Content

Error code description ................................................................................................................... 2

Logging ................................................................................................................................................. 3

Error codes ..................................................................................................................................... 149

This error code list is intended for use together with AK 98 dialysis machine, program version 1.1.0.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 1 (381)
Program version 1.1.0
C EN 2015-03-26

Error code description


An error code contains thirteen digits.
Error code example: 0933 002 000 007
The error codes are sorted in numerical order starting with the lowest number.
The following information is shown in each error code:
• Digit one and two represent the node (N) that created the alarm.
• Digit three and four represent the main error group, high level categorization (H).
• Digit five, six and seven represents the subcategory, low level categorization (L).
• Digit eight, nine and ten represents the serial number (S), unique to the combination of N, H and L.
• Digit eleven, twelve and thirteen represents complementary (silent) information.
When it is a silent alarm the first of the three digits is 0 otherwise it is 1.
The second digit shows in which operating mode the error occurred. There are ten different modes.
The last digit shows the checksum of all the information in the error code.


2 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Logging
OM_MainStateFcn 0 0x00000
OM_MenuCtrlFcn 1 0x00001
OF_Dummy 2 0x00002
OF_ApBank 3 0x00003
OF_AirDetLmp 4 0x00004
OF_ArtPrLmp 5 0x00005
OF_AttnEvalFcn 6 0x00006
OF_AttnLmp 7 0x00007
OF_TreatmentCount 8 0x00008
OF_BiCarbAcetateTime 9 0x00009
OF_BloodFlowPathLmp 10 0x0000A
OF_BlPumpKnob 11 0x0000B
OF_BlPumpLmp 12 0x0000C
OF_BPMLmp 13 0x0000D
OF_BuzzCtrl 14 0x0000E
OF_CalibFcnMainstate 15 0x0000F
OF_CalibFcnSubstate 16 0x00010
OF_ConcBank 17 0x00011
OF_ConcPresModeAlt 18 0x00012
OF_DisBank 19 0x00013
OF_DisPreset 20 0x00014
OF_E2PromScanFcn 21 0x00015
OF_FchAction 22 0x00016
OF_FluidBypLmp 23 0x00017
OF_FluidFlowPathLmp 24 0x00018
OF_HepAutoStop 25 0x00019
OF_HoldLmp 26 0x0001A
OF_InfoDispMain 27 0x0001B
OF_InfoDispSub 28 0x0001C
OF_InternalSupport 29 0x0001D
OF_KeyBoard 30 0x0001E
OF_SelectKeyReset 31 0x0001F
OF_CancelKeyReset 32 0x00020
OF_ManInputMain 33 0x00021
OF_ManInputSub 34 0x00022
OF_Mm90CondFcn 35 0x00023
OF_MuteLmp 36 0x00024
OF_NonDiffTime 37 0x00025
OF_OnOffLmp 38 0x00026
OF_PassedTime 39 0x00027
OF_PcPresetActive 40 0x00028


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 3 (381)
Program version 1.1.0
C EN 2015-03-26

OF_PcPresetCommandFcn 41 0x00029
OF_PresetCalc 42 0x0002A
OF_PrimingLmp 43 0x0002B
OF_ProfNaAcFcn 44 0x0002C
OF_ProfNaFcn 45 0x0002D
OF_ProfHCO3Fcn 46 0x0002E
OF_ProfUfFcn 47 0x0002F
OF_ProfParamFcn 48 0x00030
OF_ProtectiveSet 49 0x00031
OF_ProtVprLimSet 50 0x00032
OF_RealTime 51 0x00033
OF_Recover 52 0x00034
OF_RemainTime 53 0x00035
OF_ResetValues 54 0x00036
OF_RinseBackLmp 55 0x00037
OF_SafeMem 56 0x00038
OF_SelectLmp 57 0x00039
OF_StartStopLmp 58 0x0003A
OF_TechLmp 59 0x0003B
OF_TIMECOUNT 60 0x0003C
OF_TimeDisp 61 0x0003D
OF_TimeLmp 62 0x0003E
OF_UfAutoCal 63 0x0003F
OF_VenPrLmp 64 0x00040
OF_PresetStore 65 0x00041
OF_DiascanFcn 66 0x00042
OF_DisinfLmp 67 0x00043
OF_NextLmp 68 0x00044
OF_PreviousLmp 69 0x00045
OF_BloodIncreaseLmp 70 0x00046
OF_BloodDecreaseLmp 71 0x00047
OF_BloodMenuLmp 72 0x00048
OF_FluidMenuLmp 73 0x00049
OF_IncreaseLmp 74 0x0004A
OF_DecreaseLmp 75 0x0004B
OF_ManPrimTO 76 0x0004C
OF_RinseBackTO 77 0x0004D
OF_UfdChangeTime 78 0x0004E
OB_AcConcBankSel 179 0x000B3
OB_ApBankNr 180 0x000B4
OB_BcConcBankSel 181 0x000B5
OB_BlTubeIx 182 0x000B6
OB_CalErrNr 183 0x000B7


4 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OB_ChemBankNr 184 0x000B8


OB_CProgVer0 185 0x000B9
OB_CProgVer4 186 0x000BA
OB_CProgVer5 187 0x000BB
OB_FieldSelect 188 0x000BC
OB_LogDecade 189 0x000BD
OB_LogFieldId 190 0x000BE
OB_LogFormat0 191 0x000BF
OB_LogFormat1 192 0x000C0
OB_LogFormat2 193 0x000C1
OB_LogSel0 194 0x000C2
OB_LogSel1 195 0x000C3
OB_LogSel2 196 0x000C4
OB_LogType0 197 0x000C5
OB_LogType1 198 0x000C6
OB_LogType2 199 0x000C7
OB_ProfHCO3Type 200 0x000C8
OB_ProfModelSel 201 0x000C9
OB_ProfNaAcType 202 0x000CA
OB_ProfNaType 203 0x000CB
OB_ProfParamSel 204 0x000CC
OB_ProfUFType 205 0x000CD
OB_ProfUFTypeSet 206 0x000CE
OB_TransSelect 207 0x000CF
OB_ValveSelect 208 0x000D0
OB_DiascanMeasMode 209 0x000D1
OB_DiascanWatsonGender 210 0x000D2
OB_DiascanIndex 211 0x000D3
OB_ExtVersionReport0 212 0x000D4
OB_ExtVersionReport1 213 0x000D5
OB_ExtVersionReport2 214 0x000D6
OB_ExtVersionReport3 215 0x000D7
OB_ExtVersionReport4 216 0x000D8
OB_ExtVersionReport5 217 0x000D9
OB_ExtVersionReport6 218 0x000DA
OB_MenuChemBankNr 219 0x000DB
OB_MenuHeatBankNr 220 0x000DC
OB_MenuRinseBankNr 221 0x000DD
OB_AssPrimBankSel 222 0x000DE
OB_AirDetectorStatus 223 0x000DF
OB_DevSoftwareAttnConfirmedOnce 224 0x000E0
OB_UFV_Second_Deviation_Alarm 225 0x000E1
OD_Dummy 226 0x000E2


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 5 (381)
Program version 1.1.0
C EN 2015-03-26

OD_DiffTimeACBC 227 0x000E3


OD_StartTimeACBC 228 0x000E4
OD_StopTimeACBC 229 0x000E5
OD_HardFaultLR 230 0x000E6
OD_HardFaultPC 231 0x000E7
OD_HardFaultPSR 232 0x000E8
OD_HardFaultCFSR 233 0x000E9
OI_AccACBCTime 234 0x000EA
OI_AccNbrTreatmentsBiCart 235 0x000EB
OI_AccNbrTreatmentsLiqBic 236 0x000EC
OI_AirDetTxAdjust 237 0x000ED
OI_BattFailedCnt 238 0x000EE
OI_BattMinutes 239 0x000EF
OI_BicarbC1 240 0x000F0
OI_BicarbC2 241 0x000F1
OI_BicarbC3 242 0x000F2
OI_BicarbC4 243 0x000F3
OI_BicarbC5 244 0x000F4
OI_BicarbC6 245 0x000F5
OI_BicarbC7 246 0x000F6
OI_BicarbC8 247 0x000F7
OI_BicarbC9 248 0x000F8
OI_BicarbC10 249 0x000F9
OI_BicarbC11 250 0x000FA
OI_BicarbC12 251 0x000FB
OI_BlFlLowLimSupFlow 252 0x000FC
OI_BloodFlSet 253 0x000FD
OI_BloodFlSetHL 254 0x000FE
OI_BloodFlSetLl 255 0x000FF
OI_BlTubeDiam 256 0x00100
OI_AssistedPriming 257 0x00101
OI_BPMMeasInt 258 0x00102
OI_BPMSystPrHighRange 259 0x00103
OI_BPMSystPrLowRange 260 0x00104
OI_BPMDiastPrHighRange 261 0x00105
OI_BPMDiastPrLowRange 262 0x00106
OI_BPMMeanPrHighRange 263 0x00107
OI_BPMMeanPrLowRange 264 0x00108
OI_BPMPulseHighRange 265 0x00109
OI_BPMPulseLowRange 266 0x0010A
OI_CalErrCoeff 267 0x0010B
OI_CalErrOffs 268 0x0010C
OI_CalErrCorr 269 0x0010D


6 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OI_CalRefSet 270 0x0010E


OI_ChemMixRatio 271 0x0010F
OI_ChemRinseTemp 272 0x00110
OI_CondASet 273 0x00111
OI_CondBHl 274 0x00112
OI_CondBLl 275 0x00113
OI_CondBSet 276 0x00114
OI_CondBStart 277 0x00115
OI_CondBStop 278 0x00116
OI_CondHl 279 0x00117
OI_CondLl 280 0x00118
OI_CondSet 281 0x00119
OI_CondStart 282 0x0011A
OI_CondStop 283 0x0011B
OI_DegassPrSet 284 0x0011C
OI_DiagBloodPumpRPMSet 285 0x0011D
OI_DiagHepFreq 286 0x0011E
OI_Dose 287 0x0011F
OI_DoseA 288 0x00120
OI_DoseB 289 0x00121
OI_EnteredCalPoints 290 0x00122
OI_HCO3Set 291 0x00123
OI_HCO3StartStopDelta 292 0x00124
OI_HCO3StartValue 293 0x00125
OI_HCO3StopValue 294 0x00126
OI_HepSet 295 0x00127
OI_HepStopLim 296 0x00128
OI_DaysSinceLastDisinf 297 0x00129
OI_HPGSet 298 0x0012A
OI_IntegerDiffTimeACBC 299 0x0012B
OI_MinUfRate 300 0x0012C
OI_NaSet 301 0x0012D
OI_NaStartStopDelta 302 0x0012E
OI_NaStartValue 303 0x0012F
OI_NaStopValue 304 0x00130
OI_NaAcSet 305 0x00131
OI_NaAcStartStopDelta 306 0x00132
OI_NaAcStartValue 307 0x00133
OI_NaAcStopValue 308 0x00134
OI_NextReadyDay 309 0x00135
OI_NonDiffTime 310 0x00136
OI_PassedTime 311 0x00137
OI_PassedTimeIsol 312 0x00138


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 7 (381)
Program version 1.1.0
C EN 2015-03-26

OI_PrevBloodFlSet 313 0x00139


OI_PrimingUfRate 314 0x0013A
OI_PrimVolLimit 315 0x0013B
OI_ExtraPrimVolLimit 316 0x0013C
OI_RecirUfRate 317 0x0013D
OI_TubeSetFilled 318 0x0013E
OI_RecirTime 319 0x0013F
OI_ProfHCO3RemainTime 320 0x00140
OI_ProfHCO3Time 321 0x00141
OI_ProfNaAcRemainTime 322 0x00142
OI_ProfNaAcTime 323 0x00143
OI_ProfNaRemainTime 324 0x00144
OI_ProfNaTime 325 0x00145
OI_ProfUfFirstCalc 326 0x00146
OI_ProfUFCnt2Sec 327 0x00147
OI_ProfUfDelta 328 0x00148
OI_ProfUFRate 329 0x00149
OI_ProfUFStart 330 0x0014A
OI_ProfUFStop 331 0x0014B
OI_ProfUFTime 332 0x0014C
OI_ProfUFStepsManual 333 0x0014D
OI_ProfUFIntervalsManual 334 0x0014E
OI_ProgFlowCnt 335 0x0014F
OI_RbVolSet 336 0x00150
OI_RemainTime 337 0x00151
OI_RemainTimeDiff 338 0x00152
OI_RemainTimeIsol 339 0x00153
OI_RemainTimeUfCal 340 0x00154
OI_RinseTemp 341 0x00155
OI_RinseTimeMm90 342 0x00156
OI_ServTempHL 343 0x00157
OI_ServTempLL 344 0x00158
OI_SetTimeDiff 345 0x00159
OI_SetTimeIsolAndDiff 346 0x0015A
OI_SetTimeIsol 347 0x0015B
OI_SnArtTimeSet 348 0x0015C
OI_SnVenTimeSet 349 0x0015D
OI_StrokeVolLim 350 0x0015E
OI_TempHL 351 0x0015F
OI_TempLL 352 0x00160
OI_TempEstTempDrop 353 0x00161
OI_TempSet 354 0x00162
OI_TimeNowDayOfMonth 355 0x00163


8 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OI_TimeNowDayOfWeek 356 0x00164


OI_TimeNowHour 357 0x00165
OI_TimeNowMin 358 0x00166
OI_TimeNowMonth 359 0x00167
OI_TimeNowYear 360 0x00168
OI_UFCalRefVol 361 0x00169
OI_UFDFlushDone 362 0x0016A
OI_UfRateSet 363 0x0016B
OI_UfRateSetDiff 364 0x0016C
OI_UfRateSetIsol 365 0x0016D
OI_UfVolSetDiff 366 0x0016E
OI_UfVolSetIsol 367 0x0016F
OI_UfVolSetIsolAndDiff 368 0x00170
OI_UPSMinutes 369 0x00171
OI_VenPrHl 370 0x00172
OI_VenPrLl 371 0x00173
OI_Ch1SetFlow 372 0x00174
OI_DesiredVarFlowMotorPos 373 0x00175
OI_ProfUFIntervals 374 0x00176
OI_ProfUFRemVol 375 0x00177
OI_ProfUFStartVol 376 0x00178
OI_ProfUFStepDelta 377 0x00179
OI_ProfUFStepNo 378 0x0017A
OI_ProfUFStopVol 379 0x0017B
OI_ProfUfStepTime 380 0x0017C
OI_ProfUFVolume 381 0x0017D
OI_ConcStandbyFch 382 0x0017E
OI_ConcStandbyPrim 383 0x0017F
OI_TimeInVarFlowCalDiag 384 0x00180
OI_DiascanAlarmError 385 0x00181
OI_DiascanClearance 386 0x00182
OI_DiascanClearBuffer 387 0x00183
OI_DiascanCondSet 388 0x00184
OI_DiascanDistrVolume 389 0x00185
OI_DiascanKAlarmLimit 390 0x00186
OI_DiascanKMean 391 0x00187
OI_DiascanKt 392 0x00188
OI_DiascanKtForecast 393 0x00189
OI_DiascanKtV 394 0x0018A
OI_DiascanKtVForecast 395 0x0018B
OI_DiascanMeasTime 396 0x0018C
OI_DiascanTarget 397 0x0018D
OI_DiascanTargetAlarm 398 0x0018E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 9 (381)
Program version 1.1.0
C EN 2015-03-26

OI_DiascanTargetTime 399 0x0018F


OI_DiascanUpdateBuffer 400 0x00190
OI_DiascanWatsonAge 401 0x00191
OI_DiascanWatsonHeight 402 0x00192
OI_DiascanWatsonWeight 403 0x00193
OI_DiascanWatsonVolume 404 0x00194
OI_DiascanKAlarmLimitSet 405 0x00195
OI_DiascanTargetSet 406 0x00196
OI_DiascanTargetAlarmSet 407 0x00197
OI_TotalExtraPrimSetVol 408 0x00198
OI_VerifyConcPerformed 409 0x00199
OI_PrimVolAch 410 0x0019A
OI_TreatmentStarted 411 0x0019B
OI_DiascanNoOfReschedules 412 0x0019C
OI_DiascanRescheduledTime 413 0x0019D
OI_BlPumpStopCounterRecCopy 414 0x0019E
OI_DaysSinceLastUFDChange 415 0x0019F
OI_HepPumpStartInPrim 416 0x001A0
OI_FchCondAttnIssued 417 0x001A1
OI_FchCondAttnConfirmed 418 0x001A2
OI_BloodFlADOvr 419 0x001A3
OI_BloodFlSetHlADOvr 420 0x001A4
OI_BloodFlSetLlRaw 421 0x001A5
OI_BlTubeCoeffNom 422 0x001A6
OI_LEDDispTime 423 0x001A7
OI_DiascanSodiumLimitTime 424 0x001A8
OI_DiascanAlarmHandled 425 0x001A9
OI_NbrOf65MinsSinceNewYear 426 0x001AA
OI_TimeNowDaysSince84 427 0x001AB
OI_DisAutoMode 428 0x001AC
OI_SelectedTherapyInfo 429 0x001AD
OI_PrimDetBloodLockOccured 430 0x001AE
OI_Dummy 431 0x001AF
O_Dummy 432 0x001B0
O_DummyTO 433 0x001B1
O_AcKey 434 0x001B2
O_AcModeAck 435 0x001B3
O_AcRq 436 0x001B4
O_ActivateADToP 437 0x001B5
O_ActivateOperatorReset 438 0x001B6
O_ActivateTopProc 439 0x001B7
O_ActivateTopProcess 440 0x001B8
O_AirDetAlBypassEnded 441 0x001B9


10 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_AirDetAlBypass 442 0x001BA


O_AirDetFluidRqAttn 443 0x001BB
O_AirDetInactiveAttn 444 0x001BC
O_AirDetKey 445 0x001BD
O_AirDetActive 446 0x001BE
O_AirDetPrimVolAch 447 0x001BF
O_AllLampsOnRq 448 0x001C0
O_AnyKeyPressed 449 0x001C1
O_ApBank1Active 450 0x001C2
O_ApBank2Active 451 0x001C3
O_ApBank3Active 452 0x001C4
O_ApBank1Load 453 0x001C5
O_ApBank2Load 454 0x001C6
O_ApBank3Load 455 0x001C7
O_ApBank1Sel 456 0x001C8
O_ApBank2Sel 457 0x001C9
O_ApBank3Sel 458 0x001CA
O_ApPrBuildAttn 459 0x001CB
O_ApPrimDetBloodAlarm 460 0x001CC
O_ApPrTestAttn 461 0x001CD
O_ArtPrAutoLimAlarm 462 0x001CE
O_ArtPrGrdOK 463 0x001CF
O_ArtPrKey 464 0x001D0
O_ArtPrTestAck 465 0x001D1
O_ArtPrTestAttnRq 466 0x001D2
O_ArtPumpFreq 467 0x001D3
O_AttnKey 468 0x001D4
O_AttnRq 469 0x001D5
O_AutoHeatActive 470 0x001D6
O_AutoHeatKey 471 0x001D7
O_AutoHeatKeyPulse 472 0x001D8
O_AutoHeatStart 473 0x001D9
O_AutoOffAfterDis 474 0x001DA
O_AutoOffAfterDisTo 475 0x001DB
O_AutoOffAfterDisReqToP 476 0x001DC
O_AutoOnNotPossible 477 0x001DD
O_AutoOnHeatSts 478 0x001DE
O_AutoOnImpAttnRq 479 0x001DF
O_AutoOnRinseSts 480 0x001E0
O_AutoOnDisSts 481 0x001E1
O_AutoPrimRdy 482 0x001E2
O_AutoPrimRq 483 0x001E3
O_AutoRinseActive 484 0x001E4


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 11 (381)
Program version 1.1.0
C EN 2015-03-26

O_AutoRinseKey 485 0x001E5


O_AutoRinseKeyPulse 486 0x001E6
O_AutoRinseStart 487 0x001E7
O_BattActAlarm 488 0x001E8
O_BatteryFailedAttn 489 0x001E9
O_BattFailedAttnRq 490 0x001EA
O_BattlowActAlarm 491 0x001EB
O_BCKey 492 0x001EC
O_BcModeAck 493 0x001ED
O_BcRq 494 0x001EE
O_BiCartAlt 495 0x001EF
O_BiCartEmptyRq 496 0x001F0
O_BicHolderNotInstalledAttn 497 0x001F1
O_BlFlDispAct 498 0x001F2
O_BlLeakCleanAttn 499 0x001F3
O_BlLeakKey 500 0x001F4
O_BlPumpCoverAlarm 501 0x001F5
O_BlPumpCoverOpen 502 0x001F6
O_BlPumpFixed 503 0x001F7
O_BlPumpKey 504 0x001F8
O_BlPumpKnobAct 505 0x001F9
O_BlPumpOvldAttn 506 0x001FA
O_BlPumpOnZero 507 0x001FB
O_BlPumpStartRq 508 0x001FC
O_BlPumpStopClampRq 509 0x001FD
O_BlPumpStopCounter 510 0x001FE
O_BlPumpStopRq 511 0x001FF
O_BlPumpZeroSet 512 0x00200
O_BlTubeChange 513 0x00201
O_BlTubeChangeAttn 514 0x00202
O_BlTubeChangeOk 515 0x00203
O_BPSegCh 516 0x00204
O_BMDiagExit 517 0x00205
O_BPMAlarmAck 518 0x00206
O_BPMComErrorAttn 519 0x00207
O_BPMKey 520 0x00208
O_BPMNoCuffAttn 521 0x00209
O_BPMNoAttn 522 0x0020A
O_BPMNoAlarm 523 0x0020B
O_BPMNoMeasAlarm 524 0x0020C
O_BuzzErrRegToP 525 0x0020D
O_ByvaOpen 526 0x0020E
O_CalCalcError 527 0x0020F


12 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_CalCalcOk 528 0x00210


O_CalEEPromBadCRC 529 0x00211
O_CalEEPromBadId 530 0x00212
O_CalEEPromCheckOk 531 0x00213
O_CalEEPromInitError 532 0x00214
O_CalEEPromInitOk 533 0x00215
O_CalEEPromReadError 534 0x00216
O_CalEEPromReadOk 535 0x00217
O_CalEEPromStoreError 536 0x00218
O_CalEEPromStoreOk 537 0x00219
O_CalEEPromUpdateError 538 0x0021A
O_CalEEPromUpdateOk 539 0x0021B
O_CalInitOk 540 0x0021C
O_CalSampleOk 541 0x0021D
O_CCPrimAttn 542 0x0021E
O_CdsExtAlarm 543 0x0021F
O_CentralChemAttn 544 0x00220
O_CentralChemRq 545 0x00221
O_ChemAck 546 0x00222
O_ChemDisRq 547 0x00223
O_ChemFlagClearRq 548 0x00224
O_ChemFlagSet 549 0x00225
O_ChemFlagNotSet 550 0x00226
O_ChemKey 551 0x00227
O_ChemRdy 552 0x00228
O_ChemStsSel 553 0x00229
O_CitricModeRq 554 0x0022A
O_CleanCartModeRq 555 0x0022B
O_CleanRdy 556 0x0022C
O_ClinAutComSupAlarm 557 0x0022D
O_ComDiagRq 558 0x0022E
O_ConcBank0 559 0x0022F
O_ConcBank1 560 0x00230
O_ConcBank2 561 0x00231
O_ConcRqAttn 562 0x00232
O_CondAlarm 563 0x00233
O_CondAlarmLow 564 0x00234
O_CondAlarmHigh 565 0x00235
O_ConcAlarm 566 0x00236
O_CondAAlarm 567 0x00237
O_CondBAlarm 568 0x00238
O_CondAdj 569 0x00239
O_CondBuzzerRq 570 0x0023A


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 13 (381)
Program version 1.1.0
C EN 2015-03-26

O_CondKey 571 0x0023B


O_ContBlPumpTest 572 0x0023C
O_CountTime 573 0x0023D
O_InitCountTime 574 0x0023E
O_StopCountTime 575 0x0023F
O_DiaConByp 576 0x00240
O_DiaConDia 577 0x00241
O_DiaConNotInBypassAttn 578 0x00242
O_DiaConOne 579 0x00243
O_DiaConRemove1Attn 580 0x00244
O_DiaConRemove2Attn 581 0x00245
O_DiaConReplace1Attn 582 0x00246
O_DiaConReplace2Attn 583 0x00247
O_DiaConAbortAttn 584 0x00248
O_DiaConFinalAttachAttn 585 0x00249
O_DiaConRemoveSELECTAck 586 0x0024A
O_DiaConReplaceSELECTAck 587 0x0024B
O_DiaConAbortAck 588 0x0024C
O_DiaConFinalAttachAck 589 0x0024D
O_DiaConStsRq 590 0x0024E
O_DiagUFTarationRq 591 0x0024F
O_DialyzerFlowReq 592 0x00250
O_DisBank1Active 593 0x00251
O_DisBank2Active 594 0x00252
O_DisBank4Active 595 0x00253
O_DisBank5Active 596 0x00254
O_DisBank6Active 597 0x00255
O_DisBank7Active 598 0x00256
O_DisBank1Load 599 0x00257
O_DisBank2Load 600 0x00258
O_DisBank4Load 601 0x00259
O_DisBank5Load 602 0x0025A
O_DisBank6Load 603 0x0025B
O_DisBank7Load 604 0x0025C
O_DisBank4Sel 605 0x0025D
O_DisBank5Sel 606 0x0025E
O_DisBank6Sel 607 0x0025F
O_DisBank7Sel 608 0x00260
O_DiscBlPumpStop 609 0x00261
O_DiscontinueAck 610 0x00262
O_DiscontinueAckTo 611 0x00263
O_DiscontinueModeRq 612 0x00264
O_DisinfDecalcExit 613 0x00265


14 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_DisinfDecalcTimeSetRq 614 0x00266


O_DisinfInterruptAttn 615 0x00267
O_DisinfRinseRq 616 0x00268
O_DisIsConc 617 0x00269
O_DispApPrBuild 618 0x0026A
O_DispApPrTest 619 0x0026B
O_DispArtPrTest 620 0x0026C
O_DispBatteryFailed 621 0x0026D
O_DispBlLeakClean 622 0x0026E
O_DispBlPumpOvld 623 0x0026F
O_DispBlTubeChange 624 0x00270
O_DispDiaConRemove1 625 0x00271
O_DispDiaConRemove2 626 0x00272
O_DispDiaConRemoveAny 627 0x00273
O_DispDiaConReplace1 628 0x00274
O_DispDiaConReplace2 629 0x00275
O_DispDiaConReplaceAny 630 0x00276
O_DispDiaConAbort 631 0x00277
O_DispDiaConFinalAttach 632 0x00278
O_DispLangLoadFailed 633 0x00279
O_DispLowHeatingTemp 634 0x0027A
O_DispRecover 635 0x0027B
O_DispRemoveCC 636 0x0027C
O_DispTimeToClean 637 0x0027D
O_DispTimeToDecalc 638 0x0027E
O_DispTimeToDisinf 639 0x0027F
O_DispUfParaZero 640 0x00280
O_DispVenPrTest 641 0x00281
O_DispWarningText 642 0x00282
O_DisRinseRdy 643 0x00283
O_DisRinseRq 644 0x00284
O_DivaOpen 645 0x00285
O_DrainDirect 646 0x00286
O_DrainRq 647 0x00287
O_DwellForever 648 0x00288
O_DwellTimeForever 649 0x00289
O_EnterServiceMode 650 0x0028A
O_ErrorsPrdExit 651 0x0028B
O_EvfAcAltSetHoldDelay 652 0x0028C
O_EvfAcAltSetWaitDelay 653 0x0028D
O_EvfAirDetAttnLatch 654 0x0028E
O_EvfAutoPrimMode 655 0x0028F
O_EvfBcAltSetHoldDelay 656 0x00290


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 15 (381)
Program version 1.1.0
C EN 2015-03-26

O_EvfBcAltSetWaitDelay 657 0x00291


O_EvfCondAlarmLatch 658 0x00292
O_EvfFluidByp 659 0x00293
O_EvfTempAlarmLatch 660 0x00294
O_EvfVprLimitDelay 661 0x00295
O_EvfWaitWhenEnterService 662 0x00296
O_EvfWarningText 663 0x00297
O_ExtFchINVATest 664 0x00298
O_ExtFchSel 665 0x00299
O_FchCondAttnRq 666 0x0029A
O_FchFreeze 667 0x0029B
O_Field0Sel 668 0x0029C
O_Field1Sel 669 0x0029D
O_Field2Sel 670 0x0029E
O_Field3Sel 671 0x0029F
O_FlashSETText 672 0x002A0
O_FluidBypassRbRdy 673 0x002A1
O_FluidBypKey 674 0x002A2
O_FluidBypRq 675 0x002A3
O_FluidBypToAttn 676 0x002A4
O_FluidDirectRbRdy 677 0x002A5
O_FluidPrepOffRq 678 0x002A6
O_FluidPrepOnRq 679 0x002A7
O_FluidRdyAttn 680 0x002A8
O_FlushAttn 681 0x002A9
O_FlushAttnDelayTo 682 0x002AA
O_FlushTimeOK 683 0x002AB
O_FlushTubeInactive 684 0x002AC
O_FlushTubeTo 685 0x002AD
O_FMDiagExit 686 0x002AE
O_FmDisMode 687 0x002AF
O_GreenLanePresence 688 0x002B0
O_HeatAck 689 0x002B1
O_HeatCitricModeRq 690 0x002B2
O_HeatCitricRq 691 0x002B3
O_HeatCleanCartModeRq 692 0x002B4
O_HeatDisRq 693 0x002B5
O_HeatKey 694 0x002B6
O_HepPumpFreq 695 0x002B7
O_HepPumpOvldTest 696 0x002B8
O_HepPumpRq 697 0x002B9
O_HepSetToZeroAttn 698 0x002BA
O_HepStopLimAch 699 0x002BB


16 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_HepZero 700 0x002BC


O_HepZeroAcked 701 0x002BD
O_HoldKey 702 0x002BE
O_HoldSeq 703 0x002BF
O_UICancelPressed 704 0x002C0
O_UISkipPressed 705 0x002C1
O_CancelCleared 706 0x002C2
O_CancelPressed 707 0x002C3
O_CalibOther 708 0x002C4
O_HydraulicRestartAttn 709 0x002C5
O_I2CDiagPrdExit 710 0x002C6
O_I2CScanIDError 711 0x002C7
O_I2CEEPromInitError 712 0x002C8
O_I2CScanInitError 713 0x002C9
O_I2CScanNext 714 0x002CA
O_I2CScanReady 715 0x002CB
O_I2CScanUpdateError 716 0x002CC
O_InitInPressReply 717 0x002CD
O_InPressAlarm 718 0x002CE
O_InPressDisinfAlarm 719 0x002CF
O_InTMTDiaConInBypassAlarm 720 0x002D0
O_InTMTDiaConInBypSecAlarm 721 0x002D1
O_InternalSupportMode 722 0x002D2
O_InternalCalib 723 0x002D3
O_UFCalib 724 0x002D4
O_UFReferenceVolumeSet 725 0x002D5
O_UFCalibComplete 726 0x002D6
O_UFCalibInternal 727 0x002D7
O_UFCalibVerify 728 0x002D8
O_InternalUFCal 729 0x002D9
O_IntUFCalRemainTimeRq 730 0x002DA
O_IsolUfZeroTimeAttn 731 0x002DB
O_IsolUfModeKeyPulse 732 0x002DC
O_IsolUfModeRq 733 0x002DD
O_IsolUfResetRq 734 0x002DE
O_IsolUfStartAttn 735 0x002DF
O_IsolUfSelected 736 0x002E0
O_JumpToUFDisplay 737 0x002E1
O_LampsOnCheckRq 738 0x002E2
O_LinTrans 739 0x002E3
O_LogicalOffQueryOk 740 0x002E4
O_LogicalOffQueryNotOk 741 0x002E5
O_LogParDispInh 742 0x002E6


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 17 (381)
Program version 1.1.0
C EN 2015-03-26

O_LowHeatingTempAttn 743 0x002E7


O_MachineModeStartup 744 0x002E8
O_MachineModeFunctionalCheck 745 0x002E9
O_MachineModeBloodLinePrep 746 0x002EA
O_MachineModePreTreatment 747 0x002EB
O_MachineModeTreatment 748 0x002EC
O_MachineModeDiscontinueTreatment 749 0x002ED
O_MachineModePostTreatment 750 0x002EE
O_MachineModeDisinfection 751 0x002EF
O_MachineModeBloodLinePrepDurDisinf 752 0x002F0
O_MachineModeService 753 0x002F1
O_BmBloodLinePrepDurDisinfEnter 754 0x002F2
O_MaintDispRq 755 0x002F3
O_MaintResetRq 756 0x002F4
O_MaintSuppKey0 757 0x002F5
O_MaintSuppKey1 758 0x002F6
O_MoveToNextState 759 0x002F7
O_TimeToClean 760 0x002F8
O_TimeToDecalc 761 0x002F9
O_TimeToDisinf 762 0x002FA
O_ToDisinfDecalcExit 763 0x002FB
O_ToExtFchExit 764 0x002FC
O_ToMM95StoreValueExit 765 0x002FD
O_MaintAccessOk 766 0x002FE
O_MonitorShutDownRq 767 0x002FF
O_NegTmpAttn 768 0x00300
O_NewAccCleanTime 769 0x00301
O_NewAccDecalcCounter 770 0x00302
O_NoAir 771 0x00303
O_NoConc 772 0x00304
O_NoDilutionAttn 773 0x00305
O_NonDiffTimeResetAck 774 0x00306
O_NoServiceParInh 775 0x00307
O_OnOffKey 776 0x00308
O_OptrcActiveRq 777 0x00309
O_OstrcActiveRq 778 0x0030A
O_OvernightDisinf 779 0x0030B
O_PassedTimeResetAck 780 0x0030C
O_PcDefaultPreset 781 0x0030D
O_PcPresetCalc 782 0x0030E
O_PermBfrEraseRq 783 0x0030F
O_PresetActive 784 0x00310
O_PrimDet 785 0x00311


18 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_PrimDetPRq 786 0x00312


O_PrimDetWRq 787 0x00313
O_PrimDetCurrSaveRq 788 0x00314
O_PrimingKey 789 0x00315
O_PrimVolAchAttn 790 0x00316
O_ProfACBCModeChangeRq 791 0x00317
O_ProfAcBcModeRdy 792 0x00318
O_ProfHCO3Loaded 793 0x00319
O_ProfHCO3Standby 794 0x0031A
O_ProfModel1Active 795 0x0031B
O_ProfModel2Active 796 0x0031C
O_ProfModel3Active 797 0x0031D
O_ProfModelSel 798 0x0031E
O_ProfNaAcLoaded 799 0x0031F
O_ProfNaAcStandby 800 0x00320
O_ProfNaLoaded 801 0x00321
O_ProfNaStandby 802 0x00322
O_ProfParamHCO3ChangeRq 803 0x00323
O_ProfParamNaAcChangeRq 804 0x00324
O_ProfParamNaChangeRq 805 0x00325
O_ProfParamUFChangeRq 806 0x00326
O_ProfTmtTimeOk 807 0x00327
O_ProfUFCalc 808 0x00328
O_ProfUFLoaded 809 0x00329
O_ProgVerDispActive 810 0x0032A
O_ProgVerRdy 811 0x0032B
O_ProtInfoRq 812 0x0032C
O_ProtValResetRdy 813 0x0032D
O_PumpDiagRq 814 0x0032E
O_PumpDutyCycleTestRq 815 0x0032F
O_RcAutoPrimOffRq 816 0x00330
O_RcFluidBypOnRq 817 0x00331
O_RcIsolUfModeOffRq 818 0x00332
O_RcIsolUfModeOnRq 819 0x00333
O_RcStartStopOffRq 820 0x00334
O_RcStartStopOnRq 821 0x00335
O_RcVolumeControlRq 822 0x00336
O_RdyForMainReset 823 0x00337
O_RealTimeRdy 824 0x00338
O_RecoverDisableRq 825 0x00339
O_RemoveCCAttn 826 0x0033A
O_ResetAc 827 0x0033B
O_ResetActive 828 0x0033C


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 19 (381)
Program version 1.1.0
C EN 2015-03-26

O_ResetBc 829 0x0033D


O_ResetFluid 830 0x0033E
O_ResetKey 831 0x0033F
O_ResetQueryNotOk 832 0x00340
O_ResetQueryOk 833 0x00341
O_RinseAck 834 0x00342
O_RinseBackInitRq 835 0x00343
O_RinseBackKey 836 0x00344
O_RinseBackRdy 837 0x00345
O_RinseBackRq 838 0x00346
O_RinseKey 839 0x00347
O_RinseKeyAck 840 0x00348
O_RinseRdy 841 0x00349
O_RinseRq 842 0x0034A
O_RinseTempOff 843 0x0034B
O_RunTimeMeter 844 0x0034C
O_RecirculationRq 845 0x0034D
O_RecirculationOnRq 846 0x0034E
O_RecirStopRq 847 0x0034F
O_RCDirectRq 848 0x00350
O_RCTimeCount 849 0x00351
O_RCVolAch 850 0x00352
O_RCAch 851 0x00353
O_ResetAccUFVol 852 0x00354
O_ResetQdRecir 853 0x00355
O_RecirTimeOut 854 0x00356
O_IsolRecir 855 0x00357
O_SaveLoggingSetupRq 856 0x00358
O_Select0 857 0x00359
O_Select1 858 0x0035A
O_Select2 859 0x0035B
O_Select3 860 0x0035C
O_SelectExternal 861 0x0035D
O_SelectInternal 862 0x0035E
O_SelectKey 863 0x0035F
O_SelectKeyDisinf 864 0x00360
O_SelectKeyDisinf2 865 0x00361
O_SelectLmpAttnRq 866 0x00362
O_SelectLmpRq 867 0x00363
O_SelectSeq 868 0x00364
O_SendGetPADRefVal 869 0x00365
O_SendSetPADRefVal 870 0x00366
O_SendGetWADRefVal 871 0x00367


20 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_SendSetWADRefVal 872 0x00368


O_Sense12VoltRq 873 0x00369
O_ServFluidPathStatusRq 874 0x0036A
O_ServiceActive 875 0x0036B
O_ServiceSel 876 0x0036C
O_SetNowTime 877 0x0036D
O_ShutDownQueryNotOk 878 0x0036E
O_ShutDownQueryOk 879 0x0036F
O_ShutDownRdy 880 0x00370
O_SnPrSupervAlarm 881 0x00371
O_StartStopAct 882 0x00372
O_StartStopKey 883 0x00373
O_SavePressed 884 0x00374
O_StartSts 885 0x00375
O_StartTimeACBCRdy 886 0x00376
O_StartUfCalib 887 0x00377
O_StartUFStopBack 888 0x00378
O_StopDuringTmt 889 0x00379
O_StopStsToAttn 890 0x0037A
O_StoreCADRefError 891 0x0037B
O_StoreCADRefOK 892 0x0037C
O_StoreEstTempDropOK 893 0x0037D
O_SystemDiagExit 894 0x0037E
O_TarationRq 895 0x0037F
O_TechErr 896 0x00380
O_TechErrBlPumpEn 897 0x00381
O_TechErrBmStop 898 0x00382
O_TechErrFmStop 899 0x00383
O_TechErrFmFluidStop 900 0x00384
O_TechErrHepStop 901 0x00385
O_TechErrHideRdy 902 0x00386
O_TechErrHideRq 903 0x00387
O_TechErrRecoverRq 904 0x00388
O_TechErrSelectLmpRq 905 0x00389
O_TechErrStop 906 0x0038A
O_TechKey 907 0x0038B
O_TempBuzzerRq 908 0x0038C
O_TimeAdj 909 0x0038D
O_TimeToCleanAttn 910 0x0038E
O_TimeToDecalcAttn 911 0x0038F
O_TimeToDisinfAttn 912 0x00390
O_TmpAutoLimAlarm 913 0x00391
O_TmtActive 914 0x00392


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 21 (381)
Program version 1.1.0
C EN 2015-03-26

O_TmtRdy 915 0x00393


O_TmtRdyRb 916 0x00394
O_TmtRdyRbOff 917 0x00395
O_TmtResetRq 918 0x00396
O_TmtTimePos 919 0x00397
O_TmtTimePosDiff 920 0x00398
O_TmtTimePosIsol 921 0x00399
O_IsolTimeSet 922 0x0039A
O_TmtTimePosTotal 923 0x0039B
O_ToAwaitPumpSelectionExit 924 0x0039C
O_ToBMDiagExit 925 0x0039D
O_ToCalibExit 926 0x0039E
O_ToFMDiagExit 927 0x0039F
O_ToI2CScanExit 928 0x003A0
O_ToSelectModeExit 929 0x003A1
O_ToSelectServiceExit 930 0x003A2
O_ToggleOn 931 0x003A3
O_ToggleOff 932 0x003A4
O_ToTransducerSelectExit 933 0x003A5
O_ToUfDisplayResultExit 934 0x003A6
O_ToUtilityExit 935 0x003A7
O_ToWaitEnterSensExit 936 0x003A8
O_ToWaitErrorSelectExit 937 0x003A9
O_ToWaitI2CSelectExit 938 0x003AA
O_ToWaitSensSampleSelExit 939 0x003AB
O_ToWaitSysDiagSelectExit 940 0x003AC
O_ToWaitSampleExit 941 0x003AD
O_ToWaitUFCalSelectExit 942 0x003AE
O_TransBfrEraseRq 943 0x003AF
O_TransCalibExit 944 0x003B0
O_TransErrReset 945 0x003B1
O_UFCalEEPromErrorActive 946 0x003B2
O_UFCalibExit 947 0x003B3
O_UfCalRemainTimeRq 948 0x003B4
O_UfCalVerRemainTimeRq 949 0x003B5
O_UfCalVerifyReq 950 0x003B6
O_UFCheckEEProm 951 0x003B7
O_UFDflushAttn 952 0x003B8
O_UFDisplayTime 953 0x003B9
O_UFDrainRefVol 954 0x003BA
O_UFDrainRefVolReq 955 0x003BB
O_UFEEPromReadyActive 956 0x003BC
O_UFInitEEProm 957 0x003BD


22 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_UfIntCalStoreEEProm 958 0x003BE


O_UfModeRq 959 0x003BF
O_UfModeStartRq 960 0x003C0
O_UFNegAttn 961 0x003C1
O_UFAchAttn 962 0x003C2
O_UFAchDispAttn 963 0x003C3
O_UfParAdj 964 0x003C4
O_UfParaZeroAttn 965 0x003C5
O_UfParImpSetAttn 966 0x003C6
O_UfParDiffImpSetAttn 967 0x003C7
O_UfParIsolImpSetAttn 968 0x003C8
O_UfRateLowAttn 969 0x003C9
O_UfRateSetDiffZero 970 0x003CA
O_UfRateSetIsolZero 971 0x003CB
O_UFStoreEEProm 972 0x003CC
O_UpdateAddress 973 0x003CD
O_UPSActAttn 974 0x003CE
O_ValuesAreReset 975 0x003CF
O_ValveINVA 976 0x003D0
O_ValveAIVA 977 0x003D1
O_ValveREVA 978 0x003D2
O_ValveFLVA 979 0x003D3
O_ValveDRVA 980 0x003D4
O_ValveRIVA 981 0x003D5
O_ValveZEVA 982 0x003D6
O_ValveDIVA 983 0x003D7
O_ValveDIVAP 984 0x003D8
O_ValveTAVA 985 0x003D9
O_ValveEVVA 986 0x003DA
O_ValveBYVA 987 0x003DB
O_ValveBYVAP 988 0x003DC
O_ValveTest 989 0x003DD
O_VenPrAutoLimAlarm 990 0x003DE
O_VenPrKey 991 0x003DF
O_VenPrTestAck 992 0x003E0
O_VenPrTestAttnRq 993 0x003E1
O_VprLimAdj 994 0x003E2
O_VprLimUpdate 995 0x003E3
O_WaitCalEEPromStoreOk 996 0x003E4
O_WaitForEvaluation 997 0x003E5
O_WaitEEPromStoreError 998 0x003E6
O_WaitEEPromStoreErrorSupport 999 0x003E7
O_WaitForNewDisRq 1000 0x003E8


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 23 (381)
Program version 1.1.0
C EN 2015-03-26

O_WaitUFCalEEPromStoreOk 1001 0x003E9


O_WaitUFEEPromStoreError 1002 0x003EA
O_WaitUFEEPromStoreErrorSupp 1003 0x003EB
O_WaitUFInternalReady 1004 0x003EC
O_WaitUFInternalError 1005 0x003ED
O_WaitUFInternalErrorSupp 1006 0x003EE
O_WarningTextDispRq 1007 0x003EF
O_WPrimDetBlood 1008 0x003F0
O_WrChemBankNr 1009 0x003F1
O_PrsAutoCalcRq 1010 0x003F2
O_PrsDefaultToP 1011 0x003F3
O_ToSelectDefaultType 1012 0x003F4
O_ToPrsModeSelectExit 1013 0x003F5
O_ToSelectPreGroupEntExit 1014 0x003F6
O_PrsChainSelected 1015 0x003F7
O_PrsAdjKeyboard 1016 0x003F8
O_PrsSelectRdy 1017 0x003F9
O_PrsStoreErr 1018 0x003FA
O_PrsCalcErr 1019 0x003FB
O_ApPrsCalcErr 1020 0x003FC
O_ApCalcErr 1021 0x003FD
O_PrsStoreRdy 1022 0x003FE
O_PrsDataRdy 1023 0x003FF
O_PrsGroupTop 1024 0x00400
O_PrsFromPRq 1025 0x00401
O_PrsEndOfGroup 1026 0x00402
O_PrsExtRecCalcErr 1027 0x00403
O_PrsExtRecCalcRdy 1028 0x00404
O_PrsInvalid 1029 0x00405
O_PrsMultipleAdjustInChain 1030 0x00406
O_LowDialysisFlowRq 1031 0x00407
O_LowDialysisFlowAttn 1032 0x00408
O_DispBackFiltration 1033 0x00409
O_AltlangLoadError 1034 0x0040A
O_AltlangLoadErrorAck 1035 0x0040B
O_AltlangLoadErrorAttn 1036 0x0040C
O_DispInletCondHighAttn 1037 0x0040D
O_BattFailAccepted 1038 0x0040E
O_OvernightRinseActive 1039 0x0040F
O_MachineOffDelay 1040 0x00410
O_OvernightRinseDeactivated 1041 0x00411
O_VarFlowRegIneffRq 1042 0x00412
O_VarFlowRegIneffAttn 1043 0x00413


24 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_VarFlowDiagRq 1044 0x00414


O_VarFlowCalibRq 1045 0x00415
O_AbortVFlDiagOrCalRq 1046 0x00416
O_EndVFlCalibRq 1047 0x00417
O_StoreVarFlowMotorCalOK 1048 0x00418
O_StoreVarFlowMotorCalFailure 1049 0x00419
O_ToSelectCalibGeneral 1050 0x0041A
O_VariableFlowCalibReady 1051 0x0041B
O_ProfUFStepProfile 1052 0x0041C
O_ConcStandbyFch 1053 0x0041D
O_ConcStandbyModeTo 1054 0x0041E
O_ConcStandbyPrim 1055 0x0041F
O_ConcStandbyRq 1056 0x00420
O_ConcStandbyMode 1057 0x00421
O_VarFlowDisplayNormal 1058 0x00422
O_VarFlowDisplayCond 1059 0x00423
O_VarFlowDisplayTemp 1060 0x00424
O_VarFlowDispCalDiagTime 1061 0x00425
O_Diascan30Min 1062 0x00426
O_Diascan60Min 1063 0x00427
O_DiascanAlarm 1064 0x00428
O_DiascanCondMeasReq 1065 0x00429
O_DiascanDispFailAttn 1066 0x0042A
O_DiascanDispGeneralAttn 1067 0x0042B
O_DiascanDispKtVAchievedAttn 1068 0x0042C
O_DiascanDispMeasFailAttn 1069 0x0042D
O_DiascanDistrVolMod 1070 0x0042E
O_DiascanEnoughTime 1071 0x0042F
O_DiascanFailAttn 1072 0x00430
O_DiascanGeneralAttn 1073 0x00431
O_DiascanGeneralAttnConf 1074 0x00432
O_DiascanGeneralAttnRq 1075 0x00433
O_DiascanKAlarm 1076 0x00434
O_DiascanKAlarmAttn 1077 0x00435
O_DiascanKtVAchieved 1078 0x00436
O_DiascanKtVAchievedAttn 1079 0x00437
O_DiascanKtVAchievedAttnConf 1080 0x00438
O_DiascanMeasAbort 1081 0x00439
O_DiascanMeasActive 1082 0x0043A
O_DiascanMeasFailed 1083 0x0043B
O_DiascanMeasFailAttn 1084 0x0043C
O_DiascanMeasFailAttnConf 1085 0x0043D
O_DiascanMeasModeOn 1086 0x0043E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 25 (381)
Program version 1.1.0
C EN 2015-03-26

O_DiascanMeasTimer 1087 0x0043F


O_DiascanModeOff 1088 0x00440
O_DiascanModeOn 1089 0x00441
O_DiascanModeTimer 1090 0x00442
O_DiascanNotActiveAttn 1091 0x00443
O_DiascanSingle 1092 0x00444
O_DiascanOkToMeas 1093 0x00445
O_DiascanPreMeasTimer 1094 0x00446
O_DiascanResetRdy 1095 0x00447
O_DiascanResetResult 1096 0x00448
O_DiascanTargetAlarm 1097 0x00449
O_DiascanTargetAlarmAttn 1098 0x0044A
O_DiascanTimeToMeas 1099 0x0044B
O_DiascanWarning 1100 0x0044C
O_HydraulicRestartConfirm 1101 0x0044D
O_HydraulicRestart2Attn 1102 0x0044E
O_DispHydraulicRestart 1103 0x0044F
O_InPressAlarmActive 1104 0x00450
O_WRO300HHeatAttn 1105 0x00451
O_WRO300HPrepAttn 1106 0x00452
O_WRO300HConfirm 1107 0x00453
O_WRO300HInclInHeat 1108 0x00454
O_WRO300HInclInHeatSelected 1109 0x00455
O_DispManStartLFHModeAttn 1110 0x00456
O_DispManStopLFHModeAttn 1111 0x00457
O_ManStartLFHModeAttn 1112 0x00458
O_ManStartLFHModeAck 1113 0x00459
O_ManStopLFHModeAttn 1114 0x0045A
O_ExtDisLFHNotAvailAttn 1115 0x0045B
O_DispExtDisLFHNotAvailAttn 1116 0x0045C
O_DispExtDisinfPrepAttn 1117 0x0045D
O_W300CommFailureAttn 1118 0x0045E
O_DelayINVAAttn 1119 0x0045F
O_DisinfectionKey 1120 0x00460
O_NextLmpOn 1121 0x00461
O_PreviousLmpOn 1122 0x00462
O_IncreaseLmpOn 1123 0x00463
O_DecreaseLmpOn 1124 0x00464
O_MenuMaintDisp 1125 0x00465
O_HoldLmpRq 1126 0x00466
O_PreviousLmpRq 1127 0x00467
O_NextLmpRq 1128 0x00468
O_IncreaseLmpRq 1129 0x00469


26 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_DecreaseLmpRq 1130 0x0046A


O_MenuDisBank1Sel 1131 0x0046B
O_MenuDisBank2Sel 1132 0x0046C
O_MenuDisBank4Sel 1133 0x0046D
O_MenuDisBank6Sel 1134 0x0046E
O_MenuDisBank7Sel 1135 0x0046F
O_MenuDisBank9Sel 1136 0x00470
O_DrainKey 1137 0x00471
O_DrainAck 1138 0x00472
O_DrainAckCmd 1139 0x00473
O_ProfParamRdy 1140 0x00474
O_ShowWelcomeTo 1141 0x00475
O_AcConcMenuActive 1142 0x00476
O_BcConcMenuActive 1143 0x00477
O_AcCursorOk 1144 0x00478
O_BcCursorOk 1145 0x00479
O_EvfDisinfectionKeyDelay 1146 0x0047A
O_DisinfectionKeyPulse 1147 0x0047B
O_ProfUFActivated 1148 0x0047C
O_ProfNaActivated 1149 0x0047D
O_ProfHCO3Activated 1150 0x0047E
O_ProfNaAcActivated 1151 0x0047F
O_ProfModel4Active 1152 0x00480
O_ProfModelActivated 1153 0x00481
O_ChemErrorActive 1154 0x00482
O_ProfUFKeyAck 1155 0x00483
O_ProfNaKeyAck 1156 0x00484
O_ProfNaAcKeyAck 1157 0x00485
O_ProfHco3KeyAck 1158 0x00486
O_ProfUFActive 1159 0x00487
O_ProfNaActive 1160 0x00488
O_ProfNaAcActive 1161 0x00489
O_ProfHco3Active 1162 0x0048A
O_OnOffBtnPresRecently 1163 0x0048B
O_OnOffBtnPresRecentlyTo 1164 0x0048C
O_ConPatAlarm 1165 0x0048D
O_ConPatAlarmConfirmed 1166 0x0048E
O_SettingOpened 1167 0x0048F
O_SettingChanged 1168 0x00490
O_DisableIdleCheck 1169 0x00491
O_TOMenuActive 1170 0x00492
O_IdleTOMenu 1171 0x00493
O_IdleTOMenuTo 1172 0x00494


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 27 (381)
Program version 1.1.0
C EN 2015-03-26

O_IdleSetting 1173 0x00495


O_IdleSettingTo 1174 0x00496
O_AlarmActive 1175 0x00497
O_PrimContSel 1176 0x00498
O_PrimExtraSel 1177 0x00499
O_PrimNewSel 1178 0x0049A
O_ConPatContSel 1179 0x0049B
O_ConPatInstrTOActive 1180 0x0049C
O_RecirInstrTOActive 1181 0x0049D
O_ResetPrimVolRq 1182 0x0049E
O_ResetExtraPrimVolRq 1183 0x0049F
O_RinseBackContSel 1184 0x004A0
O_RinseBackNewSel 1185 0x004A1
O_RinseBackDisconPatSel 1186 0x004A2
O_RinseBackMenuAct 1187 0x004A3
O_RinseBackSelected 1188 0x004A4
O_AskEmptyBicartRq 1189 0x004A5
O_EmptyBiCartAllowed 1190 0x004A6
O_BiCartEmptyManualRq 1191 0x004A7
O_DispEmptyBicQuestion 1192 0x004A8
O_ResetRinseBackVolRq 1193 0x004A9
O_PrimInstrTOActive 1194 0x004AA
O_RinseBackInstrTOActive 1195 0x004AB
O_EmptyBicTOActive 1196 0x004AC
O_BackPressedInTO 1197 0x004AD
O_ApBankSelected 1198 0x004AE
O_ApFinishedTOActive 1199 0x004AF
O_VerifyConcRq 1200 0x004B0
O_DisconPatRq 1201 0x004B1
O_DispDisconPatConf 1202 0x004B2
O_DispDisconPat 1203 0x004B3
O_DisconPatConfTOActive 1204 0x004B4
O_NewAttnRq 1205 0x004B5
O_NewAlarmRq 1206 0x004B6
O_AttnAck 1207 0x004B7
O_AlarmAck 1208 0x004B8
O_InactivateADToPW 1209 0x004B9
O_ThreeSecSelect 1210 0x004BA
O_ProfParamUFModifyRq 1211 0x004BB
O_VerifyConcPerformed 1212 0x004BC
O_LogMem0 1213 0x004BD
O_LogMem1 1214 0x004BE
O_LogMem2 1215 0x004BF


28 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_ConcStandbyManual 1216 0x004C0


O_EmptyBiCartAsked 1217 0x004C1
O_ExtraPrimVolAchAttn 1218 0x004C2
O_AssistedPriming 1219 0x004C3
O_DispExtraPrimVolAttn 1220 0x004C4
O_RinseBackZeroTime 1221 0x004C5
O_TreatmentStarted 1222 0x004C6
O_DisconPatRdy 1223 0x004C7
O_RinseBackExtraSel 1224 0x004C8
O_RinseBackExtraTOActive 1225 0x004C9
O_NoBlDetOverride 1226 0x004CA
O_DoNotStartBloodPump 1227 0x004CB
O_ConnectPatientOff 1228 0x004CC
O_RinseBackStopped 1229 0x004CD
O_HepSetToZeroDispAttn 1230 0x004CE
O_NewPrimingRq 1231 0x004CF
O_RecirSel 1232 0x004D0
O_RecirContSel 1233 0x004D1
O_ConnectPatientSel 1234 0x004D2
O_ConnectPatientRq 1235 0x004D3
O_ConPatStart 1236 0x004D4
O_NewRinseBackRq 1237 0x004D5
O_ResetBlPumpSpeedPrim 1238 0x004D6
O_NaAcAdj 1239 0x004D7
O_NaAdj 1240 0x004D8
O_Hco3Adj 1241 0x004D9
O_DiascanBufferUpdated 1242 0x004DA
O_ProfModelChange 1243 0x004DB
O_BloodPumpSpeedIsReset 1244 0x004DC
O_DispTmtFinishedAttnConf 1245 0x004DD
O_BPMMeasRdyAttn 1246 0x004DE
O_BPMConfirmAttn 1247 0x004DF
O_BPMConfirmAlarm 1248 0x004E0
O_BPMDispMeasRdyAttn 1249 0x004E1
O_CSBM_DelayAlarm 1250 0x004E2
O_CSBM_AlarmTo 1251 0x004E3
O_CSBM_FluidPathOkTo 1252 0x004E4
O_DisinfInProgAttnRq 1253 0x004E5
O_DisinfInProgAttn 1254 0x004E6
O_DisinfDispInProgAttn 1255 0x004E7
O_EvfInactivateADPrim 1256 0x004E8
O_NewCircuitOff 1257 0x004E9
O_NewCircuitRq 1258 0x004EA


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 29 (381)
Program version 1.1.0
C EN 2015-03-26

O_NBCResendTherapy 1259 0x004EB


O_NBCResendTherapy_DO 1260 0x004EC
O_NoBloodLockRq 1261 0x004ED
O_NBCDisconPat 1262 0x004EE
O_NBCRinseBack 1263 0x004EF
O_NBCRinseBackRq 1264 0x004F0
O_NBCPrimingRq 1265 0x004F1
O_DispNBCDisconPat 1266 0x004F2
O_ReleaseBlLockTo 1267 0x004F3
O_HepPumpStopRq 1268 0x004F4
O_NewCircuitStopUfRq 1269 0x004F5
O_DisconPatFlBypRq 1270 0x004F6
O_DispConPatTodo 1271 0x004F7
O_DispRecirTodo 1272 0x004F8
O_RecirDeaIsol 1273 0x004F9
O_ConPatTodoActive 1274 0x004FA
O_RecirTodoActive 1275 0x004FB
O_ConnectPatientOnRq 1276 0x004FC
O_ProfDeactivateAttn 1277 0x004FD
O_ProfDeactivate 1278 0x004FE
O_ProfNaChangedAttn 1279 0x004FF
O_ProfNaChangedAttnWait 1280 0x00500
O_ProfNaChangedConfirm 1281 0x00501
O_ProfHCO3ChangedAttn 1282 0x00502
O_ProfHCO3ChangedAttnWait 1283 0x00503
O_ProfHCO3ChangedConfirm 1284 0x00504
O_ProfUFRChangedAttn 1285 0x00505
O_ProfUFRChangedAttnWait 1286 0x00506
O_ProfUFRChangedConfirm 1287 0x00507
O_DiascanDeactivateAttn 1288 0x00508
O_DiascanDeactivateTo 1289 0x00509
O_DiascanDeactivate 1290 0x0050A
O_ConPatRecTo 1291 0x0050B
O_RecirRecTo 1292 0x0050C
O_AirDetFluidRqAttn2 1293 0x0050D
O_UfParMod 1294 0x0050E
O_IsolUfParMod 1295 0x0050F
O_SettingDelayTo 1296 0x00510
O_DiascanDelayTo 1297 0x00511
O_DiascanRescheduled 1298 0x00512
O_DiascanFluidAdj 1299 0x00513
O_DiascanMeasFailTriggered 1300 0x00514
O_DiascanKAlarmDispAttn 1301 0x00515


30 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_DiascanNotPossibleTo 1302 0x00516


O_DiascanPausedAttn 1303 0x00517
O_DiascanPausedTo 1304 0x00518
O_DiascanDistrVolAdj 1305 0x00519
O_DiascanDistrDelayTo 1306 0x0051A
O_ProfParamUFAdj 1307 0x0051B
O_ProfParamUFDelayTo 1308 0x0051C
O_TmtRdyConf 1309 0x0051D
O_DispCentralChemAttn 1310 0x0051E
O_TimeToChUfdDays 1311 0x0051F
O_TimeToChUfdDaysAttn 1312 0x00520
O_TimeToChUfdDisinf 1313 0x00521
O_TimeToChUfdDisinfAttn 1314 0x00522
O_TimeToChUfdHypo 1315 0x00523
O_TimeToChUfdHypoAttn 1316 0x00524
O_WaitForUFDEvaluation 1317 0x00525
O_ResetUfdReminder 1318 0x00526
O_HepPresetsReq 1319 0x00527
O_HepPumpStartInPrim 1320 0x00528
O_ChemValTstFinished 1321 0x00529
O_IsolUfVolMod 1322 0x0052A
O_UfVolMod 1323 0x0052B
O_ShowingSelIntCalib 1324 0x0052C
O_CCLatchAttn 1325 0x0052D
O_DiascanResultLogged 1326 0x0052E
O_BlPumpStopClampOpen 1327 0x0052F
O_SagsTestAttn 1328 0x00530
O_AutoHeatStartIsSentToP 1329 0x00531
O_DispFchCondAttnRq 1330 0x00532
O_AutoRinseStartIsSentToP 1331 0x00533
O_FchCondAttnConfirmed 1332 0x00534
O_FchBlTestAttnIssued 1333 0x00535
O_FchBlTestAttnRq 1334 0x00536
O_FchBlTestAttnConfirmed 1335 0x00537
O_DispFchBlTestAttnRq 1336 0x00538
O_FchVClampTestAttnIssued 1337 0x00539
O_FchAClampTestAttnIssued 1338 0x0053A
O_FchVClampTestAttnRq 1339 0x0053B
O_FchAClampTestAttnRq 1340 0x0053C
O_FchVClampTestAttnConfirmed 1341 0x0053D
O_FchAClampTestAttnConfirmed 1342 0x0053E
O_DispFchVClampTestAttnRq 1343 0x0053F
O_DispFchAClampTestAttnRq 1344 0x00540


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 31 (381)
Program version 1.1.0
C EN 2015-03-26

