Professional Documents
Culture Documents
Program
Objectives
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 2
Quality Control: Why Is It Important?
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 3
ACR 2018 DM QC Manual
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 4
2018 ACR DM QC Manual
Purpose Use
• Standardized QC Testing for all • Can Only Be Used for Sites Performing Standard
2D and DBT Exams Only.
Manufacturers that are Clinically
Relevant • No Contrast Enhancement – If the site uses I-
View, they must use Hologic’s QC Manual for
• Standardized Test Frequencies all QC Testing
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 6
Medical Equipment Evaluation (MEE)
• To Verify System Meets • Verify System • Visual Indication of Use • For Flat Paddles, Chest
Section 900.12(b) of Anchoring of Large or Small Focal Wall Edge shall be
FDA’s Final Rule for Spot (Pre and Post Straight and Parallel to
Mammography • Can Recover from Exposure) the Edge of the Image
Power Interruption Receptor and not
• To Comply with MEE- • Visual Indication of deflect more than 1cm
Only Required Tests • Grid is Out for Target Material Used when compressed.
Magnification (Pre and Post
Procedures Exposure) • Manual Selection of
mAs
• Magnification Value is • Compression Device
Between 1.4 to 2.0 • Manual Technique
• Power and Fine Factors shall be
• Collimator Light Output Adjustments displayed prior to
is Minimum 160 Lux making an exposure
(15 Foot-Candles) at • Compression Paddles
100 cm or maximum – Sizes and Special • AEC Technique Factors
SID (whichever is less) Purpose Displayed after the
Exposure has been
Made
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 7
ACR DM Phantom Image Quality
• To Verify Image Acquisition 18 Test Objects Present • Images shall be taken for both 2D • CNR and SNR - 2D Only
Chain is Providing and DBT. (If DBT is installed) The • Measurements are Made
Consistent Image Quality • 6 Fibers Window Width (WW) and Window Manually
Level (WL) settings shall be • SNR ≥ 40.0
• 6 Speck Groups optimized for visualization and • CNR ≥ 2.0
• To Verify No Clinically recorded for future evaluations. • CNR to be ≥ 85% of
Significant Artifacts Exist • 6 Masses Previous Year’s CNR
for both 2D and DBT • Minimum Score of 2 Fibers, 3 • Distance Measurement - 2D and
modes. Speck Groups, and 2 Masses DBT
Required for 2D and DBT images.
• Use Capture App Digital Ruler
The DBT image is scored using the
• To Verify that SNR and slice in which the test objects are
to Verify Phantom Width
CNR Values are Adequate best visualized. • 70.0 mm ± 14.0 mm
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 8
DBT Z-Resolution (New for 2018 QC Manual)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 9
DBT Z-Resolution (cont.)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 10
Spatial Resolution
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 11
DBT Volume Coverage
(New for 2018 QC Manual)
• Test to assure that the entire breast volume
is imaged during DBT acquisition
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 12
DBT Volume Coverage (cont.)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 13
AEC Performance
• Pass/Fail Criteria:
• 4 cm SNR ≥ 40
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 14
Average Glandular Dose (AGD)
• Pass/Fail Criteria:
• AGD ≤ 3.0 - Failure must be Corrected
Immediately
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 15
Unit Checklist
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 16
AWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 17
AWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 18
AWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 19
RWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 20
RWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 21
RWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 22
RWS Monitor QC
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 23
RWS Monitor QC (Monthly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 24
RWS Monitor QC (Monthly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 25
AWS Monitor QC (Monthly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 26
RWS Monitor QC (Monthly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 27
Manufacturer Calibrations (if applicable)
• Applicable if manufacturer’s
documentation includes routine
calibration(s)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 28
Collimation Assessment
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 29
MEE or Troubleshooting – Collimation Assessment
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 30
MEE or Troubleshooting – Beam Quality (HVL)
• Pass/Fail Criteria:
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 31
MEE or Troubleshooting – kVp Accuracy & Reproducibility
• Pass/Fail Criteria:
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 33
Troubleshooting – Ghost Image Evaluation
• Pass/Fail Criteria:
• Ghosting Index must be less than ± 0.3
• Failures must be corrected prior to clinical use
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 34
MP Involvement in
Equipment
Adjustments,
Changes, or Repairs
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 35
ACR 2018 DM QC Tests - Technologist
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 36
ACR DM Phantom Image Quality (Weekly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 37
Compression Thickness Verification (Monthly)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 38
Visual Checklist (Monthly)
• *Locks operate
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 39
Compression Force Verification (Semi-Annually)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 40
Manufacturer’s Detector Calibration (if applicable)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 41
Repeat Analysis (Optional)
ACR 2018 Digital Mammography Quality Control Program Internal Use Only 42
Thank You