Professional Documents
Culture Documents
Assistant Professor
American International University-Bangladesh
Spring 2015-16
Introduction
What is Measurement?
• The measurement of a given quantity is essentially an act or the result of comparison between the quantity
(whose magnitude is unknown) and a predefined standard.
• Since two quantities are compared, the result is expressed in numerical values.
• In the measurement process the property of an object under consideration is compared with some standard.
• The numerical quantity must have a definite unit for meaningful presentation.
Acceleration Density
Velocity Viscosity
Displacement Composition
Force-Weight pH
Pressure Humidity
Torque Temperature
Volume Heat/Light Flux
Mass Flux Density
Flow rate Current
Level Voltage
Power
Introduction
The Basic Requirements for Meaningful Measurement:
• The standard used for comparison purposes are accurately defined and should be commonly accepted.
• The apparatus and the method adopted must be provable.
Recent Developments:
Introduction
Methods of Measurements
a) Direct Method and b) Indirect Method
Direct Method: In the direct method the unknown quantity is directly compared against standard.
The result is a numerical number with a unit.
Example: Measurement of length and mass etc.
A float-type petrol tank level indicator as sketched in Figure. Here, the change in
petrol level moves a potentiometer arm, and the output signal consists of a
proportion of the external voltage source applied across the two ends of the
potentiometer. The energy in the output signal comes from the external power
source. The indirect method is instrument dependent and accurate.
Mechanical Instruments:
• It shows the reading corresponding to mechanical variables
• These instruments are slow and have inertia effect
Electrical Instruments:
• Faster but normally have mechanical deflection system and requires maximum 0.5 sec to operate
Electronic Instruments:
• These days most of the scientific and industrial measurement systems are electronic.
Example: Cathode Ray Oscilloscope (CRO) and Biomedical Instruments etc.
Classification of Instruments:
1) Absolute Instrument:
• These instruments give the value of the electrical quantity in terms of absolute quantities (or some
constants) of the instruments and their deflections.
• In this type of instruments no calibration or comparison with other instruments is necessary.
• They are generally not used in laboratories and are seldom used in practice by electricians and engineers.
• They are mostly used as means of standard measurements and are maintained by national laboratories
and similar institutions.
• Some of the examples of absolute instruments are:
Tangent galvanometer, Raleigh current balance, Absolute electrometer
2) Secondary Instrument:
• They are direct reading instruments. The quantity to be measured by these instruments can be determined
from the deflection of the instruments.
• They are often calibrated by comparing them with either some absolute instruments or with those which
have already been calibrated.
• The deflections obtained with secondary instruments will be meaningless until it is not calibrated.
• These instruments are used in general for all laboratory purposes.
• Some of the very widely used secondary instruments are: ammeters, voltmeter, wattmeter, energy meter
(watt-hour meter), ampere-hour meters etc.
• The diagram shown above has two permanent magnets which are called the stationary part of the instrument and
the moving part which is between the two permanent magnets that consists of pointer.
• The deflection of the moving coil is directly proportion to the current.
• Thus the torque is proportional to the current which is given by the expression Td ∝ G.I or Td = GI
where Td is the deflecting torque. G is proportionality constant which depends upon the strength of the
magnetic field and the number of turns in the coil.
The pointer deflects between the two opposite forces produced by the spring and the magnets. And the resulting
direction of the pointer is in the direction of the resultant force. The value of current is measured by the deflection
angle θ, and the value of K and G.
• In opposite to deflection type of instruments, the null or zero type electrical measuring instruments tend to
maintain the position of pointer stationary.
• They maintain the position of the pointer stationary by producing opposing effect.
Thus for the operation of null type instruments following steps are required:
(1) Value of opposite effect should be known in order to calculate the value of unknown quantity.
(2) Detector shows accurately the balance and the unbalance condition accurately.
The detector should also have the means for restoring force.
• The null detector is a current galvanometer whose deflection is proportional to the unbalance EMF i.e. the
difference between “a-b” of slide wire and the unknown EMF Ex.
• As soon as the two are equal there is no current through galvanometer and it shows ZERO deflection thereby
indicating null condition.
• From analog signals the values are sampled and discretized as indicated .
• An ADC with sample and hold is required to get the digital output from analog signal as shown.
• In the reverse process a DAC is required to convert digital signal to analog.
• Present day digital instrument are becoming popular and many industrial instrumentation PLC or digital
computer is used for instrumentation.
Functions of Instruments
a) Indicating Function: contains s pointer and scale.
b) Recording Function: for dynamic conditions the variables with time variations are recorded.
c) Controlling Function: for industrial process control.
Description:
1) Primary sensing element:
• It contains a transducer or sensor.
• Transducer converts energy from one form to another form. Normally, a physical quantity to an electrical
quantity.
• The physical quantity is first sensed and then detected by an element and finally an output of
different analogous form is developed.
2) Variable conversion element:
• The primary sensing element produces output that may be voltage, frequency or other quantity that
need to be converted into useful form by the variable conversion element. Such as A/D converter.
3) Variable manipulation element: It changes the level of the signal amplifies or signal conditioning.
4) Data transmission element: After the variable is manipulated it is transmitted to the data presentation element
through wire or wireless.
5) Data presentation element: The variable under measurement is qualified and to be conveyed to the
personnel for observation
Interfering Input: These are quantities or variables to which the instrument is highly sensitive. The instrument or
system is not desired to respond to interfering inputs. But the output fluctuates due to interfering input because of
their working principle, design and other factors.
