Unicorn lyte

(Na⁺, K⁺, Cl⁻, Ca²⁺, Ph)


Unicorn lyte

• Unicorn lyte 3 (Na⁺, K⁺, Cl⁻)

• Unicorn lyte 5 (Na⁺, K⁺, Cl⁻,


Ca²⁺, Ph)
Unicorn lyte
 SPECIFICATIONS :
 Method : Multi-Biosensor Technology
 Sample Volume : 120-150 µl
 Througput : 80 Test/Hour
 Sample type : Serum, Urine, Whole blood
 Display : 4 line LCD with backlight
 Printer : Inbuilt thermal printer
 3 RS 232 port : For upgrade software, for barcode, for data
output
 Working temperature : 15-30°C
Unicorn lyte
 Power requirement : 1 KVA Online UPS with isolated
transformer, Earth voltage ≤ 5 Volt
 Reagent pack : 1270 ml
 Standard A : 530 ml
 Standard B : 210 ml
 Flush Solution : 220 ml
 Reference solution : 310 ml
Unicorn lyte

The Unicorn lyte analyzer is required to be turn on all the


times.

If you do not use analyzer for a long time, you need to


remove reagent pack and then perform priming with DI
water and drain the all liquid from tubes.
Unicorn lyte
How to use MBM correctly ? Please operate
according to following requirements.
 Store the MBM card at 2 – 30⁰C.
 If you take the MBM from refrigerator then
restore it to room temperature before install
in instrument. During this period do not tear
the aluminium foil bag.(about 25 minutes).
Unicorn lyte
 Installation
 Install reagent pack

Make sure that reagent pack


connected properly
Unicorn lyte
 Install pump tube

Install pump tube on plunger and ensure it should not be


overstretched or overlapped
Unicorn lyte
 Install MBM

MBM

Push the tongue of MBM box Install new MBM in MBM box
and open it
Unicorn lyte
 Install MBM

Insert it properly in box Push it gently to close and


lock
Unicorn lyte
 Switch on analyzer
 After switch on the instrument, analyzer shows READY screen

READY
Not Cal’ V23.1.40
27-Sep-2019
13:07:53

Press NO key to enter


in Menu
Unicorn lyte
1 PRINT 1 CALIBRATION After Reset The
2 SAMPLE/QC 2 PRIMING Electrode Please
close the door
3 OPERATION 3 CHANGE MBM And Press YES
4 MAINTENANCE > 4 CHANGE PACK >

PRESS 3, THE PRESS 2, FOR PRIMING PRESS YES, So


ANALYZER WILL BE AND REPEAT SAME FOR ANALYZER WILL
ENTERED INTO 3 TIMES AND CHECK ALL TAKE ACTIVATION
‘OPERATION’ MENU 4 TUBES SHOLD BE AND CLAIBRATION
FILLED WITH LIQUID. PROCESS
THEN PRESS 3, CHENGE AUTOMATICALLY
MBM DISPLAY WILL
SHOW :
Unicorn lyte
 After calibration analyzer will show calibration slope on display

K: 62.2
Na: 61.5
Cl: -67.4
OK

 The slope range is 35 -80. After calibration now it is ready for


sampling
Unicorn lyte
 RFID Details

After completion Reagent pack and MBM installation check


and take picture of RFID details
Unicorn lyte
 Sampling :
 After the calibration of the analyzer it is in the status of
“READY”, lift the sample door, the prompt, “Probe in Serum?”
will be displayed, press YES and pump will begin to aspirate.
Unicorn lyte
 After beep sound take out the sample and clean probe with
lint-free tissue and close the sample door
Unicorn lyte
 After closing the sample door a brief countdown will begin.
“Analyzing…” will be displayed. Upon completion of analysis,
the test results will displayed and printed

Analyzing… ID: 23 Serum


35 K: 2.17 mmol/L
Na: 118.7 mmol/L
ID: 23 Cl: - 79.6 mmol/L

Note : Probe should not touch the bottom surface of sample tube
Unicorn lyte
 Cleaning Procedure : (after 300 samples or 30 days
whichever is earlier

Press 4 Press > key


Unicorn lyte

Put EZ cleaner and It will take 10 minutes


Press 5 press YES to complete
Unicorn lyte
 CALIBRATION PROCEDURE :

In READY mode press


NO key for menu

Press 4 Press 1 and enter Press 1


service code 666
Unicorn lyte

Input target Press 4, auto


value from Chart Press YES
factors
and press YES,
Then go in
Maintenance →
Auto Factors
Unicorn lyte

Press YES Enter 1569 in


Press YES
service code
Unicorn lyte

Lift probe and aspirate


Press YES Press YES QCB-1. Every level QCB
should be tested three
times. New factors will
be saved automatically
Unicorn lyte

To check new factors


in Manual mode
Unicorn lyte

THANK YOU