You are on page 1of 5

Report No.

Special project 173


3130 FM 521 Fresno, Texas 77545, USA Phone: 281-431-2561 Fax: 281-431-1655 email: gmt.analytical.services@champ-tech.com

TO: Noel Brissard Date: August 6, 2008

FROM: Lijie Feng Cc: G. Cunningham


Nellie Chu W. Goss
Ming Xie L. Rodsjo
G. Cooper
J. Venter
A. Thompson

Residual Analysis of Cortron RN-400Q in Qatar Brine using HPLC/FLD


District: Corrosion and EHBS

EXECUTIVE SUMMARY

A LC/MS method was developed and qualified for determination of residual Cortron RN-400Q
in Qatar brine on December, 2007 (See Report AS20071926). An alternative method by
HPLC/FLD was requested by TS&D Corrosion and EHBS. The residual RN-400Q HPLC/FLD
method is successfully developed along with the use of a standard of RN-400Q, low TDS and
high TDS Qatar brines. A known concentration samples were prepared in the low and high TDS
brines respectively. The analysis results showed the samples in high TDS and low TDS brine
yielded 100±3 % recovery. The results are summarized in Table 1.

EXPERIMENTAL

The standards were prepared by dissolving RN-400Q standard in 5: 5 Acetonitrile:Water at 5.6,


11.2, 22.4 and 33.6 µg/mL, respectively. Both the standards and samples were analyzed using an
Agilent 1100 Series HPLC equipped with an Agilent 1200 Series FLD. Experimental
parameters and instrument conditions are summarized in Method 1 below.

Method 1. HPLC/FLD method for residual determination of Cortron RN-400Q in Qatar brine
Column: Zorbax Eclipse XDB C8, 5 µm, 4.6 x 150 mm
Mobile phase A: 0.1% Acetic acid in water
Mobile phase B: Acetonitrile
Gridient: T (min) %A %B
0.0 90 10
30.0 0 100
35.0 0 100
35.1 90 10

1
Report No. Special project 173
3130 FM 521 Fresno, Texas 77545, USA Phone: 281-431-2561 Fax: 281-431-1655 email: gmt.analytical.services@champ-tech.com

42.0 90 10
Flow rate: 1 ml/min
Column temperature: Ambient
Injection Volume: 50µL
FLD settings PMT 12, response time 4 sec, Excitation 248 nm, Emission 418 nm,
Emission after 14 minutes, 364 nm.
Diluent: 5:5 ACN: H2O

RESULTS AND DISCUSSION

The LC/FLD analysis results for RN400Q in Brine are summarized in table 1. The overlay
chromatogram of the standards and the corresponding calibration curves are shown in Figure 1
through Figure 4. The chromatograms of RN400Q in low TDS brine and high TDS brine are
shown in Figure 5 and Figure 6.

Table 1: LC/FLD Analysis Results


Peak Identification Experimental Recovery
in RN400Q Concentration (ppm)
222 ppm RN400Q in Peak 1 (RT = 11.5 min) 225 101.6%
low TDS Brine Peak 2 (RT = 13.0 min) 224 100.9%
Peak 3 (RT = 14.8 min) 223 100.5%
288 ppm RN400Q in Peak 1 (RT = 11.5 min) 281 97.6%
High TDS Brine Peak 2 (RT = 13.0 min) 282 97.9%
Peak 3 (RT = 14.8 min) 280 97.3%

2
Report No. Special project 173
3130 FM 521 Fresno, Texas 77545, USA Phone: 281-431-2561 Fax: 281-431-1655 email: gmt.analytical.services@champ-tech.com

Figure 1: Overlay Chromatogram of RN-400Q Calibration Standard


FLD1 A, Ex=248, Em=418, TT (RN400Q072908LJF CAL CURVE\RN400Q 2008-07-30 16-15-09\LJF3.D)
FLD1 A, Ex=248, Em=418, TT (RN400Q072908LJF CAL CURVE\RN400Q 2008-07-30 16-15-09\LJF4.D)
FLD1 A, Ex=248, Em=418, TT (RN400Q072908LJF CAL CURVE\RN400Q 2008-07-30 16-15-09\LJF5.D)
FLD1 A, Ex=248, Em=418, TT (RN400Q072908LJF CAL CURVE\RN400Q 2008-07-30 16-15-09\LJF6.D)
LU

Peak 1
80
Std1: 5.6 ug/ml
Std2: 11.2 ug/ml
70
Std3: 22.4 ug/ml

Peak 2
60 Std4: 33.6 ug/ml

50

40

30

20
Peak 3

10

0 5 10 15 20 25 30 min

Figure 2: Calibration Curve of Peak 1 (RT = 11.5 min)


P eak 1 , F L D 1 A
A rea = 2 3 .5 7 5 3 5 8 9 *A m t -4 .8 1 2 0 8 2 2
A re a R el. R es % (1 ): -2 .8 0 6 3 e-1
4
600
3
400
2
200 1

0 C orrelation : 0 .9 9 9 9 1
0 20 A m o u n t[n g /u l]

3
Report No. Special project 173
3130 FM 521 Fresno, Texas 77545, USA Phone: 281-431-2561 Fax: 281-431-1655 email: gmt.analytical.services@champ-tech.com

Figure 3: Calibration Curve of Peak 2 (RT = 13.0 min)


P eak 2 , F L D 1 A
A rea = 1 7 .1 6 5 6 7 5 8 *A m t -2 .5 4 9 5 6 5 3
A re a R el. R es % (1 ): 1 .2 8 0 7 e-1

500 4

400
3
300
2
200
1
100
0 C orrelation : 0 .9 9 9 9 5
0 20 A m o u n t[n g /u l]

Figure 4: Calibration Curve of Peak 3 (RT = 14.8 min)


P eak 3 , F L D 1 A
A rea = 3 .0 8 9 1 4 9 4 2 *A m t -0 .4 7 3 3 5 8 1
A re a R el. R es % (1 ): -1 .6 4 0
100
4
80
60 3

40 2
1
20

0 C orrelation : 0 .9 9 9 9 7
0 20 A m o u n t[n g /u l]

4
Report No. Special project 173
3130 FM 521 Fresno, Texas 77545, USA Phone: 281-431-2561 Fax: 281-431-1655 email: gmt.analytical.services@champ-tech.com

Figure 5: Chromatograms of the Prepared Samples in Low TDS Brine at 222 ppm (20
Times Dilution)
FLD1A,Ex=248,Em=418,TT(RN400Q072908LJFCALCURVE\RN400Q2008-07-3016-15-09\LJF7.D)
LU
25

Peak1
Peak2
20

15

10
Peak3

0
0 5 10 15 20 25 30 min

Figure 6: Chromatograms of the Prepared Samples in High TDS Brine at 288 ppm (20
Times Dilution)
FLD1A,Ex=248,Em=418,TT(RN400Q073108LJF\RN400Q2008-07-3117-15-30\LJF5.D)
LU
Peak1

30
Peak2

25

20

15

10
Peak3

0 5 10 15 20 25 30 min