O_FchDivaTavaTestAttnIssued 1345 0x00541


O_FchDivaTavaTestAttnRq 1346 0x00542
O_FchDivaTavaTestAttnConfirmed 1347 0x00543
O_DispFchDivaTavaTestAttnRq 1348 0x00544
O_UpdateExtraRecirUFR 1349 0x00545
O_RinseBackOn 1350 0x00546
O_TurnOffBPM 1351 0x00547
O_HepComplAttn 1352 0x00548
O_HepComplDispAttn 1353 0x00549
O_DiascanKAlarmAttnConf 1354 0x0054A
O_DiascanTargetAlarmDispAttn 1355 0x0054B
O_DiascanTargetAlarmAttnConf 1356 0x0054C
O_TreatmentCounted 1357 0x0054D
O_CalcDisinfDecalcFinished 1358 0x0054E
O_BypDurRecAttn 1359 0x0054F
O_RecirRunning 1360 0x00550
O_ExtraRecirRunning 1361 0x00551
O_BypassDuringRecirTo 1362 0x00552
O_DiascanUFTimerTo 1363 0x00553
O_UFSensorCompare 1364 0x00554
O_RecirStopped 1365 0x00555
O_ExtraRecirStopped 1366 0x00556
O_ValveFIVA 1367 0x00557
O_ValveCBVA 1368 0x00558
O_ValveCHVA 1369 0x00559
O_ValveHBVA 1370 0x0055A
O_SrvDiagHbvaOpen 1371 0x0055B
O_StartRelayTest 1372 0x0055C
O_StartZeroPrTest 1373 0x0055D
O_StartPosHighPrTest 1374 0x0055E
O_StartNegHighPrTest 1375 0x0055F
O_PrTestRdy 1376 0x00560
O_PrTestInitRdy 1377 0x00561
O_PrTestCancel 1378 0x00562
O_RecirCont 1379 0x00563
O_ExtraRecirCont 1380 0x00564
O_DisinfPostTreatment 1381 0x00565
O_DisinfButtonFlash 1382 0x00566
O_RecirStarted 1383 0x00567
O_ExtraRecirStarted 1384 0x00568
O_RinseBackLmpFlash 1385 0x00569
O_RinseBackStartMan 1386 0x0056A
O_ProfUfRateLowAttn 1387 0x0056B


32 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_ProfUfRateHighAttn 1388 0x0056C


O_BPSpeedResetRB 1389 0x0056D
O_UfsAttnOnly 1390 0x0056E
O_UfsDevAcked 1391 0x0056F
O_NewBloodCircuitMenuAct 1392 0x00570
O_ExtraRinseBackActive 1393 0x00571
O_PrimBloodDetected 1394 0x00572
O_PrimNoBloodDetected 1395 0x00573
O_DispExtDisinfPrepAttnConf 1396 0x00574
O_ExtDisLFHNotAvailAttnConf 1397 0x00575
O_AltlangLoadErrorAttnConf 1398 0x00576
O_ExtraPrimVolAchAttnConf 1399 0x00577
O_BlLeakCleanAttnConf 1400 0x00578
O_LowHeatingTempAttnConf 1401 0x00579
O_ContBlPumpTestAttnConf 1402 0x0057A
O_DispApPrBuildAttnConf 1403 0x0057B
O_CentralChemAttnConf 1404 0x0057C
O_DiascanFailAttnConf 1405 0x0057D
O_BPMMeasRdyAttnConf 1406 0x0057E
O_BPMAttnConf 1407 0x0057F
O_BPMAlarmConf 1408 0x00580
O_RemoveCCAttnConf 1409 0x00581
O_UFAchAttnConf 1410 0x00582
O_BmFchIsExecuting 1411 0x00583
O_CmFchIsExecuting 1412 0x00584
O_FmFchIsExecuting 1413 0x00585
O_BmFchPrimDetTstIsExecuting 1414 0x00586
O_CloseArterialClamp 1415 0x00587
O_CloseVenousClamp 1416 0x00588
O_FchHepPumpAttnRq 1417 0x00589
O_FchHepPumpAttnIssued 1418 0x0058A
O_FchHepPumpAttnConf 1419 0x0058B
O_HepPumpDisabledAttnRq 1420 0x0058C
O_HepPumpDisabledAttnConf 1421 0x0058D
O_UFNegAttnConf 1422 0x0058E
O_UFNegDispAttnConf 1423 0x0058F
O_DevSoftwareAttnRq 1424 0x00590
O_DevSoftwareAttnDone 1425 0x00591
O_ToCalib 1426 0x00592
O_UtilityStart 1427 0x00593
O_DiagnoseStart 1428 0x00594
O_TransducerSelect 1429 0x00595
O_SensFirstReferenceValue 1430 0x00596


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 33 (381)
Program version 1.1.0
C EN 2015-03-26

O_SensLastReferenceValue 1431 0x00597


O_RefPointSelected 1432 0x00598
O_DisinfStopCalledFor 1433 0x00599
O_DisinfStartCalledFor 1434 0x0059A
O_FchTypeIsBasicFch 1435 0x0059B
O_SendNewTherapyInfo 1436 0x0059C
O_ServiceCalibrationOngoing 1437 0x0059D
O_ServiceDiagnoseOngoing 1438 0x0059E
O_ServiceUtilityOngoing 1439 0x0059F
O_ServiceOperationFinished 1440 0x005A0
O_StartExtFCHInteractive 1441 0x005A1
O_StartExtFCHStandAlone 1442 0x005A2
O_StartExtFCHEnableLeakageTest 1443 0x005A3
O_StartServiceDrain 1444 0x005A4
O_UtilityExtFCHStandAloneOngoing 1445 0x005A5
O_UtilityDrainOngoing 1446 0x005A6
O_UtilitySelectPressed 1447 0x005A7
O_StartI2CScan 1448 0x005A8
O_DiagnoseI2CScanOngoing 1449 0x005A9
O_DiagnoseI2CScanComplete 1450 0x005AA
O_DiagnoseUFTaration 1451 0x005AB
O_DiagnoseUFSensorValues 1452 0x005AC
O_DiagnoseStartLampsTest 1453 0x005AD
O_DiagnoseLampsTestOngoing 1454 0x005AE
O_DiagnoseBloodPumpTestOngoing 1455 0x005AF
O_DiagnoseHeparinPumpTestOngoing 1456 0x005B0
O_DiagnoseVariableFlow 1457 0x005B1
O_DiagnoseFluidPumps 1458 0x005B2
O_DiagnoseValves 1459 0x005B3
O_DiagnoseUpdatePressed 1460 0x005B4
O_DiagnoseUpdateOngoing 1461 0x005B5
O_NightLightMode 1462 0x005B6
O_DisconPatActive 1463 0x005B7
O_BiCartIsEmpty 1464 0x005B8
O_NBCResendTherapy2 1465 0x005B9
O_ProtectiveShutDownQuery 1466 0x005BA
O_UFCalibrateSkipStabilization 1467 0x005BB
O_NaSetReset 1468 0x005BC
O_HCO3SetReset 1469 0x005BD
O_EEIAccDecalcCounterUpdatedAfterTreatment 1470 0x005BE
O_BypButtonNotReadyForDirectFlow 1471 0x005BF
O_HardFault 1472 0x005C0
O_UfModeStartRqHold 1473 0x005C1


34 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

O_WaitAirDetTstRdy 1474 0x005C2


O_AirAlarmAction 1475 0x005C3
OR_Dummy 1476 0x005C4
OW_ChemRinseTime 1477 0x005C5
OW_CitricFlushTime 1478 0x005C6
OW_DwellTime 1479 0x005C7
OW_LogIndex0 1480 0x005C8
OW_LogIndex1 1481 0x005C9
OW_LogIndex2 1482 0x005CA
OW_RemainTimeDiffMsec 1483 0x005CB
OW_RemainTimeIsolMsec 1484 0x005CC
OW_RemainTimeUfCalMsec 1485 0x005CD
OW_TimeInVarFlowCalDiagMSec 1486 0x005CE
OW_DiascanKt 1487 0x005CF
OW_DiascanLastCalcTime 1488 0x005D0
OW_DiascanNonDiffTime 1489 0x005D1
OW_DiascanTimeAfterMeas 1490 0x005D2
PE_BTRC_start 1491 0x005D3
PE_FchBmRestartToC 1492 0x005D4
PE_AirDetResolve 1493 0x005D5
PE_AirDetConfirm 1494 0x005D6
PE_BTRC_end 1495 0x005D7
PE_CO91_start 1496 0x005D8
PE_CO91_end 1497 0x005D9
PE_FTRC_start 1498 0x005DA
PE_EvvaOpen 1499 0x005DB
PE_FchFmPart2RestartToC 1500 0x005DC
PE_FchFmRestartToC 1501 0x005DD
PE_FlSwHeatRelTst 1502 0x005DE
PE_HydraulicRestartToC 1503 0x005DF
PE_StartSlaveModeToC 1504 0x005E0
PE_TempTst 1505 0x005E1
PE_ZEVAActClose 1506 0x005E2
PE_ZEVAActOpen 1507 0x005E3
PE_FTRC_end 1508 0x005E4
PE_OTRC_start 1509 0x005E5
PE_AcModeRq 1510 0x005E6
PE_AirDetTst 1511 0x005E7
PE_AprTst 1512 0x005E8
PE_BcModeRq 1513 0x005E9
PE_BlLeakTst 1514 0x005EA
PE_BlPumpTst 1515 0x005EB
PE_BuzzTst 1516 0x005EC


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 35 (381)
Program version 1.1.0
C EN 2015-03-26

PE_ChemFlagRq 1517 0x005ED


PE_ChemRq 1518 0x005EE
PE_ChemStart 1519 0x005EF
PE_Cond1Tst 1520 0x005F0
PE_Cond2Tst 1521 0x005F1
PE_CurrValvesTst 1522 0x005F2
PE_CurrValvesTst_B 1523 0x005F3
PE_DisIntAttn 1524 0x005F4
PE_DiscontinueRq 1525 0x005F5
PE_DrainRq 1526 0x005F6
PE_DrainStart 1527 0x005F7
PE_FchComRestartToC 1528 0x005F8
PE_HeatRq 1529 0x005F9
PE_HeatStart 1530 0x005FA
PE_InitAc 1531 0x005FB
PE_InitBc 1532 0x005FC
PE_InitInPressRq 1533 0x005FD
PE_LogicalOffQuery 1534 0x005FE
PE_MainReset 1535 0x005FF
PE_MaintRq 1536 0x00600
PE_PresetReset 1537 0x00601
PE_PrimDetTst 1538 0x00602
PE_PrimingChangeRq 1539 0x00603
PE_ProfHCO3Cmd 1540 0x00604
PE_ProfNaAcCmd 1541 0x00605
PE_ProfNaCmd 1542 0x00606
PE_ProfUFCmd 1543 0x00607
PE_PrsDefaultErrToC 1544 0x00608
PE_PrsDefaultRdyToC 1545 0x00609
PE_PrsToCDone 1546 0x0060A
PE_RelayTst 1547 0x0060B
PE_ResetCancel 1548 0x0060C
PE_ResetQuery 1549 0x0060D
PE_RinseRq 1550 0x0060E
PE_RinseStart 1551 0x0060F
PE_SNDNModeChanged 1552 0x00610
PE_ShutDown 1553 0x00611
PE_ShutDownCancel 1554 0x00612
PE_ShutDownQuery 1555 0x00613
PE_SnPrSupervAlarm 1556 0x00614
PE_VClampTst 1557 0x00615
PE_ValvesTst 1558 0x00616
PE_VprTst 1559 0x00617


36 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PE_ZeroReset 1560 0x00618


PE_OTRC_end 1561 0x00619
PE_AClampTst 1562 0x0061A
P_Dummy 1563 0x0061B
P_ArtPrAutoLimAlarm 1564 0x0061C
P_ArtPrGrdOK 1565 0x0061D
P_BadFluid 1566 0x0061E
P_BcRq 1567 0x0061F
P_BlLeakAlBl 1568 0x00620
P_BlLeakAlBypass 1569 0x00621
P_BlLeakAlNoBl 1570 0x00622
P_AlLmpTst 1571 0x00623
P_BloodInColdStart 1572 0x00624
P_BlPumpCoverOpen 1573 0x00625
P_BlPumpStopAlarm 1574 0x00626
P_BlPumpTestStart 1575 0x00627
P_BlPumpTestStop 1576 0x00628
P_BmFchPart1 1577 0x00629
P_BmFchPart2 1578 0x0062A
P_BmHighAlarmMode 1579 0x0062B
P_BmHighStopAlarmMode 1580 0x0062C
P_BmLowAlarmMode 1581 0x0062D
P_BmModeRestored 1582 0x0062E
P_BmNoAlarmMode 1583 0x0062F
P_BubbleTrapHigh 1584 0x00630
P_BubbleTrapLow 1585 0x00631
P_ByvaOpen 1586 0x00632
P_ChemFlagSet 1587 0x00633
P_ChemStart 1588 0x00634
P_CloseArterialClamp 1589 0x00635
P_CloseVenousClamp 1590 0x00636
P_ColdStart 1591 0x00637
P_DisConPat 1592 0x00638
P_CondRegOn 1593 0x00639
P_DiaConByp 1594 0x0063A
P_DiaConDia 1595 0x0063B
P_DiaConErrorInFch 1596 0x0063C
P_DiaConOne 1597 0x0063D
P_DisInt 1598 0x0063E
P_DivaOpen 1599 0x0063F
P_DivaClosedTavaOpen 1600 0x00640
P_DivaClosedTavaClosed 1601 0x00641
P_DrainStart 1602 0x00642


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 37 (381)
Program version 1.1.0
C EN 2015-03-26

P_EvvaOpen 1603 0x00643


P_FlowSwitchFlow 1604 0x00644
P_FluidDirect 1605 0x00645
P_FmDisMode 1606 0x00646
P_FmFchMode 1607 0x00647
P_FmFchPart1 1608 0x00648
P_FmFchPart2 1609 0x00649
P_FmHighAlarmMode 1610 0x0064A
P_FmLowAlarmMode 1611 0x0064B
P_HeaterCEnabledRq 1612 0x0064C
P_HeaterBypRq 1613 0x0064D
P_HeatStart 1614 0x0064E
P_HighTempTest 1615 0x0064F
P_INVAActClose 1616 0x00650
P_LogicalOff 1617 0x00651
P_Lund 1618 0x00652
P_NegTmpAttnRq 1619 0x00653
P_NoAir 1620 0x00654
P_PrimDetBlood 1621 0x00655
P_PrimDetBloodLock 1622 0x00656
P_PrimDetBloodLockOccured 1623 0x00657
P_PrimCurrentStoreOk 1624 0x00658
P_PrimCurrentStoreError 1625 0x00659
P_RegCond 1626 0x0065A
P_RegHeaterStart 1627 0x0065B
P_RegHeaterStop 1628 0x0065C
P_RegHpg 1629 0x0065D
P_RinseStart 1630 0x0065E
P_SnModeStart 1631 0x0065F
P_StopBloodPump 1632 0x00660
P_Taration 1633 0x00661
P_TavaOpen 1634 0x00662
P_TempBadFluid 1635 0x00663
P_TempRegOn 1636 0x00664
P_TmpAutoLimAlarm 1637 0x00665
P_Utv 1638 0x00666
P_VenDataRestored 1639 0x00667
P_VenPrAutoLimAlarm 1640 0x00668
P_WarmStart 1641 0x00669
P_Cond2Tst 1642 0x0066A
P_CVTOpenDIVA 1643 0x0066B
P_CVTOpenBYVA 1644 0x0066C
P_CVTCloseAll 1645 0x0066D


38 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

P_CVTReady 1646 0x0066E


P_CVTOpenDRVA 1647 0x0066F
P_CVTOpenREVA 1648 0x00670
P_CVTOpenZEVA 1649 0x00671
P_CVTOpenRIVA 1650 0x00672
P_CVTOpenINVA 1651 0x00673
P_CVTOpenFLVA 1652 0x00674
P_CVTCloseAll_B 1653 0x00675
P_CVTReady_B 1654 0x00676
P_CVTOpenCHVA 1655 0x00677
P_DRVAOpen 1656 0x00678
P_CondLoBadFluid 1657 0x00679
P_AirAlarm 1658 0x0067A
P_BackFiltrationAttnRq 1659 0x0067B
P_ProfNaAcLoaded 1660 0x0067C
P_ProfNaLoaded 1661 0x0067D
P_ProfHCO3Loaded 1662 0x0067E
P_CurrValvesStsConf 1663 0x0067F
P_RIVAOpen 1664 0x00680
P_CBVAOpen 1665 0x00681
P_FluidPathLeakTst 1666 0x00682
P_StickLatchPosSup 1667 0x00683
P_ChvaOpen 1668 0x00684
P_DiaCon1Status 1669 0x00685
P_DiaCon2Status 1670 0x00686
P_InletCondDataExists 1671 0x00687
P_InletCondSupInst 1672 0x00688
P_DiaConRemove1Attn 1673 0x00689
P_DiaConRemove2Attn 1674 0x0068A
P_DiaConReplace1Attn 1675 0x0068B
P_DiaConReplace2Attn 1676 0x0068C
P_DiaConAbortAttn 1677 0x0068D
P_DiaConFinalAttachAttn 1678 0x0068E
P_CFchComTstStartRq 1679 0x0068F
P_CFchComTstAbortRq 1680 0x00690
P_DiascanInstalled 1681 0x00691
P_DiascanCondMeas 1682 0x00692
P_DiascanRestored 1683 0x00693
P_DiascanReceiveData 1684 0x00694
P_DiascanDoseError 1685 0x00695
P_FluidPathLeakTstDisable 1686 0x00696
P_ConcBankRecv 1687 0x00697
P_ArtClampAvail 1688 0x00698


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 39 (381)
Program version 1.1.0
C EN 2015-03-26

P_DoCValvesCurrTest 1689 0x00699


P_SNDNModeChanged 1690 0x0069A
P_VenLowAlarm 1691 0x0069B
P_LowTempTestActive 1692 0x0069C
P_InletTempHighAttnRq 1693 0x0069D
P_FCHTempTestsRdy 1694 0x0069E
P_HepPresetsAvail 1695 0x0069F
P_HepBolusVolumeAvail 1696 0x006A0
P_HepFlowRateAvail 1697 0x006A1
P_HeatInitiate 1698 0x006A2
P_PdAlarm 1699 0x006A3
P_SagsTestActive 1700 0x006A4
P_SagsTestHPGCtrlHigh 1701 0x006A5
P_RinseInitiate 1702 0x006A6
P_ChemInitiate 1703 0x006A7
P_DrainInitiate 1704 0x006A8
P_DivaTavaTestRdy 1705 0x006A9
P_BlPumpTstRdy 1706 0x006AA
P_UfsAlarm 1707 0x006AB
P_UfsAlarmOverride 1708 0x006AC
P_UFSFchStart 1709 0x006AD
P_UFSFchReady 1710 0x006AE
P_UFSEvalFlowInc 1711 0x006AF
P_UFSTmtRetarRq 1712 0x006B0
P_UFSStartMainFlowMeas 1713 0x006B1
P_TempTestCoolDown 1714 0x006B2
P_HighTempTestDone 1715 0x006B3
P_VprHiAl 1716 0x006B4
P_VprLoAl 1717 0x006B5
P_AirDetPAl 1718 0x006B6
P_AirDetWAl 1719 0x006B7
P_AprLoAl 1720 0x006B8
P_AprHiAl 1721 0x006B9
P_TmpLoAl 1722 0x006BA
P_TmpHiAl 1723 0x006BB
P_CondAl 1724 0x006BC
P_TempHiAl 1725 0x006BD
P_TempLoAl 1726 0x006BE
P_UfsAl 1727 0x006BF
P_FCH_TEST_BLPUMP_STARTED 1728 0x006C0
P_Fch_VCLAMP_Passed 1729 0x006C1
P_Fch_ACLAMP_Passed 1730 0x006C2
P_Fch_PRIMDET_Passed 1731 0x006C3


40 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

P_FCH_TEST_AIRDET_STARTED 1732 0x006C4


P_AirDetHigh 1733 0x006C5
P_AirDetLow 1734 0x006C6
P_AirDetTstRdy 1735 0x006C7
P_AttentionConfirmed 1736 0x006C8
P_FrontIoCVersionOK 1737 0x006C9
P_PowerIoCVersionOK 1738 0x006CA
P_GeneralIoCVersionOK 1739 0x006CB
P_UIVersionOK 1740 0x006CC
P_ProtectiveVersionOK 1741 0x006CD
P_ProtectiveVersionCheckOK 1742 0x006CE
P_UIVersionCheckOK 1743 0x006CF
P_RinseCmdOneTimeTrue 1744 0x006D0
P_DrainCmdOneTimeTrue 1745 0x006D1
P_HepPumpTestStart 1746 0x006D2
P_FCH_TEST_HEP_PASSED 1747 0x006D3
P_HeatCmdOneTimeTrue 1748 0x006D4
P_ChemCmdOneTimeTrue 1749 0x006D5
P_IncorrAcidicConc 1750 0x006D6
P_IncorrBicarbonateConc 1751 0x006D7
P_IncorrComposition 1752 0x006D8
P_AirDetActive 1753 0x006D9
P_CHVAOpenRq 1754 0x006DA
P_AutoOneTimeTrue 1755 0x006DB
P_AutoOneTimeTrue2 1756 0x006DC
P_RequestPlay 1757 0x006DD
P_BPMHighSystolicAlarm 1758 0x006DE
P_BPMHighDiastolicAlarm 1759 0x006DF
P_BPMHighMAPAlarm 1760 0x006E0
P_BPMHighPulseAlarm 1761 0x006E1
P_BPMLowSystolicAlarm 1762 0x006E2
P_BPMLowDiastolicAlarm 1763 0x006E3
P_BPMLowMAPAlarm 1764 0x006E4
P_BPMLowPulseAlarm 1765 0x006E5
P_SnPrSupervAlarm 1766 0x006E6
P_BuzzerAttention 1767 0x006E7
P_UFVDeviation 1768 0x006E8
PB_BattFatalErrCnt 1769 0x006E9
PB_BattWeakErrCnt 1770 0x006EA
PB_PProgVer0 1771 0x006EB
PB_PProgVer5 1772 0x006EC
PB_PProgVer6 1773 0x006ED
PB_PProgVer7 1774 0x006EE


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 41 (381)
Program version 1.1.0
C EN 2015-03-26

PB_PCProgVer0 1775 0x006EF


PB_PWProgVer0 1776 0x006F0
PB_WProgVer0 1777 0x006F1
PB_WProgVer5 1778 0x006F2
PB_WProgVer6 1779 0x006F3
PB_WProgVer7 1780 0x006F4
PB_WCProgVer0 1781 0x006F5
PB_WPProgVer0 1782 0x006F6
PI_Dummy 1783 0x006F7
PI_AirDetAirPErrCnt 1784 0x006F8
PI_AirDetAirCErrCnt 1785 0x006F9
PI_AirDetNoAirErrCnt 1786 0x006FA
PI_AprPrCoeff 1787 0x006FB
PI_AprPrOffset 1788 0x006FC
PI_ArtPrErrCnt 1789 0x006FD
PI_BlLeakCoeff 1790 0x006FE
PI_BlLeakMeas 1791 0x006FF
PI_BlLeakOffLevelErrCnt 1792 0x00700
PI_BlLeakOffset 1793 0x00701
PI_BlLeakOnLevelErrCnt 1794 0x00702
PI_BlLeakOffStableErrCnt 1795 0x00703
PI_BlLeakOnStableErrCnt 1796 0x00704
PI_BlPumpEdge1ErrCnt 1797 0x00705
PI_BlPumpEdge2ErrCnt 1798 0x00706
PI_BlPumpPStopErrCnt 1799 0x00707
PI_BlPumpRotate1ErrCnt 1800 0x00708
PI_BlPumpRotate2ErrCnt 1801 0x00709
PI_BlPumpRPMSet 1802 0x0070A
PI_BlPumpTach1ErrCnt 1803 0x0070B
PI_BlPumpTach2ErrCnt 1804 0x0070C
PI_BlPumpWStopErrCnt 1805 0x0070D
PI_Bm1ErrCnt 1806 0x0070E
PI_Bm2ErrCnt 1807 0x0070F
PI_BmRestartCnt 1808 0x00710
PI_BuzzPOnErrCnt 1809 0x00711
PI_BuzzPOffErrCnt 1810 0x00712
PI_BuzzSilentErrCnt 1811 0x00713
PI_BuzzUIOnErrCnt 1812 0x00714
PI_BuzzUIOffErrCnt 1813 0x00715
PI_ComErrCnt 1814 0x00716
PI_ComRestartCnt 1815 0x00717
PI_CondHighDiffErrCnt 1816 0x00718
PI_CondHighLevelErrCnt 1817 0x00719


42 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PI_CondHighStableErrCnt 1818 0x0071A


PI_CondLowDiffErrCnt 1819 0x0071B
PI_CondLowLevelErrCnt 1820 0x0071C
PI_CondLowStableErrCnt 1821 0x0071D
PI_CondNoAchHighErrCnt 1822 0x0071E
PI_CondNoAchLowErrCnt 1823 0x0071F
PI_CondP 1824 0x00720
PI_CondPCoeff 1825 0x00721
PI_CondPOffset 1826 0x00722
PI_CondSet 1827 0x00723
PI_CondTempP 1828 0x00724
PI_CondTempPCoeff 1829 0x00725
PI_CondTempPHeatCoeff 1830 0x00726
PI_CondTempPHeatOffset 1831 0x00727
PI_CondTempPOffset 1832 0x00728
PI_DiaPrCoeff 1833 0x00729
PI_DiaPresMeas 1834 0x0072A
PI_DiaPrMeas 1835 0x0072B
PI_DiaPrOffset 1836 0x0072C
PI_FlowOffErrCnt 1837 0x0072D
PI_FlowOn1ErrCnt 1838 0x0072E
PI_Fm1ErrCnt 1839 0x0072F
PI_Fm2ErrCnt 1840 0x00730
PI_FmRestartCnt 1841 0x00731
PI_HeaterOff1ErrCnt 1842 0x00732
PI_HeaterOff2ErrCnt 1843 0x00733
PI_HeaterOff3ErrCnt 1844 0x00734
PI_HeaterOff4ErrCnt 1845 0x00735
PI_HeaterOn1ErrCnt 1846 0x00736
PI_HeaterOn2ErrCnt 1847 0x00737
PI_HeaterOn3ErrCnt 1848 0x00738
PI_HeaterOn4ErrCnt 1849 0x00739
PI_HeaterOn5ErrCnt 1850 0x0073A
PI_HepPumpTacho 1851 0x0073B
PI_HighVprTestCount 1852 0x0073C
PI_HpgPrSet 1853 0x0073D
PI_PrimDetBloodPErrCnt 1854 0x0073E
PI_PrimDetBloodCErrCnt 1855 0x0073F
PI_PrimDetCalibCurrErrCnt 1856 0x00740
PI_PrimDetCurrErrCnt 1857 0x00741
PI_PrimDetEmptyP1ErrCnt 1858 0x00742
PI_PrimDetEmptyP2ErrCnt 1859 0x00743
PI_PrimDetEmptyC1ErrCnt 1860 0x00744


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 43 (381)
Program version 1.1.0
C EN 2015-03-26

PI_PrimDetEmptyC2ErrCnt 1861 0x00745


PI_PrimDetRecVolt 1862 0x00746
PI_RegHeaterSet 1863 0x00747
PI_RelayOff1ErrCnt 1864 0x00748
PI_RelayOff2ErrCnt 1865 0x00749
PI_RelayOn1ErrCnt 1866 0x0074A
PI_RelayOn2ErrCnt 1867 0x0074B
PI_RelayOn3ErrCnt 1868 0x0074C
PI_TempHighDiffErrCnt 1869 0x0074D
PI_TempHighLevelErrCnt 1870 0x0074E
PI_TempHighStableErrCnt 1871 0x0074F
PI_TempLowDiffErrCnt 1872 0x00750
PI_TempNoAchHigh1ErrCnt 1873 0x00751
PI_TempNoAchHigh2ErrCnt 1874 0x00752
PI_TempMeas 1875 0x00753
PI_TempDelta 1876 0x00754
PI_ValveBubLevelErrCnt 1877 0x00755
PI_ValveFillupErrCnt 1878 0x00756
PI_ValveHighLeakErrCnt 1879 0x00757
PI_ValveHighLevel1ErrCnt 1880 0x00758
PI_ValveHighLevel2ErrCnt 1881 0x00759
PI_ValveHighStableErrCnt 1882 0x0075A
PI_ValveLowLeakErrCnt 1883 0x0075B
PI_ValveLowLevel1ErrCnt 1884 0x0075C
PI_ValveLowLevel2ErrCnt 1885 0x0075D
PI_ValveLowStableErrCnt 1886 0x0075E
PI_ValveNoAchHigh1ErrCnt 1887 0x0075F
PI_ValveNoAchHigh2ErrCnt 1888 0x00760
PI_ValveNoAchLow1ErrCnt 1889 0x00761
PI_ValveNoAchLow2ErrCnt 1890 0x00762
PI_VenPrHl 1891 0x00763
PI_VenPrLl 1892 0x00764
PI_VClampOpen1ErrCnt 1893 0x00765
PI_VClampOpen2ErrCnt 1894 0x00766
PI_VClampOpen3ErrCnt 1895 0x00767
PI_VClampPCloseErrCnt 1896 0x00768
PI_VClampCCloseErrCnt 1897 0x00769
PI_VenPr 1898 0x0076A
PI_VenPrCoeff 1899 0x0076B
PI_VenPrOffset 1900 0x0076C
PI_VenPrErrCnt 1901 0x0076D
PI_AirDetTstCnt 1902 0x0076E
PI_AprTstCnt 1903 0x0076F


44 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PI_BlLeakTstCnt 1904 0x00770


PI_BlPumpTstCnt 1905 0x00771
PI_BuzzTstCnt 1906 0x00772
PI_Cond1TstCnt 1907 0x00773
PI_Cond2TstCnt 1908 0x00774
PI_FlSwHeatRelTstCnt 1909 0x00775
PI_PrimDetTstCnt 1910 0x00776
PI_RelayTstCnt 1911 0x00777
PI_TempTstCnt 1912 0x00778
PI_ValvesTstCnt 1913 0x00779
PI_VClampTstCnt 1914 0x0077A
PI_VprTstCnt 1915 0x0077B
PI_CurrVaConfirmErrCnt 1916 0x0077C
PI_CurrVaCloseErrCnt 1917 0x0077D
PI_CurrVaOpenDIVACPErrCnt 1918 0x0077E
PI_CurrVaOpenDIVACErrCnt 1919 0x0077F
PI_CurrVaOpenDIVAPErrCnt 1920 0x00780
PI_CurrVaOpenBYVACPErrCnt 1921 0x00781
PI_CurrVaOpenBYVACErrCnt 1922 0x00782
PI_CurrVaOpenBYVAPErrCnt 1923 0x00783
PI_CurrVaOpenTAVAErrCnt 1924 0x00784
PI_CurrVaOpenEVVAErrCnt 1925 0x00785
PI_CurrVaOpenErrErrCnt 1926 0x00786
PI_CurrValvesTstCnt 1927 0x00787
PI_CondCHighStableErrCnt 1928 0x00788
PI_CondTimeOutErrCnt 1929 0x00789
PI_CurrValvesTstCnt_B 1930 0x0078A
PI_BPMCurrentPressure 1931 0x0078B
PI_ProfNaAcStart 1932 0x0078C
PI_ProfNaAcStop 1933 0x0078D
PI_ProfNaStart 1934 0x0078E
PI_ProfNaStop 1935 0x0078F
PI_ProfNaStartOld 1936 0x00790
PI_ProfNaStopOld 1937 0x00791
PI_ProfHCO3Start 1938 0x00792
PI_ProfHCO3Stop 1939 0x00793
PI_ProfHCO3StartOld 1940 0x00794
PI_ProfHCO3StopOld 1941 0x00795
PI_ValveNoAchChem1ErrCnt 1942 0x00796
PI_ValveHpgLevelChem1ErrCnt 1943 0x00797
PI_ValveHpgLevelChem2ErrCnt 1944 0x00798
PI_ValveNegLeakChem1ErrCnt 1945 0x00799
PI_ValveNegLeakChem2ErrCnt 1946 0x0079A


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 45 (381)
Program version 1.1.0
C EN 2015-03-26

PI_ValvePosLeakChem1ErrCnt 1947 0x0079B


PI_ValvePosLeakChem2ErrCnt 1948 0x0079C
PI_CondPLog 1949 0x0079D
PI_InletCondValue 1950 0x0079E
PI_DiascanCondMeas 1951 0x0079F
PI_DiascanPhase 1952 0x007A0
PI_DiascanClearance 1953 0x007A1
PI_DiascanError 1954 0x007A2
PI_DiascanClearanceAvail 1955 0x007A3
PI_DiascanNextStateTime 1956 0x007A4
PI_CurrVaOpenCHVAErrCnt 1957 0x007A5
PI_HepSyringeDiam 1958 0x007A6
PI_HepBolusVolume 1959 0x007A7
PI_HepFlowRate 1960 0x007A8
PI_HepBolusVol 1961 0x007A9
PI_HepTotMeasBolusVolAcc 1962 0x007AA
PI_HepMeasVolAcc 1963 0x007AB
PI_ArtPrProt 1964 0x007AC
PI_UFSTarFailedErrCnt 1965 0x007AD
PI_UFSMainFailedErrCnt 1966 0x007AE
PI_UFSEvalFailedErrCnt 1967 0x007AF
PI_UFSFchTimeOutErrCnt 1968 0x007B0
PI_AClampTstCnt 1969 0x007B1
PI_ProtectiveStartUpFlagStatus 1970 0x007B2
PI_SystemStartUpFlagStatus 1971 0x007B3
PI_UIStartUpFlagStatus 1972 0x007B4
PI_AClampOpen1ErrCnt 1973 0x007B5
PI_AClampOpen2ErrCnt 1974 0x007B6
PI_AClampOpen3ErrCnt 1975 0x007B7
PI_AClampPCloseErrCnt 1976 0x007B8
PI_AClampCCloseErrCnt 1977 0x007B9
PI_Hep_P_Req_freq 1978 0x007BA
PI_HepPumpErrorCnt 1979 0x007BB
PI_ProfilingDummy 1980 0x007BC
PI_CHVACurrErrCnt 1981 0x007BD
PI_ConcPumpDirBackwardErrCnt 1982 0x007BE
PI_ConcPumpDirForwardCErrCnt 1983 0x007BF
PI_ConcPumpDirBackward2ErrCnt 1984 0x007C0
PI_ConcPumpDirForwardPErrCnt 1985 0x007C1
PI_ConcPumpDirForwardCPErrCnt 1986 0x007C2
PI_ConcPumpDirBackward3ErrCnt 1987 0x007C3
PI_AlLmpTstCountDownTimer 1988 0x007C4
PI_ValveLimitedFchBubLevelErrCnt 1989 0x007C5


46 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PI_ValveLimitedFchNoAchLowErrCnt 1990 0x007C6


PI_ValveLimitedFchNoAchHighErrCnt 1991 0x007C7
PI_ValveLimitedFchLeakErrCnt 1992 0x007C8
PW_HepPumpTachoDiagSum 1993 0x007C9
PW_INVAOvr_COMMON 1994 0x007CA
PW_INVAOvr_BLOOD 1995 0x007CB
PW_INVAOvr_FLUID 1996 0x007CC
PW_INVAOvr_COND 1997 0x007CD
PW_INVAOvr_TEMP 1998 0x007CE
PR_Dummy 1999 0x007CF
W_CloseVenousClamp 2000 0x007D0
W_NoAir 2001 0x007D1
W_PrimDetBlood 2002 0x007D2
W_StopBloodPump 2003 0x007D3
W_CloseArterialClamp 2004 0x007D4
FM_MAINSTATEFCN 2005 0x007D5
FF_Dummy 2006 0x007D6
FF_AdState 2007 0x007D7
FF_BiCartValve 2008 0x007D8
FF_BloodLeakTst 2009 0x007D9
FF_ComDiag 2010 0x007DA
FF_CondReg 2011 0x007DB
FF_DiaPr 2012 0x007DC
FF_DirectValve 2013 0x007DD
FF_DisHist 2014 0x007DE
FF_DisinfCtrl 2015 0x007DF
FF_DisinfectValves 2016 0x007E0
FF_DisrinseValves 2017 0x007E1
FF_FlMeasFcn 2018 0x007E2
FF_FlowPumpReg 2019 0x007E3
FF_FluidComp 2020 0x007E4
FF_HeaterEnable 2021 0x007E5
FF_HeaterReg 2022 0x007E6
FF_HPG 2023 0x007E7
FF_InValve 2024 0x007E8
FF_NegUFCheck 2025 0x007E9
FF_MainStateFcn 2026 0x007EA
FF_MM90AdState 2027 0x007EB
FF_OverheatSup 2028 0x007EC
FF_RemDisTimeFcn 2029 0x007ED
FF_SucPumpReg 2030 0x007EE
FF_TarationValves 2031 0x007EF
FF_UFBoxMode 2032 0x007F0


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 47 (381)
Program version 1.1.0
C EN 2015-03-26

FF_UFGain 2033 0x007F1


FF_UFLoggingFcn 2034 0x007F2
FF_UFRateFiltFcn 2035 0x007F3
FF_UFReg 2036 0x007F4
FF_UFVolAcc 2037 0x007F5
FF_ValveCheckFcn 2038 0x007F6
FF_VarFlowMotor 2039 0x007F7
FF_DiascanCtrl 2040 0x007F8
FF_WRO300H 2041 0x007F9
FF_RemDisTimeFcn2 2042 0x007FA
FF_DispRemDisTimeFcn 2043 0x007FB
FF_ChemicalValves 2044 0x007FC
FD_Dummy 2045 0x007FD
FB_Dummy 2046 0x007FE
FB_TarOk 2047 0x007FF
FB_UFLogErrorCode 2048 0x00800
FB_Callfactor 2049 0x00801
FI_Dummy 2050 0x00802
FI_AdRefCoeff 2051 0x00803
FI_AdRefCoeffDiv3 2052 0x00804
FI_Ch1Rate 2053 0x00805
FI_Ch1RateAvg 2054 0x00806
FI_Ch1RateFilt 2055 0x00807
FI_Ch2Rate 2056 0x00808
FI_CondARate 2057 0x00809
FI_CondRate 2058 0x0080A
FI_CWPIntegHeatTemp 2059 0x0080B
FI_DegassPr 2060 0x0080C
FI_DegassPrCoeff 2061 0x0080D
FI_DegassPrOffset 2062 0x0080E
FI_DegassPrTransTestCnt 2063 0x0080F
FI_DegassPrUnstableCnt 2064 0x00810
FI_DiaFlow 2065 0x00811
FI_EstHeatTime 2066 0x00812
FI_EstRinseTime 2067 0x00813
FI_FlowPumpCurrent 2068 0x00814
FI_FlowPumpCycle 2069 0x00815
FI_HeaterCycle 2070 0x00816
FI_HeaterCycleNoLim 2071 0x00817
FI_HeaterEfficiency 2072 0x00818
FI_HeatSucPumpPid 2073 0x00819
FI_NegUFTime 2074 0x0081A
FI_OldHPGPr 2075 0x0081B


48 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

FI_HeatFlowPumpMaxPid 2076 0x0081C


FI_HeatOldSucPumpPid 2077 0x0081D
FI_HeatSucPumpMaxPid 2078 0x0081E
FI_HPGPr 2079 0x0081F
FI_HPGPrCoeff 2080 0x00820
FI_HPGPrOffset 2081 0x00821
FI_InvCondBValue 2082 0x00822
FI_InvTempValue 2083 0x00823
FI_MainFlFiltConst 2084 0x00824
FI_MM90CondA 2085 0x00825
FI_MM90CondAAvg 2086 0x00826
FI_MM90CondACoeff 2087 0x00827
FI_MM90CondAOffset 2088 0x00828
FI_MM90CondB 2089 0x00829
FI_MM90CondBAvg 2090 0x0082A
FI_MM90CondBCoeff 2091 0x0082B
FI_MM90CondBOffset 2092 0x0082C
FI_MM90CondTempA 2093 0x0082D
FI_MM90CondTempACoeff 2094 0x0082E
FI_MM90CondTempAHeatCoeff 2095 0x0082F
FI_MM90CondTempAHeatOffset 2096 0x00830
FI_MM90CondTempAOffset 2097 0x00831
FI_MM90CondTempB 2098 0x00832
FI_MM90CondTempBCoeff 2099 0x00833
FI_MM90CondTempBDisp 2100 0x00834
FI_MM90CondTempBHeatCoeff 2101 0x00835
FI_MM90CondTempBHeatOffset 2102 0x00836
FI_MM90CondTempBOffset 2103 0x00837
FI_CondPAvg 2104 0x00838
FI_OldTempMeas 2105 0x00839
FI_PressRegAdj 2106 0x0083A
FI_PrevUFVol 2107 0x0083B
FI_RegTemp 2108 0x0083C
FI_RegTempCoeff 2109 0x0083D
FI_RegTempHeatCoeff 2110 0x0083E
FI_RegTempHeatOffset 2111 0x0083F
FI_RegTempOffset 2112 0x00840
FI_RelPumpSpeed 2113 0x00841
FI_RelPumpASpeed 2114 0x00842
FI_RelPumpBSpeed 2115 0x00843
FI_RinseOldPid 2116 0x00844
FI_RinseRefTemp 2117 0x00845
FI_SucPumpCurrent 2118 0x00846


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 49 (381)
Program version 1.1.0
C EN 2015-03-26

FI_SucPumpCycle 2119 0x00847


FI_TarTimer 2120 0x00848
FI_UFDFlushTime 2121 0x00849
FI_UFCalHLStatPr 2122 0x0084A
FI_UFCalIsolVol 2123 0x0084B
FI_UFCalIsolVolDiff 2124 0x0084C
FI_UFCalLinDev 2125 0x0084D
FI_UFCalLLStatPr 2126 0x0084E
FI_UFCalVol 2127 0x0084F
FI_UFCalVolDiff 2128 0x00850
FI_UFRate 2129 0x00851
FI_UFRateCtrl 2130 0x00852
FI_UFRateDisp 2131 0x00853
FI_UFRateFiltConst 2132 0x00854
FI_UFRateSNFiltConst 2133 0x00855
FI_UFRateRegSet 2134 0x00856
FI_UFRateSet 2135 0x00857
FI_UFVol 2136 0x00858
FI_UFVolRecir 2137 0x00859
FI_TotalUFVolRecir 2138 0x0085A
FI_AccUFVolRecir 2139 0x0085B
FI_DispUFVolRecir 2140 0x0085C
FI_UfVolDiff 2141 0x0085D
FI_UfVolIsol 2142 0x0085E
FI_UFVolRateHL 2143 0x0085F
FI_UFVolRateLL 2144 0x00860
FI_UFVolSet 2145 0x00861
FI_ValvesCurrent 2146 0x00862
FI_VFCRefCoeff 2147 0x00863
FI_InvCondAValue 2148 0x00864
FI_FiltCondA 2149 0x00865
FI_FiltCondB 2150 0x00866
FI_FixTarAtGreenLineCnt 2151 0x00867
FI_FixTarAtTreatStartCnt 2152 0x00868
FI_InletCondSupInst 2153 0x00869
FI_InletCondValue 2154 0x0086A
FI_InletCondDataExists 2155 0x0086B
FI_CVlvTstOpenedDIVA 2156 0x0086C
FI_CVlvTstOpenedBYVA 2157 0x0086D
FI_CVlvTstClosedAll 2158 0x0086E
FI_CVlvTstOpenedDRVA 2159 0x0086F
FI_CVlvTstOpenedREVA 2160 0x00870
FI_CVlvTstOpenedZEVA 2161 0x00871


50 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

FI_CVlvTstOpenedRIVA 2162 0x00872


FI_CVlvTstOpenedINVA 2163 0x00873
FI_CVlvTstOpenedFLVA 2164 0x00874
FI_CVlvTstOpenedCHVA 2165 0x00875
FI_CVlvTstClosedAll_B 2166 0x00876
FI_CVlvTstAckedRdy 2167 0x00877
FI_CVlvTstAckedRdy_B 2168 0x00878
FI_Ch1RateAvgAvg 2169 0x00879
FI_ClosedDec 2170 0x0087A
FI_OpenInc 2171 0x0087B
FI_FchStepsDec 2172 0x0087C
FI_FchStepsInc 2173 0x0087D
FI_OldFlow 2174 0x0087E
FI_MaxiSteps 2175 0x0087F
FI_ServStepsDec 2176 0x00880
FI_ServStepsInc 2177 0x00881
FI_Ch1SetFlowOld 2178 0x00882
FI_VarFlowFixedStepRemSteps 2179 0x00883
FI_VarFlowFixedStepRemTime 2180 0x00884
FI_TimeAfterLastTarAttempt 2181 0x00885
FI_TimeToNextTarAttempt 2182 0x00886
FI_VarFlowMotorPosition 2183 0x00887
FI_VarFlowCalibInternErrorCode 2184 0x00888
FI_VarFlowCalibOldCalValue 2185 0x00889
FI_VarFlowCalibNewCalValue 2186 0x0088A
FI_DwellOK 2187 0x0088B
FI_HeatDisFlushOK 2188 0x0088C
FI_HeatDisTestOK 2189 0x0088D
FI_HeatFlushOK 2190 0x0088E
FI_RinseOk 2191 0x0088F
FI_CCPrimOK 2192 0x00890
FI_ChemFillupOK 2193 0x00891
FI_SHC_FlushChemValve 2194 0x00892
FI_ShortHeatCitricReady 2195 0x00893
FI_RinseBefHeatDone 2196 0x00894
FI_MaxiStepsReached 2197 0x00895
FI_VarFlowRegReadyCounter 2198 0x00896
FI_VarFlowTrendAnalysisOfFlow 2199 0x00897
FI_VarFlowStabilityOfFlow 2200 0x00898
FI_VarFlowDeltaAvgAvg 2201 0x00899
FI_DiascanTarTimer 2202 0x0089A
FI_WRO300HRecoveryData 2203 0x0089B
FI_Ch2FlowErrorCnt 2204 0x0089C


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 51 (381)
Program version 1.1.0
C EN 2015-03-26

FI_Ch2Calc 2205 0x0089D


FI_MMValveCurrent 2206 0x0089E
FI_InvPredFiltTempBToTempP 2207 0x0089F
FI_StartFlushRevaCount 2208 0x008A0
FI_DiascanPreStepDur 2209 0x008A1
FI_DiascanPostStepDur 2210 0x008A2
FI_CleanCartWaterActive 2211 0x008A3
FI_NoOfPassedFch 2212 0x008A4
FI_RecirCounter 2213 0x008A5
FI_NoOfFirstTarFailed 2214 0x008A6
FI_RecirVolAch 2215 0x008A7
FI_HPGPRFILT 2216 0x008A8
FI_UFSFchIncSteps 2217 0x008A9
FI_StartTarTimer 2218 0x008AA
FI_UFSFchNoOfSteps 2219 0x008AB
FI_OldCondARate 2220 0x008AC
FI_FlowMotorStepsLeft 2221 0x008AD
FI_BLinePrimed 2222 0x008AE
FI_CHVATestResult 2223 0x008AF
FI_MeasuredDegassPress 2224 0x008B0
FI_MeasuredCondB 2225 0x008B1
FI_MeasuredDiaPres 2226 0x008B2
FI_DiaPressMeasureDone 2227 0x008B3
FI_RunCtrlOnDiaPressure 2228 0x008B4
FI_BatteryVoltage 2229 0x008B5
FI_DisinfTypeId 2230 0x008B6
FI_DisinfTypeStartStop 2231 0x008B7
FI_NCrazyUFRates 2232 0x008B8
FI_dwellTimeCombHeat 2233 0x008B9
FI_MeasuredSucPumpRegVarOffs 2234 0x008BA
FI_CleanCartAActive 2235 0x008BB
F_Dummy 2236 0x008BC
F_DummyTO 2237 0x008BD
F_BlLeakMeasRq 2238 0x008BE
F_BubbleTrapRq 2239 0x008BF
F_ByvaCloseCmd 2240 0x008C0
F_DiaPresMeasRq 2241 0x008C1
F_TempMeasRq 2242 0x008C2
F_SagsStsRq 2243 0x008C3
F_DiaPrAch 2244 0x008C4
F_ServTempAch 2245 0x008C5
F_SucDutyCycleOk 2246 0x008C6
F_TavaCloseCmd 2247 0x008C7


52 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_AbortIfBicLatchClosed 2248 0x008C8


F_AirLeakMajorAttn 2249 0x008C9
F_AivaCurrMeasRdy 2250 0x008CA
F_AivaRetestRq 2251 0x008CB
F_AivaSts 2252 0x008CC
F_AlarmBypassRq 2253 0x008CD
F_APDiaClosed 2254 0x008CE
F_AttentionActive 2255 0x008CF
F_AlarmActive 2256 0x008D0
F_AutoRunSts 2257 0x008D1
F_BattActivated 2258 0x008D2
F_BattLowActivated 2259 0x008D3
F_BattSts 2260 0x008D4
F_BCPrimAttn 2261 0x008D5
F_BCPrimAttnRq 2262 0x008D6
F_BCPrimRq 2263 0x008D7
F_BCSts 2264 0x008D8
F_BeforeAIVATo 2265 0x008D9
F_BiCartEmptying 2266 0x008DA
F_BiCartEmptyPressAch 2267 0x008DB
F_BicartEmptyRdy 2268 0x008DC
F_BiCartEmptyWaitTo 2269 0x008DD
F_BiCartEmptyPowdNoCond 2270 0x008DE
F_BiCartExtraFlushTo 2271 0x008DF
F_BiCartFlushTo 2272 0x008E0
F_BicartFlushRdy 2273 0x008E1
F_BicartFlvaCounter 2274 0x008E2
F_BiCartLatchClosedTo 2275 0x008E3
F_BicartPostPrimRdy 2276 0x008E4
F_BicartPreRinseRdy 2277 0x008E5
F_BiCartPrimTo 2278 0x008E6
F_BiCartRdy 2279 0x008E7
F_BiCHeatCheck 2280 0x008E8
F_BicHolderNotClosedAttn 2281 0x008E9
F_BiCInvaCloseRq 2282 0x008EA
F_BicLatchAttnTo 2283 0x008EB
F_BicLatchClosedRq 2284 0x008EC
F_BicLatchSupRq 2285 0x008ED
F_BicLatchOpen 2286 0x008EE
F_BicPrePrimRdy 2287 0x008EF
F_BicPressAch 2288 0x008F0
F_BicPressNotAch 2289 0x008F1
F_BiCPrimCount 2290 0x008F2


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 53 (381)
Program version 1.1.0
C EN 2015-03-26

F_BiCPrimHeatOffRq 2291 0x008F3


F_BiCPrimNormFlowRq 2292 0x008F4
F_BicPrimRdy 2293 0x008F5
F_BicPrimWait4flow 2294 0x008F6
F_BiCStartUpNormFlow 2295 0x008F7
F_BlkDilFillupTo 2296 0x008F8
F_BlLeakCleanAttn 2297 0x008F9
F_BlLeakTstRdy 2298 0x008FA
F_BloodPumpStop 2299 0x008FB
F_BoilingWaterFlag 2300 0x008FC
F_BPumpFillupTo 2301 0x008FD
F_BubbleTestRdy 2302 0x008FE
F_BubbleTrapHLReq 2303 0x008FF
F_HeaterBypass 2304 0x00900
F_HeaterBypassTo 2305 0x00901
F_ByvaDilFillupTo 2306 0x00902
F_BYVADIVACloseRq 2307 0x00903
F_BYVADIVANormal 2308 0x00904
F_BYVADIVANormalRq 2309 0x00905
F_BYVAEvacReq 2310 0x00906
F_ByvaFillupTo 2311 0x00907
F_BYVAOpenRq 2312 0x00908
F_BYVAReq 2313 0x00909
F_ByvaServOpen 2314 0x0090A
F_ByvaSts 2315 0x0090B
F_ChvaSts 2316 0x0090C
F_CCAttachTo 2317 0x0090D
F_CCCondCheckTo 2318 0x0090E
F_CCEmptyTo 2319 0x0090F
F_CCEmptyRinseTo 2320 0x00910
F_CCFillCondOk 2321 0x00911
F_CCFillCondFailed 2322 0x00912
F_CCFivaPathFillTo 2323 0x00913
F_CCFlowTo 2324 0x00914
F_CCLatchAttn 2325 0x00915
F_CCNormalFlowTo 2326 0x00916
F_CCOutletFillTo 2327 0x00917
F_CCPressureTestStart 2328 0x00918
F_CCPressureTestTo 2329 0x00919
F_CCPrimAttn 2330 0x0091A
F_CCPrimCondOk 2331 0x0091B
F_CCPrimRdy 2332 0x0091C
F_CCHeatUp 2333 0x0091D


54 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_CCHeatUpTo 2334 0x0091E


F_CCRivaPathFillTo 2335 0x0091F
F_CeByvaFillupTo 2336 0x00920
F_CeChemFillup15SecTo 2337 0x00921
F_CeChemFillupCount 2338 0x00922
F_CeEvvaFillupTo 2339 0x00923
F_CeExpChamberFillupTo 2340 0x00924
F_CeFinalFillupTo 2341 0x00925
F_CeFlowThrottleCleanTo 2342 0x00926
F_CeRevaFillupTo 2343 0x00927
F_CeRivaFillupTo 2344 0x00928
F_CeZevaFillupTo 2345 0x00929
F_CFchComTstReady 2346 0x0092A
F_CFchComTstAborted 2347 0x0092B
F_DoFchComTestProcRdy 2348 0x0092C
F_CFchComTestVarFlowRdy 2349 0x0092D
F_FchComTestVarFlowProcRdy 2350 0x0092E
F_VarFlowTestProcActive 2351 0x0092F
F_ChemDisSts 2352 0x00930
F_ChemDilFillupCount 2353 0x00931
F_ChemFillupCount 2354 0x00932
F_ChemFillupRdy 2355 0x00933
F_ChemFillupRdyLatch 2356 0x00934
F_ChemFillupRdyLatchClear 2357 0x00935
F_ChemFillupRdyLatchSet 2358 0x00936
F_ChemInactivation 2359 0x00937
F_CirculatingDwellStateActive 2360 0x00938
F_CirculatingDwellProcVar 2361 0x00939
F_CirculationCount 2362 0x0093A
F_CirculationDrvaTo 2363 0x0093B
F_CirculationMainTo 2364 0x0093C
F_CirculationTo 2365 0x0093D
F_CitricDwellSts 2366 0x0093E
F_ClChemDisFlag 2367 0x0093F
F_CleanCartActive 2368 0x00940
F_CleanCartAActive 2369 0x00941
F_CleanCartPrimSts 2370 0x00942
F_ConcPumpFlush 2371 0x00943
F_ConcFillStateActive 2372 0x00944
F_ConcFillupProcVar 2373 0x00945
F_CondAch 2374 0x00946
F_CondBcMode 2375 0x00947
F_CondinState4 2376 0x00948


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 55 (381)
Program version 1.1.0
C EN 2015-03-26

F_CondLow 2377 0x00949


F_CondRateToHigh 2378 0x0094A
F_CondRegStart 2379 0x0094B
F_ContinueTo 2380 0x0094C
F_CWPIntegHeat 2381 0x0094D
F_DegassLvlAch 2382 0x0094E
F_DegassPressBuildRq 2383 0x0094F
F_DegassPrTransTestRetry 2384 0x00950
F_DegassPrTransTestRdy 2385 0x00951
F_DegassPrTransTestRq 2386 0x00952
F_DelayedDiaConDia 2387 0x00953
F_DiaConError 2388 0x00954
F_DiagnosticsReq 2389 0x00955
F_DiaFluid 2390 0x00956
F_DialyzEvacReq 2391 0x00957
F_DialyzerReq 2392 0x00958
F_DiffUF 2393 0x00959
F_DilFillupTimeCorr 2394 0x0095A
F_DilFillStateActive 2395 0x0095B
F_DilFillupProcVar 2396 0x0095C
F_DisActive 2397 0x0095D
F_DisCondHighAttn 2398 0x0095E
F_DisCondLowAttn 2399 0x0095F
F_DisCtrlRdyToStart 2400 0x00960
F_DisinfCtrlActive 2401 0x00961
F_DisinfCtrlErr 2402 0x00962
F_DisinfCtrlRdy 2403 0x00963
F_DisinfOk 2404 0x00964
F_DisinfRestart 2405 0x00965
F_DisCondErr 2406 0x00966
F_DisTimeOff 2407 0x00967
F_DIVAReq 2408 0x00968
F_DivaServOpen 2409 0x00969
F_DivaSts 2410 0x0096A
F_DIVAZEVAReq 2411 0x0096B
F_Drain5SecTo 2412 0x0096C
F_Drain20SecTo 2413 0x0096D
F_Drain30SecTo 2414 0x0096E
F_Drain150SecTo 2415 0x0096F
F_DrainBefChem 2416 0x00970
F_DrainBefChemProcVar 2417 0x00971
F_DrainBefChem5SecTo 2418 0x00972
F_DrainBefChem20SecTo 2419 0x00973


56 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_DrainBefChem30SecTo 2420 0x00974


F_DrainBefChem55SecTo 2421 0x00975
F_DrainBefChem150SecTo 2422 0x00976
F_DrainBefChemDisRdy 2423 0x00977
F_DrainByvaTo 2424 0x00978
F_DrainDivaRinse10UfdTo 2425 0x00979
F_DrainDivaTo 2426 0x0097A
F_DrainEvvaTo 2427 0x0097B
F_DrainExtra 2428 0x0097C
F_DrainMainLineUfdTo 2429 0x0097D
F_DrainMainLineTo 2430 0x0097E
F_DrainMixingChamberATo 2431 0x0097F
F_DrainRdy 2432 0x00980
F_DrainProcVar 2433 0x00981
F_DrainRevaBlkRinse10UfdTo 2434 0x00982
F_DrainRevaBlkTo 2435 0x00983
F_DrainRevaBlkUfdTo 2436 0x00984
F_DrainRevaLineTo 2437 0x00985
F_DrainRivaTo 2438 0x00986
F_DrainStartTo 2439 0x00987
F_DrainSts 2440 0x00988
F_DrainTavaUfdTo 2441 0x00989
F_DrainZevaTo 2442 0x0098A
F_DrvaSts 2443 0x0098B
F_DutyCycleTestOk 2444 0x0098C
F_DwellCirculationRdy 2445 0x0098D
F_DwellCirculationRq 2446 0x0098E
F_DwellRecirc1TempOk 2447 0x0098F
F_DwellRecirc2To 2448 0x00990
F_DwellTimeTo 2449 0x00991
F_Evf1UFCalibPrepComplete 2450 0x00992
F_Evf2UFCalibPrepComplete 2451 0x00993
F_Evf3UFCalibPrepComplete 2452 0x00994
F_Evf4UFCalibPrepComplete 2453 0x00995
F_Evf5UFCalibPrepComplete 2454 0x00996
F_Evf6UFCalibPrepComplete 2455 0x00997
F_Evf7UFCalibPrepComplete 2456 0x00998
F_Evf7UFCalibPrepComplUfd 2457 0x00999
F_Evf8UFCalibPrepComplete 2458 0x0099A
F_Evf8UFCalibPrepComplUfd 2459 0x0099B
F_Evf9UFCalibPrepComplete 2460 0x0099C
F_Evf1UFCh1FlowCal 2461 0x0099D
F_Evf1UFDiffCal 2462 0x0099E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 57 (381)
Program version 1.1.0
C EN 2015-03-26

F_Evf2UFDiffCal 2463 0x0099F


F_Evf1UFIsolVerify 2464 0x009A0
F_Evf1UFMainflOffset 2465 0x009A1
F_Evf2UFMainflOffset 2466 0x009A2
F_Evf1UFVerify 2467 0x009A3
F_EvfDegassingLevel 2468 0x009A4
F_EvfDrainRefVol 2469 0x009A5
F_Evf1CompleteUFTaration 2470 0x009A6
F_Evf2CompleteUFTaration 2471 0x009A7
F_EvfHeaterOnDelay 2472 0x009A8
F_EvfHpgAttnTo 2473 0x009A9
F_EvfPumpDutyTime 2474 0x009AA
F_EvfUFCalibPrep 2475 0x009AB
F_EvvaDilFillupTo 2476 0x009AC
F_EvvaDisOpen 2477 0x009AD
F_EvvaFillupTo 2478 0x009AE
F_EvvaServActiveRq 2479 0x009AF
F_EvvaServOpen 2480 0x009B0
F_EvvaOpenDelay 2481 0x009B1
F_EvvaOpenDelayTO 2482 0x009B2
F_ExpChamberFillupTo 2483 0x009B3
F_FixedDiaPresMinus100Req 2484 0x009B4
F_FixedFlowAndSetReq 2485 0x009B5
F_FchActive 2486 0x009B6
F_FchCondHigh 2487 0x009B7
F_FchControlCondMode 2488 0x009B8
F_FchDegassPrTestMode 2489 0x009B9
F_FchFluidPrepMode 2490 0x009BA
F_FchReady 2491 0x009BB
F_FchSlaveMode 2492 0x009BC
F_FchWaitMode 2493 0x009BD
F_FillPickUpTubeTo 2494 0x009BE
F_FillupCheckCount 2495 0x009BF
F_FillupCheck1To 2496 0x009C0
F_FillupCheck2To 2497 0x009C1
F_FillupPressOk 2498 0x009C2
F_FilterFillupTO 2499 0x009C3
F_FinalCompTo 2500 0x009C4
F_FinalFillupTo 2501 0x009C5
F_FivaRetestRq 2502 0x009C6
F_FivaSts 2503 0x009C7
F_Flow700ml 2504 0x009C8
F_FlowDutyCycleError 2505 0x009C9


58 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_FlowDutyCycleOk 2506 0x009CA


F_FlowDutyCycleTestRq 2507 0x009CB
F_FlowOpenAivaFlvaReq 2508 0x009CC
F_FlowOpenAivaRevaReq 2509 0x009CD
F_FlowPumpDecoupl 2510 0x009CE
F_FlowStoped 2511 0x009CF
F_FlowStopOpenAivaRevaReq 2512 0x009D0
F_FlowTempCondSetReq 2513 0x009D1
F_FlowThrottleFillupTo 2514 0x009D2
F_FlowThrottleCleanTo 2515 0x009D3
F_FluidDirect 2516 0x009D4
F_FluidStatusBypassFlow 2517 0x009D5
F_FluidStatusDiaFlow 2518 0x009D6
F_FlushDivaTavaTo 2519 0x009D7
F_FlushEvvaTo 2520 0x009D8
F_FlushMixingChamberATo 2521 0x009D9
F_FlushRampMixingChamberATo 2522 0x009DA
F_FlushRevaTo 2523 0x009DB
F_FlushRivaTo 2524 0x009DC
F_FlushRivaStartTo 2525 0x009DD
F_FlushSmothMixingChamberATo 2526 0x009DE
F_FlushZevaByvaTo 2527 0x009DF
F_FlPrepStopReq 2528 0x009E0
F_FlPrepCh1RateFloPmp2Req 2529 0x009E1
F_FlPrepNoUFFloPmp2Req 2530 0x009E2
F_FlPrepNoUFFloPmp19Req 2531 0x009E3
F_FlPrepUFRateFloPmp2Req 2532 0x009E4
F_FlPrepUFRateFloPmp5Req 2533 0x009E5
F_FlPrepInactiveFloPmp0Req 2534 0x009E6
F_FlPrepInactiveFloPmp2Req 2535 0x009E7
F_FluidPrepOff 2536 0x009E8
F_FluidPrepOnRq 2537 0x009E9
F_FluidPrepOn 2538 0x009EA
F_FluidPrepRdy 2539 0x009EB
F_FluidPrepRq 2540 0x009EC
F_FluidPrepTo 2541 0x009ED
F_FlvaSts 2542 0x009EE
F_HeatBubbleTrapFillTo 2543 0x009EF
F_HeatByvaTo 2544 0x009F0
F_HeatCoolOffOk 2545 0x009F1
F_HeatCoolOff2To 2546 0x009F2
F_HeatDisFlushRdy 2547 0x009F3
F_HeatDisSts 2548 0x009F4


