Professional Documents
Culture Documents
with a few measurements on the therapy authors here, and this paper is intended to
machine. The dose is separated into pni- present the essentials of the discussions at
mary and scatter contributions, as has been the workshop.
described by Clarkson,’16 with allowances Another method of dose calculation
for the variation of exposure across the utilizing the tissue maximum ratio (TMR)
field and non-nominal source-skin distance definition of the dose at depth has been
(SSD) in parts of the field. A computer developed at the Memorial Hospital in
method embodying the same principles has New York.’4 This method is designed to be
been in use at the Princess Margaret Hos- independent of energy and is in use at a
pital, Toronto.5 The method described here number of institutions; however, it will
can be used to calculate the dose at any not be presented in the discussions here.
point in the patient within or outside the
useful beam except near the edges of the CONSIDERATIONS IN CALCULATION
field. It may be considered a standard of
IHE I’ROII.EM
comparison for approxirn ation methods
without further experimental verification. In the treatment of Hodgkin’s disease,
These discussions of calculative methods, large fields with outside borders of up to
I 5upported in part b National Cancer Institute Grants CA 08971, Cooperative Hodgkin’s I)isease Clinical Trial, and CA ioc,
1’he Radiological Physics Center.
2The Radiological Physics Center, The University ofTexas at Houston, NI. I). Anderson Hospital and Tumor Institute, Houston,
lexas. R. Golden now at Youngstown Hospital Association, Youngstown, Ohio.
3 The Princess Margaret Hospital, Toronto, Ontario, Canada.
I Argonne Cancer Research Hospital, The University ofChicago, Chicago, Illinois.
I Saroni Tumor Institute, Mount Zion Hospital, San F rancisco, California.
S Michael Reese Hospital, Chicago, Illinois; now at Milwaukee County General Hospital, Milwaukee, Wisconsin.
7 Stanford University Medical School, Palo Alto, California; now at Prince ofWales Hospital, Randwick, Australia.
‘Memorial Hospital, New York, New York; now at Wessex Regional Department of Nuclear Medicine, Southampton General
Hospital, England.
‘The University of Texas at Houston, M. D. Anderson Hospital and Tumor Institute, Houston, Texas.
10 Walter Reed Hospital, Bethesda, Maryland; now at the School of Medicine, The University of North Carolina, Chapel Hill, North
Carolina.
11 Thomas Jefferson University Hospital, Philadelphia, Pennsylvania.
30
\OL. 117, No. I Calculation of Dose in Hodgkin’s Disease 3’
Downloaded from www.ajronline.org by 157.92.4.75 on 09/05/14 from IP address 157.92.4.75. Copyright ARRS. For personal use only; all rights reserved
0 5 0 IS 20
1111111111! 1111111111
cm.
l’Ic. 1. Typical mantle field for radiation treatnient of the neck and thorax. + = positions on surface of pa-
tient; 0 = the projection at mid-depth.
4o Xo cm. are blocked to shield regions radiations usually are used, including
where the radiation is not desired, such as cobalt 6o gamma rays and 2, 4,6, or 8 MV
the lungs. The resulting treatment field will roentgen rays, with SSD of 8o to o cm.
differ in shape from patient to patient; \Vith highei’ energy photons, there is
thus, it is not practical to have isodose danger of underdosing superficial regions
curves available for each field shape and in which electronic build-up has not been
patient contour. An example in Figure i achieved, and some clinicians question the
shows the outline of a field and a number use of 6 and 8 MV roentgen ra\-s for this
of points chosen for calculation. A cross- reason. This danger may be more apparent
sectional view of the same treatment plan than real, since with 8 MV roentgen rays
which includes the central axis is shown in the dose at 0.5 cm. depth is 90% of the
Figure 2. The departure from nominal maximum with a 30 X3o cm. field, and 8o%
SSD (g), relating to a point of calculation of maximum with a ioXio cm. field.’
