HARMONIC® Shears

Assembly Guide

Instrument
Use the torque wrench (already
Assembly mounted to the shaft) to tighten the
instrument onto the Hand Piece. Turn
the torque wrench clockwise while
While holding the holding only the gray Hand Piece until it
Hand Piece in a clicks twice.
vertical orientation,
• Note: Do not use any other means
attach the Hand Piece
than the torque wrench to attach or
to the instrument by
detach the instrument from the hand
rotating the instrument
piece.
onto the Hand Piece
in a clockwise rotation • Note: Do not torque the instrument by

Hand Piece
(finger tight only). hand without the torque wrench.
• Note: Hold only the gray Hand Piece
handle while applying the torque
wrench.

Pre-Run Testing

Connect the assembled Hand Piece


and instrument to the generator and
The test prompt screen appears when the initiation sequence is complete. Activate the
turn the generator power on. Follow
instrument with the jaws open for 2 seconds to run test until the HARMONIC® ready screen is
On-Screen Instructions to complete
displayed.
initiation.

For complete indications, contraindications, warnings, precautions and adverse reactions,


please reference full package insert.