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 59 (381)
Program version 1.1.0
C EN 2015-03-26

F_HeatDisTestSts 2549 0x009F5


F_HeatDisTestRdy 2550 0x009F6
F_HeaterActiveSts 2551 0x009F7
F_HeaterDisable 2552 0x009F8
F_HeaterOff 2553 0x009F9
F_HeaterOffRq 2554 0x009FA
F_HeaterEff1300 2555 0x009FB
F_HeaterEff1750 2556 0x009FC
F_HeatEvva1To 2557 0x009FD
F_HeatEvva2To 2558 0x009FE
F_HeatFlushReady 2559 0x009FF
F_HeatFlushStartTimeCorr 2560 0x00A00
F_HeatBubbleFillSts 2561 0x00A01
F_HeatFlushSts 2562 0x00A02
F_HeatFlushTimeCorr 2563 0x00A03
F_HeatFreezeFlowPump 2564 0x00A04
F_HeatNoInPressDelayTo 2565 0x00A05
F_HeatNoInPressTo 2566 0x00A06
F_HeatPressRegMeasTo 2567 0x00A07
F_HeatRecirc1TempOk 2568 0x00A08
F_HeatRecirc1TimeCorr 2569 0x00A09
F_HeatRecirc1To 2570 0x00A0A
F_HeatRecirc2Ch2To 2571 0x00A0B
F_HeatRecirc2ContSts 2572 0x00A0C
F_HeatRecirc2DwellTo 2573 0x00A0D
F_HeatRecirc2MainTo 2574 0x00A0E
F_HeatRecirc2MainUfdTo 2575 0x00A0F
F_HeatRecirc2Sts 2576 0x00A10
F_HeatRecirc2TempOk 2577 0x00A11
F_HeatRecirc2TimeCorr 2578 0x00A12
F_HeatRecirc2To 2579 0x00A13
F_HeatRivaTo 2580 0x00A14
F_HeatSucPumpPidOk 2581 0x00A15
F_HeatPumpVoltMeasTo 2582 0x00A16
F_HeatSupTempOk 2583 0x00A17
F_HeatTavaInitPath 2584 0x00A18
F_HeatTavaTo 2585 0x00A19
F_HeatTestRestart 2586 0x00A1A
F_HeatZevaTo 2587 0x00A1B
F_CCHeatUpInit 2588 0x00A1C
F_DecVenPrAlarm 2589 0x00A1D
F_HLDrainReq 2590 0x00A1E
F_HLFillupReq 2591 0x00A1F


60 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_HpgAttnAlarm 2592 0x00A20


F_HpgHigh 2593 0x00A21
F_HPGNegative 2594 0x00A22
F_HPGPrStable 2595 0x00A23
F_HPGTimeout 2596 0x00A24
F_HydraulicRestart 2597 0x00A25
F_IllegalUFRateRqdAttn 2598 0x00A26
F_InactiveActive 2599 0x00A27
F_InactiveUfBoxReq 2600 0x00A28
F_InitForDiffTar 2601 0x00A29
F_InPressOK 2602 0x00A2A
F_InRivaClosedState 2603 0x00A2B
F_IntegHeatAttn 2604 0x00A2C
F_InletFlushTo 2605 0x00A2D
F_IntUFCalCoeffErr 2606 0x00A2E
F_IntUFCalOffsetErr 2607 0x00A2F
F_IntUFFlTempCondSetReq 2608 0x00A30
F_InValveOpen 2609 0x00A31
F_InValveOpenDelayed 2610 0x00A32
F_InvaSts 2611 0x00A33
F_IsolUF 2612 0x00A34
F_IsolUfNoFlowRq 2613 0x00A35
F_IsolUFReq 2614 0x00A36
F_LLDrainReq 2615 0x00A37
F_LLFillupReq 2616 0x00A38
F_LowHeatingTempAttn 2617 0x00A39
F_MainFlAch 2618 0x00A3A
F_MainFlBiCart 2619 0x00A3B
F_MajorAirleakAttnCo 2620 0x00A3C
F_MajorAirleakAttnTo 2621 0x00A3D
F_MajorAirleakAttnOff 2622 0x00A3E
F_MaxFlowOpenAivaReq 2623 0x00A3F
F_MaxFlAch 2624 0x00A40
F_MedFlAch 2625 0x00A41
F_MinFlAch 2626 0x00A42
F_MM90ConcAttn 2627 0x00A43
F_MM90ConcAAttn 2628 0x00A44
F_MM90ConcBAttn 2629 0x00A45
F_MM90CondAch 2630 0x00A46
F_MM90CondOkATime 2631 0x00A47
F_MM90CondOkBTime 2632 0x00A48
F_MM90CondAh 2633 0x00A49
F_MM90CondAl 2634 0x00A4A


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 61 (381)
Program version 1.1.0
C EN 2015-03-26

F_MM90CondAAh 2635 0x00A4B


F_MM90CondAAl 2636 0x00A4C
F_MM90CondCtrlErr 2637 0x00A4D
F_MM90CondACtrlErr 2638 0x00A4E
F_MM90CondBCtrlErr 2639 0x00A4F
F_NegUFKeyTo 2640 0x00A50
F_NoCondModeSel 2641 0x00A51
F_NoDiaFluid 2642 0x00A52
F_NoDilutionAttn 2643 0x00A53
F_NoDilution1To 2644 0x00A54
F_NoDilution2To 2645 0x00A55
F_OffsetTarRq 2646 0x00A56
F_OpenFiva 2647 0x00A57
F_OpenRiva 2648 0x00A58
F_OverheatAlHeater 2649 0x00A59
F_OverHeatAlNorm 2650 0x00A5A
F_PreChemicalDrain 2651 0x00A5B
F_PrimLowRateTo 2652 0x00A5C
F_PrepFillupReq 2653 0x00A5D
F_ProgressiveHLFillupReq 2654 0x00A5E
F_PumpSpeedRatioAlarm 2655 0x00A5F
F_PumpStoppedFlush 2656 0x00A60
F_RdyToShutdown 2657 0x00A61
F_RefMeasFivaRdy 2658 0x00A62
F_RefMeasRdy 2659 0x00A63
F_RegCondAch 2660 0x00A64
F_RegOnHPGAcc 2661 0x00A65
F_RegOnHPGRq 2662 0x00A66
F_RegTempFail 2663 0x00A67
F_RegTempStable 2664 0x00A68
F_RemoveCCAttn 2665 0x00A69
F_ResetRiva 2666 0x00A6A
F_RevaDilFillupTo 2667 0x00A6B
F_RevaFillupTo 2668 0x00A6C
F_RevaPathFillTo 2669 0x00A6D
F_RevaSts 2670 0x00A6E
F_RinseBicartFlvaPath 2671 0x00A6F
F_RinseByvaTo 2672 0x00A70
F_RinseEvvaPath 2673 0x00A71
F_RinseEvvaTo 2674 0x00A72
F_RinseProcVar 2675 0x00A73
F_RinseRivaPath 2676 0x00A74
F_RinseRivaTo 2677 0x00A75


62 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_RinseRevaTo 2678 0x00A76


F_RinseStateActive 2679 0x00A77
F_RinseTavaPath 2680 0x00A78
F_RinseZevaPath 2681 0x00A79
F_Rinse10SecTo 2682 0x00A7A
F_Rinse20SecTo 2683 0x00A7B
F_Rinse30SecTo 2684 0x00A7C
F_Rinse50SecTo 2685 0x00A7D
F_Rinse60SecTo 2686 0x00A7E
F_Rinse120SecTo 2687 0x00A7F
F_RinseBefDisable 2688 0x00A80
F_RinseBefHeatTo 2689 0x00A81
F_RinseBefHeatUfdTo 2690 0x00A82
F_RinseCycleReady 2691 0x00A83
F_RinseRdy 2692 0x00A84
F_RinseSts 2693 0x00A85
F_RinseTavaTo 2694 0x00A86
F_RinseTime 2695 0x00A87
F_RinseZevaTo 2696 0x00A88
F_RivaDilFillupTo 2697 0x00A89
F_RivaFillupTo 2698 0x00A8A
F_RIVAOpenRq 2699 0x00A8B
F_RivaSts 2700 0x00A8C
F_RivaUFCalib 2701 0x00A8D
F_SagsTestAttn 2702 0x00A8E
F_SeqUFVolResetRdy 2703 0x00A8F
F_ServActive 2704 0x00A90
F_ServDiaPrReq 2705 0x00A91
F_ServStateReq 2706 0x00A92
F_SetChannel1RateReq 2707 0x00A93
F_SetConstantFlowReq 2708 0x00A94
F_SetConstantVoltageReq 2709 0x00A95
F_SetConstLowLevelPressReq 2710 0x00A96
F_SetDiaPress0mmReq 2711 0x00A97
F_SetDiaPressMinus400mmReq 2712 0x00A98
F_SetDiaPrToCalHLStatPrReq 2713 0x00A99
F_SetDiaPrToCalLLStatPrReq 2714 0x00A9A
F_SetToCalPIDValueReq 2715 0x00A9B
F_SetUFRateMinus300mlReq 2716 0x00A9C
F_SetUFRateMinus50mlReq 2717 0x00A9D
F_SetUFRateReq 2718 0x00A9E
F_ShutOffCoolingTo 2719 0x00A9F
F_StartIntUFCalib 2720 0x00AA0


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 63 (381)
Program version 1.1.0
C EN 2015-03-26

F_StartUpSequenceRdy 2721 0x00AA1


F_ChemFlagReceived 2722 0x00AA2
F_StickAAttnRq 2723 0x00AA3
F_StickAAttnTo 2724 0x00AA4
F_StickAInPos 2725 0x00AA5
F_StickAInRq 2726 0x00AA6
F_StickAOutRq 2727 0x00AA7
F_StickBAttnRq 2728 0x00AA8
F_StickBAttnTo 2729 0x00AA9
F_StickBInPos 2730 0x00AAA
F_StickBInRq 2731 0x00AAB
F_StickBInstAttnRq 2732 0x00AAC
F_StickBOutRq 2733 0x00AAD
F_StopPumpReq 2734 0x00AAE
F_SucDutyCycleTestRq 2735 0x00AAF
F_SucPumpDecoupl 2736 0x00AB0
F_SuctionDutyCycleError 2737 0x00AB1
F_SucPumpNormalRq 2738 0x00AB2
F_SucPumpUFTarBiCPrim 2739 0x00AB3
F_SucPumpUFPrim 2740 0x00AB4
F_TaraCmd 2741 0x00AB5
F_TaraRdyCmd 2742 0x00AB6
F_TarationRdy 2743 0x00AB7
F_TarationRq 2744 0x00AB8
F_TarErr 2745 0x00AB9
F_TarOk 2746 0x00ABA
F_TavaDisOpen 2747 0x00ABB
F_TavaCloseBypPrim 2748 0x00ABC
F_TavaCloseBypCmd 2749 0x00ABD
F_TavaCloseRq 2750 0x00ABE
F_TAVAReq 2751 0x00ABF
F_TavaServOpen 2752 0x00AC0
F_TempAch 2753 0x00AC1
F_TempAH 2754 0x00AC2
F_TempAL 2755 0x00AC3
F_TmtActive 2756 0x00AC4
F_UFAccCh1ZeroRdy 2757 0x00AC5
F_UFBeenCalibrated 2758 0x00AC6
F_UFBoxMode0Req 2759 0x00AC7
F_UFBoxMode1Selected 2760 0x00AC8
F_UFBoxMode2Selected 2761 0x00AC9
F_UFBoxMode3Selected 2762 0x00ACA
F_UFBoxMode4Selected 2763 0x00ACB


64 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_UFCalCoeffLimErr 2764 0x00ACC


F_UFCalErr 2765 0x00ACD
F_UFCalMailErr 2766 0x00ACE
F_UFCalEstStatPrRdy 2767 0x00ACF
F_UFCalibBelowHL 2768 0x00AD0
F_UFCalibBelowLL 2769 0x00AD1
F_UFCalibHLReached 2770 0x00AD2
F_UFCalibLLReached 2771 0x00AD3
F_UFCalibration 2772 0x00AD4
F_UFCalOffsetLimErr 2773 0x00AD5
F_UFCalPrAch 2774 0x00AD6
F_UFCalRdy 2775 0x00AD7
F_UFCalReady 2776 0x00AD8
F_UFCalWaitState 2777 0x00AD9
F_UFCalUserAborted 2778 0x00ADA
F_UFCellCondLimitAch 2779 0x00ADB
F_UFCh1CalRdy 2780 0x00ADC
F_UFCh1FlCalError 2781 0x00ADD
F_UFDiffCalError 2782 0x00ADE
F_UFDiffTarRq 2783 0x00ADF
F_UfdPrimeCounter 2784 0x00AE0
F_UfdPrimeInDelay 2785 0x00AE1
F_UfdPrimeOutDelay 2786 0x00AE2
F_UfdPrimeInsideTo 2787 0x00AE3
F_UfdPrimeOutsideTo 2788 0x00AE4
F_UFDrainRefVolFinished 2789 0x00AE5
F_UFEEPromCheckOk 2790 0x00AE6
F_UFEEPromInitOk 2791 0x00AE7
F_UFEEPromStoreOk 2792 0x00AE8
F_UFEEPromStoreError 2793 0x00AE9
F_UFGainActive 2794 0x00AEA
F_UFIsolVerifyError 2795 0x00AEB
F_UFLogRdyToShutDown 2796 0x00AEC
F_UfLogBufferStored 2797 0x00AED
F_UFGainAttn 2798 0x00AEE
F_UfAchievedAttn 2799 0x00AEF
F_UfVolAchieved 2800 0x00AF0
F_UFNegAttn 2801 0x00AF1
F_UFRateFiltFreezeRq 2802 0x00AF2
F_UFTarHPGPrStab 2803 0x00AF3
F_UFTarHPGPrStabWaitTo 2804 0x00AF4
F_UFCalIsolErr 2805 0x00AF5
F_UFInternalError 2806 0x00AF6


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 65 (381)
Program version 1.1.0
C EN 2015-03-26

F_UFInternalReady 2807 0x00AF7


F_UFIsolVerifyRdy 2808 0x00AF8
F_UFIsolVerStartRdy 2809 0x00AF9
F_UFLinErr 2810 0x00AFA
F_UFLinVerNextFlow 2811 0x00AFB
F_UFLinVerMeasRdy 2812 0x00AFC
F_UFLinVerRdy 2813 0x00AFD
F_UFLogAddErrorRecordRq 2814 0x00AFE
F_UFLogAddRetarationRecRq 2815 0x00AFF
F_UFLogAddTarationNotOkRecRq 2816 0x00B00
F_UFLogAddTarationOkRecRq 2817 0x00B01
F_UFMainFlOffsetCalRdy 2818 0x00B02
F_UFMainFlOffsetMeasError 2819 0x00B03
F_RecirVolAch 2820 0x00B04
F_UFMainFlOffsetSts 2821 0x00B05
F_UFCh1CalPhaseSts 2822 0x00B06
F_UFDiffCalSts 2823 0x00B07
F_UFVerifyPhaseSts 2824 0x00B08
F_UFBoxMode1Sts 2825 0x00B09
F_UFBoxMode2Sts 2826 0x00B0A
F_UFBoxMode3Sts 2827 0x00B0B
F_UFBoxMode4Sts 2828 0x00B0C
F_UFWaitSts 2829 0x00B0D
F_UFStabilizeSts 2830 0x00B0E
F_UFOffTarRq 2831 0x00B0F
F_UFTarActive 2832 0x00B10
F_UFTarNotAllowed 2833 0x00B11
F_UFTarPostProcAct 2834 0x00B12
F_UFTarPostProcRdy 2835 0x00B13
F_UFTarPreProcAct 2836 0x00B14
F_UFTarpreProcRdy 2837 0x00B15
F_UFTarProcAct 2838 0x00B16
F_UFTarProcRdy 2839 0x00B17
F_UFTarRdy 2840 0x00B18
F_UFTarRIVAOpenTO 2841 0x00B19
F_ValveCurrMeasRdy 2842 0x00B1A
F_ValveOpen 2843 0x00B1B
F_ValveOpen0 2844 0x00B1C
F_ValveOpen1 2845 0x00B1D
F_ValveOpen2 2846 0x00B1E
F_ValveOpen3 2847 0x00B1F
F_ValveOpen4 2848 0x00B20
F_ValveOpen5 2849 0x00B21


66 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_ValveOpen6 2850 0x00B22


F_ValveOpen7 2851 0x00B23
F_ValveOpen8 2852 0x00B24
F_ValveOpen9 2853 0x00B25
F_ValveOpen10 2854 0x00B26
F_ValveTestSts 2855 0x00B27
F_UFVerifyError 2856 0x00B28
F_UFVerifyRdy 2857 0x00B29
F_UFVerStartRdy 2858 0x00B2A
F_UFVolResetRdy 2859 0x00B2B
F_UPSSts 2860 0x00B2C
F_UPSActivated 2861 0x00B2D
F_WaitForShutDown 2862 0x00B2E
F_WaitForPTarationTo 2863 0x00B2F
F_WaitForNotPTarationTo 2864 0x00B30
F_WrChemDisFlag 2865 0x00B31
F_WROTurnedOn 2866 0x00B32
F_ZEVACloseRq 2867 0x00B33
F_ZevaDilFillupTo 2868 0x00B34
F_ZevaFillupTo 2869 0x00B35
F_ZEVAOpenRq 2870 0x00B36
F_ZEVAReq 2871 0x00B37
F_ZevaSts 2872 0x00B38
F_LowDialysisFlowWaitTo 2873 0x00B39
F_LDFDelayPostTarTo 2874 0x00B3A
F_HighDialysisFlowWaitTo 2875 0x00B3B
F_IntUFCoeffNotUpdated 2876 0x00B3C
F_CVlvTstOpenedDIVA 2877 0x00B3D
F_CVlvTstOpenedBYVA 2878 0x00B3E
F_CVlvTstClosedAll 2879 0x00B3F
F_WaitForTMPStableTo 2880 0x00B40
F_CVlvTstOpenedDRVA 2881 0x00B41
F_CVlvTstOpenedREVA 2882 0x00B42
F_CVlvTstOpenedZEVA 2883 0x00B43
F_CVlvTstOpenedRIVA 2884 0x00B44
F_CVlvTstOpenedINVA 2885 0x00B45
F_CVlvTstOpenedFLVA 2886 0x00B46
F_CVlvTstOpenedCHVA 2887 0x00B47
F_CVlvTstClosedAll_B 2888 0x00B48
F_CVlvTstAckedRdy 2889 0x00B49
F_CVlvTstAckedRdy_B 2890 0x00B4A
F_CVlvTstOpenDRVARq 2891 0x00B4B
F_CVlvTstOpenREVARq 2892 0x00B4C


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 67 (381)
Program version 1.1.0
C EN 2015-03-26

F_CVlvTstOpenZEVARq 2893 0x00B4D


F_CVlvTstOpenRIVARq 2894 0x00B4E
F_CVlvTstOpenINVARq 2895 0x00B4F
F_CVlvTstOpenFLVARq 2896 0x00B50
F_CVlvTstOpenCHVARq 2897 0x00B51
F_INVAOverridePower 2898 0x00B52
F_INVAOverrideMannHeat 2899 0x00B53
F_INVAOverride 2900 0x00B54
F_FchProlongPowerAttn 2901 0x00B55
F_FchProlongMannHeatAttn 2902 0x00B56
F_FchResetRdyCOMMON 2903 0x00B57
F_FchResetRdyBLOOD 2904 0x00B58
F_FchResetRdyFLUID 2905 0x00B59
F_FchResetRdyCOND 2906 0x00B5A
F_FchResetRdyTEMP 2907 0x00B5B
F_NegDiaPressOK 2908 0x00B5C
F_CondUFDFlushOK 2909 0x00B5D
F_InletCondTooHigh 2910 0x00B5E
F_InletCondSupInst 2911 0x00B5F
F_InletCondHighConfirm 2912 0x00B60
F_InletCondDataExists 2913 0x00B61
F_InletCondHighAttn 2914 0x00B62
F_InletCondAccepted 2915 0x00B63
F_InletCondNotAccepted 2916 0x00B64
F_CVTOpenDIVA 2917 0x00B65
F_CVTOpenBYVA 2918 0x00B66
F_CVTCloseAll 2919 0x00B67
F_CVTReady 2920 0x00B68
F_CVTOpenDRVA 2921 0x00B69
F_CVTOpenREVA 2922 0x00B6A
F_CVTOpenZEVA 2923 0x00B6B
F_CVTOpenRIVA 2924 0x00B6C
F_CVTOpenINVA 2925 0x00B6D
F_CVTOpenFLVA 2926 0x00B6E
F_CVTOpenCHVA 2927 0x00B6F
F_CVTCloseAll_B 2928 0x00B70
F_CVTReady_B 2929 0x00B71
F_HeatCCEmptyTo 2930 0x00B72
F_HeatLowInletPressDelayTo 2931 0x00B73
F_StopHeating 2932 0x00B74
F_DRVAOpenFT 2933 0x00B75
F_FirstTimeOpen 2934 0x00B76
F_RegTempStableNew 2935 0x00B77


68 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_CloseDrvaSF 2936 0x00B78


F_OpenDrvaSF 2937 0x00B79
F_FineTuneFlowInc 2938 0x00B7A
F_FineTuneFlowDec 2939 0x00B7B
F_SmallFlowDiff 2940 0x00B7C
F_StableSetFlow 2941 0x00B7D
F_AdjustingDrvaSF 2942 0x00B7E
F_VarFlowFixedStepRemTimeZero 2943 0x00B7F
F_VarFlowRegIneffTo 2944 0x00B80
F_VarFlowRegEffTo 2945 0x00B81
F_VarFlowRegDelayPostTarTo 2946 0x00B82
F_VarFlowMotorCtrlIdle 2947 0x00B83
F_VarFlowCalibSuccess 2948 0x00B84
F_VarFlowCalibNotSuccess 2949 0x00B85
F_FluidTempTestRdy 2950 0x00B86
F_VarFlowMotorCalibrating 2951 0x00B87
F_VFlMotorCalibRdySuccess 2952 0x00B88
F_VFlMotorCalibRdyFailure 2953 0x00B89
F_CurrMeasFlowInServiceValid 2954 0x00B8A
F_VarFlowCalibValOutRange 2955 0x00B8B
F_VarFlowMotorRegInactive 2956 0x00B8C
F_MachineOffDelayWaitTo 2957 0x00B8D
F_InitRamp 2958 0x00B8E
F_FilterFillup 2959 0x00B8F
F_FilterFilled 2960 0x00B90
F_DiascanMeasActive 2961 0x00B91
F_DiascanMeasDone 2962 0x00B92
F_DiascanMeasRdy 2963 0x00B93
F_DiascanNegStep 2964 0x00B94
F_DiascanPosStep 2965 0x00B95
F_DiascanReceiveRdy 2966 0x00B96
F_DiascanStepActive 2967 0x00B97
F_Diascan_AF_START 2968 0x00B98
F_Diascan_AF_WAIT 2969 0x00B99
F_Diascan_AF_RAISECOND 2970 0x00B9A
F_Diascan_AF_NORMALFLOWEST 2971 0x00B9B
F_Diascan_AF_LOWERCOND 2972 0x00B9C
F_Diascan_AF_CALC 2973 0x00B9D
F_Diascan_AF_END 2974 0x00B9E
F_Diascan_AF_DONE 2975 0x00B9F
F_Diascan_AF_FAIL 2976 0x00BA0
F_Diascan_AF_NOTVALID 2977 0x00BA1
F_DiascanProtConfirmOk 2978 0x00BA2


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 69 (381)
Program version 1.1.0
C EN 2015-03-26

F_HpgStableTo 2979 0x00BA3


F_WRO300HRUNTimeout 2980 0x00BA4
F_WRO300HLFHTimeout 2981 0x00BA5
F_WRO300HRUNRequested 2982 0x00BA6
F_WRO300HLFHRequested 2983 0x00BA7
F_ExtDisinfectionSts 2984 0x00BA8
F_ExtDisinfRdy 2985 0x00BA9
F_ExtDisinfAbort 2986 0x00BAA
F_ExtDisinfExtPrepSts 2987 0x00BAB
F_ExtDisinfIntPrepSts 2988 0x00BAC
F_ExtDisinfLFHStarted 2989 0x00BAD
F_ExtDisinfLFHRdy 2990 0x00BAE
F_ExtDisinfPrepAttnRq 2991 0x00BAF
F_ExtDStartLFHModeTimeoutTo 2992 0x00BB0
F_IssueAttLFHnotAvail 2993 0x00BB1
F_LowFlowHeatProcFlowSts 2994 0x00BB2
F_W300RemColdPrepTimeZero 2995 0x00BB3
F_W300RemWarmPrepTimeZero 2996 0x00BB4
F_W300RemWaterDistrTimeZero 2997 0x00BB5
F_W300RemLowFlowHeatTimeZero 2998 0x00BB6
F_ExtDisinfComProcLFHRq 2999 0x00BB7
F_ExtDisinfComProcLFHEndRq 3000 0x00BB8
F_ManStartLFHModeAttnRq 3001 0x00BB9
F_DoingActiveDisinfSts 3002 0x00BBA
F_DisIntHasOccured 3003 0x00BBB
F_CloseINVADueToW300 3004 0x00BBC
F_W300ProtStsStandby 3005 0x00BBD
F_W300ProtStsHeat 3006 0x00BBE
F_W300ProtStsRun 3007 0x00BBF
F_W300ProtStsLFH 3008 0x00BC0
F_W300ProtStsCentrChem 3009 0x00BC1
F_W300ProtStsLowFlowRun 3010 0x00BC2
F_W300ProtLFHFeatureStsKnown 3011 0x00BC3
F_W300ProtLFHFeatureInstalled 3012 0x00BC4
F_ICINVAWRO300LFH 3013 0x00BC5
F_W300ProtHEATFeatureStsKnown 3014 0x00BC6
F_W300ProtHEATFeatureInstalled 3015 0x00BC7
F_W300ProtLFRunFeatureStsKnown 3016 0x00BC8
F_W300ProtLFRunFeatureInstalled 3017 0x00BC9
F_W300ProtCenChFeatureStsKnown 3018 0x00BCA
F_W300ProtCenChFeatureInstalled 3019 0x00BCB
F_WROStsShouldBeKnown 3020 0x00BCC
F_W300CommFailure 3021 0x00BCD


70 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_InitINVAOpenDelayTimeout 3022 0x00BCE


F_CbvaSts 3023 0x00BCF
F_StickPosValveTestOK 3024 0x00BD0
F_ReverseDir 3025 0x00BD1
F_StickBOutIfChemRq 3026 0x00BD2
F_StickBInNoDelayRq 3027 0x00BD3
F_BiCLatchClosedNoDelayRq 3028 0x00BD4
F_CloseBicLatchAttnNoDelayRq 3029 0x00BD5
F_StickAInNoDelayRq 3030 0x00BD6
F_StickCEmpty 3031 0x00BD7
F_ChemInStickC 3032 0x00BD8
F_StickBIsUsed 3033 0x00BD9
F_FluidPathLeakTstEnable 3034 0x00BDA
F_StickCInNoDelayRq 3035 0x00BDB
F_FCHInfoStickAttnRq 3036 0x00BDC
F_FluidPathLeakTestAttnRq 3037 0x00BDD
F_CbvaCurrMeasRdy 3038 0x00BDE
F_ValveTestRdy 3039 0x00BDF
F_Ch2FlowError 3040 0x00BE0
F_EmptyCCATo 3041 0x00BE1
F_StickStatusActive 3042 0x00BE2
F_FluidCircuitAlarm 3043 0x00BE3
F_FLVACurrTestOpen 3044 0x00BE4
F_FlvaCurrMeasRdy 3045 0x00BE5
F_CValvesCurrTestFinished 3046 0x00BE6
F_SucPumpLowLimReached 3047 0x00BE7
F_IncDiaFlowTo500Alarm 3048 0x00BE8
F_BackFlushUfdTo 3049 0x00BE9
F_BackFlushUFDActive 3050 0x00BEA
F_ToggleFiva 3051 0x00BEB
F_HeatRecirc2ByvaZevaTo 3052 0x00BEC
F_FillupCoolOffTo 3053 0x00BED
F_ZevaFillCitricUfdTo 3054 0x00BEE
F_ByvaFillCitricUfdTo 3055 0x00BEF
F_ExpChambFillCitricUfdTo 3056 0x00BF0
F_CCOutletFillUfdTo 3057 0x00BF1
F_AbortDisHighHpg 3058 0x00BF2
F_NewPredictedTempPSample 3059 0x00BF3
F_AIVACurrTestActive 3060 0x00BF4
F_FCHTempTestsRdy 3061 0x00BF5
F_StartFlushRevaCount 3062 0x00BF6
F_RevaFlush 3063 0x00BF7
F_BackFlushUFDDrainTo 3064 0x00BF8


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 71 (381)
Program version 1.1.0
C EN 2015-03-26

F_IsolUFActive 3065 0x00BF9


F_EvvaOpenHighPres 3066 0x00BFA
F_LatchedDiaPresTooHigh 3067 0x00BFB
F_PressureCoolDownTo 3068 0x00BFC
F_PressureCoolDownTo2 3069 0x00BFD
F_FluidFlowOkForUFCalib 3070 0x00BFE
F_AdjustFluidFlowForUFCal 3071 0x00BFF
F_HighCondIsDesired 3072 0x00C00
F_InternSuppProc2AdjFlow 3073 0x00C01
F_InternSuppProc2AdjInit 3074 0x00C02
F_InternSuppProc2RdyAdj 3075 0x00C03
F_RinseBLineTimerReady 3076 0x00C04
F_HeaterCoolDownAch 3077 0x00C05
F_CCPrimCoolDownTo 3078 0x00C06
F_CCPrimCoolDown 3079 0x00C07
F_CCHeatCoolDown 3080 0x00C08
F_CCHeatCoolDownTo 3081 0x00C09
F_DiascanCalc 3082 0x00C0A
F_SagsTestAttnTo 3083 0x00C0B
F_SagsTestDrainAttnTo 3084 0x00C0C
F_SagsTestInitActive 3085 0x00C0D
F_SagsTestDrainInitActive 3086 0x00C0E
F_RinseBLineDrainBTo 3087 0x00C0F
F_RinseBLineDrainInitTo 3088 0x00C10
F_HeatFlushNoInPressDelayTo 3089 0x00C11
F_RinseBlineActive 3090 0x00C12
F_SetCCWaterActive 3091 0x00C13
F_CleanCartWaterActive 3092 0x00C14
F_OneTimeRq 3093 0x00C15
F_FirstUFTarDone 3094 0x00C16
F_RestartFluidPrepTo 3095 0x00C17
F_DisinfFillupComplAttnRq 3096 0x00C18
F_FirstUFTarFailed 3097 0x00C19
F_MaxUFTarAttempts 3098 0x00C1A
F_TavaCloseBypTar 3099 0x00C1B
F_DontAdjustFlow 3100 0x00C1C
F_HeatRevaFillAutoDisTo 3101 0x00C1D
F_HeatUFDFillAutoDisTo 3102 0x00C1E
F_FillFlvaTo 3103 0x00C1F
F_SetflowInstalled 3104 0x00C20
F_UfdInstalled 3105 0x00C21
F_KeepTAVAClosedAfterTar 3106 0x00C22
F_TAVACloseDelayTo 3107 0x00C23


72 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_ConsiderDiaPressure 3108 0x00C24


F_HpgNorm2AttnTo 3109 0x00C25
F_MM90TarCondLimLowAC 3110 0x00C26
F_MM90TarCondLimLowBC 3111 0x00C27
F_HbvaSts 3112 0x00C28
F_ValveOpenFIVA 3113 0x00C29
F_ValveOpenCBVA 3114 0x00C2A
F_ValveOpenCHVA 3115 0x00C2B
F_ValveOpenHBVA 3116 0x00C2C
F_ChvaServOpen 3117 0x00C2D
F_HbvaCurrTestOpen 3118 0x00C2E
F_HbvaCurrMeasRdy 3119 0x00C2F
F_FirstUFTarRdy 3120 0x00C30
F_UFSFchStart 3121 0x00C31
F_UFSFchStop 3122 0x00C32
F_UFSFchStarted 3123 0x00C33
F_UFSFchStopped 3124 0x00C34
F_UFSFchReady 3125 0x00C35
F_UFSFchActive 3126 0x00C36
F_WaitUfTarPreProcRdyTo 3127 0x00C37
F_WaitPostUfTarTo 3128 0x00C38
F_HBVAInstalled 3129 0x00C39
F_UFSFchIncFlow 3130 0x00C3A
F_SendBYVATarStsToP 3131 0x00C3B
F_UFSStartMainFlowMeas 3132 0x00C3C
F_InletOpen 3133 0x00C3D
F_TempTestCoolRdy 3134 0x00C3E
F_OpenINVAandHBVA 3135 0x00C3F
F_InpsSts 3136 0x00C40
F_InPressNotOK 3137 0x00C41
F_RinseBefHeatInit 3138 0x00C42
F_RinseBefHeatRdy 3139 0x00C43
F_RinseBefHeatNoInPrTo 3140 0x00C44
F_UFSFchAbort 3141 0x00C45
F_DelayCondAlarmUFSFchTo 3142 0x00C46
F_DelayTimeForUFS 3143 0x00C47
F_NoDelayCondAlarmUFSFch 3144 0x00C48
F_WaitForStableFluidTo 3145 0x00C49
F_UFSFchRecover 3146 0x00C4A
F_UFSFchFlowRecoverRdy 3147 0x00C4B
F_DelaySelectProcTo 3148 0x00C4C
F_CondLoBadFluid 3149 0x00C4D
F_CondHiBadFluid 3150 0x00C4E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 73 (381)
Program version 1.1.0
C EN 2015-03-26

F_PrepareWaitForStickTo 3151 0x00C4F


F_CirculationRivaTo 3152 0x00C50
F_DrainBLineTo 3153 0x00C51
F_RinseBLineDrainATo 3154 0x00C52
F_WaterColumnCCATo 3155 0x00C53
F_WaterColumnCCBTo 3156 0x00C54
F_EmptyCCBTo 3157 0x00C55
F_EmptyCCB2To 3158 0x00C56
F_EmptyCCTo 3159 0x00C57
F_IgnoreLowTemp 3160 0x00C58
F_FillupPrSetTrueOnes 3161 0x00C59
F_CCHeatupSetTrueOnes 3162 0x00C5A
F_OpenBicLatchAct 3163 0x00C5B
F_DelayNoFlowTo 3164 0x00C5C
F_DegassPrOk 3165 0x00C5D
F_DrvaCurrMeasRdy 3166 0x00C5E
F_PresetCRCOKFlag 3167 0x00C5F
F_UfTarationZeroFlowFinished 3168 0x00C60
F_UfTarationDiffFlowFinished 3169 0x00C61
F_UfTarationFinishedOk 3170 0x00C62
F_UFTarSeqProc 3171 0x00C63
F_TarationTO 3172 0x00C64
F_RinseBLineBicTo 3173 0x00C65
F_DilDisCtrlRdyToStart 3174 0x00C66
F_CentrDisCtrlRdyToStart 3175 0x00C67
F_DisCondHighAttnRq 3176 0x00C68
F_DisCondLowAttnRq 3177 0x00C69
F_UFSFchAborted 3178 0x00C6A
F_EmptyCCBufdTo 3179 0x00C6B
F_EmptyCCB2ufdTo 3180 0x00C6C
F_EmptyCCAufdTo 3181 0x00C6D
F_WaterColumnCCBufdTo 3182 0x00C6E
F_DisinfCtrlRdyCount 3183 0x00C6F
F_UFTarUnstableFlow 3184 0x00C70
F_UFSTmtRetarRq 3185 0x00C71
F_OpenFLVAToPrimeBLineTO 3186 0x00C72
F_PrimingBLine 3187 0x00C73
F_BLinePrimed 3188 0x00C74
F_HeaterCoolDownAchTrueOnes 3189 0x00C75
F_HeatCCEmptyNoInPressDelayTo 3190 0x00C76
F_ReverseDir2 3191 0x00C77
F_ShortHeatCitricReady 3192 0x00C78
F_SHC_RinseRevaLineTo 3193 0x00C79


74 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_SHC_RinseBLineTo 3194 0x00C7A


F_SHC_RinseALineAndMixChambTo 3195 0x00C7B
F_SHC_RinseRivaLineTo 3196 0x00C7C
F_SHC_PrimeBLineWithCitricTo 3197 0x00C7D
F_SHC_RinseMainLineTo 3198 0x00C7E
F_SHC_RinseZevaByvaLineTo 3199 0x00C7F
F_SHC_RinseFivaLineTo 3200 0x00C80
F_SHC_CitricFillupMainLineTo 3201 0x00C81
F_SHC_DisinfCtrlRdyToStart 3202 0x00C82
F_SHC_FlushChemValve 3203 0x00C83
F_FlowPumpRefPressMinus50 3204 0x00C84
F_FlowPumpRefPressMinus250 3205 0x00C85
F_FlowPumpRefPressMinus300 3206 0x00C86
F_flowPumpRegConst2 3207 0x00C87
F_flowPumpRegConstHeat3 3208 0x00C88
F_FlowPumpRefDutyCycle44Perc 3209 0x00C89
F_FlowPumpRefDutyCycle20Perc 3210 0x00C8A
F_FlowPumpRefDutyCycle18Perc 3211 0x00C8B
F_FlowPumpRefDutyCycle12Perc 3212 0x00C8C
F_FlowPumpRefDutyCycle4Perc 3213 0x00C8D
F_HeaterRef60DegC 3214 0x00C8E
F_HeaterRef50DegC 3215 0x00C8F
F_HeaterRefPresettedHeatTemp 3216 0x00C90
F_SucPumpRefPressMinus100 3217 0x00C91
F_SucPumpRefPressMinus200 3218 0x00C92
F_SucPumpRefPressZero 3219 0x00C93
F_SucPumpUseRegConst3 3220 0x00C94
F_SucPumpUseRegConst5 3221 0x00C95
F_InletCondMeasRq 3222 0x00C96
F_ShortHeatCitricSts 3223 0x00C97
F_FillupFinalFlushRatioTo 3224 0x00C98
F_FlushChemValveLine1To 3225 0x00C99
F_FlushChemValveLine2To 3226 0x00C9A
F_FlushChemValveLine3To 3227 0x00C9B
F_PreRinseChemValveLineTo 3228 0x00C9C
F_RinseChemValveLineTo 3229 0x00C9D
F_StopPumpsBeforeCHVAtestTo 3230 0x00C9E
F_PostRinseChemValveLineTo 3231 0x00C9F
F_CHVACurrentTestStartToP 3232 0x00CA0
F_ByvaDivaTarClosed 3233 0x00CA1
F_CHVACurrentTestStartSent 3234 0x00CA2
F_CHVACurrentTestResultFromP 3235 0x00CA3
F_PresetCRCArrived 3236 0x00CA4


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 75 (381)
Program version 1.1.0
C EN 2015-03-26

F_CHVACurrentTestOkFromP 3237 0x00CA5


F_CHVACurrentTestErrorFromP 3238 0x00CA6
F_CHVACurrentTestOk 3239 0x00CA7
F_CHVACurrentTestError 3240 0x00CA8
F_CHVAPressureTestActive1 3241 0x00CA9
F_CHVAPressureTestActive2 3242 0x00CAA
F_CHVAPressureTestPressStable 3243 0x00CAB
F_CHVAPressureTestPressOk 3244 0x00CAC
F_CHVAPressureTestPressError 3245 0x00CAD
F_CHVAPressureTestOk 3246 0x00CAE
F_CHVAPressureTestError 3247 0x00CAF
F_CHVAPumpSlowdownDelayTo 3248 0x00CB0
F_ChC_FillPickUpTubeTo 3249 0x00CB1
F_ChC_ExpChamberFillupTo 3250 0x00CB2
F_ChC_ExpChambFillCitricUfdTo 3251 0x00CB3
F_ChC_FlowThrottleCleanTo 3252 0x00CB4
F_ChC_RivaFillupTo 3253 0x00CB5
F_ChC_ZevaFillupTo 3254 0x00CB6
F_ChC_ZevaFillCitricUfdTo 3255 0x00CB7
F_ChC_ByvaFillupTo 3256 0x00CB8
F_ChC_ByvaFillCitricUfdTo 3257 0x00CB9
F_ChC_RevaFillupTo 3258 0x00CBA
F_ChC_EvvaFillupTo 3259 0x00CBB
F_ChC_PrepareWaitForStickTo 3260 0x00CBC
F_ChC_FinalFillupTo 3261 0x00CBD
F_ChC_FillupCoolOffTo 3262 0x00CBE
F_ChC_RinseToFlvaRatioTo 3263 0x00CBF
F_ChC_FillupFinalFlushRatioTo 3264 0x00CC0
F_ChC_ChemFillupCount 3265 0x00CC1
F_ChC_DisinfCtrlRdyCount 3266 0x00CC2
F_CHVATestOk 3267 0x00CC3
F_CHVATestError 3268 0x00CC4
F_SHC_RinseMainLineUFDTo 3269 0x00CC5
F_SHC_DetectCitricTo 3270 0x00CC6
F_SHC_RinseEvvaLineTo 3271 0x00CC7
F_FlowPumpRefFlow500 3272 0x00CC8
F_FlowPumpRefFlow800 3273 0x00CC9
F_SHC_CitricFillupRevaTo 3274 0x00CCA
F_SHC_HeatAndCitricRecircTavaTo 3275 0x00CCB
F_SHC_HeatDistrTavaTo 3276 0x00CCC
F_CondBMeasureRq 3277 0x00CCD
F_SHC_DiaPresMeasureRq 3278 0x00CCE
F_SucPumpRefPressPlus100 3279 0x00CCF


76 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

F_SHC_CitFillMainLineCoolDownTo 3280 0x00CD0


F_HeaterRefPresetHeatTempMinus3DegC 3281 0x00CD1
F_SHC_HeatAndCitricRecircEvvaTo 3282 0x00CD2
F_SHC_HeatAndCitricRecircRivaTo 3283 0x00CD3
F_SHC_HeatAndCitricRecircZevaByvaTo 3284 0x00CD4
F_SHC_HeatAndCitricRecircCycleCount 3285 0x00CD5
F_SHC_ExtraRinseMainPathTo 3286 0x00CD6
F_SucPumpRefHeatDistrPressure 3287 0x00CD7
F_SHC_HeatDistrEvvaTo 3288 0x00CD8
F_SHC_HeatDistrRivaTo 3289 0x00CD9
F_SHC_HeatDistrZevaByvaTo 3290 0x00CDA
F_SHC_HeatDistrCycleCount 3291 0x00CDB
F_SHC_DilluteRemCitricInBLineTo 3292 0x00CDC
F_SHC_HeatDistrMeasDone 3293 0x00CDD
F_SHC_FlushChemValveTo 3294 0x00CDE
F_DelayNoFlow2To 3295 0x00CDF
F_ACondIsLower 3296 0x00CE0
F_LiquidFlush 3297 0x00CE1
F_SHC_RinseBefHeatSts 3298 0x00CE2
F_SHC_FillupSts 3299 0x00CE3
F_SHC_RecircSts 3300 0x00CE4
F_SHC_HeatDistrSts 3301 0x00CE5
F_SHC_FinalSts 3302 0x00CE6
F_SHC_NoInPress1 3303 0x00CE7
F_SHC_NoInPress2 3304 0x00CE8
F_SHC_NoInPress3 3305 0x00CE9
F_SHC_NoInPress4 3306 0x00CEA
F_SHC_NoInPress5 3307 0x00CEB
F_SHC_CondNotDetected 3308 0x00CEC
F_RequestRestartToP 3309 0x00CED
F_SHC_CitricDetectionStart 3310 0x00CEE
F_CCPrimAttnDisable 3311 0x00CEF
F_SucPumpRefDutyCycle12Perc 3312 0x00CF0
F_SucPumpRefDutyCycle3Perc 3313 0x00CF1
F_RivaFillup5sTo 3314 0x00CF2
F_RivaFillup5sCount 3315 0x00CF3
F_SHC_HeatDistrStartTo 3316 0x00CF4
F_SHC_HeatDistrStartCount 3317 0x00CF5
F_SHC_HeaterInefficientTo 3318 0x00CF6
F_SucPumpRefDutyCycle46Perc 3319 0x00CF7
F_SetDialysisFlowChange 3320 0x00CF8
F_HeatRecircSts 3321 0x00CF9
F_DialysisFlowChangeTarationRq 3322 0x00CFA


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 77 (381)
Program version 1.1.0
C EN 2015-03-26

F_SagsTestInitRecovery 3323 0x00CFB


F_ToggleFiva_15s 3324 0x00CFC
F_SHC_AdjustRegHeatVar 3325 0x00CFD
F_BasicOrLimitedFCHReceived 3326 0x00CFE
F_DisableUFAcc 3327 0x00CFF
F_SucPumpRefDutyCycle6Perc 3328 0x00D00
F_SucPumpRefDutyCycle70Perc 3329 0x00D01
F_MeasuredSucPumpRefDutyCycle 3330 0x00D02
F_ChangeBackRefTemp 3331 0x00D03
F_IncreaseRefTemp1 3332 0x00D04
F_IncreaseRefTemp2 3333 0x00D05
F_SucPumpLowFlowRiva 3334 0x00D06
F_SucPumpLowFlowZeva 3335 0x00D07
F_SucPumpDtclRampComp 3336 0x00D08
F_HalfSucPumpDtclRampComp 3337 0x00D09
F_SucPumpDtclRampStarted 3338 0x00D0A
F_StartMeasFlowPumpDtcl 3339 0x00D0B
F_DisHistNotRdy 3340 0x00D0C
F_LowPdDegassRq 3341 0x00D0D
F_LowPdEvaluate 3342 0x00D0E
F_LowPdDegassActive 3343 0x00D0F
FR_Dummy 3344 0x00D10
FR_AccCh1 3345 0x00D11
FR_Ch1Rate 3346 0x00D12
FR_CH1RateFilt 3347 0x00D13
FR_Ch2Rate 3348 0x00D14
FR_CalCh1RCoeff 3349 0x00D15
FR_CalCh1ROffset 3350 0x00D16
FR_CalCh2RCoeff 3351 0x00D17
FR_CalCh2ROffset 3352 0x00D18
FR_CalUFrTolerance 3353 0x00D19
FR_CalUFrOffset 3354 0x00D1A
FR_Ch1RCoeff 3355 0x00D1B
FR_Ch1ROffset 3356 0x00D1C
FR_Ch1RTarationVar 3357 0x00D1D
FR_Ch2RCoeff 3358 0x00D1E
FR_Ch2ROffset 3359 0x00D1F
FR_Ch2RTarationVar 3360 0x00D20
FR_FlCh1Cl 3361 0x00D21
FR_FlCh2Cl 3362 0x00D22
FR_FlDifCl 3363 0x00D23
FR_FlDifMean 3364 0x00D24
FR_FlCh1Mean 3365 0x00D25


78 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

FR_FlCh2Mean 3366 0x00D26


FR_MaxUFVol 3367 0x00D27
FR_UFLogCh1Offset 3368 0x00D28
FR_UFLogCh2Offset 3369 0x00D29
FR_UFLogUFRateTolerance 3370 0x00D2A
FR_UfRate 3371 0x00D2B
FR_UFrOffset 3372 0x00D2C
FR_UFRateFilt 3373 0x00D2D
FR_UFRateFiltCtrl 3374 0x00D2E
FR_UFRateSNFilt 3375 0x00D2F
FR_UFRateTarationVar 3376 0x00D30
FR_UFRTolerance 3377 0x00D31
FR_UFVol 3378 0x00D32
FR_UFRawVol 3379 0x00D33
FR_UFVolRecir 3380 0x00D34
FR_UFRawVolRecir 3381 0x00D35
FW_BiCartExtraFlushTo 3382 0x00D36
FW_CCFillTime 3383 0x00D37
FW_CCPrimTime 3384 0x00D38
FW_CWPIntegHeatFlushStartTime 3385 0x00D39
FW_DilFillCorrTime 3386 0x00D3A
FW_DilFillTime 3387 0x00D3B
FW_DrainExtra 3388 0x00D3C
FW_DrainTime 3389 0x00D3D
FW_HeatCitricFillTime 3390 0x00D3E
FW_ConcChemFillTime 3391 0x00D3F
FW_CentrChemFillTime 3392 0x00D40
FW_FlushDivaTavaTo 3393 0x00D41
FW_FlushZevaByvaTo 3394 0x00D42
FW_HeatCitricFlushStartTime 3395 0x00D43
FW_HeatCoolOff2To 3396 0x00D44
FW_HeatDisTestTime 3397 0x00D45
FW_HeatFlushStartTime 3398 0x00D46
FW_HeatFlushTime 3399 0x00D47
FW_HeatIncOutFlowTime 3400 0x00D48
FW_HeatRecirc1Time 3401 0x00D49
FW_MaxHeatCitricFlushStartTime 3402 0x00D4A
FW_MinChemRinseCycleTime 3403 0x00D4B
FW_RemChemDwellTime 3404 0x00D4C
FW_RemFillupTime 3405 0x00D4D
FW_RemHeatTime 3406 0x00D4E
FW_RemRinseTime 3407 0x00D4F
FW_RinseCycleReady 3408 0x00D50


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 79 (381)
Program version 1.1.0
C EN 2015-03-26

FW_RinseTavaTo 3409 0x00D51


FW_RinseZevaTo 3410 0x00D52
FW_UFDFlushTime 3411 0x00D53
FW_UFDFlushTime2 3412 0x00D54
FW_FchStepsTimer 3413 0x00D55
FW_FlowTimer 3414 0x00D56
FW_ServStepsTimer 3415 0x00D57
FW_DisinfStepsTimer 3416 0x00D58
FW_TreatmentStepsTimer 3417 0x00D59
FW_TimeInSecAfterStartOrRec 3418 0x00D5A
FW_TreatmentForceAdjTimer 3419 0x00D5B
FW_attnListNbr 3420 0x00D5C
FW_lastIssuedAttn 3421 0x00D5D
FW_ConvIntoIntOverFlow1 3422 0x00D5E
FW_ConvIntoIntOverFlow2 3423 0x00D5F
FW_ConvIntoIntOverFlow3 3424 0x00D60
FW_WRO300HRUNTime 3425 0x00D61
FW_WRO300HLFHTime 3426 0x00D62
FW_W300RemColdPrepTime 3427 0x00D63
FW_W300RemWarmPrepTime 3428 0x00D64
FW_W300RemWaterDistrTime 3429 0x00D65
FW_W300RemLowFlowHeatTime 3430 0x00D66
FW_RemHeatTime2 3431 0x00D67
FW_RemWROStartUpTime 3432 0x00D68
FW_DisplayRemFillupTime 3433 0x00D69
FW_DisplayRemChemDwellTime 3434 0x00D6A
FW_DisplayRemRinseTime 3435 0x00D6B
FW_DisplayRemHeatTime 3436 0x00D6C
FW_TreatmentForceAdjTimerSec 3437 0x00D6D
FW_GoodVarFlowFlTimer 3438 0x00D6E
FW_GoodVarFlowFlTimerSec 3439 0x00D6F
FW_InitINVAOpenDelayTime 3440 0x00D70
FW_ErrorCpuNumber 3441 0x00D71
FW_ErrorModClass 3442 0x00D72
FW_ErrorModNum 3443 0x00D73
FW_ErrorErrNum 3444 0x00D74
FW_EmptyChemLineRatioTo 3445 0x00D75
FW_FillupFinalFlushRatioTo 3446 0x00D76
FW_RinseToFlvaRatioTo 3447 0x00D77
FW_SHC_HeatAndCitricRecircCycleCount 3448 0x00D78
FW_SHC_HeatDistrCycleCount 3449 0x00D79
FW_SHC_ExtraRinseMainPathTo 3450 0x00D7A
FW_SHC_HeatDistrStartTo 3451 0x00D7B


80 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

BF_Dummy 3452 0x00D7C


BF_AutoPriming 3453 0x00D7D
BF_AutoRinse 3454 0x00D7E
BF_BlPumpControl 3455 0x00D7F
BF_BloodFlow 3456 0x00D80
BF_Clamps 3457 0x00D81
BF_BPM 3458 0x00D82
BF_BPMService 3459 0x00D83
BF_HeparinPump 3460 0x00D84
BF_SnTimeControl 3461 0x00D85
BF_StrokeVol 3462 0x00D86
BF_VenousPressure 3463 0x00D87
BF_Volume 3464 0x00D88
BF_AssistedPrimingTO 3465 0x00D89
BF_PressureTest 3466 0x00D8A
BF_AirDetector 3467 0x00D8B
BB_Dummy 3468 0x00D8C
BB_SnModeStart 3469 0x00D8D
BB_BPMIntervalMeasurementActive 3470 0x00D8E
BD_Dummy 3471 0x00D8F
BI_Dummy 3472 0x00D90
BI_AccExtraPrimVol 3473 0x00D91
BI_AccPrimVol 3474 0x00D92
BI_AccConPatVol 3475 0x00D93
BI_AccRinseBackVol 3476 0x00D94
BI_ApBlPumpFillFl 3477 0x00D95
BI_ApBlPumpPrimFl 3478 0x00D96
BI_ApBlPumpSetFlow 3479 0x00D97
BI_ApDisplayTime 3480 0x00D98
BI_ApTotRemTime 3481 0x00D99
BI_ApFillTime 3482 0x00D9A
BI_ApFlush1time 3483 0x00D9B
BI_ApFlush2time 3484 0x00D9C
BI_ApLevelAdjTime 3485 0x00D9D
BI_ApPrBuildTime 3486 0x00D9E
BI_ApPrTestTime 3487 0x00D9F
BI_ApPrimTime 3488 0x00DA0
BI_ApTubeFillTime 3489 0x00DA1
BI_ApUFRateSet 3490 0x00DA2
BI_ApVenClosedTime 3491 0x00DA3
BI_ApVenOpenTime 3492 0x00DA4
BI_BloodAccVol 3493 0x00DA5
BI_BloodFl 3494 0x00DA6


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 81 (381)
Program version 1.1.0
C EN 2015-03-26

BI_BloodFlActSet 3495 0x00DA7


BI_BloodPumpRPM 3496 0x00DA8
BI_BPMInitCuffPr 3497 0x00DA9
BI_DiagHepEstAcc 3498 0x00DAA
BI_DiagHepRealAcc 3499 0x00DAB
BI_EstAcc 3500 0x00DAC
BI_EstMicroLitre 3501 0x00DAD
BI_ArtPr 3502 0x00DAE
BI_MeanBloodFl 3503 0x00DAF
BI_MeasVol 3504 0x00DB0
BI_MonitorTemp 3505 0x00DB1
BI_PrimDetCurrent 3506 0x00DB2
BI_RinseBlPumpSetFlow 3507 0x00DB3
BI_RinseVenClosedTime 3508 0x00DB4
BI_RinseVenOpenTime 3509 0x00DB5
BI_SnActArtPhTime 3510 0x00DB6
BI_SnActVenPhTime 3511 0x00DB7
BI_SnStrokeVol 3512 0x00DB8
BI_TimeTooLowFlow 3513 0x00DB9
BI_VenPr 3514 0x00DBA
BI_VenPrCoeff 3515 0x00DBB
BI_VenPrInfoDisp 3516 0x00DBC
BI_VenPrOffset 3517 0x00DBD
BI_TotalPrimSetVol 3518 0x00DBE
BI_AccExtraRbVol 3519 0x00DBF
BI_HighVprLowUF 3520 0x00DC0
BI_HighVprLowUFSN 3521 0x00DC1
BI_HepAccTotVol 3522 0x00DC2
BI_HeparinBolusRate 3523 0x00DC3
BI_HeparinBolusVolumeSet 3524 0x00DC4
BI_HepTotAccBolusVol 3525 0x00DC5
BI_HeparinBolusMotorCtrl 3526 0x00DC6
BI_EstBolusMicroLitre 3527 0x00DC7
BI_EstBolusAcc 3528 0x00DC8
BI_EstTmtAcc 3529 0x00DC9
BI_AccBolusVolFromP 3530 0x00DCA
BI_HepMeasVolAcc 3531 0x00DCB
BI_DeltaBolusCnt 3532 0x00DCC
BI_HepMeasBolusAcc 3533 0x00DCD
BI_TickBolusAcc 3534 0x00DCE
BI_AccTotBolusVolFromP 3535 0x00DCF
BI_HepMeasVolAccFromP 3536 0x00DD0
BI_DeltaCnt 3537 0x00DD1


82 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

BI_TickAcc 3538 0x00DD2


BI_EstTotBolusAcc 3539 0x00DD3
BI_AccADVol 3540 0x00DD4
BI_ArtPrProtFilt 3541 0x00DD5
BI_BloodFlNom 3542 0x00DD6
BI_BLoodACCQBLastRun 3543 0x00DD7
BI_TargetPrimVol 3544 0x00DD8
B_Dummy 3545 0x00DD9
B_AccBloodVolZeroed 3546 0x00DDA
B_AccPrimVolZeroed 3547 0x00DDB
B_ApActive 3548 0x00DDC
B_ApDiaClosedRq 3549 0x00DDD
B_ApDialysisFluidAvail 3550 0x00DDE
B_ApEarlyDiaFlStart 3551 0x00DDF
B_ApFillRdy 3552 0x00DE0
B_ApFillTubeRdy 3553 0x00DE1
B_ApFluidDirectRq 3554 0x00DE2
B_ApFluidTubeAttn 3555 0x00DE3
B_ApFlush1FcnRdy 3556 0x00DE4
B_ApFlush1Rdy 3557 0x00DE5
B_ApFlush2Act 3558 0x00DE6
B_ApFlush2FcnRdy 3559 0x00DE7
B_ApInitRdy 3560 0x00DE8
B_ApLevelAdjRdy 3561 0x00DE9
B_ApLevelAdjustActive 3562 0x00DEA
B_ApPrBuildAttn 3563 0x00DEB
B_ApPrHiLim 3564 0x00DEC
B_ApPrimFcnRdy 3565 0x00DED
B_ApPrimRdy 3566 0x00DEE
B_ApPrTestAct 3567 0x00DEF
B_ApPrTestAttn 3568 0x00DF0
B_ApPrTestRdy 3569 0x00DF1
B_ApVenClRdy 3570 0x00DF2
B_ApWaitForFlDirectTo 3571 0x00DF3
B_ApWaitForFmAttnRq 3572 0x00DF4
B_ApWaitForPressTestDispRq 3573 0x00DF5
B_ApVenOpRdy 3574 0x00DF6
B_ArbPrHiLim 3575 0x00DF7
B_ArtClampClosed 3576 0x00DF8
B_VenArtClampTestRdy 3577 0x00DF9
B_ArtPhPrLimit 3578 0x00DFA
B_ArtPhTimeOut 3579 0x00DFB
B_AutoPrimRdy2 3580 0x00DFC


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 83 (381)
Program version 1.1.0
C EN 2015-03-26

B_AutoRinseInitRdy 3581 0x00DFD


B_BlPumpActive 3582 0x00DFE
B_BlPumpFreqSet 3583 0x00DFF
B_BlPumpOverload 3584 0x00E00
B_BlPumpOvldAlarm 3585 0x00E01
B_BlPumpSts 3586 0x00E02
B_BlPumpTestStart 3587 0x00E03
B_BlPumpTestStop 3588 0x00E04
B_BmFuncTestRdy 3589 0x00E05
B_BPMStartSingleMeasurement 3590 0x00E06
B_BPMStopSingleMeasurement 3591 0x00E07
B_BPMStartIntervalMeasurement 3592 0x00E08
B_BPMMeasResultRdy 3593 0x00E09
B_BPMActivating 3594 0x00E0A
B_BPMDeactivating 3595 0x00E0B
B_BPMActivationFailed 3596 0x00E0C
B_BPMDeactivationFailed 3597 0x00E0D
B_BPMFunctionCheckCompleted 3598 0x00E0E
B_BPMIsActive 3599 0x00E0F
B_BPMIsEnabled 3600 0x00E10
B_BPMEnabledDisabledReceived 3601 0x00E11
B_BPMDeflatedTimeExpired 3602 0x00E12
B_BPMInMeasurementDoneState 3603 0x00E13
B_BPMInMeasurementOngoingState 3604 0x00E14
B_BPMInIdleState 3605 0x00E15
B_BPMInWaitingForEnableState 3606 0x00E16
B_BPMAirLeakTestStart 3607 0x00E17
B_BPMAirLeakTestRdy 3608 0x00E18
B_BPMInflationTestStart 3609 0x00E19
B_BPMInflationTestRdy 3610 0x00E1A
B_BPMPressureTestStart 3611 0x00E1B
B_BPMPressureTestRdy 3612 0x00E1C
B_BPMTestFailed 3613 0x00E1D
B_BPMServiceActive 3614 0x00E1E
B_BPMAlarmActive 3615 0x00E1F
B_BPMAlarmCleared 3616 0x00E20
B_BPMErrorFromIO 3617 0x00E21
B_BPMPressTooHigh 3618 0x00E22
B_BPMPulseTooLow 3619 0x00E23
B_BPMNoCuff 3620 0x00E24
B_BPMNoMeasurement 3621 0x00E25
B_BPMBodyMove 3622 0x00E26
B_BPMError 3623 0x00E27