(A), at depth (d), is illustrated. For treat- With 6 MV roentgen rays, the dose at 0.5
ments of this type, high energy photon cm. depth is 92% of the maximum with a
J. H. Cundiff et al. JANUARY, 1973
necessary to take account of beam charac- The variation of SSD within the field of
teristics, just as with smaller fields of treatment is a source of error in the cal-
regular shape. With the latter, the off-axis culation considered. If the nominal SSD is
diminution of the beam is implicit in the 130 cm. and a portion of the skin is 4 cm.
isodose curves; for the calculation of dose more distant, then an error of about 6% is
to the points in an irregular field for which introduced if the patient is considered flat
isodose curves do not exist, however, the at 130 cm.
diminution across the beam must be al-
INFLUENCE OF COLLIMATORS AND BLOCKING
lowed for explicitly. In Table i, some
typical beam profiles in air are shown for The setting of the primary collimators
large fields, indicating that points off the influences the exposure rate significantly,
TABLE I
BEAM PROFILE IN AIR, RELATIVE TO THE CENTRAL AXIS FOR LARGE FIELDS*
2 MV 6 MV
Co’#{176} Co#{176}#{176} Roentgen Rays Roentgen Rays
Distance from
Side of Field
ioo cm. SSD 130 cm. SSD
(cm.) 150 cm. TSD 160 cm. TSD
35X35 cm. Trim- 53X52 cm. Blocked
35X35 cm. 40X40 cm.
mers 6 cm. to 34X34 cm.
* Measurements were in air on particular machines. The results are not applicable generally to other similar machines.
ROL. 117, No. I Calculation of Dose in Hodgkin’s Disease 33
but the presence of secondary blocking, mary and scattered radiation with al-
such as for the lungs, may increase or de- lowance for the off-axis diminution and
crease the exposure rate slightly-usually non-nominal SSD. The primary radiation
less than I %. This occurs because of the to a point of interest is calculated using the
large contribution ofscatter radiation from tissue-air ratio (TAR) for zero field size,
Downloaded from www.ajronline.org by 157.92.4.75 on 09/05/14 from IP address 157.92.4.75. Copyright ARRS. For personal use only; all rights reserved
the collimating parts near the source and and the scatter radiation is calculated from
the relatively small contribution of scatter scatter-air ratios (SAR). The TAR is de-
from secondary collimators at a distance fined as the ratio of the absorbed dose at a
from the source. Therefore, it is sufficient given point in a phantom to the absorbed
that the exposure rate in air or absorbed dose which would be measured at the same
dose rate in free space’#{176}from a machine be point in free air within a volume of the
known or calculated with the primary phantom material just large enough to
collimators as they are set in treatment provide the maximum electronic buildup at
prior to the addition of secondary blocking. the point of measurement.1’ The SAR is
defined as the difference between the TAR
for a given field and the TAR at the same
GENERAL METHOD
depth for zero field size.’3 The data for
The general method consists of a sum- these calculations are given in Table ii.
mation at each point of interest of the pri- The absorbed dose at an arbitrary point
TABLE hA
TISSUE-AIR RATIOS AND SCATTER-AIR RATIOS
11. Cobalt 60
(cm.) 0 cm.2 100 cm.2 2 cm. cm. 6 cm. 8 cm. io cm. 15 cm. 20 cm. 25 cm.
