You are on page 1of 4

MDL Manufactur Primary_Int Primary_Int PM Level

System ID Vityl CMOD Model intrtp Order


Number er erval_Value erval_Type Int

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 740 2555 days 3


cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 750 2555 days 3
cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 760 2555 days 3
cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 770 2555 days 3

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 780 2555 days 3


cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 790 2555 days 3
cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 800 2555 days 3
cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 810 2555 days 3

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 820 2555 days 3

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 830 2555 days 3

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 840 2555 days 3

cc893761dCMOD0006MDL-BFF1BSensorMedi3100B 7 850 2555 days 3


Secondary_I Secondary_I
PM Level
nterval_Val nterval_Typ
String
ue e

PM Level 3
PM Level 3
PM Level 3
PM Level 3

PM Level 3
PM Level 3
PM Level 3
PM Level 3

PM Level 3

PM Level 3

PM Level 3

PM Level 3
Question Name

Driver controller adjustments – Pulse Width Modulator (PWM) pots and switches: (Press Start/Stop.) Check that the LED on th
Driver controller adjustments – Pulse Width Modulator (PWM) pots and switches: (Press Start/Stop) Check that the LED on the
Driver controller adjustments – Pulse Width Modulator (PWM) pots and switches: (Press Start/Stop.) Check that the LED on PW
Final test – Ensure that the voltage drops by 2.0 volts from the starting voltage (approximately 7V) to verify the function of the

Final test – Check that the oscillator stops and the Oscillator Overheat and Oscillator Stopped LEDs illuminate after △P drops b

Final test – Check that, when the ground is removed, the oscillator automatically starts running again.
Final test – Make sure the power leads to the Elapsed Time Meter and the alarm beepers are connected and that the meter is
Patient circuit calibration procedure – Check that the reading is 39-43 cmH2O.

Performance checks – (0-2000 ft; 26-44 cmH2O) Check the following parameters, using the appropriate altitude range and ver

Performance checks – (2000-4000 ft; 26-44 cmH2O) Check the following parameters, using the appropriate altitude range and

Performance checks – (4000-6000 ft; 26-44 cmH2O) Check the following parameters, using the appropriate altitude range and

Performance checks – (6000-8000 ft; 26-44 cmH2O) Check the following parameters, using the appropriate altitude range and
Question
Expected
Type Lower Upper
Answer Range UoM
(Pass/Fail, Range Range Checklist
(Pass/Fail, (Range Comment
Free Text, (Range (Range Name
Yes/No only)
Yes/No, only) only)
only)
Range)

Pass/Fail Pass SensorMedics Corp 3100B Annual 7


Pass/Fail Pass SensorMedics Corp 3100B Annual 7
Pass/Fail Pass SensorMedics Corp 3100B Annual 7
Pass/Fail Pass SensorMedics Corp 3100B Annual 7

Pass/Fail Pass

SensorMedics Corp 3100B Annual 7


Pass/Fail Pass SensorMedics Corp 3100B Annual 7
Pass/Fail Pass SensorMedics Corp 3100B Annual 7
Range 39 43 cmH2O SensorMedics Corp 3100B Annual 7

Range

113 145 cmH2O SensorMedics Corp 3100B Annual 7

Range

104 135 cmH2O SensorMedics Corp 3100B Annual 7

Range

95 125 cmH2O SensorMedics Corp 3100B Annual 7

Range

90 120 cmH2O SensorMedics Corp 3100B Annual 7

You might also like