84 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

B_BPMReleasePressureTo 3624 0x00E28


B_ExtraPrimVolAch 3625 0x00E29
B_HepPumpOvldAlarm 3626 0x00E2A
B_HepPumpSts 3627 0x00E2B
B_HepPumpTestOvld 3628 0x00E2C
B_KickStartBtrc 3629 0x00E2D
B_MeanFlowCalcRdy 3630 0x00E2E
B_MeanFlowInitRdy 3631 0x00E2F
B_PrimVolAch 3632 0x00E30
B_RdyToShutDown 3633 0x00E31
B_RegHighVprAch 3634 0x00E32
B_RinseBackRdy 3635 0x00E33
B_RinseBackVolAch 3636 0x00E34
B_RinseVenClRdy 3637 0x00E35
B_RinseVenOpRdy 3638 0x00E36
B_SnAlarmPhase 3639 0x00E37
B_SnArtPhase 3640 0x00E38
B_SnVenPhase 3641 0x00E39
B_StrokeVolAlarm 3642 0x00E3A
B_TempTimerTo 3643 0x00E3B
B_TooLowBloodFlow 3644 0x00E3C
B_VenClampClosed 3645 0x00E3D
B_ClampCloseOpenFailure 3646 0x00E3E
B_VclampStart 3647 0x00E3F
B_AclampStart 3648 0x00E40
B_ConnectPatVolAch 3649 0x00E41
B_VenClampFirstTime 3650 0x00E42
B_VenPhPrLimit 3651 0x00E43
B_VenPhTimeOut 3652 0x00E44
B_VenPrAlH 3653 0x00E45
B_SNActive 3654 0x00E46
B_ApFluidBypassRq 3655 0x00E47
B_ResetPrimVolRq 3656 0x00E48
B_ExtraRbVolAch 3657 0x00E49
B_RinseBackVolReset 3658 0x00E4A
B_BloodCircuitAlarm 3659 0x00E4B
B_ApPressureTest 3660 0x00E4C
B_ApWaitForFlush 3661 0x00E4D
B_HighVprLowUF 3662 0x00E4E
B_HeparinBolusActive 3663 0x00E4F
B_HepPumpBolusStopTo 3664 0x00E50
B_PrimingActive 3665 0x00E51
B_RinsebackActive 3666 0x00E52


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 85 (381)
Program version 1.1.0
C EN 2015-03-26

B_DNActive 3667 0x00E53


B_BlPumpOverloadSignal 3668 0x00E54
BR_Dummy 3669 0x00E55
BW_Dummy 3670 0x00E56
BW_PassedBolusTime 3671 0x00E57
BW_PassedTime 3672 0x00E58
EEB_AcUserLblAlt1 3673 0x00E59
EEB_ChemBankNr 3674 0x00E5A
EEB_HeaterEfficiency 3675 0x00E5B
EEB_LogDecade 3676 0x00E5C
EEB_LogFieldId 3677 0x00E5D
EEB_LogFormat0 3678 0x00E5E
EEB_LogFormat1 3679 0x00E5F
EEB_LogFormat2 3680 0x00E60
EEB_LogSel0 3681 0x00E61
EEB_LogSel1 3682 0x00E62
EEB_LogSel2 3683 0x00E63
EEB_LogType0 3684 0x00E64
EEB_LogType1 3685 0x00E65
EEB_LogType2 3686 0x00E66
EEB_ChemDisAlt 3687 0x00E67
EEB_HeatDisAlt 3688 0x00E68
EEB_ProfModel1NaAcProfType 3689 0x00E69
EEB_ProfModel1NaProfType 3690 0x00E6A
EEB_ProfModel1HCO3ProfType 3691 0x00E6B
EEB_ProfModel1UFProfType 3692 0x00E6C
EEB_ProfModel2NaAcProfType 3693 0x00E6D
EEB_ProfModel2NaProfType 3694 0x00E6E
EEB_ProfModel2HCO3ProfType 3695 0x00E6F
EEB_ProfModel2UFProfType 3696 0x00E70
EEB_ProfModel3NaAcProfType 3697 0x00E71
EEB_ProfModel3NaProfType 3698 0x00E72
EEB_ProfModel3HCO3ProfType 3699 0x00E73
EEB_ProfModel3UFProfType 3700 0x00E74
EEB_ActiveDisinfRinseAlt 3701 0x00E75
EEI_ACBPumpSpeedWindow 3702 0x00E76
EEI_ACBPUMPSPEEDWINDOWHLLVL1 3703 0x00E77
EEI_ACBPUMPSPEEDWINDOWHLLVL2 3704 0x00E78
EEI_ACBPUMPSPEEDWINDOWHLLVL3 3705 0x00E79
EEI_AcetateCoeffAlt1 3706 0x00E7A
EEI_AcetateCoeffAlt2 3707 0x00E7B
EEI_AcetateCoeffAlt3 3708 0x00E7C
EEI_AcetateOffsetAlt1 3709 0x00E7D


86 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEI_AcetateOffsetAlt2 3710 0x00E7E


EEI_AcetateOffsetAlt3 3711 0x00E7F
EEI_AcMolCoeffAlt1 3712 0x00E80
EEI_AcMolCoeffAlt2 3713 0x00E81
EEI_AcMolCoeffAlt3 3714 0x00E82
EEI_AcMolOffsetAlt1 3715 0x00E83
EEI_AcMolOffsetAlt2 3716 0x00E84
EEI_AcMolOffsetAlt3 3717 0x00E85
EEI_AdRefCoeff 3718 0x00E86
EEI_BCAPumpSpeedWindow 3719 0x00E87
EEI_BCAPumpSpeedWindowHLLvl1 3720 0x00E88
EEI_BCAPUMPSPEEDWINDOWHLLVL2 3721 0x00E89
EEI_BCAPUMPSPEEDWINDOWHLLVL3 3722 0x00E8A
EEI_BCBPumpSpeedWindow 3723 0x00E8B
EEI_BCBPumpSpeedWindowHLLvl1 3724 0x00E8C
EEI_BCBPUMPSPEEDWINDOWHLLVL2 3725 0x00E8D
EEI_BCBPUMPSPEEDWINDOWHLLVL3 3726 0x00E8E
EEI_BicarbC1Alt1 3727 0x00E8F
EEI_BicarbC1Alt2 3728 0x00E90
EEI_BicarbC1Alt3 3729 0x00E91
EEI_BicarbC2Alt1 3730 0x00E92
EEI_BicarbC2Alt2 3731 0x00E93
EEI_BicarbC2Alt3 3732 0x00E94
EEI_BicarbC3Alt1 3733 0x00E95
EEI_BicarbC3Alt2 3734 0x00E96
EEI_BicarbC3Alt3 3735 0x00E97
EEI_BicarbC4Alt1 3736 0x00E98
EEI_BicarbC4Alt2 3737 0x00E99
EEI_BicarbC4Alt3 3738 0x00E9A
EEI_BicarbC5Alt1 3739 0x00E9B
EEI_BicarbC5Alt2 3740 0x00E9C
EEI_BicarbC5Alt3 3741 0x00E9D
EEI_BicarbC6Alt1 3742 0x00E9E
EEI_BicarbC6Alt2 3743 0x00E9F
EEI_BicarbC6Alt3 3744 0x00EA0
EEI_BicarbC7Alt1 3745 0x00EA1
EEI_BicarbC7Alt2 3746 0x00EA2
EEI_BicarbC7Alt3 3747 0x00EA3
EEI_BicarbC8Alt1 3748 0x00EA4
EEI_BicarbC8Alt2 3749 0x00EA5
EEI_BicarbC8Alt3 3750 0x00EA6
EEI_BicarbC9Alt1 3751 0x00EA7
EEI_BicarbC9Alt2 3752 0x00EA8


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 87 (381)
Program version 1.1.0
C EN 2015-03-26

EEI_BicarbC9Alt3 3753 0x00EA9


EEI_BicarbC10Alt1 3754 0x00EAA
EEI_BicarbC10Alt2 3755 0x00EAB
EEI_BicarbC10Alt3 3756 0x00EAC
EEI_BicarbC11Alt1 3757 0x00EAD
EEI_BicarbC11Alt2 3758 0x00EAE
EEI_BicarbC11Alt3 3759 0x00EAF
EEI_BicarbC12Alt1 3760 0x00EB0
EEI_BicarbC12Alt2 3761 0x00EB1
EEI_BicarbC12Alt3 3762 0x00EB2
EEI_BlFlLowLimSupFlow 3763 0x00EB3
EEI_BlFlLowLimSupTime 3764 0x00EB4
EEI_BloodFlRampCoeff 3765 0x00EB5
EEI_BloodFlSetHLLvl1 3766 0x00EB6
EEI_BloodFlSetHlLvl2 3767 0x00EB7
EEI_BloodFlSetHlLvl3 3768 0x00EB8
EEI_BloodFlSetInit 3769 0x00EB9
EEI_BloodFlSetLLLvl1 3770 0x00EBA
EEI_BloodFlSetLlLvl2 3771 0x00EBB
EEI_BloodFlSetLlLvl3 3772 0x00EBC
EEI_BlPumpBypassSetFlow 3773 0x00EBD
EEI_BlPumpBypassSetFlowHLLvl1 3774 0x00EBE
EEI_BLPUMPBYPASSSETFLOWHLLVL2 3775 0x00EBF
EEI_BLPUMPBYPASSSETFLOWHLLVL3 3776 0x00EC0
EEI_BlTubeDiam0 3777 0x00EC1
EEI_BlTubeDiam1 3778 0x00EC2
EEI_BlTubeDiam2 3779 0x00EC3
EEI_ChemMixRatioAlt1 3780 0x00EC4
EEI_ChemMixRatioAlt2 3781 0x00EC5
EEI_ChemMixRatioAlt3 3782 0x00EC6
EEI_ChemMixRatioAlt6 3783 0x00EC7
EEI_ChemMixRatioAlt7 3784 0x00EC8
EEI_ChemRinseTempAlt1 3785 0x00EC9
EEI_ChemRinseTempAlt2 3786 0x00ECA
EEI_ChemRinseTempAlt3 3787 0x00ECB
EEI_ChemRinseTempAlt6 3788 0x00ECC
EEI_ChemRinseTempAlt7 3789 0x00ECD
EEI_DegassPr 3790 0x00ECE
EEI_EstTempDrop 3791 0x00ECF
EEI_HCO3SetOff 3792 0x00ED0
EEI_HepSetHlLvl1 3793 0x00ED1
EEI_HepSetInit 3794 0x00ED2
EEI_HepSetLlLvl1 3795 0x00ED3


88 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEI_HepStopLimHLLvl1 3796 0x00ED4


EEI_HepStopLimHLLvl2 3797 0x00ED5
EEI_HepStopLimHlLvl3 3798 0x00ED6
EEI_HepStopLimInit 3799 0x00ED7
EEI_HepStopLimLLLvl1 3800 0x00ED8
EEI_HepStopLimLLLvl2 3801 0x00ED9
EEI_HepStopLimLlLvl3 3802 0x00EDA
EEI_HeatingTemp 3803 0x00EDB
EEI_HeatingTempMM90 3804 0x00EDC
EEI_HepSyringeDiam 3805 0x00EDD
EEI_MinUfRateHLLvl1 3806 0x00EDE
EEI_MINUFRATEHLLVL2 3807 0x00EDF
EEI_MinUfRateHlLvl3 3808 0x00EE0
EEI_MinUfRateInit 3809 0x00EE1
EEI_MinUfRateLLLvl1 3810 0x00EE2
EEI_MINUFRATELLLVL2 3811 0x00EE3
EEI_MinUfRateLlLvl3 3812 0x00EE4
EEI_NaAcSetOff 3813 0x00EE5
EEI_NaSetOff 3814 0x00EE6
EEI_PrimUfRate 3815 0x00EE7
EEI_PrimUfRateHLLvl1 3816 0x00EE8
EEI_PRIMUFRATEHLLVL2 3817 0x00EE9
EEI_PrimUfRateHlLvl3 3818 0x00EEA
EEI_PrimUfRateLLLvl1 3819 0x00EEB
EEI_PRIMUFRATELLLVL2 3820 0x00EEC
EEI_PrimUfRateLlLvl3 3821 0x00EED
EEI_RbFlowSet 3822 0x00EEE
EEI_RbVolSet 3823 0x00EEF
EEI_RemainTimeDiffInit 3824 0x00EF0
EEI_RemainTimeHlLvl1 3825 0x00EF1
EEI_REMAINTIMEHLLVL2 3826 0x00EF2
EEI_RemainTimeHlLvl3 3827 0x00EF3
EEI_RemainTimeIsolInit 3828 0x00EF4
EEI_RemainTimeLlLvl1 3829 0x00EF5
EEI_REMAINTIMELLLVL2 3830 0x00EF6
EEI_RemainTimeLlLvl3 3831 0x00EF7
EEI_RinseCycleTime 3832 0x00EF8
EEI_RinseTimeMM90HLLvl1 3833 0x00EF9
EEI_RinseTimeMm90HlLvl3 3834 0x00EFA
EEI_RinseTimeMM90Init 3835 0x00EFB
EEI_RinseTimeMM90LLLvl1 3836 0x00EFC
EEI_RINSETIMEMM90LLLVL2 3837 0x00EFD
EEI_RinseTimeMm90LlLvl3 3838 0x00EFE


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 89 (381)
Program version 1.1.0
C EN 2015-03-26

EEI_ServTempHLInit 3839 0x00EFF


EEI_ServTempLLInit 3840 0x00F00
EEI_SnArtTimeHLLvl1 3841 0x00F01
EEI_SnArtTimeHLLvl2 3842 0x00F02
EEI_SnArtTimeHlLvl3 3843 0x00F03
EEI_SnArtTimeLLLvl1 3844 0x00F04
EEI_SnArtTimeLLLvl2 3845 0x00F05
EEI_SnArtTimeLlLvl3 3846 0x00F06
EEI_SnArtTimeSetInit 3847 0x00F07
EEI_SnVenTimeHLLvl1 3848 0x00F08
EEI_SNVENTIMEHLLVL2 3849 0x00F09
EEI_SnVenTimeHlLvl3 3850 0x00F0A
EEI_SnVenTimeLLLvl1 3851 0x00F0B
EEI_SNVENTIMELLLVL2 3852 0x00F0C
EEI_SnVenTimeLlLvl3 3853 0x00F0D
EEI_SnVenTimeSetInit 3854 0x00F0E
EEI_TempHLLvl1 3855 0x00F0F
EEI_TempLLLvl1 3856 0x00F10
EEI_TempSet 3857 0x00F11
EEI_TimeNowYear 3858 0x00F12
EEI_UFRateHLCoeff 3859 0x00F13
EEI_UFRATEHLCOEFFHLLVL1 3860 0x00F14
EEI_UFRATEHLCOEFFHLLVL2 3861 0x00F15
EEI_UFRATEHLCOEFFHLLVL3 3862 0x00F16
EEI_UfRateHlLvl1 3863 0x00F17
EEI_UFRATEHLLVL2 3864 0x00F18
EEI_UfRateHlLvl3 3865 0x00F19
EEI_UfRateLlLvl1 3866 0x00F1A
EEI_UFRATELLLVL2 3867 0x00F1B
EEI_UfRateLlLvl3 3868 0x00F1C
EEI_UfRateSetDiffInit 3869 0x00F1D
EEI_UfRateSetIsolInit 3870 0x00F1E
EEI_UfVolSetDiffInit 3871 0x00F1F
EEI_UfVolSetHlLvl1 3872 0x00F20
EEI_UFVOLSETHLLVL2 3873 0x00F21
EEI_UfVolSetManUpperLimit 3874 0x00F22
EEI_UfVolSetIsolInit 3875 0x00F23
EEI_UfVolSetLlLvl1 3876 0x00F24
EEI_UFVOLSETLLLVL2 3877 0x00F25
EEI_StrokeVolLim 3878 0x00F26
EEI_ACCCLEANTIME 3879 0x00F27
EEI_ApBlPumpFillFlAlt1 3880 0x00F28
EEI_ApBlPumpPrimFlAlt1 3881 0x00F29


90 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEI_ApFillTimeAlt1 3882 0x00F2A


EEI_ApFlush1TimeAlt1 3883 0x00F2B
EEI_ApFlush2TimeAlt1 3884 0x00F2C
EEI_ApLevelAdjTimeAlt1 3885 0x00F2D
EEI_ApMaxPrDropAlt1 3886 0x00F2E
EEI_ApPrBuildTimeAlt1 3887 0x00F2F
EEI_ApPrimCycleTimeAlt1 3888 0x00F30
EEI_ApPrimTimeAlt1 3889 0x00F31
EEI_ApPrTestTimeAlt1 3890 0x00F32
EEI_ApTubeTimeAlt1 3891 0x00F33
EEI_ApUFRateSetAlt1 3892 0x00F34
EEI_ChemCondLimLLAlt1 3893 0x00F35
EEI_ChemCondLimHLAlt1 3894 0x00F36
EEI_ChemCondLimLLAlt2 3895 0x00F37
EEI_ChemCondLimHLAlt2 3896 0x00F38
EEI_ChemCondLimLLAlt3 3897 0x00F39
EEI_ChemCondLimHLAlt3 3898 0x00F3A
EEI_ChemCondLimLLAlt5 3899 0x00F3B
EEI_ChemCondLimHLAlt5 3900 0x00F3C
EEI_ChemCondLimLLAlt6 3901 0x00F3D
EEI_ChemCondLimHLAlt6 3902 0x00F3E
EEI_ChemCondLimLLAlt7 3903 0x00F3F
EEI_ChemCondLimHLAlt7 3904 0x00F40
EEI_MaxUfRate 3905 0x00F41
EEI_TimeDisinfMin 3906 0x00F42
EEI_TimeDisinfHour 3907 0x00F43
EEI_TimeDisinfDayOfMonth 3908 0x00F44
EEI_TimeDisinfMonth 3909 0x00F45
EEI_TimeDisinfYear 3910 0x00F46
EEI_TimeBetweenCleaning 3911 0x00F47
EEI_BlPumpPCoeffVprTst 3912 0x00F48
EEI_HPGMaxLimit 3913 0x00F49
EEI_PumpSpeedRatioWindow 3914 0x00F4A
EEI_PumpSpeedRatioWindowHLLvl1 3915 0x00F4B
EEI_PumpSpeedRatioWindowHLLVL2 3916 0x00F4C
EEI_PumpSpeedRatioWindowHLLVL3 3917 0x00F4D
EEI_BattFailedCnt 3918 0x00F4E
EEI_LowDialysisFlow 3919 0x00F4F
EEI_TempMaxHL 3920 0x00F50
EEI_TempMinLL 3921 0x00F51
EEI_MinUfRateHL 3922 0x00F52
EEI_MinUFRateLL 3923 0x00F53
EEI_BloodFlHighLim 3924 0x00F54


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 91 (381)
Program version 1.1.0
C EN 2015-03-26

EEI_BPMDiastPrHigh 3925 0x00F55


EEI_BPMDiastPrLow 3926 0x00F56
EEI_BPMInitCuffPr 3927 0x00F57
EEI_BPMMeanPrHigh 3928 0x00F58
EEI_BPMMeanPrLow 3929 0x00F59
EEI_BPMMeasInt 3930 0x00F5A
EEI_BPMPulseHigh 3931 0x00F5B
EEI_BPMPulseLow 3932 0x00F5C
EEI_BPMSystPrHigh 3933 0x00F5D
EEI_BPMSystPrLow 3934 0x00F5E
EEI_ConcPumpSpeedFlushRec 3935 0x00F5F
EEI_FlushRecHPGPress 3936 0x00F60
EEI_ProfModel1UFRateDiff 3937 0x00F61
EEI_ProfModel2UFRateDiff 3938 0x00F62
EEI_ProfModel3UFRateDiff 3939 0x00F63
EEI_OvernightRinseDelay 3940 0x00F64
EEI_FlowSteps 3941 0x00F65
EEI_SetFlowLLLvl1 3942 0x00F66
EEI_SetFlowLLLvl2 3943 0x00F67
EEI_SetFlowLLLvl3 3944 0x00F68
EEI_SetFlowLL 3945 0x00F69
EEI_SetFlow 3946 0x00F6A
EEI_SetFlowHL 3947 0x00F6B
EEI_SetFlowHLLvl3 3948 0x00F6C
EEI_SetFlowHLLvl2 3949 0x00F6D
EEI_SetFlowHLLvl1 3950 0x00F6E
EEI_ChemConcPercentAlt1 3951 0x00F6F
EEI_ChemConcPercentAlt2 3952 0x00F70
EEI_ChemConcPercentAlt3 3953 0x00F71
EEI_ChemConcPercentAlt6 3954 0x00F72
EEI_ConcStandbyFch 3955 0x00F73
EEI_ConcStandbyPrim 3956 0x00F74
EEI_ProfModel1UFIntervals 3957 0x00F75
EEI_ProfModel2UFIntervals 3958 0x00F76
EEI_ProfModel3UFIntervals 3959 0x00F77
EEI_VarFlowHistReplaceIndex 3960 0x00F78
EEI_VarFlowHistVal0 3961 0x00F79
EEI_VarFlowHistVal1 3962 0x00F7A
EEI_VarFlowHistVal2 3963 0x00F7B
EEI_VarFlowHistVal3 3964 0x00F7C
EEI_VarFlowHistVal4 3965 0x00F7D
EEI_DiascanDistrVolume 3966 0x00F7E
EEI_DiascanKAlarmLimit 3967 0x00F7F


92 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEI_DiascanMeasMode 3968 0x00F80


EEI_DiascanMinCdi 3969 0x00F81
EEI_DiascanStepDuration 3970 0x00F82
EEI_DiascanStepHeight 3971 0x00F83
EEI_DiascanTarget 3972 0x00F84
EEI_DiascanTargetAlarm 3973 0x00F85
EEI_LowFlowHeatPress 3974 0x00F86
EEI_InstRTCType 3975 0x00F87
EEI_ExtraRbVolSet 3976 0x00F88
EEI_RecirBloodFlSetInit 3977 0x00F89
EEI_PrimVolLimitSS 3978 0x00F8A
EEI_PrimVolLimitBS 3979 0x00F8B
EEI_RecirTimeInit 3980 0x00F8C
EEI_RecirUFRInit 3981 0x00F8D
EEI_RecirUfVolSetInit 3982 0x00F8E
EEI_ExtraPrimVolLimitBS 3983 0x00F8F
EEI_ExtraPrimVolLimitSS 3984 0x00F90
EEI_ConnectPatVolLimit 3985 0x00F91
EEI_RecirDiaFlow 3986 0x00F92
EEI_HPGMaxDisinfLimit 3987 0x00F93
EEI_DaysBetweenUfdReplaceLim 3988 0x00F94
EEI_DisinfCountUfdReplaceLim 3989 0x00F95
EEI_HypoCountUfdReplaceLim 3990 0x00F96
EEI_TimeUfdReplaceMin 3991 0x00F97
EEI_TimeUfdReplaceHour 3992 0x00F98
EEI_TimeUfdReplaceDayOfMonth 3993 0x00F99
EEI_TimeUfdReplaceMonth 3994 0x00F9A
EEI_TimeUfdReplaceYear 3995 0x00F9B
EEI_DisinfCountUfdReplace 3996 0x00F9C
EEI_HypoCountUfdReplace 3997 0x00F9D
EEI_RinseClampOpenTime 3998 0x00F9E
EEI_RinseClampCloseTime 3999 0x00F9F
EEI_ZevaPressReleaseLim 4000 0x00FA0
EEI_ExtraRecirUfVolSetInit 4001 0x00FA1
EEI_ExtraRecirTimeInit 4002 0x00FA2
EEI_ExtraRecirUFRInit 4003 0x00FA3
EEI_NoOfPassedFch 4004 0x00FA4
EEI_NoOfFirstTarFailed 4005 0x00FA5
EEI_TreatmentOverview 4006 0x00FA6
EEI_AccDecalcCounterBiCart 4007 0x00FA7
EEI_AccDecalcCounterLiqBic 4008 0x00FA8
EEI_TreatmBetweenDecalcBiCart 4009 0x00FA9
EEI_TreatmBetweenDecalcLiqBic 4010 0x00FAA


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 93 (381)
Program version 1.1.0
C EN 2015-03-26

EEI_AirDetOvrTimeTmt 4011 0x00FAB


EEI_AirDetOvrTimePrim 4012 0x00FAC
EEI_BcHco3Init 4013 0x00FAD
EEI_BcHco3LL 4014 0x00FAE
EEI_BcHco3HL 4015 0x00FAF
EEI_BcNaInit 4016 0x00FB0
EEI_BcNaLL 4017 0x00FB1
EEI_BcNaHL 4018 0x00FB2
EEI_BcConcAlt 4019 0x00FB3
EEI_ProfModel1NaStart 4020 0x00FB4
EEI_ProfModel2NaStart 4021 0x00FB5
EEI_ProfModel3NaStart 4022 0x00FB6
EEI_ProfModel1NaStartMin 4023 0x00FB7
EEI_ProfModel2NaStartMin 4024 0x00FB8
EEI_ProfModel3NaStartMin 4025 0x00FB9
EEI_ProfModel1NaStartMax 4026 0x00FBA
EEI_ProfModel2NaStartMax 4027 0x00FBB
EEI_ProfModel3NaStartMax 4028 0x00FBC
EEI_ProfModel1NaStop 4029 0x00FBD
EEI_ProfModel2NaStop 4030 0x00FBE
EEI_ProfModel3NaStop 4031 0x00FBF
EEI_ProfModel1NaStopMin 4032 0x00FC0
EEI_ProfModel2NaStopMin 4033 0x00FC1
EEI_ProfModel3NaStopMin 4034 0x00FC2
EEI_ProfModel1NaStopMax 4035 0x00FC3
EEI_ProfModel2NaStopMax 4036 0x00FC4
EEI_ProfModel3NaStopMax 4037 0x00FC5
EEI_ProfModel1HCO3Start 4038 0x00FC6
EEI_ProfModel2HCO3Start 4039 0x00FC7
EEI_ProfModel3HCO3Start 4040 0x00FC8
EEI_ProfModel1HCO3StartMin 4041 0x00FC9
EEI_ProfModel2HCO3StartMin 4042 0x00FCA
EEI_ProfModel3HCO3StartMin 4043 0x00FCB
EEI_ProfModel1HCO3StartMax 4044 0x00FCC
EEI_ProfModel2HCO3StartMax 4045 0x00FCD
EEI_ProfModel3HCO3StartMax 4046 0x00FCE
EEI_ProfModel1HCO3Stop 4047 0x00FCF
EEI_ProfModel2HCO3Stop 4048 0x00FD0
EEI_ProfModel3HCO3Stop 4049 0x00FD1
EEI_ProfModel1HCO3StopMin 4050 0x00FD2
EEI_ProfModel2HCO3StopMin 4051 0x00FD3
EEI_ProfModel3HCO3StopMin 4052 0x00FD4
EEI_ProfModel1HCO3StopMax 4053 0x00FD5


94 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEI_ProfModel2HCO3StopMax 4054 0x00FD6


EEI_ProfModel3HCO3StopMax 4055 0x00FD7
EEI_UfVolSetIsolManUpperLimit 4056 0x00FD8
EEI_MinPrimingVol 4057 0x00FD9
EEI_LowPdDegassPrOffset 4058 0x00FDA
EEI_NegDegassPrDiffError 4059 0x00FDB
EEI_LowPdActivationLimit 4060 0x00FDC
EEI_LowPdDeactivationLimit 4061 0x00FDD
EEI_LowPdActivationTime 4062 0x00FDE
EEI_LowPdDeactivationTime 4063 0x00FDF
EEL_Dummy 4064 0x00FE0
EEL_ApDialysisFluidAvail 4065 0x00FE1
EEL_BicartAlt1 4066 0x00FE2
EEL_BicartAlt2 4067 0x00FE3
EEL_BicartAlt3 4068 0x00FE4
EEL_BicHolderInstalled 4069 0x00FE5
EEL_ChemDisSts 4070 0x00FE6
EEL_ClinAutComSupAvail 4071 0x00FE7
EEL_DisIsConcAlt1 4072 0x00FE8
EEL_DisIsConcAlt2 4073 0x00FE9
EEL_DisIsConcAlt3 4074 0x00FEA
EEL_DisIsConcAlt6 4075 0x00FEB
EEL_DisIsConcAlt7 4076 0x00FEC
EEL_HepPumpAvail 4077 0x00FED
EEL_ImmediateFmStart 4078 0x00FEE
EEL_LogParDispInh 4079 0x00FEF
EEL_MM90Installed 4080 0x00FF0
EEL_RbFluidBypass 4081 0x00FF1
EEL_RemoveCleanCart 4082 0x00FF2
EEL_RinseBackAvail 4083 0x00FF3
EEL_PrsMustCalculate 4084 0x00FF4
EEL_CentralChemRqAlt1 4085 0x00FF5
EEL_CentralChemRqAlt2 4086 0x00FF6
EEL_CentralChemRqAlt3 4087 0x00FF7
EEL_IsolUfHandl 4088 0x00FF8
EEL_RS232 4089 0x00FF9
EEL_BicFlEmpty 4090 0x00FFA
EEL_ApEarlyDiaFlStartAlt1 4091 0x00FFB
EEL_ApEarlyDiaFlStartAlt2 4092 0x00FFC
EEL_ApEarlyDiaFlStartAlt3 4093 0x00FFD
EEL_FlushTubeAttnAct 4094 0x00FFE
EEL_OvernightDisinfAlt1 4095 0x00FFF
EEL_OvernightDisinfAlt2 4096 0x01000


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 95 (381)
Program version 1.1.0
C EN 2015-03-26

EEL_OvernightDisinfAlt3 4097 0x01001


EEL_UfdInstalled 4098 0x01002
EEL_BloodPumpStop 4099 0x01003
EEL_BPMInstalled 4100 0x01004
EEL_BPMAlarm 4101 0x01005
EEL_ProfInstalled 4102 0x01006
EEL_SetflowInstalled 4103 0x01007
EEL_NoWaterCons 4104 0x01008
EEL_AutoPrimType 4105 0x01009
EEL_WRO300HInstalled 4106 0x0100A
EEL_AutoHeatDisinfWRO300H 4107 0x0100B
EEL_WRO300HCommMode 4108 0x0100C
EEL_VerifyConc 4109 0x0100D
EEL_LogMem0 4110 0x0100E
EEL_LogMem1 4111 0x0100F
EEL_LogMem2 4112 0x01010
EEL_RbNoBloodDetect 4113 0x01011
EEL_OpenSettingReminder 4114 0x01012
EEL_MenuClosedAttentions 4115 0x01013
EEL_BPMMeasReadyAttn 4116 0x01014
EEL_RecirCSBM 4117 0x01015
EEL_VenClampOpen 4118 0x01016
EEL_GravityRinseInstalled 4119 0x01017
EEL_RunTimeCodepageSwtich 4120 0x01018
EEL_SucPumpRunningRecirc 4121 0x01019
EEL_HepComplAttn 4122 0x0101A
EEL_HBVAInstalled 4123 0x0101B
EEL_DiascanInstalled 4124 0x0101C
EEL_SnMode 4125 0x0101D
EEL_BasicFchEnabled 4126 0x0101E
EEL_CanisterSup 4127 0x0101F
EEL_HeatAlt4AutoRestart 4128 0x01020
EEL_CheckSumErrorOccurred 4129 0x01021
EER_Dummy 4130 0x01022
EEW_Dummy 4131 0x01023
EEW_ChemRinseTime 4132 0x01024
EEW_ChemRinseTimeAlt1 4133 0x01025
EEW_ChemRinseTimeAlt2 4134 0x01026
EEW_ChemRinseTimeAlt3 4135 0x01027
EEW_ChemRinseTimeAlt6 4136 0x01028
EEW_ChemRinseTimeAlt7 4137 0x01029
EEW_ChemRinseTimeUfdAlt1 4138 0x0102A
EEW_ChemRinseTimeUfdAlt2 4139 0x0102B


96 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EEW_ChemRinseTimeUfdAlt3 4140 0x0102C


EEW_ChemRinseTimeUfdAlt6 4141 0x0102D
EEW_ChemRinseTimeUfdAlt7 4142 0x0102E
EEW_CitricFlushTimeAlt1 4143 0x0102F
EEW_CitricFlushTimeAlt2 4144 0x01030
EEW_CitricFlushTimeAlt4 4145 0x01031
EEW_CitricFlushTimeAlt6 4146 0x01032
EEW_CitricFlushTimeAlt7 4147 0x01033
EEW_DwellTimeAlt1 4148 0x01034
EEW_DwellTimeAlt2 4149 0x01035
EEW_DwellTimeAlt3 4150 0x01036
EEW_DwellTimeAlt6 4151 0x01037
EEW_DwellTimeAlt7 4152 0x01038
EEW_LogIndex0 4153 0x01039
EEW_LogIndex1 4154 0x0103A
EEW_LogIndex2 4155 0x0103B
EEW_UFStopTimeTo 4156 0x0103C
EEW_BiCartExtraFlushTo 4157 0x0103D
EEW_HeatCoolOffToAlt1 4158 0x0103E
EEW_HeatCoolOffToAlt2 4159 0x0103F
EEW_HeatCoolOffToAlt3 4160 0x01040
EEW_HeatFlushCycleToAlt1 4161 0x01041
EEW_HeatFlushCycleToAlt2 4162 0x01042
EEW_HeatFlushCycleToAlt3 4163 0x01043
EEW_TimeBetweenDisinf 4164 0x01044
EEW_AutoOffAfterDisTime 4165 0x01045
EEW_WRO300HColdPrepTime 4166 0x01046
EEW_WRO300HLFHPrepTime 4167 0x01047
EEW_W300SelfHeatCoolTime 4168 0x01048
EEW_WaterDistrHeatDelayTime 4169 0x01049
EEW_LowFlowHeatTime 4170 0x0104A
EEW_StartLFHModeTimeout 4171 0x0104B
EEW_InitialINVAOpenDelay 4172 0x0104C
EEW_DwellTimeAlt4 4173 0x0104D
EEW_HeatFlushCycleToAlt4 4174 0x0104E
EEW_ExtraRinseBefHeatToAlt4 4175 0x0104F
C_Dummy 4176 0x01050
C_EvfWarmStart 4177 0x01051
C_InitProtValRq 4178 0x01052
C_InvalidRecovery 4179 0x01053
C_LogComFail 4180 0x01054
C_MonitorResetRq 4181 0x01055
C_MonitorTempAttn 4182 0x01056


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 97 (381)
Program version 1.1.0
C EN 2015-03-26

C_MonitorTempNewMeasure 4183 0x01057


C_RecoverAlarm 4184 0x01058
C_RecoverNoFluid 4185 0x01059
C_RecoverRdyToShutdown 4186 0x0105A
C_ResetEnabled 4187 0x0105B
C_RunTimeAvail 4188 0x0105C
C_RunTimeWriteBusy 4189 0x0105D
C_SafeMemCheckSumErr 4190 0x0105E
C_SafeMemCheckSumRdy 4191 0x0105F
C_SaveLoggingSetupPend 4192 0x01060
C_StoreBattFailedCnt 4193 0x01061
C_SDTLogDataBufferFull 4194 0x01062
C_WarmStart 4195 0x01063
C_WarmStartNoFluidRq 4196 0x01064
C_WarmStartRq 4197 0x01065
C_WarmStartRqAck 4198 0x01066
C_WriteNewAccCleanTime 4199 0x01067
C_WriteNewAccDecalcBiCartCount 4200 0x01068
C_WriteNewAccDecalcLiqBicCount 4201 0x01069
C_RecoverDataCollection 4202 0x0106A
CI_Dummy 4203 0x0106B
CB_Dummy 4204 0x0106C
CR_Dummy 4205 0x0106D
CW_Dummy 4206 0x0106E
CD_Dummy 4207 0x0106F
CD_RTMTimeMs 4208 0x01070
CD_RTMTime 4209 0x01071
CD_RTMTimeTmp 4210 0x01072
CF_AnalogConnCheck 4211 0x01073
CF_MonitorTemp 4212 0x01074
CF_Recover 4213 0x01075
CF_RunTimeMeter 4214 0x01076
CF_SafeMem 4215 0x01077
PEEL_BcMode 4217 0x01079
PEEL_SnMode 4218 0x0107A
PEEL_PhAvail 4219 0x0107B
PEEL_NegTmpAttnAct 4220 0x0107C
PEEL_AprPresetHLAvail 4221 0x0107D
PEEL_InletCondCheckActive 4222 0x0107E
PEEL_DiascanInstalled 4223 0x0107F
PEEL_ArterialClampAvail 4224 0x01080
PEEL_LoudSpeaker 4225 0x01081
PEEI_BlPumpStopTime 4227 0x01083


98 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PEEI_AirDetOvrTimePrim 4228 0x01084


PEEI_AirDetOvrTimeTmt 4229 0x01085
PEEI_NotUsed 4230 0x01086
PEEI_BlLeakOverrideTime 4231 0x01087
PEEI_AprLL 4232 0x01088
PEEI_AprHL 4233 0x01089
PEEI_AprLLInit 4234 0x0108A
PEEI_AprHLInit 4235 0x0108B
PEEI_AprWidth 4236 0x0108C
PEEI_AprLowPart 4237 0x0108D
PEEI_AprMargin 4238 0x0108E
PEEI_AcOffsetAlt1 4239 0x0108F
PEEI_AcCoeffAlt1 4240 0x01090
PEEI_AcOffsetAlt2 4241 0x01091
PEEI_AcCoeffAlt2 4242 0x01092
PEEI_AcOffsetAlt3 4243 0x01093
PEEI_AcCoeffAlt3 4244 0x01094
PEEI_AcNaHL 4245 0x01095
PEEI_AcNaLL 4246 0x01096
PEEI_AcNaInit 4247 0x01097
PEEI_AcConcAlt 4248 0x01098
PEEI_BcC1Alt1 4249 0x01099
PEEI_BcC2Alt1 4250 0x0109A
PEEI_BcC3Alt1 4251 0x0109B
PEEI_BcC4Alt1 4252 0x0109C
PEEI_BcC5Alt1 4253 0x0109D
PEEI_BcC6Alt1 4254 0x0109E
PEEI_BcC7Alt1 4255 0x0109F
PEEI_BcC8Alt1 4256 0x010A0
PEEI_BcC9Alt1 4257 0x010A1
PEEI_BcC10Alt1 4258 0x010A2
PEEI_BcC11Alt1 4259 0x010A3
PEEI_BcC12Alt1 4260 0x010A4
PEEI_BcC1Alt2 4261 0x010A5
PEEI_BcC2Alt2 4262 0x010A6
PEEI_BcC3Alt2 4263 0x010A7
PEEI_BcC4Alt2 4264 0x010A8
PEEI_BcC5Alt2 4265 0x010A9
PEEI_BcC6Alt2 4266 0x010AA
PEEI_BcC7Alt2 4267 0x010AB
PEEI_BcC8Alt2 4268 0x010AC
PEEI_BcC9Alt2 4269 0x010AD
PEEI_BcC10Alt2 4270 0x010AE


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 99 (381)
Program version 1.1.0
C EN 2015-03-26

PEEI_BcC11Alt2 4271 0x010AF


PEEI_BcC12Alt2 4272 0x010B0
PEEI_BcC1Alt3 4273 0x010B1
PEEI_BcC2Alt3 4274 0x010B2
PEEI_BcC3Alt3 4275 0x010B3
PEEI_BcC4Alt3 4276 0x010B4
PEEI_BcC5Alt3 4277 0x010B5
PEEI_BcC6Alt3 4278 0x010B6
PEEI_BcC7Alt3 4279 0x010B7
PEEI_BcC8Alt3 4280 0x010B8
PEEI_BcC9Alt3 4281 0x010B9
PEEI_BcC10Alt3 4282 0x010BA
PEEI_BcC11Alt3 4283 0x010BB
PEEI_BcC12Alt3 4284 0x010BC
PEEI_BcNaHL 4285 0x010BD
PEEI_BcNaLL 4286 0x010BE
PEEI_BcHco3HL 4287 0x010BF
PEEI_BcHco3LL 4288 0x010C0
PEEI_BcNaInit 4289 0x010C1
PEEI_BcHco3Init 4290 0x010C2
PEEI_BcConcAlt 4291 0x010C3
PEEI_PhLL 4292 0x010C4
PEEI_PhHL 4293 0x010C5
PEEI_PhLLInit 4294 0x010C6
PEEI_PhHLInit 4295 0x010C7
PEEI_TmpLL 4296 0x010C8
PEEI_TmpHL 4297 0x010C9
PEEI_TmpLLInit 4298 0x010CA
PEEI_TmpHLInit 4299 0x010CB
PEEI_TmpWidth 4300 0x010CC
PEEI_TmpLowPart 4301 0x010CD
PEEI_TmpIsolWidth 4302 0x010CE
PEEI_TmpIsolLowPart 4303 0x010CF
PEEI_VprLL 4304 0x010D0
PEEI_VprHL 4305 0x010D1
PEEI_VprLLInit 4306 0x010D2
PEEI_VprHLInit 4307 0x010D3
PEEI_VprWidth 4308 0x010D4
PEEI_VprLowPart 4309 0x010D5
PEEI_VprLLSnAutoLim 4310 0x010D6
PEEI_VprHLSnAutoLim 4311 0x010D7
PEEI_VprAutoLimAlarmTime 4312 0x010D8
PEEI_AprHLimit 4313 0x010D9


100 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

PEEI_CondDiffLimit 4314 0x010DA


PEEI_VprMaxHL 4315 0x010DB
PEEI_VprMinLL 4316 0x010DC
PEEI_AprMaxHL 4317 0x010DD
PEEI_AprMinLL 4318 0x010DE
PEEI_PhMaxHL 4319 0x010DF
PEEI_PhMinLL 4320 0x010E0
PEEI_BPMInitCuffPr 4321 0x010E1
PEEI_ProfModel1HCO3Start 4322 0x010E2
PEEI_ProfModel1HCO3Stop 4323 0x010E3
PEEI_ProfModel1NaAcStart 4324 0x010E4
PEEI_ProfModel1NaAcStop 4325 0x010E5
PEEI_ProfModel1NaStart 4326 0x010E6
PEEI_ProfModel1NaStop 4327 0x010E7
PEEI_ProfModel2HCO3Start 4328 0x010E8
PEEI_ProfModel2HCO3Stop 4329 0x010E9
PEEI_ProfModel2NaAcStart 4330 0x010EA
PEEI_ProfModel2NaAcStop 4331 0x010EB
PEEI_ProfModel2NaStart 4332 0x010EC
PEEI_ProfModel2NaStop 4333 0x010ED
PEEI_ProfModel3HCO3Start 4334 0x010EE
PEEI_ProfModel3HCO3Stop 4335 0x010EF
PEEI_ProfModel3NaAcStart 4336 0x010F0
PEEI_ProfModel3NaAcStop 4337 0x010F1
PEEI_ProfModel3NaStart 4338 0x010F2
PEEI_ProfModel3NaStop 4339 0x010F3
PEEI_BPMInflationMode 4340 0x010F4
PEEI_DiascanMaxStep 4341 0x010F5
PEEI_DiascanDoseMax 4342 0x010F6
PEEI_DiascanDoseMin 4343 0x010F7
PEEI_LoudSpeakerVolume 4344 0x010F8
PEEI_HepSyringeDiam 4345 0x010F9
PEEI_HeparinFlowRate 4346 0x010FA
PEEI_HepSetLLLvl1 4347 0x010FB
PEEI_HepSetLLLvl2 4348 0x010FC
PEEI_HepSetLLLvl3 4349 0x010FD
PEEI_HepSetHLLvl1 4350 0x010FE
PEEI_HepSetHLLvl2 4351 0x010FF
PEEI_HepSetHLLvl3 4352 0x01100
PEEI_PdSupLimit 4353 0x01101
PEEI_UfsStartVolAlarm 4354 0x01102
PEEI_UfsVolAlarmWideningRate 4355 0x01103
EPB_dummy 4359 0x01107


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 101 (381)
Program version 1.1.0
C EN 2015-03-26

EPB_AcConc1Type 4366 0x0110E


EPB_AcConc2Type 4367 0x0110F
EPB_AcConc3Type 4368 0x01110
EPB_AcConcTypeHl 4369 0x01111
EPB_AcUserLblAlt1 4370 0x01112
EPB_AcUserLblAlt2 4384 0x01120
EPB_AcUserLblAlt3 4398 0x0112E
EPB_ApBankNr 4412 0x0113C
EPB_ApBankNrHL 4413 0x0113D
EPB_ApBankNrLL 4414 0x0113E
EPB_ApUserLblAlt1 4415 0x0113F
EPB_ApUserLblAlt2 4429 0x0114D
EPB_ApUserLblAlt3 4443 0x0115B
EPB_BcConc1Type 4457 0x01169
EPB_BcConc2Type 4458 0x0116A
EPB_BcConc3Type 4459 0x0116B
EPB_BcConcTypeHl 4460 0x0116C
EPB_BcUserLblAlt1 4461 0x0116D
EPB_BcUserLblAlt2 4475 0x0117B
EPB_BcUserLblAlt3 4489 0x01189
EPB_ChemBankNr 4503 0x01197
EPB_ChemDisAlt 4504 0x01198
EPB_ChemDisUserLblAlt1 4505 0x01199
EPB_ChemDisUserLblAlt2 4519 0x011A7
EPB_ChemDisUserLblAlt3 4533 0x011B5
EPB_ChemDisUserLblAlt5 4547 0x011C3
EPB_ChemDisUserLblAlt6 4561 0x011D1
EPB_ChemDisUserLblAlt7 4575 0x011DF
EPB_DiaTypeLL 4589 0x011ED
EPB_DiaTypeHL 4590 0x011EE
EPB_DiaType 4591 0x011EF
EPB_DisinfAltHL 4592 0x011F0
EPB_DisinfAltLL 4593 0x011F1
EPB_HeatDisAlt 4594 0x011F2
EPB_HeaterEfficiency 4595 0x011F3
EPB_HeaterEfficiencyHl 4596 0x011F4
EPB_HeaterEfficiencyLl 4597 0x011F5
EPB_LangSel 4598 0x011F6
EPB_LangSelHl 4599 0x011F7
EPB_LangSelLl 4600 0x011F8
EPB_LogDecade 4601 0x011F9
EPB_LogFieldId 4602 0x011FA
EPB_LogFormat0 4603 0x011FB


102 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPB_LogFormat1 4604 0x011FC


EPB_LogFormat2 4605 0x011FD
EPB_LogSel0 4606 0x011FE
EPB_LogSel1 4607 0x011FF
EPB_LogSel2 4608 0x01200
EPB_LogType0 4609 0x01201
EPB_LogType1 4610 0x01202
EPB_LogType2 4611 0x01203
EPB_ViewAngleBgph 4612 0x01204
EPB_ViewAngleInfoDisp 4613 0x01205
EPB_Zero 4614 0x01206
EPB_ProfTypeLL 4615 0x01207
EPB_ProfTypeHL 4616 0x01208
EPB_ProfModel1NaAcProfType 4617 0x01209
EPB_ProfModel1NaProfType 4618 0x0120A
EPB_ProfModel1HCO3ProfType 4619 0x0120B
EPB_ProfModel1UFProfType 4620 0x0120C
EPB_ProfModel2NaAcProfType 4621 0x0120D
EPB_ProfModel2NaProfType 4622 0x0120E
EPB_ProfModel2HCO3ProfType 4623 0x0120F
EPB_ProfModel2UFProfType 4624 0x01210
EPB_ProfModel3NaAcProfType 4625 0x01211
EPB_ProfModel3NaProfType 4626 0x01212
EPB_ProfModel3HCO3ProfType 4627 0x01213
EPB_ProfModel3UFProfType 4628 0x01214
EPB_ProfUFTypeLL 4629 0x01215
EPB_ProfUFTypeHL 4630 0x01216
EPB_HepSyringeType 4631 0x01217
EPB_ActiveDisinfRinseAltLL 4646 0x01226
EPB_ActiveDisinfRinseAltHL 4647 0x01227
EPB_ActiveDisinfRinseAlt 4648 0x01228
EPI_ACBPumpSpeedWindowLLLvl0 4649 0x01229
EPI_ACBPumpSpeedWindowHLLvl0 4650 0x0122A
EPI_ACBPumpSpeedWindow 4651 0x0122B
EPI_ACBPumpSpeedWindowHLLvl3 4652 0x0122C
EPI_ACBPumpSpeedWindowHLLvl2 4653 0x0122D
EPI_ACBPumpSpeedWindowHLLvl1 4654 0x0122E
EPI_AcConcAlt 4655 0x0122F
EPI_AcConcAltHL 4656 0x01230
EPI_AcConcAltLL 4657 0x01231
EPI_AcetateCoeffAlt1 4658 0x01232
EPI_AcetateCoeffAlt2 4659 0x01233
EPI_AcetateCoeffAlt3 4660 0x01234


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 103 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_AcetateOffsetAlt1 4661 0x01235


EPI_AcetateOffsetAlt2 4662 0x01236
EPI_AcetateOffsetAlt3 4663 0x01237
EPI_AcMolCoeffAlt1 4664 0x01238
EPI_AcMolCoeffAlt2 4665 0x01239
EPI_AcMolCoeffAlt3 4666 0x0123A
EPI_AcMolOffsetAlt1 4667 0x0123B
EPI_AcMolOffsetAlt2 4668 0x0123C
EPI_AcMolOffsetAlt3 4669 0x0123D
EPI_AdRefCoeff 4670 0x0123E
EPI_AdRefCoeffHl 4671 0x0123F
EPI_AdRefCoeffLl 4672 0x01240
EPI_AirDetAlarmResolvTimePrim 4673 0x01241
EPI_AirDetAlarmResolvTimeTmt 4674 0x01242
EPI_AirDetAlarmResolvTimeHLLvl0 4675 0x01243
EPI_AirDetAlarmResolvTimeLLLvl0 4676 0x01244
EPI_ADOvrTmtTimeHLLv10 4677 0x01245
EPI_ADOvrTmtTimeLLLv10 4678 0x01246
EPI_ADOvrTmtTimeHLLv11 4679 0x01247
EPI_ADOvrTmtTimeLLLv11 4680 0x01248
EPI_ADOvrTmtTimeHLLv12 4681 0x01249
EPI_ADOvrTmtTimeLLLv12 4682 0x0124A
EPI_ADOvrTmtTimeHLLv13 4683 0x0124B
EPI_ADOvrTmtTimeLLLv13 4684 0x0124C
EPI_AirDetOvrTmtTimeHL 4685 0x0124D
EPI_AirDetOvrTmtTimeLL 4686 0x0124E
EPI_ApBlPumpFillFlLL 4687 0x0124F
EPI_ApBlPumpFillFlHL 4688 0x01250
EPI_ApBlPumpFillFl 4689 0x01251
EPI_ApBlPumpPrimFlLL 4690 0x01252
EPI_ApBlPumpPrimlHL 4691 0x01253
EPI_ApBlPumpPrimFl 4692 0x01254
EPI_ApFillTimeLL 4693 0x01255
EPI_ApFillTimeHL 4694 0x01256
EPI_ApFillTime 4695 0x01257
EPI_ApFlush1TimeLL 4696 0x01258
EPI_ApFlush1TimeHL 4697 0x01259
EPI_ApFlush1Time 4698 0x0125A
EPI_ApFlush2TimeLL 4699 0x0125B
EPI_ApFlush2TimeHL 4700 0x0125C
EPI_ApFlush2Time 4701 0x0125D
EPI_ApLevelAdjTimeLL 4702 0x0125E
EPI_ApLevelAdjTimeHL 4703 0x0125F


104 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_ApLevelAdjTime 4704 0x01260


EPI_ApMaxPrDropLL 4705 0x01261
EPI_ApMaxPrDropHL 4706 0x01262
EPI_ApMaxPrDrop 4707 0x01263
EPI_ApPrBuildTimeLL 4708 0x01264
EPI_ApPrBuildTimeHL 4709 0x01265
EPI_ApPrBuildTime 4710 0x01266
EPI_ApPrimCycleTimeLL 4711 0x01267
EPI_ApPrimCycleTimeHL 4712 0x01268
EPI_ApPrimCycleTime 4713 0x01269
EPI_ApPrimTimeLL 4714 0x0126A
EPI_ApPrimTimeHL 4715 0x0126B
EPI_ApPrimTime 4716 0x0126C
EPI_ApPrTestTimeLL 4717 0x0126D
EPI_ApPrTestTimeHL 4718 0x0126E
EPI_ApPrTestTime 4719 0x0126F
EPI_AprHLInit 4720 0x01270
EPI_AprHLLvl0 4721 0x01271
EPI_AprLLInit 4722 0x01272
EPI_AprLLLvl0 4723 0x01273
EPI_AprLowPart 4724 0x01274
EPI_AprLowPartHLLvl0 4725 0x01275
EPI_AprLowPartLLLvl0 4726 0x01276
EPI_NotUsed 4727 0x01277
EPI_NotUsedHLLvl0 4728 0x01278
EPI_NotUsedLLLvl0 4729 0x01279
EPI_AprPrCalHL 4730 0x0127A
EPI_AprPrCalLL 4731 0x0127B
EPI_AprPrCoeffHL 4732 0x0127C
EPI_AprPrCoeffLL 4733 0x0127D
EPI_AprPrOffsetHL 4734 0x0127E
EPI_AprPrOffsetLL 4735 0x0127F
EPI_AprWidth 4736 0x01280
EPI_AprWidthHLLvl0 4737 0x01281
EPI_AprWidthLLLvl0 4738 0x01282
EPI_ApTubeTimeLL 4739 0x01283
EPI_ApTubeTimeHL 4740 0x01284
EPI_ApTubeTime 4741 0x01285
EPI_ApUFRateSet 4742 0x01286
EPI_ApUFRateSetLL 4743 0x01287
EPI_ApUFRateSetHL 4744 0x01288
EPI_ARinseBlPumpFlLL 4745 0x01289
EPI_ARinseBlPumpFlHL 4746 0x0128A


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 105 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_ARinseBlPumpFl 4747 0x0128B


EPI_BCAPumpSpeedWindowLLLvl0 4748 0x0128C
EPI_BCAPumpSpeedWindowHLLvl0 4749 0x0128D
EPI_BCAPumpSpeedWindow 4750 0x0128E
EPI_BCAPumpSpeedWindowHLLvl3 4751 0x0128F
EPI_BCAPumpSpeedWindowHLLvl2 4752 0x01290
EPI_BCAPumpSpeedWindowHLLvl1 4753 0x01291
EPI_BCBPumpSpeedWindowLLLvl0 4754 0x01292
EPI_BCBPumpSpeedWindowHLLvl0 4755 0x01293
EPI_BCBPumpSpeedWindow 4756 0x01294
EPI_BCBPumpSpeedWindowHLLvl3 4757 0x01295
EPI_BCBPumpSpeedWindowHLLvl2 4758 0x01296
EPI_BCBPumpSpeedWindowHLLvl1 4759 0x01297
EPI_BcConcAlt 4760 0x01298
EPI_BcConcAltHL 4761 0x01299
EPI_BcConcAltLL 4762 0x0129A
EPI_BicarbC1Alt1 4763 0x0129B
EPI_BicarbC1Alt2 4764 0x0129C
EPI_BicarbC1Alt3 4765 0x0129D
EPI_BicarbC2Alt1 4766 0x0129E
EPI_BicarbC2Alt2 4767 0x0129F
EPI_BicarbC2Alt3 4768 0x012A0
EPI_BicarbC3Alt1 4769 0x012A1
EPI_BicarbC3Alt2 4770 0x012A2
EPI_BicarbC3Alt3 4771 0x012A3
EPI_BicarbC4Alt1 4772 0x012A4
EPI_BicarbC4Alt2 4773 0x012A5
EPI_BicarbC4Alt3 4774 0x012A6
EPI_BicarbC5Alt1 4775 0x012A7
EPI_BicarbC5Alt2 4776 0x012A8
EPI_BicarbC5Alt3 4777 0x012A9
EPI_BicarbC6Alt1 4778 0x012AA
EPI_BicarbC6Alt2 4779 0x012AB
EPI_BicarbC6Alt3 4780 0x012AC
EPI_BicarbC7Alt1 4781 0x012AD
EPI_BicarbC7Alt2 4782 0x012AE
EPI_BicarbC7Alt3 4783 0x012AF
EPI_BicarbC8Alt1 4784 0x012B0
EPI_BicarbC8Alt2 4785 0x012B1
EPI_BicarbC8Alt3 4786 0x012B2
EPI_BicarbC9Alt1 4787 0x012B3
EPI_BicarbC9Alt2 4788 0x012B4
EPI_BicarbC9Alt3 4789 0x012B5


106 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_BicarbC10Alt1 4790 0x012B6


EPI_BicarbC10Alt2 4791 0x012B7
EPI_BicarbC10Alt3 4792 0x012B8
EPI_BicarbC11Alt1 4793 0x012B9
EPI_BicarbC11Alt2 4794 0x012BA
EPI_BicarbC11Alt3 4795 0x012BB
EPI_BicarbC12Alt1 4796 0x012BC
EPI_BicarbC12Alt2 4797 0x012BD
EPI_BicarbC12Alt3 4798 0x012BE
EPI_BlLeakCalHL 4799 0x012BF
EPI_BlLeakCalLL 4800 0x012C0
EPI_BlLeakCoeffHL 4801 0x012C1
EPI_BlLeakCoeffLL 4802 0x012C2
EPI_BlLeakOffsetHL 4803 0x012C3
EPI_BlLeakOffsetLL 4804 0x012C4
EPI_BlLeakOverrideTime 4805 0x012C5
EPI_BlLeakOverrideTimeHLLvl0 4806 0x012C6
EPI_BlLeakOverrideTimeLLLvl0 4807 0x012C7
EPI_RecirBloodFlSetLLLvl0 4808 0x012C8
EPI_RecirBloodFlSetHlLvl0 4809 0x012C9
EPI_BloodFlSetLLLvl0 4810 0x012CA
EPI_BloodFlSetHlLvl0 4811 0x012CB
EPI_BloodFlSetLLLvl1 4812 0x012CC
EPI_BloodFlSetLLLvl2 4813 0x012CD
EPI_BloodFlSetLLLvl3 4814 0x012CE
EPI_BloodFlSetLL 4815 0x012CF
EPI_RecirBloodFlSet 4816 0x012D0
EPI_BloodFlSet 4817 0x012D1
EPI_BloodFlSetHLLvl3 4818 0x012D2
EPI_BloodFlSetHLLvl2 4819 0x012D3
EPI_BloodFlSetHLLvl1 4820 0x012D4
EPI_BlFlLowLimSupFlow 4821 0x012D5
EPI_BlFlLowLimSupFlowLlLvl0 4822 0x012D6
EPI_BlFlLowLimSupFlowHlLvl0 4823 0x012D7
EPI_BlFlLowLimSupTimeLlLvl0 4824 0x012D8
EPI_BlFlLowLimSupTimeHlLvl0 4825 0x012D9
EPI_BlFlLowLimSupTime 4826 0x012DA
EPI_BlPumpBypassSetFlowLLLvl0 4827 0x012DB
EPI_BlPumpBypassSetFlowHLLvL0 4828 0x012DC
EPI_BlPumpBypassSetFlowLL 4829 0x012DD
EPI_BlPumpBypassSetFlow 4830 0x012DE
EPI_BlPumpBypassSetFlowHLLvl3 4831 0x012DF
EPI_BlPumpBypassSetFlowHLLvl2 4832 0x012E0


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 107 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_BlPumpBypassSetFlowHLLvl1 4833 0x012E1


EPI_BlPumpPCoeffVprTst 4834 0x012E2
EPI_BlPumpPCoeffVprTstHL 4835 0x012E3
EPI_BlPumpPCoeffVprTstLL 4836 0x012E4
EPI_BlPumpStopTime 4837 0x012E5
EPI_BlPumpStopTimeHLLvl0 4838 0x012E6
EPI_BlPumpStopTimeLLLvl0 4839 0x012E7
EPI_BlTubeCoeffDef 4840 0x012E8
EPI_BlTubeDiam0 4841 0x012E9
EPI_BlTubeDiam1 4842 0x012EA
EPI_BlTubeDiam2 4843 0x012EB
EPI_BlTubeDiamDef 4844 0x012EC
EPI_BlTubeDiamHl 4845 0x012ED
EPI_BlTubeDiamLl 4846 0x012EE
EPI_ChemCondLimLL 4847 0x012EF
EPI_ChemCondLimHL 4848 0x012F0
EPI_ChemMixRatioLL 4849 0x012F1
EPI_ChemMixRatioHL 4850 0x012F2
EPI_ChemMixRatioAlt1 4851 0x012F3
EPI_ChemMixRatioAlt2 4852 0x012F4
EPI_ChemMixRatioAlt3 4853 0x012F5
EPI_ChemMixRatioAlt6 4854 0x012F6
EPI_ChemMixRatioAlt7 4855 0x012F7
EPI_ChemRinseTempLL 4856 0x012F8
EPI_ChemRinseTempHL 4857 0x012F9
EPI_ChemRinseTempAlt1 4858 0x012FA
EPI_ChemRinseTempAlt2 4859 0x012FB
EPI_ChemRinseTempAlt3 4860 0x012FC
EPI_ChemRinseTempAlt6 4861 0x012FD
EPI_ChemRinseTempAlt7 4862 0x012FE
EPI_CondCalHL 4863 0x012FF
EPI_CondCalLL 4864 0x01300
EPI_CondCoeffHl 4865 0x01301
EPI_CondCoeffLl 4866 0x01302
EPI_CONDLLLVL0 4867 0x01303
EPI_CONDHLLVL0 4868 0x01304
EPI_CondOffsetHl 4869 0x01305
EPI_CondOffsetLl 4870 0x01306
EPI_CondTempCoeffHl 4871 0x01307
EPI_CondTempCoeffLl 4872 0x01308
EPI_CondTempDisCalHL 4873 0x01309
EPI_CondTempDisCalLL 4874 0x0130A
EPI_CondTempHeatCoeffHl 4875 0x0130B


108 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_CondTempHeatCoeffLl 4876 0x0130C


EPI_CondTempHeatOffsetHl 4877 0x0130D
EPI_CondTempHeatOffsetLl 4878 0x0130E
EPI_CondTempOffsetHl 4879 0x0130F
EPI_CondTempOffsetLl 4880 0x01310
EPI_CondTempTMTCalHL 4881 0x01311
EPI_CondTempTMTCalLL 4882 0x01312
EPI_DegassPrLL 4883 0x01313
EPI_DegassPrHL 4884 0x01314
EPI_DegassPr 4885 0x01315
EPI_DegassPrCalHL 4886 0x01316
EPI_DegassPrCalLL 4887 0x01317
EPI_DegassPrCoeffHl 4888 0x01318
EPI_DegassPrCoeffLl 4889 0x01319
EPI_DegassPrOffsetHl 4890 0x0131A
EPI_DegassPrOffsetLl 4891 0x0131B
EPI_DiagCondLL 4892 0x0131C
EPI_DiagCondHL 4893 0x0131D
EPI_DiagDegassPrLL 4894 0x0131E
EPI_DiagDegassPrHL 4895 0x0131F
EPI_DiagHPGLL 4896 0x01320
EPI_DiagHPGHL 4897 0x01321
EPI_DiagTempLL 4898 0x01322
EPI_DiagTempHL 4899 0x01323
EPI_DiaPrCalHL 4900 0x01324
EPI_DiaPrCalLL 4901 0x01325
EPI_DiaPrCoeffHL 4902 0x01326
EPI_DiaPrCoeffLL 4903 0x01327
EPI_DiaPrOffsetHL 4904 0x01328
EPI_DiaPrOffsetLL 4905 0x01329
EPI_EXTRAPRIMVOLLIMITBS 4906 0x0132A
EPI_EXTRAPRIMVOLLIMITBSHLLVL0 4907 0x0132B
EPI_EXTRAPRIMVOLLIMITBSLLLVL0 4908 0x0132C
EPI_EXTRAPRIMVOLLIMITSS 4909 0x0132D
EPI_EXTRAPRIMVOLLIMITSSHLLVL0 4910 0x0132E
EPI_EXTRAPRIMVOLLIMITSSLLLVL0 4911 0x0132F
EPI_ConnectPatVolLimitLLLvl0 4912 0x01330
EPI_ConnectPatVolLimitHLLvl0 4913 0x01331
EPI_ConnectPatVolLimit 4914 0x01332
EPI_HCO3SetInit 4915 0x01333
EPI_HCO3SETHLLVL0 4916 0x01334
EPI_HCO3SETLLLVL0 4917 0x01335
EPI_HCO3SetOff 4918 0x01336