0.5 1.000 1.036 .014 .026 .037 .048 .o8 .076 .085 .089
I . 965 I . 028 . 025 . 048 . o66 . 078 . 089 . 107 . I I8 . i 23
2 .905 1.002 .045 .o8o .102 .ii6 .127 .149 .160 .167
3 .845 .97I .061 .103 .130 .147 .i6i .184 .198 .205
4 .792 .938 .071 .121 .151 .170 .i86 .210 .228 .239
S .742 .902 .076 .134 .i66 .189 .206 .235 .255 .266
6 .694 .864 .o8o .14I .176 .201 .219 .252 .272 .284
7 .6o .825 .o8i .143 .181 .209 .229 .267 .290 .304
8 .6o8 .785 .o8o .142 .i8 .214 .236 .278 .301 .315
9 .570 .746 .078 .140 .183 .216 .240 .284 .312 .327
10 .534 .707 .075 .136 .i8i .215 .242 .288 .318 .336
II .501 .671 .071 .132 .178 .213 .241 .289 .322 .339
12 .469 .635 .069 .128 .174 .210 .239 .290 .324 .342
13 .440 .602 .o66 .124 .170 .207 .237 .290 .325 .345
‘4 .412 .570 .063 .120 .i68 .204 .235 .288 .326 .347
Is .386 .539 .060 .ii6 .162 .200 .231 .286 .325 .347
i6 .361 .511 .o8 .112 .157 .196 .227 .283 .322 .346
‘7 .338 .484 .o6 .io8 .153 .191 .223 .279 .318 .343
i8 .317 .459 .054 .104 .148 .i86 .218 .275 .313 .339
‘9 .297 .433 .052 .101 .144 .181 .213 .270 .309 .335
20 .278 .409 .049 .097 .139 .176 .207 .265 .305 .329
TABLE JIB
TISSUE-AIR RATIOS AND SCATTER-AIR RATIOS
(cm.) 0 cm.2 100 cm.2 2 cm. 4 cm. 6 cm. 8 cm. 10 cm. 15 cm. 20 cm.* 25 cm.*
J2 1.000 1.037 .013 .026 .039 .052 .062 .073 .077 .079
2.0 .952 1.021 .035 .055 .073 .o88 .098 .110 .115 .ii8
3.0 .899 .998 .048 .081 .102 .ii8 .130 .143 .150 .154
4.0 .847 .966 .o58 .097 .125 .145 .157 .174 .i88 .195
5.0 .799 .933 .063 .110 .141 .i6 .179 .200 .214 .222
6.o .753 .900 .067 .119 .155 .181 .196 .220 .237 .24#{244}
7.0 .7I0 .866 .o68 .125 .163 .193 .2II .237 .257 .268
8.o .670 .832 .069 .128 .170 .201 .222 .252 .272 .28
9.0 .632 .800 .069 .131 .174 .207 .230 .263 .283 .296
10.0 .598 .764 .o68 .132 .177 .210 .236 .271 .291 .30
11.0 .564 .7I .067 .II .178 .213 .240 .278 .300 .I5
12.0 .529 .698 .o66 .130 .178 .215 .242 .283 .308 .I8
13.0 .500 .666 .o6 .128 .177 .214 .243 .288 .325
14.0 .471 .636 .063 .125 .175 .212 .24 .289 .I6
15.0 .444 .6og .062 .122 .171 .210 .241 .290 .316
i6.o .419 .576 .o6o .118 .168 .206 .239 .289 .316 .325
17.0 .395 .549 .059 .114 .163 .202 .235 .287 .314 .323
18.0 .373 .42I .057 .110 .159 .197 .230 .284 .312 .320
19.0 .351 .496 .o6 .107 .154 .19’ .226 .28! .310 .318
20.0 .331 .47 .054 .104 .150 .188 .221 .276 .306 .3i
S Extrapolated data.
Notes: (I) Table by Mary Peterson ofRosewood General Hospital and Robert Golden ofthe AAPM Radiological Physics Center fronT
measurements at Wadley Institute, Dallas, Texas.
(21 This machine produced a depth dose of 6o.8 per cent at io cm. depth for a ioX so cm. field at 8o cm. target-skin distance
and a tissue-maximum rato ofo.738 at io cm. depth for a loX 10 cm. field at 8c cm. target-axis distance.
A, as shown in Figures i and , is calcu- Symbolically, with the brackets having the
lated as follows: same meaning as above, the absorbed dose
x [transition to absorbed dose in a small mass of tissue for treatment field sizej
TABLE IIC
TISSUE-AIR RATIOS AND SCATTER-AIR RATIOS
(cm.) 0 cm.2 100 cm.2 2 cm. cm. 6 cm. 8 cm. 10 cm. 15 cm. 20 cm. 25 cm.