Modifying Input: They are defined as inputs which cause change in I/O relationships for either desired input or
interfering inputs or both. These are closure to interfering input.
Let’s consider a Manometer as an example to explain the above terms.
• U-tube manometer is used to measure the differential pressure of fluid or gas (but not atmospheric
pressure) using manometric filed.
Response with respect to Desired Input:
The differential pressure,
P = P1 – P2 = g. h. (ρm – ρf)
Where,
ρm , ρr are the density of Mercury and Fluid respectively.
g is the gravitational force, h is the height difference
If unknown input is gas then, ρm >> ρg. So the equation becomes,
P = P1 – P2 = g. h. ρm
Hence, P ∝ ℎ
Here, P1 and P2 are desired input
“ Static characteristics are normally obtained from static calibration. It affords the opportunity to check
the instrument against a known standard and subsequently to find out accuracy and error”
Apart from the above listed characteristics, there are few others static characteristics.
True Value:
• True value refers to a value that should be obtained if the quantity under consideration is measured by an
“exemplar method”, that is a method agreed upon by experts as being sufficiently accurate for the purpose to
which the data will ultimately be put to use.
• It is not possible to determine the true of quantity by experiment means.
• True value may be defined as the average value of an infinite number of measured values when average
deviation due to various contributing factor will approach to zero.
Measured Value:
• Measured value is defined as the approximated value of true value.
• It can be found out by taking means of several measured reading during an experiment, by applying suitable
approximations on physical conditions.
Static Error:
• Static Error is defined as the difference of the measured value and the true value of the quantity.
• Mathematically it can be expressed as follows.
Static Error (or Absolute Error), = − =
= Measured value = True value
Problem 2: A voltage has a true value of 1.50V. An analog indicating instrument with a scale range of 0-2.5V
shows a voltage of 1.46V. What are the absolute error and correction? Express the error as a function of the
true value and full scale deflection.
Answer:
Absolute Error, δA = Am− At = 1.46 − 1.50 = − 0.04 V.
Remember, Absolute Correction, δc = − δA
Absolute Correction, δc = − δA = −(− 0.04 V) = 0.04 V.
δA −0.04
εr = (absolute error/true value) = X100% = X100% = − 2.67%
At 1.50
−0.04
Relative error as a percentage of full scale deflection = X100% = − 1.6%
2.50
Error Calibration :
• Error calibration means that an instrument has been calibrated against a suitable standard and its static error is
determined at number of points on its scale.
• Calibration process involves a comparison of the particular instrument with either (a) a primary instrument (b) a
secondary standard with a higher accuracy than the instrument to be calibrated or (c) an instrument of known
accuracy.
Reproducibility :
• It is a degree of closeness with which a given value may be repeatedly measured.
• An instrument with perfect reproducibility has no drift.
Precision:
• Precision indicates the repeatability or reproducibility of an instrument
• For a true value of 100V a voltmeter reads 104,103,105,103 and 105V. The instrument cannot be depended
on more than ±5% accuracy but the precision in only 1%., i.e. deviation of 1 V from the average value.
Linearity:
• The linearity is defined as the maximum deviation from the linear characteristics as a percentage of the full scale
output.
• It also measures how constant is the ratio between input and output. It is usually expressed in %.
Accurate but
Non-linear
Linear but not accurate
This is better than Non-linear. Because Does not provide same voltage variation with same range of speed in case of
this can be compensated speed measurement
Measurement And Instrument, Spring 2015-16 Ahmed Mortuza Saleque -24-
Hysteresis:
• Hysteresis effects show up different output values for the same input during loading and unloading.
Threshold:
• It is clear that in some instrument if the input is increased gradually from zero there will be some value below
which no output change is detected, this minimum value defines the threshold of the instrument.
Sensitivity:
• It measures the ratio between the change in the output and the corresponding change in the input (i.e. gain).
Resolution:
• It measures the minimum change in the input that is reflected in the output.
Dead Time :
• Dead time is defined as the time required by a measurement system to begin to respond to a change in
measurand.
Dead Zone :
• It is defined as the maximum input variation that is not reflected
in the output.
Drift:
• It measures the variation in the output that is not caused by a variation in the input.
• Typically Drift is caused by temperature. Sometimes by humidity, moisture i.e. it is caused by environmental
factors.
• Drifts in measurement and instrument systems are classifier into three categories.
(i) Zero Drift (ii) Zonal Drift or Sensitivity Drift (iii) Span Drift or Zero Drift plus Sensitivity Drift
Zero Drift:
• It describes the effect where the zero reading of an instrument is modified by a change in ambient condition.
• This causes a constant error that exists over the full range of measurement of the instrument.
Example: A voltmeter has sensitivity of 2000 Ω/V is used to measure voltage of a circuit having an output
resistance of 1kΩ and an open circuit voltage of 6V at its 10V scale. Find the error in measurement.
Errors in Measurement:
Accuracy of an instrument depends on
(a) Material (b) Design (c) Workmanship
R1 = 37 Ω ± (5× )= 37 ± 1.85 Ω
R2 = 75 Ω ± (5× )= 75 ± 3.75 Ω
R3 = 50 Ω ± (5× )= 50 ± 2.50 Ω
Problem: The resistance of a circuit is found by measuring current flowing and power loss in a circuit. Find the
limiting errors in the measurement of resistance when the limiting errors in the measurement of power and current
are respectively ± 1.5% and ± 1%.
