen φ 0.5 @ 50 Hz


550 var @ L1 549 var -1,00 var 0,60 var ±6,50 var Pass
550 var @ L2 549 var -1,00 var 0,60 var ±6,50 var Pass
550 var @ L3 549 var -1,00 var 0,60 var ±6,50 var Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 9 of 10
CIRCUTOR S.A.
Dpto. Laboratorio Calibración y Ensayos
Vial Sant Jordi s/n (08232 -Viladecavalls)
Spain
Tel: 937452900

Certificate number: 23401 Date: 28th March 2022

Nominal Mesured Error Calibration Device Compliance


value value uncertainty specification specifications

220 V, 5 A, sen φ 0.5 @ 60 Hz


550 var @ L1 550 var 0,00 var 0,60 var ±6,50 var Pass
550 var @ L2 550 var 0,00 var 0,60 var ±6,50 var Pass
550 var @ L3 550 var 0,20 var 0,68 var ±6,50 var Pass

110 V, 2.5 A, sen φ 0.7 @ 50 Hz


192,5 var @ L1 192,0 var -0,500 var 0,076 var ±2,925 var Pass
192,5 var @ L2 192,0 var -0,500 var 0,076 var ±2,925 var Pass
192,5 var @ L3 192,0 var -0,500 var 0,076 var ±2,925 var Pass

220 V, 2.5 A, sen φ 0.7 @ 50 Hz


385 var @ L1 384 var -1,00 var 0,59 var ±4,85 var Pass
385 var @ L2 384 var -1,00 var 0,59 var ±4,85 var Pass
385 var @ L3 384 var -1,00 var 0,59 var ±4,85 var Pass

110 V, 2.5 A, sen φ 0.5 @ 50 Hz


137,5 var @ L1 137,0 var -0,500 var 0,071 var ±2,375 var Pass
137,5 var @ L2 137,0 var -0,500 var 0,071 var ±2,375 var Pass
137,5 var @ L3 137,0 var -0,500 var 0,071 var ±2,375 var Pass

220 V, 2.5 A, sen φ 0.5 @ 50 Hz


275 var @ L1 274 var -1,00 var 0,58 var ±3,75 var Pass
275 var @ L2 274 var -1,00 var 0,58 var ±3,75 var Pass
275 var @ L3 274 var -1,00 var 0,58 var ±3,75 var Pass

The results of this Certificate refer only to the moment and conditions in which the measurements were made.
The results of the measurement and declaration of compliance with the specification refer only to the instrument calibrated.

This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory and ENAC.

Page 10 of 10