1.5 1.000 I .029 .010 .021 .031 .042 .05I .o6 .069 .071
2 . 997 I .027 .011 .02I .032 .043 .052 .064 .069 .071
3 . 965 I .009 .016 .032 .047 .o8 .067 .o8; .089 .oco
4 .926 .981 .020 .040 .057 .071 .082 .099 .io8 .111
5 .88 . 952 . 024 .049 .071 .087 .100 .117 .126 .I0
7 . 802 . 895 . 035 .070 .og8 .120 .14 I .155 .ib6 .171
8 .76 I . 866 . 040 .0’79 .111 .I34 .149 .172 .184 .189
9 .722 . 837 . 043 .087 .121 .145 .162 .i88 .200 .206
I0 .687 . 8o8 . 046 .091 .127 .151 .170 .199 .212 .217
14 .567 . 694 . 047 .09.; .134 .i6 .187 .224 .241 .254
15 . 540 . 668 . 047 .094 .i;6 .i68 .191 .226 .247 .262
16 .513 . 642 . 048 .095 .137 .171 .19 .229 .252 .269
17 .488 . 618 . 048 .096 .138 .171 .194 .22 .256 .274
18 . 464 .593 . 048 .o#{231}6 . I37 . 170 . 194 . 234 . 259 . 278
19 . 443 . 570 .048 .094 .134 .167 .192 .22 .259 .278
20 .422 . 546 . 047 .091 .130 .164 .189 .230 .258 .278
DA [D1 X B][T’R) x G]
on isocentnic techniques.
Fixed
phantom.
source-skin distance; calibration in
. [ ( )I SAD’
SPD [TAR)d + Sd].
(2)
23.
l’ic. . The method of obtaining the average scatter-air ratio from equal sectors centering on point A.
Regions under heavy blocking are considered as contributing no scatter.
TABLE 111
CALCULATION OF AVERAGE SCATTER-AIR RATIO TO POiNT A IN FIGURES 2 AND 3
I + . 208 +.208
2 + .218 +.218
3 + . 306 +.238 - .26) +.275
4 + . 297 +.297
5 + .316 +.280 - .302 +.294
30 +132 +.132
31 +.138 +.138
32 + . 149 +.‘49
33 + . 164 +.164
34 +.,88 7.216 +.188
35 +.200 Average SAR = = .200 +.200
36 + . 202 36 +.202
area is that area which gives the same TAR UNCERTAINTIES IN THE GENERAL METHOD
as the general method. The equivalent
area is irr2, where r, the radius, is selected Measured/
Calculated
from Table ii to give the scatter-air ratio
Downloaded from www.ajronline.org by 157.92.4.75 on 09/05/14 from IP address 157.92.4.75. Copyright ARRS. For personal use only; all rights reserved
COMPARISON OF CALCULATIONS
ofthe field such as point A. The low density
AND MEASUREMENTS
of lung beneath the lung blocks does not
fully attenuate the scatter from distant
The general method of calculation is parts ofthe open field, as indicated in Item
proposed as a definitive method, not re- I of this table. Item 2 arises because of the
quining further experimental verification. assumption tables
in that
scatter-air all
Accordingly, 3 approaches have been taken parts of the field are at the same SSD; the
to challenge the calculative method: (i)
general method allows for the effect of non-
uncertainties in the general method; (2)
nominal SSD for the primary radiation but
measurements in an idealized phantom; not for the scattered radiation. Item 3 is an
and (,) measurements in an anatomic estimate of the effect of the limited thick-
phantom. ness of scattering media beneath or ad-
jacent to such the neck or axilla,
areas as
UNCERTAINTIES IN THE GENERAL METHOD
which fail to provide full scatter to the
A summary of types of known uncer- point of calculation. Some transmission of
tainties in the general calculative method, primary radiation occurs through the lung
together with a measurement or estimate blocks with subsequent
scatter to points
of each, appears in Table v. These entries of calculation. Since lung blocks usually
are appropriate to points in the open parts attenuate the primary beam to about 5%
transmission, it is estimated that scattered
TABLE IV radiation contni buted by non-zero primary
under shielding contributes on the order
EQUIVALENT RADII AND AREAS FOR THE POINTS NOTED
IN FIGURE I DERIVED BY THE GENERAL METHOD of I % of the total absorbed dose at a point
in the open field (Item ). These uncer-
. Equivalent Radius* Equivalent Area* tainties are estimates, or partial estimates,
I olnt
(cm.) (cm.2) and they are added to obtain an oven-all
uncertainty of about ±3% in the correct-
A 7.00 154
ness of the calculation. In this context,
B 9.19 265
C 8.43 223 “uncertainty” does not have a rigorous
D 5.76 104 definition, but it is likely that errors due
E 5.20 85 to the above causes would fall within the
F 8.00 20I
stated range 95% of the time; if this is so,
#{149}
These values are applicable only to the example presented in the standard error due to the above causes
Figure 1. would be less than ±2%.
