TSU-E03015
2003.JULY.3
1. P B S . P C B
SW1: 1 This setting is meaning for REV of BD ASSY.
2 This setting is not regard as machine’s working.
3
4
5
6 Set the DIP SW setting like next.
7 SDU-1200,(-01) 5:OFF,6:OFF,7:OFF,8:ON
8 SDU-2200,(-02) 5:OFF,6:OFF,7:ON,8:OFF
2. D B F . P C B
SW1: 1 This setting is meaning for REV of BD ASSY.
2 This setting is not regard as machine’s working.
3
4
5
6 This setting is meaning for DASH NO. of BD ASSY.
7 0000→NO DASH No., 0001→-01, 0010→-02,0011→-03(0:OFF,1:ON)
8 -01,-02,-03 is meaning for DBF20-01,-02,-03 each other.
3. V E C . P C B
SW1: 1
2 You should set to be all OFF.
3
4
5
6
7 7:This setting is always ON.
8 8:This setting is always ON.
5. D O P / C F M . P C B
SW1: 1
This setting is meaning for REV of BD ASSY.
2
This setting is not regard as machine’s working.
3
4
5
This setting is meaning for DASH NO. of BD ASSY.
6
5,6,7 is 000→ NO DASH.No,001→-01,010→-02(0 is OFF,1 is ON)
7
8
SW2: 1
This setting is meaning for REV of BD ASSY.
2
This setting is not regard as machine’s working.
3
4
5
Set the setting all of OFF.
6
7
8
SW3: 1
1:OFF,2:OFF,3:ON,4:ON,5:OFF,6:OFF,7:ON,8:ON
2
Confirm these setting should be done.
3
4
5
6
7
8
SW4: 1
Set the setting to 1:OFF,2:OFF
2
7. C P B . P C B