You are on page 1of 6

FPD17-R1

Specifications

S517-E065
2007.02
Contents

1. Introduction··························································································································1
2. Features································································································································2
3. Precautions before installing the system···············································································2
4. Configuration ·······················································································································2
4.1 Standard Configuration ············································································································································· 2
4.2 Optional Accessories··················································································································································· 2
5. Specifications ·······················································································································2
5.1 Specifications ······························································································································································· 2
6. Environment·························································································································3
6.1 Temperature & humidity ···································································································································· 3
6.2 System calorific value ··········································································································································· 3
7. Power supply························································································································3
7.1 Power unit ································································································································································ 3
8. Dimensions ··························································································································4
9. Remote Maintenance············································································································4
9.1 Functions ································································································································································· 4
S517-1081

1. Introduction
FPD17-R1 is a flat panel X-ray detector unit, which is composed of Flat Panel Detector
(FPD) with direct conversion method and the FPD control PSU rack
FPD17-R1 is embedded in a general radiography system to acquire images on Chest Stand
and Table radiography. FPD is built in Radiography Stand and/or FPD Radiography table.
FPD converts incident X-ray to electronic signal directly, converts electronic signal to
digital image, and transfers digital image to digital imaging console.

[1]PSU Rack
[1]PSUラック

[2][2]フラットパネルセンサ
Flat panel sensor

[3] Cable
[3]ケーブル

Option
オプション
[4] Flat panel sensor unit
[4]フラットパネルセンサユニット

Flat panel sensor


フラットパネルセンサ
Power supply unit
電源ユニット

Cable
ケーブル

-1-
S517-1081

2. Features
FPD17-R1 offers the following features.
- X-Ray detector with a direct conversion method of high spatial frequency, it makes possible to acquire
the high-quality image.
- It allows adding another FPD (Optional feature), and it needs no additional PSU rack.

3. Precautions before installing the system


Prior to installing FPD17-R1, please read the following precautions such as Power Supply system, Air
conditioning system.
1. Power Supply system
This system includes the device which is required 24 hours non stop power supply.
Follow the Power conditions referring to “7. Power Supply”.

2. Air conditioning system


The temperature and humidity in the exam room must be controlled properly for this system.
Follow the proper environmental conditions referring to “6.Environmental Conditions”.

4. Configuration
4.1 Standard Configuration
1. Flat Panel Detector 1 set
2. PSU rack 1 set
3. cables 1 set
4.2 Optional Accessories
Additional Flat Panel Detector unit 1 set

5. Specifications
5.1 Specifications
(1) FPD

Item Contents
X-ray transform method Direct-conversion method
Conversion material a-Se
Effective field of view size (side/diagonal) 43.2cm / 61cm
Pixel pitch 150μm
Number of pixels 2,880 x 2,880 =8,294,400 pixels
Limit spatial frequency 3.3 lp/mm
Gradation 16bit (65,534 step)
Cooling system Water circulation cooling system
Dimensions of detector (W) x (H) x (D) 530 x 530 x 70 mm
Weight Approx 20kg

(2) PSU

Item Contents
Dimensions (W) x (H) x (D) 900 x 1750 x 400 mm
Weight Approx 265kg

-2-
S517-1081

6. Environment
6.1 Temperature & humidity
Temperature 20 to 27℃ Continuous air-conditioning for 24 hrs
Humidity 15 to 75% (Non condensing)
-The air-conditioning system is required in the examination room to meet the environmental
condition by the continuous air-conditioning for 24hrs.
6.2 System calorific value
Standard (1-FPD) 500Kcal/hr Continuous air-conditioning for 24 hrs
Optional (2-FPDs) 700Kcal/hr Continuous air-conditioning for 24 hrs

7. Power supply
7.1 Power unit

Rated supply power Voltage


Note
(kVA) (V, Hz)
24 hours non stop
4 Single phase 200V power supply

Single phase
200V
4KVA or more

Breaker

PSU
24 hours
non stop
power supply

-3-
S517-1081

8. Dimensions

Dimensions
Item Description
(unit: mm)
FPD main unit L530×W530×H70 Installed in BR-120F/FT
FPD main unit L530×W530×H70 Installed in BK-200F
PSU L400×W900×H1750 FPD power control unit

9. Remote Maintenance
FPD17-R1 is equipped with Remote Maintenance as standard implementation.
Remote Maintenance allows the followings;
- Daily check gets more efficient.
- Preventions and early detection of errors.
- System down time can be reduced when an error occurs.
The maintenance contract is required to use Remote Maintenance feature.

9.1 Functions

Maintenance contents
Devices Functions Descriptions
FPD main unit Correction data check Check each correction data exists on FPD control PC in
order to prevent errors.
FPD temperature Analyze FPD temperature log file.
monitor
Temperature and Monitor the temperature and humidity at the exam room.
Humidity monitor

-4-