40 J. H. Cundiff et al. JANUARY, 1973
C 10.0 1.00
utilizing the field shape and points of mea-
1) io.o .97 surement indicated in Figure i . The dose at
F . 5.2 .98 mid-depth was compared with the dose
F o. 1.00
calculated by the general method of cal-
G 10.0 1.00
culation. The first column in Table V1B12
H io.o .97
is the ratio of the dose calculated to that
Average 0.99±0.02 measured, considering as muscle density
the pants of the phantom exposed to the
B. .lnatomic Phantom (Grant, 1970)
open radiation field; the second column
corrects the scatter calculations for in-
Dose CalculatedDose Measured
cursions of lung (at density 0.33) into the
Point Phantom Phantom exposed area. This correction was neces-
. i nhomogenei ties i nhomogenei ties sar’ in this experiment because the Alder-
not considered considered son phantom used had abnormally large
lungs resulting in incomplete shielding of
1.04± .02 1 .02± .02
13 1.02± .0! 1 .00± .01 the lung area with lead. One should not
C 1 .05 ± .03 I .03± .03 need to make this correction for a clinical
I) .96± .o .95±0 situation since the lung normally is com-
F 1.02± .01 1.02± .01
tray, the skin dose is 5% of the sub- lead shot, No. 7, to fill a styrofoam cavity
cutaneous dose at maximum for a field size 7.5 cm. deep. This thickness is equivalent
of I2X12 cm. and a skin-to-tray distance to more than cm. of solid lead. The
of 5 cm. ;18 however,
I for a 30 X3o cm. field, cavity is cut in the styrofoam at the tray
the skin dose is about 70% that at maxi- distance with a hot electrical wire sup-
Downloaded from www.ajronline.org by 157.92.4.75 on 09/05/14 from IP address 157.92.4.75. Copyright ARRS. For personal use only; all rights reserved
mum.27 The relative skin dose depends not ported at the target distance, as the other
only on field size and distance but also end of the taut wire is made to trace the
upon the material and type of shadow lung outline on the roentgenogram at the
tray. For any given field size on skin-to- film distance.7 In the latter method, the
tray distance, a solid plastic tray gives the edges of the blocking are parallel to the
greatest dose to the skin. The use of plastic rays, but since the styrofoam holding the
or metal trays in the form of an open mesh shot is opaque to light, the outline of the
may improve skin sparing somewhat; how- lung blocks is not seen upon the skin. A
even, in those instances where optimum method which utilizes lead shot, but per-
skin sparing is desired, a material of inter- mits visualization of the light field upon
mediate atomic number may be placed on the patient, has been described.’9”0 The
the side of the shadow tray nearest the shot is glued together in a thermoplastic
patient. For a 30X30 cm. field at i cm. material using polyethylene foam as a form
distance, the dose to the skin is reduced to cut with divergent sides in a manner sim-
about o% of that at the maximum by the ilar to that described above. Alternatively,
addition of copper, tin, or leaded glass on a low melting point alloy molded into a
the underside of the shadow tray.27 solid block may be used.6
a: 9
. 1.c
#{149} #{149}0
.8
F
Toward
Cranium
t I
T9
C-2 C- C- C-es C-7 T-1 T-3 T-5 T-5T-T-( ‘10 T’U
* 5 7 T-1 2 VERTEBRAL
BODY
I’!G. 5. The dose to the spinal cord from equally weighted anterior and posterior mantle treatment fields. The
dotted lines are normalized to 1.00 at the mid-thickness of the mid-mediastinum (such as point C in Fig.
I); the solid lines are normalized to 1.00 at the mid-thickness of the upper mediastinum (such as point B
in Fig. I). The circles are a continuation of the solid lines, measured with no cervical shielding in place.