Solution:
We know, P = I2R
0%
=> R = P
=> lnR = lnP − 2 lnI
&2 &3
=> = −2
1 2 &1 3 &1
1
=> = ± 1.50 + 2 × 1.0 = ±3.50%
1
Problem: The solution for the unknown resistance of a Wheatstone bridge is, RX = (R2*R3)/R1. Where, R1 = 100±
0.5%Ω, R2 = 1000± 0.5%Ω and R3 = 842±0.5%Ω. Determine the magnitude of the unknown resistance and limiting
error in percent and in ohm for the unknown resistance RX.
Solution:
RX = (R2*R3)/R1 = (1000*842)/100 = 8420Ω
Relative limiting error of unknown resistance
1' 1% 1 1
= + − = ± 0.5 + 0.5 − 0.5 = ±0.5%
1' 1% 1 1
Types of Error:
Systematic Error:
(i) Instrumental Error:
• These errors may be due to wrong construction, calibration of the measuring instruments.
• These types of error may be arises due to friction or may be due to hysteresis.
• In order to minimize the instrumental errors in measurement various correction factors must be
applied and in extreme condition instrument must be re-calibrated carefully.
(ii) Environmental Error:
• This type of error arises due to instrument’s external conditions.
• External condition includes temperature, pressure, humidity or magnetic filed.
(iii) Observational Error:
• As the name suggests these types of error are due to wrong observations.
• The wrong observations may be due to parallax.
Systematic Error affects ACCURACY
Random Error:
• After calculating all systematic errors, it is found that there are still some errors in measurement are left.
• These errors are known as random errors.
• Some of the reasons of the appearance of these errors are known but still some reasons are unknown.
• Hence we cannot fully eliminate these kinds of error
Dynamic Characteristics:
The dynamic characteristics are
(a) Speed of Response (b) Measurement lag (c) Fidelity (d) Dynamic Error
a) Speed of response: It is defined as the rapidity with which a measurement system responds to changes in the
measurement quantity.
b) Measurement lag: It is the retardation or delay in the response of a measurement system.
c) Fidelity: It is defined as the degree to which a measurement system indicates changes in the measured quantity
without dynamic error.
d) Dynamic error: It is the differences between the true value of the quantity under measurement changing with
time and the value indicated by the measurement system with no static error.
Dynamic analysis:
a) Time domain analysis & b) Frequency domain analysis.
• For analysis linear time invariant systems are considered.
• For non-linear systems linearization is required.
Lecture on
Measurement of Resistances
Prepared By
Mehedi Azad Shawon
Assistant Professor, Dept. of EEE
American International University-Bangladesh (AIUB)
1
Lecture Outcomes
By the end of this lecture, the students should be able to:
2
Resistance Measurement
What are the major significance of resistance?
Since resistance are used in electrical circuits for variety of useful tasks, the
property of resistances play a significant role for determining the
performance parameter of various circuit elements. Thus it is highly
important to have a reasonable accurate information about the magnitude
of resistance in any electrical circuit.
How did you need to measure resistance with high level of accuracy?
But for accurate values specially for resistances with very low and very high
values, we need specific methods. In this lecture we will discuss various
methods of resistance measurement.
3
Classification of Resistances
4
Choice of Methods
The choice of a suitable method of measuring resistance depends on several different
factors.
The range of resistance to be measured is the most important factor.
In any particular case the choice of method will be limited to one or two,
depending on the equipment available.
The different methods of resistance measurement can be divided into two classes,
1. laboratory methods &
2. workshop methods
This division again depends on the accuracy to which the measurement is to be made.
5
Classification of Resistance Measurement Methods
Where as figure 3(b) is suitable for measuring resistance of low values within the
ranges of medium value resistance. Both techniques require one voltmeter and
ammeter each. The readings from these are used to calculate the unknown
resistance using Ohm`s law R=V/I.
Where,
Rm = measured value,
Ra = ammeter internal resistance and,
R = true value of the resistance
8
Measurement of Low Resistance
Ammeter-Voltmeter (AV) Method-2
Now in figure 3, the ammeter measures the sum of current through voltmeter and
resistance, hence the measured resistances will be
Where,
Rm = measured value,
Rv = Voltmeter internal resistance and,
R = true value of the resistance
9
Example 1: Measurement of low resistance: AV method
A voltmeter of 600 Ω resistance and ammeter of 0.8 Ω resistance are used to measure two
unknown resistances by AV method. If the voltmeter reads 40 V and ammeter reads 120 mA
in both the cases, calculate the percentage error in the
values of measured resistances if
(a) The voltmeter is put across the resistance and the ammeter connected in series with
the supply
(b) The voltmeter is connected in the supply side and ammeter connected directly in
series with the resistance.
10
Example 1: Measurement of low resistance: AV method
(Answered)
(Answered)
11
Measurement of Medium Resistance (1Ω < R < 100kΩ)
Wheatstone Bridge Method
The Wheatstone bridge is the most commonly used circuit for measurement of
medium range resistances. The Wheatstone bridge consists of four resistance arms,
together with a battery (voltage source) and a galvanometer (null detector)
Where,
P & Q is called ratio arm,
S = Standard resistance (can be variable type),
R = Unknown resistance
12
Measurement of Medium Resistance
Error in Wheatstone Bridge Method
4. Errors may creep into measurement due to resistances of leads and contacts.
This effect is however, negligible unless the unknown resistance is of very low
value.
13
Measurement of Medium Resistance
The upper limit is set by the reduction in sensitivity to unbalance caused by high
resistance values.
