BS-200 Endpoint
BS-200 Endpoint
• Preface
• Basic parameter setting
• Reagent setting
• Calibration and QC setting
Preface
• Here we set ALB as the example for all
Endpoint items on
No. will be generated automatically by the system, eg 1 for the first one.
ALB
1
Full Name of the item, can be blank if the user don’t want to input anything.
ALB
1
Standard No. of the item, can be blank or involved with the bar code.
ALB
1
Reac. Type of the item, can be referred from the reagent insert as ...
Basic parameter for ALB
Endpoint
Endpoint
Pri. Wave & Sec. Wave should be selected as reagent insert indicates …
Basic parameter for ALB
Endpoint
578
Endpoint
578
Endpoint
578
Ascending
Endpoint
578
Ascending
Reac. Time & Incuba. Time for the reaction, refer to ...
Basic parameter for ALB
R1 Sample
minute cycle
0.5 2
1 4
2 8
3 12
4 16
5 20
… …
Basic parameter for ALB
Endpoint
578
Ascending
0 20
Reac. Time is set as 0 and 20, no Incuba. Time for mono reagent items.
ALB
1
Endpoint
578
Ascending
0 20
Endpoint
578
Ascending
0 20
g/l
Endpoint
578
Ascending
0 20
g/l
Endpoint
578
Ascending
0 20
g/l
Endpoint
578
Ascending
0 20
g/l
Endpoint
578
Ascending
0 20
g/l
Endpoint
578
Ascending
0 20
g/l
0
Absorbance of R1 Blank (mono rgt.) and Mixed Rgt. Blank (dual rgt.),
set according to experience or reagent insert; if no idea, can be blank.
ALB 300
1
3
Endpoint
578
Ascending
0 20
g/l
Endpoint
578 7
Ascending
0 20
g/l
Endpoint
578 7
Ascending
0 20
g/l
0
Linearity Limit, Substrate Limit and Factor should be set for the
Kinetics tests, so here we can leave them as blank.
ALB 300
1
3
Endpoint
578 7
Ascending
0 20
g/l
Prozone check is set for the turbidimetric tests, which can also be blank.
Reagent parameter for ALB