Measurements were with 2 MV roentgen rays in an Alderson phantom at 150 cm. SSD.
from an upper mantle treatment delivering MID-DEPTH IN THE CENTER OF THE FIELD
point in an irregularly shaped radiation RAD. THERAPY & NUCLEAR MED., 1968, 102,
199-206.
field. This method is recommended either I I. GOLDEN, R. Dose to the Spinal Cord in Mantle
as a routine method or as a standard Treatments at Various institutions
Reviewed
method against which more rapid approxi- by the RPC. Included in Hodgkin’s
Workshop
Downloaded from www.ajronline.org by 157.92.4.75 on 09/05/14 from IP address 157.92.4.75. Copyright ARRS. For personal use only; all rights reserved
mate methods may be judged for correct- Notebook, September 17, 1970, as paper H.
This book is available upon request to the
ness.
Radiological Physics Center, M. D. Anderson
In addition, limited consideration is
Hospital & Tumor Institute, Houston, Texas.
given to the problems of secondary block- 12. GRANT, W. H. Unpublished data, 1970.
ing, matching adjacent fields, dose to 13. GUPTA, S. K., and CUNNINGHAM, J. R. Measure-
spinal cord, and dose to the gonads or ment of tissue-air ratios and scatter functions
for large field sizes, for cobalt-6o gamma radia-
fetus.
tion. Brit. 7. Radio!., 1966, 39, 7-I I.
Robert J. Shalek, Ph.D. 14. HOLT, J. G., LAUGHLIN, J. S., and MORONEY,
Radiological Physics Center
J. P. Extension of concept of tissue-air ratios
TAR to high energy x-ray beams. Radiology,
The University ofTexas M. D. Anderson Hospital
and Tumor Institute at Houston
1970, 96, 437446.
15. ICRU. International Commission of Radiologi-
6723 Bertner Avenue
Houston, Texas 77025
cal Units and Measurements, Report iod,
Handbook 87, 1963, U. S. National Bureau of
Standards.
REFERENCES
i6. JOHNS, H. E., and CUNNINGHAM, J. R. The
I. ANDERSON, R. E., D’ANGIO, G. J., and KAHN, Physics of Radiology. Third edition. Charles
F. M. Dosimetry of irregularly shaped radia- C Thomas, Publisher, Springfield, Ill., 1969.
tion therapy fields. Parts 1 and II. Radiology, 17. JOHNSON, R. E. Private communication, 1970.
1969, 92, 1092-I 100. 18. LANZL, L. H. Unpublished data, 1969.
2. A review ofSuppl. No. 10. Depth dose tables for 19. MARUYAMA, Y., MOORE, V. C., BURNS, D., and
use in radiotherapy. Brit. 7. Radiol., 1968, i, HILGER, M. T. Individualized lung shields
932936. constructed from lead shot embedded in plas-
3. CLARKSON, J. R. Note on depth doses in fields of tic. Radiology, 1969, 92, 634-635.
irregular shape. Brit. 7. Radiol., I 941 , 14, 20. MARUYAMA, Y., WREDE, D., VAN ARSDALE, E.,
265-268. SAYEG, J., and ENGLES, E. P. Comments on
4. COVINGTON, E., and BAKER, A. S. Dosimetry of shielding by lead shot method. Radiology,
scattered radiation to fetus. 7A.M.il., 1969, 1972, 102, 445.
209, 414415. 21. MEURK, M. L., GREEN, J. P., NUSSBAUM, H.,
5. CUNNINGHAM, J. R. Calculation of dose due to and VAETH, J. M. Phantom dosimetry study
irregular fields by computer. To be published. of shaped cobalt-6o fields in treatment of
6. DEMIDECKI, A. J., FELDMAN, A., BRADFIELD, Hodgkin’s disease. Radiology, 1968, 9!, 554-
J. S., KINZIE, J. J., and POWERS, W. E. Evolu- 558.
tion ofsystem for field-shaping in radiotherapy 22. MEURK, M. L. Unpublished data, 1969.
utilizing low-melting alloys. Phys. Med. Biol., 23. MEURK, M. L. Unpublished data, 1970.