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 109 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_HCO3SETOFFHLLVL0 4919 0x01337


EPI_HCO3SETOFFLLLVL0 4920 0x01338
EPI_HeatingTemp 4921 0x01339
EPI_HeatingTempLL 4922 0x0133A
EPI_HeatingTempHl 4923 0x0133B
EPI_HeatingTempMM90 4924 0x0133C
EPI_HeatingTempMM90LL 4925 0x0133D
EPI_HeatingTempMM90HL 4926 0x0133E
EPI_HepSetHLLvl0 4927 0x0133F
EPI_HepSetLLLvl0 4928 0x01340
EPI_HepSetLLLvl1 4929 0x01341
EPI_HepSetLLLvl2 4930 0x01342
EPI_HepSetLLLvl3 4931 0x01343
EPI_HepSet 4932 0x01344
EPI_HepSetHLLvl3 4933 0x01345
EPI_HepSetHLLvl2 4934 0x01346
EPI_HepSetHLLvl1 4935 0x01347
EPI_HepStopLimLLLvl0 4936 0x01348
EPI_HepStopLimHLLvL0 4937 0x01349
EPI_HepStopLimLLLvl1 4938 0x0134A
EPI_HepStopLimLLLvl2 4939 0x0134B
EPI_HepStopLimLLLvl3 4940 0x0134C
EPI_HepStopLimSet 4941 0x0134D
EPI_HepStopLimHLLvl3 4942 0x0134E
EPI_HepStopLimHLLvl2 4943 0x0134F
EPI_HepStopLimHLLvl1 4944 0x01350
EPI_HepSyringeDiam 4945 0x01351
EPI_HepSyringeDiamHl 4946 0x01352
EPI_HepSyringeDiamLl 4947 0x01353
EPI_HPGMaxLimit 4948 0x01354
EPI_HPGMaxLimitHL 4949 0x01355
EPI_HPGMaxLimitLL 4950 0x01356
EPI_HPGPrCalHL 4951 0x01357
EPI_HPGPrCalLL 4952 0x01358
EPI_HPGPrCoeffHl 4953 0x01359
EPI_HPGPrCoeffLl 4954 0x0135A
EPI_HPGPrOffsetHl 4955 0x0135B
EPI_HPGPrOffsetLl 4956 0x0135C
EPI_IntSuppCondLL 4957 0x0135D
EPI_IntSuppCondHL 4958 0x0135E
EPI_IntSuppTempLL 4959 0x0135F
EPI_IntSuppTempHL 4960 0x01360
EPI_MinDiagFreqArt 4961 0x01361


110 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_MaxDiagFreqArt 4962 0x01362


EPI_MinDiagFreqHep 4963 0x01363
EPI_MaxDiagFreqHep 4964 0x01364
EPI_MinUfRateLLLvl0 4965 0x01365
EPI_MinUfRateHLLLl0 4966 0x01366
EPI_MinUfRateLLLvl1 4967 0x01367
EPI_MinUfRateLLLvl2 4968 0x01368
EPI_MinUfRateLLLvl3 4969 0x01369
EPI_MinUfRate 4970 0x0136A
EPI_MinUfRateHLLvl3 4971 0x0136B
EPI_MinUfRateHLLvl2 4972 0x0136C
EPI_MinUfRateHLLvl1 4973 0x0136D
EPI_MM90AdRefCoeffHl 4974 0x0136E
EPI_MM90AdRefCoeffLl 4975 0x0136F
EPI_MonitorTempAttnLim 4976 0x01370
EPI_MonitorTempErrorLim 4977 0x01371
EPI_NAACSETHLLVL0 4978 0x01372
EPI_NaAcSetInit 4979 0x01373
EPI_NAACSETLLLVL0 4980 0x01374
EPI_NaAcSetOff 4981 0x01375
EPI_NAACSETOFFHLLVL0 4982 0x01376
EPI_NAACSETOFFLLLVL0 4983 0x01377
EPI_NASETHLLVL0 4984 0x01378
EPI_NaSetInit 4985 0x01379
EPI_NASETLLLVL0 4986 0x0137A
EPI_NaSetOff 4987 0x0137B
EPI_NASETOFFHLLVL0 4988 0x0137C
EPI_NASETOFFLLLVL0 4989 0x0137D
EPI_OverHeatTempCalHL 4990 0x0137E
EPI_OverHeatTempCalLL 4991 0x0137F
EPI_OverHeatTempCoeffHL 4992 0x01380
EPI_OverHeatTempCoeffLL 4993 0x01381
EPI_OverheatTempLimLLLvl0 4994 0x01382
EPI_OverheatTempLimHLLvl0 4995 0x01383
EPI_OverheatTempLim 4996 0x01384
EPI_OverheatTempLimHLLvl3 4997 0x01385
EPI_OverheatTempLimHLLvl2 4998 0x01386
EPI_OverheatTempLimHLLvl1 4999 0x01387
EPI_OverHeatTempOffsetHL 5000 0x01388
EPI_OverHeatTempOffsetLL 5001 0x01389
EPI_PAdRefCoeffHl 5002 0x0138A
EPI_PAdRefCoeffLl 5003 0x0138B
EPI_pHCalHL 5004 0x0138C


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 111 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_pHCalLL 5005 0x0138D


EPI_pHCoeffHL 5006 0x0138E
EPI_pHCoeffLL 5007 0x0138F
EPI_PhHLInit 5008 0x01390
EPI_PhHLLvl0 5009 0x01391
EPI_PhLLInit 5010 0x01392
EPI_PhLLLvl0 5011 0x01393
EPI_pHOffsetHL 5012 0x01394
EPI_pHOffsetLL 5013 0x01395
EPI_PrimDetCurrentHL 5014 0x01396
EPI_PrimDetCurrentLL 5015 0x01397
EPI_PrimUfRateLLLvl0 5016 0x01398
EPI_PrimUfRateHLLvl0 5017 0x01399
EPI_PrimUfRateLLLvl1 5018 0x0139A
EPI_PrimUfRateLLLvl2 5019 0x0139B
EPI_PrimUfRateLLLvl3 5020 0x0139C
EPI_PrimUfRate 5021 0x0139D
EPI_PrimUfRateHLLvl3 5022 0x0139E
EPI_PrimUfRateHLLvl2 5023 0x0139F
EPI_PrimUfRateHLLvl1 5024 0x013A0
EPI_PRIMVOLLIMIT 5025 0x013A1
EPI_PRIMVOLLIMITHLLVL0 5026 0x013A2
EPI_PRIMVOLLIMITLLLVL0 5027 0x013A3
EPI_PrimVolLimitSS 5028 0x013A4
EPI_PrimVolLimitSSHLLVL0 5029 0x013A5
EPI_PrimVolLimitSSLLLVL0 5030 0x013A6
EPI_PrimVolLimitBS 5031 0x013A7
EPI_PrimVolLimitBSHLLVL0 5032 0x013A8
EPI_PrimVolLimitBSLLLVL0 5033 0x013A9
EPI_PumpSpeedRatioWindowLLLvl0 5034 0x013AA
EPI_PumpSpeedRatioWindowHLLvl0 5035 0x013AB
EPI_PumpSpeedRatioWindow 5036 0x013AC
EPI_PumpSpeedRatioWindowHLLvl3 5037 0x013AD
EPI_PumpSpeedRatioWindowHLLvl2 5038 0x013AE
EPI_PumpSpeedRatioWindowHLLvl1 5039 0x013AF
EPI_RBFLOWHLLVL0 5040 0x013B0
EPI_RBFLOWLLLVL0 5041 0x013B1
EPI_RBFLOWSET 5042 0x013B2
EPI_RBVOLSET 5043 0x013B3
EPI_RbVolSetHl 5044 0x013B4
EPI_RbVolSetLl 5045 0x013B5
EPI_RegTempCoeffHl 5046 0x013B6
EPI_RegTempCoeffLl 5047 0x013B7


112 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_RegTempDisCalHL 5048 0x013B8


EPI_RegTempDisCalLL 5049 0x013B9
EPI_RegTempHeatCoeffHl 5050 0x013BA
EPI_RegTempHeatCoeffLl 5051 0x013BB
EPI_RegTempHeatOffsetHl 5052 0x013BC
EPI_RegTempHeatOffsetLl 5053 0x013BD
EPI_RegTempOffsetHl 5054 0x013BE
EPI_RegTempOffsetLl 5055 0x013BF
EPI_RegTempTMTCalHL 5056 0x013C0
EPI_RegTempTMTCalLL 5057 0x013C1
EPI_RecirUFTimeLLLVL0 5058 0x013C2
EPI_RecirUFTimeHLLVL0 5059 0x013C3
EPI_RecirTimeInit 5060 0x013C4
EPI_RecirUFRLLLvl0 5061 0x013C5
EPI_RecirUFRHLLvl0 5062 0x013C6
EPI_RecirUFRInit 5063 0x013C7
EPI_REMAINTIMELLLVL0 5064 0x013C8
EPI_REMAINTIMELLLVL1 5065 0x013C9
EPI_REMAINTIMELLLVL2 5066 0x013CA
EPI_RemainTimeLLLvl3 5067 0x013CB
EPI_REMAINTIMEHLLVL0 5068 0x013CC
EPI_REMAINTIMEHLLVL1 5069 0x013CD
EPI_REMAINTIMEHLLVL2 5070 0x013CE
EPI_RemainTimeHLLvl3 5071 0x013CF
EPI_RemainTime 5072 0x013D0
EPI_RemainTimeDiffInit 5073 0x013D1
EPI_RemainTimeIsolInit 5074 0x013D2
EPI_RinseCycleTime 5075 0x013D3
EPI_RINSECYCLETIMEHL 5076 0x013D4
EPI_RINSECYCLETIMELL 5077 0x013D5
EPI_RinseTimeMM90LLLvl0 5078 0x013D6
EPI_RinseTimeMM90LLLvl1 5079 0x013D7
EPI_RinseTimeMM90LLLvl2 5080 0x013D8
EPI_RinseTimeMM90LLLvl3 5081 0x013D9
EPI_RinseTimeMM90 5082 0x013DA
EPI_RinseTimeMM90HLLvl0 5083 0x013DB
EPI_RinseTimeMM90HLLvl3 5084 0x013DC
EPI_RinseTimeMM90HLLvl2 5085 0x013DD
EPI_RinseTimeMM90HLLvl1 5086 0x013DE
EPI_SnArtTimeLLLvl0 5087 0x013DF
EPI_SnArtTimeHLLvl0 5088 0x013E0
EPI_SnArtTimeLLLvl1 5089 0x013E1
EPI_SnArtTimeLLLvl2 5090 0x013E2


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 113 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_SnArtTimeLLLvl3 5091 0x013E3


EPI_SnArtTimeSet 5092 0x013E4
EPI_SnArtTimeHLLvl3 5093 0x013E5
EPI_SnArtTimeHLLvl2 5094 0x013E6
EPI_SnArtTimeHLLvl1 5095 0x013E7
EPI_SnVenTimeLLLvl0 5096 0x013E8
EPI_SnVenTimeHLLvl0 5097 0x013E9
EPI_SnVenTimeLLLvl1 5098 0x013EA
EPI_SnVenTimeLLLvl2 5099 0x013EB
EPI_SnVenTimeLLLvl3 5100 0x013EC
EPI_SnVenTimeSet 5101 0x013ED
EPI_SnVenTimeHLLvl3 5102 0x013EE
EPI_SnVenTimeHLLvl2 5103 0x013EF
EPI_SnVenTimeHLLvl1 5104 0x013F0
EPI_StrokeVolLim 5105 0x013F1
EPI_StrokeVolLimHlLvl0 5106 0x013F2
EPI_StrokeVolLimLlLvl0 5107 0x013F3
EPI_TempLLLvl0 5108 0x013F4
EPI_TempHLLvl0 5109 0x013F5
EPI_TempLLLvl1 5110 0x013F6
EPI_TempLLLvl2 5111 0x013F7
EPI_TempLLLvl3 5112 0x013F8
EPI_TempLL 5113 0x013F9
EPI_TEMPSET 5114 0x013FA
EPI_TempHL 5115 0x013FB
EPI_TempHLLvl3 5116 0x013FC
EPI_TempHLLvl2 5117 0x013FD
EPI_TempHLLvl1 5118 0x013FE
EPI_TimeBetweenCleaning 5119 0x013FF
EPI_TimeBetweenCleanLL 5120 0x01400
EPI_TimeBetweenCleanHL 5121 0x01401
EPI_TreatmBetweenDecalcBiCart 5122 0x01402
EPI_TreatmBetweenDecalcLiqBic 5123 0x01403
EPI_TreatmentsBetweenDecalcLL 5124 0x01404
EPI_TreatmentsBetweenDecalcHL 5125 0x01405
EPI_TmpHLInit 5126 0x01406
EPI_TmpHLLvl0 5127 0x01407
EPI_TmpIsolLowPart 5128 0x01408
EPI_TmpIsolWidth 5129 0x01409
EPI_TmpLLInit 5130 0x0140A
EPI_TmpLLLvl0 5131 0x0140B
EPI_TmpLowPart 5132 0x0140C
EPI_TmpLowPartHLLvl0 5133 0x0140D


114 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_TmpLowPartLLLvl0 5134 0x0140E


EPI_TmpWidth 5135 0x0140F
EPI_TmpWidthHLLvl0 5136 0x01410
EPI_TmpWidthLLLvl0 5137 0x01411
EPI_TMTOverheatLimLlLvl0 5138 0x01412
EPI_TMTOverheatLimHLLvl0 5139 0x01413
EPI_TMTOverheatLim 5140 0x01414
EPI_TMTOverheatLimHLLvl3 5141 0x01415
EPI_TMTOverheatLimHLLvl2 5142 0x01416
EPI_TMTOverheatLimHLLvl1 5143 0x01417
EPI_UFCalRefVolHL 5144 0x01418
EPI_UFCalRefVolLL 5145 0x01419
EPI_UfsStartVolAlarmHL 5146 0x0141A
EPI_UfsStartVolAlarmLL 5147 0x0141B
EPI_UfsStartVolAlarm 5148 0x0141C
EPI_UfsVolAlarmWideningRateHL 5149 0x0141D
EPI_UfsVolAlarmWideningRateLL 5150 0x0141E
EPI_UfsVolAlarmWideningRate 5151 0x0141F
EPI_UFRATEHLLVL0 5152 0x01420
EPI_UFRATEHLLVL1 5153 0x01421
EPI_UFRATEHLLVL2 5154 0x01422
EPI_UFRATEHLLVL3 5155 0x01423
EPI_UFRateLlLvl0 5156 0x01424
EPI_UFRATELLLVL1 5157 0x01425
EPI_UFRATELLLVL2 5158 0x01426
EPI_UFRATELLLVL3 5159 0x01427
EPI_UfRateHLCoeffLLLvl0 5160 0x01428
EPI_UfRateHLCoeffHLLvl0 5161 0x01429
EPI_UfRateHLCoeff 5162 0x0142A
EPI_UfRateHLCoeffHLLvl3 5163 0x0142B
EPI_UfRateHLCoeffHLLvl2 5164 0x0142C
EPI_UfRateHLCoeffHLLvl1 5165 0x0142D
EPI_UFRATESETDIFFINIT 5166 0x0142E
EPI_UFRATESETISOLINIT 5167 0x0142F
EPI_UFVOLSETDIFFINIT 5168 0x01430
EPI_UFVOLSETHLLVL0 5169 0x01431
EPI_UFVOLSETHLLVL1 5170 0x01432
EPI_UFVOLSETHLLVL2 5171 0x01433
EPI_UFVOLSETHLLVL3 5172 0x01434
EPI_UFVOLSETISOLINIT 5173 0x01435
EPI_RecirUFVolSetLLLVL0 5174 0x01436
EPI_RecirUFVolSetHLLVL0 5175 0x01437
EPI_RecirUFVolSetINIT 5176 0x01438


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 115 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_UFVOLSETLLLVL0 5177 0x01439


EPI_UFVOLSETLLLVL1 5178 0x0143A
EPI_UFVOLSETLLLVL2 5179 0x0143B
EPI_UFVOLSETLLLVL3 5180 0x0143C
EPI_VenPrCalHL 5181 0x0143D
EPI_VenPrCalLL 5182 0x0143E
EPI_VenPrCoeffHL 5183 0x0143F
EPI_VenPrCoeffLL 5184 0x01440
EPI_VenPrOffsetHL 5185 0x01441
EPI_VenPrOffsetLL 5186 0x01442
EPI_VprAutoLimAlarmTime 5187 0x01443
EPI_VprAutoLimAlarmTimeHLLvl0 5188 0x01444
EPI_VprAutoLimAlarmTimeLLLvl0 5189 0x01445
EPI_VprHLInit 5190 0x01446
EPI_VprHLLvl0 5191 0x01447
EPI_VprHLSnAutoLim 5192 0x01448
EPI_VprLLInit 5193 0x01449
EPI_VprLLLvl0 5194 0x0144A
EPI_VprLLSnAutoLim 5195 0x0144B
EPI_VprLowPart 5196 0x0144C
EPI_VprLowPartHLLvl0 5197 0x0144D
EPI_VprLowPartLLLvl0 5198 0x0144E
EPI_VprWidth 5199 0x0144F
EPI_VprWidthHLLvl0 5200 0x01450
EPI_VprWidthLLLvl0 5201 0x01451
EPI_WAdRefCoeffHl 5202 0x01452
EPI_WAdRefCoeffLl 5203 0x01453
EPI_AprHLPresetInit 5204 0x01454
EPI_AprHLPresetHigh 5205 0x01455
EPI_AprHLPresetLow 5206 0x01456
EPI_RecirDiaFlowLL 5207 0x01457
EPI_RecirDiaFlowHL 5208 0x01458
EPI_RecirDiaFlow 5209 0x01459
EPI_DialysisFlowLL 5210 0x0145A
EPI_DialysisFlowHL 5211 0x0145B
EPI_LowDialysisFlow 5212 0x0145C
EPI_CondDiffLimit 5213 0x0145D
EPI_CondDiffHlLimit 5214 0x0145E
EPI_CondDiffLlLimit 5215 0x0145F
EPI_VprMaxHL 5216 0x01460
EPI_VprMinLL 5217 0x01461
EPI_AprMaxHL 5218 0x01462
EPI_AprMinLL 5219 0x01463


116 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_TempMaxHL 5220 0x01464


EPI_TempMinLL 5221 0x01465
EPI_PHMaxHl 5222 0x01466
EPI_PHMinLl 5223 0x01467
EPI_MinUfRateHL 5224 0x01468
EPI_MinUfRateLL 5225 0x01469
EPI_BloodFlHighLim 5226 0x0146A
EPI_BPMDiastPrHL 5227 0x0146B
EPI_BPMDiastPrLL 5228 0x0146C
EPI_BPMDiastPrHigh 5229 0x0146D
EPI_BPMDiastPrLow 5230 0x0146E
EPI_BPMInitCuffPr 5231 0x0146F
EPI_BPMInitCuffPrHL 5232 0x01470
EPI_BPMInitCuffPrLL 5233 0x01471
EPI_BPMMeanPrHL 5234 0x01472
EPI_BPMMeanPrLL 5235 0x01473
EPI_BPMMeanPrHigh 5236 0x01474
EPI_BPMMeanPrLow 5237 0x01475
EPI_BPMMeasIntHL 5238 0x01476
EPI_BPMMeasIntLL 5239 0x01477
EPI_BPMMeasInt 5240 0x01478
EPI_BPMPulseHL 5241 0x01479
EPI_BPMPulseLL 5242 0x0147A
EPI_BPMPulseHigh 5243 0x0147B
EPI_BPMPulseLow 5244 0x0147C
EPI_BPMSystPrHL 5245 0x0147D
EPI_BPMSystPrLL 5246 0x0147E
EPI_BPMSystPrHigh 5247 0x0147F
EPI_BPMSystPrLow 5248 0x01480
EPI_BPMSystPrHighRange 5249 0x01481
EPI_BPMSystPrLowRange 5250 0x01482
EPI_BPMDiastPrHighRange 5251 0x01483
EPI_BPMDiastPrLowRange 5252 0x01484
EPI_BPMMeanPrHighRange 5253 0x01485
EPI_BPMMeanPrLowRange 5254 0x01486
EPI_BPMPulseHighRange 5255 0x01487
EPI_BPMPulseLowRange 5256 0x01488
EPI_ConcPumpSpeedFlushRecLL 5257 0x01489
EPI_ConcPumpSpeedFlushRecHL 5258 0x0148A
EPI_ConcPumpSpeedFlushRec 5259 0x0148B
EPI_FlushRecHPGPressLL 5260 0x0148C
EPI_FlushRecHPGPressHL 5261 0x0148D
EPI_FlushRecHPGPress 5262 0x0148E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 117 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_ProfHCO3StartHL 5263 0x0148F


EPI_ProfHCO3StartLL 5264 0x01490
EPI_ProfHCO3StopHL 5265 0x01491
EPI_ProfHCO3StopLL 5266 0x01492
EPI_ProfNaStartHL 5267 0x01493
EPI_ProfNaStartLL 5268 0x01494
EPI_ProfNaStopHL 5269 0x01495
EPI_ProfNaStopLL 5270 0x01496
EPI_ProfUFRateDiffHL 5271 0x01497
EPI_ProfUFRateDiffLL 5272 0x01498
EPI_ProfModel1HCO3Start 5273 0x01499
EPI_ProfModel1HCO3Stop 5274 0x0149A
EPI_ProfModel1NaAcStart 5275 0x0149B
EPI_ProfModel1NaAcStop 5276 0x0149C
EPI_ProfModel1NaStart 5277 0x0149D
EPI_ProfModel1NaStop 5278 0x0149E
EPI_ProfModel1UFRateDiff 5279 0x0149F
EPI_ProfModel2HCO3Start 5280 0x014A0
EPI_ProfModel2HCO3Stop 5281 0x014A1
EPI_ProfModel2NaAcStart 5282 0x014A2
EPI_ProfModel2NaAcStop 5283 0x014A3
EPI_ProfModel2NaStart 5284 0x014A4
EPI_ProfModel2NaStop 5285 0x014A5
EPI_ProfModel2UFRateDiff 5286 0x014A6
EPI_ProfModel3HCO3Start 5287 0x014A7
EPI_ProfModel3HCO3Stop 5288 0x014A8
EPI_ProfModel3NaAcStart 5289 0x014A9
EPI_ProfModel3NaAcStop 5290 0x014AA
EPI_ProfModel3NaStart 5291 0x014AB
EPI_ProfModel3NaStop 5292 0x014AC
EPI_ProfModel3UFRateDiff 5293 0x014AD
EPI_OvernightRinseDelay 5294 0x014AE
EPI_OvernightRinseDelayLL 5295 0x014AF
EPI_OvernightRinseDelayHL 5296 0x014B0
EPI_FlowSteps 5297 0x014B1
EPI_FlowStepsLL 5298 0x014B2
EPI_FlowStepsHL 5299 0x014B3
EPI_SetFlowLLLvl1 5300 0x014B4
EPI_SetFlowLLLvl2 5301 0x014B5
EPI_SetFlowLLLvl3 5302 0x014B6
EPI_SetFlowLL 5303 0x014B7
EPI_SetFlow 5304 0x014B8
EPI_SetFlowHL 5305 0x014B9


118 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_SetFlowHLLvl3 5306 0x014BA


EPI_SetFlowHLLvl2 5307 0x014BB
EPI_SetFlowHLLvl1 5308 0x014BC
EPI_ChemConcPercentLL 5309 0x014BD
EPI_ChemConcPercentHL 5310 0x014BE
EPI_ChemConcPercentAlt1 5311 0x014BF
EPI_ChemConcPercentAlt2 5312 0x014C0
EPI_ChemConcPercentAlt3 5313 0x014C1
EPI_ChemConcPercentAlt6 5314 0x014C2
EPI_ProfUFIntervalsHL 5315 0x014C3
EPI_ProfUFIntervalsLL 5316 0x014C4
EPI_ProfModel1UFIntervals 5317 0x014C5
EPI_ProfModel2UFIntervals 5318 0x014C6
EPI_ProfModel3UFIntervals 5319 0x014C7
EPI_ConcStandbyFch 5320 0x014C8
EPI_ConcStandbyHL 5321 0x014C9
EPI_ConcStandbyLL 5322 0x014CA
EPI_ConcStandbyPrim 5323 0x014CB
EPI_BPMInflationMode 5324 0x014CC
EPI_BPMInflationModeLL 5325 0x014CD
EPI_BPMInflationModeHL 5326 0x014CE
EPI_DiascanDistrVolume 5327 0x014CF
EPI_DiascanDistrVolumeHL 5328 0x014D0
EPI_DiascanDistrVolumeLL 5329 0x014D1
EPI_DiascanDoseMax 5330 0x014D2
EPI_DiascanDoseMaxHL 5331 0x014D3
EPI_DiascanDoseMaxLL 5332 0x014D4
EPI_DiascanDoseMin 5333 0x014D5
EPI_DiascanDoseMinHL 5334 0x014D6
EPI_DiascanDoseMinLL 5335 0x014D7
EPI_DiascanKAlarmLimit 5336 0x014D8
EPI_DiascanKAlarmLimitHL 5337 0x014D9
EPI_DiascanKAlarmLimitLL 5338 0x014DA
EPI_DiascanMaxStep 5339 0x014DB
EPI_DiascanMaxStepHL 5340 0x014DC
EPI_DiascanMaxStepLL 5341 0x014DD
EPI_DiascanMeasMode 5342 0x014DE
EPI_DiascanMeasModeHL 5343 0x014DF
EPI_DiascanMeasModeLL 5344 0x014E0
EPI_DiascanMinCdi 5345 0x014E1
EPI_DiascanMinCdiHL 5346 0x014E2
EPI_DiascanMinCdiLL 5347 0x014E3
EPI_DiascanStepDuration 5348 0x014E4


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 119 (381)
Program version 1.1.0
C EN 2015-03-26

EPI_DiascanStepDurationLL 5349 0x014E5


EPI_DiascanStepDurationHL 5350 0x014E6
EPI_DiascanStepHeight 5351 0x014E7
EPI_DiascanStepHeightHL 5352 0x014E8
EPI_DiascanStepHeightLL 5353 0x014E9
EPI_DiascanTarget 5354 0x014EA
EPI_DiascanTargetHL 5355 0x014EB
EPI_DiascanTargetLL 5356 0x014EC
EPI_DiascanTargetAlarm 5357 0x014ED
EPI_DiascanTargetAlarmHL 5358 0x014EE
EPI_DiascanTargetAlarmLL 5359 0x014EF
EPI_DiascanWatsonAgeLL 5360 0x014F0
EPI_DiascanWatsonHL 5361 0x014F1
EPI_DiascanWatsonLL 5362 0x014F2
EPI_LowFlowHeatPressLL 5363 0x014F3
EPI_LowFlowHeatPressHL 5364 0x014F4
EPI_LowFlowHeatPress 5365 0x014F5
EPI_InstRTCTypeLL 5366 0x014F6
EPI_InstRTCTypeHL 5367 0x014F7
EPI_InstRTCType 5368 0x014F8
EPI_LoudSpeakerVolumeLL 5369 0x014F9
EPI_LoudSpeakerVolumeHL 5370 0x014FA
EPI_LoudSpeakerVolume 5371 0x014FB
EPI_EXTRARBVOLSET 5372 0x014FC
EPI_ExtraRbVolSetHl 5373 0x014FD
EPI_ExtraRbVolSetLl 5374 0x014FE
EPI_AprMargin 5375 0x014FF
EPI_AprMarginHLLvl0 5376 0x01500
EPI_AprMarginLLLvl0 5377 0x01501
EPI_HPGMaxDisinfLimit 5378 0x01502
EPI_HPGMaxDisinfLimitHL 5379 0x01503
EPI_HPGMaxDisinfLimitLL 5380 0x01504
EPI_DaysBetweenUfdReplaceLim 5381 0x01505
EPI_DaysBetweenUfdReplaceLimLL 5382 0x01506
EPI_DaysBetweenUfdReplaceLimHL 5383 0x01507
EPI_DisinfCountUfdReplaceLim 5384 0x01508
EPI_DisinfCountUfdReplaceLimLL 5385 0x01509
EPI_DisinfCountUfdReplaceLimHL 5386 0x0150A
EPI_HypoCountUfdReplaceLim 5387 0x0150B
EPI_HypoCountUfdReplaceLimLL 5388 0x0150C
EPI_HypoCountUfdReplaceLimHL 5389 0x0150D
EPI_RinseClampOpenTime 5390 0x0150E
EPI_RinseClampOpenTimeLL 5391 0x0150F


120 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPI_RinseClampOpenTimeHL 5392 0x01510


EPI_RinseClampCloseTime 5393 0x01511
EPI_RinseClampCloseTimeLL 5394 0x01512
EPI_RinseClampCloseTimeHL 5395 0x01513
EPI_ZevaPressReleaseLim 5396 0x01514
EPI_ZevaPressReleaseLimLL 5397 0x01515
EPI_ZevaPressReleaseLimHL 5398 0x01516
EPI_PDSupLimit 5399 0x01517
EPI_PDSupLimitLL 5400 0x01518
EPI_PDSupLimitHL 5401 0x01519
EPI_ExtraRecirUfVolSetLlLvl0 5402 0x0151A
EPI_ExtraRecirUfVolSetHlLvl0 5403 0x0151B
EPI_ExtraRecirUfVolSetInit 5404 0x0151C
EPI_ExtraRecirUFTimeLLLvl0 5405 0x0151D
EPI_ExtraRecirUFTimeHLLvl0 5406 0x0151E
EPI_ExtraRecirTimeInit 5407 0x0151F
EPI_ExtraRecirUFRLLLvl0 5408 0x01520
EPI_ExtraRecirUFRHLLvl0 5409 0x01521
EPI_ExtraRecirUFRInit 5410 0x01522
EPI_NoOfPassedFch 5411 0x01523
EPI_NoOfFirstTarFailed 5412 0x01524
EPI_TreatmentOverview 5413 0x01525
EPI_TreatmentOverviewLL 5414 0x01526
EPI_TreatmentOverviewHL 5415 0x01527
EPI_TempSetLL 5416 0x01528
EPI_NoOfStepsInUFSEvalPhase 5417 0x01529
EPI_UFVOLSETHLLVL4 5418 0x0152A
EPL_dummy 5419 0x0152B
EPL_FALSE 5420 0x0152C
EPL_TRUE 5421 0x0152D
EPL_WarningTextActive 5422 0x0152E
EPR_dummy 5423 0x0152F
EPR_ACaAlt1 5424 0x01530
EPR_ACH3COOAlt1 5425 0x01531
EPR_AClAlt1 5426 0x01532
EPR_AHAlt1 5427 0x01533
EPR_AKAlt1 5428 0x01534
EPR_AMgAlt1 5429 0x01535
EPR_ANaAlt1 5430 0x01536
EPR_ACaCl2Alt1 5431 0x01537
EPR_ACaCl2H2OAlt1 5432 0x01538
EPR_ACH3COOHAlt1 5433 0x01539
EPR_ACH3COONaAlt1 5434 0x0153A


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 121 (381)
Program version 1.1.0
C EN 2015-03-26

EPR_ACH3COONaH2OAlt1 5435 0x0153B


EPR_AKClAlt1 5436 0x0153C
EPR_AMgCl2Alt1 5437 0x0153D
EPR_AMgCl2H2OAlt1 5438 0x0153E
EPR_ANaClAlt1 5439 0x0153F
EPR_BCH3COOAlt1 5440 0x01540
EPR_BClAlt1 5441 0x01541
EPR_BNaAlt1 5442 0x01542
EPR_BCH3COONaAlt1 5443 0x01543
EPR_BCH3COONaH2OAlt1 5444 0x01544
EPR_BHCO3Alt1 5445 0x01545
EPR_BNaHCO3Alt1 5446 0x01546
EPR_BNaClAlt1 5447 0x01547
EPR_CaAlt1 5448 0x01548
EPR_CH3COOAlt1 5449 0x01549
EPR_ClAlt1 5450 0x0154A
EPR_KAlt1 5451 0x0154B
EPR_MgAlt1 5452 0x0154C
EPR_NaAlt1 5453 0x0154D
EPR_CaCl2Alt1 5454 0x0154E
EPR_CaCl2H2OAlt1 5455 0x0154F
EPR_CH3COONaAlt1 5456 0x01550
EPR_CH3COONaH2OAlt1 5457 0x01551
EPR_KClAlt1 5458 0x01552
EPR_MgCl2Alt1 5459 0x01553
EPR_MgCl2H2OAlt1 5460 0x01554
EPR_NaClAlt1 5461 0x01555
EPR_ACaAlt2 5462 0x01556
EPR_ACH3COOAlt2 5463 0x01557
EPR_AClAlt2 5464 0x01558
EPR_AHAlt2 5465 0x01559
EPR_AKAlt2 5466 0x0155A
EPR_AMgAlt2 5467 0x0155B
EPR_ANaAlt2 5468 0x0155C
EPR_ACaCl2Alt2 5469 0x0155D
EPR_ACaCl2H2OAlt2 5470 0x0155E
EPR_ACH3COOHAlt2 5471 0x0155F
EPR_ACH3COONaAlt2 5472 0x01560
EPR_ACH3COONaH2OAlt2 5473 0x01561
EPR_AKClAlt2 5474 0x01562
EPR_AMgCl2Alt2 5475 0x01563
EPR_AMgCl2H2OAlt2 5476 0x01564
EPR_ANaClAlt2 5477 0x01565


122 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPR_BCH3COOAlt2 5478 0x01566


EPR_BClAlt2 5479 0x01567
EPR_BNaAlt2 5480 0x01568
EPR_BCH3COONaAlt2 5481 0x01569
EPR_BCH3COONaH2OAlt2 5482 0x0156A
EPR_BHCO3Alt2 5483 0x0156B
EPR_BNaHCO3Alt2 5484 0x0156C
EPR_BNaClAlt2 5485 0x0156D
EPR_CaAlt2 5486 0x0156E
EPR_CH3COOAlt2 5487 0x0156F
EPR_ClAlt2 5488 0x01570
EPR_KAlt2 5489 0x01571
EPR_MgAlt2 5490 0x01572
EPR_NaAlt2 5491 0x01573
EPR_CaCl2Alt2 5492 0x01574
EPR_CaCl2H2OAlt2 5493 0x01575
EPR_CH3COONaAlt2 5494 0x01576
EPR_CH3COONaH2OAlt2 5495 0x01577
EPR_KClAlt2 5496 0x01578
EPR_MgCl2Alt2 5497 0x01579
EPR_MgCl2H2OAlt2 5498 0x0157A
EPR_NaClAlt2 5499 0x0157B
EPR_ACaAlt3 5500 0x0157C
EPR_ACH3COOAlt3 5501 0x0157D
EPR_AClAlt3 5502 0x0157E
EPR_AHAlt3 5503 0x0157F
EPR_AKAlt3 5504 0x01580
EPR_AMgAlt3 5505 0x01581
EPR_ANaAlt3 5506 0x01582
EPR_ACaCl2Alt3 5507 0x01583
EPR_ACaCl2H2OAlt3 5508 0x01584
EPR_ACH3COOHAlt3 5509 0x01585
EPR_ACH3COONaAlt3 5510 0x01586
EPR_ACH3COONaH2OAlt3 5511 0x01587
EPR_AKClAlt3 5512 0x01588
EPR_AMgCl2Alt3 5513 0x01589
EPR_AMgCl2H2OAlt3 5514 0x0158A
EPR_ANaClAlt3 5515 0x0158B
EPR_BCH3COOAlt3 5516 0x0158C
EPR_BClAlt3 5517 0x0158D
EPR_BNaAlt3 5518 0x0158E
EPR_BCH3COONaAlt3 5519 0x0158F
EPR_BCH3COONaH2OAlt3 5520 0x01590


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 123 (381)
Program version 1.1.0
C EN 2015-03-26

EPR_BHCO3Alt3 5521 0x01591


EPR_BNaHCO3Alt3 5522 0x01592
EPR_BNaClAlt3 5523 0x01593
EPR_CaAlt3 5524 0x01594
EPR_CH3COOAlt3 5525 0x01595
EPR_ClAlt3 5526 0x01596
EPR_KAlt3 5527 0x01597
EPR_MgAlt3 5528 0x01598
EPR_NaAlt3 5529 0x01599
EPR_CaCl2Alt3 5530 0x0159A
EPR_CaCl2H2OAlt3 5531 0x0159B
EPR_CH3COONaAlt3 5532 0x0159C
EPR_CH3COONaH2OAlt3 5533 0x0159D
EPR_KClAlt3 5534 0x0159E
EPR_MgCl2Alt3 5535 0x0159F
EPR_MgCl2H2OAlt3 5536 0x015A0
EPR_NaClAlt3 5537 0x015A1
EPR_CH1RCoeffHL 5538 0x015A2
EPR_CH1RCoeffLL 5539 0x015A3
EPR_CH1ROffsetHL 5540 0x015A4
EPR_CH1ROffsetLL 5541 0x015A5
EPR_CH2RCoeffHL 5542 0x015A6
EPR_CH2RCoeffLL 5543 0x015A7
EPR_CH2ROffsetHL 5544 0x015A8
EPR_CH2ROffsetLL 5545 0x015A9
EPR_UFROffsetHL 5546 0x015AA
EPR_UFROffsetLL 5547 0x015AB
EPR_UFRToleranceHL 5548 0x015AC
EPR_UFRToleranceLL 5549 0x015AD
EPR_TenThousand 5550 0x015AE
EPR_Zero 5551 0x015AF
EPW_DummyTo 5552 0x015B0
EPW_PutteLL 5553 0x015B1
EPW_PutteHL 5554 0x015B2
EPW_Putte 5555 0x015B3
EPW_AirEvacTo 5556 0x015B4
EPW_ArtClClosErrCntTo 5557 0x015B5
EPW_ArtClClosRtrTo 5558 0x015B6
EPW_ArtClOpenErrCntTo 5559 0x015B7
EPW_ArtClOpenRtrTo 5560 0x015B8
EPW_ArtClPeckTo 5561 0x015B9
EPW_AutoOffAfterDisTime 5562 0x015BA
EPW_AutoOffAfterDisTimeHL 5563 0x015BB


124 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPW_AutoOffAfterDisTimeLL 5564 0x015BC


EPW_BcFillTo 5565 0x015BD
EPW_BeforeAIVATo 5566 0x015BE
EPW_BiCartEmptyWaitTo 5567 0x015BF
EPW_BiCartExtraFlushTo 5568 0x015C0
EPW_BiCartExtraFlushToLL 5569 0x015C1
EPW_BiCartExtraFlushToHL 5570 0x015C2
EPW_BiCartFlushTo 5571 0x015C3
EPW_BiCartLatchClosedTo 5572 0x015C4
EPW_BiCartPrimTo 5573 0x015C5
EPW_BiCPrimCount 5574 0x015C6
EPW_BicPrimWait4flow 5575 0x015C7
EPW_BiCStartUpNormFlow 5576 0x015C8
EPW_BicartFlvaCounter 5577 0x015C9
EPW_BicLatchAttnTo 5578 0x015CA
EPW_BlkDilFillupTo 5579 0x015CB
EPW_BPumpFillupTo 5580 0x015CC
EPW_BlPumpStopCounter 5581 0x015CD
EPW_BPMClearTimerTo 5582 0x015CE
EPW_BPMComRetryTimer 5583 0x015CF
EPW_BPMComFailTimer 5584 0x015D0
EPW_BPMComRetryTmtTimer 5585 0x015D1
EPW_BPMContModeTimer 5586 0x015D2
EPW_BPMNoCuffTo 5587 0x015D3
EPW_BypDisTo 5588 0x015D4
EPW_ByvaDilFillupTo 5589 0x015D5
EPW_ByvaFillupTo 5590 0x015D6
EPW_CCAttachTo 5591 0x015D7
EPW_CCCondCheckTo 5592 0x015D8
EPW_CCEmptyTo 5593 0x015D9
EPW_CCEmptyRinseTo 5594 0x015DA
EPW_CCFivaPathFillTo 5595 0x015DB
EPW_CCFlowTo 5596 0x015DC
EPW_CCNormalFlowTo 5597 0x015DD
EPW_CCOutletFillTo 5598 0x015DE
EPW_CCPressureTestTo 5599 0x015DF
EPW_CCHeatUpTo 5600 0x015E0
EPW_CCRivaPathFillTo 5601 0x015E1
EPW_ChemDilFillupCount 5602 0x015E2
EPW_ChemFillupCount 5603 0x015E3
EPW_CeChemFillup15SecTo 5604 0x015E4
EPW_ChemRinseTimeLL 5605 0x015E5
EPW_ChemRinseTimeUfdLL 5606 0x015E6


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 125 (381)
Program version 1.1.0
C EN 2015-03-26

EPW_ChemRinseTimeHL 5607 0x015E7


EPW_ChemRinseTimeAlt1 5608 0x015E8
EPW_ChemRinseTimeAlt2 5609 0x015E9
EPW_ChemRinseTimeAlt3 5610 0x015EA
EPW_ChemRinseTimeAlt6 5611 0x015EB
EPW_ChemRinseTimeAlt7 5612 0x015EC
EPW_ChemRinseTimeUfdAlt1 5613 0x015ED
EPW_ChemRinseTimeUfdAlt2 5614 0x015EE
EPW_ChemRinseTimeUfdAlt3 5615 0x015EF
EPW_ChemRinseTimeUfdAlt6 5616 0x015F0
EPW_ChemRinseTimeUfdAlt7 5617 0x015F1
EPW_CirculationCount 5618 0x015F2
EPW_CirculationDrvaTo 5619 0x015F3
EPW_CirculationMainTo 5620 0x015F4
EPW_CirculationTo 5621 0x015F5
EPW_ContinueTo 5622 0x015F6
EPW_DelayedDiaConDia 5623 0x015F7
EPW_DegFillTo 5624 0x015F8
EPW_DrainByvaTo 5625 0x015F9
EPW_DrainDivaRinse10UfdTo 5626 0x015FA
EPW_DrainDivaTo 5627 0x015FB
EPW_DrainDivaUfdTo 5628 0x015FC
EPW_DrainEvvaTo 5629 0x015FD
EPW_DrainMainLineUfdTo 5630 0x015FE
EPW_DrainMainLineTo 5631 0x015FF
EPW_DrainMixingChamberATo 5632 0x01600
EPW_DrainRevaBlkRinse10UfdTo 5633 0x01601
EPW_DrainRevaBlkTo 5634 0x01602
EPW_DrainRevaBlkUfdTo 5635 0x01603
EPW_DrainRevaLineTo 5636 0x01604
EPW_DrainRivaTo 5637 0x01605
EPW_DrainStartTo 5638 0x01606
EPW_DrainTavaUfdTo 5639 0x01607
EPW_DrainZevaTo 5640 0x01608
EPW_Drain5SecTo 5641 0x01609
EPW_Drain20SecTo 5642 0x0160A
EPW_Drain30SecTo 5643 0x0160B
EPW_Drain150SecTo 5644 0x0160C
EPW_DwellRecirc2To 5645 0x0160D
EPW_DwellTimeLL 5646 0x0160E
EPW_DwellTimeHL 5647 0x0160F
EPW_DwellTimeAlt1 5648 0x01610
EPW_DwellTimeAlt2 5649 0x01611


126 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPW_DwellTimeAlt3 5650 0x01612


EPW_DwellTimeAlt6 5651 0x01613
EPW_DwellTimeAlt7 5652 0x01614
EPW_DwellTimeHlLvl0 5653 0x01615
EPW_EmptyRevaTo 5654 0x01616
EPW_EmptyRivaTo 5655 0x01617
EPW_EvlAcAltSetHoldDelay 5656 0x01618
EPW_EvlAcAltSetWaitDelay 5657 0x01619
EPW_EvlAcAltWaitRqDelay 5658 0x0161A
EPW_EvlAirDetAttnLatch 5659 0x0161B
EPW_EvlAutoPrimMode 5660 0x0161C
EPW_EvlBcAltSetHoldDelay 5661 0x0161D
EPW_EvlBcAltSetWaitDelay 5662 0x0161E
EPW_EvlBcAltWaitRqDelay 5663 0x0161F
EPW_Ev1CompleteUFTaration 5664 0x01620
EPW_Ev2CompleteUFTaration 5665 0x01621
EPW_EvlCondAlarmLatch 5666 0x01622
EPW_EvlCondSetHoldDelay 5667 0x01623
EPW_EvlCondSetWaitDelay 5668 0x01624
EPW_EvlCondWaitRqDelay 5669 0x01625
EPW_EvlDegassingLevel 5670 0x01626
EPW_EvlDegassingLevelUfd 5671 0x01627
EPW_EvlDrainRefVol 5672 0x01628
EPW_EvlFluidByp 5673 0x01629
EPW_UfModeStartRq 5674 0x0162A
EPW_EvlInfoDispHoldControl 5675 0x0162B
EPW_EvlPumpDutyTime 5676 0x0162C
EPW_EvlRinseKeyDelay 5677 0x0162D
EPW_EvlTempAlarmLatch 5678 0x0162E
EPW_EvlUfCalibPrep 5679 0x0162F
EPW_EvlVprLimitDelay 5680 0x01630
EPW_EvlWarmStart 5681 0x01631
EPW_EvlWarningText 5682 0x01632
EPW_Evl1UfCalibPrepComplete 5683 0x01633
EPW_Evl2UfCalibPrepComplete 5684 0x01634
EPW_Evl3UfCalibPrepComplete 5685 0x01635
EPW_Evl4UfCalibPrepComplete 5686 0x01636
EPW_Evl5UfCalibPrepComplete 5687 0x01637
EPW_Evl6UfCalibPrepComplete 5688 0x01638
EPW_Evl7UfCalibPrepComplete 5689 0x01639
EPW_Evl7UfCalibPrepComplUfd 5690 0x0163A
EPW_Evl8UfCalibPrepComplete 5691 0x0163B
EPW_Evl8UfCalibPrepComplUfd 5692 0x0163C


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 127 (381)
Program version 1.1.0
C EN 2015-03-26

EPW_Evl9UfCalibPrepComplete 5693 0x0163D


EPW_Evl1UfCh1FlowCal 5694 0x0163E
EPW_Evl1UfDiffCal 5695 0x0163F
EPW_Evl2UfDiffCal 5696 0x01640
EPW_Evl1UfIsolVerify 5697 0x01641
EPW_Evl1UfMainflOffset 5698 0x01642
EPW_Evl2UfMainflOffset 5699 0x01643
EPW_Evl1UfVerify 5700 0x01644
EPW_EvvaDilFillupUfdTo 5701 0x01645
EPW_EvvaFillupUfdTo 5702 0x01646
EPW_EvvaDilFillupTo 5703 0x01647
EPW_EvvaFillupTo 5704 0x01648
EPW_EvvaOpenDelayTO 5705 0x01649
EPW_ExpChamberFillupTo 5706 0x0164A
EPW_FillPickUpTubeTo 5707 0x0164B
EPW_FillupCheckCount 5708 0x0164C
EPW_FillupCheck1To 5709 0x0164D
EPW_FillupCheck2To 5710 0x0164E
EPW_FilterFillupTO 5711 0x0164F
EPW_FinalCompTo 5712 0x01650
EPW_FinalFillupTo 5713 0x01651
EPW_FlowThrottleCleanTo 5714 0x01652
EPW_FlowThrottleFillupTo 5715 0x01653
EPW_FluidPrepTo 5716 0x01654
EPW_FlushAttnDelayTo 5717 0x01655
EPW_FlushMixingChamberATo 5718 0x01656
EPW_FlushMixingChamberAUfdTo 5719 0x01657
EPW_FlushEvvaTo 5720 0x01658
EPW_FlushEvvaUfdTo 5721 0x01659
EPW_FlushRampMixingChamberATo 5722 0x0165A
EPW_FlushRevaTo 5723 0x0165B
EPW_FlushRivaStartTo 5724 0x0165C
EPW_FlushRivaTo 5725 0x0165D
EPW_FlushSmothMixingChamberATo 5726 0x0165E
EPW_FlushSmothMixChamberAUfdTo 5727 0x0165F
EPW_FlushTubeTo 5728 0x01660
EPW_FlvaFillupTo 5729 0x01661
EPW_HeatBubbleTrapFillTo 5730 0x01662
EPW_HeatByvaTo 5731 0x01663
EPW_HeatCoolOffTo 5732 0x01664
EPW_HeatCoolOffToLL 5733 0x01665
EPW_HeatCoolOffToHL 5734 0x01666
EPW_HeatFlushCycleTo 5735 0x01667


128 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPW_HeatFlushCycleAlt1ToLL 5736 0x01668


EPW_HeatFlushCycleAlt2ToLL 5737 0x01669
EPW_HeatFlushCycleAlt3ToLL 5738 0x0166A
EPW_HeatFlushCycleToHL 5739 0x0166B
EPW_HeaterOnDelayTo 5740 0x0166C
EPW_HeaterBypassTo 5741 0x0166D
EPW_HeatEvva1To 5742 0x0166E
EPW_HeatEvva2To 5743 0x0166F
EPW_HeatFlushNoInPressTo 5744 0x01670
EPW_HeatingOffTo 5745 0x01671
EPW_HeatNoInPressDelayTo 5746 0x01672
EPW_HeatNoInPressTo 5747 0x01673
EPW_HeatPressRegMeasTo 5748 0x01674
EPW_HeatRecirc1To 5749 0x01675
EPW_HeatRecirc2Ch2To 5750 0x01676
EPW_HeatRecirc2MainTo 5751 0x01677
EPW_HeatRecirc2MainUfdTo 5752 0x01678
EPW_HeatRecirc2To 5753 0x01679
EPW_HeatRivaTo 5754 0x0167A
EPW_HeatPumpVoltMeasTo 5755 0x0167B
EPW_HeatTavaTo 5756 0x0167C
EPW_HeatZevaTo 5757 0x0167D
EPW_HpgAttnTo 5758 0x0167E
EPW_InletFlushTo 5759 0x0167F
EPW_InPressAlarmTo 5760 0x01680
EPW_InPressDisinfAlarmTo 5761 0x01681
EPW_MajorAirleakAttnCo 5762 0x01682
EPW_LogIndex0 5763 0x01683
EPW_LogIndex1 5764 0x01684
EPW_LogIndex2 5765 0x01685
EPW_MajorAirleakAttnTo 5766 0x01686
EPW_MajorAirleakAttnOff 5767 0x01687
EPW_NegUFKeyTo 5768 0x01688
EPW_NoDilution1To 5769 0x01689
EPW_NoDilution2To 5770 0x0168A
EPW_PrimLowRateTo 5771 0x0168B
EPW_RestrictorDisTo 5772 0x0168C
EPW_RevaDilFillupTo 5773 0x0168D
EPW_RevaFillupTo 5774 0x0168E
EPW_RevaPathFillTo 5775 0x0168F
EPW_RinseBefHeatTo 5776 0x01690
EPW_RinseBefHeatUfdTo 5777 0x01691
EPW_RinseByvaTo 5778 0x01692


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 129 (381)
Program version 1.1.0
C EN 2015-03-26

EPW_RinseEvvaTo 5779 0x01693


EPW_RinseRivaTo 5780 0x01694
EPW_RinseRevaTo 5781 0x01695
EPW_Rinse10SecTo 5782 0x01696
EPW_Rinse20SecTo 5783 0x01697
EPW_Rinse30SecTo 5784 0x01698
EPW_Rinse50SecTo 5785 0x01699
EPW_Rinse60SecTo 5786 0x0169A
EPW_Rinse120SecTo 5787 0x0169B
EPW_RivaDilFillupTo 5788 0x0169C
EPW_RivaFillupTo 5789 0x0169D
EPW_SagsDisTo 5790 0x0169E
EPW_ShutOffCoolingTo 5791 0x0169F
EPW_StickAAttnTo 5792 0x016A0
EPW_StickBAttnTo 5793 0x016A1
EPW_TarValveDisTo 5794 0x016A2
EPW_TimeBetweenDisinf 5795 0x016A3
EPW_TimeBetweenDisinfLL 5796 0x016A4
EPW_TimeBetweenDisinfHL 5797 0x016A5
EPW_UFDisplayResult1delay 5798 0x016A6
EPW_UFDisplayResult2delay 5799 0x016A7
EPW_UfdPrimeCounter 5800 0x016A8
EPW_UfdPrimeInDelay 5801 0x016A9
EPW_UfdPrimeOutDelay 5802 0x016AA
EPW_UfdPrimeInsideTo 5803 0x016AB
EPW_UfdPrimeOutsideTo 5804 0x016AC
EPW_UFStopTimeTo 5805 0x016AD
EPW_UFStopTimeToHl 5806 0x016AE
EPW_UFStopTimeToLl 5807 0x016AF
EPW_UFTarHPGPrStab 5808 0x016B0
EPW_UFTarHPGPrStabWaitTo 5809 0x016B1
EPW_UFTarRIVAOpenTO 5810 0x016B2
EPW_EvlWaitWhenEnterService 5811 0x016B3
EPW_WaitForPTarationTo 5812 0x016B4
EPW_ApWaitForFlDirectTo 5813 0x016B5
EPW_ZevaDilFillupTo 5814 0x016B6
EPW_ZevaFillupTo 5815 0x016B7
EPW_LowDialysisFlowWaitTo 5816 0x016B8
EPW_HighDialysisFlowWaitTo 5817 0x016B9
EPW_LDFDelayPostTarTo 5818 0x016BA
EPW_WaitForTMPStableTo 5819 0x016BB
EPW_TempTimerTo 5820 0x016BC
EPW_HeatCCEmptyTo 5821 0x016BD


130 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPW_HeatLowInletPressDelayTo 5822 0x016BE


EPW_VarFlowRegIneffTo 5823 0x016BF
EPW_VarFlowRegEffTo 5824 0x016C0
EPW_VarFlowRegDelayPostTarTo 5825 0x016C1
EPW_MachineOffDelayWaitTo 5826 0x016C2
EPW_ConcStandbyModeTo 5827 0x016C3
EPW_PrimRateCtrlWaitTo 5828 0x016C4
EPW_Diascan30MinTimer 5829 0x016C5
EPW_Diascan60MinTimer 5830 0x016C6
EPW_DiascanMeasTimer 5831 0x016C7
EPW_DiascanPreMeasTimer 5832 0x016C8
EPW_HpgStableTo 5833 0x016C9
EPW_WRO300HColdPrepTimeLL 5834 0x016CA
EPW_WRO300HColdPrepTimeHL 5835 0x016CB
EPW_WRO300HColdPrepTime 5836 0x016CC
EPW_WRO300HLFHPrepTimeLL 5837 0x016CD
EPW_WRO300HLFHPrepTimeHL 5838 0x016CE
EPW_WRO300HLFHPrepTime 5839 0x016CF
EPW_W300SelfHeatCoolTimeLL 5840 0x016D0
EPW_W300SelfHeatCoolTimeHL 5841 0x016D1
EPW_W300SelfHeatCoolTime 5842 0x016D2
EPW_WaterDistrHeatDelayTimeLL 5843 0x016D3
EPW_WaterDistrHeatDelayTimeHL 5844 0x016D4
EPW_WaterDistrHeatDelayTime 5845 0x016D5
EPW_LowFlowHeatTimeLL 5846 0x016D6
EPW_LowFlowHeatTimeHL 5847 0x016D7
EPW_LowFlowHeatTime 5848 0x016D8
EPW_StartLFHModeTimeoutLL 5849 0x016D9
EPW_StartLFHModeTimeoutHL 5850 0x016DA
EPW_StartLFHModeTimeout 5851 0x016DB
EPW_InitialINVAOpenDelayLL 5852 0x016DC
EPW_InitialINVAOpenDelayHL 5853 0x016DD
EPW_InitialINVAOpenDelay 5854 0x016DE
EPW_WRO300HQuestionTime 5855 0x016DF
EPW_BPMContModeDelayTimer 5856 0x016E0
EPW_ShowWelcomeTo 5857 0x016E1
EPW_OnOffBtnPresRecentlyTo 5858 0x016E2
EPW_EvlDisinfectionKeyDelay 5859 0x016E3
EPW_IdleTOMenuTime 5860 0x016E4
EPW_IdleSettingTime 5861 0x016E5
EPW_EvlEmptyBicartMode 5862 0x016E6
EPW_WROConfirmTo 5863 0x016E7
EPW_ThreeSecDelayTo 5864 0x016E8


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 131 (381)
Program version 1.1.0
C EN 2015-03-26

EPW_EmptyCCATo 5865 0x016E9


EPW_CSBM_AlarmTo 5866 0x016EA
EPW_CSBM_FluidPathOkTo 5867 0x016EB
EPW_InactivateADPrimTo 5868 0x016EC
EPW_ReleaseBlLockTo 5869 0x016ED
EPW_DiascanDeactivateTo 5870 0x016EE
EPW_ConPatRecTo 5871 0x016EF
EPW_RecirRecTo 5872 0x016F0
EPW_SettingDelayTo 5873 0x016F1
EPW_DiascanDelayTo 5874 0x016F2
EPW_DiascanNotPossibleTo 5875 0x016F3
EPW_DiascanPausedTo 5876 0x016F4
EPW_DiascanDistrDelayTo 5877 0x016F5
EPW_ProfParamUFDelayTo 5878 0x016F6
EPW_DiscontinueAckTo 5879 0x016F7
EPW_BackFlushUfdTo 5880 0x016F8
EPW_HeatRecirc2ByvaZevaTo 5881 0x016F9
EPW_FillupCoolOffTo 5882 0x016FA
EPW_ZevaFillCitricUfdTo 5883 0x016FB
EPW_ByvaFillCitricUfdTo 5884 0x016FC
EPW_ExpChambFillCitricUfdTo 5885 0x016FD
EPW_CCOutletFillUfdTo 5886 0x016FE
EPW_StartFlushRevaCount 5887 0x016FF
EPW_BackFlushUFDDrainTo 5888 0x01700
EPW_PressureCoolDownTo 5889 0x01701
EPW_PressureCoolDownTo2 5890 0x01702
EPW_HepPumpBolusStopTimer 5891 0x01703
EPW_CCPrimCoolDownTo 5892 0x01704
EPW_CCHeatCoolDownTo 5893 0x01705
EPW_SagsTestAttnTo 5894 0x01706
EPW_SagsTestDrainAttnTo 5895 0x01707
EPW_RinseBLineDrainBTo 5896 0x01708
EPW_RinseBLineDrainInitTo 5897 0x01709
EPW_HeatFlushNoInPressDelayTo 5898 0x0170A
EPW_MaxUFTarAttempts 5899 0x0170B
EPW_BPMNewMeasDelayTimer 5900 0x0170C
EPW_BypassDuringRecirTo 5901 0x0170D
EPW_DiascanUFTimerTo 5902 0x0170E
EPW_HeatRevaFillAutoDisTo 5903 0x0170F
EPW_HeatUFDFillAutoDisTo 5904 0x01710
EPW_FillFlvaTo 5905 0x01711
EPW_TAVACloseDelayTimerTo 5906 0x01712
EPW_HpgNorm2AttnTo 5907 0x01713


132 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

EPW_WaitUfTarPreProcRdyTo 5908 0x01714


EPW_WaitPostUfTarTo 5909 0x01715
EPW_RinseBefHeatNoInPrTo 5910 0x01716
EPW_DelayCondAlarmUFSFchTo 5911 0x01717
EPW_WaitForStableFluidTo 5912 0x01718
EPW_DelaySelectProcTo 5913 0x01719
EPW_CeFinalFillupTo 5914 0x0171A
EPW_PrepareWaitForStickTo 5915 0x0171B
EPW_CirculationRivaTo 5916 0x0171C
EPW_DrainBLineTo 5917 0x0171D
EPW_RinseBLineDrainATo 5918 0x0171E
EPW_WaterColumnCCATo 5919 0x0171F
EPW_WaterColumnCCBTo 5920 0x01720
EPW_EmptyCCBTo 5921 0x01721
EPW_EmptyCCB2To 5922 0x01722
EPW_EmptyCCTo 5923 0x01723
EPW_BPMReleasePressureTimer 5924 0x01724
EPW_DelayNoFlowTo 5925 0x01725
EPW_RinseBLineBicTo 5926 0x01726
EPW_DisinfCtrlRdyCount 5927 0x01727
EPW_EmptyCCBufdTo 5928 0x01728
EPW_EmptyCCB2ufdTo 5929 0x01729
EPW_EmptyCCAufdTo 5930 0x0172A
EPW_WaterColumnCCBufdTo 5931 0x0172B
EPW_TarationTO 5932 0x0172C
EPW_OpenFLVAToPrimeBLineTO 5933 0x0172D
EPW_HeatCCEmptyNoInPressDelayTo 5934 0x0172E
EPW_SHC_RinseRevaLineTo 5935 0x0172F
EPW_SHC_RinseBLineTo 5936 0x01730
EPW_SHC_RinseALineAndMixChambTo 5937 0x01731
EPW_SHC_RinseRivaLineTo 5938 0x01732
EPW_SHC_PrimeBLineWithCitricTo 5939 0x01733
EPW_SHC_RinseMainLineTo 5940 0x01734
EPW_SHC_RinseZevaByvaLineTo 5941 0x01735
EPW_SHC_RinseFivaLineTo 5942 0x01736
EPW_SHC_RinseMainLineUFDTo 5943 0x01737
EPW_SHC_DetectCitricTo 5944 0x01738
EPW_SHC_CitricFillupMainLineTo 5945 0x01739
EPW_SHC_RinseEvvaLineTo 5946 0x0173A
EPW_SHC_CitricFillupRevaTo 5947 0x0173B
EPW_SHC_HeatAndCitricRecircTavaTo 5948 0x0173C
EPW_SHC_HeatDistrTavaTo 5949 0x0173D
EPW_SHC_CitFillMainLineCoolDownTo 5950 0x0173E


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 133 (381)
Program version 1.1.0
C EN 2015-03-26

