The document summarizes the results of a system accuracy test (SAT) conducted on furnace GRIEVE with ID # CAP-6342. The test was performed according to specification AMS2750 using test sensors CAP-6570 and FLUKE 724 with instrument CAP-6950. The SAT results show that the indicator and recorder readings were within the allowed ±4°F tolerance compared to the test instrument readings, so the furnace passed the SAT.
The document summarizes the results of a system accuracy test (SAT) conducted on furnace GRIEVE with ID # CAP-6342. The test was performed according to specification AMS2750 using test sensors CAP-6570 and FLUKE 724 with instrument CAP-6950. The SAT results show that the indicator and recorder readings were within the allowed ±4°F tolerance compared to the test instrument readings, so the furnace passed the SAT.
The document summarizes the results of a system accuracy test (SAT) conducted on furnace GRIEVE with ID # CAP-6342. The test was performed according to specification AMS2750 using test sensors CAP-6570 and FLUKE 724 with instrument CAP-6950. The SAT results show that the indicator and recorder readings were within the allowed ±4°F tolerance compared to the test instrument readings, so the furnace passed the SAT.
FURNACE ID # CAP-6342 TEST SENSOR: CAP-6570 CLASS: 3 TEST INSTRUMENT: FLUKE 724 CAP-6950 INSTRUMENTATION TYPE: TYPE C PROCEDURE: QAP-2750-001
SYSTEM ACCURACY TEST (SAT) ZONA DE CONTROL DATOS DE TEMPERATURAS
INDICATOR RECORDER INDICATOR RECORDER
DATE 10/14/17 10/14/17 TIME 11:30:00 AM 11:30:00 AM SET TEMPERATURE 400 400 SETPT CONTROLADOR SETPT CONTROLADOR INSTRUMENT INDICATED TEMPERATURE A 399.80 399.70 TEMP CONTROLADOR TEMP TERMOM GRAFIC CORRECTION FACTOR #1 B 0 0 CORRECC. CALIBRACION CORRECC. CALIBRACION CORRECTION FACTOR #2 B1 0 0 CORRECTION FACTOR #3 B2 0 0 A + B + B1 + B2 = C C 399.80 399.70 TEST INSTRUMENT READING D 400.90 400.90 TEMP DE CALIBRATOR TEMP DE CALIBRATOR TEST T/C CORRECTION FACTOR D-C E 1.10 1.20 TEST INSTRUMENT CORRECTION FACTOR F 0 0 CORRECC. CALIBRATOR CORRECC. CALIBRATOR TRUE TEMPERATURE D + E + F = G G 402.00 402.10 SAT DIFFERENCE C - G -2.2 -2.4 4 F 4 F PASS / FAIL PASS PASS PASS / FAIL PASS / FAIL TECHNICIAN: A. PORRAS A. PORRAS TECHNICIAN TECHNICIAN