Sample Heater/Cooler Temperature Accuracy

A C Q U I S I TI O N I N F O R M A TI O N

System Name: 2695_2489_ISM_QDa72 Acquired By: ryan


Processing Method: Temperature Accuracy Test PM Project Name: SQT_E_2695_2489_QDa_P_ISM_RC2
Result Set Id: 5404 Sample Kit 1 Lot Number: W12351314
Sample Set Id: 5394 Sample Kit 1 Exp. Date: 12/23/2014 12:00:00 AM
Sample Set Name: SHC Temperature Acc SSM Column Serial Number: W5123541
Sample Set Start Date: 12/23/2014 2:38:08 PM EST
TE S T D E S C R I P TI O N

Temperature Accuracy is determined by measuring the temperature within the instrument at a specific location
(the sensor) and comparing it against the set-point programmed on the instrument. The test requires a calibrated
temperature measuring device that is accurate to ± 1.0 C. Up to two temperature set-points may be tested.

The Absolute Temperature Inaccuracy should be ± 4.0 C of set point temperature.

Sample Heater/Cooler Temperature Accuracy Results


Temperature Absolute Temperature
Set Temperature
Sample Name Accuracy Serial Number Temperature Accuracy
Temperature Observed
Specification Inaccuracy Pass/Fail
1 5.0 SHC Temperature 1 Set Point +/- 4.0 C 1234567 5.0 0.0 PASS
2 10.0 SHC Temperature 2 Set Point +/- 4.0 C 1234567 11.0 1.0 PASS

Printed 11:30:22 AM US/Eastern 12/29/2014 Page: 1 of 1


Waters Project Name and Version: SQT_E_2695_2489_QDa_P_ISM Version 1.00