EPW_FlushChemValveLine1To 5951 0x0173F


EPW_CHVAPumpSlowdownDelayTo 5952 0x01740
EPW_FlushChemValveLine2To 5953 0x01741
EPW_FlushChemValveLine3To 5954 0x01742
EPW_PreRinseChemValveLineTo 5955 0x01743
EPW_RinseChemValveLineTo 5956 0x01744
EPW_StopPumpsBeforeCHVAtestTo 5957 0x01745
EPW_PostRinseChemValveLineTo 5958 0x01746
EPW_SHC_HeatAndCitricRecircEvvaTo 5959 0x01747
EPW_SHC_HeatAndCitricRecircRivaTo 5960 0x01748
EPW_SHC_HeatAndCitricRecircZevaByvaTo 5961 0x01749
EPW_SHC_HeatDistrEvvaTo 5962 0x0174A
EPW_SHC_HeatDistrRivaTo 5963 0x0174B
EPW_SHC_HeatDistrZevaByvaTo 5964 0x0174C
EPW_SHC_DilluteRemCitricInBLineTo 5965 0x0174D
EPW_SHC_FlushChemValveTo 5966 0x0174E
EPW_ChC_FillPickUpTubeTo 5967 0x0174F
EPW_ChC_ExpChamberFillupTo 5968 0x01750
EPW_ChC_ExpChambFillCitricUfdTo 5969 0x01751
EPW_ChC_FlowThrottleCleanTo 5970 0x01752
EPW_ChC_RivaFillupTo 5971 0x01753
EPW_ChC_ZevaFillupTo 5972 0x01754
EPW_ChC_ZevaFillCitricUfdTo 5973 0x01755
EPW_ChC_ByvaFillupTo 5974 0x01756
EPW_ChC_ByvaFillCitricUfdTo 5975 0x01757
EPW_ChC_RevaFillupTo 5976 0x01758
EPW_ChC_EvvaFillupTo 5977 0x01759
EPW_ChC_EvvaFillupUfdTo 5978 0x0175A
EPW_ChC_PrepareWaitForStickTo 5979 0x0175B
EPW_ChC_FinalFillupTo 5980 0x0175C
EPW_ChC_FillupCoolOffTo 5981 0x0175D
EPW_ChC_ChemFillupCount 5982 0x0175E
EPW_ChC_DisinfCtrlRdyCount 5983 0x0175F
EPW_DelayNoFlow2To 5984 0x01760
EPW_SHC_HeatDistrStartCount 5985 0x01761
EPW_SHC_HeaterInefficientTo 5986 0x01762
EPW_RivaFillup5sTo 5987 0x01763
EPW_RivaFillup5sCount 5988 0x01764
EPD_RTMCalHL 5989 0x01765
EPD_RTMCalLL 5990 0x01766
BMTRCBMTraco 5991 0x01767
BMTRCTracoBloodProc 5992 0x01768
BMTRCDialysisProc 5993 0x01769


134 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

BMTRCBlPumpTestProc 5994 0x0176A


BMTRCVenArtClampTestProc 5995 0x0176B
BMTRCAirDetTestProc 5996 0x0176C
BMTRCBloodModeProc 5997 0x0176D
BMTRCBlPumpDnProc 5998 0x0176E
BMTRCClampsDnProc 5999 0x0176F
BMTRCVenPrVarianceProc 6000 0x01770
BMTRCBlPumpSnProc 6001 0x01771
BMTRCClampsSnProc 6002 0x01772
BMTRCSingleNeedleProc 6003 0x01773
BMTRCBloodFlowProc 6004 0x01774
BMTRCMeanFlowSnProc 6005 0x01775
BMTRCAutoPrimingProc 6006 0x01776
BMTRCApArtClampProc 6007 0x01777
BMTRCApFlowProc 6008 0x01778
BMTRCApFillingProc 6009 0x01779
BMTRCApPressureTestProc 6010 0x0177A
BMTRCApPrimingProc 6011 0x0177B
BMTRCApFlush1Proc 6012 0x0177C
BMTRCApFlush2Proc 6013 0x0177D
BMTRCApLevelAdjustProc 6014 0x0177E
BMTRCApTubeFillProc 6015 0x0177F
BMTRCInitApEarlyFm 6016 0x01780
BMTRCAutoRinseBackProc 6017 0x01781
BMTRCRunRinseBackProc 6018 0x01782
BMTRCRinseStandbyClampProc 6019 0x01783
BMTRCARinseFlowProc 6020 0x01784
BMTRCHeparinPumpProc 6021 0x01785
BMTRCVolumeProc 6022 0x01786
BMTRCStrokeVolProc 6023 0x01787
BMTRCBPMProc 6024 0x01788
BMTRCBPMServiceProc 6025 0x01789
BMTRCBmErrorProc 6026 0x0178A
BMTRCKickStartProc 6027 0x0178B
BMTRCPressureTestProc 6028 0x0178C
BMTRCHepPumpTestProc 6029 0x0178D
CMTRCCMRootProc 6030 0x0178E
CMTRCCMTraco 6031 0x0178F
CMTRCAnalogConnCheckProcess 6032 0x01790
CMTRCRecoverControlProc 6033 0x01791
CMTRCSafeMemWriteProc 6034 0x01792
FDTRCFDTransCtrlProc 6035 0x01793
FDTRCSelectProc 6036 0x01794


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 135 (381)
Program version 1.1.0
C EN 2015-03-26

FDTRCHeaterEnableCtrlProc 6037 0x01795


FDTRCRinseBLineProc 6038 0x01796
FDTRCSagsTestProc 6039 0x01797
FDTRCSagsTestDrainProc 6040 0x01798
FDTRCValveTestProc 6041 0x01799
FDTRCDisinfCtrlProc 6042 0x0179A
FDTRCDisinfectionProc 6043 0x0179B
FDTRCTracoDisIntCheckProc 6044 0x0179C
FDTRCConcFillupProc 6045 0x0179D
FDTRCConcFillupChemProc 6046 0x0179E
FDTRCCentrFillupProc 6047 0x0179F
FDTRCDilFillupProc 6048 0x017A0
FDTRCCleanCartPrimProc 6049 0x017A1
FDTRCAutoRunProc 6050 0x017A2
FDTRCCleanCartFillProc 6051 0x017A3
FDTRCCitricDwellProc 6052 0x017A4
FDTRCRinseProc 6053 0x017A5
FDTRCHeatDisTestProc 6054 0x017A6
FDTRCHeatDisFlushProc 6055 0x017A7
FDTRCDrainProc 6056 0x017A8
FDTRCDrainBefChemDisProc 6057 0x017A9
FDTRCCirculatingDwellProc 6058 0x017AA
FDTRCInteruptDisProc 6059 0x017AB
FDTRCDisinfectionTimeProc 6060 0x017AC
FDTRCDisinfectionTimeProc2 6061 0x017AD
FDTRCDisinfectionDispTimeProc 6062 0x017AE
FDTRCChemRinseDispTimeProc 6063 0x017AF
FDTRCHeatCitricDispTimeProc 6064 0x017B0
FDTRCStickAResolverProc 6065 0x017B1
FDTRCStickBResolverProc 6066 0x017B2
FDTRCBicLatchResolverProc 6067 0x017B3
FDTRCChemFillupRdyLatchProc 6068 0x017B4
FDTRCRinseBefHeatProc 6069 0x017B5
FDTRCDisinfHistoryProc 6070 0x017B6
FDTRCWRO300HDisComProc 6071 0x017B7
FDTRCExtDisinfComProc 6072 0x017B8
FDTRCExtDisinfectionProc 6073 0x017B9
FDTRCLowFlowHeatPumpProc 6074 0x017BA
FDTRCDisIntHistoryProc 6075 0x017BB
FDTRCHpgCtrlProc 6076 0x017BC
FDTRCEvacuationValveCtrlProc 6077 0x017BD
FDTRCShortHeatCitricProc 6078 0x017BE
FDTRCCHVATestProc 6079 0x017BF


136 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

FDTRCCHVACurrentTestProc 6080 0x017C0


FDTRCCHVAPressureTestProc 6081 0x017C1
FDTRCShortHeatCitricTimeProc 6082 0x017C2
FFTRCFFTransCtrlProc 6083 0x017C3
FFTRCFchMainProc 6084 0x017C4
FFTRCSucPumpRegProc 6085 0x017C5
FFTRCFlowPumpRegProc 6086 0x017C6
FFTRCCondRegProc 6087 0x017C7
FFTRCDirectValvesProc 6088 0x017C8
FFTRCCurrValvesTestProc 6089 0x017C9
FFTRCZeroValveProc 6090 0x017CA
FFTRCInletValveProc 6091 0x017CB
FFTRCDisinfectValvesProc 6092 0x017CC
FFTRCCurrTestDisValvesProc 6093 0x017CD
FFTRCFchComTestVarFlowProc 6094 0x017CE
FFTRCHeaterRegProc 6095 0x017CF
FFTRCHeaterEnableProc 6096 0x017D0
FFTRCHeaterBypassProc 6097 0x017D1
FFTRCHpgCtrlProc 6098 0x017D2
FFTRCDegassPrTestProc 6099 0x017D3
FFTRCAPFlowPathCtrlProc 6100 0x017D4
FFTRCStickProc 6101 0x017D5
FFTRCStickModeProc 6102 0x017D6
FFTRCStickAResolverProc 6103 0x017D7
FFTRCStickBResolverProc 6104 0x017D8
FFTRCBicLatchResolverProc 6105 0x017D9
FFTRCMonitorOffProc 6106 0x017DA
FFTRCCValvesTestProc 6107 0x017DB
FFTRCCValvesTestCtrlProc 6108 0x017DC
FFTRCBiCartValveProc 6109 0x017DD
FFTRCProlongFchPowerProc 6110 0x017DE
FFTRCProlongFchW300HeatProc 6111 0x017DF
FFTRCFchCommonTestCtrlProc 6112 0x017E0
FFTRCDoFchCommTestProc 6113 0x017E1
FFTRCVarFlowTestProc 6114 0x017E2
FFTRCCValvesCurrTestProc 6115 0x017E3
FMTRCFmTransCtrlProc 6116 0x017E4
FSTRCFSTransCtrlProc 6117 0x017E5
FSTRCServiceProc 6118 0x017E6
FSTRCDegassingLevelProc 6119 0x017E7
FSTRCValveProcDisrinseValves 6120 0x017E8
FSTRCUFCalibrationProc 6121 0x017E9
FSTRCDrainRefVolProc 6122 0x017EA


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 137 (381)
Program version 1.1.0
C EN 2015-03-26

FSTRCUFCalibPrepProc 6123 0x017EB


FSTRCUFMainFlOffsetProc 6124 0x017EC
FSTRCUFCh1FlowCalProc 6125 0x017ED
FSTRCUFDiffCalProc 6126 0x017EE
FSTRCVerifyUFProc 6127 0x017EF
FSTRCUFIsolVerifyProc 6128 0x017F0
FSTRCInternalSupportProc 6129 0x017F1
FSTRCInternalSupportProc2 6130 0x017F2
FSTRCDiagnosticProc 6131 0x017F3
FSTRCCompleteUFTarationProc 6132 0x017F4
FSTRCPumpDutyCycleTestProc 6133 0x017F5
FSTRCPreUFTarProc 6134 0x017F6
FSTRCUfTarProc 6135 0x017F7
FSTRCDrainProc 6136 0x017F8
FSTRCFlowPathProc 6137 0x017F9
FSTRCDIVAEvvaProc 6138 0x017FA
FSTRCBYVAEvvaProc 6139 0x017FB
FSTRCSuctionPumpServiceProc 6140 0x017FC
FSTRCHeaterEnableProc 6141 0x017FD
FSTRCUFBoxAndW300ServComProc 6142 0x017FE
FSTRCUFBox 6143 0x017FF
FSTRCFlPrepProc 6144 0x01800
FSTRCValveProcPrepValves 6145 0x01801
FSTRCValveProc 6146 0x01802
FSTRCVarFlowMotorCtrlProc 6147 0x01803
FSTRCVFMCtrlNotPresetProc 6148 0x01804
FSTRCVFMCtrlNormalProc 6149 0x01805
FTTRCFTTransCtrlProc 6150 0x01806
FTTRCPrepCtrlProc 6151 0x01807
FTTRCCondModeProc 6152 0x01808
FTTRCFluidPrepRdyProc 6153 0x01809
FTTRCUFSFunctionCheckProc 6154 0x0180A
FTTRCUFSFchRecoverTimerProc 6155 0x0180B
FTTRCFluidTempTestRdyProc 6156 0x0180C
FTTRCBiCartRdyProc 6157 0x0180D
FTTRCFluidPrepProc 6158 0x0180E
FTTRCInletCtrlProc 6159 0x0180F
FTTRCHeaterCtrlProc 6160 0x01810
FTTRCMainFlCtrlProc 6161 0x01811
FTTRCCondCtrlProc 6162 0x01812
FTTRCBCProdProc 6163 0x01813
FTTRCBiCartEmptyProc 6164 0x01814
FTTRCBiCartPrimProc 6165 0x01815


138 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

FTTRCBCAttnResProc 6166 0x01816


FTTRCStickAResolverProc 6167 0x01817
FTTRCStickBResolverProc 6168 0x01818
FTTRCBicLatchResolverProc 6169 0x01819
FTTRCUFCtrlProc 6170 0x0181A
FTTRCFilterFillupCtrlProc 6171 0x0181B
FTTRCSucPumpCtrlProc 6172 0x0181C
FTTRCSucPumpBypassProc 6173 0x0181D
FTTRCSucPumpPrimCtrlProc 6174 0x0181E
FTTRCSucPumpTMTCtrlProc 6175 0x0181F
FTTRCSucPumpNormalUFRateProc 6176 0x01820
FTTRCSucPumpLowUFRateProc 6177 0x01821
FTTRCSucPumpZeroUFRateProc 6178 0x01822
FTTRCSucPumpDisConCtrlProc 6179 0x01823
FTTRCUFModeProc 6180 0x01824
FTTRCUFLogProc 6181 0x01825
FTTRCDelayedDiaConStsProc 6182 0x01826
FTTRCUFTarSeqProc 6183 0x01827
FTTRCUFTarPreProc 6184 0x01828
FTTRCUFTarProc 6185 0x01829
FTTRCUFTarPostProc 6186 0x0182A
FTTRCUFTarFirstProc 6187 0x0182B
FTTRCFlMeasProc 6188 0x0182C
FTTRCUFRateFiltProc 6189 0x0182D
FTTRCNegUFChkProc 6190 0x0182E
FTTRCFlMeasInactiveTarProc 6191 0x0182F
FTTRCFlMeasUFModeProc 6192 0x01830
FTTRCFlMeasUFModeTarProc 6193 0x01831
FTTRCFlMeasIsolUFModeProc 6194 0x01832
FTTRCFlMeasIsolUFModeTarProc 6195 0x01833
FTTRCMiscCtrlProc 6196 0x01834
FTTRCHeaterEnableCtrlProc 6197 0x01835
FTTRCHpgCtrlProc 6198 0x01836
FTTRCValveCtrlProc 6199 0x01837
FTTRCRIVACtrlProc 6200 0x01838
FTTRCZEVACtrlProc 6201 0x01839
FTTRCTAVACtrlProc 6202 0x0183A
FTTRCDirectValveCtrlProc 6203 0x0183B
FTTRCDirectValveNormProc 6204 0x0183C
FTTRCDirectValveTarProc 6205 0x0183D
FTTRCMajorAirLeakAttnProc 6206 0x0183E
FTTRCAPFlowPathCtrlProc 6207 0x0183F
FTTRCMonitorOffProc 6208 0x01840


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 139 (381)
Program version 1.1.0
C EN 2015-03-26

FTTRCEvvaOpenDelayProc 6209 0x01841


FTTRCInletWaterCondProc 6210 0x01842
FTTRCDiascanCtrlProc 6211 0x01843
FTTRCKeepTAVAClosedAfterTarProc 6212 0x01844
FTTRCDialysisFlowChangeTarationProc 6213 0x01845
OMTRCOmMachineModeCtrlProc 6214 0x01846
OPTRCTopProc 6215 0x01847
OPTRCPresetModeSelectProc 6216 0x01848
OPTRCManualPresetProc 6217 0x01849
OPTRCPresetChainHandlingProc 6218 0x0184A
OPTRCExtPresetProc 6219 0x0184B
OPTRCPcPresetProc 6220 0x0184C
OPTRCPcPresetCalcProc 6221 0x0184D
OPTRCPcPresetTotalCalcProc 6222 0x0184E
OPTRCPcPresetChainCalcProc 6223 0x0184F
OPTRCPcPresetDefaultProc 6224 0x01850
OPTRCExtSendProc 6225 0x01851
OPTRCExtReceiveProc 6226 0x01852
OPTRCExtReceiveTotalCalcProc 6227 0x01853
OPTRCExtReceiveChainCalcProc 6228 0x01854
OPTRCExtReceiveCalcStatusProc 6229 0x01855
OPTRCDefaultPresetProc 6230 0x01856
OPTRCAllDefaultProc 6231 0x01857
OPTRCSelectKeySeqProc 6232 0x01858
OPTRCHoldKeySeqProc 6233 0x01859
OPTRCNextLampProc 6234 0x0185A
OPTRCPreviousLampProc 6235 0x0185B
OPTRCIncreaseLampProc 6236 0x0185C
OPTRCDecreaseLampProc 6237 0x0185D
OSTRCTopProc 6238 0x0185E
OSTRCSelectModeProc 6239 0x0185F
OSTRCFChProc 6240 0x01860
OSTRCPresetProc 6241 0x01861
OSTRCSelectingServiceProc 6242 0x01862
OSTRCCalibProc 6243 0x01863
OSTRCTransducerProc 6244 0x01864
OSTRCCalibOtherProc 6245 0x01865
OSTRCVariableFlowCalibProc 6246 0x01866
OSTRCVFCalibDisplayCtrlProc 6247 0x01867
OSTRCEstTempDropAdjProc 6248 0x01868
OSTRCRunTimeMeterProc 6249 0x01869
OSTRCCalibLinTransProc 6250 0x0186A
OSTRCWaitEEPromTestProc 6251 0x0186B


140 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OSTRCWaitSampleProc 6252 0x0186C


OSTRCWaitOffsCalCalcOkProc 6253 0x0186D
OSTRCWaitEnterSensitivityProc 6254 0x0186E
OSTRCWaitSensitivitySelectProc 6255 0x0186F
OSTRCWaitSensSampleSelectProc 6256 0x01870
OSTRCWaitSenseSampleProc 6257 0x01871
OSTRCWaitSenseCalCalcOkProc 6258 0x01872
OSTRCWaitEEPromInitProc 6259 0x01873
OSTRCWaitCalEEPromStoreOkProc 6260 0x01874
OSTRCEEPromStoreConfirmProc 6261 0x01875
OSTRCPrimDetCalProc 6262 0x01876
OSTRCWaitCalCalcErrorConfirmProc 6263 0x01877
OSTRCWaitUFCalSelectProc 6264 0x01878
OSTRCWaitUFEEPromTestProc 6265 0x01879
OSTRCWaitUFCalReadyProc 6266 0x0187A
OSTRCWaitUFCalEEPromStoreOkProc 6267 0x0187B
OSTRCUFEEPromStoreConfirmProc 6268 0x0187C
OSTRCWaitUFBeenCalibratedProc 6269 0x0187D
OSTRCUFWaitInternalProc 6270 0x0187E
OSTRCUFConfirmNotAllowedProc 6271 0x0187F
OSTRCWaitOUFEEPromReadyProc 6272 0x01880
OSTRCWaitUFInternalConfirmProc 6273 0x01881
OSTRCVerifyUFProc 6274 0x01882
OSTRCVerifyUFCalibProc 6275 0x01883
OSTRCUFDisplayResult1Proc 6276 0x01884
OSTRCUFDisplayResult2Proc 6277 0x01885
OSTRCWaitUFCalEEPromStoOkSuppProc 6278 0x01886
OSTRCWaitDiagnosticSelectionProc 6279 0x01887
OSTRCAwaitFMDiagSelectionProc 6280 0x01888
OSTRCVarFlowDiagProc 6281 0x01889
OSTRCVarFlowDiagCtrlProc 6282 0x0188A
OSTRCVFDiagDisplayCtrlProc 6283 0x0188B
OSTRCPumpsDiagProc 6284 0x0188C
OSTRCUFDiagProc 6285 0x0188D
OSTRCCompleteUFTarationProc 6286 0x0188E
OSTRCCompletedUFTarationProc 6287 0x0188F
OSTRCAwaitBMDiagSelectionProc 6288 0x01890
OSTRCLampDiagProc 6289 0x01891
OSTRCAwaitPumpSelectionProc 6290 0x01892
OSTRCArtPumpDiagProc 6291 0x01893
OSTRCAwaitHepPumpDiagSelectionProc 6292 0x01894
OSTRCHepPumpDiagProc 6293 0x01895
OSTRCWaitSystemDiagSelectionProc 6294 0x01896


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 141 (381)
Program version 1.1.0
C EN 2015-03-26

OSTRCWaitI2CSelectProc 6295 0x01897


OSTRCI2CScanProc 6296 0x01898
OSTRCI2CScanReadyConfirmProc 6297 0x01899
OSTRCI2CAddressUpdateConfirmProc 6298 0x0189A
OSTRCWaitI2CUpdateProc 6299 0x0189B
OSTRCI2CInitConfirmProc 6300 0x0189C
OSTRCWaitI2CInitProc 6301 0x0189D
OSTRCI2CCalEEPromErrorConfirmProc 6302 0x0189E
OSTRCUtilitySelectProc 6303 0x0189F
OSTRCDrainProc 6304 0x018A0
OSTRCAwaitExtFchStartProc 6305 0x018A1
OSTRCExtFchStandAloneProc 6306 0x018A2
OSTRCExtFchInteractiveProc 6307 0x018A3
OSTRCCalibBypassProc 6308 0x018A4
OSTRCFMBypassProc 6309 0x018A5
OSTRCTransducerSupportProc 6310 0x018A6
OSTRCWaitCalEEPromStoreOkSuppProc 6311 0x018A7
OSTRCUFWaitInternalSupportProc 6312 0x018A8
OSTRCSelectKeySeqProc 6313 0x018A9
OSTRCHoldKeySeqProc 6314 0x018AA
OSTRCLampSeqProc 6315 0x018AB
OSTRCFluidPathStatusProc 6316 0x018AC
OSTRCFMPumpDiagLampProc 6317 0x018AD
OSTRCResetProc 6318 0x018AE
OSTRCTimeDispProc 6319 0x018AF
OSTRCNextLmpProc 6320 0x018B0
OSTRCPreviousLmpProc 6321 0x018B1
OSTRCIncreaseLmpProc 6322 0x018B2
OSTRCDecreaseLmpProc 6323 0x018B3
OSTRCUFCalDisplayProc 6324 0x018B4
OSTRCPumpsDiagHbvaHandleProc 6325 0x018B5
OSTRCCancelButtonProc 6326 0x018B6
OTTRCTracoOpcomProc 6327 0x018B7
OTTRCOperatorsCommunicationProc 6328 0x018B8
OTTRCMonitorControlProc 6329 0x018B9
OTTRCMonitorModeProc 6330 0x018BA
OTTRCModeSelectProc 6331 0x018BB
OTTRCUfIsolUfModeProc 6332 0x018BC
OTTRCDisinfRinseModeProc 6333 0x018BD
OTTRCAutoDisinfActiveProc 6334 0x018BE
OTTRCMaintenanceSelProc 6335 0x018BF
OTTRCTreatmentStartStopProc 6336 0x018C0
OTTRCFluidBypProc 6337 0x018C1


142 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OTTRCAutoPrimModeProc 6338 0x018C2


OTTRCAutoPrimBankSelectProc 6339 0x018C3
OTTRCFluidPrepRqProc 6340 0x018C4
OTTRCDiscontinueModeProc 6341 0x018C5
OTTRCRinseBackModeProc 6342 0x018C6
OTTRCHepPumpModeProc 6343 0x018C7
OTTRCProfModeProc 6344 0x018C8
OTTRCProfNaAcModeProc 6345 0x018C9
OTTRCProfNaModeProc 6346 0x018CA
OTTRCProfHCO3ModeProc 6347 0x018CB
OTTRCProfUFModeProc 6348 0x018CC
OTTRCProfModelSelectProc 6349 0x018CD
OTTRCProfValueProc 6350 0x018CE
OTTRCProfNaAcProc 6351 0x018CF
OTTRCProfNaProc 6352 0x018D0
OTTRCProfHCO3Proc 6353 0x018D1
OTTRCProfUFProc 6354 0x018D2
OTTRCConcStandbyModeProc 6355 0x018D3
OTTRCDiascanModeProc 6356 0x018D4
OTTRCDiascanMeasProc 6357 0x018D5
OTTRCDiascanTimerProc 6358 0x018D6
OTTRCDiascanPreMeasProc 6359 0x018D7
OTTRCModeAckProc 6360 0x018D8
OTTRCChemFlagAckProc 6361 0x018D9
OTTRCAcModeAckProc 6362 0x018DA
OTTRCBcModeAckProc 6363 0x018DB
OTTRCRinseAckProc 6364 0x018DC
OTTRCChemAckProc 6365 0x018DD
OTTRCHeatAckProc 6366 0x018DE
OTTRCDrainAckProc 6367 0x018DF
OTTRCDiscontinueAckProc 6368 0x018E0
OTTRCInitInPressReplyProc 6369 0x018E1
OTTRCMm90SupportProc 6370 0x018E2
OTTRCConcBankSelProc 6371 0x018E3
OTTRCMm90CondSetCalcProc 6372 0x018E4
OTTRCChemDisBankSelectProc 6373 0x018E5
OTTRCAlarmLatchProc 6374 0x018E6
OTTRCTempAlarmLatchProc 6375 0x018E7
OTTRCCondAlarmLatchProc 6376 0x018E8
OTTRCAttnHandlingProc 6377 0x018E9
OTTRCAirDetInactiveAttnProc 6378 0x018EA
OTTRCInPressAlarmProc 6379 0x018EB
OTTRCInPressAlarmCtrlProc 6380 0x018EC


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 143 (381)
Program version 1.1.0
C EN 2015-03-26

OTTRCInPressActiveAlarmProc 6381 0x018ED


OTTRCInPressActiveAlarmCtrlProc 6382 0x018EE
OTTRCWRO300HNotRdyAttnProc 6383 0x018EF
OTTRCWRO300HNotRdyAttnSubProc 6384 0x018F0
OTTRCWRO300HPrepAttnProc 6385 0x018F1
OTTRCWRO300HLFHNOTAvailAttnProc 6386 0x018F2
OTTRCWRO300CommFailureProc 6387 0x018F3
OTTRCUPSActAttnProc 6388 0x018F4
OTTRCBattActAlarmProc 6389 0x018F5
OTTRCBattlowActAlarmProc 6390 0x018F6
OTTRCArtPrTestAckProc 6391 0x018F7
OTTRCVenPrTestAckProc 6392 0x018F8
OTTRCArtPrAutoLimAlarmProc 6393 0x018F9
OTTRCVenPrAutoLimAlarmProc 6394 0x018FA
OTTRCTmpAutoLimAlarmProc 6395 0x018FB
OTTRCBlLeakCleanAttnProc 6396 0x018FC
OTTRCBatteryFailedAttnProc 6397 0x018FD
OTTRCLowDialysisFlowAttnProc 6398 0x018FE
OTTRCBlPumpOvldFchProc 6399 0x018FF
OTTRCBlPumpCoverOpenAlarmProc 6400 0x01900
OTTRCBlTubeChangeAttnProc 6401 0x01901
OTTRCSnPrSupervAlarmProc 6402 0x01902
OTTRCConcRqAttnProc 6403 0x01903
OTTRCAirDetFluidRqAttnProc 6404 0x01904
OTTRCDiaConNotInBypassProc 6405 0x01905
OTTRCDisinfInteruptAttnProc 6406 0x01906
OTTRCDiaConRemoveAttnProc 6407 0x01907
OTTRCDiaConReplaceAttnProc 6408 0x01908
OTTRCDiaConRemoveAttnProc1 6409 0x01909
OTTRCDiaConRemoveAttnProc2 6410 0x0190A
OTTRCDiaConReplaceAttnProc1 6411 0x0190B
OTTRCDiaConReplaceAttnProc2 6412 0x0190C
OTTRCDiaConAbortAttnProc 6413 0x0190D
OTTRCDiaConFinalAttachAttnProc 6414 0x0190E
OTTRCManW300StartStopAttnProc 6415 0x0190F
OTTRCInTMTDiaConInBypAlarmProc 6416 0x01910
OTTRCCleanCartLatchAttnProc 6417 0x01911
OTTRCCleanCartPrimAttnProc 6418 0x01912
OTTRCLowHeatingTempAttnProc 6419 0x01913
OTTRCNoDilutionAttnProc 6420 0x01914
OTTRCRemoveCleanCartAttnProc 6421 0x01915
OTTRCApPrBuildAttnProc 6422 0x01916
OTTRCApPrTestAttnProc 6423 0x01917


144 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OTTRCApPrimDetBloodAlarmProc 6424 0x01918


OTTRCBicHolderNotInstAttnProc 6425 0x01919
OTTRCUfParaZeroAttnProc 6426 0x0191A
OTTRCSagsTestAttnProc 6427 0x0191B
OTTRCHepSetToZeroAttnProc 6428 0x0191C
OTTRCCentralChemAttnProc 6429 0x0191D
OTTRCPrimVolAchAttnProc 6430 0x0191E
OTTRCPrimVolAchActionProc 6431 0x0191F
OTTRCTimeBetwDisinfDecalcProc 6432 0x01920
OTTRCCountTimeInBicarbProc 6433 0x01921
OTTRCCountTimeInACBCProc 6434 0x01922
OTTRCFlushTubeAttnProc 6435 0x01923
OTTRCUFDflushAttnProc 6436 0x01924
OTTRCNegTmpAttnProc 6437 0x01925
OTTRCNegUFAttnProc 6438 0x01926
OTTRCBPMAttnProc 6439 0x01927
OTTRCBPMAlarmProc 6440 0x01928
OTTRCDisinfInProgAttnProc 6441 0x01929
OTTRCUFAchTooEarlyAttnProc 6442 0x0192A
OTTRCHydraulicRestartAttnProc 6443 0x0192B
OTTRCDiascanDeactivateAttnProc 6444 0x0192C
OTTRCProfDeactivateAttnProc 6445 0x0192D
OTTRCProfNaChangedAttnProc 6446 0x0192E
OTTRCProfHCO3ChangedAttnProc 6447 0x0192F
OTTRCProfUFRChangedAttnProc 6448 0x01930
OTTRCFchHighCondAttnProc 6449 0x01931
OTTRCFchBlTestAttnProc 6450 0x01932
OTTRCFchVClampTestAttnProc 6451 0x01933
OTTRCFchAClampTestAttnProc 6452 0x01934
OTTRCFchDivaTavaTestAttnProc 6453 0x01935
OTTRCTimeProc 6454 0x01936
OTTRCTmtTimeProc 6455 0x01937
OTTRCRemainingTimePosEvalProc 6456 0x01938
OTTRCNonDiffTimeProc 6457 0x01939
OTTRCClinAutComSupProc 6458 0x0193A
OTTRCAltLangWarningProc 6459 0x0193B
OTTRCVarFlowRegIneffAttnProc 6460 0x0193C
OTTRCDiascanAttnProc 6461 0x0193D
OTTRCDiascanGeneralAttnProc 6462 0x0193E
OTTRCDiascanMeasFailAttnProc 6463 0x0193F
OTTRCDiascanFailAttnProc 6464 0x01940
OTTRCDiascanTargetAchievedProc 6465 0x01941
OTTRCDiascanNotActiveAttnProc 6466 0x01942


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 145 (381)
Program version 1.1.0
C EN 2015-03-26

OTTRCDiascanKLowAttnProc 6467 0x01943


OTTRCDiascanKtVLowAttnProc 6468 0x01944
OTTRCDiascanProfUFAttnProc 6469 0x01945
OTTRCWRO300HInclInHeatProc 6470 0x01946
OTTRCTreatOverviewCtrlProc 6471 0x01947
OTTRCTO_Welcome_CtrlProc 6472 0x01948
OTTRCTO_OnOffBtn_CtrlProc 6473 0x01949
OTTRCTO_Idle_CtrlProc 6474 0x0194A
OTTRCTO_IdleSetting_CtrlProc 6475 0x0194B
OTTRCUfAndTimeAdjTimerProc 6476 0x0194C
OTTRCDiascanTmtSettingsAdjProc 6477 0x0194D
OTTRCDiascanDistrVolAdjProc 6478 0x0194E
OTTRCProfParamUFAdjProc 6479 0x0194F
OTTRCTimeBetweenUfdChangeProc 6480 0x01950
OTTRCChemValTstStatusProc 6481 0x01951
OTTRCHeparinCompletedAttn 6482 0x01952
OTTRCBypassDuringRecirAttn 6483 0x01953
OTTRCFchHepPumpAttnProc 6484 0x01954
OTTRCHepPumpDisabledAttnProc 6485 0x01955
OTTRCBlPumpStartStopProc 6486 0x01956
OTTRCPrimUIProc 6487 0x01957
OTTRCConnectPatientProc 6488 0x01958
OTTRCRecirculationProc 6489 0x01959
OTTRCNewBloodCircuitProc 6490 0x0195A
OTTRCRinseBackUIProc 6491 0x0195B
OTTRCBlPumpStartStopTmtProc 6492 0x0195C
OTTRCBlPumpStopClampRqProc 6493 0x0195D
OTTRCBlPumpStartStopApProc 6494 0x0195E
OTTRCBlPumpStartStopRbProc 6495 0x0195F
OTTRCBlPumpTechErrProc 6496 0x01960
OTTRCBlPumpKnobProc 6497 0x01961
OTTRCProtectiveSetValueProc 6498 0x01962
OTTRCFluidValuesProc 6499 0x01963
OTTRCCondSetProc 6500 0x01964
OTTRCAcConcAltSetProc 6501 0x01965
OTTRCBcConcAltSetProc 6502 0x01966
OTTRCVprLimitValuesProc 6503 0x01967
OTTRCVprLimitUpdateProc 6504 0x01968
OTTRCWarningTextProc 6505 0x01969
OTTRCKeyboardProc 6506 0x0196A
OTTRCSelectKeySeqProc 6507 0x0196B
OTTRCDisinfectionKeyDelayProc 6508 0x0196C
OTTRCDisconnectPatientProc 6509 0x0196D


146 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

OTTRCBiCartEmptyAllowedProc 6510 0x0196E


OTTRCBiCartEmptyRqProc 6511 0x0196F
OTTRCDoNotStartBloodPumpProc 6512 0x01970
OTTRCResetRqProc 6513 0x01971
OTTRCDisRinseResetProc 6514 0x01972
OTTRCAutoOffAfterDisProc 6515 0x01973
OTTRCMachineOffProc 6516 0x01974
OTTRCMachineOffNormalProc 6517 0x01975
OTTRCMachineOffDisinfProc 6518 0x01976
OTTRCLogicalOffRecoverProc 6519 0x01977
OTTRCRealTimeClockProc 6520 0x01978
OTTRCRealTimeNormClockProc 6521 0x01979
OTTRCRealTimeAutoInactiveProc 6522 0x0197A
OTTRCRealTimeAutoHeatProc 6523 0x0197B
OTTRCRealTimeAutoRinseProc 6524 0x0197C
OTTRCRealTimeOvernightRinseProc 6525 0x0197D
OTTRCSetNowTimeProc 6526 0x0197E
OTTRCLampControlProc 6527 0x0197F
OTTRCOnOffLmpProc 6528 0x01980
OTTRCAttnLmpProc 6529 0x01981
OTTRCTechLmpProc 6530 0x01982
OTTRCBlPumpLmpProc 6531 0x01983
OTTRCBlPumpLmpTmtProc 6532 0x01984
OTTRCBlPumpLmpApProc 6533 0x01985
OTTRCBlPumpLmpRbProc 6534 0x01986
OTTRCBlPumpLmpTechErrProc 6535 0x01987
OTTRCFluidBypLmpProc 6536 0x01988
OTTRCStartStopLmpProc 6537 0x01989
OTTRCPrimingLmpProc 6538 0x0198A
OTTRCTimeLmpProc 6539 0x0198B
OTTRCSelectLmpProc 6540 0x0198C
OTTRCBloodFlowPathLmpProc 6541 0x0198D
OTTRCFluidFlowPathLmpProc 6542 0x0198E
OTTRCFlowPathFluidPrepModeProc 6543 0x0198F
OTTRCFlowPathIsolUfModeProc 6544 0x01990
OTTRCFlowPathUfModeProc 6545 0x01991
OTTRCRinseBackLmpProc 6546 0x01992
OTTRCBuzzAttnControlProc 6547 0x01993
OTTRCTimeDisplayControlProc 6548 0x01994
OTTRCTimeDispDisinfProc 6549 0x01995
OTTRCBPMLmpProc 6550 0x01996
OTTRCDisinfLmpProc 6551 0x01997
OTTRCFluidMenuLmpProc 6552 0x01998


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 147 (381)
Program version 1.1.0
C EN 2015-03-26

OTTRCBloodMenuLmpProc 6553 0x01999


OTTRCNextLmpProc 6554 0x0199A
OTTRCPreviousLmpProc 6555 0x0199B
OTTRCIncreaseLmpProc 6556 0x0199C
OTTRCDecreaseLmpProc 6557 0x0199D
OTTRCHoldLmpProc 6558 0x0199E

Version: 1.3 Fri Mar 20 08:10:11 2015


148 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Error codes
0115 000 XXX XXX, CANOPEN RX/TX ERROR
Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The RX or TX FIFO is full due to overload or other reasons.

Technical description of conditions for occurring:


- Contents are pushed to the FIFO at higher speed than they
are pulled from the FIFO. The error index indicates where
in the code the problem occurs.

Some possible explanations why the error has occurred:


- Bus overload or Protective exhausted. Serial log problems.

0115 001 000 XXX, CANOPEN MAILBOX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available mailbox for sending a message.

Technical description of conditions for occurring:


- Messages failed to be sent on the bus at the pace required
by Protective.

Some possible explanations why the error has occurred:


- Bus overload.

0115 002 000 XXX, CANOPEN MAILBOX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available mailbox for sending a message.

Technical description of conditions for occurring:


- Messages failed to be sent on the bus at the pace required
by Protective.

Some possible explanations why the error has occurred:


- Bus overload.

0117 001 000 XXX, LEAKAGE TEST STATUS NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Protective has not received FLUID_PATH_LEAKAGE_TEST_STATUS
from Control.

Some possible explanations why the error has occurred:


- Software Error, CAN problems.

0117 004 001 XXX, CPU C OR UI DOES NOT SEND SHUT DOWN ACKNOWLEDGE
Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 149 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Control or UI has not acknowledged that Protective will
shut down the machine.

Technical description of conditions for occurring:


- Protective has informed Control and UI that the machine
will shut down, but Protective has not got an acknowledge
from one or both nodes. This can for example happen if any
of nodes are in a condition where they can not communicate
with other nodes.

Some possible explanations why the error has occurred:


- Software Error, CAN problems, CPU Errors.

0117 004 003 XXX, SHUT DOWN FAILURE


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- The machine was not switched off within 5 seconds after
shut down request to IO.

0117 004 004 XXX, CPU C DOES NOT SEND SHUT DOWN QUERY RESPONSE
Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- No response to query within 5 seconds.

0117 004 005 XXX, CPU C DOES NOT SEND SHUT DOWN QUERY RESPONSE
Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- No response to query within 5 seconds.

0117 004 006 XXX, CPU C DID NOT APPROVE RESET REQUEST
Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- This type of request should is not expected to be denied by
CPU C.

0117 004 007 XXX, THE CPU P CHEM FLAG INCORRECTLY SET
Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Some possible explanations why the above error has
occurred: Software Error Corrupt safeRAM.

0117 004 008 XXX, CPU P STAMGR ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- A single request of signal bmfchpart1, bmfchpart2,


150 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

fmfchpart1 or fmfchpart2 has been performed. Some possible


explanations why the above error has occurred: Software
Error.

0117 004 009 XXX, MAINTENANCE RESET ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- The control system requested a reset to exit to function
check. The protective system was not in maintenance mode or
was not in disinfection with a finished disinfection.

Technical description of conditions for occurring:


- Software error.

0117 004 010 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 011 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 012 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 013 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 151 (381)
Program version 1.1.0
C EN 2015-03-26

system was not received by Protective.

0117 004 014 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 015 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 016 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 017 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 018 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 019 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


152 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 020 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 021 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 004 022 XXX, STARTUP SIGNAL NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Faulty configuration, Software Error or CAN Network Error.

Technical description of conditions for occurring:


- An expected signal required to correctly start up the SDL
system was not received by Protective.

0117 007 000 XXX, DEAIRATING CHAMBER HIGH LEVEL


Unconditional error, fluid path and many blood related
features are stopped but the blood pump can still be run.

General description of conditions for occurring:


- The deairating chamber bubble level has remained high for
too long time during treatment.

Some possible explanations why the error has occurred:


- EVVA leakage or faulty sensor.

0117 029 001 XXX, SETTING EDIT OPEN ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to open editing of a setting when another setting is
still open for edit.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 153 (381)
Program version 1.1.0
C EN 2015-03-26

0117 029 002 XXX, SETTING EDIT CLOSE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to close edit of a setting that is not open for
editing.

0117 037 001 XXX, APR SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit APR limits when editing of that setting is
locked.

0117 038 002 XXX, PRESET ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The preset bicarbonate concentrate bank is not set to a
valid value. Corrupt SAFE RAM (please make a default
preset).

0117 038 003 XXX, CONC ALT EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit concentration alternative when editing of
that setting is locked.

0117 038 004 XXX, CONC NA EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit Na concentration when editing of that
setting is locked.

0117 038 005 XXX, CONC HCO3 EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.


154 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Trying to edit HCO3 concentration when editing of that
setting is locked.

0117 038 006 XXX, CONC ALT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Not supported concentration bank alternative.

0117 040 001 XXX, TMP SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit TMP lower limit when editing of that setting
is locked.

0117 040 002 XXX, TMP SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit TMP higher limit when editing of that
setting is locked.

0117 040 003 XXX, TMP SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- TMP limits outside allowed range for lower limit or larger
than high limit.

0117 040 004 XXX, TMP SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- TMP limits outside allowed range for higher limit or
smaller than low limit.

0117 041 001 XXX, VPR SET EDIT ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 155 (381)
Program version 1.1.0
C EN 2015-03-26

during the last 30 days it is treated as an unconditional


error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Trying to edit VPR limits when editing of that setting is
locked.

0117 046 001 XXX, DIASCAN ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The total sodium limit has been exceeded in CPU P. No more
Diascan measurements are possible.

0117 047 000 XXX, PRESSURE TEST ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Problem with negative pressure during Pressure Test.
Negative pressure could not be obtained or too big
difference between VPR readings on C and P bus and APR
reading.

Some possible explanations why the error has occurred:


- Pump malfunction, leaking tubes or valve problems.

0117 047 001 XXX, PRESSURE TEST ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Problem with positive pressure during Pressure Test.
Positive pressure could not be obtained or too big
difference between VPR readings on C and P bus and APR
reading.

Some possible explanations why the error has occurred:


- Pump malfunction, leaking tubes or valve problems.

0117 047 002 XXX, PRESSURE TEST ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Problem with pressure variance during Pressure Realy Test.
Too big difference compared to reference values in Relay
test.

Some possible explanations why the error has occurred:


- Leaking tubes or valve problems.

0117 048 001 XXX, DIALYSIS PRESSURE TOO HIGH ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- This error is generated in treatment if the dialysis
pressure measured by dialysis pressure sensor is above the
value of "Dialysis pressure Sup" preset.


156 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0117 048 002 XXX, DIALYSIS PRESSURE TOO HIGH ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- This error is generated in priming if the dialysis pressure
measured by dialysis pressure sensor is above the value of
"Dialysis pressure Sup" preset.

0117 048 003 XXX, DIALYSIS PRESSURE TOO HIGH ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- This error is generated in rinseback if the dialysis
pressure measured by dialysis pressure sensor is above the
value of "Dialysis pressure Sup" preset.

0117 048 004 XXX, DIALYSIS PRESSURE TOO HIGH ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- This silent error is generated in disinfection if the
dialysis pressure measured by dialysis pressure sensor is
above the value of "Dialysis pressure Sup" preset.

0117 048 005 XXX, DIALYSIS PRESSURE TOO HIGH ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- This silent error is generated in service and preset mode
if the
- dialysis pressure measured by dialysis pressure sensor is
above
- the value of "Dialysis pressure Sup" preset.

Some possible explanations why the error has occurred:


- Fluid path leakage.

0117 049 001 XXX, UF VOLUME DEVIATION TWICE DURING THE SAME TREATMENT
Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The difference between control and protective accumulated
UF volumes has exceeded both the normal and the widened
alarm limit. The operator has been informed by alarms
during treatment, this silent error is for loggin purposes.

Some possible explanations why the error has occurred:


- Too long time between disinfections.
- Uncalibrated or broken UF cell. Please perform calibration.
- Broken UF flow meters.
- Malfunctioning General IO Protective.
- Fluid path leakage.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 157 (381)
Program version 1.1.0
C EN 2015-03-26

0117 049 002 XXX, UF RATE DEVIATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The protective system has recorded UF rate values beyond
the alarm limits, even after an extra taration.

Some possible explanations why the error has occurred:


- Broken UF protective flow meters.
- Fluid path leakage.

0117 049 003 XXX, TOO LARGE UF RATE DIFFERENCE BETWEEN CONTROL AND PROTECTIVE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The protective system has recorded a too large UF rate
difference between the control UF cell and the protective
UF flow meters.

Some possible explanations why the error has occurred:


- Too long time between disinfections.
- Uncalibrated or broken UF cell. Please perform calibration.
- Broken UF flow meters.
- Malfunctioning General IO Protective.
- Fluid path leakage.

0117 049 005 XXX, TOO LONG TIME HAS PASSED SINCE LAST TARATION
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The UF cell and protective flow meters need to be tarated
regularly during treatment, but no automatic taration
atttempt has been done in the last 45 minutes.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 000 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Alarm Enable, BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.


158 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0117 050 001 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Systolic High. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 002 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Systolic Low. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 003 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Diastolic High. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 004 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Diastolic Low. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 159 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Software Error.

0117 050 005 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Mean High. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 006 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Mean Low. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 007 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Pulse High. BPM Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 008 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Pulse Low. BPM Editing attempted when edit locked.


160 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 050 009 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Requested high value is lower than current low limit.

Technical description of conditions for occurring:


- Relative size not checked by UI.

0117 050 010 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Requested low value is higher than current high limit.

Technical description of conditions for occurring:


- Relative size not checked by UI.

0117 050 011 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Requested limit too low. Preset value exceeded.

Technical description of conditions for occurring:


- Preset limit not checked by UI.

0117 050 012 XXX, BPM SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Requested limit too high. Preset value exceeded.

Technical description of conditions for occurring:


- Preset limit not checked by UI.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 161 (381)
Program version 1.1.0
C EN 2015-03-26

0117 051 000 XXX, BLLINE SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- BLLINE Editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

Some possible explanations why the error has occurred:


- Software Error.

0117 051 001 XXX, BLLINE SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Requested bllod line diameter not in line with any of the
diameters specified by presets.

Some possible explanations why the error has occurred:


- Software Error.

0117 052 001 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling Na Start Manual value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 002 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling Na Stop Manual value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 003 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more


162 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

during the last 30 days it is treated as an unconditional


error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling Na Start Model value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 004 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling Na Stop Model value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 005 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling HCO3 Start Manual value editing attempted when
edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 006 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling HCO3 Stop Manual value editing attempted when
edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 007 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 163 (381)
Program version 1.1.0
C EN 2015-03-26

- Profiling HCO3 Start Model value editing attempted when


edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 008 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling HCO3 Stop Model value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 009 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling model number editing attempted when edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 010 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Mode Manual value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 011 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Mode Model value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.


164 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0117 052 012 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Steps Manual value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 013 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Steps Model value editing attempted when edit
locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 014 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Intervals Manual value editing attempted when
edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.

0117 052 015 XXX, PROFILING SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Profiling UF Intervals Model value editing attempted when
edit locked.

Technical description of conditions for occurring:


- UI request when not unlocked.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 165 (381)
Program version 1.1.0
C EN 2015-03-26

0117 053 000 XXX, BPM Software Error


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Confirmation of alarm not related to BPM.

0117 053 001 XXX, UNEXPECTED BPM CONFIRMATION


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- Confirmation of BPM alarm currently not active.

Technical description of conditions for occurring:


- May occur as a transient error if an alarm is confirmed by
the user at the same time as the alarm is cleared by a new
measurement. May also occur due to Software Error.

0117 053 002 XXX, BPM Software Error


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- BPM Measurement received before limits received.

Technical description of conditions for occurring:


- Software Error.

0117 054 000 XXX, MACHINE OVERHEATED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The machine is too hot. Disconnect patient (if applicable)
and shut down the machine immediately.

0118 020 000 XXX, UFS SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Transient Error. More detailed infomration is logged for
Gambro SW developer.

0118 020 005 XXX, UF SUPERVISION SENSORS NOT POWER SUPPLIED


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Protective system has requested the protective UF sensors
power supply to be switched on. But power status was
reported to be off.


166 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- 24V power supply malfunction.
- Power IO C failure/internal software failure.

0118 020 006 XXX, UF SUPERVISION SENSORS NOT PRESENT


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The UF Supervision does not detect one or both of the
Protective UF sensors. Information about which of the
sensors is missing is presented in this error codes extra
information. E.g. "No 1" for missing Protective UF sensor
1.

Technical description of conditions for occurring:


- GIOP reports that is is not able to access the UFS sensors
opto boards EE memory over I2C bus.

Some possible explanations why the error has occurred:


- One or both of the Protective UF sensors are disconnected.
- One or both of the Protective UF sensors or attatched cable
is broken.

0118 020 007 XXX, UF SUPERVISION INCORRECT 24 VOLTAGE


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The UF Supervision flow meters are supplied with 24 V. If
the power supplied is not 24 V the measurements may not be
accurate and this error code is generated.

Technical description of conditions for occurring:


- GIOP reports 24V supply voltage to be outside
specification.

Some possible explanations why the error has occurred:


- 24V power supply malfunction.
- Power IO C failure/internal software failure.
- UFS sensor/sensor cable short circuit.
- General IOP voltage supervision circuit failure.

0118 020 008 XXX, SUPERVISION COMMUNICATION ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Protective and control UF flow data is not updated or
accessible with the expected update rate.

Technical description of conditions for occurring:


- Protective and control UF flow data is supplied to the
protective system over the CAN bus. It the values are not
updated in a timely manner, the input data to the
protective UF supervision is not vaild. That in its turn
makes the UF supervision estimation invalid.

Some possible explanations why the error has occurred:


- CAN bus failure.
- Power IO C failure/internal software failure.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 167 (381)
Program version 1.1.0
C EN 2015-03-26

- UF cell failure/internal software failure.


- Protective UF sensors malfunction.
- Protective internal software failure.

0118 020 011 XXX, LEAKAGE/CONTAMINATION DETECTED DURING PROTECTIVE TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- A self calibration (taration) was run. It failed as the
taration constant comparision has failed. The taration
constant deviates too much compared to previous constants.

Technical description of conditions for occurring:


- Every time a new taration constant is measured and
calculated at end of a taration, its value is compared to
previous values. The deviation between consecutive
measurements of taration constants must not deviate beyond
a threshold value. This test has failed.

Some possible explanations why the error has occurred:


- Protective UF sensors broken or dirty.
- There is air passing the protective UF sensors.
- Leakage through: RIVA, DIVA or TAVA.

0118 020 012 XXX, LEAKAGE/CONTAMINATION DETECTED DURING PROTECTIVE TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Taration constant for first taration in treatment system
differs to much to value from the taration constant
measured in function check.

Technical description of conditions for occurring:


- The taration constant for the first taration in a treatment
is checked agansted the taration constant in function
check. If the difference is too large which could indicata
a fluid path problem. The test has failed.

Some possible explanations why the error has occurred:


- Protective UF sensors broken or dirty.
- There is air passing the protective UF sensors.
- Leakage through: RIVA, DIVA or TAVA.

0118 020 013 XXX, LEAKAGE/CONTAMINATION DETECTED DURING PROTECTIVE TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Taration constant for protective system differs to much to
value from the one at the start of treatment.

Technical description of conditions for occurring:


- The taration constant for the first taration in a treatment
plays the role of a reference value. Therefore any new
taration constant measurement is followed by a comparison
with the constant from first taration in treatment. This
test has failed.


168 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Protective UF sensors broken or dirty.
- There is air passing the protective UF sensors.
- Leakage through: RIVA, DIVA or TAVA.

0118 020 014 XXX, FLOW OUT OF RANGE DURING PROTECTIVE TARATION
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- At least one of the UF protective flow meters have recorded
a value of the main flow outside the allowed range
[300,700] ml/min.

Technical description of conditions for occurring:


- The protective UF flow meters have a limited measurement
flow range, within which the system should always operate.
One or two meters show flow meter values outside of this
range.

Some possible explanations why the error has occurred:


- Uncalibrated or broken. pressure regulator.
- Dirty flow restrictor.
- Dirty outlet restrictor.
- Dirty or broken flow pump.
- Dirty or broken suction pump.
- Broken or uncalibrated degassing pressure transducer.
- RIVA: leakage.
- Broken Protective UF sensor/s.
- Broken Control UF Cell.

0118 020 015 XXX, UNSTABLE FLOW DURING PROTECTIVE TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- A protective UF self calibration (taration) was run. It
failed because the measured differential flow (UF rate)
signal varied too much. A new self calibration that
followed a first attempt did also fail for the same reason.
Standard deviation of protetive UF rate too high during
taration measurement.

Technical description of conditions for occurring:


- The standard deviation of the UF rate is outside the
required range during tarations.

Some possible explanations why the error has occurred:


- There is air passing the protective UF sensors.
- There is too much air in the inlet water.
- Valves: RIVA, DIVA or TAVA are leaking
- Protective UF sensors: the sensor is dirty.
- Protective UF sensors: the sensor is broken
- Flow restrictor: the restrictor is dirty
- Flow pump: the pump is not working properly
- Degassing chamber: the o-ring at the top of the chamber is
not in its correct position fluid is leaking.
- Degassing pressure transducer: not working properly.
- Broken Control UF Cell, feeding dialysis fluid regulation
with unstable flow data.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 169 (381)
Program version 1.1.0
C EN 2015-03-26

0118 020 025 XXX, UNEXPECTED ERROR DURING PROTECTIVE TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- A protective UF self calibration (taration) was run. It
failed due to an unexpected error. A new self calibration
that followed a first attempt did also fail for the same
reason.

Technical description of conditions for occurring:


- The calculation of a new taration offset constant could not
be done.
- The taration procedure encountered an unexpected error that
should never occur.
- Function ta_StopTaration() in tara_UFS.c returned 4 or 7.
- The CPU P did not receive enough flow updates during
taration.
- The CPU P received corrupt flow updates ( extremely large
numbers or NaN ).

Some possible explanations why the error has occurred:


- The calculation of new taration offset constants failed due
to an Software Error leading to insuficient flow data
measurement points.
- An Software Error resulted in an unexpected and normally
impossible error at end of the taration procedure.
- The UF Protective sensors or FM I/O board are/is broken and
generates corrupt data.

0119 000 000 XXX, PRESET CHECK SUM ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The check sum of the protective system presets is not
correct.

Some possible explanations why the error has occurred:


- Software Error.

0119 000 001 XXX, NOT ALL PRESETS RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The protective system has not received all required
presets.

Some possible explanations why the error has occurred:


- Ui and Protective versions are not compatible.

0119 000 002 XXX, STARTUP PROBLEM, PRESETS NOT RECEIVED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The protective system has not received all required
presets.


170 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- All presets were not received within a timeout period of 1
minute.

0119 001 000 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Checksum error in safe memory, detected at startup.

Some possible explanations why the error has occurred:


- Broken CPU P safe memory or machine not properly
initialized.

0119 001 001 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Safe memory read error (except CRC errors).

Some possible explanations why the error has occurred:


- Broken CPU P safe memory or incorrect internal
configuration.

0119 001 002 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Data could not be written to the safe memory.

Some possible explanations why the error has occurred:


- Broken CPU P safe memory or incorrect internal
configuration.

0119 001 003 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Safe memory task timeout error.

Some possible explanations why the error has occurred:


- Software Error - not possible to communicate with
task_safe_mem.

0119 001 004 XXX, SAFE MEMORY ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- The safe memory was empty and has been initialized with
default values.

Some possible explanations why the error has occurred:


- This should be a transient error that should occur the
first time the machine is started. In this case old data
does not exist.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 171 (381)
Program version 1.1.0
C EN 2015-03-26

0119 001 005 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The safe memory did not comply with the current format and
has been re-initialized with default values.

Some possible explanations why the error has occurred:


- This should be a transient error that should occur the
first time the machine is started with new incompatible
software. In this case either old data can not be mapped to
the data used by the new software version.

0119 001 006 XXX, SAFE MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Malloc failed. Lack of memory in system.

0119 002 002 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 003 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 004 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.


172 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 005 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 006 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 007 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 008 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 173 (381)
Program version 1.1.0
C EN 2015-03-26

0119 002 009 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 010 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Machine has restarted due to lost heartbeat from the given
node.

Technical description of conditions for occurring:


- The indicated node has stopped working correctly. When
detected by protective a restart was executed.

0119 002 020 XXX, POWER IO P RESTARTED - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- A watchdog reset in Power IO P has caused the machine to
restart.

Technical description of conditions for occurring:


- Software Error in Power IO P.

0119 003 002 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 003 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.


174 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 004 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 005 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 006 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 007 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 008 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.
General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 175 (381)
Program version 1.1.0
C EN 2015-03-26

- Heartbeast lost or heartbeat never detected for a node


during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 009 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 003 010 XXX, HEARTBEAT ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Heartbeast lost or heartbeat never detected for a node
during startup of the machine.

Technical description of conditions for occurring:


- The startup of the given node does not work as expected, it
either does no start or performs an unexpected reset during
the startup phase of the machine.

0119 004 000 XXX, PROTECTIVE VERSION CONTROL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- At least one of the other nodes in the system has a version
not accepted by Protective.

Technical description of conditions for occurring:


- Version of at least one other node does not match the
Protective version.

Some possible explanations why the error has occurred:


- Faulty software configuration on machine.

0131 001 XXX XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.


176 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0131 002 XXX XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0131 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0131 022 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

0131 023 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

0131 024 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

0131 025 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 177 (381)
Program version 1.1.0
C EN 2015-03-26

0131 026 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

0131 027 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

0131 028 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

0131 029 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

0131 030 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

0131 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0131 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0131 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.


178 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

0131 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0131 049 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

0131 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

0132 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0134 001 XXX XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 179 (381)
Program version 1.1.0
C EN 2015-03-26

0134 002 XXX XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0134 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0134 022 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

0134 023 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

0134 024 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.


180 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

0134 025 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.

0134 026 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

0134 027 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

0134 028 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

0134 029 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

0134 030 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 181 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

0134 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0134 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0134 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

0134 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0134 049 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

0134 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.


182 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.G409.

0135 003 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency.

0135 022 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 022 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

Technical description of conditions for occurring:


- MCO Init failed.

0135 022 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 022 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 2.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 022 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info from Node 2.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 183 (381)
Program version 1.1.0
C EN 2015-03-26

0135 023 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 023 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

Technical description of conditions for occurring:


- MCO Init failed.

0135 023 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 023 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 023 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 3.

0135 024 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.


184 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 024 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

Technical description of conditions for occurring:


- MCO Init failed.

0135 024 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 024 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 024 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 4.

0135 025 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 025 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info from Node 5.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 185 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- MCO Init failed.

0135 025 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 025 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 025 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 5.

0135 026 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 026 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

Technical description of conditions for occurring:


- MCO Init failed.

0135 026 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info from Node 6.


186 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 026 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 026 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 6.

0135 027 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 027 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

Technical description of conditions for occurring:


- MCO Init failed.

0135 027 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 027 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info from Node 7.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 187 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 027 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 7.

0135 028 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 028 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

Technical description of conditions for occurring:


- MCO Init failed.

0135 028 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 028 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 8.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 028 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info from Node 8.


188 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0135 029 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 029 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

Technical description of conditions for occurring:


- MCO Init failed.

0135 029 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 029 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 029 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 9.

0135 030 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 189 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0135 030 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

Technical description of conditions for occurring:


- MCO Init failed.

0135 030 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0135 030 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0135 030 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node A.

0135 042 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 042 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK


190 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

after request to SDO Server node 2.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 042 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Time out.

0135 042 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Out of Memory.

0135 042 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0135 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0135 043 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Toggle bit not altered.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 191 (381)
Program version 1.1.0
C EN 2015-03-26

0135 043 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 043 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Time out.

0135 043 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Out of Memory.

0135 043 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0135 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0135 044 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


192 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- CANopen stack detected a CANopen client or server NACK


after request to SDO Server node 4.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 044 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 044 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Time out.

0135 044 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Out of Memory.

0135 044 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0135 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 193 (381)
Program version 1.1.0
C EN 2015-03-26

0135 045 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 045 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 045 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Time out.

0135 045 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Out of Memory.

0135 045 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.


194 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0135 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0135 049 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 049 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 049 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Time out.

0135 049 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Out of Memory.

0135 049 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 195 (381)
Program version 1.1.0
C EN 2015-03-26

after request to SDO Server node 9.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0135 049 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

0135 050 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 050 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Toggle bit not altered.

0135 050 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Time out.

0135 050 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Out of Memory.

0135 050 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more


196 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

during the last 30 days it is treated as an unconditional


error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0135 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

0220 004 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0220 005 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0220 006 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 197 (381)
Program version 1.1.0
C EN 2015-03-26

0220 008 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- received data length is not as defined.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0220 012 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0220 013 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received air detector command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0220 015 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received BPTT command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0220 018 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received BPM command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.


198 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0221 004 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- (0 > data > 1) when setting blood pump enable/disable.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 004 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting blood pump line diameter.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 005 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- (CLOSE > data > OPEN) when setting arterial clamp
open/close.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 005 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- (CLOSE > data > OPEN) when setting venous clamp open/close.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 199 (381)
Program version 1.1.0
C EN 2015-03-26

0221 006 001 XXX, RECEIVED DATA INVALID


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heparin pump syringe
diameter.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 006 002 XXX, RECEIVED DATA INVALID


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- Invalid valie when setting heparin pump enable/disable.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 008 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Corrupt data received from CAN.

0221 008 004 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- Data outside enum range.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0221 008 005 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and


200 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- Data outside enum range.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0221 009 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0221 009 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0221 010 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 201 (381)
Program version 1.1.0
C EN 2015-03-26

0221 011 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0221 011 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0221 013 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received air detector command is invalid.