The lower limit for measurement is set by the resistance of the connecting leads
and by contact resistance at the binding posts
14
Measurement of Low Resistance
Kelvin Bridge Method
The Kelvin bridge or Thompson bridge is used for measuring the unknown resistances
having a value less than 1Ω. It is the modified form of the Wheatstone Bridge.
When the galvanometer is connected to point ‘m’, the lead resistance r is added to the standard
resistance S. Thereby the very low indication obtains for unknown resistance R. And if the
galvanometer is connected to point n then the r adds to the R, and hence the high value of unknown
resistance is obtained. Thus, at point n and m either very high or very low value of unknown resistance
is obtained. 15
Measurement of Low Resistance
Kelvin Bridge Method (cont…)
So, instead of connecting the galvanometer from point, m and n we chose any
intermediate point say d where the resistance of lead r is divided into two equal parts,
i.e., r1 and r2
As
The above equation shows that if the galvanometer connects at
point d then the resistance of lead will not affect their results.
16
Measurement of Medium Resistance
17
Measurement of Low Resistance
Kelvin Double Bridge Method
The ratio of the arms p and q are used to connect the galvanometer at the right place
between the point j and k. The j and k reduce the effect of connecting lead. The P and
Q is the first ratio of the arm and p and q is the second arm ratio.
18
Measurement of Low Resistance
Kelvin Double Bridge Method
The ratio of p/q is made equal to the P/Q.
Under balance condition zero current
flows through the galvanometer. The
potential difference between the point a
and b is equivalent to the voltage drop
between the points Eamd.
19
Measurement of Low Resistance
The above equation is the working equations of the Kelvins bridge. The equation
shows that the result obtains from the Kelvin double bridge is free from the impact
of the connecting lead resistance.
For obtaining the appropriate result, it is very essentials that the ratio of their arms
is equal. The unequal arm ratio causes the error in the result. Also, the value of
resistance r should be kept minimum for obtaining the exact result.
The thermo-electric EMF induces in the bridge during the reading. This effect can be
reduced by measuring the resistance with the reverse battery connection. The real
value of the resistance obtains by takings the means of the two.
20
Measurement of High Resistance (>100 kΩ)
21
Measurement of High Resistance (>100 kΩ)
23
Measurement of High Resistance (>100 kΩ)
24
Measurement of High Resistance (>100 kΩ)
Loop Test:
There are several methods for locating the faults in underground cables. However, two
popular methods known as loop tests are :
(i) Murray loop test
(ii) Varley loop test
These simple tests can be used to locate the earth fault or short-circuit fault in
underground cables provided that a sound cable runs along the faulty cable. Both these
tests employ the principle of Wheatstone bridge for fault location.
26
Detection of Fault Location in U/G cables
Murray Loop Test:
Figure sss shows the circuit diagram for locating the earth fault by Murray loop test.
Here AB is the sound cable and CD is the faulty cable; the earth fault occurs at point F.
The far end D of the faulty cable is joined to the far end B of the sound cable through
a low resistance link.
Two variable resistances P and Q
are joined to ends A and C
respectively and serve as the
ratio arms of the Wheatstone
bridge.
Let
Distance of fault point from test end is Figure 10: Equivalent circuit of
Murray Loop Test
28
Detection of Fault Location in U/G cables
Varley Loop Test:
The Varley loop test is also used to locate earth fault or short-circuit fault in U/G cables.
This test also employs Wheatstone bridge principle. It differs from Murray loop test in that
here the ratio arms P and Q are fixed resistances. Balance is obtained by adjusting the
variable resistance S
31
Detection of Fault Location in U/G cables
Tutorial on Loop Test:
Problem 2: In a test for a fault to earth on a 500 m length of cable having a resistance of
1 Ω per 1000 m, the faulty cable is looped with a sound cable of the same length but
having a resistance of 2·25 Ω per 1000 m. The resistance of the other two arms of the
testing network at balance are in the ratio 2·75 : 1. Calculate the distance of the fault from
the testing end of the cable.
32
Detection of Fault Location in U/G cables
Tutorial on Loop Test:
Problem 3: Varley loop test is performed to locate an earth fault on a 20 km long cable.
The resistance per km of the single conductor is 20 Ω. The loop is completed with a similar
healthy conductor. At balance, the variable resistance connected to the faulty conductor is
200 Ω. The fixed resistors have equal values. Calculate the distance of the fault from the
test end
33
References:
[1] Measurement & Instrumentation Principle by Alan S. Morris
34
W4
Lecture on
AC Bridge
Prepared By
Mehedi Azad Shawon
Lecturer, Dept. of EEE
American International University-Bangladesh (AIUB)
1
Lecture Outcomes
By the end of this lecture, the students should be able to:
2
AC Bridge
The AC bridge is a natural outgrowth of the Wheatstone Bridge. It consists of
four arms each having an impedance, a AC source of excitation and a balance
detector.
Bridges with AC excitation are used to measure the unknown parameters such
as Inductance, Capacitance, Quality factor, Dissipation Factor etc.
The AC bridge circuits are commonly used for the following applications:
Phase Shifting
Measuring the frequency of audio signal
Providing feedback path for the oscillator and amplifiers
Filtering out desirable signals etc.
