QUALIS INDONESIA
Kornite Akedit sl l{asional
LK-trg- IDTT
Calibration Laboratory
alis
lEd Oftce & faberatory Marketi4gOffre
-lL Pei::4;*ran no. 17, Gandasari, Kec, JatiuwunB, Tanterang 15137 - INDONESIA Artha Gading Niaga Blok H no. 1-2, Kelapa Gading Jakarta Ut3ra 14240 - INDONESIA
T: {A21} 55}€5 2583/8611 F: (021) 5565 2489 ll www.qualis-indonesia.com T: (021) 4585 0885 ll F: (021) 4585 0661 ll www.qualis-indonesia.com

t ert te
Sertt
Qtl0622ls0t497

IDEFJTTTAS PEMIUK IDENTITASATAT


Owner ldenity Equipment tdentity
Narna PT QUALIS INDONESIA Nama Digital Thermometer
Norne Name

Alamat Jl. Padjajaran No.17, Ds. Gandasari, Kec. Merek UNI-T


Addrrs Jatruwung, Tangerang Brand

Tanggal Diterima 2l Juni 2022 Tipe/Nomor Seri uT 320D / C180025475


Date of Aeceptonce Type/Seriol Number (aE/18/ELl11s)

Tan*eal Dikalibrasi 23 Juni2A22 RentanE Ukur 0 - 1000 0c


Date o! Colibroted Measur|ng Ronqe

Lokasi Kalibrasi PT QUALIS INDONESIA Resolusi 0.Loc


LocGtion oJ Calibrated Resolution

Metode Kalibrasi T5Q-CAL-IK-052 Alat Standar QE{TT lcA-Au


Colibrotion Method Stondord Equipment
Standar Acuan Euramet CG 11 Vers 2.0 Te rtelus u r Sl melalui standar
Refierence Standard Troceability
LK-150-IDN

Hasit lbtibrasi
Cotibrotion Resuft
Hasil kalibrasi diperoleh dari rata-rata pengukuran yang terdapat pada lampiran. Ketidakpastian ini merupakan ketidakpastian
bentangan yang diperoleh dari sumber-sumber ketidakpastian tipe A dan tipe B sesuai dengan.iCGM 100 :2OOS "Evaluotion and
Measurement Doto-Guide to The Expression of Uncertainty in Meosurement" pada tingkat kepercayaan 95% dan faktor cakupan k = 2.
The .€llbrstian resu/ts sre obtoined from the overage medsurements contdined in the dppendix. The calibration uncertainty is expanded t'rom type A dnd type B
uncettointy sources in accordoflce with ICGM 1OA : 2OO8 "EvoluGfion *nd Measuremeat Datd-Gt ide to The Expression of llncettoin+, in Measurefient" with d confideflce
level oJ 95% and coveroge foctor k = 2.

Tangerang, 24 Juni2022
ManaEer Laboratorium Kalibrasi
Calibration Manager

TSQ-CAL-FM{60 Page I ol 2

Ca&ration and TestingSewicesComplieswith SNI tSO/lEC 77OE:2O77 Requirements


The results relate only to the items tested. This report shall not be reproduced except in full.
Do not copy this report without written consent from authorized personnel of PT. Qualis lndonesia
PT. QUALIS INDONESIA
Xomite Akreditasi Nasional
LK-269-IDN
Calibration Laboratory
alis

No. Sertifi k at I Cenifrcne No. Qll O622IAAL a9T

IAMPIRAN SERTIF! KAT KATIBRASI


Attachme nt o! Colibration Ceftifi cation

Nama
DigitalThermometer
Temoerature Ruane t 1zt.et 1.1) "c
Room Temperature

cLBoo2647 5 IAE/ L8 / EL/ L13)


Kelembaban ' (s8 t 3.4) % RH
Humidifi

HASU- I(AUS8AS
Result of Calibration

c TX.(1000oC)

kttins Alat Penuniukan standard Penuniukan Alat Koreksi


let value Stsndard Reading lntrument Reoding Correction
se or oa oa

0 -0.19 -0.5 0.31


100 99.87 99.s n )-7

200 199.85 199.5 0.35


300 299.8L 300.2 -0.39
400 399.83 400.8 -0.97
500 4s9.83 s(8.7 -0.87
500 599.83 600.4 -0.57
700 699.83 700.5 -0"67

800 799.8L 80r-1 -1.29

900 899.81. 901.4 -1.59


1000 999.81 1001.2 -1.39

Ketidakpastian/Uncertainq, (95%) eC t : 0.99

-FM-050
Page 2 of 3

The results relate only to the items tested. This report shall not be reproduced except in full.
Do not copy this report without written consent f rom authorized personnel of PT. Qualis lndonesia
PT. Qt .{LIS N{DONESIA
LK-269-iDN
Calibration Laboratorv
alis
No, Sertifikat ! Certificote No. qllffizZlffiL497

r T2 (1000 oC)

Penuniukan
Setting Alat sta nda rd Penuniukan Alat Koreksi
Set value Stondord Reoding lntrument Reading Correction
oa sc oa ec

0 -0.23 {.6 0.37


100 99.86 99.6 0.26
200 199.84 199.8 0.04
3ffi 299.83 300.3 -D.47
4CICI
lqq e" 400.7 -s.87
500 499.83 500.s -0.67
600 s99.83 600.8 -0.97
700 699.83 701.6 -1.77
800 799.81 801.4 -1.59
900 899.81 901.7 -1.89
1000 999.81 1001.4 -1.59
Ketidakpastian/Uncertainty (95%) sC t 0.99

tS

Page 3 of 3
The results relate only to the items tested. This report shall not be reproduced except in full.
Do not copy this report without written consent from authorized personnel of PT. Qualis lndonesia
PT. Qualis lndonesia ll Testing, Certification, lnspection, Calibration
Head Office & Laboratory
Jl. Paj:jaran No. 17, Gandasari, Kec. Jatiuwung, Tangerang iakarta utara 142rlo - INDONESIA
T: (021) 5565 2s83/86llF:tO21) 5565 2489 ll www.qualis qualis-indonesia.com

You might also like