Technical description of conditions for occurring:


- (AIR_LOW > data > AIR_HIGH) when setting air detector
transmitter level.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 015 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received BPTT command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting BPTT Enable/Disable.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 018 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.


202 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The data in the received BPM command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting BPM Active.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0221 018 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received BPM command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting BPM request enable disable.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0222 008 001 XXX, GENERAL CALIBRATION FAILURE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Received unknown transducer id.

Technical description of conditions for occurring:


- Received unknown transducer id.

Some possible explanations why the error has occurred:


- Internal SW error.

0222 008 002 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate venous pressure p.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could be read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0223 008 006 XXX, VENOUS PRESSURE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- I2C scanning procedure for venous pressure failed.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 203 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Failed to read/write epprom data for venous pressure.

Some possible explanations why the error has occurred:


- Epprom is broken.
- I2C interface is broken.
- Internal Error.

0223 008 007 XXX, EEPROM WRITE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Data could not be written to the EEPROM.

Some possible explanations why the error has occurred:


- Epprom is broken.
- I2C interface is broken.
- Internal Error.

0223 020 000 XXX, AIR DETECTOR TRANSMITTER ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not set DAC value.

Some possible explanations why the error has occurred:


- DAC is broken.
- I2C communication error.
- Software error.

0225 019 000 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate ADC.

Technical description of conditions for occurring:


- ADC could not be configured and calibrated.

Some possible explanations why the error has occurred:


- SPI communication error.
- ADC is broken.
- Software error.

0225 019 001 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC sampling error.

Technical description of conditions for occurring:


- ADC stopped sending data ready signal or SW error occured.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.


204 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0225 019 002 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0225 019 003 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0225 019 004 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0225 019 005 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC offset is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0225 019 006 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC slope is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0227 014 000 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 205 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- PZ24V_ERROR, Voltage out of range.

0227 014 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ12VN_ERROR, Voltage out of range.

0227 014 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ12V_ERROR, Voltage out of range.

0227 014 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ5V0_ERROR,Voltage out of range.

0227 014 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ3V3_ERROR, Voltage out of range.

0227 014 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PZ24VSAFE_ERROR, Voltage out of range.

0227 014 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ12VBPM_ERROR, Voltage out of range.

0227 014 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ2V5_ERROR, Voltage out of range.

0227 014 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CVREF_ERROR, Voltage out of range.


206 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0227 014 009 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PVREF_ERROR, Voltage out of range.

0231 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0231 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0231 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0232 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 207 (381)
Program version 1.1.0
C EN 2015-03-26

0234 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0234 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0234 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0320 004 004 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received transducer command is invalid.

Technical description of conditions for occurring:


208 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- CANOpen offset is unknown. Probably because of interask


communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.

0320 004 011 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received calibration command is invalid.

Technical description of conditions for occurring:


- The received data length is not as defined.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.

0320 005 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The data in the received UFS command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.

0320 011 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0321 004 003 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received blood leak detector command is
invalid.

Technical description of conditions for occurring:


- (EDT_DISABLE > data > EDT_ENABLE) when setting blood leak
detector transmitter level.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 209 (381)
Program version 1.1.0
C EN 2015-03-26

0321 005 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received UFS command is invalid.

Technical description of conditions for occurring:


- (FM1 > data > FM2) when setting the UFS factors.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0321 005 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received UFS command is invalid.

Technical description of conditions for occurring:


- (FM1 > data > FM2) when setting the UFS factors.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0321 007 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0321 007 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.


210 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0321 008 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0321 009 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0321 009 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0322 004 000 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Could not calibrate blood leak detector.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 211 (381)
Program version 1.1.0
C EN 2015-03-26

- I2C communication error.


- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0322 004 001 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not calibrate conductivity cell p.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0322 004 002 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not calibrate dialysis pressure.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0322 004 005 XXX, GENERAL CALIBRATION FAILURE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Received unknown transducer id.

Technical description of conditions for occurring:


- Received unknown transducer id.

Some possible explanations why the error has occurred:


- Internal SW error.

0322 004 006 XXX, BLOOD LEAK DETECTOR CALIBRATION FAILURE


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Could not save new calibration data for Bloodleak Detector.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.


212 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0322 004 007 XXX, DIALYSIS PRESSURE CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Dialysis Pressure.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0322 004 008 XXX, TEMP P CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp P (Treatment).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0322 004 009 XXX, TEMP P CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp p
(Disinfection).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0322 004 010 XXX, COND P CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 213 (381)
Program version 1.1.0
C EN 2015-03-26

The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Cond P.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0322 004 012 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received calibration command is invalid.

Technical description of conditions for occurring:


- Invalid data received.

Some possible explanations why the error has occurred:


- Software Error in NODE_GENERAL_IO_P Node.
- Control software error.

0323 014 000 XXX, EEPROM WRITE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Error when sending data to EEPROM.

Some possible explanations why the error has occurred:


- EEPROM is broken.
- I2C is broken.
- Software error.

0324 006 000 XXX, SENSOR ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Bubble trap sensors invalid inputs.

Technical description of conditions for occurring:


- Both sensors are active only one is allowed.

Some possible explanations why the error has occurred:


- Sensor is broken.
- Software Error.

0325 015 000 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC could not be calibrated.


214 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- ADC could not be configured and calibrated.

Some possible explanations why the error has occurred:


- SPI communication error.
- ADC is broken.
- Software Error.

0325 015 001 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC sampling error.

Technical description of conditions for occurring:


- ADC stopped sending data ready signal or SW error occured.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.

0325 015 002 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Internal software error.

Some possible explanations why the error has occurred:


- Software Error.

0325 015 003 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Internal software error.

Some possible explanations why the error has occurred:


- Software Error.

0325 015 004 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Internal software error.

Some possible explanations why the error has occurred:


- Software Error.

0325 015 005 XXX, ADC CALIBRATION OUT OF LIMITS


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Offset is out limit.

Technical description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 215 (381)
Program version 1.1.0
C EN 2015-03-26

- ADC channel is sent as additional information.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.

0325 015 006 XXX, ADC CALIBRATION OUT OF LIMITS


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Slope is out limit.

Technical description of conditions for occurring:


- ADC channel is sent as additional information.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.

0327 012 000 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ24V_ERROR.

0327 012 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ12VN_ERROR.

0327 012 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ12V_ERROR.

0327 012 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ5V0_ERROR.


216 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0327 012 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ3V3_ERROR.

0327 012 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PREF_ERROR.

0327 012 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ2V5_ERROR.

0327 012 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CVREF_ERROR.

0331 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0331 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 217 (381)
Program version 1.1.0
C EN 2015-03-26

information together with another error and is by that


reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0331 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0332 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0334 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0334 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.


218 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0334 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0401 013 001 XXX, HEARTBEAT ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- Protective heartbeat was lost.

Some possible explanations why the error has occurred:


- Protective error.
- CAN Error.

0420 005 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the buzzer command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0420 006 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The data in the heater command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 219 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0420 007 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the relay command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0420 008 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the valves command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0420 012 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0420 013 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the power control command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.


220 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0420 016 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the LED command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0420 017 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the concentrate pump command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.

0421 005 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the buzzer command is invalid.

Technical description of conditions for occurring:


- (BUZZER_STATE_OFF > data > BUZZER_STATE_ON) when setting
buzzer.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 006 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the heater command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heater relay
enable/disable.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 221 (381)
Program version 1.1.0
C EN 2015-03-26

0421 007 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the relay command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting safety relay
enable/disable.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 008 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the valves command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting valves open/close.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 009 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0421 010 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


222 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0421 011 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0421 011 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0421 013 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the power control command is invalid.

Technical description of conditions for occurring:


- Error when setting set supervision.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 013 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the power control command is invalid.

Technical description of conditions for occurring:


- Error when setting startup timer set.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 223 (381)
Program version 1.1.0
C EN 2015-03-26

- Corrupt data received from CAN.

0421 013 004 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the power control command is invalid.

Technical description of conditions for occurring:


- Invalid status green command.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 016 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- (LAMP_STATUS_OFF > data > (LAMP_STATUS_FLASH when setting
arterial clamp open/close.

Some possible explanations why the error has occurred:


- Software Error in POWER IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0421 017 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The data in the received concentrate pump command is
invalid.

Technical description of conditions for occurring:


- Invalid set direction command.

Some possible explanations why the error has occurred:


- Software Error in POWER IO P Node.
- Software Error in Protective Node.
- Corrupt data received from CAN.

0427 014 000 XXX, VOLTAGE ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- Battery Voltage too low.

Technical description of conditions for occurring:


- PBAT3M_ERROR.

Some possible explanations why the error has occurred:


- Backup battery is defect, removed or not charged. Real Time


224 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Clock setting may be lost if main power is switched off.

0427 014 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PPSU12M_ERROR.

0427 014 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ24PM_ERROR.

0427 014 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ24CM_ERROR.

0427 014 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PC3V3M_ERROR.

0427 014 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PC5V0M_ERROR.

0427 014 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PCREFM_ERROR.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 225 (381)
Program version 1.1.0
C EN 2015-03-26

0427 014 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PREFM_ERROR.

0427 014 009 XXX, POWER SWITCH ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Problem with Power switch.

Technical description of conditions for occurring:


- There is a short-circuit in power switch Q49. PZ24PM
measure 24 V before the power switch is enabled.

Some possible explanations why the error has occurred:


- The inrush current might have been too high.

0431 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0431 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0431 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.


226 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Unspecified condition.

0432 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0434 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0434 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0434 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 227 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0500 000 000 XXX, CASE ERROR O/FM_MAINSTATEFCN.


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error. The main state of the machine (for example
treatment or disinfection) is faulty.

0500 000 100 XXX, PRESET VALUE INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Invalid heat efficiency.

Technical description of conditions for occurring:


- EB[EEB_HeaterEfficiency] > 3.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 202 XXX, VERSION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- The version of one of the nodes is not ok.

0500 000 500 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Mesg_Wait error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 502 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Mesg_Ack_Sender error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 503 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and


228 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

the blood pump can not be run.

General description of conditions for occurring:


- K_Mesg_Send_Wait general error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 504 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Mesg_Send error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 505 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Task_Wait error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 506 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Task_Wake error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 507 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- K_Mesg_Send_Wait timeout error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 508 XXX, CMX OPERATING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX signaling error.

Some possible explanations why the error has occurred:


- Software Error.

0500 000 600 XXX, FCH TYPE DISAGREEMENT


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 229 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Extended Fch from Control & Basic Fch from Protective.

0502 000 001 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 002 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 003 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 004 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 005 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 006 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 007 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 008 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.


230 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0502 000 009 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 010 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 011 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 012 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 013 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 014 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 015 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 016 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0502 000 017 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 231 (381)
Program version 1.1.0
C EN 2015-03-26

- Software Error.

0502 003 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 006 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 007 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 008 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 008 002 XXX, BLOOD PUMP TACHO ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump tacho countup value is incorrect.

Technical description of conditions for occurring:


- Tacho count up value is out of limit (20000).

0502 009 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 009 004 XXX, MEASURED BLOOD FLOW OUT OF LIMITS


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The measured blood flow has differed too much from the
blood flow set value.

Technical description of conditions for occurring:


- BRI(BI_BloodFlActSet) BRI(BI_BloodFl) < 600 or
BRI(BI_BloodFlActSet) BRI(BI_BloodFl) > 600.

Some possible explanations why the error has occurred:


- The blood pump unit is broken.

0502 010 000 XXX, SOFTWARE ERROR


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- Software Error.


232 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0502 010 002 XXX, ESTIMATED AND MEASURED HEPARIN VOL. DIFFER TOO MUCH
Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- Conditional error, confirm by zeroset heparin flow. The
measured accumulated heparin volume differ more than 2 ml
from the estimated accumulated volume.

Technical description of conditions for occurring:


- BRI(BI_MeasVol)BRI(BI_EstAcc)<=2 or
(BRI(BI_MeasVol)BRI(BI_EstAcc)>=2.

Some possible explanations why the error has occurred:


- The heparin pump belt is broken.

0502 010 003 XXX, ESTIMATED AND MEASURED HEPARIN BOLUS VOL. DIFFER TOO MUCH
Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- Conditional error, confirm by setting bolus volume and
heparin flow rate to zero. The estimated and measured bolus
heparin volume in the control system differ more than 10%.

0502 011 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 011 002 XXX, NUMBER OF BLOOD PUMP TACHO PULSES ERROR
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump tacho countup value is incorrect.

Technical description of conditions for occurring:


- Tacho count up value is out of limit (20000).

0502 014 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 014 001 XXX, ARTERIAL PHASE TIME OUT OF LIMITS ERROR
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 014 002 XXX, VENOUS PHASE TIME OUT OF LIMITS ERROR.
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 015 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 233 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0502 016 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 017 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 017 001 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0502 018 000 XXX, SOFTWARE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Software Error.

0503 000 001 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 002 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 003 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 004 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 005 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.


234 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0503 000 006 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 007 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 008 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 009 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 010 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 011 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 012 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 013 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 014 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 235 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0503 000 015 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 016 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 000 017 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 001 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 002 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 003 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 004 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 005 001 XXX, DIA CONNECTOR SENSORS ERROR IN SAGS TEST (SENSORS TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Dia connector sensors are not working properly.

Technical description of conditions for occurring:


- During the dia connector sensors test the operator was


236 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

asked to remove from or attach the connectors to the safety


bypass block 3 times. After operator's confirmation about
the removing or attaching the dia connectors, no
confirmation about it from dia connector sensors had come.

Some possible explanations why the error has occurred:


- The mechanical dia connector sensors may be stucked.

0503 006 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 007 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 010 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 018 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 021 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 022 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 027 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 030 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 237 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0503 031 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 031 001 XXX, BICART HOLDER PRESET ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- BiCart holder is preset "Not Installed" and should be
detected as closed but it is detected open.

Some possible explanations why the error has occurred:


- Check preset for BiCart holder.

0503 034 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 035 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 037 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0503 037 001 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA current test error from Protective: Failure closing
the valve, current still above limit.

0503 037 002 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA current test error from Protective:Failure opening the
valve when requested by P and C.

0503 037 003 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


238 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CHVA current test error from Protective: Valve still opened
after P ordered close.

0503 037 004 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA current test error from Protective: Valve does not
open when ordered by P.

0503 037 005 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA current test error from Protective: Valve still opened
after C ordered close.

0503 037 006 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA current test error. Test result from Protective
unknown.

0503 037 007 XXX, CHVA TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CHVA pressure test error. Unable reach stable -300 mmHq
Degassing pressure. Or keep Degassing pressure below -150
mmHq for 5 s. CHVA might be open.

0503 071 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 072 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 073 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 080 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 239 (381)
Program version 1.1.0
C EN 2015-03-26

The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0503 080 001 XXX, VARIABLE FLOW CALIBRATION VALUE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The checksum of the EEPROM of control CPU is incorrect Make
a default preset, restart the machine and calibrate the
variable flow motor.

Some possible explanations why the error has occurred:


- Broken safeRAM (batteryRAM).

0503 080 002 XXX, VARIABLE FLOW CALIBRATION VALUE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The calibrated variable flow motor value was out of range.
It has been automatically changed to a default value. Make
a default preset, restart the machine and calibrate the
variable flow motor.

Some possible explanations why the error has occurred:


- Software Error Broken safeRAM (batteryRAM).

0503 080 003 XXX, VARIABLE FLOW HISTORY UPDATE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The variable flow history could not be updated due to some
reason below. The checksum is incorrect. Make a default
preset, restart the machine and calibrate the variable flow
motor.

Some possible explanations why the error has occurred:


- Broken safeRAM (batteryRAM).

0503 080 004 XXX, VARIABLE FLOW CALIBRATION VALUE UPDATE ERROR
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The calibrated value for the variable flow history could
not be updated due to some reason below. The checksum is
incorrect. Make a default preset, restart the machine and
calibrate the variable flow motor.

Some possible explanations why the error has occurred:


- Broken safeRAM (batteryRAM).

0503 080 005 XXX, MAXIMUM OF NUMBER OF STEPS TO INCREASE


FLOW DURING UFS FUNCTION CHECK
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


240 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- During functions check of UF Supervision the dialysis flow
shall be increased with 50 ml/min. If this is not achieved
by increasing the variable step motor with 4 step a
technical error is generated.

0504 008 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 008 006 XXX, HEATER EFFICIENCY CHECK ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 009 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 009 001 XXX, FLOW PUMP CURRENT ERROR


Unconditional error, fluid path (DIVA/TAVA and fluid pumps)
is stopped. The blood pump can still be used.

General description of conditions for occurring:


- Flow pump current too high.

Technical description of conditions for occurring:


- FRI(FI_FlowPumpCurren) > 1500mA.

Some possible explanations why the error has occurred:


- Pump motor broken.

0504 014 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 014 001 XXX, SUCTION PUMP CURRENT ERROR


Unconditional error, fluid path (DIVA/TAVA and fluid pumps)
is stopped. The blood pump can still be used.

General description of conditions for occurring:


- Suction pump current too high.

Technical description of conditions for occurring:


- FRI(FI_SucPumpCurrent) > 1500mA.

Some possible explanations why the error has occurred:


- Pump motor broken.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 241 (381)
Program version 1.1.0
C EN 2015-03-26

0504 020 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 020 011 XXX, UF CELL TARATION NOT APPROVED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- A self calibration (taration) was run. It failed as the
values of the variables FRR(FR_UFRTOLERANCE) and
FRR(FR_UFROFFSET) had changed too much since the last self
calibration. A new self calibration was run after 5
minutes. It did also fail for the same reason.

Technical description of conditions for occurring:


- CalibrDiffFlowNew differed more than 1 ml/min from
CalibrDiffFlowOld A new self calibration was done after 5
minutes. CalibrDiffFlow again differed more than 1.0 ml/min
from CalibrDiffFlowNew AND CalibrDiffFlow again differed
more than 1.0 ml/min from CalibrDiffFlowOld.

Some possible explanations why the error has occurred:


- UF cell broken or dirty There is air through the UF cell
Leakage through: RIVA, ZEVA, DIVA or TAVA.

0504 020 012 XXX, UF CELL TARATION NOT APPROVED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- A self calibration (taration) was run. It failed as the
value of the variable FRR(FR_UFRTOLERANCE) differed too
much from the value that was stored in the EEPROM or
estimated at the last performed self calibration.

Technical description of conditions for occurring:


- CALIBR DIFF FLOWnew differed more than 7.5 ml/min from
CALIBR DIFF FLOWold.

Some possible explanations why the error has occurred:


- UF cell broken or dirty There is air through the UF cell
Leakage through: RIVA, ZEVA, DIVA or TAVA.

0504 020 014 XXX, TOO LOW SELF CALIBRATION FLOW


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow measuring unit measured a flow through channel 1
that was less than 100 ml/min during self calibration
(taration).

Technical description of conditions for occurring:


- FRR(FR_CH1RATE) < 100.0 ml/min.

Some possible explanations why the error has occurred:


- Uncalibrated or broken. pressure regulator Dirty flow


242 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

restrictor. Dirty outlet restrictor Dirty or broken flow


pump:. Dirty or broken suction pump. Broken or uncalibrated
degassing pressure transducer. ZEVA: leakage RIVA: leakage.

0504 020 015 XXX, UNSTABLE FLOW SIGNALS DURING SELF CALIBRATION
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The self calibration (taration) failed because at least one
of the measured flow signals varied too much. A new self
calibration that was done 5 minutes later did also fail
because at least one of the measured flow signals varied
too much.

Technical description of conditions for occurring:


- Variance of the flow signals were calculated The
differential flow during either the offset calibration or
during the differential flow varied too much.

Some possible explanations why the error has occurred:


- Air is flowing through flow channel 1. There is too much
air in the inlet water Valves: RIVA, ZEVA, diva or TAVA are
leaking Flow cell: the flow cell is dirty. Flow cell: the
flow cell is broken Flow restrictor: the restrictor is
dirty Flow pump: the pump is not working properly Degassing
chamber: the oring at the top of the chamber is not in its
correct position fluid is leaking. Degassing pressure
transducer: not working properly.

0504 020 016 XXX, SELF CALIBRATION TIME OUT


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The self calibration (taration) executed for a too long
time.

Technical description of conditions for occurring:


- Self calibration time > max. self calibration time (max.
self calibration time = 15s).

0504 020 022 XXX, UNSTABLE UF CELL SIGNAL


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Unstable signal caused by bad transducer or unstable flow.

0504 020 023 XXX, UNSTABLE UF CELL SIGNAL


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Unstable signal caused by unstable flow.

0504 020 024 XXX, UNSTABLE SIGNAL DURING UF CELL CALIBRATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 243 (381)
Program version 1.1.0
C EN 2015-03-26

- The UF cell calibration failed because at least one of the


measured flow signals varied too much.

Some possible explanations why the error has occurred:


- Air is flowing through flow channel 1. There is too much
air in the inlet water Valves: RIVA, ZEVA, diva or TAVA are
leaking Flow cell: the flow cell is dirty. Flow cell: the
flow cell is broken Flow restrictor: the restrictor is
dirty Flow pump: the pump is not working properly Degassing
chamber: the oring at the top of the chamber is not in its
correct position. fluid is leaking. Degassing pressure
transducer: not working properly.

0504 020 025 XXX, UNSTABLE SIGNAL DURING UF OFFSET CALIBRATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The channel 1 flow signal and the channel 2 flow signal
varied too much during the main flow offset calibration in
service mode. There should be no flow through the measuring
cells.

Technical description of conditions for occurring:


- The variance of the channel 1 flow and the variance of the
channel 2 flow were out of limit.

Some possible explanations why the error has occurred:


- The flow cell is dirty The flow cell is broken
BYVA/DIVA/TAVA leakage.

0504 020 026 XXX, UF CELL LINEARITY TEST FAILED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The UF transducer failed the linearity test.

0504 020 027 XXX, UF CELL LINEARITY TEST FAILED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The UF transducer failed the linearity test.

0504 020 028 XXX, UF CELL LINEARITY TEST FAILED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The UF transducer failed the linearity test.

0504 020 029 XXX, UF CALIBRATION REFERENCE VOLUME OUT OF LIMIT


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The reference volume was outside its specified limits
during the UF calibration in service mode.

Technical description of conditions for occurring:


244 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- ORI(OI_UFCalRefVol) < PI(EPI_UFCalRefVolLl) or


ORI(OI_UFCalRefVol) > PI(EPI_UFCalRefVolHl).

Some possible explanations why the error has occurred:


- Memory RAM corrupt.

0504 020 031 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 020 032 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 028 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0504 028 003 XXX, HIGH INLET WATER CONDUCTIVITY


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Inlet water conductivity is too high.

0505 006 002 XXX, HIGH PRESSURE DURING FUNCTION CHECK


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- High pressure in monitor during function check.

Technical description of conditions for occurring:


- FRI(FI_HPGPr)>450mmHg for more than 1,5 seconds.

Some possible explanations why the error has occurred:


- Fluid path obstruction HPG broken/uncalibrated.

0505 006 003 XXX, UNSTABLE DEGASSING PRESSURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Unstable degassing pressure during function check. The
degassing pressure has been measured twice with an interval
of 15 seconds. The difference between the two measurements
was more than 20mmHg.

Technical description of conditions for occurring:


- oldDegassPr FRI(FI_DegassPr) ) < 20mmHg or oldDegassPr
FRI(FI_DegassPr) ) > 20mmHg.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 245 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Leakage on top of degassing chamber. Check the oring
Unstable/broken degassing transducer RIVA leakage.

0505 006 004 XXX, TOO BIG DIFFERENCE BETWEEN HPG AND DEGASS PRESSURE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The difference between HPG and degassing pressure has been
too big during function check of degassing transducer.

Technical description of conditions for occurring:


- FRI(FI_HPGPr) FRI(FI_DegassPr) 10mmHg.

Some possible explanations why the error has occurred:


- Leakage on top of degassing chamber. Check the oring.
Unstable/broken degassing transducer RIVA leakage.

0505 008 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0505 009 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0505 009 001 XXX, FLOW PUMP CURRENT ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Flow pump current too low.

Technical description of conditions for occurring:


- FRI(FI_FlowPumpCurrent) < 50mA.

Some possible explanations why the error has occurred:


- Air in flow pump Broken/unconnected flow pump The flow pump
is decoupled. FRL(F_FlowPumpDecoupl)=TRUE ( The flow pump
could not follow the flow pump motor rotation. The magnetic
coupling disconnected the pump from the motor. The pump
might be dirty or the flow path might be blocked).

0505 009 002 XXX, DEGASS PRESSURE CONTROL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow pump could not create the preset degassing
pressure.

Technical description of conditions for occurring:


- FRI(FI_DegassPr) + EI(EEI_DegassPr) < 60mmHg or
FRI(FI_DegassPr) + EI(EEI_DegassPr) > 60mmHg.


246 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- DRVA leakage Flow pump/motor broken The preset degassing
pressure is too low The degassing pressure transducer is
broken.

0505 014 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0505 018 002 XXX, CURRENT TEST ERROR, AIVA VALVE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through the AIVA valve has been measured with
valve closed and then with valve opened. The current
difference has not been large enough.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent) refCurrent < 200mA.

Some possible explanations why the error has occurred:


- Broken/unconnected magnet on AIVA valve.

0505 018 004 XXX, BUBBLE CHAMBER (DEAIRATING CHAMBER) ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- AIVA and REVA opened. Suction pump active and fluid tubes
connected to safety bypass. Low level in bubble chamber is
not achieved within 90 sec. The status from level sensor
indicates high level in chamber.

Technical description of conditions for occurring:


- PRL(P_BubbleTrapHigh)=TRUE.

Some possible explanations why the error has occurred:


- The AIVA or REVA could not be opened. Broken float in
bubble chamber.

0505 018 005 XXX, BUBBLE CHAMBER (DEAIRATING CHAMBER) ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- AIVA and REVA opened. Suction pump active and fluid tubes
connected to safety bypass. Low level in bubble chamber is
not achieved within 90 sec. The status from level sensor
indicates middle level in chamber.

Technical description of conditions for occurring:


- PRL(P_BubbleTrapHigh)=FALSE.

Some possible explanations why the error has occurred:


- The AIVA or REVA could not be opened. Broken float in
bubble chamber.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 247 (381)
Program version 1.1.0
C EN 2015-03-26

0505 018 006 XXX, NOT POSIBLE TO REGULATE HPG PRESSURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Impossible to regulate HPG to 100mmHg in AIVA test (120
sec. testtime).

Technical description of conditions for occurring:


- FRI(FI_HPGPR) could not be regulated to 100mmHg.

Some possible explanations why the error has occurred:


- Air in fluid path. Broken suction pump.

0505 018 007 XXX, CURRENT TEST ERROR , FIVA VALVE (1)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through FIVA valve has been measured with valve
closed and then with valve opened, despite the UFD has been
preseted NotInst. The current difference has been large
enough in order to detect that FIVA valve is installed.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent)refCurrent > 200mA and not
EL(EEL_UfdInstalled).

Some possible explanations why the error has occurred:


- The UFD has been preseted NotInst instead of Inst, despite
the FIVA valve and consequently the UFD is installed.

0505 018 008 XXX, CURRENT TEST ERROR , FIVA VALVE (2)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through FIVA valve has been measured with valve
closed and then with valve opened, with the UFD preseted
Installed. The current difference has not been large
enough.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent)refCurrent < 200mA and
EL(EEL_UfdInstalled).

Some possible explanations why the error has occurred:


- Broken/unconnected magnet on FIVA valve. The UFD has been
preseted Inst instead of NotInst, despite there is no FIVA
valve and consequently no UFD installed.

0505 018 011 XXX, CURRENT TEST ERROR, FLVA VALVE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through FLVA valve has been measured with valve
closed and then with valve opened. The current difference
has not been large enough. Broken/unconnected magnet on
FLVA valve.


248 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0505 018 012 XXX, CURRENT TEST ERROR , HBVA VALVE (1)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through HBVA valve has been measured with valve
closed and then with valve opened, despite the HBVA has
been preseted NotInst. The current difference has been
large enough in order to detect that HBVA valve is
installed.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent)-refCurrent > 200mA and not
EL(EEL_HBVAInstalled).

Some possible explanations why the error has occurred:


- The HBVA has been preseted NotInst instead of Inst, despite
the HBVA valve is installed.

0505 018 013 XXX, CURRENT TEST ERROR , HBVA VALVE (2)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through HBVA valve has been measured with valve
closed and then with valve opened, with the HBVA preseted
Installed. The current difference has not been large
enough.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent)-refCurrent < 200mA and EL(EEL_
HBVAInstalled).

Some possible explanations why the error has occurred:


- Broken/unconnected magnet on HBVA valve.
- Broken control driver board.
- The HBVA has been preseted Inst instead of NotInst, despite
there is no HBVA valve.

0505 020 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0505 028 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 008 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 249 (381)
Program version 1.1.0
C EN 2015-03-26

0506 009 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 009 001 XXX, DEGASS PRESSURE CONTROL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow pump could not create the preset degassing
pressure.

Technical description of conditions for occurring:


- FRI(FI_DegassPr) > (EI(EEI_DegassPr)60)mmHg.

Some possible explanations why the error has occurred:


- DRVA leakage Flow pump/motor broken The preset degassing
pressure is too low The degassing pressure transducer is
broken.

0506 009 002 XXX, FLOW PUMP CURRENT ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Flow pump current too low.

Technical description of conditions for occurring:


- FRI(FI_FlowPumpCurrent) < 50mA.

Some possible explanations why the error has occurred:


- Air in flow pump Broken/unconnected flow pump The flow pump
is decoupled. FRL(F_FlowPumpDecoupl)=TRUE ( The flow pump
could not follow the flow pump motor rotation. The magnetic
coupling disconnected the pump from the motor. The pump
might be dirty or the flow path might be blocked).

0506 014 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 020 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 020 001 XXX, INCORRECT FLOW FROM CHANNEL 2


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Flow deviation in channel 2.


250 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- The flow deviation in channel 2 (compared with channel 1 +
UFrate) exceeded 100ml/min for 30 seconds a second time
during the same treatment. The first Time this occur, a
forced taration is performed.

Some possible explanations why the error has occurred:


- UFcell broken.

0506 025 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 026 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 028 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 033 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 036 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0506 069 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0507 006 001 XXX, HIGH PRESSURE DURING SAGS TEST


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- High pressure in monitor during SAGS test. Main flow
through RIVA.

Technical description of conditions for occurring:


- FRI(FI_HPGPr)>450mmHg for more than 5 seconds.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 251 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- RIVA unable to open Fluid path obstruction HPG
broken/uncalibrated.

0507 008 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0507 008 001 XXX, REFERENCE TEMP ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The reference temperature could not be set.

0507 009 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0507 009 001 XXX, FAILED TO CALCULATE RAMP DATA


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- In circulation dwell state the ramp data could not be
calculated.

0507 009 002 XXX, REFERENCE PRESSURE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- In control degassing pressure state the reference pressure
could not be calculated.

0507 009 003 XXX, PID VALUE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- In control flowpump on constant speed state the PID value
could not be calculated.

0507 009 004 XXX, REFERENCE_FLOW_ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0507 014 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


252 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0507 014 001 XXX, REFERENCE PRESSURE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- In state control HPG pressure the reference pressure could
not be set.

0507 014 002 XXX, REFERENCE FLOW ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- In state run pump on constant flow the reference flow could
not be set.

0507 018 001 XXX, CURRENT TEST ERROR, DRVA VALVE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through the DRVA valve has been measured with
valve closed and then with valve opened. The current
difference has not been large enough.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent) refCurrent < 200mA.

Some possible explanations why the error has occurred:


- Broken/unconnected magnet on DRVA valve.

0507 018 009 XXX, CURRENT TEST ERROR, DRVA VALVE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through the DRVA valve has been measured with
valve closed and then with valve opened. The current
difference is too big.

Technical description of conditions for occurring:


- FRI(FI_ValvesCurrent) refCurrent > 200mA.

0507 018 010 XXX, CURRENT TEST ERROR, CBVA VALVE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The current through CBVA valve has been measured with valve
closed and then with valve opened. The current difference
has not been large enough. Broken/unconnected magnet on
CBVA valve.

0507 020 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 253 (381)
Program version 1.1.0
C EN 2015-03-26

- Software Error.

0507 028 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0507 028 001 XXX, TOO LOW CHEMICAL MIXING RATIO


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The chemical mixing ratio is too low.

Technical description of conditions for occurring:


- ORI(OI_ChemMixRatio) < 9.

Some possible explanations why the error has occurred:


- Incorrect preset of concentration preset. Software Error.

0508 008 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 009 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 009 001 XXX, FLOW PUMP CURRENT ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Flow pump current too low.

Technical description of conditions for occurring:


- FRI(FI_FlowPumpCurrent) < 50mA.

Some possible explanations why the error has occurred:


- Air in flow pump Broken/unconnected flow pump The flow pump
is decoupled. FRL(F_FlowPumpDecoupl)=TRUE ( The flow pump
could not follow the flow pump motor rotation. The magnetic
coupling disconnected the pump from the motor. The pump
might be dirty or the flow path might be blocked).

0508 009 002 XXX, DEGASS PRESSURE CONTROL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow pump could not create the preset degassing
pressure.


254 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- FRI(FI_DegassPr) > (EI(EEI_DegassPr)60)mmHg.

Some possible explanations why the error has occurred:


- DRVA leakage Flow pump/motor broken The preset degassing
pressure is too low The degassing pressure transducer is
broken.

0508 009 004 XXX, UNSTABLE MAIN FLOW


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Unstable main flow during UF cell calibration.

Technical description of conditions for occurring:


- The main flow has been outside limit (+/ 5ml/min).

Some possible explanations why the error has occurred:


- Broken UF cell Leakage in fluid path Air in fluid path The
flow pump is decoupled. FRL(F_FlowPumpDecoupl)=TRUE ( The
flow pump could not follow the flow pump motor rotation.
The magnetic coupling disconnected the pump from the motor.
The pump might be dirty or the flow path might be blocked).

0508 014 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 020 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 023 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 028 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0508 068 000 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 255 (381)
Program version 1.1.0
C EN 2015-03-26

0509 000 001 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 002 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 003 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 004 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 005 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 006 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 007 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 008 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 009 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


256 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Software Error.

0509 000 011 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 012 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 013 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 014 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 015 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 016 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 000 017 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 043 000 XXX, DISPLAY INFO HANDLING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 043 001 XXX, DISPLAY INFO HANDLING ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 257 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0509 045 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 045 001 XXX, INCORRECT BLOOD TUBE DIAMETER (PRESET VALUE)
Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- The parameter for the set value of the blood tube diameter
is out of limit.

Technical description of conditions for occurring:


- ORI(OI_BlTubeDiam) < PI(EPI_BlTubeDiamLl) or
ORI(OI_BlTubeDiam) > PI(EPI_BlTubeDiamHl).

Some possible explanations why the error has occurred:


- Software Error Read/write memory error (represet and
restart machine).

0509 049 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 050 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 051 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 054 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 060 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


258 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0509 069 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 072 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 073 000 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 073 001 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 073 002 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 074 000 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 259 (381)
Program version 1.1.0
C EN 2015-03-26

0509 076 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 079 000 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 079 001 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 079 002 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 080 001 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 084 000 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 084 001 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional


260 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Software Error.

0509 088 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 089 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 089 006 XXX, REAL TIME CLOCK ERROR (READ)


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- During the startup, the function detected that the RTC
clock was not set.

Some possible explanations why the error has occurred:


- Real time clock backup battery empty or powercycled. Power
IO C board could also have been replaced.

0509 089 007 XXX, REAL TIME CLOCK ERROR (READ)


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- RTU Clock deviation.

Some possible explanations why the error has occurred:


- During the startup, the function detected that the Power IO
C RTC drifted towards control CPU clock.

0509 090 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 093 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 094 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 261 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0509 095 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 096 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0509 096 001 XXX, UNKNOWN PROFILING MODEL


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Unknown profiling model.

0510 088 001 XXX, AIR DETECTOR TEST ERROR, CPU P


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P air detector does not detect "air" when the
transmitted intensity is lowered during Fch.

Technical description of conditions for occurring:


- PRI(PI_AirDetAirPErrCnt) > 10.

0510 088 002 XXX, AIR DETECTOR TEST ERROR, CPU C


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C air detector does not detect "air" when the
transmitted intensity is lowered during Fch.

Technical description of conditions for occurring:


- PRI(PI_AirDetAirCErrCnt) > 10.

0510 088 003 XXX, AIR DETECTOR TEST ERROR, CPU P AND C
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- One or both of CPU P and CPU C's air detectors does not
sense alarm free status when the transmitted intensity is
increased to normal level. This error code should only
occure if anything is done with the air detector during the
low intensity test phase, since this condition (alarm free
status) was already fulfilled at the start of the test.

Technical description of conditions for occurring:


- PRI(PI_AirDetNoAirErrCnt) > 10.

Some possible explanations why the error has occurred:


- Low level in venous drip chamber. Check the level and
restart the monitor.


262 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0510 088 008 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING (FIRST TEST)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump flow
value is too low AND the RPM is too low (first check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpEdge1ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Blood pump not connected to BM I/O board, Pump rotation
indicator AND tacho not connected to BM I/O. Blood pump
electronics broken.

0510 088 009 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING (SECOND TEST)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump flow
value is too low AND the RPM is too low (second check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpEdge2ErrCnt) > 4.

Some possible explanations why the error has occurred:


- Blood pump not connected to BM I/O board, Pump rotation
indicator AND tacho (P604) not connected to BM I/O. Blood
pump electronics broken.

0510 088 010 XXX, BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to stop by CPU P. The measured
values from flow and/or RPM indicator show that the blood
pump is still running.

Technical description of conditions for occurring:


- PRI(PI_BlPumpPStopErrCnt) > 4.

Some possible explanations why the error has occurred:


- Broken BM I/O board.

0510 088 011 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump flow
value is OK but the RPM is too low (first check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpRotate1ErrCnt) > 10.

Some possible explanations why the error has occurred:


- The rotation indicator is broken or unconnected to BM I/O
board. Broken BM I/O board.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 263 (381)
Program version 1.1.0
C EN 2015-03-26

0510 088 012 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump flow
value is OK but the RPM is too low (second check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpRotate2ErrCnt) > 4.

Some possible explanations why the error has occurred:


- The rotation indicator is broken or unconnected. Broken BM
I/O board.

0510 088 013 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump RPM is
OK but flow value is too low (first check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpTach1ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Blood pump tacho broken or unconnected to BM I/O board.

0510 088 014 XXX, BLOOD PUMP TEST ERROR, PUMP NOT RUNNING
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to run. The blood pump RPM is
OK but flow value is too low (second check).

Technical description of conditions for occurring:


- PRI(PI_BlPumpTach2ErrCnt) > 4.

Some possible explanations why the error has occurred:


- Blood pump tacho broken or unconnected to BM I/O board.

0510 088 015 XXX, BLOOD PUMP TEST ERROR, PUMP NOT STOPPED (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The blood pump was commanded to stop by CPU C. The measured
values from flow and/or RPM indicator shows that the blood
pump is still running.

Technical description of conditions for occurring:


- PRI(PI_BlPumpWStopErrCnt) > 4.

Some possible explanations why the error has occurred:


- Broken BM I/O board.

0510 088 016 XXX, COMMUNICATION ERROR BETWEEN PROTECTIVE and CONTROL (BM1)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C does not agree on that the blood monitor function
check part 1 is ready.


264 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- PRI(PI_Bm1ErrCnt) > 2.

Some possible explanations why the error has occurred:


- Broken protective CPU board.

0510 088 017 XXX, COMMUNICATION ERROR BETWEEN PROTECTIVE and CONTROL (BM2)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C does not agree on that the blood monitor function
check part 2 is ready.

Technical description of conditions for occurring:


- PRI(PI_Bm2ErrCnt) > 2.

Some possible explanations why the error has occurred:


- Broken protective CPU board.

0510 088 018 XXX, BUZZER TEST ERROR, BUZZER NOT ON PROTECTIVE
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The microphone can not detect that the CPU P buzzer is on.

Technical description of conditions for occurring:


- PRI(PI_BuzzPOnErrCnt) > 3.

Some possible explanations why the error has occurred:


- CPU P buzzer is broken. Microphone broken.

0510 088 019 XXX, BUZZER TEST ERROR, BUZZER NOT OFF PROTECTIVE
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The microphone can not detect that the CPU P buzzer is off.

Technical description of conditions for occurring:


- PRI(PI_BuzzPOffErrCnt) > 3.

Some possible explanations why the error has occurred:


- It is not possible to mute buzzer.

0510 088 020 XXX, BUZZER TEST ERROR, BUZZER NOT SILENT
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The microphone can not detect that both the CPU P and the
CPU UI buzzer are off when the buzzer test starts.

Technical description of conditions for occurring:


- PRI(PI_BuzzSilentErrCnt) > 3.

Some possible explanations why the error has occurred:


- It is not possible to mute buzzer.

0510 088 021 XXX, BUZZER TEST ERROR, BUZZER NOT ON (CPU UI)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The microphone can not detect that the CPU UI buzzer is on.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 265 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- PRI(PI_BuzzUIOnErrCnt) > 3.

Some possible explanations why the error has occurred:


- CP UI Buzzer broken.

0510 088 022 XXX, BUZZER TEST ERROR, BUZZER NOT OFF (CPU UI)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The microphone can not detect that the CPU UI buzzer is
off.

Technical description of conditions for occurring:


- PRI(PI_BuzzUIOffErrCnt) > 3.

Some possible explanations why the error has occurred:


- It is not possible to mute buzzer.

0510 088 023 XXX, COMMUNICATION ERROR BETWEEN CPU C AND P (COM)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C does not agree on that the common function check is
ready.

Technical description of conditions for occurring:


- PRI(PI_ComErrCnt) > 2.

0510 088 042 XXX, PRIMING DETECTOR TEST ERROR,


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- In priming detector test, Protective fails to read stored
priming detector calibration current.

Technical description of conditions for occurring:


- PRI(PI_PrimDetCalibCurrErrCnt) > 4.

0510 088 043 XXX, PRIMING DETECTOR TEST ERROR, CURRENT OUT OF LIMIT
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The difference between the stored and the measured priming
detector current was too big.

Technical description of conditions for occurring:


- PRI(PI_PrimDetCurrErrCnt) > 4.

Some possible explanations why the error has occurred:


- Priming detector is uncalibrated. Priming detector board is
broken. Priming detector transmitter electronics broken.

0510 088 044 XXX, PRIMING DETECTOR TEST ERROR, P SENSES BLOOD
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P sensed "blood" when the priming detector test is
started and the transmitter intensity is high.


266 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- PRI(PI_PrimDetEmptyP1ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector not connected. Priming detector is not
empty or priming detector dirty.

0510 088 045 XXX, PRIMING DETECTOR TEST ERROR, C SENSES BLOOD
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C sensed "blood" when the priming detector test is
started and the transmitter intensity is high.

Technical description of conditions for occurring:


- PRI(PI_PrimDetEmptyW1ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector not connected. Priming detector is not
empty or priming detector dirty.

0510 088 046 XXX, PRIMING DETECTOR TEST ERROR, NO BLOOD DETECTED (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P does not sense "blood" when the priming detector
intensity is lowered.

Technical description of conditions for occurring:


- PRI(PI_PrimDetBloodPErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector board is broken.

0510 088 047 XXX, PRIMING DETECTOR TEST ERROR, NO BLOOD DETECTED (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C does not sense "blood" when the priming detector
intensity is lowered.

Technical description of conditions for occurring:


- PRI(PI_PrimDetBloodWErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector board is broken.

0510 088 048 XXX, PRIMING DETECTOR TEST ERROR, P SENSES BLOOD
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P sensed "blood" when the priming detector test
intensity is high again after the low intensity test.

Technical description of conditions for occurring:


- PRI(PI_PrimDetEmptyP2ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector is not empty or priming detector dirty.
- Priming detector board is broken.
- Priming detector transmitter electronics broken (IC15, Q9,


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 267 (381)
Program version 1.1.0
C EN 2015-03-26

Q10, IC12, V62) on Front IO board.


- Priming detector receiver electronics broken (IC30, V38,
V39, IC59, V52, V53) on Front IO board.
- Software error.

0510 088 049 XXX, PRIMING DETECTOR TEST ERROR, C SENSES BLOOD
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C sensed "blood" when the priming detector test
intensity is high again after the low intensity test.

Technical description of conditions for occurring:


- PRI(PI_PrimDetEmptyW2ErrCnt) > 10.

Some possible explanations why the error has occurred:


- Priming detector not connected. Priming detector is not
empty or priming detector dirty.

0510 088 050 XXX, SAFETY RELAY TEST ERROR, REFERENCE CURRENT TOO HIGH
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- At the beginning of the relay test the current is measured
when the relay is OFF (this current value is used as
reference current in the following tests). The reference
current measured was too high.

Technical description of conditions for occurring:


- PRI(PI_RelayOff1ErrCnt) > 4.

Some possible explanations why the error has occurred:


- The safety relay is broken The valve current measuring
system is broken.

0510 088 051 XXX, SAFETY RELAY TEST ERROR, CURRENT DIFFERENCE TOO LARGE
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The difference between the reference current and the second
current value, measured when the relay is OFF, is too high
(the safety relay is OFF and BYVA is open).

Technical description of conditions for occurring:


- PRI(PI_RelayOff2ErrCnt) > 4.

Some possible explanations why the error has occurred:


- The safety relay is broken The valve current measuring
system is broken.

0510 088 052 XXX, SAFETY RELAY TEST ERROR, CURRENT


DIFFERENCE TOO SMALL
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The difference between the reference current and the first
current value, measured when the relay is ON, is too small
(safety relay is ON and BYVA is open).

Technical description of conditions for occurring:


- PRI(PI_RelayOn1ErrCnt) > 4.


268 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- BYVA is not connected (or broken). The safety relay is
broken. The valve current measuring system is broken.

0510 088 053 XXX, SAFETY RELAY TEST ERROR, CURRENT DIFFERENCE TOO SMALL
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The difference between the reference current and the second
current value, measured when the relay is ON, is too small
(safety relay is ON and BYVA is open).

Technical description of conditions for occurring:


- PRI(PI_RelayOn2ErrCnt) > 4.

Some possible explanations why the error has occurred:


- BYVA is not connected (or broken). The safety relay is
broken. The valve current measuring system is broken.

0510 088 054 XXX, SAFETY RELAY TEST ERROR, CURRENT DIFFERENCE TOO LARGE
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The difference between the reference current and the third
current value, measured when the relay is ON, is too high
(safety relay is ON and BYVA is closed.

Technical description of conditions for occurring:


- PRI(PI_RelayOn3ErrCnt) > 4.

Some possible explanations why the error has occurred:


- The valve current measuring system is broken.

0510 088 069 XXX, VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN AT START
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The venous clamp is closed when the clamp test starts (the
clamp is commanded open).

Technical description of conditions for occurring:


- PRI(PI_VClampOpen1ErrCnt) > 4 BRL(B_VenClampClosed)=TRUE.

Some possible explanations why the error has occurred:


- The venous clamp is stucked in closed position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 070 XXX, VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The venous clamp does not open after it was closed by CPU
P.

Technical description of conditions for occurring:


- PRI(PI_VClampOpen2ErrCnt) > 4 BRL(B_VenClampClosed)=TRUE.

Some possible explanations why the error has occurred:


- The venous clamp is stucked in closed position or clamp


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 269 (381)
Program version 1.1.0
C EN 2015-03-26

position is not correctly reported by IO, check CAN


signals.

0510 088 071 XXX, VENOUS CLAMP TEST ERROR, CLAMP NOT OPEN (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The venous clamp does not open after it was closed by CPU
C.

Technical description of conditions for occurring:


- PRI(PI_VClampOpen3ErrCnt) > 4 BRL(B_VenClampClosed)=TRUE.

Some possible explanations why the error has occurred:


- The venous clamp is stucked in closed position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 072 XXX, VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P can not close the venous clamp.

Technical description of conditions for occurring:


- PRI(PI_VClampPCloseErrCnt) > 10
BRL(B_VenClampClosed)=FALSE.

Some possible explanations why the error has occurred:


- The venous clamp is stucked in open position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 073 XXX, VENOUS CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C can not close the venous clamp.

Technical description of conditions for occurring:


- PRI[PI_VClampCCloseErrCnt] > 10
- ORL[O_FchVClampTestAttnConfirmed]=TRUE.

Some possible explanations why the error has occurred:


- The venous clamp is stucked in open position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 122 XXX, VENOUS ZERO PRESSURE TEST ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Zero pressure was not achieved during function check.
Attention issued and confirmed but problem remains.

Technical description of conditions for occurring:


- ((PRI[PI_VenPrErrCnt] > 3) && (VenPrErrWasAcked)).

Some possible explanations why the error has occurred:


- Connector (tube) still connected to the machine or sensors
not calibrated.


270 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0510 088 132 XXX, ARTERIAL CLAMP TEST ERROR, CLAMP NOT OPEN AT START
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The arterial clamp test failed, either when opening or when
closing the clamp.

Technical description of conditions for occurring:


- PRI[PI_AClampOpen1ErrCnt] > 4
- ORL[O_FchAClampTestAttnConfirmed].

Some possible explanations why the error has occurred:


- The arterial clamp is stucked in closed position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 133 XXX, ARTERIAL CLAMP TEST ERROR, CLAMP NOT OPEN (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The arterial clamp does not open after it was closed by CPU
P.

Technical description of conditions for occurring:


- PRI(PI_AClampOpen2ErrCnt) > 4
- ORL[O_FchAClampTestAttnConfirmed].

Some possible explanations why the error has occurred:


- The arterial clamp is stucked in closed position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 134 XXX, ARTERIAL CLAMP TEST ERROR, CLAMP NOT OPEN (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The arterial clamp does not open after it was closed by CPU
C.

Technical description of conditions for occurring:


- PRI(PI_AClampOpen3ErrCnt) > 4
- ORL[O_FchAClampTestAttnConfirmed].

Some possible explanations why the error has occurred:


- The arterial clamp is stucked in closed position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 135 XXX, ARTERIAL CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU P)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU P can not close the arterial clamp.

Technical description of conditions for occurring:


- PRI(PI_AClampPCloseErrCnt) > 10
- ORL[O_FchAClampTestAttnConfirmed].

Some possible explanations why the error has occurred:


- The arterial clamp is stucked in open position or clamp
position is not correctly reported by IO, check CAN


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 271 (381)
Program version 1.1.0
C EN 2015-03-26

signals.

0510 088 236 XXX, ARTERIAL CLAMP TEST ERROR, CLAMP NOT CLOSED (CPU C)
Unconditional error, blood path is stopped.

General description of conditions for occurring:


- CPU C can not close the arterial clamp.

Technical description of conditions for occurring:


- PRI(PI_AClampCCloseErrCnt) > 10
- ORL[O_FchAClampTestAttnConfirmed].

Some possible explanations why the error has occurred:


- The arterial clamp is stucked in open position or clamp
position is not correctly reported by IO, check CAN
signals.

0510 088 243 XXX, ARTERIAL ZERO PRESSURE TEST ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Zero pressure was not achieved during function check.
Attention issued and confirmed but problem remains.

Technical description of conditions for occurring:


- ((PRI[PI_ArtPrErrCnt] > 3) && ArtPrErrWasAcked).

Some possible explanations why the error has occurred:


- Connector (tube) still connected to the machine or sensors
not calibrated.

0511 088 004 XXX, BLOOD LEAK DETECTOR TEST ERROR, VALUE TOO HIGH
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The blood leak detector value is too high when the LED is
off.

Technical description of conditions for occurring:


- PRI(PI_BlLeakOffLevelErrCnt) > 4 Blood leak detector value
> 2mV measured by ADC (CPU P).

Some possible explanations why the error has occurred:


- Blood leak detector uncalibrated. Blood leak detector unit
broken. Too much external light in blood leak detector
(Dialysis machine open for service). Broken/uncalibrated
A/D system.

0511 088 005 XXX, BLOOD LEAK DETECTOR TEST ERROR, VALUE OUT OF LIMIT
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The blood leak detector value is out of the limits when the
LED is lit.

Technical description of conditions for occurring:


- PRI(PI_BlLeakOnLevelErrCnt) > 4 Blood leak detector value <
10 or blood leak detector value > +60 (PRI(PI_BlLeakMeas)).


272 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Blood leak detector uncalibrated. Blood leak detector
dirty. Blood leak detector broken. Broken/uncalibrated A/D
system.

0511 088 006 XXX, BLOOD LEAK DETECTOR TEST ERROR, UNSTABLE OFF
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The blood leak detector value is unstable when the LED is
turned off.

Technical description of conditions for occurring:


- PRI(PI_BlLeakOffStableErrCnt) > 4 Blood leak detector
variance >5.

Some possible explanations why the error has occurred:


- Blood leak detector broken. Broken/uncalibrated A/D
system.

0511 088 007 XXX, BLOOD LEAK DETECTOR TEST ERROR, UNSTABLE ON
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The blood leak detector value is unstable when the LED is
turned on.

Technical description of conditions for occurring:


- PRI(PI_BlLeakOnStableErrCnt) > 4 Blood leak detector
variance >5.

Some possible explanations why the error has occurred:


- Blood leak detector broken. Broken/uncalibrated A/D system.

0511 088 024 XXX, CONDUCTIVITY TEST ERROR, B AND P DIFFERS TOO MUCH
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The conductivity difference between CPU C and CPU P is too
large in the conductivity test high phase. The difference
between CondA (FRI(FI_MM90CONDA)) and CondP (PRI(PI_CondP))
is outside limit +/ 1mS/cm The difference between CondA
(FRI(FI_MM90CONDA)) and CondB (FRI(FI_MM90CONDB)) is
outside limit +/ 0.3mS/cm.

Technical description of conditions for occurring:


- PRI(PI_CondHighDiffErrCnt) > 4 The difference between CondB
(FRI(FI_MM90CONDB)) and CondP (PRI(PI_CondP)) is outside
limit +/ 1mS/cm.

Some possible explanations why the error has occurred:


- Conductivity cell B is broken or uncalibrated Conductivity
cell P is broken or uncalibrated Conductivity cell A is
broken or uncalibrated The A/D converters are uncalibrated.

0511 088 025 XXX, CONDUCTIVITY TEST ERROR, CONDUCTIVITY TOO LOW
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 273 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The conductivity measured by CPU P is not high enough in
the conductivity test high phase.

Technical description of conditions for occurring:


- PRI(PI_CondHighLevelErrCnt) > 4 Conductivity value from
CondP(PRI(PI_CondP)).

Some possible explanations why the error has occurred:


- Incorrect concentrate used (not possible to regulate
conductivity >20mS/cm) Air leakage (pickup tube, tube,
filter) Conductivity cell B is broken or uncalibrated
Conductivity cell P is broken or uncalibrated Conductivity
cell A is broken or uncalibrated A/D converter CPU P is
uncalibrated.

0511 088 026 XXX, CONDUCTIVITY TEST ERROR, CONDUCTIVITY NOT STABLE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The conductivity value measured in CPU P is not stable in
the high conductivity test.

Technical description of conditions for occurring:


- PRI(PI_CondHighStableErrCnt) > 4.

Some possible explanations why the error has occurred:


- Too much air in fluid system Conductivity cell B is broken
or uncalibrated Conductivity cell P is broken or
uncalibrated Conductivity cell A is broken or uncalibrated.

0511 088 027 XXX, CONDUCTIVITY TEST ERROR, B AND P DIFFERS TOO MUCH
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The conductivity difference between CPU C and CPU P is too
large in the conductivity test low phase.

Technical description of conditions for occurring:


- PRI(PI_CondLowDiffErrCnt) > 4. The difference between
CondB(FRI(FI_MM90CONDB)) and CondP(PRI(PI_CondP)) is
outside limit +/ 1mS/cm The difference between CondA
(FRI(FI_MM90CONDA)) and CondP (PRI(PI_CondP)) is outside
limit +/ 1mS/cm The difference between CondA
(FRI(FI_MM90CONDA)) and CondB (FRI(FI_MM90CONDB))
(PRI(PI_CondP)) is outside limit +/ 0.3mS/cm.

Some possible explanations why the error has occurred:


- Conductivity cell B is broken or uncalibrated Conductivity
cell P is broken or uncalibrated Conductivity cell A is
broken or uncalibrated The A/D converters are uncalibrated.

0511 088 028 XXX, CONDUCTIVITY TEST ERROR, CONDUCTIVITY TOO HIGH
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


274 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The conductivity measured by CPU P is not low enough in the
conductivity test low phase.

Technical description of conditions for occurring:


- PRI(PI_CondLowLevelErrCnt) > 4 Conductivity value from
CondP (PRI(PI_CondP)) was >5 mS/cm.

Some possible explanations why the error has occurred:


- Conductivity cell B is broken or uncalibrated Conductivity
cell P is broken or uncalibrated Conductivity cell A is
broken or uncalibrated.

0511 088 029 XXX, CONDUCTIVITY TEST ERROR, CONDUCTIVITY NOT STABLE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The conductivity value measured in CPU P is not stable in
the low conductivity test.

Technical description of conditions for occurring:


- PRI(PI_CondLowStableErrCnt) > 4.

Some possible explanations why the error has occurred:


- Too much air in fluid system Conductivity cell B is broken
or uncalibrated Conductivity cell P is broken or
uncalibrated Conductivity cell A is broken or uncalibrated.

0511 088 030 XXX, FLOW SWITCH ERROR, FLOW DETECTED (INVA CLOSED)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow switch senses fluid flow when INVA is closed.

Technical description of conditions for occurring:


- PRI(PI_FlowOffErrCnt) > 3 PRL(P_FLOWSWITCHFLOW) = TRUE.

Some possible explanations why the error has occurred:


- Flow switch broken. Float stucked in upper position.

0511 088 031 XXX, FLOW SWITCH ERROR, NO FLOW (INLET PRESSURE OK)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The flow switch does not sense flow when the inlet pressure
(CPU C) is OK.

Technical description of conditions for occurring:


- PRI(PI_FlowOn1ErrCnt) > 3 PRL(P_FLOWSWITCHFLOW) = FALSE.

Some possible explanations why the error has occurred:


- Flow switch broken. Float stucked in lower position.

0511 088 033 XXX, COMMUNICATION ERROR BETWEEN PROTECTIVE and CONTROL (FM1)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 275 (381)
Program version 1.1.0
C EN 2015-03-26

- CPU C does not agree on that the fluid monitor function


check part 1 is ready.

Technical description of conditions for occurring:


- PRI(PI_Fm1ErrCnt) > 2.

0511 088 034 XXX, COMMUNICATION ERROR BETWEEN PROTECTIVE AND CONTROL (FM2)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C does not agree on that the fluid monitor function
check part 2 is ready.

Technical description of conditions for occurring:


- PRI(PI_Fm2ErrCnt) > 2.

0511 088 055 XXX, TEMPERATURE TEST ERROR, TOO HIGH DIFFERENCE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The temperature difference between CPU C and CPU P is too
large in the temperature test high phase. TempB is measured
in conductivity cell B.

Technical description of conditions for occurring:


- PRI(PI_TempHighDiffErrCnt) > 10 TempCTempP >1.3C or
TempCTempP <0.6C.

Some possible explanations why the error has occurred:


- Temperature sensors broken or not calibrated.

0511 088 056 XXX, TEMPERATURE TEST ERROR, TEMPERATURE TOO LOW
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The measured temperature (tempP) is lower than the test
limit in the high temperature test. The temperature
measured by tempP was below 41C even though the temperature
had been increased a second time by CPU C.

Technical description of conditions for occurring:


- PRI(PI_TempHighLevelErrCnt) > 4 tempP < 41C.

Some possible explanations why the error has occurred:


- Temperature sensors broken or not calibrated.

0511 088 057 XXX, TEMPERATURE TEST ERROR, TEMPERATURE NOT STABLE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The measured temperature value is not stable in the high
temperature test .

Technical description of conditions for occurring:


- PRI(PI_TempHighStableErrCnt) > 4.

Some possible explanations why the error has occurred:


276 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Temperature sensors broken or not calibrated. Flow


problems.

0511 088 058 XXX, TEMPERATURE TEST ERROR, TOO HIGH DIFFERENCE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The temperature difference between CPU C and CPU P is too
large in the temperature test low phase. TempC is measured
in conductivity cell B.

Technical description of conditions for occurring:


- PRI(PI_TempLowDiffErrCnt) > 4.

Some possible explanations why the error has occurred:


- Temperature sensors broken or not calibrated.

0511 088 060 XXX, DIVA/TAVA TEST ERROR, DEAIRATING LEVEL NOT HIGH
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- High level in the deairating chamber is not achieved at the
start of valve leakage (DIVA/TAVA) test.

Technical description of conditions for occurring:


- PRI(PI_ValveFillupErrCnt) > 4.

Some possible explanations why the error has occurred:


- Bad/no connection to level switch in deairating chamber.
Level switch broken. EVVA is not open. TAVA is not closed.
FIVA is malfunctioning (if the UFDkit is installed).

0511 088 061 XXX, DIVA/TAVA TEST ERROR, LEAKAGE IN HIGH PRESSURE TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The dialysis pressure was measured twice with
DIVA/TAVA/EVVA closed. The difference between the two
pressure measurements was too big.

Technical description of conditions for occurring:


- PRI(PI_ValveHighLeakErrCnt) > 4 The dialysis pressure
changed more than +/30mmHg between the two measurements.

Some possible explanations why the error has occurred:


- There is a leakage between the dialyzer connector and the
bypass nipple. Check the o ring in the dialyzer connector.
The deairating chamber is leaking DIVA/TAVA/EVVA/REVA is
leaking There is a leakage in the flow meter on the front
of the machine FIVA is leaking (if the UFDkit is
installed).

0511 088 062 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C had regulated a dialysis fluid pressure at +200mmHg,


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 277 (381)
Program version 1.1.0
C EN 2015-03-26

measured at HPG. The dialysis fluid pressure, measured at


the dialysis pressure transducer, was outside 200+/20 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveHighLevel1ErrCnt) > 4 Dialysis fluid pressure
outside 200+/20 mmHg.

Some possible explanations why the error has occurred:


- DIVA does not open The HPG transducer or the dialysis
pressure transducer is uncalibrated.

0511 088 063 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C had regulated a dialysis fluid pressure at +200mmHg,
measured at HPG. The dialysis fluid pressure, measured at
the dialysis pressure transducer, was outside 200+/20 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveHighLevel2ErrCnt) > 4 Dialysis fluid pressure
outside 200+/20 mmHg.

Some possible explanations why the error has occurred:


- The HPG transducer or the dialysis pressure transducer is
uncalibrated. Unstable dialysis flow.

0511 088 064 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE UNSTABLE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The dialysis pressure is not stable in the high pressure
test. Regulated pressure approx. +200mmHg (HPG).

Technical description of conditions for occurring:


- PRI(PI_ValveHighStableErrCnt) > 4 The dialysis fluid
pressure variance was >20.