3
AC Bridge
General Equation for a balanced AC Bridge
………. (1)
……. (2)
……. (3)
Substituting the values from (2) & (3) into (1) >
2nd condition ∠ 𝜽𝟏 + 𝜽𝟒 = ∠ 𝜽𝟐 + 𝜽𝟑
1) Maxwell’s Bridge
a) Maxwell’s Inductance Bridge (measurement of unknown L using standard L)
b) Maxwell’s Inductance-Capacitance Bridge (measurement of unknown L with low
Q-factor using standard C)
1) Hay’s Bridge (measurement of unknown L using high Q-factor using standard C & R)
7
Measurement of Self Inductance
1(a) Maxwell’s Inductance Bridge
Here,
Thus,
It can be seen from the above two equations that the values of L1 and R1 do not
depend on the supply frequency (𝜔). Thus, any kind of frequency variations or even
harmonic distortions in the power supply will not affect the measurement process.
10
Measurement of Self Inductance
1(b) Maxwell’s Inductance-Capacitance Bridge
Quality Factor or the Q-Factor of a lossy inductor is used to indicate how closely the
real inductor comes to behave as an ideal inductor. Q-factor is a measure of the
efficiency of the inductor
The Q-factor of an inductor is defined as the ratio of its inductive reactance to its
resistance at a given frequency.
The above relation for the inductor Q factor indicates that this bridge is not
suitable for measurement of inductor values with high Q factors, since in that
case, the required value of R4 for achieving balance becomes impracticably
high.
11
Measurement of Self Inductance
1(b) Maxwell’s Inductance-Capacitance Bridge
Advantages:
1. The balance equations are independent of each other, thus the two
variables C4 and R4 can be varied independently.
2. Final balance equations are independent of frequency.
3. The unknown quantities can be denoted by simple expressions involving
known quantities.
Disadvantages:
1. The bridge, for its operation, requires a standard variable capacitor, which
can be very expensive if high accuracies are asked for. In such a case, fixed
value capacitors are used and balance is achieved by varying R4 and R2.
2. This bridge is limited to measurement of low Q inductors (1< Q < 10).
3. Maxwell’s inductance-capacitance bridge is also unsuited for coils with very
low value of Q since this bridge finds difficult to obtain balance while
measuring such low Q inductors.
12
Measurement of Self Inductance
(2) Hay’s Bridge
13
Measurement of Self Inductance
(2) Hay’s Bridge Under balanced condition
14
Measurement of Self Inductance
(2) Hay’s Bridge By separating the real part
15
Measurement of Self Inductance
(2) Hay’s Bridge
The above expression of L1 & R1 contain the frequency term (𝜔). Therefore,
the supply frequency must be known accurately for using the Hay’s bridge.
Quality Factor:
The above equation signifies that this Hay’s bridge is suitable for measurement of
unknown inductance with high Q-factor where the resistance R4 does not need to
be impracticably high value as like Maxwell’s Bridge. 16
Measurement of Self Inductance
(2) Hay’s Bridge
Since the Hay’s Bridge is used for measurement of unknown inductance with high
Q-factor, the above equation can be simplified as follows:
17
Measurement of Self Inductance
(2) Hay’s Bridge
Advantages:
1. Gives simple expression of unknown inductance for Q-factor > 10.
Disadvantages:
1. Not suited for measurement of any inductor having Q-factor less than 10.
18
Quality Parameters of Real Capacitor
The quantifying parameters often used to describe performance of a capacitor are
ESR, its dissipation factor (DF), Quality Factor (Q-factor) and Loss Tangent (tan d)
It can thus be deduced that the Q can be expressed as the ratio of the capacitive
reactance to the ESR at the frequency of interest.
A high quality capacitor (high Q-factor) will thus have low values of dissipation
factor (DF) and loss tangent (tan δ), i.e. less losses.
20
Measurement of Capacitance
(3) Schering Bridge
Schering bridges are most popularly used these days in industries for measurement
of capacitance, dissipation factor, and loss angles
Dissipation Factor:
By separating the
imaginary part:
23
Measurement of Frequency
(3) Wien’s Bridge
In most bridges, the parameters are so chosen that,
R1 = R2 = R and C1 = C2 = C
25
Tutorials: AC Bridge
Problem 2:
26
Tutorials: AC Bridge
Problem 3: A Maxwell’s Inductance-Capacitance bridge is used for calculation of unknown
inductive impedance. The bridge parameter at balanced condition are as follows:
Pure resistance arms = 2.5 kΩ and 50 kΩ. In between these two resistors, the third arm
has a capacitor of value 0.012 μF in parallel with a resistor of value 235 kΩ. Find the series
equivalent of the unknown impedance.
27
Tutorials: AC Bridge
Problem 4: The four arms of a bridge are connected as follows:
Arm AB: A choke coil L1 with an equivalent series resistance r1
Arm BC: A noninductive resistance R3
Arm CD: A mica capacitor C4 in series a non-inductive resistance R4
Arm DA: A noninductive resistance R2
When the bridge is supplied from a source of 450 Hz is given between terminals A and
C and the detector is connected between nodes B and D, balance is obtained the following
conditions: R2 = 2400 Ω, R3 = 600 Ω, C4 = 0.3 μF and R4 = 55.4 Ω. Series resistance of the
capacitor is 0.5 Ω. Calculate the resistance and inductance of the choke coil.
28
Tutorials: AC Bridge
Problem 4: The four arms of a bridge are connected as follows:
Arm AB: A choke coil L1 with an equivalent series resistance r1
Arm BC: A noninductive resistance R3
Arm CD: A mica capacitor C4 in series a non-inductive resistance R4
Arm DA: A noninductive resistance R2
When the bridge is supplied from a source of 450 Hz is given between terminals A and
C and the detector is connected between nodes B and D, balance is obtained the following
conditions: R2 = 2400 Ω, R3 = 600 Ω, C4 = 0.3 μF and R4 = 55.4 Ω. Series resistance of the
capacitor is 0.5 Ω. Calculate the resistance and inductance of the choke coil.
