You are on page 1of 4

Drawings RC400G-1

Residual Chlorine Analyzer


(for pure water)

unit : mm
Standard solution inlet Air purge inlet
Rc1/4(female) Rc1/4(female) Maintenance Maintenance
space space
Reagent inlet
4-ø15 holes Approx. 200 530 Approx. 200
Rc1/4 (female)
for anchor bolt
100

Stand

600
Ball valve (V3)
350

Approx. 500
Strainer (S1) Maintenance
Front
space
100

20 490 20

Cross section A-A

Air pump (*3)


Converter (AP)
(CON)

Detector
(CELL)

Power terminal
Head tank
box (*1) (HD)

Metering
Meas. / Std. solution pump Drain
1450

switching valve VP40 pipe


(PU)
(V2)
Activated Ball valve Pressure reducing
charcoal filter (V11,12) valve (*2) (PR)
(*2) Ball valve
(F11) (*2) (V1) Ball valve (*2)
3-way solenoid (V4)
A A valve (SV6) Measuring
sample inlet Tap water inlet
(*2)
VP16 pipe VP16 pipe
(*2)
120
93
80

530 550 50 85 60 40
150 150

(*1) Option / ARS applies to model with arrestors


(*2) Option / AZC applies to model with auto zero calibration
(*3) Option / AP applies to model with air purge pump

F2_1E.ai

Weight: Approx. 65 kg

Unless otherwise specified,differences in the dimensions are specified as: General tolerance = ± (Criteria of tolerance class IT18 in JIS B0401) / 2.

All Rights Reserved, Copyright © 2009, Yokogawa Electric Corporation.


Subject to change without notice.
SD 12F04A01-01E 1/4
1st Edition: Dec. 1991 (YK)
3rd Edition : Apr. 2011 (YK)

P-
Wiring Diagram
l RC400G-1/ARS/AZC

Converter Dedicated cable Detector


(CON) K9332WA (CELL)
Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output Gn
M2 T2 T2
(Wired by customer)
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2

Power terminal box (*2) Pump


Dedicated cable (PU)
- Analog output - K9332WB
+ P1 P1
+ Bk
P2 P2
S W
G G G
AR2x2 -
+
G G
Solenoid valve
L2 L2 (option)
L1 L1 (SV6) (*1)
Dedicated cable
G K9460CN
V1 1
L2 Bk
L1 V2 2
W
G
CB1
G L2 Air pump
2 (option)
L1
(AP) (*3)
1 AR1 Dedicated cable
K9332WJ
A1 A1
Bk
* Case ground terminal must be A2 A2
grounded. W
* Ground

If this is not possible connect


to power source ground wire. (Note)
AR1,2 : Arrestors Dotted lines : external wiring F16.EPS

CB1 : Circuit breaker Use cable of 6 to 12 mm OD, however when connecting via power
Power Analog output terminal box option use cable of 9 to 11 mm OD for both power and
analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump

SD 12F04A01-01E 2/4
3rd Edition : Apr. 2011 (YK)

P-
Piping Diagram
l RC400G-1

Tubing materials
ø6 x ø4
polyethylene tube Converter

ø22 x ø15 Head tank CON


HD
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube

VP16 pipe
VP40 pipe CELL Measuring
tank
* option

PU Metering
pump

Ball valve
V2

COM

V1
Measuring sample (VP16)
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

Vat
*Air pump
Drain (VP40)
F1_4_1E.ai

SD 12F04A01-01E 3/4
3rd Edition : Apr. 2011 (YK)

P-
l RC400G-1/AZC with auto zero calibration

Tubing materials
ø6 x ø4
polyethylene tube Converter
ø22 x ø15 Head tank CON
HD
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube

VP16 pipe
V12
VP40 pipe CELL Measuring
tank
* option SV6 NO
3-way solenoid
valve
COM
NC Activated
charcoal
filter PU Metering
F11 pump
Ball
valve
Ball valve V11 V2
V1
Measuring sample (VP16) COM
Pressure reducing valve
V13
Calibration solution (VP16) PR
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

*Air pump Vat

Drain (VP40)
F1_4_4E.ai

Drain
VP40 pipe

Standard solution inlet


For air purge inlet
Reagent inlet
Measuring sample inlet
VP16 pipe Tap water inlet (*1)
VP16 pipe

(*1) Option /AZC applies to model with auto zero calibration


F3_1_1E.ai

SD 12F04A01-01E 4/4
3rd Edition : Apr. 2011 (YK)
P-