Some possible explanations why the error has occurred:


- There is too much air in the dialysis fluid The HPG
pressure transducer is broken The dialysis fluid pressure
transducer is broken.

0511 088 065 XXX, DIVA/TAVA TEST ERROR, LEAKAGE IN LOW PRESSURE TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The dialysis pressure was measured twice with
DIVA/TAVA/EVVA closed. The difference between the two
pressure measurements was too big.

Technical description of conditions for occurring:


- PRI(PI_ValveLowLeakErrCnt) > 4 The dialysis pressure
changed more than +/20mmHg between the two measurements.

Some possible explanations why the error has occurred:


- There is a leakage between the dialyzer connector and the
bypass nipple. Check the oring in the dialyzer connector.


278 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

The deairating chamber is leaking DIVA/TAVA/EVVA/REVA is


leaking There is a leakage in the flow meter on the front
of the machine FIVA is leaking (if the UFDkit is
installed).

0511 088 066 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C had regulated a dialysis fluid pressure at 200mmHg,
measured at HPG. The dialysis fluid pressure, measured at
the dialysis pressure transducer, was outside 200+/20 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveLowLevel1ErrCnt) > 4 Dialysis fluid pressure
outside 200+/20 mmHg.

Some possible explanations why the error has occurred:


- The HPG transducer or the dialysis pressure transducer is
uncalibrated.

0511 088 067 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C had regulated a dialysis fluid pressure at 200mmHg,
measured at HPG. The dialysis fluid pressure, measured at
the dialysis pressure transducer, was outside 200+/20 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveLowLevel2ErrCnt) > 4 Dialysis fluid pressure
outside 200+/20 mmHg.

Some possible explanations why the error has occurred:


- TAVA does not open The HPG transducer or the dialysis
pressure transducer is uncalibrated.

0511 088 068 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE PRESSURE
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The dialysis pressure is not stable in the low pressure
test. Regulated pressure approx. 200mmHg (HPG).

Technical description of conditions for occurring:


- PRI(PI_ValveLowStableErrCnt) > 4 The dialysis fluid
pressure variance was >20.

Some possible explanations why the error has occurred:


- There is too much air in the dialysis fluid The HPG
pressure transducer is broken The dialysis fluid pressure
transducer is broken.

0511 088 075 XXX, HEATER TEST ERROR, HEATER ENERGIZED


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 279 (381)
Program version 1.1.0
C EN 2015-03-26

- The heater is still energized after CPU C has regulated 0%


dutycycle.

Technical description of conditions for occurring:


- PRI(PI_HeaterOff1ErrCnt) > 3 Signal PMAINS is high (+5V).

0511 088 076 XXX, HEATER TEST ERROR, HEATER ENERGIZED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is still energized when CPU P has turned the
heater relay off.

Technical description of conditions for occurring:


- PRI(PI_HeaterOff2ErrCnt) > 3 Signal PMAINS is high (+5V).

0511 088 077 XXX, HEATER TEST ERROR, HEATER ENERGIZED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is still energized after CPU C has shut off the
heater relay.

Technical description of conditions for occurring:


- PRI(PI_HeaterOff3ErrCnt) > 3 Signal PMAINS is high (+5V).

0511 088 078 XXX, HEATER TEST ERROR, HEATER ENERGIZED


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is still energized after INVA has been closed.
No flow through the flow switch.

Technical description of conditions for occurring:


- PRI(PI_HeaterOff4ErrCnt) > 3 Signal PMAINS is high (+5V).

Some possible explanations why the error has occurred:


- The flow switch is stucked in upper position.

0511 088 079 XXX, HEATER TEST ERROR, HEATER NOT ENERGIZED
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is not energized when there is flow through to
the monitor and CPU C runs the heater with duty cycle 40 %.

Technical description of conditions for occurring:


- PRI(PI_HeaterOn1ErrCnt) > 3 Signal PMAINS is low (Zero
volt).

Some possible explanations why the error has occurred:


- Float stucked in lower position. Flow switch broken.

0511 088 080 XXX, HEATER TEST ERROR, HEATER NOT ENERGIZED
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


280 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The heater is not energized when there is flow through to
the monitor and CPU C runs the heater with duty cycle 40 %.

Technical description of conditions for occurring:


- PRI(PI_HeaterOn2ErrCnt) > 3 Signal PMAINS is low (Zero
volt).

Some possible explanations why the error has occurred:


- Float stucked in lower position. Flow switch broken.

0511 088 081 XXX, HEATER TEST ERROR, HEATER NOT ENERGIZED
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is not energized when there is flow through to
the monitor and CPU C runs the heater with duty cycle 40 %.

Technical description of conditions for occurring:


- PRI(PI_HeaterOn3ErrCnt) > 3 Signal PMAINS is low (Zero
volt).

Some possible explanations why the error has occurred:


- Float stucked in lower position. Flow switch broken. Flow
switch in incorrect position (should be mounted close to
lower silicon connector).

0511 088 082 XXX, HEATER TEST ERROR, HEATER NOT ENERGIZED
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is not energized when there is flow through to
the monitor and CPU C runs the heater with duty cycle 40 %.

Technical description of conditions for occurring:


- PRI(PI_HeaterOn4ErrCnt) > 3 Signal PMAINS is low (Zero
volt).

Some possible explanations why the error has occurred:


- Float stucked in lower position. Flow switch broken. Flow
switch in incorrect position (should be mounted close to
lower silicon connector).

0511 088 083 XXX, HEATER TEST ERROR, HEATER NOT ENERGIZED
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The heater is not energized when there is flow through to
the monitor and CPU C runs the heater with duty cycle 40 %.

Technical description of conditions for occurring:


- PRI(PI_HeaterOn5ErrCnt) > 3 Signal PMAINS is low (Zero
volt).

Some possible explanations why the error has occurred:


- Float stucked in lower position. Flow switch broken. Flow
switch in incorrect position (should be mounted close to
lower silicon connector).


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 281 (381)
Program version 1.1.0
C EN 2015-03-26

0511 088 085 XXX, NO CONDUCTIVITY VALUE SENT TO CPU P (LOW TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No conductivity value sent from CPU C to CPU P during
conductivity test.

Technical description of conditions for occurring:


- PRI(PI_CondNoAchLowErrCnt) > 0 The low test has been
running for 60 seconds with no value sent from CPU C.

Some possible explanations why the error has occurred:


- Air in fluid path. Broken CPU C conductivity cell (cell B).

0511 088 086 XXX, UNSTABLE LEVEL IN DEAIRATING CHAMBER


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The level in deairating chamber is unstable. The level
should have been high.

Technical description of conditions for occurring:


- PRI(PI_ValveBubLevelErrCnt) > 10 (UFDkit not installed)
PRI(PI_ValveBubLevelErrCnt) > 65 ( UFDkit installed).

Some possible explanations why the error has occurred:


- Air leakage. Broken level detector. EVVA is not open or
malfunctioning. TAVA is not closed or malfunctioning. FIVA
is broken (if the UFDkit is preseted). UFD is not preseted,
despite the ultrafilter and FIVA is mounted.

0511 088 087 XXX, FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVATEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No 'pressure achieved' acknowledge signal sent to CPU P
from CPU C CPU is regulating HPG pressure to 200mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveNoAchLow1ErrCnt) > 4.

Some possible explanations why the error has occurred:


- Air in fluid. Check for leakage. Leakage between the
dialysis fluid tubes and the machine. Leakage between the
dialyzer connectors and the bypass unit. Broken HPG
transducer.

0511 088 088 XXX, FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No 'pressure achieved' acknowledge signal sent to CPU P
from CPU C CPU is regulating HPG pressure to +200mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveNoAchHigh1ErrCnt) > 2.


282 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Air in fluid. Check for leakage. Leakage between the
dialysis fluid tubes and the machine. Leakage between the
dialyzer connectors and the bypass unit. Broken HPG
transducer.

0511 088 089 XXX, FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No 'pressure achieved' acknowledge signal sent to CPU P
from CPU C CPU is regulating HPG pressure to 200mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveNoAchLow2ErrCnt) > 2.

Some possible explanations why the error has occurred:


- Air in fluid. Check for leakage. Leakage between the
dialysis fluid tubes and the machine. Leakage between the
dialyzer connectors and the bypass unit. Broken HPG
transducer.

0511 088 090 XXX, FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No 'pressure achieved' acknowledge signal sent to CPU P
from CPU C CPU is regulating HPG pressure to +200mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveNoAchHigh2ErrCnt) > 2.

Some possible explanations why the error has occurred:


- Air in fluid. Check for leakage. Leakage between the
dialysis fluid tubes and the machine. Leakage between the
dialyzer connectors and the bypass unit. Broken HPG
transducer.

0511 088 092 XXX, FAILURE WHEN REGULATING TEMPERATURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU C has not been able to regulate the temperature.

Technical description of conditions for occurring:


- PRI(PI_TempNoAchHigh1ErrCnt) > 0 Unstable temperature in
conductivity cell B, (FRI(FI_MM90CondTempB).

Some possible explanations why the error has occurred:


- Air in fluid path. Unstable fluid flow or unstable fluid
pressure. Broken heater relay unit.

0511 088 093 XXX, FAILURE WHEN REGULATING TEMPERATURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 283 (381)
Program version 1.1.0
C EN 2015-03-26

- CPU C has not been able to regulate the temperature (during


retest).

Technical description of conditions for occurring:


- PRI(PI_TempNoAchHigh2ErrCnt) > 0 Unstable temperature in
conductivity cell B, (FRI(FI_MM90CondTempB).

Some possible explanations why the error has occurred:


- Air in fluid path. Unstable fluid flow or unstable fluid
pressure. Broken heater relay unit.

0511 088 095 XXX, UNSTABLE CONDUCTIVITY (HIGH TEST)


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No conductivity value sent from CPU C to CPU P due to
unstable signal.

Technical description of conditions for occurring:


- PRI(PI_CondNoAchHighErrCnt) > 0 Unstable cond. value in
conductivity cell B, FRI(FI_MM90CONDB).

Some possible explanations why the error has occurred:


- Air in fluid path Broken CPU C conductivity cell (cell B).

0511 088 097 XXX, COMMUNICATION FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- CPU P has sent a request to CPU C to set the status of BYVA
and DIVA in a specified way. CPU P has not got any
confirmation of the request.

Technical description of conditions for occurring:


- PRL[P_ArtClampAvailRecv].

Some possible explanations why the error has occurred:


- Broken CPU C board Broken CPU P board.

0511 088 098 XXX, CLOSE FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The valves DIVA, BYVA, TAVA and EVVA should be closed but
the measured valve current indicates that at least one of
them is open.

Technical description of conditions for occurring:


- PRI(PI_CurrVaCloseErrCnt) > 2 The valve current indicates
that some valve is open.

Some possible explanations why the error has occurred:


- Broken Control driver board Broken Protective driver board.

0511 088 099 XXX, DIVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


284 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- DIVA should be open but the measured valve current
indicates that it is closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenDIVACPErrCnt) > 2 The valve current
indicates that DIVA is closed.

Some possible explanations why the error has occurred:


- Broken DIVA Broken Control driver board Broken Protective
driver board.

0511 088 100 XXX, DIVA P CLOSED FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- DIVA should be closed but the measured valve current
indicates that it is open. CPU C wants to have DIVA open.
CPU P wants to have DIVA closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenDIVACErrCnt) > 2 The valve current
indicates that DIVA is open.

Some possible explanations why the error has occurred:


- Broken Control driver board Broken Protective driver board.

0511 088 101 XXX, DIVA C CLOSED FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- DIVA should be closed but the measured valve current
indicates that it is open. CPU C wants to have DIVA closed.
CPU P wants to have DIVA open.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenDIVAPErrCnt) > 2 The valve current
indicates that DIVA is open.

Some possible explanations why the error has occurred:


- Broken Control driver board Broken Protective driver board.

0511 088 102 XXX, BYVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- BYVA should be open but the measured valve current
indicates that it is closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenBYVACPErrCnt) > 2 The valve current
indicates that BYVA is closed.

Some possible explanations why the error has occurred:


- Broken BYVA Broken Control driver board Broken Protective
driver board.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 285 (381)
Program version 1.1.0
C EN 2015-03-26

0511 088 103 XXX, BYVA C OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- BYVA should be open but the measured valve current
indicates that it is closed. CPU C wants to have BYVA open.
CPU P wants to have BYVA closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenBYVACErrCnt) > 2 The valve current
indicates that BYVA is closed.

Some possible explanations why the error has occurred:


- Broken BYVA Broken Control driver board Broken Protective
driver board.

0511 088 104 XXX, BYVA P OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- BYVA should be open but the measured valve current
indicates that it is closed. CPU C wants to have BYVA
closed. CPU P wants to have BYVA open.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenBYVAPErrCnt) > 2 The valve current
indicates that BYVA is closed.

Some possible explanations why the error has occurred:


- Broken BYVA Broken Control driver board Broken Protective
driver board.

0511 088 105 XXX, TAVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- TAVA should be open but the measured valve current
indicates that it is closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenTAVAErrCnt) > 2 The valve current
indicates that TAVA is closed.

Some possible explanations why the error has occurred:


- Broken TAVA Broken Protective driver board.

0511 088 106 XXX, EVVA OPEN FAILURE DURING VALVES CURRENT TEST
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- EVVA should be open but the measured valve current
indicates that it is closed.

Technical description of conditions for occurring:


- PRI(PI_CurrVaOpenEVVAErrCnt) > 2 The valve current
indicates that EVVA is closed.


286 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Broken EVVA Broken Protective driver board.

0511 088 107 XXX, SOFTWARE ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Software Error.

0511 088 108 XXX, CONDUCTIVITY TEST ERROR,


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The conductivity value measured in CPU C is not stable in
the high conductivity test.

Technical description of conditions for occurring:


- PRI (PI_CondCHighStableErrCnt) > 4.

Some possible explanations why the error has occurred:


- Air in fluid path Conductivity cell B is broken or
uncalibrated.

0511 088 109 XXX, CONDUCTIVITY TEST ERROR,


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The high conductivity test has not passed within the 20
minutes time limit.

Technical description of conditions for occurring:


- PRI (PI_CondTimeOutErrCnt) > 0.

Some possible explanations why the error has occurred:


- Air in fluid path Conductivity cell B is broken or
uncalibrated Conductivity cell P is broken or uncalibrated
Conductivity cell A is broken or uncalibrated.

0511 088 115 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The negative pressure of 350 mmHg is not accomplished.
Probable cause is a huge air leakage somewhere between INVA
and ZEVA. Other causes could be a malfunctioning HPG sensor
or a malfunctioning suction pump.

0511 088 116 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has increased above the negative
pressure of 250 mmHg during the stabilization period of 4
seconds. Probable cause is a huge air leakage somewhere
between INVA and ZEVA.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 287 (381)
Program version 1.1.0
C EN 2015-03-26

0511 088 117 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has decreased more then 10 mmHg within
10 seconds. Probable cause is a minor air leakage in ZEVA,
BYVA, RIVA or degassing chamber.

0511 088 118 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has increased more then 10 mmHg within
10 seconds. Probable cause is a minor air leakage INVA,
AIVA, CBVA or fluid path is leaking.

0511 088 119 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has increased above the negative
pressure of 250 mmHg during the stabilization period of 4
seconds. Probable cause is a huge air leakage somewhere
between INVA and ZEVA.

0511 088 120 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has decreased more then 10 mmHg within
10 seconds. Probable cause is minor air leakage in ZEVA,
BYVA, RIVA or degassing chamber.

0511 088 121 XXX, CHEMICAL VALVE LEAKAGE TEST ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The trapped pressure has increased more then 10 mmHg within
10 seconds. Probable cause is a minor air leakage in INVA,
AIVA, CHVA or fluid path is leaking.

0511 088 128 XXX, CHVA DOES NOT OPEN CORRECTLY


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Probable cause is broken or disconnected valve.

0511 088 129 XXX, UFS FAILED TARATION


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- No successful taration during function check.

Technical description of conditions for occurring:


288 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Control has tried to perform at least two tarations during


function check but none of them have been successful.

Some possible explanations why the error has occurred:


- UF sensors uncalibrated, UFS sensors difference too large,
or flow or conductivity not stable.

0511 088 130 XXX, UFS FAILED EVALUATION PHASE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Problems were detected when measuring evaluation flow.

Technical description of conditions for occurring:


- Evaluation flow could not be calculated, difference between
main flow or evaluation flow too small or UF/UFS sensor
readings differ too much.

Some possible explanations why the error has occurred:


- UF/UFS sensors uncalibrated or flow or conductivity not
stable.

0511 088 131 XXX, UFS FUNCTION CHECK TIMEOUT (10 min)
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The UFS function check has taken too long time.

Technical description of conditions for occurring:


- The tests could not be finished within 10 minutes.

Some possible explanations why the error has occurred:


- UF/UFS sensors uncalibrated or flow or conductivity not
stable.

0511 088 136 XXX, UFS MAIN FLOW PHASE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Problems were detected when measuring main flow.

Technical description of conditions for occurring:


- UF/UFS sensor readings differs too much .

Some possible explanations why the error has occurred:


- UF sensors uncalibrated, UFS sensors difference too large,
or flow or conductivity not stable.

0511 088 237 XXX, CONCENTRATE PUMPS BACKWARD


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not backward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRI[PI_ConcPumpDirBackwardErrCnt]>4.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 289 (381)
Program version 1.1.0
C EN 2015-03-26

0511 088 238 XXX, CONCENTRATE PUMPS FORWARD BY CPU C


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not forward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRIPI_ConcPumpDirForwardCErrCnt]>4.

0511 088 239 XXX, CONCENTRATE PUMPS BACKWARD


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not backward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRI[PI_ConcPumpDirBackward2ErrCnt]>4.

0511 088 240 XXX, CONCENTRATE PUMPS FORWARD BY CPU P


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not forward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRI[PI_ConcPumpDirForwardPErrCnt]>4.

0511 088 241 XXX, CONCENTRATE PUMPS FORWARD BY CPU C AND CPU P
Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not forward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRI[PI_ConcPumpDirForwardCPErrCntt]>4.

0511 088 242 XXX, CONCENTRATE PUMPS BACKWARD


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The concentrate pump direction is not backward or no
repsonse is received from IO within 5s.

Technical description of conditions for occurring:


- PRI[PI_ConcPumpDirBackward3ErrCnt]>4.

0511 088 244 XXX, UNSTABLE LEVEL IN DEAIRATING CHAMBER


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


290 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- The level in deairating chamber is unstable. The level


should have been high.

Technical description of conditions for occurring:


- PRI( PI_ValveLimitedFchBubLevelErrCnt) > 10 (UFDkit not
installed) PRI(PI_ValveLimitedFchBubLevelErrCnt) > 65 (
UFDkit installed).

Some possible explanations why the error has occurred:


- Air leakage. Broken level detector. EVVA is not open or
malfunctioning. TAVA is not closed or malfunctioning. FIVA
is broken (if the UFDkit is preseted). UFD is not preseted,
despite the ultrafilter and FIVA is mounted.

0511 088 245 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- CPU C regulated a dialysis fluid pressure of -G306200mmHg,
measured at HPG. The dialysis fluid pressure, measured at
the dialysis pressure transducer, was not below zero mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveLimitedFchNoAchLowErrCn) > 4 Dialysis fluid
pressure not below zero mmHg.

Some possible explanations why the error has occurred:


- The HPG transducer or the dialysis pressure transducer is
uncalibrated.

0511 088 246 XXX, FAILURE WHEN REGULATING HPG PRESSURE (DIVA/TAVA TEST)
Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- CPU C regulated a dialysis fluid pressure of 200mmHg,
measured at HPG. The pressure measured at HPG was not
above 100 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveLimitedFchNoAchHighErrCnt) > 4.

Some possible explanations why the error has occurred:


- Air in fluid. Check for leakage. Leakage between the
dialysis fluid tubes and the machine. Leakage between the
dialyzer connectors and the bypass unit. Broken HPG
transducer.

0511 088 247 XXX, DIVA/TAVA TEST ERROR, DIALYSIS PRESSURE OUTSIDE LIMITS
Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- CPU C had regulated a dialysis fluid pressure at +200mmHg,
measured at HPG. The pressure measured at the dialysis
pressure transducer differs more than 50 mmHg.

Technical description of conditions for occurring:


- PRI(PI_ValveLimitedFchLeakErrCnt) > 4.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 291 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- DIVA does not open The HPG transducer or the dialysis
pressure transducer is uncalibrated.

0512 006 003 XXX, FUNCTIONAL CHECK RESTARTED


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- Unstable degassing pressure during function check. The
degassing pressure has been measured twice with an interval
of 15 seconds. The difference between the two measurements
was more than 20mmHg. Test will be restarted.

Technical description of conditions for occurring:


- oldDegassPr FRI(FI_DegassPr) ) < 20mmHg or oldDegassPr
FRI(FI_DegassPr) ) > 20mmHg.

Some possible explanations why the error has occurred:


- Leakage on top of degassing chamber. Check the oring
Unstable/broken degassing transducer RIVA leakage.

0512 088 055 XXX, FUNCTIONAL CHECK RESTARTED


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The temperature difference between CPU C and CPU P is too
large in the temperature test high phase. TempB is measured
in conductivity cell B. Test will be restarted.

Technical description of conditions for occurring:


- PRI(PI_TempHighDiffErrCnt) > 10 TempCTempP >1.3C or
TempCTempP <0.6C.

Some possible explanations why the error has occurred:


- Temperature sensors broken or not calibrated.

0512 088 056 XXX, FUNCTIONAL CHECK RESTARTED


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The measured temperature (tempP) is lower than the test
limit in the high temperature test. The temperature
measured by tempP was below 41C even though the temperature
had been increased a second time by CPU C. Test will be
restarted.

Technical description of conditions for occurring:


- PRI(PI_TempHighLevelErrCnt) > 4 tempP < 41C.

Some possible explanations why the error has occurred:


- Temperature sensors broken or not calibrated.


292 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0512 088 092 XXX, FUNCTIONAL CHECK RESTARTED


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CPU C has not been able to regulate the temperature. Test
will be restarted.

Technical description of conditions for occurring:


- PRI(PI_TempNoAchHigh1ErrCnt) > 0 Unstable temperature in
conductivity cell B, (FRI(FI_MM90CondTempB).

Some possible explanations why the error has occurred:


- Air in fluid path. Unstable fluid flow or unstable fluid
pressure. Broken heater relay unit.

0513 044 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0514 044 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0514 065 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 001 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 002 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 003 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 004 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 293 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Software Error.

0516 000 005 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 006 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 007 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 008 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 009 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 010 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 011 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 012 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.


294 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0516 000 013 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 014 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 015 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 016 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 000 017 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 002 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 002 002 XXX, SAFE MEMORY CHECKSUM ERROR (INIT)


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Error reading safe memory
- Perform the following:
- - Calibrate variable flow
- - Calibrate priming detector
- - Set auto-disinfection if used
- - Perform first a Rinse and then a Heat disinfection
- - Perform decalcification and then cleaning

Some possible explanations why the error has occurred:


- Software Error Broken safeRAM (batteryRAM).

0516 002 003 XXX, PRESET VARIABLE CHECKSUM ERROR (INIT)


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 295 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The preset variable checksum was incorrect during init
(startup) phase. Make a default preset.

Some possible explanations why the error has occurred:


- Software Error preset CRC not correct (UI CRC != CONTROL
CRC).

0516 002 005 XXX, MISSING PRESET VARIABLE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- A preset variable was missing in the signals from UI.

0516 002 008 XXX, PRESET IS UNDEFINED


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- A preset set value is defined as UNDEFINED, internal table
error.

0516 002 009 XXX, PRESET SIGNAL INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The preset data from UI is invalid.

0516 002 010 XXX, MISSING PRESETS TO CONTROL


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Control measures the time from startup until the preset
checksum arrives. If the preset checksum does not arrive
within a certain time this alarm will be raised.

Technical description of conditions for occurring:


- UI is hanging, The CAN bus is not working.

0516 002 011 XXX, MISSING PRESET VARIABLE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- A preset variable was missing in the signals from UI.

Technical description of conditions for occurring:


- All presets were not received within a timeout period of 1
minute.

0516 002 012 XXX, MISSING PRESET VARIABLE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- A preset variable was missing in the signals from UI.


296 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0516 002 013 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The program version is not valid.

Technical description of conditions for occurring:


- The version number of the software is not what it was
expected to be.

Some possible explanations why the error has occurred:


- The assembly was not correctly assembled, one or several
nodes was compiled with different version number which was
detected by UI.

0516 002 014 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- An invalid index in the preset array.

0516 002 015 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- An invalid index in the preset array.

0516 002 016 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- An invalid index in the preset array.

0516 002 017 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- An invalid index in the preset array.

0516 007 000 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0516 011 001 XXX, UNSUPPORTED SETTING


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- UI referred to an unknown setting.

Technical description of conditions for occurring:


- Control retrieves an EDIT request for a setting that is
unknown to Control. Control Message handler requested


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 297 (381)
Program version 1.1.0
C EN 2015-03-26

settings pattern for non-setting.

Some possible explanations why the error has occurred:


- Version conflict between UI and Control. Communication
error. Software error.

0516 011 002 XXX, NOT EDITING SETTING


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Attempt to alter setting not in edit mode.

Technical description of conditions for occurring:


- Control retrieves an UI_REQUEST without a prior EDIT
request.

Some possible explanations why the error has occurred:


- Setting two-stage commit misinterpretation between UI and
Control. Possibly a timeout.

0516 011 003 XXX, ALREADY EDITING


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- An EDIT request was received when a setting is already in
EDIT mode.

Technical description of conditions for occurring:


- EDIT should be closed before another EDIT request is sent.

Some possible explanations why the error has occurred:


- Program error - inconsistency between UI and Control
implementation of settings pattern.

0516 011 004 XXX, UNSUPPORTED VARIABLE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Setting referred to nonexistent variable.

Technical description of conditions for occurring:


- Each setting refers to a variable in global arrays. A
setting referred to does not exist in any array.

Some possible explanations why the error has occurred:


- Program error - could be a placeholder during development
for a setting not implemented in production.

0516 011 005 XXX, INDEX OUT OF RANGE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Incorrect mapping between setting and variable.

Technical description of conditions for occurring:


- Index of variable for setting is out of bounds for array.


298 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Program error in table mapping settings to variables.

0516 011 006 XXX, INCORRECT TYPE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Setting type not supported.

Technical description of conditions for occurring:


- Unsupported base type for setting, e.g. string settings not
supported.

Some possible explanations why the error has occurred:


- Program error in table mapping settings to variables.

0516 011 007 XXX, OUT OF RANGE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Setting value not consistent with type.

Technical description of conditions for occurring:


- Range checks are made before assigning settings. E.g. a
logic value can only be TRUE or FALSE.

Some possible explanations why the error has occurred:


- Value retrieved is out of range for setting type.

0530 000 001 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 002 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 003 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 004 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 299 (381)
Program version 1.1.0
C EN 2015-03-26

0530 000 005 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 006 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 007 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 008 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 009 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 010 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 011 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 012 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 013 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


300 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Software Error.

0530 000 014 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 015 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 016 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0530 000 017 XXX, TRACO ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Software Error.

0532 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0534 001 XXX XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 301 (381)
Program version 1.1.0
C EN 2015-03-26

Error caused the machine to restart. The operator must


confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected at start up.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 002 001 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected during excecution.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 004 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.


302 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Watchdog Triggered. Fatal error occurred and execution was
halted.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 005 XXX XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- Server Internal Error. Fatal error occurred and execution
was halted.

Some possible explanations why the error has occurred:


- Unspecified condition.

0534 007 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- Division by zero or illegal memory access. Fatal error
occurred and execution was halted. The serial number is the
upper 16 bits of the CPU register SCB->CFSR if they are non
zero.

Some possible explanations why the error has occurred:


- Use autologger or GXL to read ORD[OD_HardFaultLR],
ORD[OD_HardFaultPC], ORD[OD_HardFaultPSR],
ORD[OD_HardFaultCFSR] after restart for more information.

0534 021 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 303 (381)
Program version 1.1.0
C EN 2015-03-26

0534 041 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

0534 046 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

0534 047 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

0534 048 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

0534 049 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

0535 003 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency.

0535 021 003 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


304 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.

Technical description of conditions for occurring:


- Transmit buffer overflow. SDO message lost.

0535 021 006 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.

Technical description of conditions for occurring:


- MCO Init failed.

0535 021 008 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.

Technical description of conditions for occurring:


- Out of CAN receive filters, PDO.

0535 021 009 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.

Technical description of conditions for occurring:


- Out of CAN receive filters, SDO.

0535 021 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen emergency with extra info
from Node 1.

0535 041 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 041 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 305 (381)
Program version 1.1.0
C EN 2015-03-26

the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 041 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Time out.

0535 041 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Out of Memory.

0535 041 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0535 041 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

0535 046 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.


306 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 046 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 046 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Time out.

0535 046 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Out of Memory.

0535 046 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0535 046 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 307 (381)
Program version 1.1.0
C EN 2015-03-26

0535 047 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 047 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 047 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Time out.

0535 047 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Out of Memory.

0535 047 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.


308 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0535 047 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

0535 048 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 048 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 048 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Time out.

0535 048 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Out of Memory.

0535 048 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 309 (381)
Program version 1.1.0
C EN 2015-03-26

after request to SDO Server node 8.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0535 048 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

0535 049 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 049 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Toggle bit not altered.

0535 049 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Time out.

0535 049 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- Out of Memory.

0535 049 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more


310 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

during the last 30 days it is treated as an unconditional


error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0535 049 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 9.

0620 004 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.

0620 005 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.

0620 006 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 311 (381)
Program version 1.1.0
C EN 2015-03-26

0620 006 003 XXX, SYRINGE OVERCURRENT


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- Heparin pump signals overcurrent.

Technical description of conditions for occurring:


- Overcurrent is signal when motor is overloaded.

Some possible explanations why the error has occurred:


- Heparin Pump Broken.
- SW Error.

0620 007 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received priming detector command is
invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.

0620 008 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bptt command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.

0620 014 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0620 019 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.


312 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO P Node.

0621 004 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- (BLOOD_PUMP_OFF > data > START_WITH_RAMP) or (data =
UNUSED) when setting blood pump start/stop.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 004 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- (MAX_SET_RPM < data) when setting blood pump RPM.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 004 003 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- (MAX_SET_RPM < data) when setting blood pump flow.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 004 004 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bloodpump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting blood pump line diameter.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 313 (381)
Program version 1.1.0
C EN 2015-03-26

- Software Error in Control Node.


- Corrupt data received from CAN.

0621 005 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- (CLOSE > data > OPEN) when setting arterial clamp
open/close.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 005 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received clamp command is invalid.

Technical description of conditions for occurring:


- (CLOSE > data > OPEN) when setting venous clamp open/close.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 006 001 XXX, RECEIVED DATA INVALID


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heparin pump syringe
diameter.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 006 002 XXX, RECEIVED DATA INVALID


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heparin pump flow.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.


314 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Corrupt data received from CAN.

0621 006 004 XXX, RECEIVED DATA INVALID


Unconditional error that causes the heparin pump to be
disabled. Treatment without heparin is still possible.

General description of conditions for occurring:


- The data in the received heparin pump command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heparin pump frequency.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 007 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received priming detector command is
invalid.

Technical description of conditions for occurring:


- (0 > data > 1) when setting priming detector transmitter
level.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 007 002 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received priming detector command is
invalid.

Technical description of conditions for occurring:


- Limits out of range when setting priming detector
calibration current.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 008 001 XXX, RECEIVED DATA INVALID


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- The data in the received bptt command is invalid.

Technical description of conditions for occurring:


- (DO_NOTHING > data > RELAY_TEST) when setting bptt
enable/disable.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 315 (381)
Program version 1.1.0
C EN 2015-03-26

- Software Error in Control Node.


- Corrupt data received from CAN.

0621 010 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.

0621 010 004 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- received data length is not as defined.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 010 005 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- Enum value in the received data not arterial or venous when
enabling/disabling calibration.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0621 010 006 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- Enum value in the received data not arterial or venous when
enabling/disabling calibration.

Some possible explanations why the error has occurred:


- Software Error in NODE_FRONT_IO_C Node.


316 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Software Error in Control Node.


- Corrupt data received from CAN.

0621 011 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0621 011 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0621 012 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0621 013 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 317 (381)
Program version 1.1.0
C EN 2015-03-26

- CANOpen rx fifo is overrun beacuse of to small fifo size


combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0621 013 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0621 019 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received BPM command is invalid.

Technical description of conditions for occurring:


- (0 > data > 1) when setting BPM measure start/stop.

Some possible explanations why the error has occurred:


- Software Error in FRONT IO C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0622 010 001 XXX, GENERAL CALIBRATION FAILURE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Received unknown transducer id.

Technical description of conditions for occurring:


- Received unknown transducer id.

Some possible explanations why the error has occurred:


- Internal SW error.

0622 010 002 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate venous pressure c.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could be read.

Some possible explanations why the error has occurred:


318 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- I2C communication error.


- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0622 010 003 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate arterial pressure.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could be read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0623 004 001 XXX, BLOOD PUMP COMMUNICATION FAILURE


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Failed when reading measure data from blood pump.

Technical description of conditions for occurring:


- Ten attempts to read/write i2c message failed.

Some possible explanations why the error has occurred:


- I2C interface broken.
- Internal SW Error.
- SW error in blood pump controller.

0623 004 002 XXX, BLOOD PUMP COMMUNICATION FAILURE


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Failed when sending RPM to blood pump.

Technical description of conditions for occurring:


- Ten attempts to read/write i2c message failed.

Some possible explanations why the error has occurred:


- I2C interface broken.
- Internal SW Error.
- SW error in blood pump controller.

0623 004 003 XXX, BLOOD PUMP COMMUNICATION FAILURE


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Failed when reading version data from blood pump.

Technical description of conditions for occurring:


- Ten attempts to read/write i2c message failed.

Some possible explanations why the error has occurred:


- I2C interface broken.
- Internal SW Error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 319 (381)
Program version 1.1.0
C EN 2015-03-26

- SW error in blood pump controller.

0623 007 003 XXX, PRIMING DETECTOR COMMUNICATION FAILURE


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- Ten attempts to write DAC failed.

Technical description of conditions for occurring:


- Ten attempts to write DAC failed.

Some possible explanations why the error has occurred:


- I2C interface broken.
- SW Error.

0623 010 007 XXX, VENOUS PRESSURE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- I2C scanning procedure for venous pressure failed.

Technical description of conditions for occurring:


- Failed to read/write epprom data for venous pressure.

Some possible explanations why the error has occurred:


- Epprom is broken.
- I2C interface is broken.
- Internal Error.

0623 010 008 XXX, ARTERIAL PRESSURE ERROR


Unconditional error, blood path is stopped.

General description of conditions for occurring:


- I2C scanning procedure for arterial pressure failed.

Technical description of conditions for occurring:


- Failed to read/write epprom data for arterial pressure.

Some possible explanations why the error has occurred:


- Epprom is broken.
- I2C interface is broken.
- Internal Error.

0624 009 000 XXX, SENSOR ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No nextBurst pulse from Front IO P.

Technical description of conditions for occurring:


- No contact between NODE_FRONT_IO_C and Front IO P through
nextBurst signal.

Some possible explanations why the error has occurred:


- Line is broken.
- Internal SW error in Front IO P.
- Internal SW error in NODE_FRONT_IO_C.


320 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0625 018 000 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate ADC.

Technical description of conditions for occurring:


- ADC could not be configured and calibrated.

Some possible explanations why the error has occurred:


- SPI communication error.
- ADC is broken.
- Software error.

0625 018 001 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC sampling error.

Technical description of conditions for occurring:


- ADC stopped sending data ready signal or SW error occured.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.

0625 018 002 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0625 018 003 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0625 018 004 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 321 (381)
Program version 1.1.0
C EN 2015-03-26

0625 018 005 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC offset is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0625 018 006 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC slope is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0627 015 000 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CZ24V_ERROR, Voltage out of range.

0627 015 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PZ12VN_ERROR,Voltage out of range.

0627 015 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PZ12V_ERROR, Voltage out of range.

0627 015 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PZ5V0_ERROR, Voltage out of range.

0627 015 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


322 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- PZ3V3_ERROR, Voltage out of range.

0627 015 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PZ2V5_ERROR, Voltage out of range.

0627 015 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- PVREF_ERROR, Voltage out of range.

0627 015 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CVREF_ERROR, Voltage out of range.

0627 015 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CBBPCRNT_ERROR, Voltage out of range.

0631 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0631 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0631 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 323 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0632 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0634 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0634 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0634 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last


324 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0720 004 005 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received transducer command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in GENERAL IO C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0720 004 020 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Technical description of conditions for occurring:


- The received data length is not as defined.

Some possible explanations why the error has occurred:


- Internal SW error.

0720 010 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0721 004 006 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 325 (381)
Program version 1.1.0
C EN 2015-03-26

- The data in the received UF command is invalid.

Technical description of conditions for occurring:


- (NOT_DEFINED > data > TARATION_DIFFERENTIAL_FLOW) when
setting UF state.

Some possible explanations why the error has occurred:


- Software Error in GENERAL IO C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0721 004 007 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received UF command is invalid.

Technical description of conditions for occurring:


- (UFRVOO_ON > data > UFRVOO_RESET) when setting UF raw
volume on/off.

Some possible explanations why the error has occurred:


- Software Error in GENERAL IO C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0721 004 008 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received UF command is invalid.

Technical description of conditions for occurring:


- (UFRVOO_ON > data > UFRVOO_RESET) when setting UF raw
recirculation volume on/off.

Some possible explanations why the error has occurred:


- Software Error in GENERAL IO C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0721 004 021 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received adc command is invalid.

Some possible explanations why the error has occurred:


- Internal SW error.
- Control SW error.

0721 006 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.


326 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0721 006 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0721 006 002 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0721 007 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0721 008 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 327 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0721 008 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0722 004 000 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate conductivity cell a.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0722 004 001 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate conductivity cell b.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.


328 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0722 004 002 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate degass pressure.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0722 004 003 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate guard pressure.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0722 004 004 XXX, INVALID EEPROM CALIBRATION DATA


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate regulator temperature.

Technical description of conditions for occurring:


- The calibration data is corrupt or eeprom could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0722 004 009 XXX, DEGASS PRESSURE CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for degass pressure.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 329 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 010 XXX, HIGH GUARD PRESSURE CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for High Guard
pressure.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 011 XXX, REGULATOR TEMPERATURE CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for regulator
temperature (treatment).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 012 XXX, REGULATOR TEMPERATURE CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for regulator
temperature (disinfection).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.


330 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0722 004 013 XXX, COND A CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Cond A.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 014 XXX, TEMP A CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp A (Treatment).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 015 XXX, TEMP A CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp A
(Disinfection).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 016 XXX, COND B CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Cond B.

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 331 (381)
Program version 1.1.0
C EN 2015-03-26

sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 017 XXX, COND B CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp B (Treatment).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 018 XXX, COND B CALIBRATION FAILURE


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- Could not save new calibration data for Temp B
(Disinfection).

Technical description of conditions for occurring:


- The calibration data could not be saved to eeprom on the
sensor board.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data corrupted.
- EEPROM is broken.
- Internal SW error.

0722 004 019 XXX, GENERAL CALIBRATION FAILURE


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Received unknown transducer id.

Technical description of conditions for occurring:


- Received unknown transducer id.

Some possible explanations why the error has occurred:


- Internal SW error.

0723 005 000 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


332 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- I2C error when reading the UF Cell.

Technical description of conditions for occurring:


- I2C read function returned an error code.

Some possible explanations why the error has occurred:


- I2C interface is broken.
- Software Error.
- Error in the UF Cell.

0723 005 001 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- I2C error when reading the UF Cell.

Technical description of conditions for occurring:


- I2C read function returned an error code.

Some possible explanations why the error has occurred:


- I2C interface is broken.
- Software Error.
- Error in the UF Cell.

0723 005 002 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- I2C error when reading the UF Cell.

Technical description of conditions for occurring:


- I2C read function returned an error code three times in a
row.

Some possible explanations why the error has occurred:


- I2C interface is borken.
- Software Error.
- Error in the UF Cell.

0723 005 003 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- I2C error when reading the UF Cell calibration data.

Technical description of conditions for occurring:


- The calibration data is corrupt or EEPROM could not be
read.

Some possible explanations why the error has occurred:


- I2C communication error.
- Calibration data is corrupt.
- EEPROM is broken.
- Internal SW error.

0723 005 004 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 333 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- I2C error when sending data to the UF Cell.

Technical description of conditions for occurring:


- I2C write funtion returned an error code.

Some possible explanations why the error has occurred:


- I2C interface is broken.
- Software Error.
- Error in the UF Cell.

0723 005 005 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- I2C error when reading the UF Cell software version.

Technical description of conditions for occurring:


- I2C read function returned an error code.

Some possible explanations why the error has occurred:


- I2C interface is broken.
- Software Error.
- Error in the UF Cell.

0723 005 006 XXX, UF CELL ERROR


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- SW error when reading the UF Cell.

Technical description of conditions for occurring:


- Could not take the UF Cell I2C resource.

Some possible explanations why the error has occurred:


- UF Cell I2C resource is locked by other task.
- Internal SW error.

0723 011 000 XXX, DIASCAN CELL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Invalid crc of the data received from Diascan Cell.

Technical description of conditions for occurring:


- The data received is corrupt.

Some possible explanations why the error has occurred:


- I2C interface error.
- Software Error.
- Diascan Cell error.

0723 014 000 XXX, EEPROM WRITE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


334 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Error when writing data to EEPROM.

Some possible explanations why the error has occurred:


- EEPROM is broken.
- I2C interface error.
- Software Error.

0725 015 000 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Could not calibrate ADC.

Technical description of conditions for occurring:


- ADC could not be configured and calibrated.

Some possible explanations why the error has occurred:


- SPI communication error.
- ADC is broken.
- Software error.

0725 015 001 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC sampling error.

Technical description of conditions for occurring:


- ADC stopped sending data ready signal or SW error occured.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software Error.

0725 015 002 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0725 015 003 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0725 015 004 XXX, ADC SW ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 335 (381)
Program version 1.1.0
C EN 2015-03-26

- ADC software error.

Some possible explanations why the error has occurred:


- Software error.

0725 015 005 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC offset is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0725 015 006 XXX, ADC ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- ADC slope is out of limits.

Technical description of conditions for occurring:


- Channel is sent as additional data.

Some possible explanations why the error has occurred:


- ADC is broken.
- Software error.

0727 012 000 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ24V_ERROR.

0727 012 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ12VN_ERROR.

0727 012 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


336 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- PZ12V_ERROR.

0727 012 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ5V0_ERROR.

0727 012 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ3V3_ERROR.

0727 012 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CREF_ERROR.

0727 012 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PZ2V5_ERROR.

0727 012 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PVREF_ERROR.

0727 012 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- PVREF2_ERROR.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 337 (381)
Program version 1.1.0
C EN 2015-03-26

0731 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0731 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0731 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0732 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0734 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


338 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- The given error has resulted in a restart of the machine.


This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0734 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE


RESTARTED
Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0734 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0820 004 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received heater command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 339 (381)
Program version 1.1.0
C EN 2015-03-26

0820 005 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received concentrate pumps command is
invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 006 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received flow pumps command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 008 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received valves command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 009 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received variable flow command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 010 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


340 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 014 004 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received RTC command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0820 015 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

0820 016 001 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the power control command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 018 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received WRO 300 command is invalid.

Technical description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 341 (381)
Program version 1.1.0
C EN 2015-03-26

- CANOpen offset is unknown. Probably because of interask


communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0820 020 000 XXX, COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Can SW Error.Look for error code in the additional data
sent with the technical error.

Some possible explanations why the error has occurred:


- Wrong configuration.
- CAN is broken.

0820 020 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received UF calibration command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting UF calibration command.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0820 021 000 XXX, RECEIVED UNKNOWN COMMAND


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received battery command is invalid.

Technical description of conditions for occurring:


- CANOpen offset is unknown. Probably because of interask
communication failure.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.

0821 004 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received heater command is invalid.

Technical description of conditions for occurring:


- (0 > data > 100) when setting the heater PID value.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.


342 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0821 004 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received heater command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting heater relay
enable/disable.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 004 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received heater command is invalid.

Technical description of conditions for occurring:


- Unknown power preset.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 005 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received concentrate pumps command is
invalid.

Technical description of conditions for occurring:


- Limits out of range when setting A concentrate pump flow.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 005 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received concentrate pumps command is
invalid.

Technical description of conditions for occurring:


- Limits out of range when setting B concentrate pump flow.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 343 (381)
Program version 1.1.0
C EN 2015-03-26

- Software Error in Control Node.


- Corrupt data received from CAN.

0821 005 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received concentrate pumps command is
invalid.

Technical description of conditions for occurring:


- invalid sata when setting concentrate pump direction.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 006 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received flow pumps command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting flow pump PID value.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 006 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received flow pumps command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting suction pump PID value.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 006 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received flow pumps command is invalid.

Technical description of conditions for occurring:


- (0 > data > 1) when setting flow pump start/stop.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.


344 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Software Error in Control Node.


- Corrupt data received from CAN.

0821 006 004 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received flow pumps command is invalid.

Technical description of conditions for occurring:


- (0 > data > 1) when setting suction pump start/stop.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Node.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 008 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped.
The blood pump can still be used.

General description of conditions for occurring:


- The data in the received valves command is invalid.

Technical description of conditions for occurring:


- Limits out of range when setting valves open/close.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 010 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- Invalid value when LED state.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 010 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- Invalid value when LED state.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 345 (381)
Program version 1.1.0
C EN 2015-03-26

- Corrupt data received from CAN.

0821 010 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- Invalid value when LED state.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 010 004 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- Invalid value when LED state.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 010 005 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received LED command is invalid.

Technical description of conditions for occurring:


- Invalid value when LED state.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 011 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


346 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0821 011 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0821 012 000 XXX, MAX CMX MESSAGES ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0821 013 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

0821 013 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 347 (381)
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

0821 014 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received RTC command is invalid.

Technical description of conditions for occurring:


- (0 > data < 1) when setting alarm on/off.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 014 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received RTC command is invalid.

Technical description of conditions for occurring:


- incorrect size of variable set time.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 014 003 XXX, RECEIVED DATA INVALID


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The data in the received RTC command is invalid.

Technical description of conditions for occurring:


- incorrect size of variable set alarm time.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0821 018 001 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The data in the received WRO 300 command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting WRO activate.

Some possible explanations why the error has occurred:


348 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Software Error in NODE_POWER_IO_C Nodes.


- Software Error in Control Node.
- Corrupt data received from CAN.

0821 018 002 XXX, RECEIVED DATA INVALID


Unconditional error, fluid path (DIVA/TAVA) is stopped. The
blood pump can still be used.

General description of conditions for occurring:


- The data in the received WRO 300 command is invalid.

Technical description of conditions for occurring:


- Invalid value when setting WRO command.

Some possible explanations why the error has occurred:


- Software Error in NODE_POWER_IO_C Nodes.
- Software Error in Control Node.
- Corrupt data received from CAN.

0826 014 000 XXX, REAL TIME CLOCK ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Real time clock has not been initiated.

Technical description of conditions for occurring:


- There is no data stored in the backup register.

Some possible explanations why the error has occurred:


- First time started.
- Battery backup lost.

0826 014 005 XXX, REAL TIME CLOCK ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Invalid days.

Technical description of conditions for occurring:


- Days read from BackupMemory was invalid.

Some possible explanations why the error has occurred:


- Battery backup lost.

0826 014 006 XXX, REAL TIME CLOCK ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- Invalid time.

Technical description of conditions for occurring:


- Checksum invalid for the data read from BackupMemory.

0827 016 000 XXX, BATTERY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 349 (381)
Program version 1.1.0
C EN 2015-03-26

- Battery low invalid signal. Battery low while running on


main power.

Technical description of conditions for occurring:


- The power supply unit signals battery low while signaling
that the machine is running on main power.

Some possible explanations why the error has occurred:


- Disconnected or defect cable.
- PSU Defect.
- Power board is defect.

0827 017 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CPSU24M_ERROR.

0827 017 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CPSU12M_ERROR.

0827 017 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CZ24PM_ERROR.

0827 017 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CP3V3M_ERROR.

0827 017 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- C5V0M_ERROR.


350 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0827 017 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CPREFM_ERROR.

0827 017 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- CREFM_ERROR.

0831 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0831 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0831 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 351 (381)
Program version 1.1.0
C EN 2015-03-26

0832 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0834 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

0834 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0834 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat was stopped.


352 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Some possible explanations why the error has occurred:


- Unspecified condition.

0928 001 000 XXX, SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Setting was not acknowledged.

Technical description of conditions for occurring:


- The requested setting value was not returned as the setting
value to use.

0928 001 003 XXX, SETTING ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- Setting was not acknowledged.

Technical description of conditions for occurring:


- The requested setting value was not returned within the
timeout.

0929 001 001 XXX, PRESET ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Preset calculation function not ok.

Technical description of conditions for occurring:


- FUNCTION_SIGNATURE_ERROR.

Some possible explanations why the error has occurred:


- Ui and Control is out of sync.

0929 001 002 XXX, PRESET ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Preset calculation function not ok.

Technical description of conditions for occurring:


- PRESET_DATA_FILE_CORRUPT.

Some possible explanations why the error has occurred:


- The preset binary has been corrupted.

0931 000 000 XXX, FREQUENCY SUPERVISION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 353 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- Supervision of oscillators failed to start.

Technical description of conditions for occurring:


- Unspecified condition.

0931 000 001 XXX, FREQUENCY SUPERVISION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Frequency supervisor reported an unknown error.

Technical description of conditions for occurring:


- Unspecified condition.

0931 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_RX_P.

0931 001 001 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_RX_C.

0931 001 002 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_TX.

0931 001 003 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


354 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- A defect task was detected and heartbeat was stopped. This


error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_MASTER.

0931 001 004 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_SLAVE.

0931 001 005 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: UI.

0931 001 006 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Defect task: CANOPEN_RX_P.

0931 001 XXX XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

0931 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 355 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

0931 041 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

0931 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0931 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0931 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

0931 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0931 046 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

0931 047 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.


356 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

0931 048 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

0931 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

0932 001 001 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Main version is not valid (MAIN_VERSION_ERROR).

Technical description of conditions for occurring:


- The main system version is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly.

0932 001 002 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The operative system version is not valid
(OS_VERSION_ERROR).

Technical description of conditions for occurring:


- The operative system version is not what it was expected to
be.

Some possible explanations why the error has occurred:


- The software is not installed correctly.

0932 001 003 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Arrow.exe version is not valid (UI_VERSION_ERROR).

Technical description of conditions for occurring:


- The Arrow.exe version is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 357 (381)
Program version 1.1.0
C EN 2015-03-26

0932 001 004 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Language Pack version is not valid
(LANG_VERSION_ERROR).

Technical description of conditions for occurring:


- The Language Pack version is not what it was expected to
be.

Some possible explanations why the error has occurred:


- The software is not installed correctly.

0932 001 005 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Arrow.Launcher.exe version is not valid
(UILAUNCHER_VERSION_ERROR).

Technical description of conditions for occurring:


- The Arrow.Launcher.exe version is not what it was expected
to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly.

0932 002 001 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Main CRC is not valid (MAIN_CRC_ERROR).

Technical description of conditions for occurring:


- The Main CRC is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly or the disk is
corrupted.

0932 002 002 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The operative system CRC is not valid (OS_CRC_ERROR).

Technical description of conditions for occurring:


- The operative system CRC is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly or the disk is
corrupted.

0932 002 003 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


358 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- The Arrow.exe CRC is not valid (UI_CRC_ERROR).

Technical description of conditions for occurring:


- The Arrow.exe CRC is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly or the disk is
corrupted.

0932 002 004 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Language Pack CRC is not valid (LANG_CRC_ERROR).

Technical description of conditions for occurring:


- The Language Pack CRC is not what it was expected to be.

Some possible explanations why the error has occurred:


- The software is not installed correctly or the disk is
corrupted.

0932 002 005 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Arrow.Launcher.exe CRC is not valid
(UILAUNCHER_CRC_ERROR).

Technical description of conditions for occurring:


- The Arrow.Launcher.exe CRC is not what it was expected to
be.

Some possible explanations why the error has occurred:


- The software is not installed correctly or the disk is
corrupted.

0932 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

0933 001 001 XXX, SOFTWARE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- The Arrow.exe application (UI) is launched from the SD
card. It must be launched from the eMMC disk.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 359 (381)
Program version 1.1.0
C EN 2015-03-26

(UI_LAUNCHED_FROM_SD_CARD).

Technical description of conditions for occurring:


- The Arrow.exe application (UI) is launched from the SD
card. It must be launched from the eMMC disk.

Some possible explanations why the error has occurred:


- The hardware jumper is still set on the UI main board that
makes the machine boot from the SD card. The jumper should
be removed before running the machine in normal operation.

0933 002 000 XXX, MAIN SWITCH SHUT DOWN


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The dialysis machine was turned off using the main switch
or halt button on the back of the machine. UI could not
make a clean shutdown. Logs and information in the error
list may be corrupted.

Technical description of conditions for occurring:


- The opreator has manually shut off the machine using the
main switch or halt button on the back of the machine.

0933 002 001 XXX, SOFTWARE ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- Saved error code has been lost.

Technical description of conditions for occurring:


- Failed to open SafetyLog.txt or CRC-check of the file
failed.

Some possible explanations why the error has occurred:


- The dialysis machine was turned off using the hard reset
button. UI could not make a clean shutdown.

0933 002 002 XXX, SOFTWARE ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- Saved language setting and touch screen calibration has
been lost.

Technical description of conditions for occurring:


- Corrupt registry.

Some possible explanations why the error has occurred:


- The dialysis machine was turned off using the hard reset
button. UI could not make a clean shutdown.

0933 002 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.


360 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0933 003 002 XXX, SOFTWARE ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- History Database was repaired.

Technical description of conditions for occurring:


- Repaired database file.

Some possible explanations why the error has occurred:


- The dialysis machine was turned off using the hard reset
button. UI could not make a clean shutdown.

0933 003 003 XXX, SOFTWARE ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- History Database was regenerated.

Technical description of conditions for occurring:


- Regenerated database file.

Some possible explanations why the error has occurred:


- The dialysis machine was turned off using the hard reset
button. UI could not make a clean shutdown.

0933 003 004 XXX, SOFTWARE ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- History Database was corrupt.

Technical description of conditions for occurring:


- Corrupt database file.

Some possible explanations why the error has occurred:


- The dialysis machine was turned off using the hard reset
button. UI could not make a clean shutdown.

0933 004 000 XXX, SOFTWARE ERROR


Unconditional error that shows an alarm but the fluid and
blood paths are not stopped.

General description of conditions for occurring:


- Alarm time stamps might be wrong.

Technical description of conditions for occurring:


- Missing real time clock update at startup.

0933 005 000 XXX, MEMORY ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 361 (381)
Program version 1.1.0
C EN 2015-03-26

- UI memory test failed.

Technical description of conditions for occurring:


- UI RAM memory broken.

0934 000 000 XXX, FREQUENCY SUPERVISION ERROR – MACHINE RESTARTED


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- Supervision of oscillators failed to start.

0934 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_RX_P.

0934 001 001 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_RX_C.

0934 001 002 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.


362 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_TX.

0934 001 003 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_MASTER.

0934 001 004 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_SLAVE.

0934 001 005 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: UI.

0934 001 006 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 363 (381)
Program version 1.1.0
C EN 2015-03-26

treatment. If it has appeared more than 3 times in the last


30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Defect task: CANOPEN_RX_P.

0934 001 XXX XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0934 002 000 XXX, PROCESSOR FREQUENCY ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

0934 041 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

0934 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for


364 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

treatment. If it has appeared more than 3 times in the last


30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0934 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0934 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

0934 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0934 046 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

0934 047 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

0934 048 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 365 (381)
Program version 1.1.0
C EN 2015-03-26

30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

0934 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

0935 041 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 041 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 041 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Time out.

0935 041 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- Out of Memory.

0935 041 015 XXX, CANOPEN COMMUNICATION ERROR


366 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Conditional error that is silent and does not prevent


treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 041 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 1.

0935 042 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 042 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 042 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- Time out.

0935 042 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 367 (381)
Program version 1.1.0
C EN 2015-03-26

after request to SDO Server node 2.

Technical description of conditions for occurring:


- Out of Memory.

0935 042 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 042 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 2.

0935 043 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 043 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 043 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Time out.


368 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0935 043 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- Out of Memory.

0935 043 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 043 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 3.

0935 044 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 044 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 044 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 369 (381)
Program version 1.1.0
C EN 2015-03-26

- CANopen stack detected a CANopen client or server NACK


after request to SDO Server node 4.

Technical description of conditions for occurring:


- Time out.

0935 044 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- Out of Memory.

0935 044 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 044 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 4.

0935 045 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 045 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Toggle bit not altered.


370 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

0935 045 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Time out.

0935 045 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- Out of Memory.

0935 045 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 045 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 5.

0935 046 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 046 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 371 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 046 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Time out.

0935 046 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- Out of Memory.

0935 046 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 046 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 6.

0935 047 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


372 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- Toggle bit not altered.

0935 047 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 047 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Time out.

0935 047 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- Out of Memory.

0935 047 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 047 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node7.

0935 048 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 373 (381)
Program version 1.1.0
C EN 2015-03-26

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 048 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 048 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Time out.

0935 048 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- Out of Memory.

0935 048 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node 8.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.

0935 048 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


374 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- CANopen stack detected a CANopen client or server NACK


after request to SDO Server node 8.

0935 050 000 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 050 001 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Toggle bit not altered.

0935 050 002 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Time out.

0935 050 004 XXX, CANOPEN COMMUNICATION ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- Out of Memory.

0935 050 015 XXX, CANOPEN COMMUNICATION ERROR


Conditional error that is silent and does not prevent
treatment. If the same error has appeared 3 times or more
during the last 30 days it is treated as an unconditional
error, fluid and blood paths are stopped and the blood pump
can not be run.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

Technical description of conditions for occurring:


- General error or error in transfer between application and
CANopen.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 375 (381)
Program version 1.1.0
C EN 2015-03-26

0935 050 XXX XXX, CANOPEN COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- CANopen stack detected a CANopen client or server NACK
after request to SDO Server node A.

1020 007 001 XXX, CPU OVERLOAD ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CPU overloaded.

Technical description of conditions for occurring:


- CPU load above 95%.

Some possible explanations why the error has occurred:


- Internal SW Error.

1021 004 001 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Maximum CMX messages used.

Technical description of conditions for occurring:


- If tasks receiving CANOpen messages does not handle
messages fast enough. Maximum CMX messages may be reached
and incoming messages are lost. This can occure if CAN
packets are sent to fast to the Node.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

1021 005 000 XXX, CANOPEN RX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CANOpen rx fifo is overrun.

Technical description of conditions for occurring:


- CANOpen rx fifo is overrun beacuse of to small fifo size
combined with fast CAN packets.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

1021 005 001 XXX, CANOPEN TX ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- No available CAN controller mailbox.


376 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- The mailboxes of the MCU are full, the CAN controller has
not been able to send CAN messages on the CAN bus,

Some possible explanations why the error has occurred:


- CAN bus interferance.
- SW driver error.

1021 006 000 XXX, CMX QUEUE FULL ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- CMX queue is full.

Technical description of conditions for occurring:


- CANOpen messages are received to fast than the node can
handle. If this happends then queue might get full.
Incoming CAN messages are lost.

Some possible explanations why the error has occurred:


- CAN packets are sent to fast for this node to handle. FIFO,
Queue sizes may be to small.

1027 008 000 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z1V1PMIC1_ERROR.

1027 008 001 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z1V1PMIC2_ERROR.

1027 008 002 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z1V5PMIC_ERROR.

1027 008 003 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 377 (381)
Program version 1.1.0
C EN 2015-03-26

Technical description of conditions for occurring:


- Z1V8PMIC1_ERROR.

1027 008 004 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z1V8PMIC2_ERROR.

1027 008 005 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z3V3PMIC1_ERROR.

1027 008 006 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z3V3PMIC2_ERROR.

1027 008 007 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z5V0SYS_ERROR.

1027 008 008 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- ZVTTPMIC_ERROR.

1027 008 009 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


378 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- DDR_VREF_ERROR.

1027 008 010 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z24V_FEED_ERROR.

1027 008 011 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z3V3PROT_ERROR.

1027 008 012 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z12VDISPLAY_ERROR.

1027 008 013 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z3V3_FEED_ERROR.

1027 008 014 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z5V_FEED_ERROR.

1027 008 016 XXX, VOLTAGE ERROR


Unconditional error, fluid and blood paths are stopped and
the blood pump can not be run.

General description of conditions for occurring:


- Voltage out of range.

Technical description of conditions for occurring:


- Z1V24VREF_ERROR.


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 379 (381)
Program version 1.1.0
C EN 2015-03-26

1031 001 000 XXX, SOFTWARE TASK ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect task was detected and heartbeat was stopped. This
error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

1031 002 000 XXX, PROCESSOR FREQUENCY ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.
This error code is only given at startup as additional
information together with another error and is by that
reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

1031 003 XXX XXX, COMMUNICATION ERROR


Unconditional error that is only shown in the error list.
Machine can still be used for treatment.

General description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped. This error code is only given at startup as
additional information together with another error and is
by that reason silent.

Technical description of conditions for occurring:


- Unspecified condition.

1032 100 000 XXX, DEVELOPMENT SOFTWARE


Notification to the operator, not an error.

General description of conditions for occurring:


- The software of the node is a development version and is
not for clinical use. This error will not be presented as a
technical error on the machine, instead it will trigger an
attention to be shown.

Technical description of conditions for occurring:


- Machine has been flashed with software not intended for
clinical use.

1034 001 000 XXX, SOFTWARE TASK ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


380 (381) AK 98 Dialysis Machine - Error Codes HCEN12629 Rev. 02.2015
Program version 1.1.0
C EN 2015-03-26

- The given error has resulted in a restart of the machine.


This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect task was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Watchdog kicked in since task with lowest priority got
starved out by a task with higher priority.

1034 002 000 XXX, PROCESSOR FREQUENCY ERROR – MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A defect oscillator was detected and heartbeat was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

1034 003 XXX XXX, COMMUNICATION ERROR - MACHINE RESTARTED


Error caused the machine to restart. The operator must
confirm the alarm before the machine can be used for
treatment. If it has appeared more than 3 times in the last
30 days the machine will not allow a treatment.

General description of conditions for occurring:


- The given error has resulted in a restart of the machine.
This error messages indicates why the restart was
initiated.

Technical description of conditions for occurring:


- A CANopen communication error was detected and heartbeat
was stopped.

Some possible explanations why the error has occurred:


- Unspecified condition.

Version: TechnicalErrors 2.2


HCEN12629 Rev. 02.2015 AK 98 Dialysis Machine - Error Codes 381 (381)
Program version 1.1.0