29
Tutorials: AC Bridge
Problem 5: The four arms of a bridge are connected as follows:
AB: A capacitor C1 with an ESR r1, BC: A non-inductive resistance R3
CD: A noninductive resistance R4, DA: A capacitor C2 with an ESR r2 in series with a
resistance R2
A supply of 500 Hz is given between terminals A and C and the detector is connected
between nodes B and D. At balance, R2 = 5 Ω, R3 = 1000 Ω, R4 = 3000 Ω, C2 = 0.3 μF and r2 =
0.25 Ω. Calculate the values of C1 and r1, and also dissipation factor of the capacitor.
Solution:
30
References:
[1] Measurement & Instrumentation Principle by Alan S. Morris
31
Measurement And Instrument
Assistant Professor
American International University-Bangladesh
Spring 2015-16
• High voltage testing includes a large number of methods. Here, few methods will be considered.
• These tests are the most commonly used employing frequency of 50 Hz.
• These tests are used for routine voltage testing of motor, switch gear and other electrical equipment after
• These tests are also used in determination of dielectric constant and dielectric loss of specimens of insulation.
• Again, they are carried out on high voltage transformer, porcelain insulators and high voltage cables etc. for work
testing.
• Before high voltage overhead lines are energized, the insulation of every part connected with the lines must
• Therefore, the transmission lines and high voltage cables are subjected to a very high voltage for a
• If such voltage tests are carried out with an A.C. voltage supply, then high voltage transformer is required which
• Thus in order to overcome the difficulties encountered with high A.C. voltage supply, high voltage D.C. testing is
done.
• These high frequency disturbances are either due to switching operation or due to some external causes and
• It has been found that even though the porcelain insulator has satisfactorily withstand the sustained low
frequency high voltage tests, it breaks down at a much lower voltage at high frequencies owing to increased
• To increase reliable operation of insulators under disturbances caused by switching and arcing, they must be
• High frequency tests are also carried out for insulators used for communication purposes. Frequency ranging
from several kHz to a MHz or more are used for such purposes.
• The dangerous over voltages to which power systems are exposed to are caused by lightning.
• Lightning may produce an over voltage on an overhead line either by a direct stroke to the line or by an indirect
stroke.
• An indirect stroke occurs when a charged cloud above and near to the line, electrostatically induces charges of
• These charges are bound as long as the cloud remains near without discharging its electricity by lightning
stroke.
• But if the cloud is suddenly discharged, the induced charges in the line are no longer bound, but travel with the
velocity of light along the line to equalize the potential every where.
zero voltage.
• The danger to the electric installation comes from the fact that this over voltage acts across the insulation and
that the time taken by the voltage to rise to its peak value is very small.
• This imposes very severe voltage stresses on the insulators, transformer windings, circuit breakers, bushing
• If the voltage stress exceeds the strength of the insulation, a flash over or a puncture results causing a short
• If a power arc flows, there may be disastrous results in damage to the equipment whose repair may cost
• The following testing apparatus is used for sustained low frequency tests. This apparatus is usually common for
• High voltage testing transformers are single phase core type, oil immersed and self cooled, operated at the
same frequency as the normal working frequency at which the test specimens operate i.e. the power frequency.
• They may also be used for higher frequencies with rated voltage or for lower frequencies, if the voltages are
reduced depending upon the frequency in order to avoid saturation of the iron core.
• There is not much difference between a high voltage testing transformer and a single phase power transformer
• However, HV testing transformers are designed with a smaller value of the flux density in the core to avoid
• The testing transformers are also designed with compact and well insulated high voltage winding.
• Therefore, a single phase testing transformer may be compared with a potential transformer as regards its
construction.
• The primary winding is usually rated for rated voltages below 1 kV but might often be split up into two or more
winding sections which can be connected in series or parallel to increase voltage regulation capabilities.
• The iron core is earthed and so are the one terminal of each of the two primary and secondary windings.
• Figure (a) shows a grounded metal tank for which a HV bushing is necessary to bring the high voltage out of the
tank.
• In figure (b), the bushing is dispensed with and a co-axial cable is used for high voltage connection which results in
improved connection between testing transformer and the test object.
• In this arrangement, the active part of the transformer is placed in an isolating shell, thereby avoiding the use of
bushing and therefore reduces the height of the transformer, however, the heat dissipation is impaired with this
arrangement.
Measurement And Instrument, Spring 2015-16 Ahmed Mortuza Saleque -10-
High Voltage Measurement and Testing
High Voltage Transformers (contd.):
• In both the cases, the primary winding (LV winding) is close to the iron core and is surrounded by the secondary
• This arrangement reduces the leakage flux and increases the coupling between the two windings.
• The shape and cross section of HV winding is a reference to layout of this coil.
• The beginning (grounded end) of the HV winding is located at a side close to the core and the end close to a
sliced metal shield, which prevents too high field intensities at HV potential.
• Between both ends, the single turns are arranges in layers, therefore, form co-axial capacitors of high values.
• In order to keep the potential distribution uniform under transient conditions, the capacitance have to be made
equal. This is done by using reduced width of insulating layers with increasing diameters.
• The transformers used for insulation testing need not to be of a high kVA rating because the current taken by the
transformer is limited by inserting external resistances when the specimen under test breaks down.
• However, transformers used for cable tasting are designed to deliver large currents owing to capacitive (charging)
• Attention must be paid to regulation and cooling methods employed for such transformers.
Measurement And Instrument, Spring 2015-16 Ahmed Mortuza Saleque -11-
• Special constructional features are incorporated in high voltage transformers to eliminate corona effects as far
as possible.
• This high voltage terminal of the secondary winding and all other metal part which have a high potential with
respect to earth, must be constructed in such a way that sharp corners are avoided.
• For voltages above 500 kV, two transformers are then cascaded.
• A low voltage supply is given to the primary winding of transformer 1. The tank of this transformer is earthed.
• A lead is taken from the one end of the secondary winding and the other end is earthed.
• This lead is taken out of the transformer through a high voltage bushing which provides insulation for full
• The secondary winding is tapped at a point and another lead is taken out of the tank through the HV busing.
• One end of the secondary winding of transformer 2 is connected to the tank which is insulated from earth for a
• The other end of the secondary winding of transformer 2 is taken out with the help of a lead passing through a
• This lead forms the high voltage terminal of the transformer. The output voltage for test purpose is taken between
• This is equal to the sum of the secondary voltages of the two transformers. The output voltage can be increased
varied smoothly. Any abrupt changes in voltage will impose voltage surges on the specimen and will also affect the
accuracy of measurements.
• It is equally important that the voltage regulating device should not distort the waveform of output voltage.
• The method of voltage regulation depends upon whether a separate alternator is being used to supply the testing
• A separate alternator is usually used except in case of small testing establishments which carry out routine testing
work.
• When a separate alternator is used, the voltage regulation is carried out by variation of the alternator field current.
2. By means of a Variac
• Such instruments are essentially voltmeters, but they may be used with the help of external components to
1. One type involves two oppositely charged electrodes. One of them fixed and the other is movable. Due to
the force of attraction, the movable electrode is drawn towards the fixed one.
2. In the other type, there are forces of attraction or repulsion or both between the electrodes which cause
• In both cases, the mechanism resembles a variable capacitor and the force or torque is due to the fact that,
mechanism tend to move the moving electrode to such a position where the energy stored is maximum.
voltage.
• The stored energy is used as a basis for derivation of force and torque equations.
Linear Motion:
• In the above figure, there are two oppositely charged plates and one plate is fixed (A) and the other is movable
• Let, a potential difference of V volt is applied to the plates, then a force of attraction F exists between them
causing the movement of plate B towards plate A until this force is balanced by the spring. The capacitance
• Now, let there be a small increment dV in the applied voltage, then the plate B will move a small distance dx
towards A. When the voltage is being increased, a capacitive current flows and is given by,
⇒ . . + .
The electrical input energy is:
. . . . + .
Changed in stored energy,
1 1
+ + −
2 2
1 1
+ +2 + −
2 2
Neglecting higher order terms,
1
+
2
Or, . . + . + +
Rotational Motion:
For rotational motion, the Deflection Torque
= Angular Displacement
If the instrument is spring controlled or has a suspension then,
Controlling torque,
Where, K = Spring Constant
= Deflection angle
At steady state,
Since, the deflection is proportional to the square of voltage to be measured, the instrument can be used on both AC
and DC.
Measurement And Instrument, Spring 2015-16 Ahmed Mortuza Saleque -19-
• The breakdown voltage of a spark gap between two metal spheres may be used
• Two spheres are placed vertically one above, separated by an air gap.
• The lower sphere is earthed, the voltage between the spheres is raised till a
• The value of voltage required to spark over (break down) depends upon the
dielectric strength of the air, the size of spheres, the distance between the
• The method is simple and peak voltages may be measured from about 2 to 2500
• Sphere gaps may also be used for measurement of voltage in surge (impulse)
Magnetic Measurement
• Electrical and magnetic phenomena are closely associated with each other.
• The operating characteristics of electrical machines, apparatus and instruments are greatly influenced by the
properties of ferromagnetic materials used for their construction.
• Therefore, magnetic measurements are very important in designing and manufacturing electrical equipment.
• The principle requirements in magnetic measurements are as follows.
1. The measurement of magnetic field strength in air
2. The determination of B-H curve and hysteresis loop for soft ferromagnetic materials
3. The determination of eddy current and hysteresis loses of soft ferromagnetic materials subjected to
alternating magnetic fields
4. The testing of permanent magnets
• In magnetic measurements, the inaccuracies are due to the following reasons:
1. The conditions in the magnetic specimen under test are different from those assumed in
calculations
2. Magnetic materials are not homogeneous
3. There is no uniformity between different batches of test specimens even if such batches are of the
same composition
• Types of Tests:
1. Ballistic Tests
2. A.C. Testing
3. Steady State Tests
AC Testing:
• This test gives information about eddy current and hysteresis losses in magnetic materials.
Magnetic Measurement
Measurement of Flux Density (Ballistic Tests):
• The measurement of flux density inside a specimen can be done by winding a search coil over the specimen.
• The search coil is known as a “B-coil”.
• This search coil is then connected to a ballistic galvanometer or to a flux meter.
• A ring specimen wound with a magnetizing winding which carries a current I.
• A search coil is wound on the specimen and connected through a resistor and calibrating coil to a ballistic
galvanometer.
• The current through the magnetizing coil is reversed and
therefore the flux linkages of the search coil changing inducing
an EMF in it.
• This EMF sends a current through the ballistic galvanometer
causing it to deflect.
• Let,
Flux linking the search coil
R = Resistance of the ballistic galvanometer circuit
N = Number of turns in the search coil
t = Time taken to reverse the flux
Average EMF induced in the search coil,
2
!
And the charge,
2
"
!
Now,
Throw of the galvanometer
#$ Constant of the galvanometer
Therefore,
Charge indicated by ballistic galvanometer,
" #$
%&
or, #$
'
*+
( )
,
And the flux density,
./ )*+
-
01 2 03 ,45
Measurement And Instrument, Spring 2015-16 Ahmed Mortuza Saleque -25-
Magnetic Measurement
Determination of B-H curve:
Two methods are available for the determination of B-H curve
1. Method of Reversals
2. Step by step method
Step by Step Method:
Magnetic Measurement
Alternating Current Magnetic Testing:
• This testing is done in order to determine the iron losses in magnetic materials at different values of flux density
(B) and frequency (f) and to separate the two components of iron losses, i.e.
b) Hysteresis Losses
occurs due to hysteresis and eddy currents. This loss is called Iron or Core Loss.
• The hysteresis loss may be determined from the hysteresis loop obtained from
DC test.
• But this hysteresis loss may differ under actual working alternating magnetizing
conditions.
• The eddy current loss can be measured only under AC conditions.
• It is always convenient to measure the ambient hysteresis and eddy current losses i.e. total iron loss.
@>ABCB DE
B FB
9? watt
GH
Where,
'JK LMNO?
IC = Form Factor =
PL?QMR? SMNO?
t = Thickness of laminations, m
T = Resistivity of material, Ω m
Magnetic Measurement
Separation of Iron Loss (Contd.):
• Total iron loss per unit volume,
>
4IC < -=
9U 9: + 9? ;<-= +
3T
Hence, the total iron loss for a given specimen,
@>AB CB DE
B FB
9U > +
X./Y × ;<-=
GH
9U > +I I < -
I: <-= ? C =
• In this test, the form factor kf and the maximum flux density Bm are remain constant.
• Therefore,
9U I <+I <
Where, I IZ -=
I I? IC -=
[\
Or, I +I <
C
Magnetic Measurement
Problem: In a test on a specimen of total weight 13 kg the measured values of iron loss at a given value of peak flux
density were 17.2 W at 40 Hz and 28.9 W at 60 Hz. Estimate the values of hysteresis and eddy current losses in
w/kg at 50 Hz for the same value of peak flux density.
Solution:
Therefore,
9U 9: + 9? I <+I <
].
=> 40I + I (40)
G
@G
=> I + 40I (i)
Gaa
b.c
Similarly 60I + I (60)
G
bc
=> I + 60I (ii)
]baa
Illumination Measurement
Some Useful Definitions (Contd.):
Lumen: If a point source of one candela is placed at the center of a perfectly sphere of radius 1m, then the light
passing through one steradian or an area of 1m2 locating on the surface of the sphere is called a Lumen.
The symbol of Lumen is lm and luminous flux is φ
Luminous Intensity: It is the measure of luminous flux per unit solid angle from a point source.
Unit: lm/sr (The SI unit is candela)
Illumination or illuminance: The density of light energy incident on a surface is known as illumination or
illuminance.
h&
Illumination = is lm/m2 or lux
hP
• A photometric measurement system is a device that measures the intensity and sterance of a light beam.
• Any basic photometric measurement consists of a light source such as sunlight, LED or laser beam.
• The light is modulated with the help of light suppressor, light attenuator and filters before it is made incident
upon the detector. The detectors used are photoelectric transducers.
Optical Detectors:
Optical Sources:
1. Photo emissive cells (Gas filled or vacuum tubes)
1) Sun light : 460nm -max intensity ,UV-IR
2. Semiconductor photo-electric transducer
2) Incandescent lamp: 400nm-1100nm
Output:
3) Fluorescent sources ( sensitive to eyes)
a) Change in resistance ΔR
4) Light emitting diodes-forward biased
b) Change in current output ΔI
5) Lasers: semiconductors - GaAs
c) Change in voltage output ΔV
Illumination Measurement
Light Dependent Resistor (LDR):
• The conductivity of LDR is proportional to the intensity of light or the resistivity of LDR is inversely
proportional to the intensity of light incident upon it.
• The incident light energy releases charge carriers which increases flow of current produced by an
applied voltage.
• These semiconductors (LDR) are also called photoconductive cells.
• LDR is a slow device.
• Two most commonly used photoconductive semiconductor materials are cadmium sulphide (CdS) with a
band gap of 2.42 eV and cadmium selenide (Cd Se) with a band gap of 1.74 eV.
• Sensitivity: S = ΔR/ΔH Ω/Wm-2
ΔR = change in resistance and ΔH = change in radiation
Rt = Ri+(Rf - Ri)(1-exp(-t/τ) ), Rf =dark resistance and Ri =final resistance after application of beam.
Photo Diodes:
Illumination Measurement
Problem: The effective area of a photo diode is 0.2*10-6 m2 and the irradiance is 250W/m2. Calculate the incident
power. If the load resistance is 10kΩ and the capacitance of the diode is 2pf, find the cut-off frequency.
Solution:
Incident Power = 250*0.2*10-6 =50μW
Cut-off frequency = fc = 1/(2* π *RL*C) =1/(2* π *10*10000*2*10-12) = 8MHz.
[Photo transistors faster device (1μsec to 10μsec), Photo voltaic cell produces high voltage And Photo thyristor -
LASCR.]